Professional Documents
Culture Documents
Bmi Unit 2 - Opt
Bmi Unit 2 - Opt
com
www.eeecube.com
www.eeecube.com
An Overview . . .
Basic components of a biomedical system
Electrodes
Amplifiers
ECG
EEG
EMG
ERG
www.eeecube.com
www.eeecube.com
o
Biological
System
Transducer
Signal
Processor
Measurement
Monitoring
o
o
Diagnosis
Surgery
Feedback
Transform
Surgical
Tool
www.eeecube.com
Therapy
www.eeecube.com
www.eeecube.com
Electrodes
Employed to pick up the electrical signals of the body
Microelectrodes
Depth and Needle Electrodes
Surface Electrodes
www.eeecube.com
www.eeecube.com
Microelectrodes
Have smaller diameter
No damage to the cells during insertion
To measure the potential they, located within the cell
Reference electrode is outside the cell
Types
Metallic
Non Metallic (Micropipet)
www.eeecube.com
www.eeecube.com
Metal Microelectrode
RA
o
RS
RB
CD
CWA
CWB
RFA
RFB
RWA
RWB
EA
EB
RIN
www.eeecube.com
EC( t)
REX
www.eeecube.com
www.eeecube.com
www.eeecube.com
Equivalent Circuit
RFA
o A
CWA
EC
RWA
CD
o B
E = EA - EB
www.eeecube.com
www.eeecube.com
Since the area of the reference electrode is many times greater than
the metal electrodes tip whose area of cross section is very small, its
impedance is negligible
The impedance of the microelectrode tip is inversely proportional to
the area of the tip and frequency
When the electrode output is coupled with an amplifier, the low
www.eeecube.com
www.eeecube.com
Micropipet
RFA
ED
RT
EA
RA
RWA
RIN
EC
CWA
CD
RFB
REX
EB
www.eeecube.com
RWB
CWB
RB
www.eeecube.com
fluid
www.eeecube.com
Equivalent Circuit
RT
o A
EC
CD
o B
E
www.eeecube.com
www.eeecube.com
www.eeecube.com
www.eeecube.com
Rf
V1
V2
Ri
o
e1
o
Ri
e2
Vo
Rf
o
www.eeecube.com
www.eeecube.com
Since the input current to an ideal amplifier is zero and by the principle
of superposition, the voltage existing at the input terminal 1 is,
e1
Rf
Ri + Rf
V1
Ri
Rf + R i
Vo
Rf
Rf + R i
V2
www.eeecube.com
Therefore,
Rf
Ri + R f
(or)
V1
Ri
Rf + Ri
Vo
Vo
Rf
Ri
Rf
Rf + Ri
V2
(V2 - V1)
Thus the circuit amplifies the difference of the input signals V1 and V2
There is no virtual ground at the input to the amplifier in this circuit
www.eeecube.com
www.eeecube.com
opposite polarity
Common mode the input signals are identical both in amplitude
and phase
www.eeecube.com
www.eeecube.com
Isolation Amplifier
Used to increase the input impedance of the monitoring system in
order to isolate the patient from the bio medical instrument
Used so that any electrical fault in signal conditioning or other
circuits does affect the patient
Some isolation amplifier circuits . . .
Darlington Pair
Bootstraping Circuit
www.eeecube.com
www.eeecube.com
Darlington Pair
o - VBB
o - VCC
RB
RL
.Y
Q1
Zi
Q2
www.eeecube.com
www.eeecube.com
Darlington Pair
High input impedance with high current gain
Two transistors Q1 and Q2 are connected in common emitter
compound connecting circuits
The emitter terminal of transistor Q1 is directly connected to the
base of transistor Q2
The collectors of both transistors share a common load RL
Q2 provides negative feedback that serves to gain stabilize the
composite circuit
The input impedance is very much higher greater than that of a
single common emitter stage
www.eeecube.com
www.eeecube.com
Bootstraping Circuit
+ VBB
R1
C
Q1
o
R4
Q2
o
INPUT
R3
www.eeecube.com
Re
R2
Re
OUTPUT
www.eeecube.com
Bootstraping Circuit
A feedback network is connected in between the emitter of Q2 and
the collector of Q1
The feedback voltage created b the bootstrap voltage dividing
network R1 and R2 is injected into the collector circuit of Q1
An increase in signal level at the input of the circuit causes an
increase in signal through the divider at the collector, changing the
transistors bias point
This further increases the input impedance
www.eeecube.com
www.eeecube.com
Guard
Transformer
Modulator
.
