Professional Documents
Culture Documents
8-Port Sampler
OCT-1
for the TOC-V Series Total Organic
Carbon Analyzers
User Manual
SHIMADZU CORPORATION
PROCESS & ENVIRONMENTAL
INSTRUMENTATION DIVISION
KYOTO, JAPAN
OCT-1
Copyright Shimadzu Corporation 2002. All rights reserved, including those to reproduce this publication or
parts thereof in any form without permission in writing from Shimadzu Corporation.
Read and understand this manual before operation. Save this manual.
Information in this publication is subject to change without notice and does not represent a guarantee on the part
of the vendor.
Any errors or omissions which may have occurred in this publication despite the utmost care taken in its
production will be corrected as soon as possible, but not necessarily immediately upon detection. We appreciate
notification of any errors or omissions.
Shimadzu does not have any responsibility for errors or injuries resulting from following the instructions in this
manual.
Windows is a registered trademark of Microsoft Corporation.
OCT-1
Preface Introduction
Thank you for purchasing the OCT-1 8-Port Sampler.
Be sure to carefully read this documentation before operating the OCT-1 8-Port
Sampler for the first time. Read the Operational Precautions before operating the
sampling unit. Follow the procedures in the manual to avoid compromising safety.
CAUTION
misuse;
external factors;
Post-Sale Service
If any problems occur with the OCT-1 8-Port Sampler, inspect it and take appropriate
corrective action as described in Section 1.4.2 Troubleshooting of this manual. If the
problem persists, or symptoms are not covered in the Troubleshooting section, contact
your Shimadzu representative.
iv
OCT-1
OCT-1
Notation Conventions
Precautionary Conventions
Notation Conventions
Precautionary Conventions
The following precautionary conventions are used in this manual.
Convention
CAUTION
Note:
Description
Indicates the possibility of slight to moderate
injury or equipment damage.
Supplementary Conventions
In addition to the precautionary conventions, Tips and Reference are used to provide
additional information.
Convention
TIP:
Reference:
vi
OCT-1
Description
Indicates a particular technique, alternative
means of operation, or useful advice, etc.
Indicates a reference location for more
detailed information.
Table of Contents
Preface Introduction............................................................................ iii
Product Warranty and Post Sale Service .......................................... iv
Product Warranty ............................................................................... iv
Post-Sale Service................................................................................ iv
Hardware......................................................................................... 1
1.1
Overview............................................................................................ 2
1.2
1.3
1.3.1
1.3.2
1.4
Front View........................................................................................... 4
Rear View............................................................................................ 4
Maintenance....................................................................................... 5
1.4.1
Replacing the 8-Port Valve Rotor .......................................................
1.4.2
Troubleshooting ..................................................................................
1.4.2.1
Error Messages..........................................................................
1.4.2.2
Troubleshooting ........................................................................
1.5
5
7
7
8
1.5.1
Specifications ...................................................................................... 9
1.5.2
Standard Accessories........................................................................... 9
1.5.3
Consumable Parts .............................................................................. 10
1.5.4
Maintenance Parts ............................................................................. 10
1.5.5
Installation ......................................................................................... 10
1.5.5.1
Parts Inspection ....................................................................... 10
1.5.5.2
Installation Site ....................................................................... 10
1.5.5.3
Installation Procedure ............................................................. 11
Operation....................................................................................... 15
2.1
CS Operation.................................................................................... 16
2.1.1
TC Blank Check Measurement .........................................................
2.1.2
8-Port Sampler Measurement............................................................
2.1.2.1
Sample Preparation .................................................................
2.1.2.2
High Sensitivity Using the 8-Port Sampler.............................
TOC-VWP
16
16
16
19
Table of Contents
2.2
WS Operation.................................................................................. 21
2.2.1
Blank Check Measurement................................................................
2.2.2
8-Port Sampler Measurement ............................................................
2.2.2.1
Sample Preparation .................................................................
2.2.2.2
High Sensitivity Using the 8-Port Sampler .............................
2.3
CP Operation ................................................................................... 25
2.3.1
Instrument Setting Wizard.................................................................
2.3.2
Sample Table Setting for Calibration Curve Measurement...............
2.3.3
View Menu ........................................................................................
2.3.4
Sample Containers .............................................................................
2.3.4.1
ASI / 8-Port Sampler Vials......................................................
2.3.5
Maintenance Menu Description.........................................................
