iCAP Q
Software Manual
Revision C - 1288010
Legal Notices
2013 Thermo Fisher Scientific Inc. All rights reserved.
Viton is a registered trademark of DuPont Performance Elastomers L.L.C.. Excel, Microsoft, and
Windows are registered trademarks of Microsoft Corporation. CETAC is a trademark of SD
Acquisition, Inc..
All other trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries.
Thermo Fisher Scientific Inc. provides this document to its customers with a product purchase to use
in the product operation. This document is copyright protected and any reproduction of the whole or
any part of this document is strictly prohibited, except with the written authorization of Thermo
Fisher Scientific Inc.
Contacting Us
There are several ways to contact Thermo Fisher Scientific.
Assistance
For technical support and ordering information, please visit:
www.thermoscientific.com/ms
Service contact details are available under:
www.unitylabservice.com
For brochures, application notes and other material, please visit:
www.thermoscientific.com
For operating manuals, please contact your local Service organization.
Thermo Scientific
Contacting Us
Suggestions to the Manual
ii
Thermo Scientific
Contents
Thermo Scientific
Chapter 1
Chapter 2
Chapter 3
Configurator ...............................................................................3-1
User Interface of the Configurator Tool ............................ 3-2
Viewer Region................................................................... 3-3
Access Control Editor ....................................................... 3-4
Setting User Levels ......................................................... 3-6
Granting Access Rights .................................................. 3-7
Element Editor.................................................................. 3-9
Changing the Properties of an Element or Isotope ......... 3-9
Changing the Default Isotope ...................................... 3-11
Experiment Configurator ................................................ 3-13
Creating a New Experiment Configuration.................. 3-14
Editing the Settings of Instruments .............................. 3-17
Loading Experiment Configurations ............................ 3-20
Creating a Preset Configuration ................................... 3-21
Deleting a Configuration ............................................. 3-22
Hardware Configurator................................................... 3-24
Hardware Panel Configurator ......................................... 3-26
Molecule Editor .............................................................. 3-28
Report Editor.................................................................. 3-31
Script Editor ................................................................... 3-33
Settings ........................................................................... 3-34
Standard Editor............................................................... 3-35
Changing the Default Concentration ........................... 3-37
Creating a New Elemental Standard............................. 3-38
Creating a New Internal Standard................................ 3-39
Creating a New Isotope Dilution Standard .................. 3-42
iii
Contents
iv
Chapter 4
Chapter 5
Thermo Scientific
Contents
Thermo Scientific
Templates ...................................................................................6-1
Template Toolbar ............................................................. 6-2
Evaluation Methods .......................................................... 6-9
Color Scheme of the Periodic Table ................................ 6-11
Method Parameters ......................................................... 6-14
Analytes ....................................................................... 6-15
Acquisition Parameters................................................. 6-18
Monitor Analytes ......................................................... 6-24
Survey Scan Settings..................................................... 6-26
Interference Correction ................................................ 6-30
Standards ..................................................................... 6-32
Compounds (tQuant only) .......................................... 6-46
Peak Detection (tQuant only) ...................................... 6-50
Regions (trQuant only) ................................................ 6-60
Quantification.............................................................. 6-62
Ratios........................................................................... 6-67
Quality Control (eQuant only) .................................... 6-72
Defining Detection Limits (eQuant only) .................. 6-101
Defining QC Settings in Sample Definition (eQuant
only) .......................................................................... 6-107
Peripherals .................................................................... 6-111
Cetac ASX-520 Autosampler...................................... 6-111
Cetac ASX-260 Autosampler...................................... 6-112
Chromeleon System ................................................... 6-114
ESI SC-4Q Autosampler ............................................ 6-119
ESI Fast Option ......................................................... 6-123
SpectraSystem LC Autosampler ................................. 6-124
SpectraSystem LC Pump............................................ 6-126
Accela LC Autosampler.............................................. 6-128
Accela LC Pump ........................................................ 6-130
iCAP Q Software Manual (P/N 1288010, Revision C)
Contents
vi
Chapter 7
Chapter 8
Chapter 9
Thermo Scientific
Contents
Thermo Scientific
vii
Contents
viii
Thermo Scientific
Chapter 1
Thermo Scientific
1-1
Typographical Conventions
This section describes typographical conventions that have been
established for Thermo Fisher Scientific manuals.
Signal Words
Make sure you follow the precautionary statements presented in this
manual. The special notices appear different from the main flow of text:
NOTICE Points out possible material damage and other important
information in connection with the instrument.
Data Input
Throughout this manual, the following conventions indicate data input
and output via the computer:
1-2
For brevity, expressions such as choose File > Directories are used
rather than pull down the File menu and choose Directories.
Any button that you click on the screen is represented in bold face
letters. For example, click Close.
Thermo Scientific
Topic Headings
The following headings are used to show the organization of topics
within a chapter:
Chapter 1
Chapter Name
Thermo Scientific
1-3
Reference Documentation
In addition to this iCAP Q Software Manual, Thermo Fisher Scientific
provides the following documents for the iCAP Q instrument:
1-4
Thermo Scientific
Chapter 2
Introduction to Qtegra
This iCAP Q Software Manual describes the Thermo Scientific
Qtegra Intelligent Scientific Data Solution (ISDS) configurable
software package for elemental analyses for configuration and operation
of the Thermo Scientific iCAP Q mass spectrometer. Qtegra is a true
end-to-end solution for workflow-driven analysis. You can use this suite
of applications for a variety of Thermo Fisher Scientific products.
The main Qtegra frameworks introduced in this chapter are:
Contents
Thermo Scientific
Configurator Overview
Qtegra Overview
2-1
Introduction to Qtegra
Configurator Overview
Configurator Overview
The Configurator tool is used by the Administrator of your network and
the Manager of your laboratory. Different applets are provided to edit
general settings of the hardware and software and to configure and
adjust the Thermo Scientific Qtegra framework for your laboratory.
The user interface of the Configurator tool is shown in Figure 2-1:
Figure 2-1.
2-2
Thermo Scientific
Introduction to Qtegra
Configurator Overview
Applet
Description
Element Editor
Experiment Configurator
Hardware Configurator
Molecule Editor
Report Editor
Script Editor
Settings
Standard Editor
Thermo Scientific
2-3
Introduction to Qtegra
Configurator Overview
1. Click
2-4
to open Configurator.
Thermo Scientific
Introduction to Qtegra
Instrument Control Overview
Thermo Scientific
2-5
Introduction to Qtegra
Instrument Control Overview
Figure 2-2.
2-6
Thermo Scientific
Introduction to Qtegra
Instrument Control Overview
1. Click
Thermo Scientific
2-7
Introduction to Qtegra
Qtegra Overview
Qtegra Overview
The Qtegra tool is the main Qtegra ISDS module and is used to design,
start and stop the measurements. The Home Page tab offers access to all
pages of the Qtegra tool. By default, the Home Page opens on the
Dashboard page.
The analytical workflow is based on the design of a measurement in a
Template. Sample analysis and data acquisition is then performed in
LabBooks created from the Template with appropriate information on
the samples.
The user interface of the Qtegra tool is shown in Figure 2-3:
Figure 2-3.
1. Click
2-8
to open Qtegra.
Thermo Scientific
Introduction to Qtegra
Qtegra Software and Compliance with 21 CFR Part 11
Thermo Scientific
2-9
Introduction to Qtegra
Qtegra Software and Compliance with 21 CFR Part 11
2-10
Thermo Scientific
Chapter 3
Configurator
The Configurator tool contains all tools necessary to configure and
adjust the Qtegra framework for your laboratory.
Contents
Viewer Region
Element Editor
Experiment Configurator
Hardware Configurator
Molecule Editor
Report Editor
Script Editor
Settings
Standard Editor
To open the Configurator tool
1. Click
Thermo Scientific
to open Configurator.
3-1
Configurator
User Interface of the Configurator Tool
Labeled Components: 1=display region for applet settings, 2=Viewer region, 3=list of applets
Figure 3-1.
3-2
Thermo Scientific
Configurator
Viewer Region
Viewer Region
The Viewer region (see Figure 3-2) of the Configurator tool displays a
list of log files, such as errors and warnings.
Figure 3-2.
1. Click
to open Configurator.
Thermo Scientific
3-3
Configurator
Access Control Editor
Figure 3-3.
3-4
Thermo Scientific
Configurator
Access Control Editor
Name
Description
Administrator
Manager
User
Thermo Scientific
3-5
Configurator
Access Control Editor
Name
Description
Full Access
Read Only
Hidden
1. Click
2. Click
to open Configurator.
1. Click
2. Click
to open Configurator.
Thermo Scientific
Configurator
Access Control Editor
5. Click
Figure 3-4.
1. Click
2. Click
Thermo Scientific
to open Configurator.
3-7
Configurator
Access Control Editor
Figure 3-5.
6. Select the new access right for the user group, for example, Read
only.
The new access rights are defined for this Virtual instruments or
Virtual evaluations item.
7. Click
3-8
Thermo Scientific
Configurator
Element Editor
Element Editor
The Element editor applet of the Configurator tool gives access to the
properties of the element and isotope table.
The Element editor applet contains a list of all elements and their
properties sorted in the order of their atomic number, see Figure 3-6.
The information listed here will be used for all experiments.
Figure 3-6.
1. Click
2. Click
to open Configurator.
Element editor.
Thermo Scientific
3-9
Configurator
Element Editor
1. Click
2. Click
to open Configurator.
Element editor.
to save
1. Click
2. Click
to open Configurator.
Element editor.
Figure 3-7.
3-10
to exit the
Thermo Scientific
Configurator
Element Editor
1. Click
2. Click
to open Configurator.
Element editor.
Figure 3-8.
Thermo Scientific
3-11
Configurator
Element Editor
7. Click
3-12
Thermo Scientific
Configurator
Experiment Configurator
Experiment Configurator
The Experiment configurator applet of the Configurator tool
combines instrument sets. Each combination is saved as specific
Experiment Configuration for later use in the Qtegra tool when creating
a Template, and in Instrument Control.
NOTICE Access to this module is defined in the Access Control Editor
on page 3-4. Generally, only the Administrator and the Manager have
full access to this module.
In Experiment configurator, all virtual instruments, virtual evaluation
types and preset configurations are listed on tabbed pages, see
Figure 3-9.
Figure 3-9.
Thermo Scientific
3-13
Configurator
Experiment Configurator
1. Click
to open Configurator.
2. Click
Experiment configurator.
3-14
1. Click
to open Configurator.
2. Click
Experiment Configurator.
Thermo Scientific
Configurator
Experiment Configurator
3. Click
New to add a new Experiment Configuration, see
Figure 3-10.
Thermo Scientific
3-15
Configurator
Experiment Configurator
3-16
Thermo Scientific
Configurator
Experiment Configurator
7. Drag and drop iCAP Q from the tab Instruments to your new
Configuration.
Drop when
Thermo Scientific
3-17
Configurator
Experiment Configurator
1. Click
to open Configurator.
2. Click
Experiment Configurator.
3-18
Thermo Scientific
Configurator
Experiment Configurator
4. Click
window, see Figure 3-15.
Thermo Scientific
3-19
Configurator
Experiment Configurator
7. Change the value or click the arrow to display the drop-down menu
if available, see Figure 3-16.
.
to save the Configuration.
1. Click
to open Configurator.
2. Click
Experiment Configurator.
3. Click
to load the Experiment Configurations from the
configurations database.
3-20
Thermo Scientific
Configurator
Experiment Configurator
1. Click
to open Configurator.
2. Click
Experiment Configurator.
Thermo Scientific
3-21
Configurator
Experiment Configurator
1. Click
to open Configurator.
2. Click
Experiment Configurator.
4. Select
from the context menu to
add the Configuration to the list of Preset Configurations.
5. Click
Deleting a Configuration
In the Experiment configurator of the Configurator tool,
Configurations can be deleted from the configurations database.
1. Click
to open Configurator.
2. Click
Experiment Configurator.
3-22
Thermo Scientific
Configurator
Experiment Configurator
4. Click
to delete the selected Configuration.
The Delete Configuration dialog opens, see Figure 3-18.
5. Click
6. Click
Thermo Scientific
3-23
Configurator
Hardware Configurator
Hardware Configurator
The Hardware configurator applet of the Configurator tool gives
access to hardware databases on electrotechnical items level.
NOTICE Generally, these settings are factory-set and do not need to be
modified.
Hardware configurator (see Figure 3-19) comprises all settings of
interfaces and hardware devices.
3-24
Thermo Scientific
Configurator
Hardware Configurator
Table 3-4.
Commands
Thermo Scientific
1. Click
to open Configurator.
2. Click
Hardware configurator.
3-25
Configurator
Hardware Panel Configurator
3-26
Thermo Scientific
Configurator
Hardware Panel Configurator
Table 3-5.
Commands
1. Click
2. Click
Thermo Scientific
to open Configurator.
3-27
Configurator
Molecule Editor
Molecule Editor
The Molecule Editor applet of the Configurator tool allows the
Administrator and Manager to create molecules (or polyatomics) which
are automatically added to the Molecules tab of the Method Parameter
Analytes (see Analytes on page 6-15) and in Instrument Control (see
Analytes Tab on page 4-5).
Molecule Editor, see Figure 3-21, shows the periodic table as well as the
fields to enter elements and molecules.
3-28
Thermo Scientific
Configurator
Molecule Editor
1. Click
2. Click
to open Configurator.
Molecule Editor.
To create molecules
1. Click
2. Click
to open Configurator.
Molecule Editor.
3. Click the elements in the periodic table to select the elements for the
molecule to be created.
Alternatively, type in the relevant symbols for the elements of the
molecule, separated by a period <.>, see Figure 3-22.
Thermo Scientific
3-29
Configurator
Molecule Editor
8. For user-defined elements, enter Symbol and Mass, see Figure 3-24.
3-30
Thermo Scientific
Configurator
Report Editor
Report Editor
The Report Editor applet of the Configurator tool allows you to create
new report templates or to edit existing templates.
Report editor (see Figure 3-25) determines the layout of the reports.
Thermo Scientific
3-31
Configurator
Report Editor
1. Click
to open Configurator.
2. Click
Report editor.
1. Click
to open Configurator.
2. Click
3. Click
Report editor.
to create a new Report template.
1. Click
to open Configurator.
2. Click
3. Click
Report editor.
and browse for the report file.
3-32
Thermo Scientific
Configurator
Script Editor
Script Editor
The Script Editor applet of the Configurator tool allows you to create
and compile C# scripts for virtual instruments.
Script Editor (see Figure 3-26) is designed to be used by persons
experienced with C# scripts.
1. Click
2. Click
Thermo Scientific
to open Configurator.
Script Editor.
3-33
Configurator
Settings
Settings
The Settings applet of the Configurator tool controls default settings,
for example, the default directory path for Qtegra or the default settings
for dwell time.
Settings (see Figure 3-27) gives access to the settings database (registry).
To open Settings
1. Click
2. Click
3-34
to open Configurator.
Settings.
Thermo Scientific
Configurator
Standard Editor
Standard Editor
The Standard Editor applet of the Configurator tool gives access to the
standards database. New standard files, internal standard files and
isotope dilution standard files are created here.
Standard Editor (see Figure 3-28) shows a list of standards on the left.
On the right the associated elements and their concentration in the
standard solution are displayed as well as the periodic table.
Thermo Scientific
3-35
Configurator
Standard Editor
Table 3-7.
Commands
1. Click
2. Click
to open Configurator.
Standard Editor.
1. Click
2. Click
to open Configurator.
Standard Editor.
3. Click
.
All standards are loaded from the database.
1. Click
2. Click
to open Configurator.
Standard Editor.
3-36
Thermo Scientific
Configurator
Standard Editor
4. Click
.
The standards are saved to the database.
1. Click
2. Click
to open Configurator.
Standard Editor.
5. Click
The standard is deleted from the database.
6. Click
1. Click
2. Click
Thermo Scientific
to open Configurator.
Standard Editor.
3-37
Configurator
Standard Editor
3. Click
to open the Set Default Concentration window, see
Figure 3-30.
1. Click
2. Click
to open Configurator.
Standard Editor.
3. Click
to check the default concentration.
Change if appropriate as described in Changing the Default
3-38
Thermo Scientific
Configurator
Standard Editor
5. Click
window.
Thermo Scientific
3-39
Configurator
Standard Editor
1. Click
2. Click
to open Configurator.
Standard Editor.
3. Click
.
The drop-down menu opens, see Figure 3-32.
4. Click
Internal Standard (Isotopical) to open the Add New
Standard window.
5. Enter the Standard Name and a Standard Description.
6. Click
to add the file.
The new internal standard is added to the list on the left. An empty
page opens containing the table columns No, Isotope,
Concentration and Unit, and the periodic table of elements with all
available isotope information.
