You are on page 1of 47

1

SELECTION,
VERIFICATION AND
VALIDATION OF METHOD
08 JUL 2016
masih

Selection
2

ISO/IEC 17025:2005
Clause 5.4.2
The laboratory shall use test and/or calibration
methods, including methods for sampling, which meet
the needs of the customer and which are appropriate
for the tests and/or calibrations it undertakes

EURACHEM Guide. The Fitness for Purpose of Analytical Methods: A Laboratory Guide
to Method Validation and Related Topics, 2nd ed. 2014

Verification
3

The provision of objective evidence that a given item


fulfills specified requirements. [VIM 2.44]

EURACHEM Guide. The Fitness for Purpose of Analytical Methods: A Laboratory Guide
to Method Validation and Related Topics, 2nd ed. 2014

Validation
4

Verification, where the specified requirements are


adequate for intended use. [VIM 7]
Confirmation by examination and provision of
objective evidence that the particular requirements
for a specific intended use are fulfilled. (ISO/IEC
17025)

EURACHEM Guide. The Fitness for Purpose of Analytical Methods: A Laboratory Guide
to Method Validation and Related Topics, 2nd ed. 2014

Selection
Type of Analytical Application

Performance
Characteristic
Selectivity

Identification
Test

Quantitative
test for
impurity

Limit test
for
impurity

Quantification
of main
component

Limit of Detection

Limit of Quantitation

Working range

Trueness (bias)

Precision (repeatability
and intermediate
precision)

x signifies normally validated performance characteristics for four types of


analytical applications
EURACHEM Guide. The Fitness for Purpose of Analytical Methods: A Laboratory Guide
to Method Validation and Related Topics, 2nd ed. 2014

Definition
6

Reagent Blank - reagents used during the analytical


process (including solvents used for extraction or
digestion) are analyzed in order to determine if they
contribute to measurement signal. The signal arising
from the analyte can then be corrected accordingly.
Method Blank reagent blanks that are taken
through the whole measurement procedure or
method process

EURACHEM Guide. The Fitness for Purpose of Analytical Methods: A Laboratory Guide
to Method Validation and Related Topics, 2nd ed. 2014

Definition
7

Sample Blank these are essentially sample matrices


with no analyte present. These are difficult to obtain
but such materials are necessary to give a realistic
estimate of interferences that would be encountered
in the analysis of test samples.

EURACHEM Guide.The Fitness for Purpose of Analytical Methods: A Laboratory Guide


to Method Validation and Related Topics, 2nd ed. 2014

Performance Characteristics

Method Performance Characteristics


9

Selectivity
Limit of Detection (LOD) and Limit of Quantitation
(LOQ)
Working Range
Trueness (bias, recovery)
Precision (repeatability, intermediate precision,
reproducibility)
Ruggedness (robustness)

EURACHEM Guide. The Fitness for Purpose of Analytical Methods: A Laboratory Guide
to Method Validation and Related Topics, 2nd ed. 2014

Method Performance Characteristics


10

***Measurement Uncertainty is not a performance


characteristic but a property of the results obtained
using that method.

EURACHEM Guide. The Fitness for Purpose of Analytical Methods: A Laboratory Guide
to Method Validation and Related Topics, 2nd ed. 2014

Selectivity
11

The ability of a method to determine accurately and


specifically the analyte of interest in the presence of
other components in a sample matrix under the stated
conditions of the test. [EURACHEM Guide]

EURACHEM Guide. The Fitness for Purpose of Analytical Methods: A Laboratory Guide
to Method Validation and Related Topics, 2nd ed. 2014

Selectivity
12

What to do

How many times

Analyze test sample


containing various
suspected
interferences in the
presence of the
analytes of interest

What to calculate
from the data
Examine
interference effect.

EURACHEM Guide - Selectivity

EURACHEM Guide. The Fitness for Purpose of Analytical Methods: A Laboratory Guide
to Method Validation and Related Topics, 2nd ed. 2014

Comment
If detection or
quantification is
inhibited by the
interferences, further
method development
will be required.

Concepts on Low Concentration


Measurements
13

1. Establish a value that an analyte level is significantly


different from zero (critical value or decision limit)
2. Know the lowest concentration of the analyte that can
be detected by the method at a specified level of
confidence (limit of detection, minimum detectable
value or detection limit)
3. Establish the lowest level at which the performance is
acceptable for a typical application (limit of
quantitation)
EURACHEM Guide. The Fitness for Purpose of Analytical Methods: A Laboratory Guide
to Method Validation and Related Topics, 2nd ed. 2014

Limit of Detection
14

The lowest amount of analyte in a sample which can


be detected but not necessarily quantitated as an
exact value. [ICH Q2A and ICH Q2B, ASEAN Guide]

The lowest content that can be measured with


reasonable statistical certainty. [AOAC-PVMC]

EURACHEM Guide. The Fitness for Purpose of Analytical Methods: A Laboratory Guide
to Method Validation and Related Topics, 2nd ed. 2014

S0
(standard deviation from
results of measurement)

Limit of Detection

15

Results blank
corrected
during
routine?