2.4
TOC-VWP
25
26
26
27
27
29
WP Operation.................................................................................. 37
2.4.1
Instrument Setting Wizard.................................................................
2.4.2
Sample Table Setting for Calibration Curve Measurement...............
2.4.3
View Menu ........................................................................................
2.4.4
ASI / 8-Port Sampler Vials ................................................................
2.4.4.1
ASI / 8-Port Sampler Vials......................................................
2.4.5
Maintenance Menu Description.........................................................
21
21
21
24
37
38
38
39
39
41
1
1Hardware
This chapter provides an overview of the 8-Port Sampler, its construction,
maintenance procedures, specifications, installation instructions and the
standard parts list. The material covered in this chapter is relevant to all models
of the TOC-V series.
1.1
Overview
Provides an overview of the OCT-1 8-Port Sampler.
1.2
Operational Precautions
Details the precautions necessary to ensure safe operation.
1.3
1.4
Maintenance
Describes the maintenance and troubleshooting procedures for the OCT-1 8Port Sampler.
1.5
Reference Materials
Includes the specifications, Standard Accessories List, Parts Lists, and the
installation procedures for the 8-Port Sampler
1.1 Overview
1.1
Overview
The 8-Port Sampler (OCT-1) is a special accessory for the TOC-V Series (hereafter
referred to as TOC-V). When combined with the TOC-V, the OCT-1 provides a fully
automatic measurement system designed for use with sample containers provided by
the user.
This manual describes the handling of the 8-Port Sampler as well as the related
operations of the TOC-V. For descriptions of the TOC-V operations unrelated to the 8Port Sampler, refer to the appropriate TOC-V User Manual.
Read this user manual carefully before using the OCT-1 8-Port Sampler to ensure its
proper use. The manual should be conveniently stored for future reference.
CAUTION
Safety may be compromised if the 8-Port Sampler is used in a fashion
other than that indicated in this user manual. Pay particular attention to
the CAUTION sections to ensure safety when using the sampling unit.
OCT-1
1.2
Operational Precautions
Take the following precautions when using the OCT-1 8-Port Sampler.
CAUTION
Hand tighten the various connectors to the 8-port valve. Use of a tool
may cause deformation and leakage in the valve. <Part damage
prevention>
To clean the OCT-1 8-Port Sampler, wipe the surface with a soft cloth
moistened with water. Do not use chemicals.
OCT-1
1.3
1.3.1
Front View
8-Port Valve
1.3.2
Rear View
8-Port Sampler 1
Signal
Cable
Power
Cable
8-Port Sampler 2
Signal Cable
(with Power Supply)
OCT-1
1.4.1
1.4
Maintenance
1.4.1
1.4 Maintenance
Replacing the 8-Port Valve Rotor
The OCT-1 uses a fluororesin 8-port valve rotor. The valve rotor may become worn or
scratched after prolonged use due to the suspended particles in the samples. This wear
eventually causes leaks from the 8-port valve.
The following symptoms are associated with leaks from the 8-port valve and indicate
that it is necessary to replace the rotor.
Droplets of liquid around the 8-port valve (especially, around the upper shaft)
If the leakage is great, sample remaining in the tubing after the syringe has been filled
may flow back into the sample container. This phenomenon is most likely due to a
sample tubing, flange connection problem.
CAUTION
Switch off the power before replacing to the 8-port valve rotor.
Replacement Procedure
1.
Shut down the TOC-V unit and switch off the power.
Reference: Refer to Ending Analysis in the Operation chapter of the TOC-V User Manual for details on
shutting down the TOC-V unit.
2.
Remove the 4 screws at the top of the 8-Port Sampler, and take off the cover.
3.
Verify that the O mark on the 8-port valve cover is directed toward port No.1.
Note: If the O mark is not directed toward port No.1, the instrument was not
shut down properly. If this is the case, turn the TOC-V on again, and
shut it down properly.
OCT-1
1.4 Maintenance
1.4.1 Replacing the 8-Port Valve Rotor
4.
Remove the nut using the wrench supplied with the TOC-V.
Nut
5.
6.
7.
Note the orientation (toward port No. 1) of the flat surface of the rod. Remove
the spring-equipped rod.
8.
Note the orientation of the rotor hole position. Grasp the rotor with tweezers or
the like, and remove by pulling upwards.