3-40
Thermo Scientific
Configurator
Standard Editor
1. Click
2. Click
to open Configurator.
Standard Editor.
3. Click
to check the default concentration.
Change if appropriate as described in Changing the Default
Concentration on page 3-37.
Thermo Scientific
3-41
Configurator
Standard Editor
4. Click
.
The drop-down menu opens, see Figure 3-34.
5. Click
Internal Standard (Elemental) to open the Add New
Standard window.
6. Enter a Standard Name and a Standard Description.
7. Click
to add the file.
The new standard is added to the list on the left. An empty page
opens containing the table columns No, Element, Concentration
and Unit and the periodic table of elements with all available isotope
information.
8. Add elements to the standard table by clicking on the element in the
periodic table.
The default isotope of the element is added to the table.
Concentration and Unit are added according to the default
concentration.
9. To remove the element, click the respective element in the periodic
table again.
10. Repeat until all elements have been added.
11. Click
3-42
Thermo Scientific
Configurator
Standard Editor
1. Click
2. Click
to open Configurator.
Standard Editor.
3. Click
.
The drop-down menu opens, see Figure 3-35.
4. Click
Isotope Dilution Standard to open the Add New
Standard window.
5. Enter the Standard Name and a Standard Description.
6. Click
to add the file.
The new isotope dilution standard is added to the list on the left. An
empty page opens containing the table columns No, Element,
Concentration, Unit, Isotope 1, Isotope 2, Abundance 1,
Abundance 2 and Atomic Weight, and the periodic table of
elements with all available isotope information.
7. Click an element in the periodic table to add it to the table.
Thermo Scientific
3-43
Configurator
Standard Editor
3-44
Thermo Scientific
Chapter 4
Instrument Control
The Instrument Control tool is used to perform instrument calibrations
and to edit general instrument controls such as tune settings and
measurement modes or change stabilization times.
Contents
Control Panel
Status Panel
1. Click
Thermo Scientific
4-1
Instrument Control
User Interface of the Instrument Control Tool
Labeled Components: 1=data view region, 2=Log View region, 3=Control and Status Panel region, 4=ribbon tabs
Figure 4-1.
4-2
Thermo Scientific
Instrument Control
User Interface of the Instrument Control Tool
The ribbon tabs (4 in Figure 4-1) are displayed in accordance with the
Configuration loaded. By default the Experiment Configuration and
the Window tab are displayed. Additional tabs are added for each
instrument.
1. Click
Figure 4-2.
Thermo Scientific
4-3
Instrument Control
Data View Region
Figure 4-3.
4-4
Thermo Scientific
Instrument Control
Data View Region
1. Click
Figure 4-4.
Thermo Scientific
4-5
Instrument Control
Data View Region
1. Click
4-6
Thermo Scientific
Instrument Control
Data View Region
Figure 4-5.
Thermo Scientific
4-7
Instrument Control
Data View Region
Table 4-1.
Icon
4-8
Thermo Scientific
Instrument Control
Data View Region
1. Click
Figure 4-6.
1. Click
Thermo Scientific
4-9
Instrument Control
Data View Region
3. In the data view region, select the tab of the peripheral you wish to
view.
4-10
Thermo Scientific
Instrument Control
Experiment Configuration Ribbon Tab
Figure 4-7.
To load a Configuration
1. Click
Thermo Scientific
4-11
Instrument Control
The iCAP Q Ribbon Tab
Figure 4-8.
1. Click
Control Group
The Control group of the iCAP Q ribbon tab in Instrument Control,
see Figure 4-9, offers basic commands to switch the iCAP Q system on
and off and to start and stop the scanning of the instrument. The results
are displayed in real time.
Figure 4-9.
4-12
Thermo Scientific
Instrument Control
The iCAP Q Ribbon Tab
Description
On
Off
Run
Stop
Restart
NOTICE To start the plasma, see also Qtegra chapter Getting Ready
on page 5-6.
1. Click
4. Click
.
The plasma is switched on.
Thermo Scientific
4-13
Instrument Control
The iCAP Q Ribbon Tab
1. Click
1. Click
1. Click
1. Click
4-14
Thermo Scientific
Instrument Control
The iCAP Q Ribbon Tab
3. Click
in the group Control of the iCAP Q tab.
The confirmation window opens, see Figure 4-11.
4. Click
.
The plasma is switched off.
STD - standard mode, the mode of operation where the cell is not
pressurized
Thermo Scientific
4-15
Instrument Control
The iCAP Q Ribbon Tab
Description
CCT
KED
STD
CCTS
KEDS
STDS
Edit
Select
Apply Tune
Settings
Tune Settings
1. Click
4-16
Thermo Scientific
Instrument Control
The iCAP Q Ribbon Tab
The settings last saved for this Measurement mode are loaded into
Instrument Control.
1. Click
Thermo Scientific
in
4-17
Instrument Control
The iCAP Q Ribbon Tab
4. Select the Tune Setting you wish to view on the left, for example,
STD, see Figure 4-14.
4-18
Thermo Scientific
Instrument Control
The iCAP Q Ribbon Tab
5. Click
.
The selected settings are loaded and Last Loaded Tune Settings
shows the corresponding Tune Setting, in this example, STD, see
Figure 4-14.
Thermo Scientific
4-19
Instrument Control
The iCAP Q Ribbon Tab
7. Click
Figure 4-16.
1. Click
4-20
Thermo Scientific
Instrument Control
The iCAP Q Ribbon Tab
Thermo Scientific
4-21
Instrument Control
The iCAP Q Ribbon Tab
The drop-down list Measurement mode lists all modes defined. For
each Measurement mode you can define the Added stabilization
time [s], a delay time that is used when switching from one mode to
another within an analysis. The table Excluded Ranges shows the
Begin and End range for each excluded range. The excluded ranges refer
to protected zones of the mass spectrum which are not scanned in a
survey run.
The table Current tune setting lists a history of tune settings defined
for this Measurement mode, according to the date they were created.
All changes are automatically applied to the selected mode in the Edit
view of the data view region, no extra saving is necessary. The changes
have no effect on the currently loaded settings.
1. Click
1. Click
4-22
Thermo Scientific
Instrument Control
The iCAP Q Ribbon Tab
1. Click
Thermo Scientific
4-23
Instrument Control
The iCAP Q Ribbon Tab
4. Click
.
The Add Measurement Mode dialog opens, see Figure 4-19.
1. Click
4-24
Thermo Scientific
Instrument Control
The iCAP Q Ribbon Tab
6. Click
.
The selected measurement mode is deleted from the list.
1. Click
Display Group
The Display group of the iCAP Q ribbon tab in Instrument Control,
see Figure 4-21, allows you to select different views for the real-time
display data in both the Data Display tab and the Average Intensities
view of the data view region.
Thermo Scientific
4-25
Instrument Control
The iCAP Q Ribbon Tab
The buttons of the Display group are summarized in Table 4-4. They
take effect in all modes.
Table 4-4.
Item
Description
Calibrated
Analog
Ion
Counting
Show
Analog
Normal
Average
History
Spectrum
4-26
Thermo Scientific
Instrument Control
The iCAP Q Ribbon Tab
For details on the data view region display, see Data Display Tab on
page 4-7.
1. Click
3. Click
Calibrated,
Analog or
Ion Counting to
select the desired detector mode.
The data in the Data Display tab of the data view region is presented
according to the selection. The Average Intensities list changes
respectively.
1. Click
Wizards Group
The wizards available in the Wizards group of the iCAP Q ribbon tab
in Instrument Control, see Figure 4-22, help tuning the system.
Thermo Scientific
4-27
Instrument Control
The iCAP Q Ribbon Tab
Description
Performance
Report
Autotune
Detector Setup
4-28
Thermo Scientific
Instrument Control
The iCAP Q Ribbon Tab
1. Click
Thermo Scientific
4-29
Instrument Control
The iCAP Q Ribbon Tab
4-30
Thermo Scientific
Instrument Control
The iCAP Q Ribbon Tab
Thermo Scientific
4-31
Instrument Control
The iCAP Q Ribbon Tab
4-32
Thermo Scientific
Instrument Control
The iCAP Q Ribbon Tab
11. Drag and drop analytes from Selected analytes to Defined ratios to
define the ratios for the analytes needed to determine, for example,
the oxide ratio or the ratio of doubly charged ions, see Figure 4-27.
Thermo Scientific
4-33
Instrument Control
The iCAP Q Ribbon Tab
4-34
Thermo Scientific
Instrument Control
The iCAP Q Ribbon Tab
15. Define the mass calibration test criteria, see Figure 4-29.
Thermo Scientific
4-35
Instrument Control
The iCAP Q Ribbon Tab
4-36
Thermo Scientific
Instrument Control
The iCAP Q Ribbon Tab
1. Click
Thermo Scientific
4-37
Instrument Control
The iCAP Q Ribbon Tab
4-38
Thermo Scientific
Instrument Control
The iCAP Q Ribbon Tab
Thermo Scientific
4-39
Instrument Control
The iCAP Q Ribbon Tab
4-40
Thermo Scientific
Instrument Control
The iCAP Q Ribbon Tab
Thermo Scientific
4-41
Instrument Control
The iCAP Q Ribbon Tab
11. Define the test criteria for the Performance report, see Figure 4-36.
4-42
Thermo Scientific
Instrument Control
The iCAP Q Ribbon Tab
13. Define the mass calibration test criteria for the Performance report,
see Figure 4-37.
Figure 4-37. Defining mass calibration tests for the Performance Report Wizard
Thermo Scientific
4-43
Instrument Control
The iCAP Q Ribbon Tab
14. Select a report from the list or enter a new Name, see Figure 4-38.
Figure 4-38. Selecting Performance Report name for the new report
15. Click Next to save the Performance Report.
4-44
Thermo Scientific
Instrument Control
The iCAP Q Ribbon Tab
1. Click
Thermo Scientific
4-45
Instrument Control
The iCAP Q Ribbon Tab
4-46
Thermo Scientific
Instrument Control
The iCAP Q Ribbon Tab
Thermo Scientific
4-47
Instrument Control
The iCAP Q Ribbon Tab
7. Click Next.
The performance report requires the samples to be placed, see
Figure 4-42.
4-48
Thermo Scientific
Instrument Control
The iCAP Q Ribbon Tab
8. Place the probe into the solution if not already done so and click
Next.
The acquisition status is shown, see Figure 4-43.
Thermo Scientific
4-49
Instrument Control
The iCAP Q Ribbon Tab
4-50
Thermo Scientific
Instrument Control
The iCAP Q Ribbon Tab
1. Click
Thermo Scientific
4-51
Instrument Control
The iCAP Q Ribbon Tab
4-52
Thermo Scientific
Instrument Control
The iCAP Q Ribbon Tab
5. Click Next.
The performance report requires the samples to be placed, see
Figure 4-47.
Thermo Scientific
4-53
Instrument Control
The iCAP Q Ribbon Tab
6. Place the probe into the solution if not already done so and click
Next.
The acquisition status is shown, see Figure 4-48.
4-54
Thermo Scientific
Instrument Control
The iCAP Q Ribbon Tab
Thermo Scientific
4-55
Instrument Control
The iCAP Q Ribbon Tab
8. Select the check box Open Report if you wish to open the report,
see Figure 4-50.
Autotune Wizard
The Wizards group of the iCAP Q ribbon tab in Instrument Control
gives access to the Autotune Wizard.
An autotune procedure can be necessary when the performance of the
instrument falls below the limits specified in the performance report and
all other measures like cleaning or exchanging of the sample
introduction system have been exhausted.
Furthermore, a special Source Autotune procedure is recommended to
be run when parts of the sample introduction system like nebulizer or
cones, have been exchanged, or the system is being equipped with a
different gas type (for example O2/He mixture instead of pure He) for
operation of the QCell.
4-56
Thermo Scientific
Instrument Control
The iCAP Q Ribbon Tab
1. Click
Thermo Scientific
4-57
Instrument Control
The iCAP Q Ribbon Tab
4-58
Thermo Scientific
Instrument Control
The iCAP Q Ribbon Tab
Thermo Scientific
4-59
Instrument Control
The iCAP Q Ribbon Tab
9. Select the controls and ranges for the autotune sequence, see
Figure 4-54.
4-60
Thermo Scientific
Instrument Control
The iCAP Q Ribbon Tab
11. Select the analytes used to tune the instrument from the periodic
table and select the dwell time for each, see Figure 4-55.
Thermo Scientific
4-61
Instrument Control
The iCAP Q Ribbon Tab
4-62
Thermo Scientific
Instrument Control
The iCAP Q Ribbon Tab
15. Drag and drop analytes from Selected analytes to Defined ratios to
define the ratios for your analytes, see Figure 4-57.
Thermo Scientific
4-63
Instrument Control
The iCAP Q Ribbon Tab
4-64
Thermo Scientific
Instrument Control
The iCAP Q Ribbon Tab
18. Select an existing Autotune sequence from the list if you wish to
overwrite it or enter a new name, see Figure 4-59.
Thermo Scientific
4-65
Instrument Control
The iCAP Q Ribbon Tab
1. Click
4-66
Thermo Scientific
Instrument Control
The iCAP Q Ribbon Tab
Thermo Scientific
4-67
Instrument Control
The iCAP Q Ribbon Tab
7. Select the controls and ranges for your new autotune sequence, see
Figure 4-63.
Figure 4-63. Select controls and ranges of sequence for new autotune sequence
8. Specify appropriate default values for controls that are kept constant
during the autotune procedure.
9. Click Next.
4-68
Thermo Scientific
Instrument Control
The iCAP Q Ribbon Tab
10. Select the analytes used to tune the instrument from the periodic
table and select the dwell time for each, see Figure 4-64.
Figure 4-64. Select dwell time for analytes for new autotune sequence
11. Click Next.
Thermo Scientific
4-69
Instrument Control
The iCAP Q Ribbon Tab
12. In the tab Molecules, select molecules used to tune the instrument,
if appropriate, see Figure 4-65.
4-70
Thermo Scientific
Instrument Control
The iCAP Q Ribbon Tab
14. Drag and drop analytes from Selected analytes to Defined ratios to
define the ratios for your analyte, see Figure 4-66.
Thermo Scientific
4-71
Instrument Control
The iCAP Q Ribbon Tab
4-72
Thermo Scientific
Instrument Control
The iCAP Q Ribbon Tab
17. Enter Name and Description for the new autotune sequence, and
details about the required solution as appropriate., see Figure 4-68.
Thermo Scientific
4-73
Instrument Control
The iCAP Q Ribbon Tab
1. Click
4-74
Thermo Scientific
Instrument Control
The iCAP Q Ribbon Tab
Thermo Scientific
4-75
Instrument Control
The iCAP Q Ribbon Tab
7. Select the autotune sequence you wish to run, see Figure 4-72.
4-76
Thermo Scientific
Instrument Control
The iCAP Q Ribbon Tab
8. Click Next.
The Load the sample dialog opens, see Figure 4-73.
Thermo Scientific
4-77
Instrument Control
The iCAP Q Ribbon Tab
4-78
Thermo Scientific
Instrument Control
The iCAP Q Ribbon Tab
When the tuning process is completed, the Tune Data View waits
for you to browse through the results, see Figure 4-75.
Thermo Scientific
4-79
Instrument Control
The iCAP Q Ribbon Tab
4-80
Thermo Scientific
Instrument Control
The iCAP Q Ribbon Tab
Source Autotune
The Wizards group of the iCAP Q ribbon tab in Instrument Control
gives access to the Source Autotune wizard.
For the modi STD, CCT and KED, Source Autotune will be started
with High Matrix, for STDS, CCTS and KEDS with High Sensitivity,
or as defined, see Editing a Measurement Mode on page 4-21.
Thermo Scientific
4-81
Instrument Control
The iCAP Q Ribbon Tab
1. Click
4-82
Thermo Scientific
Instrument Control
The iCAP Q Ribbon Tab
Thermo Scientific
4-83
Instrument Control
The iCAP Q Ribbon Tab
4-84
Thermo Scientific
Instrument Control
The iCAP Q Ribbon Tab
When the tuning process is completed, the Tune Data View waits
for you to browse through the results, see Figure 4-81.
Thermo Scientific
4-85
Instrument Control
The iCAP Q Ribbon Tab
7. Click Next.
The Acquisition status dialog opens, see Figure 4-82.
4-86
Thermo Scientific
Instrument Control
The iCAP Q Ribbon Tab
8. Click Next.
The final dialog opens, see Figure 4-83.
Thermo Scientific
4-87
Instrument Control
The iCAP Q Ribbon Tab
1. Click
4-88
Thermo Scientific
Instrument Control
The iCAP Q Ribbon Tab
Thermo Scientific
4-89
Instrument Control
The iCAP Q Ribbon Tab
6. Click Next.
The periodic table window opens, see Figure 4-85.
