Specification Sheet
S/5 Avance
Innovating with you, shaping exceptional care
Features
Complete patient monitoring capabilities: respiratory gas,
hemodynamic and adequacy of anesthesia
Our state of the art electronic gas mixer with pneumatic
back-up control
Full color displays
Advanced Breathing System (ABS)
Minimal number of parts and tube connections greatly reduces
the potential for leaks and misconnects
Ease of disassembly (no tools)
Fully autoclavable and latex-free
Superior ventilation options
Volume Control, Pressure Control, PSVPro (Pressure Support
with Apnea backup mode), Synchronized Intermittent Mandatory
Ventilation (SIMV) (Volume and Pressure), electronic PEEP
Tidal volume compensation
One motion from mechanical to manual mode
Two key presses to total standby: end case
Cardiac bypass case mode
Exceptional design
Generous storage and work surface space
Bi-level work surface illumination
Integrated cable and tube management
Physical specifications
Dimensions
Height:
134.5 cm/52.9 in
Width:
72 cm/28.3 in
Depth:
73 cm/28.7 in
Weight:
Top shelf
Weight limit:
34 kg/75 lb
Width:
66 cm/26 in
Depth:
40 cm/15.75 in
Work surface
Height:
81.7 cm/32.2 in
Size:
DIN rail
Side of
machine:
34.5 cm/13.6 in
17.5 cm/6.9 in
Width:
33 cm/13 in
Depth:
26.5 cm/10.4 in
30.5 cm/12 in
Casters
Diameter:
12.5 cm/5 in
Brakes:
XXxxx Xxxx
8:48
1:58
15
l/min
15
l/min
Paw
40
cmH20
P
a
w
28
2
F
l
o
w
CO2
1.0
1.0
MV l/min
10.0
2.0
TV exp ml
1000
Off
6.2
515
-120
O2
C
O
2
/
O
2
EtCO2 mmHg
5.0
Off
EtO2 %
Off
Off
26
45
40
PEEP
Flow 120
Gases
O2
%
Total Flow
l/min
TV
ml
RR
l/min
I:E
50
2.00
500
12
1.2
PEEP
cmH2O
Off
O2 w
ET
FI
Pressure
(Psupport) range: Off, 2 to 40 cm H2O
(increments of 1 cm H2O)
Pmean
Ppeak
20
N2O
Pressure
(Plimit) range: 12 to 100 cm H2O
(increments of 1 cm H2O)
RR /min
Off
Off
12
FiCO2 mmHg
Off
Rate:
FiO2 %
49
N2O
Off
21
Sev
45 49 1.5
49 49 1.8
Inspiratory/
expiratory
ratio:
Inspiratory
time:
Trigger
window:
Flow trigger:
Tidal volume
range:
Incremental
settings:
20 to 1500 mL (Volume
Control and SIMV modes)
20 to 100 mL
(increments of 5 mL)
100 to 300 mL
(increments of 10 mL)
300 to 1000 mL
(increments of 25 mL)
1000 to 1500 mL
(increments of 50 mL)
Minute volume
range:
0 to 99.9 L/min
Pressure
(Pinspired)
range:
5 to 60 cm H2O
(increments of 1 cm H2O)
5 to 1500 mL volume delivery
Inspiration
termination
level:
2:1 to 1:8
(increments of 0.5)
0.2 to 5.0 seconds
(increments of 0.1 seconds)
(SIMV and PSV Pro)
0 to 80% (increments of 5%)
1 to 10 L/min
(increments of 0.5 L/min)
0.2 to 1 L/min
(increments of 0.2 L/min)
Integrated, electronically
controlled
Range:
OFF, 4 to 30 cm H2O
(increments of 1 cm H2O)
Ventilator performance
Pressure range
at inlet:
240 kPa to 700 kPa/
35 psig to 100 psig
Peak gas
flow:
Flow valve
range:
Flow
compensation
range:
200 mL/min to 15 L/min
Alarm settings
Tidal volume
(VTE):
Minute volume
(VE):
Low: OFF, 0 to 10 L/min
High: 0 to 30 L/min, OFF
Inspired oxygen
(FiO2):
Low: 18 to 100%
High: 19 to 100%, OFF
Apnea alarm: Mechanical ventilation ON:
< 5 mL breath measured
in 30 seconds
Ventilator accuracy
Delivery/monitoring accuracy
Volume
delivery:
Pressure
delivery:
Pressure
monitoring:
Low airway
pressure:
Ventilator components
Flow transducer
Delivery
Type:
Vaporizers:
Dimensions:
22 mm OD and 15 mm ID
Location:
Number of
positions:
Mounting:
Tool-free installation
Selectatec manifold interlocks and isolates vaporizers
Oxygen sensor
Type:
g es.
of
rnin all tim cence ents.
