Professional Documents
Culture Documents
Biosensors Electrochemical 2007 PDF
Biosensors Electrochemical 2007 PDF
13231329 (2007)
Automation and
Analytical Techniques
Key Laboratory of Analytical Chemistry (Chongqing), College of Chemistry and Chemical Engineering, Southwest University, Chongqing, Peoples
Republic of China.
* Address correspondence to this author at: College of Chemistry and
Chemical Engineering, Southwest University, Chongqing, China. Fax 86-236825-4000; e-mail tdping@swu.edu.cn.
Received December 26, 2006; accepted April 23, 2007.
Previously published online at DOI: 10.1373/clinchem.2006.085126
Tumor markers, such as -fetoprotein (AFP),1 carcinoembryonic antigen (CEA), cancer antigen 125 (CA 125), and
CA 15-3, can be found in the body (usually blood or urine)
when cancer is present (1 ). Tumor markers are mainly
used for evaluating therapy and detecting recurrence or
metastasis (2 ). Immunological methods have become the
predominant analytical techniques for quantitative detection of tumor markers (3, 4 ). Despite the high sensitivity
of conventional immunoassay methods such as RIA and
ELISA, they have some limitations, such as the short
shelf life of 125I-labeled antibody, radiation hazards, a
complicated wash procedure, and a relatively long analysis time (5, 6 ). An alternative immunosensing strategy
that is based on an electrochemical principle and does not
require an antibody-labeling reaction would be advantageous because of simple instrumentation and easy signal
quantification. A simultaneous multianalyte immunoassay, in which 2 or more analytes are measured in a
single assay, would be advantageous in the clinical laboratory to simplify testing, throughout, and reduce overall
cost per test. Various assay formats have been devised to
realize simultaneous multianalyte analysis, using either
multiple labels or spatial resolution to discriminate between the different analytes (7, 8 ). Although modification
of the individual electrodes with different biological recognition elements would enable the construction of miniaturized biosensor arrays, the directed immobilization of
functional proteins on individual microscopic regions is
still a challenge (9, 10 ).
1
Nonstandard abbreviations: AFP, -fetoprotein; CEA, carcinoembryonic
antigen; CA, cancer antigen; TEOS, tetraethoxysilane; RMS, root mean square.
1323
1324
Magnetic controlled molecular electronics and bioelectronics examine the effect of an external magnetic field on
electrochemical signals of functionalized magnetic particles associated with electrodes (11 ). Magnetic sorting
protein assay systems of various throughput have been
built and used for clinical applications (1216 ). On a
single chip, batch-type magnetic separators have been
fabricated that trap magnetic particles in flowing fluids
by use of an external magnetic field (17 ). The loading
capacity of these devices is limited, however, because
accumulation of the collected particles can restrict fluid
flow or lead to irreversible entrapment of samples, and
their use is hampered by the need to disrupt continuous
operation for sample elution. We used the physical characteristics of magnetic nanocore and the good biocompatibility of silica shell to construct a nontoxic biomimetic
interface for the immobilization of proteins. Attraction of
the functionalized magnetic nanoparticles to the probe
surface with an external magnet activates the electrical
contact between the immobilized proteins and the base
electrode, and the sensors circuit is switched on. Positioning the magnet above the cell retracts the magnetic
nanoparticles from the probe surface, and the electrochemical activity of the functional magnetic particles is
switched off.
The focus of this report is the fabrication of a novel
bead-bed immunoassay system on a microchip for multiplex measurement of 4 tumor markers, AFP, CEA, CA
125, and CA 15-3, with the switching and controlling of
electrochemical signals by means of an external magnet.
Monitoring the changes in the potential signals before and
after the antigenantibody interaction provides the basis
for the immunoassay.
We dissolved Fe(NO3)3 9H2O, Ni(NO3)2 6H2O, and glycine in distilled water (note: Fe3/Ni2 2:1, glycine/
nitrate 4:1, in molar ratio). After filtration, we heated
the transparent brown precursor solutions until a com-
Scheme 1. Diagrams of the microfluidic device (left) and the fabrication, measurement, and regeneration procedures of the immunosensor (right).
