Mode19020 20MHz Dual Trace Oscilloscope
RS Components
ADDENDUM No.1
PARA. 4.2.1.
GENERAL Please allow your oscilloscope to warm up for at least 30 minutes before performing the initial setup procedure.
ITEM 6. Set CH1-POS control to display the CAL waveform centred on the screen as shown in figure 4-1. The resultant deflection should be four divisions (+/-5%).
NOTE: Initial component drift may cause this parameter to be slightly off. This may be easily corrected by adjusting the "Y GAIN" control accessed from the underside of the oscilloscope cabinet. There is a control for channel 1 & one for channel 2.
NOTE: THIS ADJUSTMENT SHOULD BE MADE WITH A NON-CONDUCTIVE ALIGNMENT TOOL.
ITEM 11. Set VOLT/DIV to 5mV/DIV and the DC-GND-AC Switch to the GND position. Centre the trace with the CH1-POS control and rotate the amplitude VAR to the CAL position. The trace should remain in the centre. This is the "DC BALANCE" and there should be no more than .2 divisions shift.
NOTE: Initial component drift may cause this parameter to be slightly off. This may be easily corrected by adjusting the "DC BALANCE" control accessed from the underside of the oscilloscope cabinet. There is a control for channel 1 & channel 2.
If adjustment is necessary.centre the trace with the position control, rotate the amplitude VAR again to the full clockwise position. Adjust the DC BALANCE control to return the trace to the centre of the graticule. Due to interaction between controls, it may be necessary to repeat the procedure one or more times.
NOTE: THIS ADJUSTMENT SHOULD BE MADE WITH A NON-CONDUCTIVE ALIGNMENT TOOL.
Model 9020
20MHz Oscilloscope
Operator's Manual
3000-940-189B
© Beckman Industrial Corporation
Printed in Taiwan JN71PO0789
One Year Limited Warranty
Beckman Industrial Oscilloscopes are warranted against defects in parts, materials and workmanship within a period of one (1) year following the date of purchase of the oscilloscope by the original buyer. This warranty is extended by Beckman Industrial only to the original buyer or original user of the instrument, who must present proof of purchase at the time of warranty service.
ANY IMPLIED WARRANTIES ARISING OUT OF THE SALE OF A MODEL 9020 OSCILLOSCOPE, INCLUDING BUT NOT LIMITED TO IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE, ARE LIMITED TO THE ABOVE ONE (1) YEAR PERIOD. MANUFACTURER SHALL NOT BE LIABLE FOR LOSS OF USE OF THE OSCILLSCOPE OR OTHER INCIDENTAL OR CONSEQUENTIAL DAMAGES, EXPENSES OR ECONOMIC LOSS.
Some states do not allow limitations on how long implied warranties last or the exclusion or limitation of incidental or consequential damages, so the above limitations or exclusions may not apply to you. This warranty gives you specific legal rights, and you may also have other rights which vary from state to state.
Certification
Beckman Industrial certifies that this instrument was thoroughly tested and inspected and found to meet its published specifications when shipped from the factory.
SPECIAL SAFETY PRECAUTIONS
Observe the following safety precautions during oscilloscope operation. BECKMAN INDUSTRIAL CORPORATION assumes no responsibility for the operator's failure to either comply with instructions given in this manual or in the following special precautions.
INSTRUMENT GROUNDING
To m1n1m1ze shock hazard, the instrument must be connected to a three-wire electrical power system. This instrument is equipped with a three-conductor AC power cord. DO NOT connect the instrument to or operate the instrument from an electrical outlet that is not grounded. It is advisable not to use power cord adapter plugs. If this is necessary, plug power cord into a three-contact to two-contact adapter with the ground wire (green) firmly connected to an electrical ground (safety ground) at the power outlet. The power cord and AC outlet of this instrument meet International Electrotechnical Commission (IEC) safety standards.
To avoid any ELECTRICAL HAZARDS, it is recommended for safety that an isolation transformer be connected between the AC power source and the oscilloscope.
OPERATING IN AN EXPLOSIVE ATMOSPHERE
DO NOT operate instrument in the presence of flammable materials possible explosion hazard. Operation of this instrument in such as environment represents a definite safety hazard.
KEEPING AWAY FROM LIVE CIRCUITS
Operating personnel should not remove instrument covers. Component replacement and internal adjustments are best accomplished by an authorized Beckman Industrial Service Center.
For service personnel: DO NOT replace components with power cord connected. Dangerous high voltages may exist even with the power cord disconnected. To avoid injuries, always disconnect power and discharge circuits before attempting to perform any servicing other than live circuit testing or calibration.
Do not attempt oscilloscope adjustments or alignment while alone. Instead, insist on the presence of another person with the capability of administering first aid, if needed.
SUBSTITUTING PARTS OR MODIFYING INSTRUMENT
Because of the danger of introducing additional hazards, do not use substitute parts. Instead, use only factory-approved parts and procedures to service this equipment. Of special importance -- NEVER make any unauthorized modifications to the internal or external areas of the oscilloscope.
X-RAY RADIATION NOTICE
X-ray radiation emitted by the Model 9020 Oscilloscope is less than 0.5mR/hr at a distqnce of five (5) centimeters (12.7 inches) from the cathode-ray tube surface. The instrument is well shielded and meets the safety and health requirements of IEC 348.
X-ray radiation primarily depends on the cathode-ray tube characteristics and associated low voltage and high voltage circuits. Continued safe operation is ensured by periodic adjustment of both low voltage and high voltage power supplies. This is best handled by an authorized Beckman Industrial Service Center. Also be sure that when the cathode-ray tube needs to be replaced, only an identical CRT (Toshiba l50BTB3l) is installed.
Notes:
If the 9020 is turned on while the component test button is depressed, it will appear to be "dead:' To restore normal operation, press and release the component test button.
CONTENTS
Section
Para
Title
ONE
GENERAL INFORMATION
1.1 1.2 1.3 1.4 1.5 1.6
Description
Modes of Operation Vertical Deflection Time Base
AC Power Source
Performance Characteristics and Specifications
TWO
INSTALLATION
2.1 2.2 2.2.1 2.2.2
Unpacking New Oscilloscope Power Requirements
Selecting Proper Power Input Plug Checking and Adjusting Instrument
for AC Line Voltage
Hints on Selecting Operating Site Operating Environment
CRT Intensity
Operating Within Maximum Signal Input Limits
Hints for Operating Oscilloscope Additional Cushions
2.3 2.4 2.5 2.6
2.7 2.B
THREE
OPERATING CONTROLS, INDICATORS, AND CONNECTORS
3.1 3.2
Front Panel Rear Panel
FOUR
OPERATING PROCEDURES
4.1 4.2 4.2.1 4.2.2
Initial Preparation for Use
Checking Oscilloscope Operation Individual Channel Checkout Procedure Dual Trace, ALTernative Operating Mode
Checkout
Dual Trace CHOPped Operating Mode
Checkout
Dual Trace, ADD Operating Mode Checkout Component Tester Checkout
Operating Procedure Information TRIGGER COUPLING Switch Function
4.2.3
4.2.4 4.2.5 4.3 4.3.1
-i-
1-1
1-1 1-1 1-1 1-2 1-2
1-2
2-1
2-1 2-1 2-1
2-1 2-3 2-4 2-4
2-4 2-4 2-5
3-1
3-1 3-1
4-1
4-1 4-2 4-2
4-5
4-6 4-6 4-6 4-B
4-B
CONTENTS (continued)
Section
Para
Title
FOUR
OPERATING PROCEDURES (cont.)
4.3.2 4.3.3 4.3.4 4.3.5 4.3.6 4.4 4.4.1 4.4.2 4.4.3 4.4.4 4.4.5 4.4.6
AUTO/NORM Pushbutton Function TRIG SOURCE Select Function HOLD-OFF Function
Sweep DELAY Function Component Test Function Measuring Instructions
DC Voltage Measurement
AC Voltage Measurement
Period Measurement
Time Period Measurement
Rise (Fall) Time Measurement
/
Synchronization of Complex
Waveforms
Stabilizing Aperiodic Waveforms Selection of Trigger
4-8 4-8
4-8 4-9 4-9 4-9 4-9 4-10 4-12 4-13 4-16
4.4.7 4.4.8
4-17 4-18 4-18
FIVE
ADJUSTMENTS
5-1
5.1 5.2 5.3 5.4 5.5 5.6
DC Balance Adjustments
Trace Rotational Adjustments Focus and Astigmatism Adjustments Power Supply Adjustments Calibration
Probe Adjustments
5-1 5-1 5-1 5-2 5-2 5-4
SIX
MAINTENANCE AND SERVICE
6-1
6.1 6.1.1 6.1. 2 6.2
Maintenance Policy Warranty Policy Warranty Repair Policy Preventive Maintenance
6-1 6-1 6-1 6-2
SEVEN
SCHEMATICS
7-1
-ii-
Section One GENERAL INFORMATION
1.1 DESCRIPTION
.
The BECKMAN INDUSTRIAL Model 9020 Oscilloscope is rugged, easy-to-operate, highly-reliable, and provides dual-channel operation within the frequency bandwidth of DC to 20MHz at -3db with a maximum sweep time of up to 50 ns/div, Waveforms are accurately presented on the oscilloscope's 150mm rectangular CRT screen with internal graticule.
The instrument provides a multitude of operating features including waveform triggering, sweep delay, single sweep, hold off, and component testing. These plus its ease of operation make the Model 9020 an ideal instrument for diversified types of research, education, production, and development of electronic circuits and equipment.
1.2 MODES OF OPERATION
The Model 9020 Oscilloscope provides operating modes in either single or dual trace. In the single trace mode, it can produce waveforms on either channel (CHI or CH2).
While in dual trace mode, the ALT, CHOP, or ADD modes of operation can be selected. In the ALT mode, the trigger signal is derived from both channels alternately. In the CHOP mode, the trigger signal is derived from one channel. Finally the ADD mode is used for adding or subtracting the signals on CH 1 and 2 providing differential measurement waveform analysis.
The CHI (Vertical) and CH2 (Horizontal) have the same input impedance and sensitivity ranges. X-Y waveform analysis is achieved by switching the CHI signal into vertical deflection circuits and the CH2 signal into the horizontal deflection circuits.