Rotor
OCT-1
1.4.2
1.4.2
1.4 Maintenance
Troubleshooting
9.
Insert a new rotor, positioning the rotor hole to the original position.
10.
Reinsert the spring-equipped rod with the flat surface directed toward port
No. 1. Tighten the retaining bracket.
11.
Replace the 8-port valve to its original position, and secure it using the nut.
12.
Replace the cover of the 8-Port Sampler and secure it with the four screws.
13.
Troubleshooting
This section describes the various types of error messages that may be displayed when
using the OCT-1 8-Port Sampler, and the corresponding troubleshooting procedures.
1.4.2.1
Error Messages
The following error messages are displayed when a mistake is made during parameter
entry, operation, or if there is a problem with the instrument. Take the appropriate
remedial action indicated in the table.
Table 1.1 CS / WS Error Messages
Message
Description
Remedial Action
Description
Remedial Action
An operational malfunction
occurred in 8-Port Sampler 1
due to foreign matter trapped
8-Port Sampler 1 operation in the 8-port valve.
An operational malfunction
occurred in 8-Port Sampler 2
due to foreign matter trapped
8-Port Sampler 1 operation in the 8-port valve.
OCT-1
1.4 Maintenance
1.4.2 Troubleshooting
1.4.2.2
Troubleshooting
1
NO
Cables are not properly
connected.
Connect the cables properly.
YES
Poor reproducibility.
YES
Reduced reproducibility due to
foreign material.
Eliminate blockage at the tip of the
sample tubing.
NO
Is the leakage
present at the 8-port valve
and sample tubing
connection?
YES
Connection is leaking.
NO
Poor reproducibility is due to the
TOC-V unit.
Refer to the TOC-V User Manual.
OCT-1
1.5
1.5
Reference Materials
1.5.1
Specifications
Reference Materials
1.5.1 Specifications
1.5.2
Ambient temperature
5 - 35C
External dimensions
Weight
Approximately 3.5kg
Standard Accessories
Table 1.3 Table 3 Standard Accessories
Part Number
Part Name
8-Port
Sampler 1
8-Port
Sampler 2
Details
631-20265
Plug 35154
638-41481-01
638-41481-02
638-41481-03
638-41481-04
638-74004-01
Power Cable
638-74028-01
Power Cable
638-74028-02
Power/Signal Cable
M393-E228
User Manual
1*
1***
1***
* Flanged tubing in sets C and C2 both have an inner diameter of 1.0 mm. This set is
provided for use with the combustion type TOC-V.
** Connect this tubing to the common (COM) port of the 8-Port Sampler and port No. 2
on the TOC-V unit.
*** Flanged tubing in sets W and W2 both have an inner diameter of 1.5 mm. This set is
provided for use with the wet chemical type TOC-V.
**** Connect this tubing to the common (COM) port of the 8-Port Sampler and port 1on
the TOC-V unit.
OCT-1
1.5.3
Consumable Parts
Table 1.4 Consumable Parts
1.5.4
Part Number
Part Name
631-20265
Plug 35154
Maintenance Parts
Table 1.5 Maintenance Parts
Part Number Part Name
631-20265
Plug 35154
Details
Rotor (for replacement maintenance)
631-41660
Flanged Tubing W
* Connect this tubing to the common (COM) port of the 8-Port Sampler and port No. 2
on the TOC-V unit.
** Connect this tubing to the common (COM) port of the 8-Port Sampler and port 1on
the TOC-V unit.
1.5.5
Installation
1.5.5.1
Parts Inspection
Confirm that all of the components listed in Section 1.5.2 Standard Accessories
have been included in the shipping package.
1.5.5.2
Installation Site
When selecting an installation site take into account the size of the instrument, how it
will be used, and the installation conditions indicated below. Instrument damage or
measurement errors may occur based on the conditions of the installation site.
10
OCT-1
1.5
Reference Materials
1.5.5 Installation
OCT-2
TOC-V
Main Unit
675mm
This accessory is designed for use on a desk or bench, to the left of the TOC-V main
unit. The external dimensions of the instrument are shown in Figure 1.6 External
Dimensions. A space of at least 200mm is required on either side of and behind the
instrument to enable unhindered operation and maintenance. Sufficient space will also
be required for sample container placement.
OCT-1
OCT-1
TOC-V
Main Unit
TN unit
245mm
440mm
160
mm
1.5.5.3
Installation Procedure
This section describes the procedure for installing the OCT-1 8-Port Sampler.
