You are on page 1of 2

Crude Oil On-Line Analyzer

The new MOD-4100 represents a breakthrough in crude oil on-line analysis by determination of Crude Density, Concentration of Salt, Hydrogen Sulphide and Water by one single analyzer system. The analyzer comes installed in outdoor stainless steel enclosure and equipped with an integral sample conditioning system.

System Advantages
On-line analysis the quality of crude oil is important because this allows the crude to be evaluated for potential to corrode equipment and pipeline. Furthermore, as crude oil is expanding in the world's energy balance, there is an increasing need to measure with accuracy all main quality parameters such as density, concentration of salt, hydrogen sulphide and water. The sample probe is extracted anywhere along the pipeline from well head to refinery and rapid on-line analysis allows prompt corrective action when unacceptable levels of quality parameters are present. The Analyzer System is supplied on a basis of "Package concept" - factory inspected and tested, ready for immediate installation on-site. No analyzer shelter is required, and the low maintenance requirements reduce ownership costs to a minimum.

On-line Analytical System is a complete equipment set, which allows performing the following tasks in real time mode and in field conditions: Continuous sampling crude oil from pipeline or process vessel, its filtration and separation from solid particles Sample preparation for analysis of physical and chemical parameters according to specifications of analytical devices (incoming gas temperature, pressure, flow rate etc) Analysis of critical oil parameters, which are necessary for assessment of its quality and suitability for processing, transportation and use Transfer of analysis results to customers control room by means of electronic communication

MOD-4100 Crude Oil Analyzer Specification


Performance
Measuring Ranges (selectable)

Density
0-3000 kg/m3 0.1 kg/m3 1.0 kg/m3

Salt
0-1000 mg/L

H2S
0-1000 ppm

Water
0-4%

Repeatability Accuracy Response Time

20 mg/L 50 ppm ASTM 100 ppm D3230 5 minutes

0.02% 0.1%

Sample Requirements
Sample Inlet Pressure Sample Return Pressure Sample Temperature Sample Flowrate 4-12 Barg Optional sample booster pump p/n 4100-01-14 0-2 Barg Optional sample recovery system p/n 4100-01-15 10-60 oC Optional heat exchanger p/n 4100-01-16 3 L/min

Operating Environment
Ambient Temperature Humidity Storage Hazardous Area Classification Area Classification 0 to +40 oC Optional design for -40 to +70 oC 0-95% RH -40 to +70 oC

ATEX Zone 2 Class 1 Division 2 Group C&D T4

Installation Requirements
Power Supply Power Consumption Instrumentation Air Solvents requirements Nitrogen 100-130 VAC 50-60Hz 200-260 VAC 50-60 Hz 2000 VA on 220 VAC 5-6 Barg clean instrumentation air 5-10 L/min Xylene and Alcohol mixture according to ASTM D843 4-10 Barg 2000-5000 cc/min

Data Acquisition
Analogue Outputs Communication Local HMI 4-20 mA isolated outputs RS-485 Modbus, TCP/IP, Fieldbus, Profibus (optional) Built-in maintenance panel with graphical display (optional)
The Modcon logo is a trade mark of Modcon Systems Ltd. The right is reserved to amend details given in this publication without notice