YES

NO

s0 '

s0
n

Use the calculated s0 for


computing the value of LOD
and LOQ

EURACHEM Guide. The Fitness for Purpose of Analytical Methods: A Laboratory Guide
to Method Validation and Related Topics, 2nd ed. 2014

s 0 ' s0

1 1

n nb

Limit of Detection
16

Instrument Detection Limit


is based on sample analysis, often a reagent blank
presented directly to the instrument (omitting sample
preparation steps). When the reagent blanks do not go
through the whole measurement procedure, the
calculated data will give the Instrument LOD.

EURACHEM Guide. The Fitness for Purpose of Analytical Methods: A Laboratory Guide
to Method Validation and Related Topics, 2nd ed. 2014

EURACHEM Guide LOD (Instrument Detection Limit)

What to do
Replicate measurements
of reagent blanks

How many What to calculate from the


Comment
times
data
10
Calculate standard
for non-blank corrected results
deviation of the results, s0.

s0 '

or

Calculate s0.

Replicate measurements
of reagent blanks spiked
with low concentrations
of analyte

Calculate LOD

LOD 3 s0 '

This approach is
acceptable when blank
samples or test samples
at low concentrations
cannot be obtained

s0

For blank-corrected results

s0 ' s0

1
1

n nb

S0 standard deviation of result


at or near zero concentration
S0 standard deviation for
calculating LOD and LOQ
n number of replicate
observations averaged when
reporting results

nb number of blank observations


averaged when calculating blank
correction
EURACHEM Guide.The Fitness for Purpose of Analytical Methods: A Laboratory Guide
to Method Validation and Related Topics, 2nd ed. 2014

17

Limit of Detection
18

Method Detection Limit


LOD is based on the analysis of samples that have been
taken through the whole measurement procedure using
results calculated with the same equation as the test
sample.
Samples that are suitable for estimation of LOD or LOQ
are: blank samples (matrices with no detectable
analyte), test samples with concentrations of analyte
close to or below the expected LOD.
EURACHEM Guide. The Fitness for Purpose of Analytical Methods: A Laboratory Guide
to Method Validation and Related Topics, 2nd ed. 2014

EURACHEM Guide LOD (Method Detection Limit)


What to do
19Replicate measurements

of blank samples
or

How many
times
10

What to calculate from the


Comment
data
Calculate standard
for non-blank corrected results
deviation of the results, s0.
s

s0 '

Calculate s0.

Replicate measurements
of test samples with low
analyte concentration

Calculate LOD

LOD 3 s0 '

or
Replicate measurement
of reagent blanks
Or

Replicate measurements
of reagent blanks spiked
with low analyte
concentration
EURACHEM Guide. The Fitness for Purpose of Analytical Methods: A Laboratory Guide
to Method Validation and Related Topics, 2nd ed. 2014

For blank-corrected results

s0 ' s0

1
1

n nb

S0 standard deviation of result


at or near zero concentration
S0 standard deviation for
calculating LOD and LOQ
n number of replicate
observations averaged when
reporting results
nb number of blank
observations averaged when
calculating blank correction

Limit of Detection
20

LOD based on standard deviation of QC results at low


concentration levels under intermediate precision
conditions.
1.
2.

The standard deviation s0 is equal to s0


Since blank correction is no longer required, LOD
can be computed directly by using s0
multiplied by 3.

EURACHEM Guide. The Fitness for Purpose of Analytical Methods: A Laboratory Guide
to Method Validation and Related Topics, 2nd ed. 2014

Limit of Quantitation
21

The lowest level of analyte that can be determined


with acceptable performance.

EURACHEM Guide. The Fitness for Purpose of Analytical Methods: A Laboratory Guide
to Method Validation and Related Topics, 2nd ed. 2014

EURACHEM Guide LOQ


What to do
22Replicate measurements

of blank samples
or

How many
times
10

What to calculate from the


Comment
data
Calculate standard
for non-blank corrected results
deviation of the results, s0.
s

s0 '

Calculate s0.