The unit is set up to the left of the TOC-V main body as shown in Figure
1.7 Installation of the 8-Port Sampler.
TIP: When 2 OCT-1 8-Port Samplers are to be used, place both units to the left of the TOCV main unit as shown in the overhead view of Figure 1.6 External Dimensions.
8-Port Sampler
OCT-1
11
Cable Connections
The cable is connected between the main unit of the TOC-V and the 8-Port Sampler.
CAUTION
Confirm that the power to the TOC-V main unit is switched off before
performing this procedure.
Connection Procedure
1.
Remove the 4 screws at the top of the 8-Port Sampler, and take off the cover.
2.
Connect the power and signal cable to the circuit board of the 8-Port Sampler.
3.
Connect the ground wire of the power cable to the grounding terminal on the
circuit board of the OCT-1.
4.
Pass the cable through the rear panel as shown in figure 1.8, and replace the
cover of the 8-Port Sampler.
Rear Panel
12
OCT-1
5.
Connect the power cable to the power receptacle on the back of the TOC-V.
6.
Connect the signal cable to the 8-Port Sampler connection on the back of the
TOC-V.
1.5
Reference Materials
1.5.5 Installation
1.
2.
Connect the signal cable exiting 8-Port Sampler 2 to the connector of 8-Port
Sampler 1.
8-Port Sampler 2
Signal Cable
8-Port Sampler 1
3.
Remove the port number label from the 8-Port Sampler 2 unit and replace it
with the port number label provided with the unit.
Note: The port that was originally labeled as port No. 1 should be labeled as
port No. 9.
4.
OCT-1
13
1.
The sample tubing differs according to the TOC-V model that is used with the
8-Port Sampler.
TOC-V Model
2.
Connect the fitting at one end of the tubing to the common (COM) port of the
8-Port Sampler.
3.
Pass the sample tubing through the hole in the left side panel of the TOC-V
main unit. Connect the fitting at the other end of the sample tubing to port No.
2 of the 8-port valve of the TOC-V sample injector. Connect to port No. 1
when installing 8-Port Sampler 2.
4.
Sample Tubing
14
OCT-1
2
2Operation
This chapter describes the measurement procedures for the 8-Port Sampler as
they pertain to TOC-VCS, TOC-VWS, TOC-VCP, and TOC-VWP models. These
operations are similar to those of the ASI-V autosampler. Only the operations
that differ from the ASI-V are described in this chapter. For more detailed
information, refer to the documentation provided with your particular TOC-V
model.
2.1
CS Operation
The operations that differ from the ASI-V as the relate to the TOC-VCS model
are described here.
2.2
WS Operation
The operations that differ from the ASI-V as the relate to the TOC-VWS model
are described here.
2.3
CP Operation
The operations that differ from the ASI-V as the relate to the TOC-VCP model
are described here.
2.4
WP Operation
The operations that differ from the ASI-V as the relate to the TOC-VWP model
are described here.
2.1 CS Operation
2.1.1 TC Blank Check Measurement
2.1
CS Operation
This section describes measurement operations using the 8-Port Sampler. These
operations are not covered in the TOC-VCSH/CSN user manual. Operations that are not
described here are the same as those described for the autosampler (ASI-V). Refer to
the applicable item in the TOC-VCSH/CSN user manual for more information.
2.1.1
2.1.2
2.1.2.1
Sample Preparation
Type of Sampler Container
There are no restrictions on the type of sample containers that can be used with the
8-Port Sampler. It is only necessary that the containers be placed so the sample tubing
from the various ports can be inserted directly into the containers.
Note: Wash all sample containers with an appropriate laboratory detergent, and rinse
them thoroughly with water before using them for the first time.
16
OCT-1
2.1 CS Operation
2.1.2 8-Port Sampler Measurement
CS
Note: If the sample container openings are not sealed in the situations listed above,
measurement values may be influenced by such factors as absorption of
atmospheric CO2, contamination with carbon containing substances, and
evaporation of TOC constituents from the samples.
Note: In either case, wash the container with a laboratory detergent, and rinse it
thoroughly with water before use.
OCT-1
17
2.1 CS Operation
2.1.2 8-Port Sampler Measurement
1.
2.
Place a Parafilm square over the opening of the vial, and stretch it to seal it to
the edges of the vials.
