OMEGAs sanitary RTD sensors are supplied standard with 100 , Class A
sensing elements that conform to the International IEC 60751 Standard.
This standard specifies the resistance vs. temperature relationship, the
Accuracy Classes, and the color code used. This information is included
below to help in the application and use of these sanitary sensors.
For the C
RTD Tempe omplete
ra
Resistance ture vs.
Visit omeg Table
a.co
rtdtable m/
Equations
Platinum RTD resistance can be calculated using the Callendar-Van Dusen Equation as follows:
For temperatures below 0C:
For temperatures above 0C, this simplifies to:
Rt = R0 [1 + At + Bt2 + C(t-100)t3]
Rt = R0 (1 + At + Bt2)
where: A = 3.9083 x10-3 (C-1)
B = -5.775 x10-7 (C-2)
C = -4.183 x10-12 (C-4)
R0 = Resistance at 0C
t = Temperature in degrees celsius
Resistance vs. Temperature Values per IEC 60751
Temp
Resistance
Temp
(C)
()
(C)
-200
-150
-50
0
50
100
18.52
39.72
80.31
100.00
119.40
138.50
Resistance
()
Temp
(C)
Resistance
()
157.33
175.86
194.10
212.05
229.72
247.09
450
500
550
600
650
700
264.18
280.98
297.49
313.71
329.64
345.28
150
200
250
300
350
400
Accuracy Classes
There are four Accuracy Classes specified in IEC 60751, listed from least accurate to most accurate they are:
(Note: Accuracy Values are in Degrees C)
Class C = (0.60 + 0.01* t)
(-50 to 500C)
Class B = (0.30 + 0.005* t)
(-50 to 500C)
OMEGAs
RTDs ComeSanitary
Class A = (0.15 + 0.002* t)
(-30 to 300C)
Stan
with Class dard
Accuracie A
Class 13 DIN = (0.10 + 0.0017* t)
(0 to 150C)
s
where t = temperature of interest in C without regard to sign
( also known as Class AA)