Set the pump flow rate at 2.3 mL/min and detector sensitivity at 0.08 AUFS Turn on the Kipp and Zonen flatbed recorder an set the slow speed rate Flush the column with 20%methanol: 80%H20 for 5 to 10 minutes o Record the response to make sure the column is clean Inject 10 l of solution (if it is a standard, then from the least concentrated first) into the HPLC while in the LOAD position. Push the PRIME button and the valve will turn to the inject position o The detector will provide a record of the sample injected To shut down HPLC o Clean syringe with methanol (HPLC grade) o Run 100% methanol (HPLC grade) as a solvent for 5-10 minutes o Turn off pump, detector, and recorder