Waright atclave. e presetic agunction
nl
th sth
to
up
ep t au use in anae in co
.
KeDo not for able used itor.
ns. ance
tio rm
No mm be mon r
fla ly to O l fo struc perfo ance
an ua In
Onwith e man onal fecting ainten .
Se erati rs af and m tions
Op
Facto ice menda
Serv m
e
dat
reco
ice
Lock
ore
bef se
u
Ventilator screen
Display size:
31 cm/12.1 in diagonal
Pixel format:
Nex
t serv
18
16
14
12
10
8 7
6
ay
sm
flow ing of
gas eath use
low the br ds the
g
ith
nin ngs w within mmen
War er settiixturesgly reco
and b.
m
in.
poriz xic stron
L/m 13m
h va po eda g.
use. 10 at 10
Higad to hy Ohm itorin
fore 2 to C
le rcuit. mon
l be s of 0.C to 30
ua
ci ygen
w
18
man flo
ox
r to e at res of
Refe for usperatu
d
rate t tem
Calibambien
at
Battery backup
c
Te
lus
6P
USE
Y DE
ONL
ION
OPE
NO
SFLU
AL
RAT
T
OUTPU
AGENT
LOW
WARM
UP
M LO
Y
ALAR
TER
BAT
only e
Use luran
sf
De
Tec 6 Plus
Communication ports
RS-232C compatible serial interface
Ethernet
D-O device interface solutions port
USB port
Tec 7
RANE
General
Measurement
range:
4 to 60 breaths per minute
Detection
criteria:
Weight:
1.6 kg/3.7 lb
1% variation in CO2
Inspired O2 concentration
EtO2:
End-tidal O2 concentration
FiO2-EtO2:
Inspired-expired difference
O2 waveform
Non-disturbing gases:
Measurement
range:
0 to 100 %
Accuracy:
Maximum effect
on readings: CO2 < 0.2 vol %;
O2 < 2 vol %
FiCO2:
2 vol %*
CO2 waveform
Accuracy:
Measurement
range:
0 to 15% (0 to 15 kPa,
0 to 113 mmHg)
* Typical value
** Measurement not valid with O2 and N2O mixtures
Accuracy:
0.3 vol %*
3 vol %*
FiN2O > 82% alarm
Patient Spirometry
(available in S/5 AM module bay)
Pressure-volume loop
Flow-volume loop
Airway pressure and flow waveforms
Adjustable low and high alarm limits for
Ppeak, PEEPtot and MVexp
Alarms for MVexp << MVinsp and for
MVexp low. Detection through D-lite or
Pedi-lite flow sensor and gas sampler with
following specifications:
Waveform displayed
MAC value displayed
Identification
threshold:
0.15 vol %*
Agent mixture detection
Adjustable high and low alarm limits for
EtAA, FiAA
Respiration
rate:
D-lite
Pedi-lite
4 to 35
breaths/min
4 to 50
breaths/min
Tidal volume
Measurement
range:
150 to
15 to
2000 mL
300 mL
Accuracy*: 6% or 30 mL 6% or 4 mL
Minute volume
Measurement
range:
2 to 20 L/min 0.5 to
5 L/min
Accuracy*: 6%
6%
Airway pressure
Measurement
range:
-20 to +100 -20 to +100
cm H2O
cm H2O
1 cm H2O
Accuracy*: 1 cm H2O
Display
units:
cm H2O, mmHg, kPa,
mbar, hPa
Flow
Measurement
range:
1.5 to 100
L/min
0.25 to 25
L/min
I:E
Measurement
range:
1:4.5 to 2:1
Compliance
Measurement
range:
4 to 100
mL/cm H2O
1 to 100
mL/cm H2O
Current leakage
100/120 V:
< 300A
220/240 V:
< 500A
Power
Sensor specifications
Dead space:
Electrical specifications
Power input:
D-lite
Pedi-lite
9.5 mL
2.5 mL
Power cord:
1.0 cm H2O
Inlet/outlet modules
Resistance
at 30 L/min 0.5 cm H2O
at 10 L/min
220240V
Gas exchange**
(available in S/5 AM module bay)
VO2:
Oxygen consumption
VCO2:
Measurement
range:
20 to 1000 mL/min
Respiration
rate range:
System circuit
breakers:
8A
Outlets
(optional):
4 to 35 bpm
Accuracy
FiO2 < 65%:
10% or 10 mL
15% or 15 mL
120V
100V
15A
15A
Pneumatic specifications
ISO 22 mm OD and
15 mm ID
Gas supply
Pipeline input
range:
280 kPa to 600 kPa/
41 psig to 88 psig
Pipeline
connections:
DISS-male, DISS-female,
DIN 13252, AS4059,
BSPP 3/8, S90-116, or NIST
All fittings available for O2,
N2O, and Air, and contain
pipeline filter and check valve.
