Professional Documents
Culture Documents
Service Manual
Revision A
Trace MS Plus™ and Xcalibur™ are trademarks of Thermo Finnigan Corporation. Microsoft® and Windows NT® are registered
trademarks of Microsoft Corporation. Pentium® is a registered trademark of Intel Corporation.
Technical information contained in this publication is for reference purposes only and is subject to change
without notice. Every effort has been made to supply complete and accurate information; however, Thermo
Finnigan Corporation assumes no responsibility and will not be liable for any errors, omissions, damage, or
loss that might result from any use of this manual or the information contained therein (even if this
information is properly followed and problems still arise).
This publication is not part of the Agreement of Sale between Thermo Finnigan Corporation and the
purchaser of a Thermo Finnigan system. In the event of any conflict between the provisions of this
document and those contained in Thermo Finnigan Corporation’s Terms and Conditions, the provisions of
the Terms and Conditions shall govern.
Reference to System Configurations and Specifications supersede all previous information and are subject
to change without notice.
The products of Thermo Finnigan are produced under ISO 9001 accredited quality management systems.
Australia: Thermo Finnigan • P.O. Box 239 Rydalmere• Unit 20, Metro Centre • 38 – 46 South Street • Rydalmere, N.S.W. 2116 • [61] (02) 9898-9000
Austria: Spectronex GmbH • Bischoffgasse 26 • A-1120 Wein • [43] (01) 813 9993
Belgium: Thermo Finnigan BVBA • Groenenbrogerlaan 84 • B-2610 Wilrijk (Antwerpen) • [32] (03) 825 06 70
Canada: Thermo Finnigan Canada • 5716 Coopers Avenue, Unit 1 • Mississauga, Ontario • L4Z2E8 • [1] (905) 712-2258
France: Thermo Finnigan France SA • Parc Hightec Sud • 12 avenue des Tropiques • Z.A. de Courtaboeuf BP141 • F-91944 Les Ulis Cédex •
[33] (01) 69 18 88 10
Germany: Thermo Finnigan Analytische Systeme GmbH • Boschring 12 • D-63329 Egelsbach • [49] (06103) 408 0
Italy: Thermo Finnigan Italia S.p.A. • Strada Rivoltana • I-20090 Rodano (Milano) • [39] (02) 95 059 1
Japan: Thermo Finnigan K.K. • Nishi-Shinjuku Toyokuni Building, 3rd Floor • 2-5-8 Hatsudai, Shibuya-ku • Tokyo 151-0061 • [81] (03) 3372-3001
Japan: Thermo Finnigan K.K. • Esaka Grand Building • 2-3-1 Esaka-cho, Suita City • Osaka 564-0063 • [81] (06) 6387-6681
Netherlands: Thermo Finnigan BV • Druivenstraat 33 • NL – 4816 KB Breda • [31] (076) 587 8722
P.R. China: Thermo Finnigan China • Room 912-916, Ping-an Mansion. • No. 23, Jin Rong Street • Xi Cheng District • Beijing 100032 •
[86] (010) 6621 0839
Spain: Thermo Finnigan SA • Acer 30 – 32 • Edificio Sertram – Planta 2, Modulo 3 • ES-08038 Barcelona • [34] (093) 223 0918
Spain: Thermo Finnigan SA • Avenida de Valdelaparra 27 • Edificio Alcor – Planta 2a • ES-28108 Alcobendas (Madrid) • [34] (091) 657 4930
Sweden: Thermo Finnigan AB • Pyramidbacken 3 • S-141 75 Kungens Kurva (Stockholm) • [46] (08) 680 01 01
Switzerland: Spectronex AG • Rotterdam-Str 21 • Basel • CH-4002 • Switzerland • [41] (061) 331 6020
United Kingdom: Thermo Finnigan Ltd. • Paradise • Hemel Hempstead • Herts HP2 4TG • [44] (01442) 233 555
U.S.A.: Thermo Finnigan LC and LC/MS Division • 355 River Oaks Parkway • San Jose, CA 95134-1991 • [1] (408) 965-6000
Notes: The country code is enclosed in square brackets [ ]. The city code or area code is enclosed in parenthesis ( ). For countries other than the U.S.A.,
when you are dialing from within the specified country, dial the 0 of the city code. For countries other than Italy, when you are dialing from outside the country,
do not dial the 0 of the city code.