3.
Press the edges down so that they wrap around the outside of the vial.
18
OCT-1
2.1 CS Operation
2.1.2 8-Port Sampler Measurement
2.1.2.2
To ensure that the peaks are stable for generating the calibration curve,
perform 4-5 injections using the same analysis parameters prior to
performing the calibration curve analysis. For example, to generate a
400ppb calibration curve, set 400ppb as the 1st standard solution
concentration, and enter 1 as the container number. Actually place a
0ppb solution in the container for port No. 1. The objective at this
point is not to generate a calibration curve, but to inject water using the
same injection volume that will be used to generate the calibration
curve. This not only stabilizes peaks but also thoroughly washes the
sampling needle with purified water.
CS
In the 2nd row of the settings screen and using the same calibration
curve number as above, set the analysis parameters for 0ppb and
400ppb. When sample measurement starts, the calibration curve to be
used will be created with the correct standard solutions.
OCT-1
19
2.1 CS Operation
2.1.2 8-Port Sampler Measurement
20
OCT-1
2.2.1
2.2
2.2 WS Operation
Blank Check Measurement
WS Operation
This section describes measurement operations using the 8-Port Sampler. These
operations are not covered in the TOC-VWS user manual. Operations that are not
described here are the same as those described for the autosampler (ASI-V). For
details, refer to the applicable item in the TOC-VWS user manual.
2.2.1
WS
2.2.2
2.2.2.1
Sample Preparation
Type of Sample Container
There are no restrictions on the type of sample containers that can be used with the
8-Port Sampler. It is only necessary that the containers be placed so the sample tubing
from the various ports can be inserted directly into the containers.
Note: Wash all sample containers with an appropriate laboratory detergent, and rinse
them thoroughly with water before using them for the first time.
OCT-1
21
2.2 WS Operation
2.2.2 8-Port Sampler Measurement
Note: If the sample container openings are not sealed, measurement values may be
influenced by such factors as absorption of atmospheric CO2, contamination
with carbon containing substances, evaporation of TOC constituents from the
samples, etc.
Sealing of Sample Containers
Sample containers may be sealed using commercial sealing films such as Parafilm. It is
also possible to pierce the lid of the container and insert the sample tubing into the
container through this opening.
Note: In either case, wash the container with a laboratory detergent, and rinse it
thoroughly with water before use.
22
OCT-1
2.2 WS Operation
2.2.2 8-Port Sampler Measurement
1.
2.
Place a Parafilm square over the opening of the container, and stretch it to seal
it to the edges of the container.
3.
Press the edges down so that they wrap around the outside of the container.
WS
OCT-1
23
2.2 WS Operation
2.2.2 8-Port Sampler Measurement
2.2.2.2
To ensure that the peaks are stable for generating the calibration curve,
perform 4-5 injections using the same analysis parameters prior to
performing the calibration curve analysis. For example, to generate a
400ppb calibration curve, set 400ppb as the 1st standard solution
concentration, and enter 1 as the container number. Actually place a
0ppb solution in the container for port No. 1. The objective at this
point is not to generate a calibration curve, but to inject water using the
same injection volume that will be used to generate the calibration
curve. This not only stabilizes peaks but also thoroughly washes the
sampling needle with purified water.
In the 2nd row of the settings screen and using the same calibration
curve number as above, set the analysis parameters for 0ppb and
400ppb. When sample measurement starts, the calibration curve to be
used will be created with the correct standard solutions.
24
OCT-1
2.3.1
2.3
2.3 CP Operation
Instrument Setting Wizard
CP Operation
This section describes measurement operations using the 8-Port Sampler. These
operations are not covered in the TOC-VCPH/CPN user manual. Settings performed
prior to measurement in the Instrument Setting Wizard Options screen and in the
Maintenance menu are different. Operations that are not described here are the same as
those described for the autosampler (ASI-V). For details, refer to the applicable item in
the TOC-VCPH/CPN user manual.
CP
2.3.1
8-Port Sampler-1
Select this option when only one 8-Port Sampler unit is to be used. If the 8-Port
Sampler-1 option is selected, ASI and SSM are disabled.
8-Port Sampler-2
Select this option when two 8-Port Sampler units are to be used. 8-Port Sampler-2 can
only be selected after 8-Port Sampler-1 is selected.