Replicate measurements
of test samples with low
analyte concentration
or
Replicate measurements
of reagent blanks

Calculate LOQ

IUPAC default kQ value is


usually 10

For blank-corrected results

s0 ' s0

1
1

n nb

LOQ kQ s0 '

S0 standard deviation of result


at or near zero concentration

LOQ 10 s0 '

S0 standard deviation for


calculating LOD and LOQ

or

Replicate measurements
of reagent blanks spiked
with low concentrations
of analyte
EURACHEM Guide. The Fitness for Purpose of Analytical Methods: A Laboratory Guide
to Method Validation and Related Topics, 2nd ed. 2014

n number of replicate
observations averaged when
reporting results
nb number of blank
observations averaged when
calculating blank correction

Limit of Quantitation
23

LOQ based on the standard deviation of quality control


results at low concentration levels under intermediate
precision conditions.

1. The standard deviation s0 is equal to s0


2. Since blank correction is no longer required, LOQ
can be computed directly by using s0 multiplied by
10.

EURACHEM Guide. The Fitness for Purpose of Analytical Methods: A Laboratory Guide
to Method Validation and Related Topics, 2nd ed. 2014

Working Range
24

The interval over which the method provides results


with an acceptable uncertainty.
The lower end is bounded by the limit of
quantification and the upper end of the working
range is defined by concentrations at which significant
anomalies in analytical sensitivity are observed.

EURACHEM Guide. The Fitness for Purpose of Analytical Methods: A Laboratory Guide
to Method Validation and Related Topics, 2nd ed. 2014

Working Range
25

Instrument Working Range


In method validation, the instrument working range is
assessed to: confirm the relationship (e.g. linear,
curvilinear, etc) of the instrument response between the
LOQ and upper end of the instrument working range;

EURACHEM Guide. The Fitness for Purpose of Analytical Methods: A Laboratory Guide
to Method Validation and Related Topics, 2nd ed. 2014

Working Range
26

Method Working Range


Assessing method working range requires use of sample
blanks and samples with known concentrations. These
samples shall undergo the entire measurement
procedure (method) and preferably cover the whole
working range. Results are calculated based on the
method.

EURACHEM Guide. The Fitness for Purpose of Analytical Methods: A Laboratory Guide
to Method Validation and Related Topics, 2nd ed. 2014

EURACHEM Guide Working and linear range


How many
What to do
What to calculate from the data
times
1.Measure blank and
1
Plot
27calibration standards at 6Response y-axis
10 concentrations (evenly
Concentration x-axis
spaced across the range of
Visually identify approximate
interest)
linear range and upper and lower
working range boundaries. Proceed
to step 2.
2.Measure blank plus 6-10
1
Plot
calibration standards, 2 to
Response y-axis
3 times. (calibration
Concentration x-axis
standard concentrations
Calculate appropriate regression
evenly spaced across
statistics. Calculate and plot
linear range.
residuals (difference between
observed y-value and calculated yvalue predicted by the straight line,
for each x value). Linearity is
confirmed when there is random
distribution of residuals. Systematic
trends indicate non-linearity or
change in variance with level
EURACHEM Guide. The Fitness for Purpose of Analytical Methods: A Laboratory Guide
to Method Validation and Related Topics, 2nd ed. 2014

Comment
Visual confirmation if
working range is linear
or not

If the standard deviation


is proportional to
concentration, consider
using weighted
regression rather than
simple non-weighted
linear regression.
In certain circumstances, it
may be better to try to
fit a non-linear curve to
the data. Functions higher
than quadratic are
generally not advised.

EURACHEM Guide Working and linear range


How
What to do
many
What to calculate from the data
times
283.Calibrate according to
1
Plot measured concentration y-axis against
the proposed calibration
test sample concentration x-axis.
procedure. Measure
Visually identify approximate linear range
according to the
and upper and lower working range
method, blank plus RMs
boundaries.
or spiked sample blanks
Calculate appropriate regression statistics.
2 to 3 times at 6-10
Calculate and plot residuals (difference
different concentrations
between observed y-value and calculated
evenly spaced across
y-value predicted by the straight line, for
the range of interest.
each x value). Linearity is confirmed when
there is random distribution of residuals.
Systematic trends indicate non-linearity

EURACHEM Guide. The Fitness for Purpose of Analytical Methods: A Laboratory Guide
to Method Validation and Related Topics, 2nd ed. 2014

Comment
This is to assess
whether instrument
range and calibration
procedure are fit for
purpose.

Accuracy
29

The closeness of agreement between a measured


quantity value and a true quantity value of a
measurand [VIM 2.13]

Therefore, it is normally studied as two components,


trueness and precision. In addition, an increasingly
common expression of accuracy is measurement
uncertainty.