1.3 VERTICAL DEFLECTION
Both vertical input preamplifiers (CH 1 and CH 2) have Dual FET and monolithic integrated circuits to minimize drift caused by environmental conditions. Input attenuators provide 10 steps frequency-compensated RC networks for achieving accurate measurements of displayed waveforms.
Each preamplifier has diode-protected FET input circuits. Input signals are individually switched by diode gates and alternately presented to the final vertical amplifier in either the CHOP mode of operation or the ALT mode of operation. Control for the CHOP mode is achieved by a bistable multivibrator-produced 500KHz square wave. Control in the ALT mode is derived from a sweep generator produced blanking pulses. These pulses are frequency divided in two and applied to switching diode gates.
1-1
General Information
1.4 TIME BASE
The time base of the oscilloscope has 20 calibrated sweep speeds from O.lus to 0.2s with a variable control for continuously adjustable sweep rates. The X-POS magnifier control is capable of increasing the displayed sweep rate by a factor of 10.
The HOLD-OFF control alters the hold-off time between sweeps. This is used for triggering aperiodic signals where stable triggering of these complex waveforms cannot be attained with the TRIG LEVEL control alone.
The instrument triggering circuits are highly reliable with sensitivity enough to allow reliable triggering of signals up to 30MHz in frequency and with extremely small amplitudes. The TRIGGER mode can be selected from AUTOmatic Peak-to-Peak Level or NORMal Triggering. TRIGGER COUPLING and SOURCE are multi-function selector switches providing a wide range of options to achieve stable waveforms of very complex signals. A TRIG LEVEL lamp on the front panel flashes to indicate that the trigger circuit has received a trigger signal and the sweep is running.
Single sweep and reset function are used to display non re-occuring events. The TRIG LEVEL lamp indicates in this case that the sweep is awaiting a trigger signal. A simple sweep DELAY selector has 6 decade ranges from lOOns to O.ls for 100 times expansion of the signal interval.
A NORMAL/SEARCH/DELAY switch selects the sweep mode. The DELAY position allows selection of the proper delay time. A MULTIPLIER control provides precise fine adjustment of time delay interval expansion for continuously shifting starting points. IF the delay function is adjusted properly, it is possible to increase the sweep rate for expanding the desired signal interval.
1.5 AC POWER SOURCE
A power source selector fuse switch is provided on the rear panel. The correct fuse to match primary power source combined with a unique primary power selector arrangement is used to modify the power input circuit to accept a multitude power input ac voltages (refer to paragraph 2.2.1).
1.6 PERFORMANCE INSTRUMENT CHARACTERISTICS AND SPECIFICATIONS
Table 1-1 lists the performance characteristics and specifications for the Model 9020 Oscilloscope.
1-2
General Information
TABLE 1-1. PERFORMANCE CHARACTERISTICS AND SPECIFICATIONS
GENERAL,
Cathode Ray Tube:
150mm Rectangular with 8x10 division internal graticulej Approx. 2KV acceleration potential; Phosphor P31
Trace Rotation:
Corrects trace tilt; adjustment is made on right side of instrument.
Beam Finder:
Returns trace to view regardless of control settings
Z-MOD (EXT Blanking input)
DC-Coupling, All TTL low level (~O.2V)input Signal can blank any intensity trace on CRT display. Blank Range: O. 2~30V, unb1anking Voltage: ~O.5V Maximan input Voltage: :!::30V (DC+peak AC)Blanking Bandwidth ~ 2MHZ
Ramp Output:
Sawtooth wave approx. 5Vp-p
Calibrator:
1KHz (5%), O.2Vp-p (2%) Square Wave
VERTICAL DEFLECTION
Bandwidth:
DC to 20MHz (-3db)
Risetime:
17.5ns (approx.)
Overshoot:
< 3%
Calib. Deflection Factor:
10 steps; 5mV/div to 5V/div, 1-2-5 Sequence
Accuracy:
Variable Control:
5:1; max clockwise increases sensitivity five times to 1mv/div, 10MHz( -3dB-) approx. 5% error
Input impedance:
1M ohm, 35pf :!::5pf, (2%)
Input Max. Voltage:
400V (DC + AC peak);
500Vp-p AC at 1KHz or less
1-3
General Information
TABLE 1-1. PERFORMANCE CHARACTERISTICS AND SPECIFICATIONS (continued)
VERTICAL DEFLECTION (cont.)
Operating Modes:
CH1, CH2, CH1 & CH2, ALTernate, CHOPped (Approx. 500KHz)
Algebraic Addition:
CH1 + CH2; or -CH1 + CH2·
Inverter:
CH1 only
TIME BASE
Deflection Factor:
20 calibrated steps; O.lus/div to O.2s/div (1-2-5 sequence); unca1ibrated VAR control extends deflection factor to O.5s/div.
Accuracy:
Less than 3% (100C to 350C)
Expansion:
X10, Accuracy Less then 5% (0.2us, O.lus Unca1ibrated)
Single Sweep:
SINGLE - RESET switches with LED
Ho1doff:
10:1, HOLD-OFF variable control increases ho1doff time
TRIGGER
Triggering Modes:
AUTO (peak) or NORM
Source:
CH1, CH2, ALT (CH1/CH2), EXT.
Coupling:
AC, DC, LF, HF
Slope:
+ or -
Sensitivity:
Internal O.5div (20Hz - 20MHz), External 0.5Vo1ts (min)
Bandwith:
DC - 30MHz
Trigger Level:
Variable control; level indicated by LED
1-4
General Information
TABLE 1-1. PERFORMANCE CHARACTERISTICS AND SPECIFICATIONS (continued)
SWEEP DELAY:
Ranges:
10, 1, O.lms; 10, 1, O.lus.
Mode:
Normal, Search, Delay.
Multiplier:
10 turn variable (Xl to X10)
X - Y OPERATION:
X-Y Mode:
Selected by X-Y switch
Bandwidth:
DC - 2MHz (3dB), Y, CH1; X, CH2
Phase Shift:
Less than 30(100KHz)
COMPONENT TESTER:
Test Voltage:
8.6Vac (rms) (Max., Open circuit)
Test Current:
Max. 28mA (shorted)
Test Frequency:
50Hz/60Hz
Components:
Capacitor, Inductor, Diode, Transistor, Zener, etc.
AMBIENT OPERATIONAL CONDITIONS:
Normal Range:
100 to 350C; R.Humidity 80% max.
Limited Range:
00 to 500C; R.Humidity 80% max.
POWER CONSUMPTION:
Approx. 35 watts
DIMENSIONS:
310 mm (wide); 160 mm (high); 400 mm (deep).
WEIGHT:
9Kg
ACCESSORIES:
Power cord, Operator's Manual Probe (10:1/1:1).
1-5
Section Two INSTALLATION
2.1 UNPACKING NEW OSCILLOSCOPE
When the new Model 9020 Oscilloscope is received, carefully inspect unopened shipping container for any signs of damage. If damage has occurred during transit, immediately notify the carrier or dealer of this damage and file a claim. It is usually advisable not to proceed with unpacking the instrument until the carrier and/or dealer are present to view any possible damage to the shipping container. If the oscilloscope is already unpacked, save the original shipping container, so it is available for inspection, if a claim is filed.
If no damage to the shipping containers is found, open the container and remove the oscilloscope and accessories. Check shipping container contents using the inventory list packed with the equipment.
2. 2 POWER REQUIREMENTS
The instrument operates on power sources of 110, 120, 220, 240 Vac(rms) ±10%; at a frequency of 50/60 Hz single phase.
To minimize shock hazard, the instrument must be connected to a three-wire electrical power system. This instrument is equipped with a three-conductor AC power cord. DO NOT connect the instrument to or operate the instrument from an electrical outlet that is not grounded.
It is advisable not to use power cord adapter plugs. If this is necessary, plug power cord into a three-contact to two-contact adapter with the ground wire (green) firmly connected to an electrical ground (safety ground) at the power outlet.
2.2.1 Selecting Proper Power Input Plug
The power cord and AC outlet of this instrument meet International Electrotechnical Commission (IEC) safety standards. Various configurations used for three-pole power cords are illustrated in Figure 2-1. The number directly above each drawing is the international part code for that power cord. One of the illustrated types is equipped with the instrument. Other plug type (not illustrated) are available as options. Contact your local dealer or Beckman Industrial Corporation for more information.
2.2.2 Checking and Adjusting Instrument for AC Line Voltage
The oscilloscope can operate on anyone of the line voltages shown in Table 2-1. To select proper voltage on voltage selector plug proceed as follows:
1. Locate voltage selector plug (Figure 2-2) on rear panel of instrument.
2-1
Installation
o ~ 110 ~~ 0
OZ;Z;~
Line Voltage Range: 99V -121V Fuse: 0.63A
Selector: 110V
o ~ 120~~ 0 O\'z;~
Figure 2-1. Power Plug Types Available for Use with the Instrument
Line Voltage Range: 198V -242V Fuse: 0.315A
Selector: nov
Line Voltage Range: 112V - 138V Fuse: 0.63A
Selector: 120V
NEMA TYPE ANSI 73.11
LINE = BLACK
NEUTRAL = WHITE GROUND = YELLOW/GREEN
Line Voltage Range: 216V - 264V Fuse: O.315A
Selector: 240V
Figure 2-2. Select Proper Line Voltage as Indicated on Voltage Selector Plug
SCHUKO TYPE CEE7
LINE = BROWN NEUTRAL = BLUE
GROUND = YELLOW/GREEN
2. Remove ac power cord from receptacle.
3. Locate a small screwdriver insert slot between the power receptacle and the voltage selector plug.
4. Insert blade of small screwdriver into the slot and carefully pry outward until selector plug can be removed with your fingers.
2-2
Installation
5. Note fuse rating and check it with Table 2-1 for correct rating necessary with the source input power value.
For example, if you had a source power rating of 125 Vac, 60Hz, then you would select a 630mA fuse and install this fuse in the selector plug. You would then insert selector plug into fuse selector plug receptacle with the "125" located on top and press carefully but firmly until selector plug is flush with surface.
CAUTION
Oscilloscope may not operate or may appear to be defective if voltage selector plug is selected for value not matching ac power input. Be sure to check table on rear of instrument or Table 2-1 for correct fuse rating matching input ac line voltage.