1. Introduction ..................................................................................................................................1
Purpose..................................................................................................................................1
Audience ...............................................................................................................................1
Purpose
This manual has been written in order to offer assistance to field
service engineers (Customer Support Engineers) whilst working on the
Trace MS Plus Plus. Several aspects of the servicing of the Trace MS
Plus Plus are covered:
• Preventative (Routine) Maintenance.
• Fault Finding and Corrective Maintenance.
• PCB Tuning and Set-Up Procedures.
This manual is not exhaustive and further information can be found on
the TechWeb at http://www.thermofinnigan.co.uk/techweb. The
TechWeb is covered in more detail later in the manual.
Audience
Service Managers and Service / Installation Engineers.
It is intended that in the future the TechWeb will be the primary source
of technical information for the Trace MS Plus. The TechWeb is the
Product Support department’s technical support web site. It can be
found at http://www.thermofinnigan.co.uk/techweb.
After entering your name you will be taken to the Home Page.
Principle Features
The TechWeb is constantly being updated and improved and the latest
information can be found here in Technical Bulletins.
Technical Bulletins
Here you will be able to find the latest technical information released
from the factory. They will often give good hints and clues to setting
up systems and software that may not be included in the manuals.
Spares List
Links
Products
General
The TechWeb is constantly being updated but if you have any queries,
comments or spot any errors then please contact
product.support@thermofinnigan.ms.
WARNING: Check with the customer whether the instrument has been
exposed to any radioactive, toxic, or biohazardous materials. Suitable
precautions and/or decontamination may be necessary before the
preventative maintenance can begin.
Note: For more information on operating the rotary pump, refer to the
Rotary Pump User Manual, supplied with the TRACE MS PLUS system.
The rotary pump oil should be replaced at least once every six months. At
the same time as you replace the oil, you should also consider replacing the
foreline trap pellets; see Replacing the Foreline Trap pellets later in this
Section.
WARNING: Always wear gloves when changing the oil. Avoid contact
with the oil because it may contain dissolved residue from analyzed
samples. Observe appropriate disposal requirements when discarding the
old oil.
To replace the rotary pump oil, you will require E2 M1.5 Rotary Pump Oil
(P/N FM101067 (4ltr) or 5660010 (1ltr)). Follow the procedure described
below.
Foreline trap
Vent
Oil filler plug
Drain plug
Viewing window
Caution: Do not allow the pump to run for more than a few seconds
without oil.
8. Reinsert the drain plug and fill the pump reservoir up to, but not
above, the upper mark.
9. Reinsert the filler plug.
10. Gas ballast the new oil in the pump; for details, refer to the Rotary
Pump User Manual.
11. Right-click on the TRACE MS PLUS Server icon in the Windows
task-bar and choose the Pump menu option to pump-down the system.
Note: For more information on operating the rotary pump, refer to the
Rotary Pump User Manual, supplied with the TRACE MS PLUS system.
Absorbent pellets in the foreline trap help maintain the TRACE MS PLUS’s
high vacuum.
The rotary pump oil should be replaced at least once every six months. At
the same time as you replace the oil you should also consider replacing the
foreline trap pellets.
To replace the foreline trap pellets, you will require Absorbent Pellets.
Follow the procedure described below.
1. Loosen the foreline trap end cap by a 1/4 turn.
2. Right-click on the TRACE MS PLUS Server icon in the Windows
task-bar.
The server menu is displayed.
3. Choose the Vent menu option to vent the system.
4. Remove the end cap from the foreline trap to gain access to the
inner baskets of pellets.
5. Open a new can of pellets. Bake them in the GC oven for 2 hours at
250 ºC. Allow the pellets to cool and then place them in the foreline trap
basket.
6. Reassemble the foreline trap.
7. Right-click on the TRACE MS PLUS Server icon in the Windows
task-bar and choose the Pump menu option to pump-down the system.
Reference vial
Cajon union
6. Unscrew the knurled nut (Cajon union) attached to the tube of the
reference vial.
7. Remove the vial from the instrument.
8. Top up the vial, using approximately 400µl (but no more than
0.5ml) of reference compound.