EURACHEM Guide. The Fitness for Purpose of Analytical Methods: A Laboratory Guide
to Method Validation and Related Topics, 2nd ed. 2014

Trueness (bias)
30

The closeness of agreement between the average of


an infinite number of replicate measured quantity
values and a reference quantity value. [VIM 2.14]

Trueness is inversely related to systematic


measurement error which may be estimated as bias.

EURACHEM Guide. The Fitness for Purpose of Analytical Methods: A Laboratory Guide
to Method Validation and Related Topics, 2nd ed. 2014

EURACHEM Guide Trueness


How many
What to do
times
10
31Measure RM using
candidate method

What to calculate from the data


Calculate
bias,

Comment
bias taking into
account effect of
method and
laboratory bias

b x xref
%relative bias,

b (%)

x xref
xref

100

% relative recovery (apparent


recovery)

R (%)
Measure matrix
blanks/test samples
unspiked and spiked
with the analyte of
interest over a range
of concentrations

10

x
xref

100

Calculate
relative spike recovery,

R ' (%)

x 'spiked x
xspike

EURACHEM Guide. The Fitness for Purpose of Analytical Methods: A Laboratory Guide
to Method Validation and Related Topics, 2nd ed. 2014

assess net recovery of


added spike

100

EURACHEM Guide Trueness


How many
What to do
times
10
32Measure RM/test
sample using
candidate method
and alternative
method

What to calculate from the data


Calculate
bias,

alternative method
may be a reference
method

b x xref
%relative bias,

b (%)

x xref
xref

100

% relative recovery (apparent


recovery)

R (%)

x
xref

Comment

100

EURACHEM Guide. The Fitness for Purpose of Analytical Methods: A Laboratory Guide
to Method Validation and Related Topics, 2nd ed. 2014

Estimated Recovery
33

Analyte
Concentration, %

Analyte
Ratio

Unit

Mean
% Recovery

100

100%

98 102

> 10

10-1

10%

98 102

>1

10-2

1%

97 103

> 0.1

10-3

0.10%

95 105

0.01

10-4

100 ppm

90 107

0.001

10-5

10 ppm

80 110

0.0001

10-6

1 ppm

80 110

0.00001

10-7

100 ppb

80 110

0.000001

10-8

10 ppb

60 115

0.0000001

10-9

1 ppb

40 120

AOAC Manual Peer Verified Methods

Precision
34

The closeness of agreement between indications or


measured quantity values obtained by replicate
measurements on the same or similar objects under
specified conditions. [VIM 2.15]
Measurement precision is related to random
measurement error [VIM 2.19] and is a measure of
how close results are to one another.

EURACHEM Guide. The Fitness for Purpose of Analytical Methods: A Laboratory Guide
to Method Validation and Related Topics, 2nd ed. 2014

Precision
35

Repeatability
is a measure of the variability in results when a
measurement is performed by a single analyst using
the same equipment over a short timescale.

Intermediate Precision
gives an estimate of the variation in results when
measurements are made in a single laboratory but
under conditions that are more variable than
repeatability conditions (different analysts, extended
timescale, different pieces of equipment, etc).
EURACHEM Guide. The Fitness for Purpose of Analytical Methods: A Laboratory Guide
to Method Validation and Related Topics, 2nd ed. 2014

Precision
36

Reproducibility
is a measure of the variability in results between
laboratories using the same method or using different
methods but intending to measure the same quantity.

EURACHEM Guide. The Fitness for Purpose of Analytical Methods: A Laboratory Guide
to Method Validation and Related Topics, 2nd ed. 2014

EURACHEM Guide Precision


How many
What to do
times
a. same analyst
6 to 15
37and equipment,
replicates for
same laboratory, each material
short timescale
b. different
6 to 15
analysts and
replicates for
equipment, same
each material
laboratory,
extended
timescale
c. different
analysts and
equipment, same
laboratory,
extended
timescale

6 to 15 groups
of duplicate
measurements
obtained under
repeatability
conditions on
different days/
equipment for
each material

What to calculate from the


Comment
data
Calculate standard
Estimates Repeatability standard
deviation of results for each deviation sr for each material
material
Calculate standard
Estimates Intermediate Precision
deviation of results for each standard deviation sI for each
material
material

Calculate repeatability
standard deviation from
ANOVA results for each
material.
Calculate between-group
standard deviation from
ANOVA and combine with
repeatability standard
deviation for each material

EURACHEM Guide. The Fitness for Purpose of Analytical Methods: A Laboratory Guide
to Method Validation and Related Topics, 2nd ed. 2014

Estimates Repeatability standard


deviation sr for each material
Estimates Intermediate Precision
standard deviation sI for each
material

