Professional Documents
Culture Documents
tasks on the image-guided radiation therapy (IGRT) equipment. QA procedures are performed on
a daily, monthly, and annual basis to ensure image and shifting accuracy. In attempts to stay
current with safe practices, all tasks performed in this department are based on recommendations
from both the TG-179 and TG-142 report.
Daily IGRT QA:
Each morning, the therapists will perform daily QA of our on-board imaging (OBI) and
electronic portal imaging device (EPID). For the OBI, both a KV orthog pair and a CBCT are
performed separately using a cube with specifically located fiducials (Figure 1). The cube is first
aligned to the main cross-hairs and shifts to these fiducials are verified for positioning accuracy.
The main goal of this task is to ensure that the necessary shifts being requested in the imaging
software are accurately shifted inside the treatment room.
Figure 3. Equipment and images acquired in monthly QA. A: imaging cube used for
image/radiation isocenter. B: Leed phantom used in KV analysis. C: Catphan phantom used in
CBCT analysis. D: KV scaling grid. E: Las Vegas phantom used in MV imaging analysis. F: MV
portal of Las Vegas phantom.
Annual IGRT QA:
The physics staff will perform annual QA of the IGRT equipment, which is generally a little
more detailed than the monthly measurements. All IGRT tests performed during the monthly
QA are performed during the annual QA process as well. In addition, the EPID is tested for its
full range of travel SDD (Tolerance +/- 5mm). Finally, CBCT measurements will be acquired on
two different CBCT scans (standard dose head and pelvis) utilizing both the full and half fan
filters.