4. Remove the four hex nuts and remove the mounting plate.
In CI and combined sources, the source block comes away with the
mounting plate.
5. Slide the magnet assemblies off, noting which is top and bottom.
The filament is always adjacent to the North Pole.
6. Remove the filament and the trap by removing the fixing screws.
Caution: Handle the filament with particular care and store it safely.
7. For EI sources, remove the source block on the ion exit plate and
then remove the exit plate from the block.
Removed thermocouple
12. Dismantle the ion lens assembly, noting the relative lens positions
and taking care not to damage the ceramics or spacers.
Caution: You are advised NOT to remove the source heater elements for
routine cleaning, because the constant bending of its legs will eventually
lead to fractures. For this reason, ultrasonic cleaning of the repeller block is
not recommended.
All metal source components, excepting the repeller block containing the
heater element, can be cleaned as described below.
WARNING: When using solvents, all precautions for their safe use and
disposal, detailed in the appropriate National Health and Safety regulations,
should be followed.
Caution: Never try to clean the filament: this may cause damage to it.
Source Filament
Follow steps 1 to 3 of the procedure for dismantling the source for cleaning.
Remove the filament fixing screw and take out the old filament. Carefully
fit the replacement filament and refit the fixing screw. The new filament
self-aligns.
Removed filament
Source Heater
To replace a source heater:
Follow steps 1 to 10 of the procedure for dismantling the source for
cleaning.
Slacken off the two grub screws holding the heater element (only one on CI
and combined sources), and then carefully bend down one leg and retract the
used element.
Fit a new heater element into the block and carefully bend the leg back up
into place.
Tighten the grub screw(s) until the element is held firmly in place. Do not
over-tighten.
Thermocouple
To replace a source thermocouple:
Follow steps 1 to 3 of the procedure for dismantling the source for cleaning.
Remove the thermocouple fixing screw and then take out the old
thermocouple.
Carefully fit the replacement filament and refit the fixing screw.
Caution: Do not over-tighten nuts and screws because this can damage the
ceramic spacers.
R.F. generator
adjustable
capacitors
R.F. test
points and
adjustment
Caution: The Glass capacitors on the RF Generator are very fragile and
will break if over-adjusted.
Note: If you have had to adjust the Align voltage, it is likely that the resolution
will have been altered
2 Check that the tuning parameters are set to the normal values (especially
LM / HM Res sliders set to 12.5).
3 Set each tune window so that the bases of the main isotope of each peak
can be seen.
4 The base of the main isotope of each peak should be 1 amu wide.
5 If this is not the case, then remove the protection plate on the Analogue
unit.
6 Using the resolution trimpots, Low Mass Res (R69), Linearity (R76 –
usually fully acw) and Resolution (R68) set the resolution.
2 Adjust the Mass Scale trimpot (R52) until each of the peaks is in the
center of the window.
Communication
Symptom Probable Cause Remedy
No communication with the Loose or incorrectly inserted Check all the cables between the
other equipment cable Trace MS Plus and the other
peripheral systems are connected
correctly and in the right place.
For more information see the
Trace MS Plus Getting
Connected Guide.
PC System crash Reboot the PC
Unsupported communications Use the contact closure to start
the Trace MS Plus
Incorrect firmware version Check that the firmware on the
peripheral devices is latest issue
– you may need to contact the
supplier of the equipment for
more details
A transient break or fluctuations Reboot the PC
in the mains power supply has
halted Xcalibur
A malfunction has occurred in Exit Tune or Xcalibur and reset
the Trace MS Plus electronic. the instrument’s hardware, using
the switch on the rear of the
instrument.
A malfunction has occurred in Try cycling the power to the GC.