O2 controls
Method:
Supply failure
alarm:
Range: 193 kPa to 221 kPa/
28 psig to 32 psig
Sounds at maximum volume
every 10 seconds
O2 flush:
Indicator:
Flow tube
Indicator
accuracy:
Fresh gas
Flow range:
Total flow
accuracy:
O2 flow
accuracy:
Balance gas
flow accuracy: 5% or 20 mL/min of
setting (larger of) Air/N2O
O2 concentration
range:
25% to 100%
O2 concentration
accuracy:
5% V/V for flows < 1 L/min
2.5% setting for flows
> 1 L/min
Electronic mixer
response time: 500mS
(10% to 90% flow step)
Compensation: Temperature and
atmospheric pressure
compensated to standard
conditions of 20C
and 101.3 kPa
Hypoxic guard: Electronic
5% full scale
Materials
All materials in contact with patient breathing gases are free of natural rubber latex.
Environmental specifications
System operation
Operational modes
Humidity:
Altitude:
System storage
Altitude:
Oxygen cell
storage:
Absorbent
capacity:
800 g
Exhalation:
22 mm OD ISO
15 mm ID taper
Inhalation:
22 mm OD ISO
15 mm ID taper
Bag port:
22 mm OD
Bag-to-Ventilator switch
-15 to 50C/5 to 122F
10 to 95% relative humidity
500 to 800 mmHg
Type:
Bi-stable
Control:
Electromagnetic compatibility
Immunity:
Emissions:
Range:
Approvals:
UL 2601-1,
CSA C22.2 #601.1,
EN/IEC 60601-1,
CE 0197, EN 740
Tactile knob
indication at: 30 cm H2O and above
0.8 to 70 cm H2O
Adjustment range
of rotation:
0.8 to 30 cm H2O
(0 to 230)
30 to 70 cm H2O
(230 to 330)
Materials
AGSS Type
Bag
mode: 1.82 mL/cm H2O
Mechanical
mode: Automatically
compensates for
compression losses
within the absorber
and bellows assembly
Total circuit
volume:
Expiratory
resistance:
Includes Absorber
volume
Flow rate
Pexp
Bag Mode
Pressure drop
10 L/min
30 L/min
60 L/min
Note:
Pexp
Vent Mode
Pressure drop
High vacuum,
low flow with
indicator:
High vacuum
36 L/min
@ 12 in Hg
(305 mmHg)
High vacuum,
variable flow
with bag
indicator:
High vacuum
30 L/min
extract flow
@ 12 in Hg
(305 mmHg)
Low vacuum,
low flow
without
indicator:
Adjustable
Venturi with
flowmeter,
extract flow
< 36 L/min
Low vacuum,
high flow with
indicator:
Low vacuum
> 50 L/min
extract flow
DISS evac
DISS evac
12.7 mm/
0.5 in
hose barb
30 mm/
1.2 in
BSI male
threaded
Low vacuum,
low flow:
Venturi/Ejector 25 mm/
with extract
0.98 in
flow 36 L/min hose barb
Passive:
Passive or
30 mm/
external active 1.2 in
system with
M ISO taper
air break
GE Healthcare
Madison, Wisconsin U.S.A.
www.gehealthcare.com
General Electric Company reserves the right to make changes in specifications and features
shown herein, or discontinue the product described at any time without notice or obligation.
Contact your GE Representative for the most current information.
2004 General Electric Company
Datex-Ohmeda, Inc., a General Electric company going to market as GE Healthcare.
AN4199-D 9/04 Printed in USA