4-90
Thermo Scientific
Instrument Control
The iCAP Q Ribbon Tab
7. Click Next.
The Load the Sample window opens, see Figure 4-86.
Thermo Scientific
4-91
Instrument Control
The iCAP Q Ribbon Tab
9. Click Next.
The Waiting for sample uptake window opens, see Figure 4-87.
4-92
Thermo Scientific
Instrument Control
The iCAP Q Ribbon Tab
11. To begin the setup before the delay time elapsed, click Next.
The Analog Offset Determination starts, see Figure 4-88.
Thermo Scientific
4-93
Instrument Control
The iCAP Q Ribbon Tab
4-94
Thermo Scientific
Instrument Control
The iCAP Q Ribbon Tab
Thermo Scientific
4-95
Instrument Control
The iCAP Q Ribbon Tab
4-96
Thermo Scientific
Instrument Control
The iCAP Q Ribbon Tab
To perform a detector high voltage setup and cross calibration with the
Detector Setup Wizard
1. Click
Thermo Scientific
4-97
Instrument Control
The iCAP Q Ribbon Tab
4-98
Thermo Scientific
Instrument Control
The iCAP Q Ribbon Tab
6. Click Next.
The periodic table window opens, see Figure 4-94.
Thermo Scientific
4-99
Instrument Control
The iCAP Q Ribbon Tab
7. Click Next.
The Load the Sample window opens, see Figure 4-95.
4-100
Thermo Scientific
Instrument Control
The iCAP Q Ribbon Tab
9. Click Next.
The Waiting for sample uptake window opens, see Figure 4-96.
Thermo Scientific
4-101
Instrument Control
The iCAP Q Ribbon Tab
11. To begin the setup before the delay time elapsed, click Next.
The analog baseline determination starts, see Figure 4-97.
4-102
Thermo Scientific
Instrument Control
The iCAP Q Ribbon Tab
Thermo Scientific
4-103
Instrument Control
The iCAP Q Ribbon Tab
4-104
Thermo Scientific
Instrument Control
The iCAP Q Ribbon Tab
1. Click
Thermo Scientific
4-105
Instrument Control
The iCAP Q Ribbon Tab
4-106
Thermo Scientific
Instrument Control
The iCAP Q Ribbon Tab
6. Click Next.
The periodic table window opens, see Figure 4-102.
Thermo Scientific
4-107
Instrument Control
The iCAP Q Ribbon Tab
7. Click Next.
The Load the Sample window opens, see Figure 4-103.
4-108
Thermo Scientific
Instrument Control
The iCAP Q Ribbon Tab
9. Click Next.
The Waiting for sample uptake window opens, see Figure 4-104.
Thermo Scientific
4-109
Instrument Control
The iCAP Q Ribbon Tab
11. To begin the setup before the delay time elapsed, click Next.
The analog baseline and analog amplifier determination for the
analog offset determination starts, see Figure 4-105.
4-110
Thermo Scientific
Instrument Control
The iCAP Q Ribbon Tab
Thermo Scientific
4-111
Instrument Control
The iCAP Q Ribbon Tab
4-112
Thermo Scientific
Instrument Control
The iCAP Q Ribbon Tab
Thermo Scientific
4-113
Instrument Control
The iCAP Q Ribbon Tab
4-114
Thermo Scientific
Instrument Control
The iCAP Q Ribbon Tab
Thermo Scientific
4-115
Instrument Control
The iCAP Q Ribbon Tab
4-116
Thermo Scientific
Instrument Control
The iCAP Q Ribbon Tab
1. Click
Thermo Scientific
4-117
Instrument Control
The iCAP Q Ribbon Tab
4-118
Thermo Scientific
Instrument Control
The iCAP Q Ribbon Tab
5. Click Next.
The Coarse Mass Calibration window opens, see Figure 4-113.
Thermo Scientific
4-119
Instrument Control
The iCAP Q Ribbon Tab
4-120
Thermo Scientific
Instrument Control
The iCAP Q Ribbon Tab
If you do not select a setting from the list, see Figure 4-115, the tune
settings currently loaded will be used.
Thermo Scientific
4-121
Instrument Control
The iCAP Q Ribbon Tab
8. Click Next.
The Load the Sample window opens, see Figure 4-116.
4-122
Thermo Scientific
Instrument Control
The iCAP Q Ribbon Tab
9. Place your sample probe into the solution if not already done so and
click Next, see Figure 4-117.
Figure 4-117. Waiting for the sample uptake in Mass Calibration wizard
The Mass Calibration wizards applies a minimum delay time for
sample uptake in order to assure that enough sample has entered the
plasma.
10. Click Pause to delay further if more time is needed.
Click Continue when ready.
11. To begin the calibration before the delay time elapsed, click Next.
The data acquisition for the mass calibration starts.
12. In the data view region, select the iCAP Q tab.
The ribbon iCAP Q is activated.
Thermo Scientific
4-123
Instrument Control
The iCAP Q Ribbon Tab
4-124
Thermo Scientific
Instrument Control
The iCAP Q Ribbon Tab
Thermo Scientific
4-125
Instrument Control
The iCAP Q Ribbon Tab
16. Click Finish to finish the Mass Calibration and store the acquired
parameters, see Figure 4-120.
Views Group
The Views group of the iCAP Q ribbon tab in Instrument Control, see
Figure 4-121, allows you to view instrument calibrations, performance
reports and a real-time readback plot of the Control Panel parameters.
4-126
Thermo Scientific
Instrument Control
The iCAP Q Ribbon Tab
Description
Readback Plot
Performance
Report
1. Click
Thermo Scientific
4-127
Instrument Control
The iCAP Q Ribbon Tab
1. Click
4-128
Thermo Scientific
Instrument Control
The iCAP Q Ribbon Tab
1. Click
Thermo Scientific
4-129
Instrument Control
The iCAP Q Ribbon Tab
1. Click
4-130
Thermo Scientific
Instrument Control
The iCAP Q Ribbon Tab
1. Click
Thermo Scientific
4-131
Instrument Control
The iCAP Q Ribbon Tab
4-132
Thermo Scientific
Instrument Control
The iCAP Q Ribbon Tab
1. Click
Thermo Scientific
4-133
Instrument Control
Window Ribbon Tab
1. Click
To add a layout
1. Click
4-134
Thermo Scientific
Instrument Control
Window Ribbon Tab
To delete a layout
1. Click
Thermo Scientific
4-135
Instrument Control
Window Ribbon Tab
1. Click
1. Click
4-136
Thermo Scientific
Instrument Control
Control Panel
Control Panel
The Control Panel tab (Figure 4-132) of Instrument Control contains
several pages for interactive tuning and instrument monitoring.
Thermo Scientific
4-137
Instrument Control
Control Panel
1. Click
2. Click the Control Panel tab on the lower left side of the Instrument
Control window.
1. Click
2. Click the Control Panel tab on the lower left side of the Instrument
Control window.
3. Click
at the bottom of the Control Panel tab.
The Configure buttons menu opens, see Figure 4-133.
4-138
Thermo Scientific
Instrument Control
Control Panel
to move it
to move it
9. Click
1. Click
2. Click the Control Panel tab on the lower left side of the Instrument
Control window.
3. Click
at the bottom of the Control Panel tab.
The Configure buttons menu opens, see Figure 4-135.
Thermo Scientific
4-139
Instrument Control
Control Panel
6. Click an activated
page to deactivate the display.
The selection menu closes and the page is removed from the display.
Main
The Main page (see Figure 4-137) of the Control Panel tab in
Instrument Control contains the iCAP Q parameters that are most
commonly used and typically have the most effect on the performance
of the instrument.
Parameter
Description
Torch Horizontal
Position [mm]
4-140
Thermo Scientific
Instrument Control
Control Panel
Inlet
The Inlet page (Figure 4-138) of the Control Panel tab in Instrument
Control contains iCAP Q parameters for the inlet system.
Thermo Scientific
Parameter
Description
Peristaltic Pump
Peristaltic Pump
Speed [rpm]
4-141
Instrument Control
Control Panel
Table 4-8.
Parameter
Description
Spray Chamber
Temperature [C]
4-142
Thermo Scientific
Instrument Control
Control Panel
Plasma
The Plasma page (Figure 4-139) of the Control Panel tab in Instrument
Control shows the current gas flow of cool, auxiliary and nebulizer gas,
as well as the RF generator parameters.
Thermo Scientific
Parameter
Description
Argon supply
4-143
Instrument Control
Control Panel
Table 4-9.
Parameter
Description
RF Generator Supply
Voltage [V]
RF Generator
Supply Current [A]
Q-Cell
The Q-Cell page (Figure 4-140) of the Control Panel tab in Instrument
Control shows the parameters and state of the QCell.
4-144
Thermo Scientific
Instrument Control
Control Panel
The tuning parameters of the Q-Cell page are described in Table 4-10.
Table 4-10. Tuning parameters of the Q-Cell page
Parameter
Description
Advanced
The Advanced page (Figure 4-141) of the Control Panel tab in
Instrument Control contains further parameters for detector and lenses
that are important for the daily operation of the system.
Thermo Scientific
4-145
Instrument Control
Control Panel
Description
Extraction Lens 1
Positive [V]
Detector Voltage
(Analog) [V]
Detector Voltage
(Counting) [V]
Vacuum
The Vacuum page (Figure 4-142) of the Control Panel tab in
Instrument Control shows the parameters and state of the vacuum
system.
4-146
Thermo Scientific
Instrument Control
Control Panel
The tuning parameters of the Vacuum page are described in Table 4-12.
Table 4-12. Tuning parameters of the Vacuum page
Parameter
Description
Vacuum System
Cooling
The Cooling page (Figure 4-143) of the Control Panel tab in
Instrument Control shows the parameters related to the cooling system
of the instrument and to other components like the Peltier Cooling of
the spray chamber.
Thermo Scientific
4-147
Instrument Control
Control Panel
The tuning parameters of the Cooling page are described in Table 4-13.
Table 4-13. Tuning parameters of the Cooling page
4-148
Parameter
Description
Peltier Temperature
Hot Side [C]
Thermo Scientific
Instrument Control
Status Panel
Status Panel
In the Status Panel tab (Figure 4-144) of the Instrument Control tool
you manage your scripts.
1. Click
2. Click the Status Panel tab on the lower left side of the Instrument
Control window.
1. Click
2. Click the Status Panel tab on the lower left side of the Instrument
Control window.
Thermo Scientific
4-149
Instrument Control
Status Panel
3. Click
in the toolbar of Script List.
The Open dialog opens, see Figure 4-145.
4-150
Thermo Scientific
Instrument Control
Status Panel
5. Click
.
The script is loaded into the Status Panel, see Figure 4-146.
To run a script
1. Click
2. Click the Status Panel tab on the lower left side of the Instrument
Control window.
3. Select the script you wish to execute in the Script List of the Status
Panel.
4. Click
in the toolbar of Script List.
The selected script is executed.
To stop a script
1. Click
2. Click the Status Panel tab on the lower left side of the Instrument
Control window.
3. Select the script you wish to execute in the Script List of the Status
Panel.
Thermo Scientific
4-151
Instrument Control
Status Panel
4. Click
in the toolbar of Script List.
The selected script is executed.
5. Click
in the toolbar of Script List.
The script execution is stopped.
To debug a script
1. Click
2. Click the Status Panel tab on the lower left side of the Instrument
Control window.
3. Select the script you wish to debug in the Script List of the Status
Panel.
4. Click
in the toolbar of Script List.
The selected script is debugged if debugging has been activated.
To reset a script
1. Click
2. Click the Status Panel tab on the lower left side of the Instrument
Control window.
3. Select the script you wish to reset in the Script List of the Status
Panel.
4. Click
in the toolbar of Script List.
The selected script is reset.
To edit a script
1. Click
2. Click the Status Panel tab on the lower left side of the Instrument
Control window.
3. In the Script List, select the script you wish to edit.
4-152
Thermo Scientific
Instrument Control
Status Panel
4. Click
in the toolbar of Script List.
The script editor opens, see Figure 4-147.
7. Click
.
The script editor closes.
1. Click
Thermo Scientific
4-153
Instrument Control
Status Panel
2. Click the Status Panel tab on the lower left side of the Instrument
Control window.
3. Select the script you wish to delete from the Script List in the
Status Panel.
4. Click
in the toolbar of Script List.
The selected script is removed from the Script List in the Status
Panel.
4-154
Thermo Scientific
Instrument Control
Log View Region
1. Click
Thermo Scientific
4-155
Instrument Control
Log View Region
region as soon as you click anywhere outside the data region. Simply
click the remaining tab below the Control Panel tab to show the Log
View region again.
4-156
Thermo Scientific
Chapter 5
Qtegra
Qtegra is the principal tool for preparing and running measurements.
The Qtegra framework is the main Qtegra module and is used to design,
start and stop the measurements.
Contents
LabBooks Page
Templates Page
Help Page
Scheduler
Completed LabBooks
1. Click
Thermo Scientific
to open Qtegra.
5-1
Qtegra
User Interface of the Qtegra Tool
Labeled Components: 1=Home Page tab, 2=LogView tab, 3=Completed LabBooks tab, 4=Scheduler tab
Figure 5-1.
5-2
Thermo Scientific
Qtegra
User Interface of the Qtegra Tool
1. Click
to open Qtegra.
Figure 5-2.
area to hide the selected tab when the cursor leaves this area and click
to cancel.
5. Click and drag the title bar of Scheduler, Log View or Completed
LabBooks.
6. Move the cursor over the position indicators, see Figure 5-3.
Figure 5-3.
Thermo Scientific
5-3
Qtegra
User Interface of the Qtegra Tool
The selected tab is shown in the background, and the new position
of the tab is indicated as colored, see Figure 5-4.
Figure 5-4.
5-4
Thermo Scientific
Qtegra
Dashboard Page of Qtegra
Labeled Components: 1=Get Ready icon, 2=current Configuration status, 3=iCAP Q icon to change Configuration,
4=real-time display, 5=important parameters of the iCAP Q system, 6=current Configuration, 7=Get Ready progress
Figure 5-5.
Thermo Scientific
5-5
Qtegra
Dashboard Page of Qtegra
1. Click
to open Qtegra.
3. Click
.
The Dashboard page of Qtegra opens.
Getting Ready
The Get Ready function on the Dashboard page of Qtegra helps to
start the instrument. It switches on the plasma and waits for the
instrument to warm up. Then the performance report is started with
modes defined by the operator. Once the performance report is passed,
the instrument is ready for operation.
If the performance report fails, autotune or mass calibration are started
automatically, followed by the performance report, again as specified by
the operator.
1. Click
to open Qtegra.
5-6
Thermo Scientific
Qtegra
Dashboard Page of Qtegra
3. Click
.
The Get Ready dialog opens, see Figure 5-6.
Figure 5-6.
4. If you select the check box Warm up, enter the time in minutes.
5. Select the check box Use Manual Sample.
6. Enter the Wash time (s) and the Uptake time (s).
The default for both of 30 s is usually sufficient.
7. Select a check box for Measurement Modes, for example, STD.
Thermo Scientific
5-7
Qtegra
Dashboard Page of Qtegra
8. Click
.
The plasma is switched on. During warm-up the Dashboard shows
the remaining warm-up time, see Figure 5-15.
Figure 5-7.
The Verifying Performance test starts with the selected standard, see
Figure 5-8.
Figure 5-8.
Figure 5-9.
10. Move the sample probe to the wash position and click
5-8
Thermo Scientific
Qtegra
Dashboard Page of Qtegra
The washing starts. After washing, a dialog opens, see Figure 5-10.
After washing, you are asked to position the solution for mass
calibration, see Figure 5-12.
Figure 5-12. Waiting for response dialog for mass calibration solution
13. With manual sampling, aspirate your solutions according to the test
and click
.
Sample uptake starts and mass calibration is performed.
Thermo Scientific
5-9
Qtegra
Dashboard Page of Qtegra
14. Click
1. Click
to open Qtegra.
5-10
Thermo Scientific
Qtegra
Dashboard Page of Qtegra
3. Click
.
The Get Ready dialog opens, see Figure 5-14.
Thermo Scientific
5-11
Qtegra
Dashboard Page of Qtegra
9. Click
.
The plasma is switched on. During warm-up the Dashboard shows
the remaining warm-up time, see Figure 5-15.
11. Click
1. Click
to open Qtegra.
5-12
Thermo Scientific
Qtegra
Dashboard Page of Qtegra
3. Click
.
The iCAP Q system closes down and the plasma switches off.
The state of the Configuration changes to NotReady.
1. Click
to open Qtegra.
3. Click
Change Configuration.
The Select Configuration window opens.
Thermo Scientific
5-13
Qtegra
Dashboard Page of Qtegra
5-14
Thermo Scientific
Qtegra
Dashboard Page of Qtegra
1. Click
to open Qtegra.