TABLE 2-1. INSTRUMENT OPERATING VOLTAGES
VOLTAGE INDICATED ON SELECTOR PLUG
VOLTAGE TOLERANCE
FUSE
110V l25V 220V 240V
110V - l20V l12V - l35V 220V - 240V 220V - 260V
630mA 630mA 3l5mA 3l5mA
2.3 HINTS ON SELECTING OPERATING SITE
Before selecting an operating site, note the following precautions:
• Avoid locating oscilloscope next to air conditioning or heating vents.
Sudden changes in cooling or heating can cause adverse temperature changes causing possible drift and unwanted changes in measurement results.
• Avoid locations that can block instrument air vents. Locate oscilloscope in a relatively dust free atmosphere without excessive moisture and corrosive chemicals. Performance may be affected by strong electromagnetic fields that exist near large electric motors, diathermy machines, or microwave sources.
• Do not place equipment next to or in a location that is experiencing excessive vibration that can damage circuit boards and other sensitive internal components.
• Operate instrument within 3-1/2 feet (1.07 meters) of power source so that extension cables are not necessary.
2-3
Installation
2.4 OPERATING ENVIRONMENT
The normal ambient temperature range of this oscilloscope is 100 to 350C (500 to 950F). Operation of the oscilloscope outside of this temperature range may damage sensitive measurement circuits.
2.5 CRT INTENSITY
To protect your eyes from excessive strain while viewing screen, adjust intensity until screen is comfortable to view. This usually prevents eye strain and extends CRT phosphor life. When the oscilloscope is not in use, be sure to turn down the intensity as this also extends life of the CRT phosphor and avoids burning spots on the CRT face.
2.6 OPERATING WITHIN MAXIMUM SIGNAL INPUT LIMITS
The maximum input voltage of each input connector and probe input is shown in Table 2-2. NEVER apply a voltage higher than specified.
TABLE 2-2. MAXIMUM ALLOWABLE SIGNAL INPUT VOLTAGES
INPUT CONNECTOR
MAXIMUM SIGNAL INPUT VOLTAGE
CH1, CH2 inputs EXT TRIG input Z - Modulation
400V (DC + AC peak) lOOV (DC +AC peak) 30V (DC +AC peak)
2.7 HINTS FOR OPERATING OSCILLOSCOPE
Observe the following suggestions for successful instrument operation:
• NEVER place heavy objects on oscilloscope.
• NEVER operate with input voltages exceeding maximum limits.
• Do not insert wires, pins, or other metal objects into ventilation holes.
• NEVER place a hot soldering iron on or near the cabinet or especially near the CRT screen.
• Do not place a magnet or magnetic generating device near the cabinet.
• Do not move or pull oscilloscope with power cord or input probe cord.
Especially never move instrument when power cord or signal input lead is connected to a circuit.
2-4
Installation
2.8 ADDITIONAL CUSHIONS
Additional instrument foot cushions are packed inside the manual bag. The cushions, if desired, can be permanently attached to the corners of the right side of the oscilloscope.
2-5
Section Three OPERATING CONTROLS, INDICATORS, AND CONNECTORS
,
3.1 FRONT PANEL
Figure 3-1 shows the Model 9020 Oscilloscope front panel operating controls, indicators, and signal input connectors. Table 3-1 lists each of these and gives a brief functional description.
3.2 REAR PANEL
Figure 3-2 shows the rear panel signal input/output connectors and power receptacle. Table 3-2 lists each of these and gives a brief functional description.
33
22
9B
1
2
3
4
5
8f. 7A
6B
Figure 3-1. Model 9020 Oscilloscope Front Panel
3-1
Operating Controls, Indicators, and Connectors
TABLE 3-1. FRONT PANEL CONTROLS, INDICATORS, AND CONNECTORS
1
POWER pushbutton
Turns power on and off. Power ON lamp lights when power is on.
2
INTENSITY control
Controls brightness of display. Clockwise rotation increases brightness.
3
FOCUS control
After obtaining appropriate brightness with INTENSITY, adjust FOCUS for clearest line.
4
BEAM FIND pushbutton
Brings beam trace to center area of screen regardless of location.
5
COMP TEST jacks
Input banana jacks to test component function (refer to item 15).
6A/6B
CHl(Y) / CH2(X) inputs
Input BNC connector for CHI or CH2 deflection signals. Also provides deflection signal inputs in the X-Y operating mode. Input impedance is 1 megohm paralleled with capacitance of 25 picofarads.
7A/7B
DC/GND/AC switches
Selects following input coupling options for CHl(7A), CH2(7B):
DC: dc coupling, all signal are directly connected to attenuator.
GND: input signal is switched off and attenuator is grounded.
AC.: blocks dc signal component allowing only AC signal to pass into attenuator.
8A/8B
CHl/CH2 VOLT/DIV switches
CHl(Y)/CH2(X) attenuator. Selects deflection factor from 5v/div to 5mv/div (1-2-5 sequence, 10 positions). Calibrated deflection with VAR control (9A/9B) in CAL position.
3-2
Operating Controls, Indicators, and Connectors
TABLE 3-1. FRONT PANEL CONTROLS, INDICATORS, AND CONNECTORS (continued)
9A/9B VAR controls
Provides continuously variable unca1ibrated deflection between calibrated setting of VOLT/DIV switch. In counter-clockwise CAL position, to maximum clockwise increase sensitivity five times.
lOA/lOB CH1/CH2 POS controls
Sweep position adjustment for CH1 or CH2.
In X-Y mode of operation, CH1 POS controls y-axis (vertical), and POS controls x-axis (horizontal) positioning.
11 CH1,CH2 pushbutton
Selects CH1 or CH2 vertical operation, released pushbutton is CH1 operation and pressed is CH2.
12 MONO,DUAL pushbutton
Selects trace mode of operation. Released (MONO) single trace. Pressed (DUAL) both traces.
13 ALT,CHOP,ADD pushbutton
Provides three operating modes:
Pressed (CHOP) mode (all other push buttons in VERT MODE released) derives trigger signal from only one channel.
Released (ALT) mode, trigger signal is derived from both channels alternately.
Pressed (CHOP) ADD mode; with CH1,CH2 in (CH1) and MONO,DUAL in (MONO) (all other push buttons in VERT MODE released). Used to add or subtract CH1 and CH2 signals providing differential measurement waveform analysiS.
14 NORM,INV 1 pushbutton
Selects CH1 operation in normal (released mode) or INV 1 (inverted form). This combined with ALT/CHOP pushbutton (13) can add or subtract input signals.
3-3
Operating Controls, Indicators, and Connectors
TABLE 3-1. FRONT PANEL CONTROLS, INDICATORS, AND CONNECTORS (continued)
15 COMP TEST pushbu~ton
When pressed, tests: FETs, transistors, diodes, Zener diodes, inductors, transformers, capacitors, etc, connected between COMP TEST jacks (5). Characteristic traces of various components are listed in Section Four, Figure 4-4.
16 TIME/DIV switch
Twenty three positions ranging from: 2 sec/div to 0.1 us/div in 1-2-5 sequence.
17 X-Y pushbutton
Pressing applies vertical signal to the CH1(Y) input connector. Vertical screen positioning is controlled by CH1 POS control.
18 HOLD OFF control
Controls hold-off time between sweep signals to obtain stable display when triggering an aperiodic signal. The control hold off time ranges from 1 to 10 times timebase.
19 VAR control
Provides continuously variable sweep rate by a factor of 2.5. UNCAL lamp is turned on when VAR control is not in CAL position.
20 POS (PULL X 10) control
Positions the display horizontally. Pulling control out increases the displayed sweep rate by a factor of 10.
21 CAL output test signal
Provides a 1 KHz, 0.2Vp-p squarewave for testing oscilloscope and for calibrating voltage and probe freq. compensation.
3-4
Operating Controls, Indicators, and Connectors
TABLE 3-1. FRONT PANEL CONTROLS, INDICATORS, AND CONNECTORS (continued)
22 SINGLE pushbutton
Provides single sweep function after a sweep is displayed. Further sweeps can not be presented until RESET pushbutton is pressed.
An LED lamp to the right of pushbutton comes on to indicate that single sweep is triggered.
23 RESET pushbutton
In SINGLE sweep mode (22), the sweep is triggered then pressing this pushbutton resets to single sweep operation, and LED lamp goes out.
24 EXT TRIG input jack
For introducing an external signal for use as a trigger signal. TRIGGER SOURCE switch must be in EXT position for external trigger signal to work.
25 TRIGGER SOURCE switch
Four position switch:
CH 1: A sample of signal derived from CHI input connector is used as a trigger signal.
CH 2: A sample of signal derived from CH2 input connector is used as a trigger signal.
ALT: Alternative trigger which can trigger channels 1 and 2 alternately. Function is used in alternative dual trace, during vertical mode operation.
EXT: Trigger signal is obtained from EXT TRIG connector.
3-5
Operating Controls, Indicators, and Connectors
TABLE 3-1. FRONT PANEL CONTROLS, INDICATORS, AND CONNECTORS (continued)
26 TRIGGER COUPLING switch
Four position switch:
AC: Rejects dc and attenuates signals below 100Hz, accepts signals above 100Hz.
DC: Accepts all trigger signals from DC to 20MHz or greater.
LF: Accepts low frequency signals below about 10Hz.
HF: Accepts signals above 10Hz.
27 TRIGGER MODE pushbutton
Chooses trigger type:
AUTO: Automatically operates trigger action. Trigger level is obtained from trigger signal peak-to-peak value. TRIG LEVEL control is adjusted to a level within peak range of signal.
NORM: Trigger level range is obtained from +Sv to -Sv by adjusting TRIG LEVEL control.
28 TRIGGER SLOPE pushbutton
Selects triggering slope:
"+" Triggering occurs when trigger signal crosses trigger level in a positive-going direction.
"-" Triggering occurs when trigger signal crosses trigger level in a negative-going direction.
3-6
Operating Controls, Indicators, and Connectors
TABLE 3-1. FRONT PANEL CONTROLS, INDICATORS, AND CONNECTORS (continued)
29 TRIG LEVEL control
Used to select a level within signal peak range from which the sweep can be triggered at a proper input signal starting edge. When sweep operates, TRIG'D lamp comes on.