D.C.
Amp
Driver
Synchronous
Modulator
Synchronous
Demodulator
Lead Selector
+V
RA
LA
LL
RL
www.eeecube.com
Electrosurgery
Filter
Floating
Common
Rect.
&
Filter
100 kHz
Oscillator
-V
Isolated Power
Transformer
www.eeecube.com
The signals from the different leads are given to the low pass filter having a
cut off frequency of about 10 kHz
This is followed by high voltage and over voltage protection circuits
The signals are now into the lead selector switch which is used to derive the
required lead configurations
Its output is given to a d.c. amplifier ( also used for calibration purposes )
The primary of an isolated low capacitance power transformer is connected
www.eeecube.com
www.eeecube.com
aR1
R2
V1
o
bR2
Vo
R1
-
aR1
o
Vout
R2
V2
www.eeecube.com
Vo
bR2
1
www.eeecube.com
www.eeecube.com
Chopper Amplifier
Used to convert the d.c. or low frequency signal into a high
frequency signal
Then this modulated high frequency signal is amplified by
Types
Mechanical Chopper Amplifier
Non Mechanical Chopper Amplifier
www.eeecube.com
www.eeecube.com
P
o
M
o
A
C
o
Vi
o
Q
www.eeecube.com
S1
Vo
o
N
www.eeecube.com
o
Low Level
DC Input
Demodulator
AC
Amp.
PC1
Low Pass
Filter
Amplified
DC Output
PC4
Photo Diodes
PC2
PC3
Neon Bulb
(1)
www.eeecube.com
Oscillator
Neon Bulb
(2)
www.eeecube.com
www.eeecube.com
Whenever light falls on PC1 its resistance decreases and input capacitor
charges
When there is no light on PC1 and there is light on PC3, the input flows
through PC3
By the alternate incident light, a square wave is produced across the
capacitor
Its amplitude is proportional to the input and frequency is equal to the
Electrocardiography
www.eeecube.com
ECG deals with the study of the electrical activity of the heart muscles
right ventricle
The right ventricle pumps the blood into the lungs where it is purified
and oxygenated
The oxygen enriched blood enters the left atrium from which it is
pumped into the left ventricle
Then the left ventricle pumps the blood into arteries through aortic
valve for circulation throughout the body
www.eeecube.com
www.eeecube.com
Each action potential in the heart originates at the Sino - Atrial (SA)
node which is situated in the wall of the right atrium and near the
entry of the vena cava
Also called as Cardiac Pacemaker and generates impulses at the
normal rate of the heart, about 70 beats per minute at rest
The action potential contracts the atrial muscle and the impulse
spreads through the atrial wall during a period of about 0.04 second
to the Atrio Ventricular (AV) node
The AV node acts as a delay line to provide timing between the
www.eeecube.com
www.eeecube.com
www.eeecube.com
P Wave
Amplitude
Duration
(mV)
(seconds)
0.25
0.12 to 0.22
(P R interval)
R Wave
(QRS
1.60
0.07 to 0.1
0.1 to 0.5
0.05 to 0.15
complex)
T Wave
Ventricular repolarisation
(Relaxation of myocardium)
ST
(S.T interval)
Ventricular contraction
interval
U Wave
< 0.1
0.2
(T U interval)
www.eeecube.com
and electrodes
The potentials generated in the heart are conducted to the body surface
The potential distribution changes I a regular an complex manner during
each cardiac cycle
So standardised electrode positions are chosen
Types of electrode systems
Bipolar Limb Leads (or) Standard Leads
Augmented Unipolar Limb Leads