1. Click
to open Qtegra.
Thermo Scientific
5-15
Qtegra
Dashboard Page of Qtegra
3. For the real-time display, select a Standard from the drop-down list.
4. Select a Mode from the drop-down list.
5. Click
to show the toolbar for the graph and adjust the settings
to your need, see Figure 5-19.
5-16
6. Click
7. Click
Thermo Scientific
Qtegra
Dashboard Page of Qtegra
8. Right-click the diagram to open the context menu, see Figure 5-20.
Thermo Scientific
5-17
Qtegra
LabBooks Page
LabBooks Page
On the LabBooks page of Qtegra, see Figure 5-21, LabBooks are
created and opened.
5-18
Thermo Scientific
Qtegra
LabBooks Page
1. Click
to open Qtegra.
3. Click
.
The LabBooks page of Qtegra opens.
Opening a LabBook
LabBooks are opened either from the LabBooks page of Qtegra which
is described here, or from the File Manager page, see File Manager
Page on page 5-49.
1. Click
to open Qtegra.
Thermo Scientific
5-19
Qtegra
LabBooks Page
3. Below , click
.
The Browse for LabBook window opens, see Figure 5-22.
1. Click
to open Qtegra.
5-20
Thermo Scientific
Qtegra
LabBooks Page
Creating a LabBook
LabBooks are created from blank Templates, existing Templates or from
existing LabBooks on the LabBooks page of Qtegra.
1. Click
Thermo Scientific
to open Qtegra.
5-21
Qtegra
LabBooks Page
for it.
5. Click the second radio button if you wish to Create a new LabBook
from an existing LabBook and select a LabBook Name from the
drop-down list.
You can also enter a name or browse
5-22
for it.
Thermo Scientific
Qtegra
LabBooks Page
6. Click the third radio button if you wish to Create a LabBook from
a blank Template, and select an Evaluation from the drop-down
list.
7. Click
to create the new LabBook.
A new tab opens for the new LabBook.
8. If you created the LabBook from a blank Template, define the
Method Parameters.
See Method Parameters on page 6-14 for details.
9. Click Sample List to check the sample list parameters.
The final Sample List is defined by the number of samples selected
when creating a LabBook. The Sample List in the LabBook is
created from the parameters defined in Sample Definition for a
Template on page 6-135.
10. Click Automated Export to define the data for export.
See Automatic Export - Template on page 6-143 for details.
11. In the toolbar of your LabBook page, click
LabBook.
to save your
Editing a LabBook
LabBooks are edited in Qtegra. Editing a LabBook involves a number of
parameters, see LabBooks on page 7-1 for a complete description of
LabBooks.
To edit a LabBook
1. Click
to open Qtegra.
Thermo Scientific
to save your
5-23
Qtegra
LabBooks Page
Deleting a LabBook
LabBooks are deleted in the File Manager page of Qtegra.
To delete a LabBook
1. Click
to open Qtegra.
5-24
Thermo Scientific
Qtegra
LabBooks Page
4. Right-click the LabBook you wish to delete in the list on the right.
A context menu opens, see Figure 5-26.
Closing a LabBook
LabBooks are closed in Qtegra by clicking the appropriate button in the
toolbar of the LabBook or by simply closing the tab of the LabBook.
To close a LabBook
1. Click
to open Qtegra.
Thermo Scientific
5-25
Qtegra
LabBooks Page
5-26
to close your
Thermo Scientific
Qtegra
Templates Page
Templates Page
On the Templates page of Qtegra, see Figure 5-27, Templates for your
methods are created and opened.
Thermo Scientific
5-27
Qtegra
Templates Page
1. Click
to open Qtegra.
3. Click
.
The Templates page of Qtegra opens.
Opening a Template
Templates are opened either from the Templates page of Qtegra which is
described here, or from the File Manager page, see File Manager Page
on page 5-49.
1. Click
to open Qtegra.
5-28
Thermo Scientific
Qtegra
Templates Page
3. Below , click
.
The Browse for Template window opens, see Figure 5-28.
1. Click
to open Qtegra.
Thermo Scientific
5-29
Qtegra
Templates Page
Creating a Template
Templates are created from blank Templates, existing Templates or
existing LabBooks in Qtegra. For blank Templates, you need to select a
system Configuration. Configurations, including peripherals
(Instruments), are defined by your Administrator in the applet
Experiment Configurator of the Configurator tool (see Experiment
Configurator on page 3-13).
1. Click
5-30
to open Qtegra.
Thermo Scientific
Qtegra
Templates Page
Thermo Scientific
5-31
Qtegra
Templates Page
drop-down list.
You can also enter the name or browse
for it.
6. Click the third radio button if you wish to Create a new Template
from an existing LabBook and select a LabBook Name from the
drop-down list.
You can also enter the name or browse
for it.
7. Click
to create the new Template.
A new tab opens for the new Template.
8. In the tab of your template, define the Method Parameters.
See Method Parameters on page 6-14 for details.
9. Click Sample Definition to set up the sample list parameters.
See Sample Definition for a Template on page 6-135 for details.
The final Sample List is defined by the number of samples selected
when creating a LabBook. The Sample List in the LabBook is
created from the parameters in Sample Definition in the Template.
10. Click Automated Export to define the data for export. See
Automatic Export - Template on page 6-143 for details.
11. In the toolbar of your Template page, click
Template.
to save your
Editing a Template
Templates are edited in Qtegra. See Templates on page 6-1 for a
complete description of the parameters involved.
1. Click
to open Qtegra.
5-32
Thermo Scientific
Qtegra
Templates Page
to save your
Deleting a Template
Templates are deleted in the File Manager page of Qtegra.
To delete a Template
1. Click
to open Qtegra.
Thermo Scientific
5-33
Qtegra
Templates Page
Closing a Template
Templates are closed by clicking the appropriate button in the toolbar of
the Template or by simply closing the tab of the Template.
To close a Template
1. Click
to open Qtegra.
5-34
Thermo Scientific
Qtegra
Templates Page
Thermo Scientific
to close your
5-35
Qtegra
LabBook Query Page
5-36
Thermo Scientific
Qtegra
LabBook Query Page
On the resulting Query page, you select the data you wish to view, see
Figure 5-34.
Thermo Scientific
5-37
Qtegra
LabBook Query Page
1. Click
to open Qtegra.
3. Click
.
The LabBook Query page of Qtegra opens.
5-38
Thermo Scientific
Qtegra
LabBook Query Page
1. Click
to open Qtegra.
Thermo Scientific
5-39
Qtegra
LabBook Query Page
The field LabBooks displays the first entries in the list of results, see
Figure 5-37.
5-40
Thermo Scientific
Qtegra
LabBook Query Page
8. Click
to start the query.
Qtegra collects the data, see Figure 5-38.
Thermo Scientific
5-41
Qtegra
LabBook Query Page
The results are displayed when the query has been executed, see
Figure 5-39.
11. Click
if you wish to hide the units.
The columns Unit are hidden. Repeat to display the units again.
5-42
Thermo Scientific
Qtegra
LabBook Query Page
1. Click
to open Qtegra.
Thermo Scientific
5-43
Qtegra
LabBook Query Page
NOTICE Result data presets saved here will be available as well in the
Query view of a completed LabBook, see Query on page 7-32.
5. Enter a Name for the preset.
6. Enter a Description.
7. Click
.
The preset is added to the list.
To save a preset
1. Click
to open Qtegra.
7. Click
To delete presets
1. Click
to open Qtegra.
5-44
Thermo Scientific
Qtegra
LabBook Query Page
5. Click
to delete the result data preset.
The Delete Preset dialog is displayed, see Figure 5-42.
6. Click
.
The preset is deleted from the list.
Generating Reports
In the Reports view of the LabBook Query page of Qtegra, you
generate result data reports from presets.
Generally, you can structure your report to your needs, add headings,
tables and graphs to specify the presentation of the result data.
NOTICE For details on creating report presets, see Creating Reports
on page 7-41.
1. Click
to open Qtegra.
Thermo Scientific
5-45
Qtegra
LabBook Query Page
To generate a report
1. Click
to open Qtegra.
5-46
Thermo Scientific
Qtegra
LabBook Query Page
Thermo Scientific
5-47
Qtegra
LabBook Query Page
5-48
or save
as desired.
Thermo Scientific
Qtegra
File Manager Page
1. Click
to open Qtegra.
3. Click
.
The File Manager page of Qtegra opens.
1. Click
to open Qtegra.
3. Click
.
The File Manager page of Qtegra opens.
4. Select the directory Templates.
Thermo Scientific
5-49
Qtegra
File Manager Page
1. Click
to open Qtegra.
3. Click
.
The File Manager page of Qtegra opens.
4. Select the directory LabBooks.
5. Double-click the LabBook you wish to open.
The LabBook is opened in a new tab.
1. Click
to open Qtegra.
5-50
Thermo Scientific
Qtegra
File Manager Page
1. Click
to open Qtegra.
Thermo Scientific
5-51
Qtegra
File Manager Page
1. Click
to open Qtegra.
5-52
Thermo Scientific
Qtegra
File Manager Page
1. Click
to open Qtegra.
Thermo Scientific
5-53
Qtegra
File Manager Page
1. Click
to open Qtegra.
5-54
Thermo Scientific
Qtegra
File Manager Page
6. Click
.
The file is deleted.
1. Click
to open Qtegra.
Thermo Scientific
5-55
Qtegra
File Manager Page
5-56
Thermo Scientific
Qtegra
Help Page
Help Page
The Help page of Qtegra, see Figure 5-55, provides information about
Qtegra, support and tools.
1. Click
to open Qtegra.
3. Click
.
The Help page of Qtegra opens.
Thermo Scientific
5-57
Qtegra
Help Page
1. Click
to open Qtegra.
3. Click
Qtegra Help.
1. Click
to open Qtegra.
3. Click
5-58
Options.
Thermo Scientific
Qtegra
Help Page
1. Click
Thermo Scientific
to open Qtegra.
5-59
Qtegra
Help Page
3. Click
Options.
5-60
Thermo Scientific
Qtegra
Help Page
13. Select the check box Automatic next to Start Queue if you wish to
activate this feature.
If the check box has been selected, the measurement starts
immediately when a LabBook is added to the Scheduler.
14. Click
Registering Qtegra
The Tools section on the Help page of Qtegra offers access to the
license manager of Qtegra.
During first installation of Qtegra your Thermo Fisher Scientific field
service engineer enters your license information.
NOTICE It is necessary to run Qtegra as Administrator to activate
Registering.
Additionally, for controlling computers that do not have access to the
Internet, Export and Import functions are provided to export a request
which can then be sent to the licensing website. The generated license
can be imported on the controlling computer.
1. Click
to open Qtegra.
Thermo Scientific
5-61
Qtegra
Help Page
3. Click
Register Qtegra.
The Register Qtegra dialog opens loading license information, see
Figure 5-58.
5-62
Thermo Scientific
Qtegra
Help Page
Thermo Scientific
5-63
Qtegra
Help Page
4. Click
.
The Register Qtegra dialog asks you to choose your licensing mode,
see Figure 5-58.
1. Click
5-64
to open Qtegra.
Thermo Scientific
Qtegra
Help Page
3. Click
Register Qtegra.
5-65
Qtegra
Help Page
5-66
Thermo Scientific
Qtegra
Help Page
8. Click
.
Qtegra is connecting to the licensing server, see Figure 5-63.
Thermo Scientific
5-67
Qtegra
Help Page
1. Click
to open Qtegra.
3. Click
Register Qtegra.
5-68
Thermo Scientific
Qtegra
Help Page
Thermo Scientific
5-69
Qtegra
Help Page
8. Click
.
A dialog opens, see Figure 5-67.
5-70
Thermo Scientific
Qtegra
Help Page
3. Click
to navigate to the folder of your license
request file QtegraLicenseRequest.licreq and select the file.
4. Click
Thermo Scientific
5-71
Qtegra
Help Page
The license file is generated. The website opens with a link to the
license file *.lic, see Figure 5-70.
Figure 5-70. Thermo Scientific Product activation portal with download link
6. Click the link to download the *.lic file.
7. Click
8. Select the folder from where you wish to import the license file onto
your controlling Qtegra computer and click
9. Click
11. Click
Register Qtegra.
Thermo Scientific
Qtegra
Help Page
Thermo Scientific
5-73
Qtegra
Help Page
14. Click
.
The Import a Qtegra license dialog opens, see Figure 5-72.
5-74
Thermo Scientific
Qtegra
Help Page
17. Click
.
The Register Qtegra dialog closes.
18. Restart Qtegra.
Thermo Scientific
5-75
Qtegra
Scheduler
Scheduler
In the Scheduler tool of Qtegra, the measurement for a scheduled
LabBook is executed. The completed LabBook is automatically deleted
from the Scheduler and added to Completed LabBooks on page 5-81.
The Scheduler settings can be customized via the
button in the
Scheduler toolbar, or in the Tools section on the Help page of Qtegra,
see Customizing Scheduler Settings on page 5-59.
NOTICE To move the Scheduler region in Qtegra, see User Interface of
the Qtegra Tool on page 5-2.
1. Click
2. Click
to open Qtegra.
to open the Scheduler tab, see Figure 5-74.
1. Click
to open Qtegra.
5-76
Thermo Scientific
Qtegra
Scheduler
NOTICE If the check box Automatic has been selected for Start Queue
in the Options settings of the Scheduler (see Customizing Scheduler
Settings on page 5-59), the measurement starts immediately.
1. Click
to open Qtegra.
Thermo Scientific
5-77
Qtegra
Scheduler
in the toolbar
1. Click
to open Qtegra.
5-78
in the toolbar
Thermo Scientific
Qtegra
Scheduler
1. Click
to open Qtegra.
Thermo Scientific
5-79
Qtegra
Scheduler
8. Click one of the arrow buttons to move the LabBook to the desired
position in the queue, for example, to the top of the queue, see
Figure 5-77.
5-80
Thermo Scientific
Qtegra
Completed LabBooks
Completed LabBooks
Upon completion of a LabBook, the LabBook is automatically deleted
from Scheduler and added to the Completed LabBooks tab in Qtegra,
see Figure 5-78.
1. Click
to open Qtegra.
2. Click
to open the Completed LabBooks tab.
All LabBooks that have already been executed are listed.
Thermo Scientific
5-81
Qtegra
Log View Region
1. Click
2. Click
to open Qtegra.
to open the Log View tab, see Figure 5-79.
5-82
Thermo Scientific
Chapter 6
Templates
The analytical workflow for sample measurement is defined in a
Template. Templates are created in the Qtegra on page 5-1.
Templates are based on a particular Configuration which is usually
defined by the Manager (see Experiment Configurator on page 3-13)
and reflects your system setup. Each Template consists of a Method
Parameters section, a Sample Definition section, an Automatic Export
section, and a section for the Peripherals if so configured for this
Configuration.
The Method Parameters within a Template are dependent on the
evaluation method assigned to the Template (see Evaluation Methods
on page 6-9). For every application an appropriate Template can be
created.
Contents
Template Toolbar
Evaluation Methods
Method Parameters
Peripherals
1. Click
to open Qtegra.
Thermo Scientific
6-1
Templates
Template Toolbar
Template Toolbar
In the Template tab of Qtegra, Qtegra offers buttons to save, close or
run a Template, see Figure 6-1.
Figure 6-1.
Template toolbar
To save a Template
1. Click
to open Qtegra.
To close a Template
1. Click
to open Qtegra.
6-2
Thermo Scientific
Templates
Template Toolbar
1. Click
to open Qtegra.
Figure 6-2.
5. Click New LabBook if you wish to create a new LabBook from the
existing current Template.
The LabBooks view of the Home Page opens. See Creating a
LabBook on page 5-21 for further details.
6. If you wish to create a new Template from the existing current
Template, click New Template.
The Template view of the Home Page opens. See Creating a
Template on page 5-30 for further details.
1. Click
to open Qtegra.
Thermo Scientific
6-3
Templates
Template Toolbar
4. Click
.
The History window for this Templates opens, see Figure 6-3.
Figure 6-3.
5. Click
1. Click
to open Qtegra.
6-4
Thermo Scientific
Templates
Template Toolbar
5. Press <Ctrl> and select the entries you wish to compare, see
Figure 6-4.
Figure 6-4.
6. Click
to compare the selected entries.
The Comparison dialog opens, see Figure 6-5.
Figure 6-5.
Thermo Scientific
6-5
Templates
Template Toolbar
7. Select the check box Show differences only if you wish to view only
the differences.
8. Click
9. Click
1. Click
to open Qtegra.
Figure 6-6.
5. Select the Template for which you wish to export the audit trail.
6-6
Thermo Scientific
Templates
Template Toolbar
6. Click
.
The Export Audittrail dialog opens, see Figure 6-7.
Figure 6-7.
1. Click
to open Qtegra.