1325
measurement method
The manifold consists of a syringe pump directly connected to a distribution valve system with 8 ports. All
basic operations, including the antigen-antibody reaction,
immunocomplex capture, and washing, were carried out
automatically. PBS or sample was injected at 0.25 mL/min
through the reactor, and each sample to be analyzed was
introduced into the detection vessel after stabilization of
response (shift 1 mV/min) by a digital ion analyzer
(model PHS-3C, Dazhong Instruments). The detection
principle is based on the shift in the potential before and
after the antigenantibody interaction. To avoid possible
error resulting from different additions of samples and
deduct the responses induced by nonspecific adsorption,
the response of each sensor was recorded as the immunoreaction from 30 s (after the addition of samples) until
equilibrium was reached, 5 min. The control tests with
normal (negative) samples and the evaluations for clinical
specimens were performed accordingly. After each immunoassay, we regenerated the contaminated probes by
pulling out the permanent magnet and washing with pH
7.4 PBS.
(1)
1326
To arrive at the optimum conditions for antigen detection, we optimized both the preparation of magnetic
nanoparticles and the fabrication of sensors. The SiO2
provides a good surface for subsequent functionalization
with protein. The SiO2 coating is not sufficiently thick to
1327
1 (with anti-AFP)
2 (with anti-CEA)
5 (control test)
AFP
CEA
CA125
CA15--3
Negative sample
36.7
2.1
2.5
1.0
3.5
1.3
46.3
1.5
2.2
2.7
0.9
2.3
53.1
2.7
5.4
2.4
2.5
1.8
57.8
6.3
2.3
2.5
1.9
2.1
1.6
To investigate the reproducibility of the newly prepared immunoassay protocol, we repeatedly assayed 20
times for different concentrations of 4 tumor markers. The
CVs among 20 runs were 8.6%, 7.5%, and 4.3% for 2.5, 80,
and 160 g/L AFP; 6.3%, 5.4%, and 3.7% for 2.0, 100,
and 180 g/L CEA; 6.1%, 7.4%, 4.9% for 3.5, 30, and
50 kunits/L CA 125; and 3.5%, 7.8%, and 9.2% for 3.0, 120,
and 220 kunits/L CA 15-3. The prepared biomagnetic
nanoparticles dispersed uniformly, resulting in relatively
large variation. Considering this factor, these CV values
were acceptable. The low CV values indicated that the
proposed immunosensor could be regenerated and used
repeatedly, and further verified the possibility of batch
preparation of biomagnetic nanoparticles. When the biomagnetic nanoparticles were not in use, they were stored
in PBS (pH 7.4) at 4 C. No obvious change was observed
after storage for 7 days. Thus, the magnetic NiFe2O4/SiO2
nanoparticles could efficiently immobilize various proteins.
1328
References
1. Schneider J. Tumor markers in detection of lung cancer. Adv Clin
Chem 2006;42:1 41.
2. Duffy MJ. Serum tumor markers in breast cancer: are they of
clinical value? Clin Chem 2006;52:34551.
3. Merani S, Schapiro A. Current status of pancreatic islet transplantation. Clin Sci 2006;110:61125.
4. Tang D, Yuan R, Chai Y, Fu Y, Dai J, Liu Y, et al. New amperometric
and potentiometric immunosensors based on gold nanoparticles/
Tris(2,2-bipyridyl)cobalt(III) multilayer films for hepatitis B surface
antigen determinations. Biosens Bioelectron 2005;21:539 48.
5. Cherif B, Roget A, Villiers CL, Calemczuk R, Leroy V, Marche P, et
al. Clinically related protein-peptide interactions monitored in real
time on novel peptide chips by surface plasmon resonance
imaging. Clin Chem 2006;52:255 62.
15.
16.
17.
18.
19.
20.
21.
22.
1329