Chest Leads (or) Precordial Leads
Frank Lead System (or) Corrected Orthogonal Leads
www.eeecube.com
www.eeecube.com
Right Leg
Left Leg
The right leg electrode is used as ground reference electrode
www.eeecube.com
www.eeecube.com
Lead I
Lead II
Vout
www.eeecube.com
Vout
www.eeecube.com
Lead III
+
Vout
www.eeecube.com
www.eeecube.com
Lead I Position gives voltage drop from the left arm to the right arm
Lead II Position - gives voltage drop from the left leg to the right arm
Lead III Position gives voltage drop from the left leg to the left arm
www.eeecube.com
www.eeecube.com
Lead I
Right Arm
+
Left Arm
Lead III
Lead II
+
+
Left Leg
www.eeecube.com
www.eeecube.com
Along the sides of this triangle the three projections of ECG vector
are measured
Further the vector sum of the projection on all the three sides is
equal to zero
Thus, following Kirchoffs law, the R wave amplitude of Lead II is
equal to the sum of the R wave amplitudes of leads I and III
www.eeecube.com
www.eeecube.com
Lead aVR
Lead aVL
+
-
Vout
www.eeecube.com
Vout
www.eeecube.com
Lead aVF
+
Vout
www.eeecube.com
www.eeecube.com
the body
Two equal and large resistors are connected to a pair of limb electrodes
The center of this resistive network acts as central terminal and the
1 mV
Calibration
o
o
1V
www.eeecube.com
1M
1k
Lead
Selector
Defibrillator
Protection
Circuit
Output
Unit
aVR
aVL
1
OFF
Bioamp
aVF
C
CM Reduction
Amplifier
o
a.c.
Power
o
www.eeecube.com
Isolation
Transformer
Isolated
Power
Supply
Power
Supply
www.eeecube.com
Electroencephalography
Deals with the recording and study of electrical activity of the brain
Level of consciousness
Sleep disorders
Brain tumors
Epilepsy
Brain death
Multiple sclerosis
www.eeecube.com
Input
Input
Input
Stage 1
www.eeecube.com
Pre
Amplifier
Stage 2
Amplifier
Writer
Control
Unit
Stage 3
www.eeecube.com
Trace
Driver
Amplifier
Stage 4
Stage 5
www.eeecube.com
Electromyography
Science of recording and interpreting the electrical activity of
www.eeecube.com
www.eeecube.com
Oscilloscope
.
Input
EMG
Amp.
Tape
Recorder
Speaker
A.F.
Amp.
www.eeecube.com
www.eeecube.com
www.eeecube.com
www.eeecube.com
The nerve function is examined directly at the various segments of the nerve by means of
stimulating it with a brief electric shock
Latency elapsed time between the stimulating impulse and the muscles action potential
Procedure
The EMG electrode and the stimulating electrode are placed at two points on the skin,
separated by a known distance l1
When the excitation reaches the muscle, this contracts with a short twitch
The elapsed time t1, between the stimulating impulse and muscles action potential, is
measured
www.eeecube.com
Electroretinography
www.eeecube.com
All sense organs are connected to the brain but the eye has a special
relationship as the retina is the extension of the cerebral cortex
Potentials within the eye can be easily recorded because of its
exposed position
If the illumination of the retina is changed, the potential changes
slightly in a complex manner
The recording of these changes is called the electroretinogram
When light falls on the retina, the absorption of photons by photo
pigments localized in the outer segment of the retinas
photoreceptors takes place
www.eeecube.com
www.eeecube.com
www.eeecube.com