30 NORMAL, SEARCH, DELAY switch
NORMAL: Selects main sweep functions. LED lamp (at right of switch) is off.
SEARCH: Searches for a point at which the sweep will be delayed. LED lamp flashes.
DELAY: - Used in conjunction with DELAY switch whereby delayed signal is displayed.by increasing sweep rate of timebase. LED lamp stays on.
31 DELAY switch
Selected delay time after trigger point.
32 MULTIPLIER control
Provides continuously variable delay time after trigger point.
33 CRT display
l50mm rectangular screen with internal graticule 8 x 10 div and blue filter placed in front of CRT.
3-7
Operating Controls, Indicators, and Connectors
,.
z
$. $
I
DANGER ~ ~
HIGH VOLTAGE INSIDE
FUSE
._IIOT"""~"'" KLECTOII lINE
1ell'flCl'1'O~""__"
• eIK~C'T ~ co-III!PC* 111 •
-- as n
• MfIUICC"* wnw~"'" -
_ .....
...
LINE YOLTAG
49 ., ,
e~
.I!
• o®o 3
\-".aXII
~PUT
-®
2
BECKMAN INDUSTRIAL CORPORATION
AN AFFILIATE OF EMERSON ELECTRIC co INSTRUMENTATION PRODUCTS DIVISION SAN DIEGO. CA1..lFORNIA
."MAlIIIUII
~
DATA
VOl. TAGE fUSE
"'·'11" _
PII·,_ _
,..,.- - _._" -
E.~CTOf'
SERIAL NO:
o~ \ __ EgfIl __
MADE IN TAIWAN
1
Figure 3-2 Model 9020 Oscilloscope Rear Panel Connectors
TABLE 3-2. REAR PANEL CONNECTORS
I
POWER RECEPTACLE AND FUSE HOLDER
Receptacle for AC power input. Fuse Holder is used as a power source selector.
2
S - output connector
Provides ramp (sawtooth wave) output about SVp-p, which is internal sweep signal.
3
Z - modulation connector
Input terminal for external intensity modulation signal.
3-8
Section Four OPERATING PROCEDURES
4.1 INITIAL PREPARATION FOR USE
Before attempting to operate your new oscilloscope, perform the following procedure:
1. Verify that voltage selector block is in correct position with proper voltage displayed in top position on selector (refer to paragraph 2.2.2 in Section Two).
2. Insert power plug into receptacle on rear panel.
3. Place all pushbutton switches on front panel in out (released) position, except DUAL/MONO switch in VERT MODE pushbutton group to DUAL position (pushed in).
4. Set front panel controls to positions as indicated below:
• INTENSITY fully counter-clockwise.
• TRIGGER SOURCE to ALT.
• CH1 and CH2 amplitude VAR to CAL.
• DC/GND/AC to GND.
• CH1-POS and CH2-POS to midrange.
• TIME/DIV to .Sms/DIV.
• NORMAL/SEARCH/DELAY to NORMAL.
• Sweep VAR to CAL.
• TRIGGER AUTO/NORM to AUTO.
5. Plug power cord into ac power outlet. Press red POWER on pushbutton in; power ON lamp comes on.
6. Adjust INTENSITY control for comfortable view of traces. If base lines are not found, press BEAM FIND button to locate traces and adjust CH1/CH2 POS controls and horizontal POS control to center traces on CRT display.
7. Now proceed to paragraph 4.2 to checkout the functional operation of the instrument.
4-1
Operating Procedures
4.2 CHECKING OSCILLOSCOPE OPERATION
Before using the new oscilloscope, be sure to go through the procedures in this paragraph to help familiarize yourself with the front panel controls and features of the Model 9020 Oscilloscope. Also any time that you wish to check instrument operation, it is a good idea to do this procedure because it functionally checks each display mode and operation of the front panel controls. The instrument can be operated in these procedures without the need of additional test equipment by using the front panel CAL (0.2V, 1KHz) output as a signal source.
4.2.1 Individual Channel Checkout Procedure
1. Adjust front panel controls as follows:
CHANNEL 1
VOLT/DIV ••• 5OmV/DIV DC-GND-AC • • .AC
VA!!.. • • • • • • CAL CHI-POS. • • • Center
VERT MODE ••• (all pushbuttons to out)
TIME BASE
POS. • Center (pushed in)
Sweep VAR. • • • • CAL
TIME/DIV • • •••• 0.5ms/DIV
HOLDOFF ••••••••• MIN SINGLE • • • • • • • • • OUT NORMAL/SEARCH/DELAY ••• NORMAL
TRIGGER
MODE • • AUTO
SLOPE. •• + (PLUS)
TRIG LEVEL • • Center SOURCE • • • • CH 1
COUPLING • AC
2.
Press Power ON and set INTENSITY and FOCUS controls for desired baseline display.
3.
Apply CAL voltage output directly to CH1 vertical input.
4.
Adjust TRIG LEVEL for stable display. Also adjust X-POS for sweep baseline starting at first graticule line.
5.
Verify six pulses with leading edge of first and sixth pulse on first and eleventh vertical graticule lines, respectively (±10%), as shown in Figure 4-1.
6.
PLEASE REFER TO ADDENDUM No.1
7. Rotate sweep VAR control clockwise to stop. between first and eleventh graticule lines.
Verify 12 or more pulses Then move to CAL position.
4-2
Operating Procedures
I
~ce ...
•••• •••• 4O •• 4O •••••••••••••••••••••• __
, FIRST
f ELEVENTH
Figure 4-1. Six Leading Edges Shown with First and Eleventh on Graticu1e Lines
8. Set VOLT/DIV to 0.2V/DIV. Verify vertical amplitude in one division on screen.
9. Rotate amplitude VAR control clockwise to stop.
10. Verify amplitude of five divisions <±10%). The amplitude VAR control increases sensitivity five times.
11.
PLEASE REFER TO ADDENDUM No.1
12.
Adjust TRIG LEVEL control fully clockwise. Then, press in AUTO/NORM swi tch to NORM (TRIG LED goes oft) and verify no sweep.
13.
Press in SINGLE pushbutton, RESET indicator goes on and stays on.
14.
Press RESET pushbutton switch, Verify no sweep.
15.
Rotate TRIG LEVEL fully counter-clockwise.
16.
Verify one sweep, RESET indicator goes off after sweep.
17.
Set AUTO/NORM switch to AUTO and TRIG LEVEL to center position.
4-3
Operating Procedures
18. Press RESET pushbutton. Verify one sweep.
19. Disable SINGLE sweep (SINGLE switch in out position).
20. Set NORMAL/SEARC~/DELAY switch to SEARCH position (LED lamp flashes).
21. Select DELAY time switch to lms.
22. Verify that portion of displayed trace on left screen is blanked, as shown in Figure 4-2.
, o
I
f ~
1P. .•. .. . .. . . . . .. . .... .... - ro· ••• .. . . .. .. .... - ...... -.--
ro
__._. . • LL.L L.I. . .. • .I.
. .
p .... .... .. .. .... - .. .. .. .. .. .. ..... .. ...... .. ...... . ...... .. ...... - ......
f· , 9
Figure 4-2. Left Portion of Waveform Blanked
23. Rotate MULTIPLIER (10 Turns) to increase (or decrease) the blanked portion about 2-5 divisions at a desired rising or falling edge. If DELAY TIME is set too short, blanked portion will be very short. If DELAY TIME is set too long, waveform will be blanked.
24. Set NORMAL/SEARCH/DELAY mode to DELAY position (LED lamp stays on fully) •
25. Set TIME/DIV to 50us/DIV (or higher sweep rate).
26. Slightly adjust MULTIPLIF.R to position rise or fall edge in center of SCreen. Verify that waveform is expanded as shown in Figure 4-3.
4-4
Operating Procedures
1 o
J
~ ... ~
.... . . . . . ... . ... .... . ... . ... . .. - _._-
I
_L . I • . .
') . .
V
p' ... . . . . . ... . . . . .... . ... . ... . ... . ... . ...
f' . 1 9
Figure 4-3. Sweep Delay Showing Expansion of Input Signal to Desired Interval
27. Set NORMAL/SEARCH/DELAY switch to NORMAL position and TIME/DIV to 1ms/DIV.
28. Pullout horizontal POS control (magnify 10 times). Verify that two pulses with leading edge of first and second pulse are on first and eleventh vertical graticule lines, respectively (±lO%).
29. Push in horizontal POS control.
30. The checkout of Channel 2 is the same as Channell. To checkout Channel 2 repeat steps 1 through 29.
4.2.2. Dual trace, ALTernative Operating Mode Checkout
1. Connect 0.2V CAL output to both channel signal input connectors.
2. Press in on MONO/DUAL switch to obtain DUAL channel operation.
4-5
Operating Procedures
3. Move TRIGGER SOURCE switch to ALT position.
4. Perform operating timebase functions in the same way as described in paragraph 4.2.1, except in this case both channels are affected.
4.2.3. Dual Trace, CHOPped Operating Mode Checkout
1. Connect O.2V CAL output to both channel signal input connectors.
2. Press in on MONO/DUAL switch to obtain DUAL channel operation.
3. Press in on ALT/CHOP switch to obtain CHOP operating mode.
4. Move TRIGGER SOURCE switch to CHI or CH2.
5. Perform operating timebase functions in the same way as described in paragraph 4.2.1, except in this case both channels are affected.
4.2.4. Dual Trace, ADD Operating Mode Checkout
1. Connect O.2V CAL output to both channel signal input connectors.
2. Actuate MONO/DUAL switch to obtain MONO position; AI.T/CHOP switch to CHOP position; and CHl/CH2 switch to CHI.
3. Move TRIGGER SOURCE switch to ALT.
4. Perform operating timebase functions in the same way as described in paragraph 4.2.1, except in this case both channels are affected. Also ignore any instructions to change positions of VERT MODE pushbuttons and TRIG SOURCE switch, except as instructed above in this paragraph (4.2.4).