Thermo Scientific
6-7
Templates
Template Toolbar
Figure 6-8.
4. Click
.
The Content pane is hidden, see Figure 6-9.
Figure 6-9.
5. Click
6-8
Thermo Scientific
Templates
Evaluation Methods
Evaluation Methods
Qtegra offers a range of Evaluation methods (see Figure 6-10) to be
selected when creating a Template in Qtegra to accommodate any type
of analysis required.
Thermo Scientific
6-9
Templates
Evaluation Methods
The main applications for the Evaluations are summarized in Table 6-1.
Table 6-1.
Evaluation methods
Evaluation
Description
aQuant
eQuant
Raw Data
Displays the acquired raw intensities which are then used by the
different evaluations.
rQuant
tQuant
trQuant
6-10
Thermo Scientific
Templates
Color Scheme of the Periodic Table
Thermo Scientific
Item
Description
Uncolored
Series
6-11
Templates
Color Scheme of the Periodic Table
Table 6-2.
Item
Description
Block
Ionization Potential
Electronegativity
Electron Affinity
Individual
1. Click
to open Qtegra.
4. Click
6-12
Thermo Scientific
Templates
Color Scheme of the Periodic Table
6. Right-click next to the periodic table (but not on the table itself ) to
open the context menu, see Figure 6-12.
Thermo Scientific
6-13
Templates
Method Parameters
Method Parameters
In the Method Parameters section, you define all measurement settings
for your analytes. The availability of each parameter is controlled by the
type of Evaluation defined for the Template. As an example, the Method
Parameters of a tQuant Template are shown in Figure 6-13.
6-14
Method Parameter
Evaluation
Analytes
Acquisition parameters
Monitor Analytes
Interference Correction
Standards
Compounds
tQuant
Peak Detection
tQuant
Regions
trQuant
Thermo Scientific
Templates
Method Parameters
Table 6-3.
Method Parameter
Evaluation
Quantification
eQuant, aQuant
Ratios
Quality Control
eQuant
Analytes
For all Template types, the analytes to be acquired during the
measurement are selected in the Method Parameter view Analytes of
Qtegra.
Analytes can be selected from the periodic table display in the Elements
page, see Figure 6-14.
Thermo Scientific
6-15
Templates
Method Parameters
On the Molecules page, see Figure 6-15, analytes can be selected from a
tabulated list of the analyte isotopes, and matrix components can be
defined.
1. Click
to open Qtegra.
4. Click
below Method Parameters to open the Analytes view of
the Template.
6-16
Thermo Scientific
Templates
Method Parameters
1. Click
to open Qtegra.
4. Click
below Method Parameters to open the Analytes view of
the Template.
5. Select the Elements page in the Analytes view.
6. Left-click the element in the periodic table to select the default
isotope for this element.
The isotope and its default information stored in the database are
added to the Acquisition Parameters view (Acquisition Parameters
on page 6-18).
7. To deselect an isotope, click the element again.
1. Click
to open Qtegra.
4. Click
below Method Parameters to open the Analytes view of
the Template.
5. Select the Elements page in the Analytes view.
Thermo Scientific
6-17
Templates
Method Parameters
Acquisition Parameters
For all Template types, the list of analytes selected for the Template is
displayed in the Acquisition parameters view of the Qtegra tool.
Acquisition details such as dwell time and number of channels can be
defined.
6-18
Thermo Scientific
Templates
Method Parameters
Acquisition parameters
Column
Description
Identifier
Dwell Time
Displays the dwell time for the selected isotope, for example, the time
spent measuring this analyte on a single channel. By default, this value
is set to 0.01 seconds.
Recommended Settings: Typically, dwell times are related to the
expected concentration of the analyte in the samples and the tune
setting. Major analytes (ppm level) require shorter dwell times. Minor
analytes (ppt, ppb level) require longer dwell times.
Channels
Displays the number of channels used for each peak. The default
number is 1. When entering an even number, the system will
automatically enter the higher odd number.
Spacing
Thermo Scientific
6-19
Templates
Method Parameters
Table 6-4.
Acquisition parameters
Column
Description
Measurement mode
Resolution
1. Click
to open Qtegra.
4. Click
below Method Parameters to open the Acquisition
parameters view of the Template.
1. Click
to open Qtegra.
4. Click
below Method Parameters to open the Acquisition
parameters view of the Template.
6-20
Thermo Scientific
Templates
Method Parameters
5. Click
in the toolbar of the Template.
The list of Possible interferences for the currently selected row in
the table opens, see Figure 6-18.
To duplicate rows
1. Click
to open Qtegra.
4. Click
below Method Parameters to open the Acquisition
parameters view of the Template.
5. Click the gray field in front of the row or rows you wish to duplicate
to select the row or rows.
Thermo Scientific
6-21
Templates
Method Parameters
1. Click
to open Qtegra.
4. Click
below Method Parameters to open the Acquisition
parameters view of the Template.
The column Identifier lists all analytes selected for this Template.
5. Enter Dwell time (s), Channels and Spacing (u) for each analyte,
as appropriate.
With a right-click a cell you open the context menu. You can select,
for example, Fill down or Fill up, as appropriate. Then the entries
from the first selected cell are copied down or up to all cells selected.
6. In the section Advanced Parameters, enter the Number of sweeps.
This option is not available for tQuant or trQuant Templates.
7. Select a Measurement mode for each analyte.
The Measurement mode is displayed in brackets after the analyte in
6-22
Thermo Scientific
Templates
Method Parameters
1. Click
to open Qtegra.
4. Click
below Method Parameters to open the Acquisition
parameters view of the Template.
5. In the Acquisition Parameters table, right-click a cell to open the
context menu.
6. Select
to export the table.
The Save As dialog opens, see Figure 6-20.
Thermo Scientific
6-23
Templates
Method Parameters
Monitor Analytes
For all Template types except tQuant and trQuant, the Monitor
Analytes view of the Qtegra tool is available.
Delays for Uptake and Wash can be defined for the analytes added to
the table, see Figure 6-21.
1. Click
to open Qtegra.
6-24
Thermo Scientific
Templates
Method Parameters
1. Click
to open Qtegra.
7. Type in the Analyte, for example, 43Ca and click another row to
enter.
NOTICE The denomination for the analyte must exactly match that of
the Identifier in the Acquisition Parameters table.
8. Select the check boxes for Uptake to activate monitoring for the
analyte for Signal Above, see Figure 6-22.
Figure 6-22. Monitor Analytes table front section for Signal Above
9. Change the value for Signal Above and Stability (%RSD) if
appropriate.
Thermo Scientific
6-25
Templates
Method Parameters
10. Select Ignore and continue or Skip this sample from the
drop-down menu On Failure. Enter Dwell Time (s) and
Resolution.
11. Select the check boxes for Wash to activate monitoring for the
analyte for Signal Below, see Figure 6-23.
Figure 6-23. Monitor Analytes table end section for Signal Below
12. Change the value for Signal Below if appropriate.
13. Enter Dwell Time (s) and Resolution.
14. Select Ignore and continue or Abort LabBook or Abort queue
from the drop-down menu On Failure.
15. Click
6-26
Thermo Scientific
Templates
Method Parameters
Above the table, the settings are displayed as Scan Regions Graph
which is editable, see Figure 6-24.
Column
Description
Start mass
End mass
Dwell time
Spacing
Resolution
Measurement
mode
Thermo Scientific
6-27
Templates
Method Parameters
1. Click
to open Qtegra.
4. Click
below Method Parameters to open the Survey scan
settings view of the Template.
1. Click
to open Qtegra.
4. Click
below Method Parameters to open the Survey scan
settings view of the Template.
5. Drag the scan region borders in the graphic above the table to define
the Start mass and End mass of the scan region.
The new values are immediately displayed in the table below.
6. Enter the desired values for Dwell time and Spacing.
7. Select Normal or High from the drop-down menu Resolution.
8. Select a Measurement mode from the drop-down list.
9. Click
1. Click
to open Qtegra.
6-28
Thermo Scientific
Templates
Method Parameters
4. Click
below Method Parameters to open the Survey scan
settings view of the Template.
5. Right-click in the graphic right next to the scan region.
A context menu opens, see Figure 6-25.
1. Click
to open Qtegra.
Thermo Scientific
6-29
Templates
Method Parameters
4. Click
below Method Parameters to open the Survey scan
settings view of the Template.
5. In the Advanced Parameters section at the bottom, enter the
Number of sweeps to be performed during one survey scan.
6. Click
Interference Correction
For all Template types, the Interference Correction view in Qtegra
allows you to enable mathematical interference correction for the
analytes in the Template.
NOTICE Interference Correction must be enabled prior to running the
measurement to be used to correct data.
Elements selected for analysis in the Analytes view are listed in the
Interference Correction view, see Figure 6-26.
6-30
Thermo Scientific
Templates
Method Parameters
If the isotopes used in the interference correction equation are not part
of the selected analytes, Qtegra uses the settings for the Identifier
belonging to this equation in Interference Correction also for the
isotope not listed.
1. Click
to open Qtegra.
4. Click
below Method Parameters to open the Interference
correction view of the Template.
1. Click
to open Qtegra.
4. Click
below Method Parameters to open the Interference
correction view of the Template.
5. Select the Enabled check box next to the analyte to activate
interference correction.
By default, all analytes are disabled.
6. Right-click in the Correction cell to open the context menu and
select Default interference correction from the context menu.
The default formula defined for this isotope is added.
You can also double-click the cell Correction and enter a formula.
7. Define the interference correction for all analytes.
8. Click
Thermo Scientific
6-31
Templates
Method Parameters
Standards
For all Template types, the Standards view in Qtegra allows you to
define standards. This section defines all information about the
solutions used to calibrate the instrument. For aQuant and eQuant,
additionally calibration types can be defined in the Quantification view.
Once a standard is created, the elements of the standard can be selected
in the periodic table, see Figure 6-27.
Column
Description
No
Element/Isotope/Compound
6-32
Thermo Scientific
Templates
Method Parameters
Table 6-6.
Column
Description
Concentration
Unit
Displays the concentration unit for the element in the standard file. By
default, the unit is set to ppb. This default can be changed and stored.
See Setting the Default Concentration on page 6-43.
The commands of the Standards view of a Template are summarized in
Table 6-7.
Table 6-7.
Commands
Thermo Scientific
6-33
Templates
Method Parameters
The quantification standard defined here can be selected and used in the
Sample Definition for a Template on page 6-135. When defining the
calibration standards in that section, dilution factors can be applied to
the standard.
Internal standards which are used for quantification are also created
here. Their definition as internal standards is done in the
Quantification view, see Quantification on page 6-62.
1. Click
to open Qtegra.
4. Click
below Method Parameters to open the Standards view
of the Template.
Creating a New Standard
Standards created in the Standards view of a Template (or LabBook) in
Qtegra are created for the current Template (or LabBook) but can be
saved to the global database.
For eQuant and trQuant Templates, it is possible to create internal
standards.
For rQuant Templates, only isotope dilution standards can be created.
For tQuant Templates, only compound standards can be created.
NOTICE Global database standards are created in the Configurator
applet Standard Editor on page 3-35.
1. Click
to open Qtegra.
6-34
Thermo Scientific
Templates
Method Parameters
4. Click
below Method Parameters to open the Standards view
of the Template.
5. Click
6. Click
Elemental Standard to open the Add New Standard
dialog, see Figure 6-29.
Thermo Scientific
6-35
Templates
Method Parameters
1. Click
to open Qtegra.
4. Click
below Method Parameters to open the Standards view
of the Template.
6-36
Thermo Scientific
Templates
Method Parameters
5. Click
6. Click
Internal Standard (Isotopical) to open the Add New
Standard dialog, see Figure 6-32.
Thermo Scientific
6-37
Templates
Method Parameters
10. Add the isotope to the table by clicking on the element in the
periodic table.
The most abundant isotope is added to the table, see Figure 6-33.
1. Click
to open Qtegra.
6-38
Thermo Scientific
Templates
Method Parameters
4. Click
below Method Parameters to open the Standards view
of the Template.
5. Click
6. Click
Isotope Dilution Standard to open the Add New
Standard dialog, see Figure 6-35.
Figure 6-35. Add New Standard dialog for isotope dilution standards
7. Enter the Standard Name.
8. Enter a Standard Description.
9. Select the check box Create standard using analyte list to create
the standard from the list of analytes as defined in the Analytes view.
10. Click
Thermo Scientific
6-39
Templates
Method Parameters
11. Add isotopes to the table by clicking on the element in the periodic
table.
The most abundant isotope of each selected element is added to the
table as Isotope 1, see Figure 6-36.
1. Click
to open Qtegra.
6-40
Thermo Scientific
Templates
Method Parameters
4. Click
below Method Parameters to open the Standards view
of the Template.
5. Click
6. Click
Compound Standard to open the Add New Standard
dialog, see Figure 6-38.
10. Click
Thermo Scientific
6-41
Templates
Method Parameters
14. Click
in the toolbar of your Template to save the standard to
your Template.
Loading a Standard from the Global Database
It is possible to load global standards created in the Configurator tool to
your Template in Qtegra.
1. Click
to open Qtegra.
4. Click
below Method Parameters to open the Standards view
of the Template.
6-42
Thermo Scientific
Templates
Method Parameters
or
1. Click
to open Qtegra.
4. Click
below Method Parameters to open the Standards view
of the Template.
Thermo Scientific
6-43
Templates
Method Parameters
1. Click
to open Qtegra.
4. Click
below Method Parameters to open the Standards view
of the Template.
5. In the Standards list on the left, click the standard to be edited.
6. If required, change the default concentration
6-44
Thermo Scientific
Templates
Method Parameters
Deleting a Standard
Standards can be deleted from a Template in Qtegra.
To delete a standard
1. Click
to open Qtegra.
4. Click
below Method Parameters to open the Standards view
of the Template.
5. In the Standards list on the left, click the standard to be deleted.
6. Click
1. Click
to open Qtegra.
Thermo Scientific
6-45
Templates
Method Parameters
4. Click
below Method Parameters to open the Standards view
of the Template.
5. In the Standards list on the left, right-click the standard to be saved
to the global database to open the context menu, see Figure 6-42.
6-46
Thermo Scientific
Templates
Method Parameters
The columns that define the properties of the Compounds are listed in
Table 6-8.
Table 6-8.
Column
Description
Compound Name
Trace
Analyte (isotope) trace used for the compound defined in the row.
Isotope from which the species will be quantified. The drop-down list
includes all isotopes selected in the Analytes view.
Auto Detect
Blank
Normalize Trace
Retention [s]
Tolerance [s]
Internal Standard
Fit
Weighting
Forcing
Thermo Scientific
6-47
Templates
Method Parameters
Above the Compounds table, buttons are provided to add a row, delete
a row and to activate the Internal Standard column.
1. Click
to open Qtegra.
4. Click
below Method Parameters to open the Compounds
view of the Template.
To define compounds
1. Click
to open Qtegra.
4. Click
below Method Parameters to open the Compounds
view of the Template.
5. Click
Thermo Scientific
Templates
Method Parameters
11. Click
1. Click
to open Qtegra.
4. Click
below Method Parameters to open the Compounds
view of the Template.
5. Click
above the Compounds table.
Internal Standardization is activated, see Figure 6-44.
1. Click
to open Qtegra.
Thermo Scientific
6-49
Templates
Method Parameters
4. Click
below Method Parameters to open the Compounds
view of the Template.
5. Click the gray field in front of the row or rows you wish to delete.
6. Click
7. Click
6-50
Thermo Scientific
Templates
Method Parameters
Smoothing, Peak Detection and Peak Filter parameters (see Figure 6-45)
can be defined.
1. Click
to open Qtegra.
Thermo Scientific
6-51
Templates
Method Parameters
4. Click
below Method Parameters to open the Peak Detection
view of the Template.
1. Click
to open Qtegra.
4. Click
below Method Parameters to open the Peak Detection
view of the Template.
5. For Smoothing, select the check box Active to activate Smoothing.
The smoothing settings are applied to the traces.
6. Enter the Number of Points.
Determines the number of points to be used to calculate the average
value.
7. Enter the Number of Passes.
Number of times the smoothing algorithm is run.
8. Select the Smoothing Method from the drop-down menu.
Mainly used is the smoothing method Moving Mean which uses the
rolling average of the given number of points, see Figure 6-46.
6-52
Thermo Scientific
Templates
Method Parameters
1. Click
to open Qtegra.
4. Click
below Method Parameters to open the Peak Detection
view of the Template.
5. For Peak Filter, enter the Minimum Peak Height [cps].
Only peaks that meet this condition are automatically integrated.
6. For Peak Filter, enter the Minimum Peak Area [cts].
Only peaks that meet this condition are automatically integrated.
7. Click
1. Click
to open Qtegra.
4. Click
5. For Peak Detection, select ICIS from the drop-down list Selected
Integrator.
