Adjustment of Patient Table R711 Potentiometer


Mx8000 IDT, Mx8000
Hardware Informational
N/A
N/A

PURPOSE:

The purpose of this technical note is to inform the field of the calibration procedure of a new Patient Table R711
Potentiometer. Follow this procedure if:

The PTH board has been replaced with the New PTH Board Kit P/N SP717542306
You are installing a new Patient Table R711 Potentiometer

After replacement of the PTH board, the reference voltage to the R711 potentiometer is now 5V and not 10V.

VALIDITY:

Mx8000 IDT and Mx8000 systems where the PTH board has been replaced with the New PTH Board Kit
P/N SP717542306

LOGISTICS:

N/A

All pages of this document contain proprietary and confidential information of Philips Medical Systems(Cleveland), Inc. The documents are intended for current Philips Medical Systems(Cleveland), Inc. personnel or are
licensed to Philips Medical Systems(Cleveland), Inc. customer for use by the customer's in-house service employee on equipment located at the customer's designated site. Copying, disclosure to others or other use is
prohibited without the express written authorization of Philips Medical Systems(Cleveland), Inc Law Department. Report violation of these requirements to the Philips Medical Systems(Cleveland), Inc Law
Department,Highland Heights, Ohio.

Customer Bulletin CTQ1218

March 2003

Procedure
WARNING
FOLLOW ALL SAFETY GUIDELINES AS DESCRIBED IN CHAPTER 1 OF THE MX8000 IDT SERVICE MANUAL.
FAILURE TO DO SO CAN RESULT IN SEVERE PERSONAL INJURY.
After replacing a faulty horizontal position potentiometer according to Chapter Patient Table ,Section Adjustment of
Potentiometer R711 in the Mx8000 IDT Service Manual, calibrate the new potentiometer as follows:
1. Turn the system ON.
2. Move the tabletop all the way back.
3. Measure the input voltage to the potentiometer between pins 1 (brown)
and 3 (green). The value must measure 5V.
4. Measure the following voltages in this table position:
Between pins 2 (white) and 3 (green) = 4.75V0.2V
Between pins 1(brown) and 2 (white) = 0.24V0.2V

3-2

Pin 1 (brown)

Pin 3 (green)

5. If the values do not fall within this range, modify the shaft adjustment on
the new potentiometer to achieve these values. Refer to Chapter
Patient Table ,Section Adjustment of Potentiometer R711 in the
Mx8000 IDT Service Manual.
Pin 2 (white)

3-1

Figure 1: PTH Potentiometer