4.2.5. Component Tester Checkout
1. Press in COMP TEST pushbutton to place oscilloscope in component test operating mode.
2. Disconnect CHI and CH2 input connectors.
3. Insert diode or zener diode, LED, capacitor, etc., between COMP TEST banana jacks.
4. Verify that displayed waveform are similar to the Test Patterns shown in Figure 4-4.
5. When you have finished component testing, press COMP TEST pushbutton to out position. This disables component test mode of operation.
4-6
Operating Procedures
Single Passive Device
Short circuit
Resistor 680n
Single Transistor (for n-p-n type)
_j
8-C junction
8-E junction
Single diode (with cathode grounded)
_j
Silicone diode
_j
Rectifier
In-circuit component
B-E junctionH680n
_j
Germanium diode
o
o
)
Capacit or 471.1 f
Power transformer primary
Diode# 680n
8-E junction II O· l ut
E-C barrier
_j
FET (between gate and source,with source grounded)
_j
Zener diode under 8V
Zener diode beyond 12V
o
-
2 diodes antiparallel
Capacitor hd'1Kn
d
E-8 junctionllDiodelllKn Diode 114.7~f
Figure 4-4. Component Test Patterns
4-7
Operating Procedures
4.3 OPERATING PROCEDURE INFORMATION
4.3.1
TRIGGER COUPLING Switch Function
The TRIGGER COUPLING slide switch has four positionsl AC/DC/LF/HF. The correct position to use for coupling of the input signal for triggering depends upon the following information.
AC coupling removes the DC level from trigger signals and attenuates signals below 10Hz. In most cases this coupling selection works well. Only in some difficult cases where AC signals are highly complex do you need to use a different coupling position.
DC coupling connects input signals. directly to the input amplifier. With DC coupling selected, a large DC voltage component in an input signal can offset the input signal outside the trigger level range of the instrument and cause the oscilloscope to lose trigger.
HF and LF positions both provide AC coupling functions. When AC signals are complex, and it is difficult to get stable waveforms by rotating TRIG LEVEL, either HF or LF can be selected to block low frequency (HF) or bypass high frequency (LF) signals providing a much more stable trigger function.
4.3.2 AUTO/NORM Pushbutton Function
In AUTO mode, the trig level is automatically set within the input signal peak-to-peak range. A bright baseline is displayed in the absence of a trigger signal. A trigger of 30Hz or higher overrides AUTO operation and produces a presentation. Adjustment of TRIG LEVEL control may be necessary to obtain a stable display. If the trigger is less than 30Hz, NORM operation must be used. A trigger signal is always needed in NORM operation to generate a sweep.
4.3.3 TRIG SOURCE Select Function
The trigger select SOURCE switch selects trigger sources from CH1, CH2, ALT or EXT. The ALT position selects a trigger source from both channels alternately. When dual trace ALT is selected, signals from both channels are required to obtain a stable display. In ALT trigger operation, if input frequency of Channell differs much from Channel 2, the displayed waveforms may be dim and unstable.
4.3.4 HOLD-OFF Function
When the trigger signal is complex or is aperiodic, a stable waveform cannot be displayed by using the TRIG LEVEL control alone. It is necessary to rotate the HOLDOFF control to get more stable waveforms.
4-8
Operating Procedures
4.3.5 Sweep DELAY Function
By employing the sweep delay function, the start of the sweep can be selectively delayed from the trigger point O.lus to a maximum of lOms. It is therefore possible to start the sweep at nearly any point in a waveform. The interval, which follows the start of the sweep, can be greatly expanded by increasing the sweep speed. It can be expanded at least 100 times by using the DELAY switch and can be expanded to 1,000 times using the MULTIPLIER control; however, the display brightness decreases as sweep speed increases. Note that input signal jitters are somewhat increased with higher expansions of sweep speed.
Operation of the sweep delay is easy. Only three controls in the DELAY section are needed. The delay mode slide switch NORMAL/SEARCH/DELAY, the DELAY time rotary switch (selection of delay time range), and the MULTIPLIER variable control (10 Turns, fine control).
4.3.6 Component Test Function
With the built-in component tester: semiconductors, capacitors, inductors, resistors; R-L-C network or even combinations of R-L-C and semiconductors can be easily tested. Also, integrated circuits may be checked.
The tested result is displayed on the screen of the oscilloscope. The device under test can be either in or out of a circuit. When the device under test is situated in highly complex circuits, the test result may not be clear. This is due to the effects of the components and the stray capacitance in the test circuit. This problem may be overcome by comparing the result obtained from the circuit under test with the result obtained from a similar circuit which is known to be fully operative.
4.4 MEASURING INSTRUCTIONS
4.4.1 DC Voltage Measurement
1. Connect DC voltage directly to CHI input.
2. Set AC-GND-DC to GND.
3. Adjust CHI-POS until baseline coincides with one horizontal graticule lines. This is used for the zero reference level.
4. Set VOLT/DIV to an appropriate range, and set AC-GND-DC switch to DC.
5. Baseline shifts by the amplitude of DC voltage. Multiply measurement of shift divisions by value indicated on VOLT/DIV switch. For example, in Figure 4-5, when VOLT/DIV is 5OmV/DIV, the shift division is 4.2DIV, then the DC voltage (Vdc) is: 5OmV/DIV x 4.2DIV:
= 50 x 10-3 x 4.2 = 210 x 10-3
= 210 mVdc
4-9
Operating Procedures
T
4 o
I ~
I-
~
~b: ... l-
I-
- .... - .... . ..... - ..... .. . .. .. .. ..... .. ...... .. .... - ...... -
.2
IV • • · . • . . •
•• • · .. . . . ." . • .
...
.
-
if ~
10· •.• l-
· - - .. . .... .. ...... . .. .. .. .. ...... .. ..... .. .... . .... . ...
10
r or-
o
4ZERO LEVEL(REFERENCE LEVEL)
- DC LEVEL SH 1FT
Figure 4-5 DC Voltage Measurement
6. If the XlO attenuation probe is in use, the true value of the signal becomes 10 times the calculated value, or 5OmV/DIV x 4.2DIV x 10 = 2.10 Vdc.
4.4.2 AC Voltage Measurement
1. Connect signal directly to channell input.
2. Set VOLT/DIV to an appropriate range, and set AC-GND-DC switch to AC.
3. Adjust CHI-POS control to position lowest portion of waveform to coincide with horizontal graticule lines between 10 and 90.
The peak-to-peak voltage (Vp-p) is measured as shown in Figure 4-6. If VOLT/DIV is adjusted to O.lV/DIV, and wayeform amplitude is 4.4 divisions, then Vp-p = O.lV/DIV x 4.4Vp-p = 0.44Vp-p or 44OmVp-p
4-10
Operating Procedures
1 o
I ..
..
..
..
~ .• j ·
·
~ ... - ... · ... . ... ..... .. oJ ~. O. . .. - · .. -
..
j \ · if \
..
~
~
I ~ ~./ 1\
. . . · • •• . .. · ..
... . . . . '\ · . r · .. . . .. , · ..
~ j ~ ~
\ VJ \
0···· . - .. - .. - · ... . - .. ~ .... · .. - ... - ... - . ...
I .~
0/0 ~
l-
. 1 9
Figure 4-6. AC Voltage Measurement
4-11
Operating Procedures
4.4.3 Period Measurement
The duration of a signal period is determined by multiplying the horizontal distance in the signal period by the time coefficient selected on the TlME/DIV switch.
For example, as shown in Figure 4-7, the waveform which starts at point A and ends at point B is equal to one time period. The distance between A and B is equal to 2.0 division on the screen.
If, TlME/DIV is equal to O.5ms/DIV, then the period (or A to B interval) is calculated as follows:
0.5ms/DIV x 2.0 DIV = 1.0 mS
= 1. 0 x 10-3 S
Then, signal frequency (F) is
= 1/ (1.0 x 10-3) S = 1 x 103 Hz
= 1 KHz.
1 o
I ~
~
... . .... .. . . .. . ... . ..... - .. - - ...... . ...... ...... _.-_
~
f , r ~ f l f \ f ~
I I. , I . , ~l , I ,
\ \ .1 .\ . L
~-' ,. J " J" ~ '\ J ''l J 'l
t \
O· ..• ....... .. .... - ...... .. - .... .. .... - ..... . ... - .. ..... .. -.
I
r
t ,\ A
B
Figure 4-7. Period Measurement
4-12
4.4.4 Time Difference Measurement
Operating Procedures
Time difference measurements are accomplished in dual trace operation. When measuring the time difference between two signals, a position is selected on the TRIGGER SELECT switch as a source for the trigger reference signal. Two typical signals are displayed on the screen as shown in Figure 4-8.
,1
9~ .....
lI-
CHI
10·· ~.
I I
°C-T-1-+--+--+--+--+-+-+-+-+--I
t
CH2
I I
AS
Figure 4-8. Waveforms for Time Difference Measurements
Figure 4-9 shows an expanded waveform of Figure 4-8, with trigger source as CHI. It is easier to measure the horizontal divisions from point A to point B point as shown in Figure 4-9. The time difference as for other measurements is obtained by multiplying the interval between A and B by sweep speed.
Figure 4-10 is the waveform triggered by CH2 signal which is not a reference signal, and therefore impossible to measure time difference. A more precise measurement method is shown in Figure 4-11 and 4-12.
4-13
Operating Procedures
I
~r ... ........ ··f .. ...... ... .. .. .. "" .. .. -. .. ...... .. J .. .... -
~ r-,
/
10··· - ... ". . - - ........ .. ...... . ..... J - ......... ... ....... .. ......
I
~, A
B
CHI
CH2
Figure 4-9. Expanded Waveforms Allow Easier Difference Measurements (TRIGGER SOURCE set to CH 1)
I
1 ''"' ~--
0- . - . .. .. .. .. .. .. .. .. .. .. .. .. - - - - ........ ........ .. ...... .. .... -
\.
L
. . .
r ~ J
'?, - -- j-- - ...... ........ ... ...... .... . - .. .. .. ... .......... --t .. ...... 9
CHI
CH2
Figure 4-10. Improperly Expanded Waveform Making Measurement Impossible (TRIGGER SOURCE set to CH 2)
4-14
Operating Procedures
Figure 4-12 is the equal amplitude measuring method, and is achieved by rotating amplitude variable controls and CHI POS control to equalize amplitudes for both channels.