Thermo Scientific
6-53
Templates
Method Parameters
6-54
Thermo Scientific
Templates
Method Parameters
1. Click
to open Qtegra.
4. Click
5. For Peak Detection, select PPD from the drop-down list Selected
Integrator, see Figure 6-48.
Thermo Scientific
6-55
Templates
Method Parameters
1. Click
to open Qtegra.
4. Click
6-56
Thermo Scientific
Templates
Method Parameters
1. Click
to open Qtegra.
4. Click
Thermo Scientific
6-57
Templates
Method Parameters
Thermo Scientific
Templates
Method Parameters
11. Select the check box Calculate Noise as RMS if you wish to
calculate the noise according to the root mean square method.
12. Enter the value for Base Noise Limit.
This is the parameter that controls how the baseline is drawn in the
noise data. The higher the baseline noise limit value, the higher the
baseline is drawn through the noise data. The valid range is 0.0 to
100.0.
13. Enter the value for Min Scans in Baseline.
This parameter is used to calculate a baseline. A larger number
includes more data in determining an averaged baseline. The valid
range is 2 to 100.0.
14. Enter the value for Baseline Noise Rejection Factor.
This factor controls the width of the RMS noise band above and
below the peak detection baseline. This factor is applied to the raw
RMS noise values to raise the effective RMS noise used by the
software during peak detection. Qtegra responds by assigning the
left and right peak boundaries above the noise and therefore closer
to the peak apex value in seconds. This action effectively raises the
peak integration baseline above the RMS noise level. The valid range
of this factor is 0.1 to 10.0.
15. Enter the value for Percent Largest Peak.
This limits the number of peaks submitted for further processing.
Qtegra discards any detected peaks with an intensity less than the
threshold percentage of the most intense peak.
16. Enter the value for SN Threshold.
The default value is 0.5 and the valid range is 0.0 to 999.0. Qtegra
calculates the signal-to-noise ratio using only the baseline signal.
Any extraneous, minor, detected peaks are excluded from the
calculation.
17. Enter the value for Base Signal to Noise Ratio.
18. Enter the value for Valley Threshold.
19. Enter the value for Valley Depth.
20. Enter the value for Peak Signal to Noise Ratio Cut Off.
Qtegra defines this signal-to-noise level as the top of the peak edge.
For example, if the signal-to-noise at the apex is 500 and the Peak
S/N Cutoff value is 200, Qtegra will define the right and left edges
of the peak when the S/N reaches a value less than 200. The valid
range is 50.0 to 10000.0.
21. Enter the value for Background Update Rate.
Qtegra periodically recalculates the representative background signal
it uses for background subtraction. This is to compensate for the
possibility that the height of the background might change over the
Thermo Scientific
6-59
Templates
Method Parameters
1. Click
to open Qtegra.
4. Click
below Method Parameters to open the Regions view of
the Template.
To add a region
1. Click
6-60
to open Qtegra.
Thermo Scientific
Templates
Method Parameters
4. Click
below Method Parameters to open the Regions view of
the Template.
5. Click
To delete a region
1. Click
to open Qtegra.
4. Click
below Method Parameters to open the Regions view of
the Template.
5. Click the gray field in front of the row or rows you wish to delete.
6. Click
in the toolbar of the Template and confirm the message
dialog to delete the rows.
7. Click
Thermo Scientific
6-61
Templates
Method Parameters
Quantification
The Quantification view of a eQuant, aQuant or trQuant Template in
the Qtegra tool allows you to set the calibration and quantification
strategy for each analyte.
All analytes selected in the Analytes view are shown in the
Quantification view, see Figure 6-52.
6-62
Thermo Scientific
Templates
Method Parameters
The parameters that can be defined for the Quantification table are
summarized in Table 6-9.
Table 6-9.
Column
Description
Analyte
Measurement Mode
Quantify
Internal Standard
Once Internal Standards are defined they are added to the drop-down
list of the cells in the Internal Standard column. The operator may
define any internal standard isotope desired.
If Use as Internal standard is selected, this row is shown with a green
background.
Use Interpolation enables a linear regression between two
enshrouding internal standards that corrects the observed intensity of
the analyte. If only one internal standard was chosen, Qtegra
automatically selects the internal standard with the mass closest to the
analyte.
Fit Type
Weighting
Thermo Scientific
6-63
Templates
Method Parameters
Table 6-9.
Column
Description
Forcing
1. Click
to open Qtegra.
4. Click
below Method Parameters to open the Quantification
view of the Template.
1. Click
to open Qtegra.
6-64
Thermo Scientific
Templates
Method Parameters
4. Click
below Method Parameters to open the Quantification
view of the Template.
5. Select the check box Quality Control at the above the table, see
Figure 6-53.
1. Click
to open Qtegra.
4. Click
below Method Parameters to open the Quantification
view of the Template.
5. Click
in the cell of the column Internal Standard to open the
drop-down list.
Thermo Scientific
6-65
Templates
Method Parameters
6-66
Thermo Scientific
Templates
Method Parameters
14. Click
to save your Template.
The new parameters are saved to the Template.
Ratios
For all Template types, the Ratios view of a Template in the Qtegra tool
allows you to define isotopic, elemental or compound ratios for the
measurement.
For aQuant, eQuant and trQuant Templates, the Ratios view shows a
table for the ratio of Analyte 1 and Analytes 2, see Figure 6-55.
Thermo Scientific
6-67
Templates
Method Parameters
For rQuant Templates, the table in the Ratios view shows rows for each
element for which you selected more than one isotope for this Template,
see Figure 6-57.
Description
No
Ratio
Analyte 1
or
Compound 1
or
Isotope 1
Analyte 2
or
Compound 2
or
Isotope 2
1. Click
to open Qtegra.
6-68
Thermo Scientific
Templates
Method Parameters
4. Click
below Method Parameters to open the Ratios view of
the Template.
1. Click
to open Qtegra.
4. Click
below Method Parameters to open the Ratios view of
the Template.
5. Click
in a cell of the column Analyte 1 to display the list of
available isotopes.
6. Select an isotope for column Analyte 1.
The isotope selected in column Analyte 1 is the numerator in
column Ratio.
7. Click
in a cell of the column Analyte 2 to display the list of
available isotopes.
8. Select an isotope for column Analyte 2.
The isotope selected in column Analyte 2 is the denominator in
column Ratio.
The ratio of both isotopes is displayed in the column Ratio.
9. Click
1. Click
to open Qtegra.
Thermo Scientific
6-69
Templates
Method Parameters
4. Click
below Method Parameters to open the Ratios view of
the Template.
5. Click
in a cell of the column Isotope 1 to display the list of
available isotopes.
6. Select an isotope for column Isotope 1.
The isotope selected in column Isotope 1 is the numerator in
column Ratio.
7. Click
in a cell of the column Isotope 2 to display the list of
available isotopes.
8. Select an isotope for column Isotope 2.
The isotope selected in column Isotope 2 is the denominator in
column Ratio.
The ratio of both isotopes is displayed in the column Ratio.
9. Click
1. Click
to open Qtegra.
4. Click
below Method Parameters to open the Ratios view of
the Template.
5. Click
in a cell in the column Compound 1 to display the list of
available compounds.
6. Select a compound for column Compound 1.
The compound selected in column Compound 1 is the numerator
in column Ratio.
6-70
Thermo Scientific
Templates
Method Parameters
7. Click
in a cell in the column Compound 2 to display the list of
available compounds.
8. Select a compound for column Compound 2.
The compound selected in column Compound 2 is the
denominator in column Ratio.
The ratio of both compounds is displayed in the column Ratio.
9. Click
To delete rows
1. Click
to open Qtegra.
4. Click
below Method Parameters to open the Ratios view of
the Template.
5. Right-click the cell at the beginning of a row.
A context menu opens, see Figure 6-58.
Thermo Scientific
6-71
Templates
Method Parameters
6-72
Thermo Scientific
Templates
Method Parameters
1. Click
to open Qtegra.
5. Click
below Method Parameters to open the Quality
Control view of the Template.
1. Click
to open Qtegra.
4. Click
below Method Parameters to open the Quantification
view of the Template.
5. Select the check box Use Quality Control above the table
Quantification.
The Method Parameters
immediately.
1. Click
to open Qtegra.
Thermo Scientific
6-73
Templates
Method Parameters
4. Click
below Method Parameters to open the Quality
Control view of the Template.
On the left, the available Quality Control Tests are listed.
5. Click the Quality Control Test you wish to define.
On the right, the corresponding Test details and Test Parameters
are shown.
6. In the Test Parameters table, click the first cell of a column for
which you wish to change the entry and enter the new value.
7. Click anywhere in the table to enter the value.
8. Click and drag the mouse from this first entry you wish to copy over
all cells of the column to be changed with this value.
9. Right-click in the cell of the table to where you wish to copy the
value.
A context menu opens, see Figure 6-60.
1. Click
to open Qtegra.
6-74
Thermo Scientific
Templates
Method Parameters
4. Click
below Method Parameters to open the Quality
Control view of the Template.
On the left, the available Quality Control Tests are listed.
5. Click the Quality Control Test you wish to duplicate and define.
6. Click
at the top of Quality Control Tests list.
A dialog box opens, see Figure 6-61.
1. Click
to open Qtegra.
Thermo Scientific
6-75
Templates
Method Parameters
4. Click
below Method Parameters to open the Quality
Control view of the Template.
On the left, the available Quality Control Tests are listed.
5. Click the Quality Control Test you wish to delete, for example, see
Figure 6-62.
6. Click
.
The selected quality control test is immediately deleted from the list.
7. In the toolbar of your Template, click
6-76
Thermo Scientific
Templates
Method Parameters
Description
CCB
ICB
Frequency
Limits
every 10 samples
< 3 x IDL
(Instrumental
Detection Limit)
after initial
calibration
< 3 x IDL
Thermo Scientific
Purpose
6-77
Templates
Method Parameters
Description
Purpose
Frequency
Limits
MTB
user definable
user definable
PRB
Preparation Blank
(LRB in EPA Method
200.8)
< 3 x IDL
1. Click
to open Qtegra.
4. Click
below Method Parameters to open the Quality
Control view of the Template.
6-78
Thermo Scientific
Templates
Method Parameters
On the left, the available Quality Control Tests are listed, see
Figure 6-64.
Thermo Scientific
6-79
Templates
Method Parameters
11. Deselect the check box next to the analyte in the Enabled column to
skip this analyte, see Figure 6-65.
6-80
Thermo Scientific
Templates
Method Parameters
Description
CCV
ICV
Initial Calibration
Verification
Thermo Scientific
Purpose
Frequency
Limits
every 10 samples
90-110 %
90-110 %
6-81
Templates
Method Parameters
Description
Purpose
Frequency
Limits
LCS
Laboratory Control
Sample
every 20 samples
80-120 %
(EPA Method
6020A)
30-70 %
ILM05.2D
QCS
Quality Control
Standard
10%
(EPA Method
200.8)
1. Click
to open Qtegra.
6-82
Thermo Scientific
Templates
Method Parameters
4. Click
below Method Parameters to open the Quality
Control view of the Template.
On the left, the available Quality Control Tests are listed, see
Figure 6-67.
Thermo Scientific
6-83
Templates
Method Parameters
11. Deselect the check box next to the analyte in the Enabled column to
skip this analyte, see Figure 6-68.
6-84
Thermo Scientific
Templates
Method Parameters
Thermo Scientific
6-85
Templates
Method Parameters
The Paired Sample Tests available are summarized in Table 6-13. The
last two columns show typical QC requirements of the US EPA.
Table 6-13. Quality control paired sample tests
Test
type
Description
Purpose
Frequency
Limits
DUP
Duplicate
SER
Serial Dilution
1. Click
to open Qtegra.
6-86
Thermo Scientific
Templates
Method Parameters
4. Click
below Method Parameters to open the Quality
Control view of the Template.
On the left, the available Quality Control Tests are listed, see
Figure 6-70.
Thermo Scientific
6-87
Templates
Method Parameters
11. Deselect the check box next to the analyte in the Enabled column to
skip this analyte, see Figure 6-71.
6-88
Thermo Scientific
Templates
Method Parameters
Description
Purpose
Frequency
Limits
LFB
Laboratory Fortified
Blank
every 20 to 30
samples
85-115%
(EPA Method
200.8)
MXS
Matrix Spike
80-120 %
(EPA Method
6020A)
30-70 %
ILM05.2D
PDS
Thermo Scientific
6-89
Templates
Method Parameters
1. Click
to open Qtegra.
4. Click
below Method Parameters to open the Quality
Control view of the Template.
On the left, the available Quality Control Tests are listed, see
Figure 6-73.
6-90
Thermo Scientific
Templates
Method Parameters
6. Click
to change the Number of analyte failures to generate a
QC failure or enter the new value directly. For details, see Quality
Control Failure Rules on page 6-99.
7. Click
to change the Number of analyte warnings to generate
a QC failure or enter the new value directly.
8. Select the action to take place If this QC fails.
9. Select the action to take place If this QC fails again.
10. Select the action to take place If this QC fails a final time.
11. Deselect the check box next to the analyte in the Enabled column to
skip this analyte, see Figure 6-74.
Thermo Scientific
6-91
Templates
Method Parameters
Continuous Tests
In Qtegra, the Quality Control Method Parameter of the eQuant
Template offers several types of Continuous Tests, see Figure 6-75.
6-92
Thermo Scientific
Templates
Method Parameters
Two different Continuous Tests are available. The last two columns in
Table 6-15 show typical QC requirements of the US EPA.
Table 6-15. Quality control spike recovery
Test
type
Description
Purpose
Frequency
Limits
RCV
Regression Coefficient
Verification
0.95
Failure 0.9
RSV
Relative Stability
Deviation Verification
5%
Failure 10%
1. Click
to open Qtegra.
Thermo Scientific
6-93
Templates
Method Parameters
4. Click
below Method Parameters to open the Quality
Control view of the Template.
On the left, the available Quality Control Tests are listed, see
Figure 6-76.
6-94
Thermo Scientific
Templates
Method Parameters
11. Deselect the check box next to the analyte in the Enabled column to
skip this analyte, see Figure 6-77.
Figure 6-79. Quality Control Continuous Test Parameters RSV for Concentration
Thermo Scientific
6-95
Templates
Method Parameters
14. Set the value for Ignore Concentration Below and define a Unit, if
activated.
Set the Concentration Warning Limit and Concentration Failure
Limit
15. Set the value for Ignore Intensities Below, if activated.
Set the Intensity Warning Limit and Intensity Failure Limit, see
Figure 6-80.
Figure 6-80. Quality Control Continuous Test Parameters RSV for Intensity
16. In the toolbar of your Template, click
6-96
Thermo Scientific
Templates
Method Parameters
1. Click
to open Qtegra.
4. Click
below Method Parameters to open the Quality
Control view of the Template.
On the left, the available Quality Control Tests are listed, see
Figure 6-82.
Thermo Scientific
6-97
Templates
Method Parameters
6-98
Thermo Scientific
Templates
Method Parameters
Figure 6-84. Quality control details and parameters for Spike Test PDS
In the first part of Test details, the number of failures can be defined,
see Table 6-16.
Table 6-16. Settings for number of failures
Parameter
Description
To define how many analytes must fail before the flag message is
generated.
Recommended setting: 1
Thermo Scientific
The QC fails
6-99
Templates
Method Parameters
Description
This action repeats the test after a rinse step has been performed. A
failed QC will automatically trigger an identical copy of the QC
sample to be inserted into the sample list after the failed QC. This step
will be repeated once.
Recalibrate, recalculate and reacquire Upon QC failure this action will automatically insert a copy of the
from a named sample
calibration block and the QC sample immediately after the failed QC
sample. A QC pass from the repeated tests will then allow the sample
list to be resumed from a named sample which is defined in the
Sample Definition section.
Autotune, recalibrate, recalculate and Upon QC failure this action will automatically run the autotune
reacquire from a previous sample
program followed by a copy of the calibration block and the QC
sample. A QC pass from the repeated tests will then allow the sample
list to be resumed from a named sample which is defined in the
Sample Definition section.
Pause the experiment and wait for
restart
Upon QC failure, the LabBook will be aborted and the Scheduler will
continue with other scheduled LabBooks if there are any.
If a QC test fails, the first action is normally to rinse and repeat the test.
If the test fails again, it might be advisable to recalibrate and repeat or to
ignore and continue.
Each incident of this test will have exactly the same condition.
For example, once defined, whenever an CCB is defined in the sample
list, the same conditions will be used every time. The parameters can be
set separately for each of the tests, for example, CCB can use one set of
tests, whereas ICB uses a tighter set.
6-100
Thermo Scientific
Templates
Method Parameters
Thermo Scientific
Column
Description
Symbol
Concentration
Unit
6-101
Templates
Method Parameters
Description
Import
Export
NOTICE Any analytes (cells) that are not required for the LOD checks
can be left blank.