Refer to Section 4.2.1 TOC Parameter Settings in the TOC-VCPH/CPN user manual
for details on other options.
OCT-1
25
2.3 CP Operation
2.3.2 Sample Table Setting for Calibration Curve Measurement
2.3.2
Figure 1.4 Sparge / Acid Addition Dialog Box for the 8-Port Sampler
Refer to the sample table setting procedure described in Section 4.3 Measurement
Procedures and Tutorial of the TOC-VCPH/CPN user manual for more information.
2.3.3
View Menu
The View menu displayed in the Method, Sample Information Display Options and
various other screens changes when performing measurement using the 8-Port
Sampler.
26
OCT-1
2.3 CP Operation
2.3.4 Sample Containers
2.3.4
Sample Containers
When performing measurement using the 8-Port Sampler, the autosampler and the vial
setting screens are different. The description in Section 4.4.4.10 ASI Vials in the
TOC-VCPH/CPN User Manual is changed as follows.
2.3.4.1
CP
This dialog box is displayed. Select OK to save the changes and close the dialog box.
Single 8-Port Sampler
The vial setting screen appears as shown below.
Figure 1.6 Sparge / Acid Addition Dialog Box for a Single 8-Port Sampler
OCT-1
27
2.3 CP Operation
2.3.4 Sample Containers
Figure 1.7 Sparge / Acid Addition Dialog Box for Two 8-Port Samplers
A tool tip window displays the port number when the mouse is moved over a sample
container in the Port No. graphic. Click on the port number to display the current
setting.
28
OCT-1
Parameter
Description
Port number
Sparge time
2.3 CP Operation
2.3.5 Maintenance Menu Description
2.3.5
CP
Zero Detection
This function is used to detect the syringe zero point. Refer to Section 4.1.3.2 Syringe
Pump Zero Point Detection in the TOC-VCPH/CPN user manual for more details.
OCT-1
29
2.3 CP Operation
2.3.5 Maintenance Menu Description
TC Catalyst Regeneration
This function initiates the catalyst regeneration process. In TC catalyst regeneration,
several injections of 2M hydrochloric acid are added to the TC combustion tube.
Press the Start button to begin TC catalyst regeneration. Refer to Section 5.2.1
Catalyst Regeneration in the TOC-VCPH/CPN User Manual for more information.
IC Reagent Regeneration
This function initiates regeneration of the IC reagent. This selection is disabled for the
TOC-VCPN.
Press the Start button to begin IC reagent regeneration. Refer to Section Section 4.1.11
IC Analysis Preparations in the TOC-VCPH/CPN User Manual for more information.
Residue Removal
Select this command to remove substances remaining in the flow line or syringe.
Press the Start button to begin removal of the residual substances. This button is
disabled and the progress bar is active while removal is being performed.
30
OCT-1
2.3 CP Operation
2.3.5 Maintenance Menu Description
CP
Click the Set Flow Line to Wash button to open the Flow Line Wash dialog box. Select
the flow line to be washed, and click the Wash button. The various ports of the 8-Port
Samplers can be selected for individual flow line washing.
The Wash Dialog box buttons are disabled and the progress bar is activated during the
procedure.
OCT-1
31
2.3 CP Operation
2.3.5 Maintenance Menu Description
Mechanical Check
This function is used to check each operation. This item is mainly used for servicerelated matters.
TOC Tab
32
OCT-1
Screen Item
Function
Syringe Pump
Solenoid Valve
TC Slider
IC Slider
2.3 CP Operation
2.3.5 Maintenance Menu Description
CP
Figure 1.16 Mechanical Check Dialog Box - ASI / 8-Port Sampler Tab
Screen Item
Function
Vial Position
Press the Start button to begin the initialization process for the ASI or 8-Port Sampler,.
OCT-1
33
2.3 CP Operation
2.3.5 Maintenance Menu Description
Screen Item
Description
Replacement Preparation
Finish Replacement
Close
Blank Check
The blank check function is used to measure the blank system value of the instrument.
This is performed prior to conducting high sensitivity measurement or just after
replacing the TOC high sensitivity catalyst. Refer to Section 4.1.15.2 TC Blank
Check Analysis in the TOC-VCPH/CPN User Manual for more information.
34
OCT-1
2.3 CP Operation
2.3.5 Maintenance Menu Description
CP
Maintenance Log
Refer to Section 4.2.4 Maintenance History Settings of the TOC-VCPH/CPN User
Manual for more information.