the GC electronics If this is unsuccessful then
contact the local Thermo
Finnigan office
The instrument fails to begin The contact closure cable is Check that the cable connected
scanning and Xcalibur displays incorrectly fitted properly. Replace if necessary
the message ‘Waiting for contact
closure’ The wrong instrument is selected Check that the start instrument
as the start instrument in the selected is appropriate for the
instrument method file equipment you are running with
Xcalibur crashes at the start of a Xcalibur has become corrupted Delete the raw data file and
run or the raw data file is corrupted reboot the PC. Reload the
software if necessary
Loose RS 232 cable connection Consult your PC hardware
manual and check the connection
Vacuum
Symptom Probable Cause Remedy
Rotary Pump fails to turn on Pump not plugged in Connect power lead to rear of
instrument
Pump breaker activated Ensure that green breaker on rear
of instrument is depressed
Turbomolecular pump fails to Communications failure See later in this Section
turn on
Turbo pump controller cable Reconnect cable
disconnected
Pumps switch on but instrument Air leaking into system Check all O-rings, especially on
fails to reach “Vac OK” status any parts recently removed
Remove source and refit
Check interface from GC to
source – tighten if necessary
See later in this Section
Foreline Trap pellets need Refer to Section 3.3 of this
replacing – possibly damp manual
Rotary pump oil needs replacing Refer to Section 3.2 of this
manual
Rotary pump gas ballast open Close gas ballast
The rotary pump oil level is low Check the oil level on the rotary
pump. Add oil if necessary –
refer to Section 3.2 of this
manual
GC Column is broken or is Check the GC column has been
leaking at the interface installed properly
Check the ferrules on the GC
interface
Faulty Pirani Gauge Check the Pirani Gauge Set Point
– refer to Section 4.3 of this
manual
Column flow rate is too high Reduce the flow rate
CI gas valve is open or the gas Check the valve and gas line
line is leaking
Tuning
Symptom Probable Cause Remedy
Low or no signal Reference vial empty or nearly Refill Reference vial – refer to
empty Section 3.4 of this manual
Faulty reference solenoid valve Check that the solenoid
energizes when the button is
activated in the software. A
“click” can be heard
If there is no “click” then reboot
the software and try again. If it
still fails, replace the solenoid
Voltage not being supplied to a Check the readbacks in the Per
component Method Parameters table.
Ensure that each value is close to
its Set Point
Air leak on system Refer to Section 4.3 for
information on eliminating air
leaks
Reference gas pipes are blocked Check and clean the source –
or missing refer to Section 3.6 of this
manual
There are problems with the Check the RF Generator Set-Up
quadrupole or pre-filters – refer to Section 3.7 of this
manual
No peaks (or very low intensity Source lenses poorly aligned Disassemble the source, as
peaks) although the source described in Section 3.6, and
readbacks are correct check the lens alignment with the
tool
Air is leaking into the system Refer to Section 4.3 for
information on eliminating air
leaks
Detector voltage too low Increase the voltage on he Tune
page
The source is dirty Clean the source – refer to
Section 3.6 of this manual
The pre-filters are dirty Clean the pre-filters – refer to
Section 4.3 for more information
Tuning (Ctd)
Symptom Probable Cause Remedy
No peaks (or very low intensity Faulty detector supply Check that the voltage displayed
peaks) although the source on the tune page agrees with the
readbacks are correct voltage measured coming from
the supply (Note: Use a High
Voltage probe to measure this)
High mass peaks (502 and 614 Pre-filters are short-circuiting Refer to Section 4.3 for
m/z) are missing information on pre-filters
Tune peaks show signs of Bad connection in the ion source Disassemble and re-assemble the
breaking up source – refer to Section 3.6 of
this manual
Pre-filters are short-circuiting Refer to Section 4.3 for
information on pre-filters
Malfunction in system Contact Product Support for help
electronics
Changing a tuning parameter has Source plates not connected or Check the readbacks in the Per
no effect on the peaks or there is short-circuited Method Parameters table.