".__ TIME -~.."
DIFFERENCE
Figure 4-11. Equal Amplitude Measuring Method
Figure 4-12 is the superposition measuring method, and is achieved by adjusting the CHI-POS control to position the smaller amplitude waveform to the center of the larger amplitude waveform.
&:flo sn.
TIME
DIFFERENCE
SCY/o sn.
j Figure 4-12. Superposition Measuring ~1ethod
4-15
Operating Procedures
4.4.5 Rise (Fall) Time Measurement
With pulse or square waveforms, the critical factor is the risetime of the voltage step. The rise time is generally measured between 10% and 90% of the vertical pulse height. Adjust signal peak-to-peak amplitude to fall within six divisions on the screen, and symmetrically adjust it to the horizontal center line. Match rise or fall line on the waveshape to the internal graticule of the two horizontal dotted lines and within +2.4 divisions from screen center line. A properly adjusted waveform is shown in Figure 4-13.
To measure the rise time, adjust the VOLT/DIV switch with variable control and CHI POS control so that the waveform amplitude precisely coincides with 0% and 100% lines on CRT graticule, as shown in Figure 4-13.
1 o
I
~ ... : .. i ~
.... - ... . .... - .. - · ... . .. - . - .-
/
~_._ _L V. • • .. •
. . ") . · . . · . . . . ..
/..
o· ... . - .. . . . - .... V·· . . ... · ... · .... . .... - ...
I ..,
f' . 1 9
Figure 4-13. Rise or Fall Time Measurement
Risetime is defined by the product of the horizontal divisions between the waveform intersections to 10% and 90% dotted lines and the sweep time. If magnification is used, this product must be divided by 10. The falltime of a square wave can also be measured by using this method.
4-16
Operating Procedures
When risetime is measured directly from the displayed waveform on the screen, it contains two parameters. One parameters is the true risetime of the measured signal, the other is the risetime of the oscilloscope. For example,
Ts = measured risetime from CRT screen
To = oscilloscope risetime
Tms = measured signal risetime
Risetime displayed in Figure 4-13, Ts has the following relationship,
Actually, the true rise time is given by,
4.4.7 Synchronization of Complex Waveforms
In the case shown in Figure 4-14, the waveform contains two peaks that greatly differ in amplitude. This waveform will display double peaks of equal amplitude on the screen, if the trigger level is not set properly.
If trigger level is set at point L2 then both peaks will be displayed as shown here.
If trigger level is set at L1 then displayed waveform will display incorrectly with instability and with incorrect amplitude.
Figure 4-14. Waveform with Dual Peaks Displayed at Different Amplitudes
4-17
Operating Procedures
In Figure 4-14, if the TRIG LEVEL is set to L1, the waveform may be doubled, or even unstable. Operators should increase the TRIG LEVEL control until the display on the screen is synchronized.
4.4.7 Stabilizing Aperiodic Waveforms
In the case where waveforms greatly vary in pulse width, obtaining a stable display is difficult to adjust with TRIG LEVEL only. The HOLD OFF function will improve synchronization (Figure 4-15).
With an aperiodic waveform, adjusting HOLD-OFF control increases hold off time between two sweeps. If the hold-off time is increased to a proper range, an undesirable triggering level cannot generate sweep, thereby synchronizing the waveform. If adjustment of HOLDOFF and TRIG LEVEL controls do not produce a stable display, set TRIG COUPLING switch to HF position and readjust HOLD-OFF and TRIG LEVEL for improvement.
-
9~··· ".' -
1?' ...
T·-+----f--+--+--+---+--+----i--+--~
. .
-
.
Figure 4-15. Aperiodic Square Wave
4.4.8 Selection of Trigger
As an aid to achieving easier operation and stable well-synchronized waveforms, recommended instructions are shown in Table 4-1.
4-18
Operating Procedures
TABLE 4-1. TRIGGER SELECTION GUIDE
Pushbutton Switch Positions
Use HOLD-OFF Use TRIG LEVEL
VERT MODE TRIGGER TRIGGER TRIGGER if if
GROUP SOURCE COUPLE MODE wave shape is: wave shape is:
CHI CHI AC AUTO Complex,Jitter Complex
CH2 CH2 AC AUTO Complex, Jitter Complex
DUAL,ALT ALT AC,HF AUTO,NORM If Unstable Complex
DUAL, CHOP CHl,CHl AC,HF NORM
ADD CHl,CH2 AC AUTO 4-19
Section Five ADJUSTMENTS
5.1 DC BALANCE ADJUSTMENTS
Over long periods, the DC balance characteristics of the FET input of the vertical amplifier for each channel FET may drift.
If the vertical baseline position shifts greater than lmm after a 30 minute warm-up of the instrument, then adjustment of the DC BALANCE potentiometer may be necessary. Access to this adjustment can be gained from the underside of the oscilloscope cabinet.
Before adjusting, set VOLT/DIV switch on the front panel to 5mV/DIV, and input coupling to GND. Also, use a non-metallic alignment tool to adjust the DC balance potentiometer.
The front panel VAR control should be continuously readjusted while the DC balance potentiometer is adjusted.
Correct DC balance is obtained when the vertical baseline position remains steady while the VAR control is varied. Use this same procedure for both Channel 1 and Channel 2.
5.2 TRACE ROTATION ADJUSTMENTS
The trace rotation adjustment compensates for external magnetic fields that may affect alignment of the horizontal baseline with respect to the CRT graticule. When the oscilloscope is moved to a new location, trace rotation adjustment should be checked and readjusted if necessary. To adjust the trace, proceed as follows:
1. Set DC-GNC-AC coupling switch to GND
2. Adjust CHI POS control until baseline is on the center of the horizontal graticule line.
3. Using a non-metallic alignment tool, adjust TRACE ROTATION control (which is located on the Right Side of the cabinet) until baseline aligns with horizontal graticule line.
5.3 FOCUS AND ASTIGMATISM ADJUSTMENTS
Normally, the CRT has extremely good brightness and sharpness. Any reduction of the brightness or sharpness can be corrected by the FOCUS and ASTIGMATISM controls. Adjustments of focus and astigmatism can be performed as follows:
5-1
Adjustments
CAUTION
CRT brightness and sharpness will remain for very long intervals. Do not adjust ASTIGMATISM, except when necessary.
1. Turn INTENSITY control fully counter-clockwise.
2. Set vertical and timebase controls as follows:
VOLT/DIV ••••••• 1OmV/DIV
AC-GND-DC • .GND
VERT MODE • •• .All push buttons out
VARiable. • • .CAL
TRIGGER ••••••• SOURCE (CH 1); COUPLING (AC) TRIG LEVEL •••••• Center
X-Y deflection •••• pushed in
3. Set INTENSITY to observe spot.
4. Adjust FOCUS and ASTIG controls for best defined spot. ASTIG control is located on the Power Supply Board as shown in Figure 5-1.
CAUTION
High Voltage exists near ASTIG control. Adjust carefully.
5.4 POWER SUPPLY ADJUSTMENTS
There are seven regulated DC operating voltages: +24V, +5V, -12V, +140V, +260V, including -1900V and 33V for the unblanking circuit. The +24V, +5V, -12V and 33V are fixed voltages ±5% tolerance.
It is not recommended that the operator attempt to readjust these voltages. Better to refer all alignment and servicing to an authorized Beckman Industrial Service Center.
5.5 CALIBRATION
To maintain oscilloscope accuracy, perform calibration at least every 1000 hours of operation, or every six months if used infrequently.
It is not recommended that the operator attempt to perform any calibration or any servicing of this oscilloscope. Instead refer this type of activity to an authorized Beckman Industrial Service Center.
5-2
Adjustments
Figure 5-1.
Power Su 1
pp y Circuit Board
5-3
5.6 PROBE ADJUSTMENTS
Adjustments
Adjust probe frequency compensation by using following procedure (Probe is Xl/XlO attenuator):
1. Set probe to XlO position, then connect probe tip to output of CAL O.2V, 1KHz
2. Set VOLT/DIV to 5mV/DIV
3. Set AC-GND-DC to DC.
4. Adjust frequency compensating trimmer on the probe until the displayed square wave is in optimum state as shown in Figure 5-2. NOTE: When probe is in Xl position it does not require frequency compensation adjustment.
OPTIMUM
CAPACITANCE TOO SMALL
CAPACITANCE TOO LARGE
5-4
Figure 5-2. Probe Compensation
Section Six MAINTENANCE AND SERVICE
6.1 MAINTENANCE POLICY
6.1.1. Warranty Policy
The Beckman Industrial Model 9020 Oscilloscope, excluding probes, is warranted against any defects of material or workmanship which develop within a period of one (1) year following the date of purchase by the original purchaser. This warranty is extended by Beckman Industrial Corporation only to the original purchaser of the oscilloscope, who must, as a CONDITION PRECEDENT TO WARRANTY COVERAGE AND PERFORMANCE THEREUNDER BY BECKMAN INDUSTRIAL CORPORATION, complete and return the Warranty Registration Card, received on purchase of the oscilloscope.
ANY IMPLIED WARRANTIES ARISING OUT OF THE SALE OF THE BECKMAN INDUSTRIAL CORPORATION OSCILLOSCOPE, INCLUDING BUT NOT LIMITED TO IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE, ARE LIMITED IN DURATION TO THE ABOVE-STATED ONE (1) YEAR PERIOD. BECKMAN INDUSTRIAL CORPORATION SHALL NOT BE LIABLE FOR LOSS OF USE OF THE OSCILLOSCOPE OR OTHER INCIDENTAL OR CONSEQUENTIAL DAMAGES, EXPENSES, OR ECONOMIC LOSS OR FOR ANY CLAIM OR CLAIMS FOR SUCH DAMAGES, EXPENSES, OR ECONOMIC LOSS.
Some states do not allow limitations on how long implied warranties last or the exclusion or limitation of incidental or consequential damages, so the above limitations or exclusions may not apply to you.