Sparge Valve
Place a check mark next to this item to start the sparging operation in the external
sparge kit.
OCT-1
35
2.3 CP Operation
2.3.5 Maintenance Menu Description
36
OCT-1
2.4.1
2.4
2.4 WP Operation
Instrument Setting Wizard
WP Operation
This section describes measurement operations using the 8-Port Sampler. These
operations are not covered in the TOC-VWP user manual. Settings performed prior to
measurement in the Instrument Setting Wizard Options screen and in the
Maintenance menu are different. Operations that are not described here are the same as
those described for the autosampler (ASI-V). For details, refer to the applicable item in
the TOC-VWP user manual.
2.4.1
WP
8-Port Sampler-1
Select this option when only one 8-Port Sampler unit is to be used. If the 8-Port
Sampler-1 option is selected, ASI and SSM are disabled.
8-Port Sampler-2
Select this option when two 8-Port Sampler units are to be used. 8-Port Sampler-2 can
only be selected after 8-Port Sampler-1 is selected.
Refer to Section 3.3.1.2 Options Window in the TOC-VWP user manual for details
on other options.
OCT-1
37
2.4 WP Operation
2.4.2 Sample Table Setting for Calibration Curve Measurement
2.4.2
Figure 1.22 Sparge / Acid Addition Dialog Box for the 8-Port Sampler
Refer to the sample table setting procedure described in Section 4.3.2 Tutorial of the
TOC-VWP user manual for more information.
2.4.3
View Menu
The View menu displayed in the Method, Sample Information Display Options and
various other screens changes when performing measurement using the 8-Port
Sampler.
38
OCT-1
2.4 WP Operation
2.4.4 ASI / 8-Port Sampler Vials
2.4.4
2.4.4.1
This dialog box is displayed. Select OK to save the changes and close the dialog box.
WP
Figure 1.24 Sparge / Acid Addition Dialog Box for a single 8-Port Sampler
OCT-1
39
2.4 WP Operation
2.4.4 ASI / 8-Port Sampler Vials
Figure 1.25 Sparge / Acid Addition Dialog Box for Two 8-Port Samplers
A tool tip window displays the port number when the mouse is moved over a sample
container in the Port No. graphic. Click on the port number to display the current
setting.
40
OCT-1
Parameter
Description
Port number
Sparge time
2.4 WP Operation
2.4.5 Maintenance Menu Description
2.4.5
WP
Zero Detection
This function is used to detect the syringe zero point. Refer to Section 4.1.4 Syringe
Pump Zero Point Detection in the TOC-VWP User Manual for more details.
OCT-1
41
2.4 WP Operation
2.4.5 Maintenance Menu Description
Press the Start button to start the air bubble removal process
Flow Line Wash
Select this command to wash the selected flow line.
Click the Set Flow Line to Wash button to open the Flow Line Wash dialog box. Select
the flow line to be washed, and click the Wash button. The various ports of the 8-Port
Samplers can be selected for individual flow line washing.
42
OCT-1
2.4 WP Operation
2.4.5 Maintenance Menu Description
The Wash Dialog box buttons are disabled and the progress bar is activated during the
procedure.
8-Port Valve
Button
Description
Wash
Dry
Description
Wash
Drain
WP
Description
Wash
Drain
OCT-1
43
2.4 WP Operation
2.4.5 Maintenance Menu Description
Mechanical Check
This function is used to check each operation. This item is mainly used for servicerelated matters.
TOC Tab
44
OCT-1
Screen Item
Function
Syringe Pump
Solenoid Valve
2.4 WP Operation
2.4.5 Maintenance Menu Description
WP
Figure 1.32 Mechanical Check Dialog Box - ASI / 8-Port Sampler Tab
Screen Item
Function
Vial Position
Press the Start button to begin the initialization process for the ASI or 8-Port Sampler,.
OCT-1
45
2.4 WP Operation
2.4.5 Maintenance Menu Description
Screen Item
Description
Replacement Preparation
Finish Replacement
Close
Blank Check
The blank check function is used to measure the blank system value of the instrument.
This is performed prior to conducting high sensitivity measurement or just after
replacing the TOC high sensitivity catalyst. Refer to Section 4.3.3 TC Blank Check
Analysis in the TOC-VWP User Manual for more information.