a delay before peaks are changed Ensure that each value is close to
its Set Point
Pre-filters are short-circuiting Refer to Section 4.3 for
information on pre-filters
Peaks have an irregular shape The source is dirty Clean the source – refer to
Section 3.6 of this manual
Old or faulty filament Replace the source filament –
refer to Section 3.6 of this
manual
Poor tuning Check the End User’s Tuning –
refer to Trace MS Plus Getting
Started Guide
Source lenses poorly aligned Disassemble the source, as
described in Section 3.6, and
check the lens alignment with the
tool provided
Tuning (Ctd)
Symptom Probable Cause Remedy
Large peaks observed at 18, 28 Air is leaking into the system Refer to Section 4.3 for
and/or 32 m/z information on eliminating air
leaks
GC carrier gas bottle is nearly Replace the gas bottle
empty
The system is still moist Leave the system in vacuum for
following a change of column or a little longer
source
High mass peaks (502 and 614 Pre-filters are short-circuiting Refer to Section 4.3 for
m/z) are missing information on pre-filters
Tune peaks show signs of Bad connection in the ion source Disassemble and re-assemble the
breaking up source – refer to Section 3.6 of
this manual
Pre-filters are short-circuiting Refer to Section 4.3 for
information on pre-filters
Malfunction in system Contact Product Support for help
electronics
Excessive ion energy voltage or The source is dirty (ion repeller Clean the source – refer to
ion repeller voltage is required voltage >6.0 V) Section 3.6 of this manual
Pre-filters are dirty (ion energy Clean the pre-filters – refer to
voltage >3.0 V) Section 4.3 for more information
Poor tuning Check the End User’s Tuning –
refer to Trace MS Plus Getting
Started Guide
Unusual Emission, Filament or The source magnets are inserted Check the alignment of the
Source current readbacks shown incorrectly magnets: one should have the
on Tune View "N" pole visible; the other, the
"S" pole. See Section 3.6 for
details of how to disassemble
and re-assemble the source
Electrical problem within the Disassemble and re-assemble the
source source – refer to Section 3.6 of
this manual
Autotune fails even though Poorly tuned resolution pots Refer to Section 3.8 of this
manual tune has delivered an manual
acceptable beam
Tuning (Ctd)
Symptom Probable Cause Remedy
The ratio of Source current to The source is dirty Clean the source – refer to
Emission current shown on the Section 3.6 of this manual
Tune View is greater than 10:1
The source filament has Replace the source filament.
deteriorated Refer to Section 3 for details
Air is leaking into the system Refer to Section 4.3 for
information on eliminating air
leaks
The source filament is poorly Disassemble and re-assemble the
aligned source. Refer to Section 3.6 for
details
The ratio of Source current to The trap is dirty or has a poor Clean or replace the trap. Refer
Emission current shown on the connection to Section 3.6 for details
Tune View is greater than 10:1
The source magnets are inserted Check the alignment of the
incorrectly magnets: one should have the
"N" pole visible; the other, the
"S" pole. See Section 3.6 for
details of how to disassemble
and re-assemble the source
Peaks are shifted from their Poor calibration See TRACE MS PLUS Getting
nominal position Started for details of how to
calibrate the system
Malfunction in system Refer to Section 3.7 & 3.8 of
electronics this manual
MS Resolution
Symptom Probable Cause Remedy
Unresolved peaks at LM/HM Resolution Pots need adjusting Set up resolution pots. Refer to
Res at 12.5 and Ion Energy at 1.0 Section 3.8 of this manual
DC voltages on quadrupole not Run diagnostic as per later in this
being applied section. If the check fails then
contact the local Thermo
Finnigan office
Poor resolution and low Dirty pre-filter or quadrupole Clean pre-filters. If replacement
sensitivity – Ion Energy needs to quadrupole is required then
be >1.5 to get ions contact local Thermo Finnigan
office
Calibration
Symptom Probable Cause Remedy
The instrument will not calibrate Poor tuning Check the End User’s Tuning –
refer to Trace MS Plus Getting
Started Guide
Low sensitivity Refer to Low sensitivity earlier
in this section
The first peak (lowest mass) in Alter the calibration range to
the acquired spectrum is too start with a strong first peak
weak
Using the incorrect reference file Check that the reference file
being used is related to the
calibrant. In most cases this will
be Heptacosa.
Check that the reference file has
not been altered or corrupted and
that the correct masses are in the
file. Reload the software if
necessary
Poor resolution Check the LM/HM Res and Ion
Energy settings on the Tune page
Source temperature is too high or Adjust the source temperature on
too low the TRACE MS PLUS Per
Method Parameters
Air is leaking into the system Refer to Section 4.3 for
information on eliminating air
leaks
Electron energy not at 70 eV Reset value on the TRACE MS
PLUS Per Method Parameters
The quadrupole is dirty Increase the source temperature
to 300°C and leave at that
temperature for 2 hours
Masses are drifting Excessive room temperature Check the room temperature.
fluctuations Check the operation of the fan at
the rear of the instrument – refer
to Section 3.5 of this manual
Masses are drifting High fields Check for other instruments
emitting high fields in close
proximity. Change the
instrument location if necessary
Note: In the event of having difficulty calibrating the instrument, use the
tuning diagnostics to obtain further information. These are covered later in
this section.