6.1.2 Warranty Repair Policy
In the event a defect develops during the warranty period, Beckman Industrial Corporation will, at Beckman Industrial Corporation's election, repair or replace the oscilloscope with a new or reconditioned equivalent model, without charge for parts or labor. In order to obtain performance of any obligation of Beckman Industrial Corporation under the warranty, the original purchaser must notify Beckman Industrial Corporation Customer Service Department at
(213) 690-7255 of the defect, obtain a Return Authorization Number and receive instructions on where to return the defective product for the performance of service. In the event the oscilloscope must be returned, it must be sent postage prepaid, along with a handling charge of $10 to an authorized Beckman Industrial Corporation Service Center.
In the event of replacement with a new or reconditioned model, the replacement unit will continue the warranty period of the original oscilloscope for six
(6) months after the date of repair, whichever is longer. .
6-1
Maintenance and Service
This warranty shall not apply to any defect, failure or damage caused by improper use or improper or inadequate maintenance or care. Beckman Industrial Corporation shall not be obligated to furnish service under this warr~nty:
a) to repair damage resulting from attempts by personnel other than Beckman Industrial Corporation representatives to install, repair or service the product;
b) to repair damage resulting from improper use;
c) to service a product that has been modified or integrated with other products when the effect of such modification or integration increases the time or difficulty of servicing the product.
6.2 PREVENTIVE MAINTENANCE
It is important that you follow the preventive maintenance items indicated below to insure years of reliable instrument performance and in certain cases extend the life of the instrument:
• Since semiconductors, precision components, static-sensitive devices, etc. are employed in this oscilloscope, use utmost care in operation and storage.
• Clean front surface of blue CRT filter every two weeks with soft tissue moistened in window cleaning fluid. If dust should accumulate on the CRT surface, the filter can be removed* for lightly dusting or blowing off the surface of the CRT only. Be especially sure not to wipe or even touch the inside surface of the blue filter as it is coated with a soft antistatic spray. This coating should remain on the inside surface of the blue filter to protect the CRT from nearby static charges.
• Every month wipe off entire oscilloscope cabinet with a cloth dampened in a anti-static electronic equipment cleaner.
• Be sure to retain the shipping box for storage of the instrument when it is not in use. Store in an ambient temperature from -IOoC to +600C, and keep away from moisture and dust.
*CAUTION: DO NOT TOUCH OR WIPE the inside surface of the blue filter because it is coated with a soft antistatic spray. Because the coated side of the filter is unmarked, note this position when it is removed. Then when reinstalling ensure that the coated surface is again positioned to face the CRT.
6-2
Section Seven SCHEMATICS
Schematic diagrams for the Model 9020 Oscilloscope are presented on the following pages:
7-1
...... I N
All trimmers 2.5P-6P
except VClll and VC 211 3.5P-10P
_ AiT-BOARD ATTENUATORS
~ ~L-----
R120 33 -,- - - - - - - - - - - - -
--,
r I
I
I
I
I
I
I
I
I L
vms R129 1MF
AC
GND~R01 !
DC 51 115V [002
400V
= 0.0471'f I
--,
--~
r I
I
I
I
I
I
I
I
I L
AC
GND
DC
V-BOARD VERTICAL AMPLIFIER & CONTROL LOGIC lYill
R156 6RB
r--~--r-"'_____'-~--./IN----D+5V
1C140 [15B + 1[157
:133 IOn 122n ![
~~ 10/-< I22n
40'OV -
~o-~R~I~3~4~R~IN35w -=
220K 100
132 1M F
0101 FD333
D204
D401
I cb'C422 J
lOOP r----.,.__---, -<> "-
R411l~;~
lK r1.E102 '---8
IC401rw ~ ~
4013 fl C3J ©
2 11
~ -=
R409 22K
R444 6R8
R181 12K
I N4148 '22
~rllD---,
R439 1
220 -
R422 2.2K
0404
T401
~ 2~0~5Y
~~fl
lK
~ [--oR7 32 Illil
C419 R403 R402
470P 10K 15K
C5 55 [ill]
R40B R447
47K 47K
0402 D403
'" ~
'" 0 tI
"'" TO 0605 ~
'W it
-{j) R6011r1'21J ~
_"" R602ITm]
- I'"
...
r.
0: le239 C233 IOn I Z2n
~ -
[ill] o--+..,R N231/'4-+-R.".2~3"--,5
220K 100
~901 ~902 V51 'Z4V
3 -@- 0 (5
C163 TI02 0104
~516,B ~)'-2~S[~19~0~6--~~--~
~ 151 ~
150 m- RI80
R170 ; >::; ~ ~: 12K
Tl01 R165 ~ I ~ 4!0 __ -,e!. &: ~ 0102
R146 51 22K I'-' ~ i 0;~~,S. ,~
r--k--ll -v v ~:: g f--
R141 w -- '.0 VR136K6,_ Lo! 0- , __ 5_!1Q_ ~~ en Z
3.9K ~ 14ICl0';. IIV'cBc__ ..J ,- I ~ "
r r--<-' v- R172 i':'lil~;,'
OL:00 -;:i I-'A733 fl7______ R167 470 n: a: (')
co L,.,~~-".- t UK ~J+--TI03 > Dl03 ~
VR142 urI. ~5! R164
o '--,f'vVI-'-I---t--' 5 1 2 S [1906 Dl 05
1VR143 -:1 '~?Dl0l lC119 150
11 OK. CD ~ 6VB _ 22n 0+: ~~673
"147 i:S R261 ~R266 )T202
51 C126 5P n IC201 B \1--..
~1~~ ~~ SITO' (136 150 en '" 51 ~~2SCI906
R13e '--- l~R_I:~~'l' 122n &! R629 0202 U
GTIJ O-~N!lv--t---.---I~o. ~8 _~7~_ _ __ >::; ~
(TIN 12K -lIlOL R153 33 VARn-=, ~ ~ 1* : 8 g: <X R442
RIl8 t1'E_~1"2 C137 g;: y en ~: '" z ~
51 27P '33 + 101' 2: Y-~r~ Lp6s -~ [j[;:; ::;: 51
~ ~ ~ VOLT R271 '~ ~ ~
-z: R162 -= ~-'2V <X N 470,__ ~ g 0203
6RB R262 R267 ~ T20l
1[201 7 o-f~III-II~51YV-~>l):2~si[i219.Qj06L_j~0~2~05~J
R254 ~+5V 150 R434
~2VO-4---4-~+-~--~--------------------___'------'--C6~RBV-+---~--~~------~_
Tl04 R401 6R8
2SA1029[ 0405
R1B2 8:.)
150 ~
R1B9
1iKl330 VR188 R190 330
T1 05-.L'?\
2SA1029[ )
R232 lMF
R 23B 51
6RB
0410 R435
1.7
R428
51
0411 f~
lK
0412 [25~ + 101'
0408
0406
T201 R24551 ~lllQ
P429
R240
39K 1[201 R261
~ f'..A7l3 7 R262
v8~f 11 4 f----,
V\I<RJ",'4.;_2+--++-_J1 I 20201 1[257
~ ~"6V8 12zn
[248SP /l
~~ sm- [258
: VR253 ~-4: 122n
R2'~ 13_ ~c_j.
51 0407
D201 FD333
R187 150
10K
R276 /v
ik)
150 ~
11-- R283
2 lK.~ 330
DC VR282
TRIG R284
330 R280 150 ff']
R246 51
R243 39K L-
R277 6RB
+5V
T2011--
tt~
27P 33
n[259
101'
r-r-r+-
r-t-----+-t-t--,
2SC 1815Y~ ~ ~ R430 ~ 14 6
T405-( 1[4'.' R-f--
12K o· to-[
4013 ~ I-
R43B ,1, I!±-
22J h. R431 '--"I R423
(! ( '(l 12K 33K
T404
2S[IBI5Y
C42,1,,1 C420 IOl'f 221
TRACE ROTATION
VP~f71
+5V~-12V
5K 270 ITmJ
-WTO 0604
l 1 2 1~
R417 [1(4021
R427 10K 14011 )
47K ~ R't14 4¢
10K l'i:f12
T403
2SA1015Y R437 22K
~~5: R425
33K
R404 22K
r+;;;EXTl
l~l I I
.= S6011
L:;Y_..J
R424
220 R407
22K
T402 R41~r
2SA10~j5 10K 47K
R426
47K L.. R416 1
10K R260 L_--------__,j~MI-{)-12V
6R8
TZ05 2SA1029[
H-BOARD I
I LED 0802 I H-OEI DELAY-BOARD
TRIGGER GENERATOR CIRCUIT ~ <D+5V SWEEP DELAY
I -_=
R664 R667 6R8 R826 10K
+12V +5V I
1[659 6R8
22n [665 I [668 I
22n I 22n -= R825
I K
TO 1[701 I 2S[1815Y
13 ) T~) R824
I +5V
R661 .~ [8IJ
I ~T801 R823
[557B 390
[8§1J I 101'
T613
2SAI029[ I R822 R821
0607 R654 47K 47K
3K3 I DELAY
R652 G-BOARO MOD N~801
-12V I
51 1 (66~ 22n
0801
1[658 ~~
22n I
I 5 ("\.
+12V R6998 100
I o 0 [818 ~ h
VR6999 R6995 R6996 + 101'
R815
-..J +5V I (801 47P R820
lOOK 00
I -12V )u_7 6R8 10K
.::-- S~ R817
1fJ I 1i' 10K
1 R679 H 6
10K T 22n I h 1[802 2 3 14
-12V W ~g_ ..___j_ 74122 5
II' 5 I 4,-
ffiIjl:8]3] I 820P ~14 I 1[801
7474 [816
TO 1[704 06 I 101' 5 H~ ~I::" ~ l22n
5 12 I 7
IOn lOOP 6R8
W6r~ 13 I
5V6 I 1001' 5 ) ~~ 1 1[801
IE@ 7474
I II'
0602 [~O~ 9
5V6 1m S o---r,..,
[689 I
22n I 10m S 0 [~6
(EXT X-SIGNAL) TO S'wEEP SIGNAL ~
0603 ffilli~~ I
R808
+24V R6006 I 1.5K (~ < <
J lOOK I lB::illl I H-OEI lB::illl
t------D iVR80~9~;J 0707 R731 1[702
+12V [6003 R6002 ~ sso: I L_~~~ ___ ~
+ lOP IK l81l2JtZI X-Y R810 l5K z~m325Y
lOOK 35V R6001 I MULTIPLIER CIl
[6990+35V 33 o
-12V :J"
AUTO TRIG VR6999 lOP I R811 -= ~
SWEEP
SENSI TIVITY -VR698 DIP TO R961 FROM R714,VR639 TO R410 22K S
AD) [[[TIC) Ill'ZIl Q) I Q)
(REAR PANEL) M'
....