46
OCT-1
2.4 WP Operation
2.4.5 Maintenance Menu Description
Maintenance Log
Refer to Section 4.2.4 Maintenance History Settings of the TOC-VWP User Manual
for more information.
WP
Sparge Valve
Place a check mark next to this item to start the sparging operation in the external
sparge kit.
Sparge Gas Flow Control (Wet. Chem.)
Use this function to control the flow rate of sparge gas to the IC reactor. Refer to
Section 4.3.1.4 Setting Sparge Gas Flow Rate in the TOC-VWP User Manual for
more information.
Figure 1.37 Sparge Gas Flow Rate Adjustment (Wet. Chem.) Dialog Box
OCT-1
47
2.4 WP Operation
2.4.5 Maintenance Menu Description
48
OCT-1
Index
Numerics
8-Port sampler
construction ..............................................4
initialize CP ............................................33
initialize WP ...........................................45
8-Port valve rotor, replacing.............................5
E
A
Accessories, standard .......................................9
Air bubble removal
CP ...........................................................31
WP ..........................................................42
B
Blank check
CP ...........................................................34
CS ...........................................................16
WP ..........................................................46
WS ..........................................................21
C
Cable connections...........................................12
Calibration curve, sample table
CP ...........................................................26
WP ..........................................................38
Connection
cables ......................................................12
two units .................................................13
with TOC-V............................................14
Construction
front view..................................................4
rear view ...................................................4
Consumable parts ...........................................10
Conventions.....................................................vi
CP
air bubble removal..................................31
blank check .............................................34
calibration curve sample table ................26
date and time...........................................35
flow line wash.........................................31
IC reagent regeneration ..........................30
initialize ..................................................33
Instrument Setting Wizard......................25
Maintenance log .....................................35
maintenance menu ..................................29
mechanical check....................................32
operation .................................................25
residue removal ......................................30
sample container.....................................26
sparge gas flow .......................................47
sparge valve ............................................35
TC catalyst regeneration.........................30
zero detection..........................................29
CS
blank check .............................................16
high sensitivity........................................19
operation .................................................16
sample preparation..................................16
F
Flow line wash
CP........................................................... 31
WP.......................................................... 42
Front View ....................................................... 4
H
Hardware.......................................................... 1
High sensitivity
CS........................................................... 19
WS.......................................................... 24
I
IC reagent regeneration.................................. 30
Initialize
CP........................................................... 33
WP.......................................................... 45
Installation
inspection ............................................... 10
procedure................................................ 11
site .......................................................... 10
Instrument Setting Wizard
CP........................................................... 25
WP.......................................................... 37
M
Maintenance ..................................................... 5
Maintenance log
CP........................................................... 35
WP.......................................................... 47
Maintenance menu
CP........................................................... 29
WP.......................................................... 41
Maintenance parts .......................................... 10
Manual contents ............................................... v
Mechanical check
CP........................................................... 32
WP.......................................................... 44
N
Notation conventions ...................................... vi
Index-1
Index
O
Operation
CP........................................................... 25
CS........................................................... 16
WP.......................................................... 37
WS.......................................................... 21
Operational precautions ................................... 3
Overview .......................................................... 2
View menu
CP ...........................................................26
WP ..........................................................38
P
Parts
consumable............................................. 10
maintenance............................................ 10
R
Rear view ......................................................... 4
Residue removal, CP...................................... 30
S
Sample
container...........................................16, 26
loading..............................................17, 22
preparation........................................16, 21
Sparge gas flow, CP ....................................... 47
Sparge valve
CP........................................................... 35
WP.......................................................... 47
Specifications ................................................... 9
Standard accessories......................................... 9
Syringe, replace........................................34, 46
T
TC catalyst regeneration, CP.......................... 30
Troubleshooting ...........................................7, 8
Index-2
W
Warranty ..........................................................iv
WP
air bubble removal ..................................42
blank check .............................................46
calibration curve .....................................38
date and time...........................................47
flow line wash.........................................42
initialize ..................................................45
Instrument Setting Wizard......................37
maintenance log......................................47
maintenance menu ..................................41
mechanical check....................................44
sparge valve ............................................47
syringe replace ........................................46
view menu...............................................38
zero detection..........................................41
WS
blank check .............................................21
high sensitivity........................................24
operation .................................................21
sample preparation..................................21
Z
Zero detection
CP ...........................................................29
WP ..........................................................41