Low Sensitivity
Sensitivity is determined by the ratio of signal to noise. Before attempting
a diagnosis, determine whether the problem is due to a weak signal or
excessive background noise by reference to a previously know good
spectrum.
Rising total ion current (TIC) The GC column is bleeding Condition or change the column
baseline.
Air is leaking into the system Refer to Section 4.3 for
information on eliminating air
leaks
Spectral Problems
Symptom Probable Cause Remedy
The spectra are distorted or The scan speed is too low Increase the scan speed so that
skewed there are at least 6 scans across
the peak
Unstable or poor tuning Re-tune the system
Incorrect isotope peak/ratios LM/HM Res settings or Ion Check the Tune page set-up with
Energy setting too high or too the reference gas and set the
low in the tune file LM/HM Res and Ion Energy for
good resolution
Inaccurate calibration Re-calibrate – refer to the Trace
MS Plus Getting Started
Manual for details
The detector is being overloaded If any of the peaks have an ion
count of >2e6, reduce the
concentration of the sample or
reduce the detector voltage
Air is leaking into the system Refer to Section 4.3 for
information on eliminating air
leaks
Source Problems
Note. Refer to Chapter 4 for information on how to maintain and clean the
ion source.
Source not heating. The heater is malfunctioning or Check the heater connection and
is not connected replace, if necessary.
Refer to Section 3.6 for details
Note: If the Trace MS Plus is not connected to the correct power supply
then damage could occur to the system
Rebooting the PC
The most common communication problems with Xcalibur can often be
resolved by simply rebooting the PC.
To safely reboot the PC:
1. Try to save any open data files.
Depending on the nature of the problem, this may not be possible. Any
data that is not saved before rebooting the PC will be lost.
2. Close Xcalibur.
3. Exit from Windows in the usual way, selecting Restart in the window.
4. Reboot Xcalibur.
Polarity Check
Set the following ionization modes and check that the polarities of the
voltages listed are correct and close to the Set Point in the Per Method
Parameters table.
Ionization Lens 1 Lens 2 Repeller Electron
Mode Energy
EI -ve -ve +ve -ve
CI+ -ve -ve +ve -ve
CI- +ve +ve -ve -ve
Note: It is usual for the Product Support department to ask you to run these
checks when contacting them for support.
Column Connections
If press-fit unions are used on the analytical column, examine these first:
Gently tighten the GC interface column 1/16” Swagelok securing nut by a
fraction of a turn.
Gently tighten the injector column-securing nut by no more than a quarter of
a turn.
Depending on the tightness of the column, the response of the air peak
display will be delayed for between ½ and 2 minutes if the leak is here.
Supply Plumbing
Make a particular check in this area if any changes are made to the supply
plumbing after commissioning of the instrument.
If the carrier gas cylinder is near the Trace MS Plus:
1. Close the regulator valve and use a leak-checker (a helium sniffing
electronic device would be most useful) to check the cylinder thread and
all exposed seals and joints of the regulator body.
Set the cylinder regulator to at least 50 psig and repeat the leak checking
procedure from the supply to the GC inlet.
If the carrier gas supply comes from a manifold serving a large number of
systems, you can only make an effective check for leaks in new plumbing.
Once you have done this, examine whether the leak might be from another
source. Finally, if necessary, obtain a gas cylinder and plumb it directly into
the system.
GC Internal Plumbing
8. Undo the nuts holding the reference pipes into the instrument and
withdraw them fully.
9. Remove the source and slacken the screws that hold the source interface
bracket.
10. Remove the Headamp and Photomultiplier assembly from the
instrument.
11. Disconnect the 3 cables to the detector and undo the 2 screws fixing the
detector to the Housing.
12. Remove the detector assembly and store in a suitable manner.
13. Undo the 4 screws holding the Analyzer assembly to the Housing.
Support the assembly as the screws are removed.
14. Carefully withdraw the Analyzer assembly from the Housing.
3. On each Pre-filter undo the Allen grub screw on the Pre-filter and
carefully remove the Pre-filter from the assembly.