1. Click
to open Qtegra.
4. Click
below Method Parameters to open the Quality
Control view of the Template.
6-102
Thermo Scientific
Templates
Method Parameters
5. Click
.
A dialog opens, see Figure 6-86.
10. Click
to save your Template.
The detection limits are saved to the Template.
1. Click
to open Qtegra.
Thermo Scientific
6-103
Templates
Method Parameters
4. Click
below Method Parameters to open the Quality
Control view of the Template.
5. Click
.
A dialog opens, see Figure 6-87.
6-104
Thermo Scientific
Templates
Method Parameters
8. Select the *.csv file you wish to import, see Figure 6-88.
1. Click
to open Qtegra.
4. Click
below Method Parameters to open the Quality
Control view of the Template.
Thermo Scientific
6-105
Templates
Method Parameters
5. Click
.
A dialog opens, see Figure 6-89.
6-106
Thermo Scientific
Templates
Method Parameters
10. Click
Limits dialog.
1. Click
to open Qtegra.
Thermo Scientific
6-107
Templates
Method Parameters
6. Enter a Label for each row and define the columns to your needs.
NOTICE For details on the columns, see Sample Definition for a
Template on page 6-135.
7. Enter a Label and select QC for the column Sample Type for your
QC rows.
8. For your Initial Actions QC row, select a QC test type from the
drop-down list for the column QC Action.
9. Select a Standard where required by the QC test, see Figure 6-91.
6-108
Thermo Scientific
Templates
Method Parameters
Thermo Scientific
6-109
Templates
Method Parameters
6-110
Thermo Scientific
Templates
Peripherals
Peripherals
The settings for peripherals such as LC pumps or LC autosamplers can
be adjusted in the corresponding view of the Template in Qtegra.
NOTICE Peripherals are added to the Configuration in the Configurator
tool of Qtegra (see Experiment Configurator on page 3-13).
1. Click
to open Qtegra.
Thermo Scientific
6-111
Templates
Peripherals
4. Click
Figure 6-94.
6-112
Thermo Scientific
Templates
Peripherals
1. Click
to open Qtegra.
Thermo Scientific
6-113
Templates
Peripherals
Chromeleon System
Depending on the Chromeleon system configured within the applicable
hardware configuration, different icons will be displayed on the left of
the Chromeleon method editor view, for example, see Figure 6-96.
6-114
Thermo Scientific
Templates
Peripherals
Pump
Autosampler
Triggering Command
Integrated
System
ICS-900
AS-DV
Pump_InjectValve State
Load --> Inject
Integrated
System
ICS-2100
AS-DV
Pump_InjectValve State
Load --> Inject
ICS-5000
DP
AS-AP
Ultimate
3000
ICS-5000
DP, DC
with AS-DV
AS-DV
DC_InjectValve_1_State
Load --> Inject
1. Click
to open Qtegra.
Thermo Scientific
6-115
Templates
Peripherals
4. Click
Figure 6-97.
6-116
Thermo Scientific
Templates
Peripherals
5. Click
to open the Wizard to create a method for the
chromatographic separation, see Figure 6-98.
Thermo Scientific
6-117
Templates
Peripherals
7. Continue until all settings have been set and click Finish.
The Chromeleon view displays the Chromeleon Method as a text
file that can be further edited manually if required, see Figure 6-99.
6-118
Thermo Scientific
Templates
Peripherals
8. Click
to change the parameter used for the start of data
acquisition.
For some combinations of devices this can be required. A dialog
opens displaying the parameters of the chromatographic system that
can be changed, see Figure 6-100.
Thermo Scientific
6-119
Templates
Peripherals
1. Click
to open Qtegra.
5. Click
Figure 6-101.
6-120
Thermo Scientific
Templates
Peripherals
Thermo Scientific
6-121
Templates
Peripherals
6-122
Thermo Scientific
Templates
Peripherals
20. Click
1. Click
to open Qtegra.
Thermo Scientific
6-123
Templates
Peripherals
to show the
SpectraSystem LC Autosampler
The SpectraSystem LC autosampler can be operated in combination
with the SpectraSystem LC pump.
6-124
Thermo Scientific
Templates
Peripherals
1. Click
to open Qtegra.
Thermo Scientific
6-125
Templates
Peripherals
SpectraSystem LC Pump
The SpectraSystem LC pump can be operated in combination with the
SpectraSystem LC autosampler.
1. Click
to open Qtegra.
6-126
Thermo Scientific
Templates
Peripherals
4. Click
Thermo Scientific
6-127
Templates
Peripherals
Accela LC Autosampler
The Accela LC autosampler (600, 1000, 1250) can be operated in
combination with the Accela LC pump.
NOTICE Settings for Syringe size and Tray Type must be selected in the
Configurator tool by your Administrator (see Editing the Settings of
Instruments on page 3-17).
1. Click
to open Qtegra.
6-128
Thermo Scientific
Templates
Peripherals
4. Click
6-129
Templates
Peripherals
15. In the Time Event Program table, select the check box in the
desired column (TF1 to TF4) and enter the desired value in the
column Time [min].
As soon as you select a check box in the last row, another row is
added to the Time Event Program table.
16. Select the Injection Mode and enter the Loop Loading Speed
[l/sec] (L/s).
17. Select the check box On for Tray Heater/Cooler Control and enter
the Temperature (C), if desired.
18. Select the check box On for Column Oven Control and enter the
Temperature (C), if desired.
19. Enter descriptions for Bottle/Reservoir Content.
20. Click
Accela LC Pump
The Accela LC pump can be operated in combination with the Accela
LC autosampler.
NOTICE Pump model must be selected and Serial number must be
entered in the Configurator tool by your Administrator (see Editing
the Settings of Instruments on page 3-17).
1. Click
to open Qtegra.
6-130
Thermo Scientific
Templates
Peripherals
4. Click
Thermo Scientific
6-131
Templates
Peripherals
6-132
Thermo Scientific
Templates
Manual Sample Control
1. Click
to open Qtegra.
Thermo Scientific
6-133
Templates
Manual Sample Control
4. Click
Control view of the Template.
6-134
Thermo Scientific
Templates
Sample Definition for a Template
Thermo Scientific
6-135
Templates
Sample Definition for a Template
Description
Interval
Absolute
Keep together
Label
Duration
Injection Volume
Survey Runs
6-136
Thermo Scientific
Templates
Sample Definition for a Template
Description
Main Runs
Comment
Evaluate
Sample Type
For all except rQuant Templates. Definition of the sample type. See
also Data Evaluation on page 10-1.
Internal Standard
Standard
Dilution Factor
Sample Amount
Spike Amount
Amount
Final Quantity
QC Action
QC Restart
QC Reference
Special Blank
Rack Number/Tray/Block
Vial Number/Vial
Thermo Scientific
6-137
Templates
Sample Definition for a Template
1. Click
to open Qtegra.
6-138
Thermo Scientific
Templates
Sample Definition for a Template
1. Click
to open Qtegra.
Thermo Scientific
6-139
Templates
Sample Definition for a Template
1. Click
to open Qtegra.
6-140
Thermo Scientific
Templates
Sample Definition for a Template
8. Click
1. Click
to open Qtegra.
Thermo Scientific
6-141
Templates
Sample Definition for a Template
8. Enter a value for each column or select an item from the drop-down
list, as appropriate.
NOTICE For details on the columns, see Sample Definition for a
Template on page 6-135.
9. Click
6-142
Thermo Scientific
Templates
Automatic Export - Template
1. Click
to open Qtegra.
Thermo Scientific
to open
6-143
Templates
Automatic Export - Template
5. Select the check box CSV Export to define these export settings, see
Figure 6-119.
6-144
Thermo Scientific
Templates
Automatic Export - Template
13. Click
in the toolbar of Available data to save the setting to a
scheme.
The Save Export Scheme dialog opens, see Figure 6-120.
Thermo Scientific
6-145
Templates
Automatic Export - Template
19. Select the check box Export per sample if you wish to export one
file per sample.
20. Click
to save the setting to a scheme.
The Save Export Scheme dialog opens, see Figure 6-122.
1. Click
to open Qtegra.
6-146
Thermo Scientific
Templates
Automatic Export - Template
Thermo Scientific
6-147
Templates
Automatic Export - Template
6-148
Thermo Scientific
Chapter 7
LabBooks
LabBooks are based on the settings specified in the Templates on
page 6-1 in Qtegra. These setting can still be adjusted in the LabBook
before the measurements is run.
Contents
LabBook Toolbar
Summary of LabBook
Scheduling a LabBook
Log Messages
Signing
Query
Reports
Reports Editor
Creating Reports
To open a LabBook in the Qtegra tool
1. Click
to open Qtegra.
Thermo Scientific
7-1
LabBooks
LabBook Toolbar
LabBook Toolbar
In the LabBook tab of Qtegra, Qtegra offers buttons to save, close, run
or export a LabBook, see Figure 7-1.
Figure 7-1.
LabBook toolbar
To save a LabBook
1. Click
to open Qtegra.
To close a LabBook
1. Click
to open Qtegra.
Thermo Scientific
LabBooks
LabBook Toolbar
4. Click
To run a LabBook
1. Click
to open Qtegra.
1. Click
to open Qtegra.
Thermo Scientific
7-3
LabBooks
LabBook Toolbar
4. Click Create.
The Create drop-down menu opens, see Figure 7-2.
Figure 7-2.
5. Click New Template if you wish to create a new Template from the
current LabBook.
The Template view of the Home Page opens. See Creating a
Template on page 5-30 for further details.
6. If you wish to create a new LabBook from the current LabBook,
click New LabBook.
The LabBook view of the Home Page opens. See Creating a
LabBook on page 5-21 for further details.
1. Click
to open Qtegra.
7-4
Thermo Scientific
LabBooks
LabBook Toolbar
Figure 7-3.
Thermo Scientific
7-5
LabBooks
LabBook Toolbar
9. Click
to save the settings as your scheme.
The Save Export Scheme dialog opens, see Figure 7-4.
Figure 7-4.
12. Click
1. Click
to open Qtegra.
7-6
Thermo Scientific
LabBooks
LabBook Toolbar
Figure 7-5.
5. Click
1. Click
to open Qtegra.
Thermo Scientific
7-7
LabBooks
LabBook Toolbar
4. Click
.
The History dialog for this LabBook opens, see Figure 7-6.
Figure 7-6.
7-8
Thermo Scientific
LabBooks
LabBook Toolbar
6. Click
to compare the selected entries.
The Comparison dialog opens, see Figure 7-7.
Figure 7-7.
7. Select the check box Show differences only if you wish to view only
the differences.
8. Click
9. Click
1. Click
to open Qtegra.
Thermo Scientific
7-9
LabBooks
LabBook Toolbar
4. Click
.
The History dialog for this LabBook opens, see Figure 7-8.
Figure 7-8.
Figure 7-9.
7-10
Thermo Scientific
LabBooks
LabBook Toolbar
8. Click
1. Click
to open Qtegra.
Thermo Scientific
7-11
LabBooks
Color Scheme of the Periodic Table
7-12
Thermo Scientific
LabBooks
Method Parameters LabBook
Thermo Scientific
7-13
LabBooks
Summary of LabBook
Summary of LabBook
A summary page is added to each LabBook in Qtegra. This page shows
the file name, and information about Properties, Date and People for
the LabBook, see Figure 7-13.
1. Click
to open Qtegra.
7-14
Thermo Scientific
LabBooks
Sample List - LabBook
1. Click
to open Qtegra.
Thermo Scientific
7-15
LabBooks
Sample List - LabBook
1. Click
to open Qtegra.
1. Click
to open Qtegra.
5. Click the gray field in front of the row you wish to delete.
6. Click
1. Click
to open Qtegra.
5. Click
to show the comment for the selected row.
The list of comments opens below the sample list.
7-16
Thermo Scientific
LabBooks
Sample List - LabBook
1. Click
to open Qtegra.
5. Click
to show the comment for the selected row.
The list of comments opens below the sample list.
6. Click
to add a comment for the selected row.
The User Comment window opens, see Figure 7-15.
1. Click
to open Qtegra.
Thermo Scientific
7-17
LabBooks
Sample List - LabBook
5. Click
to hide the comment for the selected row.
The list of comments is hidden.
7-18
Thermo Scientific
LabBooks
Automatic Export - LabBook
Thermo Scientific
7-19
LabBooks
Scheduling a LabBook
Scheduling a LabBook
To schedule a measurement, you open a LabBook in Qtegra and run it.
Evaluation results can be accessed in a running LabBook so results can
be viewed in real time, see Evaluation Results on page 7-23.
In the Tools section on the Help page of Qtegra, you define your
Scheduler settings, see Customizing Scheduler Settings on page 5-59.
NOTICE To customize the Options of the Scheduler, you can also click
the
1. Click
to open Qtegra.
7-20
Thermo Scientific
LabBooks
Scheduling a LabBook
Thermo Scientific
7-21
LabBooks
Viewing the Result of a Measurement
1. Click
2. Click
to open Qtegra.
.
7-22
Thermo Scientific
LabBooks
Viewing the Result of a Measurement
4. Click
to show the items added to the LabBook after
measurement, see Figure 7-19.
Evaluation Results
The Evaluation Results view displays the data acquired within a
LabBook and can be viewed during the actual acquisition of a LabBook
in Qtegra.
The presentation of the evaluation results differs, according to the
Method Parameters on page 6-14 defined. In that way, Compounds
will only be shown for tQuant LabBooks whereas Survey Intensities are
found for eQuant LabBooks.
1. Click
2. Click
to open Qtegra.
.
Thermo Scientific
7-23
LabBooks
Viewing the Result of a Measurement
4. Click
to open the Evaluation Results view,
for example, see Figure 7-20.
Instrument State
The Instrument State view of a LabBook in Qtegra shows instrument
status values of the iCAP Q system for each sample in the sample list.
1. Click
2. Click
7-24
to open Qtegra.
.
Thermo Scientific
LabBooks
Viewing the Result of a Measurement
Thermo Scientific
7-25
LabBooks
Log Messages
Log Messages
The Log Messages view of a LabBook in Qtegra is added to the
LabBook after a measurement has been run for this LabBook.
The table in Log Messages contains all events with appropriate time
stamps which occur throughout the manipulation of the LabBook. All
information, warning and error messages are logged here including
information about the service concerned.
1. Click
to open Qtegra.
7-26
Thermo Scientific
LabBooks
Log Messages
1. Click
to open Qtegra.
6. Click
in the header of the column you wish to filter the display
of.
A drop-down menu opens, see, for example, Figure 7-24.
1. Click
to open Qtegra.
Thermo Scientific
7-27
LabBooks
Log Messages
6. Click
in the header of the column you wish to filter the display
of.
A drop-down menu opens, see, for example, Figure 7-25.
7-28
Thermo Scientific
LabBooks
Log Messages
11. Click
of the right column to open the drop-down menu, see
Figure 7-28.
Thermo Scientific
7-29
LabBooks
Signing
Signing
The Signing view of a LabBook in Qtegra is added to the LabBook after
a measurement has been run for this LabBook.
Signing is used to protect the acquired data and verify the operator.
Certificates are required to activate the Signing feature. These Digital
SSL certificates are issued by Trusted CA Certificate Authorities and
must be purchased separately. They are applied by your Administrator.
1. Click
2. Click
to open Qtegra.
.
7-30
Thermo Scientific
LabBooks
Signing
1. Click
2. Click
to open Qtegra.
.
Thermo Scientific
7-31
LabBooks
Query
Query
The Query view of a LabBook in Qtegra is added to the LabBook after
a measurement has been run for this LabBook.
For Query, the toolbar of the completed LabBook offers the additional
buttons Query and Reports, and the Create drop-down item New
Report is activated, see Figure 7-31.
1. Click
7-32
to open Qtegra.
Thermo Scientific
LabBooks
Query
2. Click
1. Click
2. Click
to open Qtegra.
.
Thermo Scientific
7-33
LabBooks
Query
6. In the section Rows, select the check boxes for Categories you wish
to display.
Click
1. Click
7-34
to open Qtegra.
Thermo Scientific
LabBooks
Query
2. Click
6. Select the check boxes for Rows and Columns you wish to display.
7. In the toolbar of your LabBook, click
result values.
Thermo Scientific
7-35
LabBooks
Query
1. Click
2. Click
to open Qtegra.
.
6. Select the check boxes for Rows and Columns you wish to display.
7. In the toolbar of your LabBook, click
result values.
7-36
Thermo Scientific
LabBooks
Query
Thermo Scientific
7-37
LabBooks
Query
1. Click
2. Click
to open Qtegra.
.
6. In the Results section on the right, select the preset you wish to
delete from the Presets list.
7. Click
to delete the result data preset.