EURACHEM Guide Precision


How many
What to do
times
d. different
6 to 15 groups
38analysts and
of duplicate
equipment,
measurements
different
obtained under
laboratory,
repeatability
extended
conditions in
timescale
different
laboratories for
each material

What to calculate from the


Comment
data
Calculate repeatability
Estimates Repeatability standard
standard deviation from
deviation sr for each material
ANOVA results for each
material.
Estimates Reproducibility standard
deviation sR for each material. This
Calculate betweenrequires a special inter-laboratory
laboratory standard
comparison (collaborative trial)
deviation from ANOVA
results and combine with
repeatability standard
deviation for each material

EURACHEM Guide. The Fitness for Purpose of Analytical Methods: A Laboratory Guide
to Method Validation and Related Topics, 2nd ed. 2014

Estimated Precision
39

Analyte
Concentration, %

Unit

%RSD

100

100%

1.3

10

10%

2.3

1%

2.7

0.1

0.10%

3.7

0.01

100 ppm

5.3

0.001

10 ppm

7.3

0.0001

1 ppm

11

0.00001

100 ppb

15

0.000001

10 ppb

21

0.0000001

1 ppb

30

AOAC Manual Peer Verified Methods

Ruggedness
40

The ruggedness of an analytical procedure is a


measure of its capacity to remain unaffected by
small, but deliberate variations in method parameters
and provides an indication of its reliability during
normal usage. [ICH Q2A and ICH Q2B, ASEAN
Guide]

EURACHEM Guide. The Fitness for Purpose of Analytical Methods: A Laboratory Guide
to Method Validation and Related Topics, 2nd ed. 2014

EURACHEM Guide Ruggedness


What to do
Identify variables which
41could have a significant
effect on method
performance.
Set up experiments
(analysing RMs or test
samples) to monitor the
effect on measurement
results of systematically
changing the variables.

How many times

What to calculate from the data

Most effectively
evaluated using
experimental
designs.
e.g. 7 parameters
can be studied in
8 experiments
using a
Plackett-Burman
experimental
design.

Determine the effect of each


change of condition on the
measurement results.
Rank the variables in order of
the greatest effect on method
performance.
Carry out significance tests to
determine whether observed
effects are statistically
significant.

EURACHEM Guide. The Fitness for Purpose of Analytical Methods: A Laboratory Guide
to Method Validation and Related Topics, 2nd ed. 2014

Comment
Design quality
control or modify
the method in order
to control the
critical variables,
e.g. by stating
suitable tolerance
limits in the
standard operating
procedure

Analytical Sensitivity
42

Analytical sensitivity is the change in instrument


response which corresponds to a change in the
measured quantity

EURACHEM Guide. The Fitness for Purpose of Analytical Methods: A Laboratory Guide
to Method Validation and Related Topics, 2nd ed. 2014

Analytical Sensitivity
43

Useful Applications:
1.
Theoretical analytical sensitivity is sometimes known.
(e.g. ion-selective electrode show a Nernstian
behavior.)
2.
In spectrophotometric analysis, absorbance can be
predicted from Beer-Lamberts law.

EURACHEM Guide. The Fitness for Purpose of Analytical Methods: A Laboratory Guide
to Method Validation and Related Topics, 2nd ed. 2014

*Measurement Uncertainty
44

Interval associated with the measurement result which


expresses the range of values that can reasonably be
attributed to the quantity being measured.
It is not a performance characteristic but a property
of the results obtained using the method.

EURACHEM Guide. The Fitness for Purpose of Analytical Methods: A Laboratory Guide
to Method Validation and Related Topics, 2nd ed. 2014

45

Exercise

Exercise
46

Method Validation for Determination of Fe in food


sample was done using a AAS. The expected Fe
content of the sample is around 100g/g.
a.
What are the performance characteristics that
should be included in the study?
b.
The method for routine requires duplicate analysis
for the blank and test sample. Results were
g/mL 50mL
calculated as : Fe, g/g AAS Reading,
Weight Sample, g
Calculate for LOD and LOQ using the data
obtained from the validation study.

47

Trial

Wt (g)

Method Blank
Sample Blank
Reading (g/mL) Reading (g/mL)

1.0003

0.0691

0.2070

1.0003

0.0604

0.2017

1.0003

0.0702

0.2051

1.0001

0.0756

0.1998

1.0003

0.0740

0.2097

1.0006

0.0631

0.2615

1.0000

0.0633

0.1908

1.0000

0.0482

0.2042

1.0004

0.0561

0.1907

10

1.0004

0.0509

0.2179

You might also like