Alarm Pmax: Upper pressure limit in the the Alarms window Body Wt: ETS: fControl: FiO2: fSpont: fTarget: P0.1: Patients ideal body weight, in kg (IBW) Expiratory trigger sensitivity Mandatory breathing frequency Fraction of inspired oxygen Spontaneous breathing frequency Targeted total breathing frequency Airway occlusion pressure (GALILEO only) Tube resistance compensation Patients inspiratory effort, either flow or pressure, that causes the ventilator to deliver a breath.
Startup
Alarm Pmax: 45 cmH2O (50 cmH2O in COPD) Body Wt: %MinVol: IBW (+10% if using HME) normal: 100% COPD: 90% ARDS: 120% others: 110%
TRC: Trigger:
i
Trigger: ETS: TRC:
add 20 if body temperature > 38.5C 2l/min 25% (40% in COPD) on (100%)
HAMILTON MEDICAL AG Via Crusch 8 CH-7402 Bonaduz / Switzerland Phone: (+41) 81 660 60 10 Fax: (+41) 81 660 60 20 Visit our websites: www.hamilton-medical.com www.IntelligentVentilation.org
624072/00
Adjustments / Weaning
fSpont = 0 patient is fSpont > stable*** > 60 min AND fTarget*+10 b/min (Longer period for PaCO2 < 40 mmHg OR hard to wean patients) (45 mmHg on COPD) PaCO2 > 45 mmHg NO NO (50 mmHg in COPD) NO OR P0.1 > 3 cmH2O**
YES
YES
YES
Escalation
%MinVol plus 20
De-escalation
%MinVol minus 10
(%MinVol not lower than 70%)
NO
YES
?
Target value on the ASV numerical screen ** Only with the GALILEO and on P-trigger *** stable is: fControl = 0 AND PaCO2 45 mmHg (50 mmHg on COPD) AND P0.1 3 cmH20 *
Pinsp 10 cmH2O
NO
Push to wean
%MinVol minus 10
(%MinVol not lower than 70%)
YES
consider extubation