I ()
01 H-BOARO
SWEEP GENERATOR
+5V
+5V i) (~ D
R732 22K {'l8]T
Rn4 UK
~1414 RnB R731
~ 2 6 6RB 51 ~3 R733 51 "" H
c- uuu 1[701
68f'H 2~
3 7474 J_ - ~2~2n
- 0707
~~ R725 11[702 J +1
10K [74001
[n1 _ ,__
22n'T - .tV V
T704 ) t--
- 25[181i7 BLANKI
12 9 R734 TO 0-
11 8 ~
0706 ! 071213 10 51n
R729 UK
R701 R735 :-Fr--
51 10 :-~-[~
~ 9 R730 10K [736 22n 220
070 1[701
~ 7474 8 hi 0711 I
~ '~ ~
R717 h~o 705 R71910K R737
10K 0708
6RB
1 K VR738
~ 5K R739
- R716 R742 0709 0710 H
[7~~ [740 5.6K I ~
22K L-----....lQ0 ~P [747
47P SWEEP LENGTH A8J T 47P
R702 '" H
750 oro,- ""
0702 r-r- r ------ --- - - -
H~ R743 X-FINAL AMP
R711 [l~ 100 ?,11""," I
100P
= 1 3 5 47Kq01 47P [955 22n
1[704 ~ r
8 74122 = [744 = ~EE~OR EXT-X
13114 25[1B15 27P R961 6B
11 R6001 JL..L
R710 ( L...,
m_;Z+ 10K IillJ R950
R709 [8QQ)- ~ I 12K R952 R956 51 tv ) R960 VR"~r T907
[Eg~
27K R714 V 10K T906~ (Q25[
R745 100K I VR951 l 10K /
100K 25[1815 _ R963
,__ X-PO/, R~~9 4K7 t--
[7071 R746 I 10K [i(.:c_
100K 1 I 68
1BIl@ I r 2~n
R708 2 I I
I
''" I 6R8 1[703 [40111 I R964
I I R957 R958 0905 1B2 r
I 62K 2K7 IN4148'
I , riT3 I
~)+5V I 401
[j@] R930 TO B-BOARD T903
REb01
T614
NG PULOE BOARD R552
005 Z-MOo INPUT
m T905 FROM _A_/\_
I8ID]VR91B FROM HOLD OFF
--------------
[690 [6B9
R974 6R8 -------------v~~----------------_O+24V
5601
~V=R9=6~5~~~~==~';X~10~~----~~~---+[RTD2
I
1K ,
,
1-=----:: .: ;;;0-;:; -;3;1";:""
5301 B I I
I _ BEAM FINDER : I
I - LENGTH adj : I
I L +-_.J
,---- ---- - -- J
22n
+24V
Y-BOARD
VERTICAL FINAL AMPLIFIER
Rll7 10n
TO (RT 03
3l Rll4
2 4K
C3l0 1 }'VCll1 e Lll8
100P ; 12/AH
r- (RT G 2
t. +140V
VR332 0 ~Ll39
100 ~ 121'H f7n
.ri, R335 = 250V
2 4K
1MHz AOJ
Rl36 10 ;~ ~ r_~R~l2~0~1~0--~T~3~05~4<i------------------------~r---~~~----'
n01 n03 ,\,
~ R301 68 _fy"): \l..;BF458
FROM 0104[ill (J'""').----.'V'I'-'------<~~Ir------r----I--------,.---K't:o 2S(1906 '-'
25(1906 ~616
R30l 2.4K
R304 51
R305 51
R306 24K
Rl02 68 (i ~ n02
FRO M 0105 ClTI a;(i;)}----II."I'-----'t
- 2S(1906
R307 470
Rl08 470
ad)
R833 6R8 +5vOTIJ~---------------r--~~----~
R309 (312
6R8 2P
(311 + C310 22nL 101'
IIC803] (4011 )
~ n04 R321 10 Wi n06
~:"'2S;:C1-'-90""'6 -__J'----.J___:~v:_:_-_f{~)BF4 58
11
12
R313 51
R314 51
r- F~-, I r--~ I
I 0301 R31 Ol021
I I
I 680 I R322
~~ ~~~N-------~
L JI 390
~n~ '/ b
FINO~
ltis
R319 22K
R329
TO CRTD4
[JK]
,J
RESET
R318 22K
R317 10n
R323 R325 R327
68 39K 33K
R324lC3261C328 68 47n 10P
--------------------------
CALIBRATOR
15K
R831
C-BOARD
R829 270
R830
15K
1KHz 0.2V
.ri.
CALIBRATOR
13
C836 1n r---------+-------4
R837 39K
1 KHZ ADJ
SINGLE SWEEP
R713 150
+5V illO (t;)..-----+-----~
B-BOARD TIMEBASE GENERATOR
1
5901 [9401
-----1------ ----I
¢ 0::::0 R,32 VR938
1K n"1 10K
010 TIME BASE
I 0 0 AD)
010 0 0 R937
I 12K
010 0 0 T902
I T901 ~1015Y ~2SC1815Y
010 0 0 c9341
I .,6'
010 0 0 R931 ;<ci5~F R935
1 100K 22n 47KF DBB2
R912 10KF R924 51
,- -- -------
..,. il ~l B~2 , VR936 100K
R913 20KF I b "" , SWEEP VARIABL
~I ) ~r ~r ! T903 L __________
R914 49.9KF I ~9 = (~BB2
100KF ~ : > ! 0901
R915 1N4148
R930 =
~ > ! ! UK
R916 200KF I +24V R929 68K
~I ) ! ~R733 H 1
R917 487KF I R943 6R8
I ! [ 928 +12V
~ r-------- 1 --( :> 1 OOP I
~ LHOLD OFF AD :C92~ + +zl1 C944
I --- 'VR918 101'
~- 014 047"( :>
I ~1_J_QQl<_J 35V
I ) ~9~ =
I
I ( R919 W4 R927 51 W ~
18K j~ T904 T905
I 't:! 25[1815Y 25C1815Y
I ( f---- 3SV ( II VC926 0902
~ ( (925
I = 100P / 2/22P 1N4148
~ I ( ( ( O.5I'S R941 12K -<')R739
AD) HB
~ r---
~(
,""OF' = R942
0 ) o 51
22n 1001'
0 ) ;J
0 :> J~ ~
~ BS
0 0 0 ) 1-----
01I'S-----?
010 I 010 I 010
I 010
010 I
~ [7470--------'
S403~
"
I
"
R904 S6K
1mS---
R90S 82K
R906 150K
R907 220K
R908 470K
R909 7S0K
0.1S---
(903 +
1001' +24V
-...j ,
00
G-BOARD TRIGGER AMPLIFIER
X-E XT SIGNAL 0603 R641 100
TO S6o1---+.-----~~~----------------------
XY-o[ - LEVEL X-[ENTER!
. TO S601 -+VR~6!!39M~--;-[:':.j640 22n
Ll5K ~
FROM []L.I)9[]ill:: R601 51 ~1
2SA102~JJ
R607 SKl
R614 8K2
R608 Sl r.;J
~ )T603
2S[S35E 0601
r
S604 TRIG COUPLE
AC ~~50V
DC 0------0
r-------TO 1[704 06
R682 68K
R642
2S[5l?,B 2S[i~5KB L F ~
'\ 7"'i+- [6281
~ ~, [62722n 22n1
T605 '-t_T:"~;,,,O+--+_H_F __ ~ ~ =
~~~6 51 CHl 11 10K 1
<»
~ ~
CH1/2 o
........ w--- ;RIG LEVEL
R68011K I (front panell t-------TO T704
QTI1D <D----+-_jrt--__:.____:.__+I--__::.=:.:_-----t---..-----TO 1[704 08
I I I I I
1 1 I
1
R618 ~ [~~ 33P
R609 680
T615 2S[1815Y
R687 82K
-12V
R686 OK
R685 1M
R603
51
GNO I1!'lIJ '" R605 lK R606
"'~ 3K3
R604 R611
51 680 R610 270
+--
0602
R624 lK
1 I I I I I
I S602 OAUTOINORMAL
R612 51 r.::- T604 2S[5;~
"1-_~ _ _j~02
FROM 0408 []ill crc R602 51 \l.
2SA1029[
R613 5Kl
R617 3K9
R615 8K2
R621 R623
330 3K3
EXT 0-; S60 TRIG SEL
R632 10K
[631 15P
0610 +5V
R683 22K
R684 390
1'[622 ,11'
[637 l22n
'" R635 6R8
+24V[]]I] \¥Q}-------,Ni-~-------------__j
-12V r"TI7.'51 "" L638 65i"H ~ <J.)--- rn-o-
[633
1_ 22n
[Hl TRIG 0
TO 1[402 []ill ::cr _:6::,0.j:.4 -=- J
[H2 TRIG
TO 1[402 ~Q)Gl}--------~0~60~5-----------------j
LE 0
+12V
MEMO
D-BOARD
POWER SUPPLY &CRT CIRCUIT
S-BOARD COMPONENT CIRCUIT
+24Y
R451 560K
-llY cnl
~W.H~IT:..:E_,..-+- __ --! --t..::1 0'-"0"-y---!,..------ _ocRT14
~P~AC':'3~-6:.c.J::.:Y-4.-----!----.1-----'l-----------_OCRTI -1 9KY ;;.OR.A.:;N.::;GE=--_ PAC 2
llY
CT ,-
o I
_____ ~0l.. __ _j
R446 220
BLACK -B4Y-COMPONENT TEST POWER
BLACK R504 10K
+5V
COMPONENT TEST JACK
-B5V
0509 050B
0505 0506 0507
lN4007'B
C515 22i' IKY
I
I BROWN
CD
o YSI
L _
MONO/DUAL
R551 2K2
[2]]
CHOP BLANK FROM lC4011121
1C555
120P
R581 56K
YR5B2 50K ASTIGM R58l lOOK
f"---l---<J "iY
MEMO