The Delete Preset dialog is displayed, see Figure 7-36.
8. Click
.
The preset is deleted from the list.
Reports Editor
In the Reports view of a completed LabBook in Qtegra, you generate
result data reports from presets.
You can structure your report to your needs, add headings, tables and
graphs to specify the presentation of the result data. For details on
creating report presets, see Creating Reports on page 7-41.
7-38
Thermo Scientific
LabBooks
Query
1. Click
to open Qtegra.
2. Click
To generate a report
1. Click
to open Qtegra.
2. Click
to open the
Thermo Scientific
7-39
LabBooks
Query
or save
as desired.
1. Click
2. Click
to open Qtegra.
.
7-40
to open the
Thermo Scientific
LabBooks
Query
, see
Creating Reports
You create report presets or layouts easiest from the Reports view of a
completed LabBook in Qtegra.
Generally, you can structure your report to your needs, add headings,
tables and graphs to specify the presentation of the result data.
Thermo Scientific
7-41
LabBooks
Query
7-42
Thermo Scientific
LabBooks
Query
Report modules
Report module
Function
Data
Graph
Group by
Thermo Scientific
7-43
LabBooks
Query
Table 7-1.
Report modules
Report module
Function
Text
To create a report
1. Click
2. Click
to open Qtegra.
.
Thermo Scientific
LabBooks
Query
to open the
in the reports
Thermo Scientific
7-45
LabBooks
Query
7. Click
to save the file.
The Save Report dialog opens, see Figure 7-43.
7-46
Thermo Scientific
LabBooks
Query
Thermo Scientific
7-47
LabBooks
Query
, see
7-48
Thermo Scientific
LabBooks
Query
Thermo Scientific
7-49
LabBooks
Query
, see
7-50
Thermo Scientific
LabBooks
Query
1. Click
2. Click
to open Qtegra.
.
Thermo Scientific
7-51
LabBooks
Query
7-52
Thermo Scientific
LabBooks
Query
7. Click
Thermo Scientific
7-53
LabBooks
Query
10. For the section Common on the right, enter a Name for the module
and optionally a Description.
11. Select the check box Use in Report to show this module in the
report.
NOTICE The checkbox Use in Report is provided for all modules to
show or not show each module in the report.
12. For Details, enter the text, in this example, Analytes, and define the
Font family and Font size.
Select the check box for Bold or Italic if you wish to format the text
accordingly.
Define Text color, Background color and Alignment.
For Bookmark, you can define the level, see Figure 7-54.
15. Click
7-54
Thermo Scientific
LabBooks
Query
20. Click
Thermo Scientific
7-55
LabBooks
Query
25. From the entries under Columns, select the column you wish to add
to your table.
26. On the right next to Result columns, click
item to the table.
7-56
Thermo Scientific
LabBooks
Query
29. Click
1. Click
2. Click
to open Qtegra.
.
Thermo Scientific
to open the
7-57
LabBooks
Query
6. On the left, select the report layout you wish to edit and click
.
The report editor opens for the selected report layout.
7. Under Report content, select
7-58
Thermo Scientific
LabBooks
Query
next to
Thermo Scientific
7-59
LabBooks
Query
7-60
Thermo Scientific
LabBooks
Query
19. Click the tab Layout and select Use Table Layout, see Figure 7-61.
20. Click
Thermo Scientific
7-61
LabBooks
Query
7-62
Thermo Scientific
LabBooks
Query
26. From the Available list on the left, select the rows you wish to show
in the table, see Figure 7-63.
Thermo Scientific
7-63
LabBooks
Query
29. Click
Figure 7-64. Report showing grouped list with column layout table above
30. Click
To define conditions
1. Click
2. Click
to open Qtegra.
.
to open the
6. On the left, select the report layout you wish to edit and click
.
The report editor opens for the selected report layout.
7-64
Thermo Scientific
LabBooks
Query
7. Select the module you wish to define conditions for, for example,
Group.
Conditions can be defined for the modules Group, Data and Result
Data Table.
8. For Columns, select, for example, Sample Type, and click
next
to Conditions in the lower right corner to add the item to the
Conditions table.
9. From the drop-down lists in the table Conditions, select, for
example, equal to for Operator and UNKNOWN for Condition,
see Figure 7-65.
Thermo Scientific
7-65
LabBooks
Query
11. Click
Figure 7-66. Report with list of analytes and lists of concentration per sample
Only concentration tables for the sample type UNKNOWN are
shown.
7-66
Thermo Scientific
LabBooks
Reports
Reports
The Reports view of a completed LabBook in Qtegra displays the
results of a measurement in report formats previously created.
For creation of reports, see Creating Reports on page 7-41.
To view a report
1. Click
2. Click
to open Qtegra.
.
Thermo Scientific
7-67
LabBooks
Reports
7-68
Thermo Scientific
Chapter 8
Thermo Scientific
8-1
1. Click
to open Qtegra.
4. Click
to select the Analytes view.
See Analytes on page 6-15 for a general explanation.
5. In the periodic table, select the analytes of your calibration solution
and your samples.
First, the calibration curve of known samples must be acquired for
later comparison of the intensities of analytes with this calibration
curve.
6. Click
to select the Acquisition Parameters view.
See Acquisition Parameters on page 6-18 for a general explanation.
7. Enter the Dwell time (s) for the elements of your calibration
solution and your analytes.
Typically, dwell times are related to the expected concentration of
the analyte in the sample. Higher concentrations usually require
shorter dwell times. Lower concentrations should be measured for a
longer time to improve the signal-to-noise ratio. Short dwell times
are often used for isotope dilution analysis.
8-2
Thermo Scientific
NOTICE The more sweeps are averaged, the better the measured value
should be. Drift effects might not be recognized with longer dwell times
and a small number of sweeps. Very short dwell times might impair the
duty cycle of the instrument. A good dwell time value to start from
usually is 0.01 s.
8. Enter the value for Channels and Spacing (u).
Usually with a good and stable mass calibration, the default values
should be sufficient to measure on the apex of the mass peaks.
9. Select the Measurement mode for each analyte from the drop-down
list, see Figure 8-1.
Figure 8-1.
Thermo Scientific
8-3
14. Click
18. Click
to select the Interference correction view.
Interference correction helps to minimize not-polyatomic isobaric
interferences if no other interference-free isotope is available. This
mathematical correction is suited for analytical measurements
following EPA regulations.
For details, see Interference Correction on page 6-30.
19. Click
8-4
Thermo Scientific
22. Click
25. Click
26. Select the Isotope 1 and Isotope 2 from the drop-down lists.
The Ratios page provides the option to set several user-defined ratios
which are displayed after the measurement of the LabBook.
For details, see Ratios on page 6-67.
NOTICE For details on all parameters, see Method Parameters on
page 6-14.
27. Click
1. Click
to open Qtegra.
Thermo Scientific
8-5
1. Click
to open Qtegra.
10. Click
8-6
Thermo Scientific
1. Click
to open Qtegra.
Figure 8-2.
Thermo Scientific
8-7
8. Click
to create the new LabBook.
A new tab opens for the new LabBook.
9. Check all settings.
10. Check the sample list.
11. Make sure that the settings for the autosampler are corresponding to
the actual position of vials in the autosampler.
12. In the toolbar of your LabBook, click
8-8
Thermo Scientific
1. Click
to open Qtegra.
4. Below , click
.
The Browse for LabBook window opens.
5. Select your eQuant LabBook.
6. Click
to open the LabBook.
The LabBook opens in a new tab of the Qtegra tool.
7. In the toolbar of your LabBook, click
to schedule the LabBook
for execution.
The LabBook is added to the Scheduler. If the check box Automatic
has been selected for Start Queue in the Options settings of the
Scheduler (see Customizing Scheduler Settings on page 5-59), the
measurement is started immediately.
Thermo Scientific
8-9
Concentrations
In the Evaluation Results Concentrations view of the LabBook in
Qtegra, the results of the quantitative analysis are summarized. As with
the sample list, blanks are displayed in blue, standards in yellow, QCs in
red and unknowns in white. The mean values, standard deviations (SD)
and relative standard deviations (RSD) as well as the results of each main
8-10
Thermo Scientific
Figure 8-3.
Thermo Scientific
8-11
the most common statistical data to assess the quality of the fit. The
green area in the graph represents the confidence delta at 90% while
each point is displayed with its standard deviation, see Figure 8-4.
Figure 8-4.
8-12
Thermo Scientific
By clicking
the blank correction is switched on/off. If no ZERO
STD was selected, the blank correction is done by including the
measured intensity of the different isotopes into the calibration plot
with a concentration of 0 (BLK). If one or more samples in the sample
list are indicated as ZERO STD (to perform a standard addition), the
correction is done by subtraction of the intensities determined for this
sample.
Concentration Ratios
The Evaluation Results Concentration Ratios view of the LabBook in
Qtegra shows the ratios for each pair of isotopes entered in the Method
Parameters section related to the determined concentrations, see
Figure 8-5.
Figure 8-5.
Intensities
The Evaluation Results Intensities view of the LabBook in Qtegra
displays the raw intensities. If the entries are shown in bold type, at least
one sweep within one main run was measured using the analog mode of
the detector. If the entries are displayed in blue instead of black, they
Thermo Scientific
8-13
were manually edited, for example, the result of one main run was
removed from the calculation of the average after the measurement, see
Figure 8-6.
Figure 8-6.
8-14
Thermo Scientific
Clicking
displays the mean values as well as the SD and RSD values.
In the thumbnails, filled circles indicate that the value was measured in
the analog mode, red circles represent excluded entries. The blue line in
the enlarged graph represents the mean value of the different main runs,
see Figure 8-7.
Figure 8-7.
Thermo Scientific
8-15
If more than one channel was measured for any isotope, there is also the
possibility to set the strategy how to handle the raw intensities, see
Figure 8-8.
Figure 8-8.
8-16
Strategy
Description
Average
Center
Integral
Highest
Thermo Scientific
Intensity Ratios
The Evaluation Results Intensity Ratios view of the LabBook in Qtegra
shows the data with reference to the raw intensities. Again, the context
menu offers functions to include or exclude single entries, see
Figure 8-9.
Figure 8-9.
Thermo Scientific
8-17
Survey Intensities
When a survey scan was acquired, the Evaluation Results Survey
Intensities view of the LabBook in Qtegra shows the measured
intensities of all isotopes within the defined survey scan regions, see
Figure 8-10.
8-18
Thermo Scientific
Survey Concentrations
The Evaluation Results Survey Concentrations view of the LabBook in
Qtegra is shown in Figure 8-11.
Thermo Scientific
8-19
Spectra View
The Evaluation Results Spectra View view of the LabBook in Qtegra
displays the acquired mass spectra (plots of the measured intensities
against the mass-to-charge ratio). Options to save, copy or print the
graph are offered in the context menu, see Figure 8-12.
8-20
Thermo Scientific
With the check boxes in the Details section it is possible to simulate the
natural isotopic abundances of the elements as well as of common
interferences, see Figure 8-13.
Thermo Scientific
8-21
8-22
Thermo Scientific
Chapter 9
Thermo Scientific
9-1
1. Click
to open Qtegra.
4. Click
to select the Analytes view.
See Analytes on page 6-15 for a general explanation.
5. In the periodic table, select the elements of interest for the species
present in your calibration and sample solutions.
6. Click
to select the Acquisition Parameters view.
See Acquisition Parameters on page 6-18 for a general explanation.
7. Enter the Dwell time (s) for the elements present in your
calibration and sample solutions.
The dwell time should be selected to be long enough to sufficiently
improve the signal-to-noise ratio. Dwell times that are too long
reduce the possibility to acquire enough points/values for the correct
calculation and interpolation of the peak. Seven to nine
points/values usually suffice to describe the peak shape correctly. A
good dwell time value to start from usually is 0.1 or 0.2 s. If you
wish to analyze species containing many different elements in one
measurement, the dwell times must be adjusted accordingly.
9-2
Thermo Scientific
NOTICE Dwell times for very short peaks as, for example, with Ultra
high pressure LC systems, are shorter than with customary LC systems
but equal or slightly longer as with GC or CE systems.
8. Enter the value for Channels and Spacing (u).
Default values are usually acceptable.
9. Select the Measurement mode from the drop-down list, see
Figure 9-1.
Figure 9-1.
Thermo Scientific
9-3
Figure 9-2.
12. Click
to select the Standards view.
First, the calibration curve of species with known concentrations
must be acquired for later comparison of the peak areas of unknown
samples with this calibration curve.
13. Click New to define a Standard as described in Creating a New
Standard on page 6-34.
See Standards on page 6-32 for details.
14. Define the Compounds you wish to use in your external calibration.
You can create a compound standard from the compound list if you
define the compounds first, see Compounds (tQuant only) on
page 6-46.
15. Click
16. Click
to add a line to the table and define your compound.
Your definitions for the column Compound Name will be used in
the column Compound in Standards if you create a compound
standard from the compound list. The names for Compound Name
and Compound must be identical.
For details on defining compounds, see Compounds (tQuant
only) on page 6-46.
NOTICE All settings except instrument scan dependent parameters can
still be changed after measurement.
17. Click
18. For Smoothing, select the check box Active and select Moving
Mean from the drop-down list Smoothing Method to improve the
signal-to-noise ratio.
For details, see Peak Detection (tQuant only) on page 6-50.
9-4
Thermo Scientific
19. Click
21. Click
to select the Interference correction view.
This Method Parameter provides the option to minimize
non-polyatomic isobaric interferences if no other interference-free
isotope is available.
For details, see Interference Correction on page 6-30.
NOTICE For details on all parameters, see Method Parameters on
page 6-14.
22. Click
1. Click
to open Qtegra.
Thermo Scientific
9-5
6. Click
1. Click
to open Qtegra.
8. Click
9-6
Thermo Scientific
1. Click
to open Qtegra.
Figure 9-3.
Thermo Scientific
9-7
8. Click
to create the new LabBook.
A new tab opens for the new LabBook.
9. Check all settings.
If external trigger signals are used make sure they are configured
correctly. The settings are inherited from the Template.
10. Check the sample list.
Pay special attention to Duration and Sample Type settings.
11. Make sure that the settings for the hyphenated technique are
corresponding to the actual settings, for example, the position of
vials in the LC autosampler.
12. In the toolbar of your LabBook page, click
LabBook.
9-8
to save your
Thermo Scientific
1. Click
to open Qtegra.
4. Below , click
.
The Browse for LabBook window opens.
5. Select your tQuant LabBook.
6. Click
to open the LabBook.
The LabBook opens in a new tab of the Qtegra tool.
7. In the toolbar of your LabBook, click
to schedule the LabBook
for execution.
The LabBook is added to the Scheduler. If the check box Automatic
has been selected for Start Queue in the Options settings of the
Scheduler (see Customizing Scheduler Settings on page 5-59), the
measurement is started immediately.
Thermo Scientific
9-9
9-10
Thermo Scientific
Compounds
In the Evaluation Results Compounds view of the LabBook in Qtegra,
the acquired time slices (chromatogram) are shown with the determined
peak area for the defined compounds after automatic peak detection and
integration, see Figure 9-4.
Figure 9-4.
Result chromatogram
Values determined automatically by the software are displayed in black,
values which have been changed or assigned manually are shown in blue.
Thermo Scientific
9-11
Peaks can be selected by clicking into the table cell containing the
respective peak area of a defined compound. The displayed time range
in the chromatogram automatically refocuses to the time range of the
selected peak, see Figure 9-5.
Figure 9-5.
9-12
Thermo Scientific
Clicking
at the front of a table row opens underlying information
about the signal of interest, such as Peak Start Time, Peak End Time,
Apex Retention Time, Apex Baseline Height and Apex Height above
Baseline, see Figure 9-6.
Figure 9-6.
Thermo Scientific
9-13
You can select several samples to show the several chromatograms at the
same time, either in separate panels or in one panel if you deselect
Isolated View, see Figure 9-7.
Figure 9-7.
9-14
Thermo Scientific
Peak
In the Evaluation Results Peaks view of the LabBook in Qtegra,
assignment of the peaks found in the chromatogram to the compounds
to be quantified can be revised. Chromatographic peaks which were
recognized but could not be associated to a compound are also displayed
here, see Figure 9-8.
Figure 9-8.
Thermo Scientific
9-15
In case that a peak for a defined compound has been detected, but was
not assigned correctly, this can be done manually. By right-clicking a cell
in the compound column a context menu opens, see Figure 9-9.
Figure 9-9.
9-16
Thermo Scientific
Ratios
In the Evaluation Results Ratios view of the LabBook in Qtegra, ratios
of the peak area between different compounds previously defined in the
Method Parameters section are calculated and displayed, see
Figure 9-10.
Thermo Scientific
9-17
Concentration
In the Evaluation Results Concentration view of the LabBook in
Qtegra, the acquired fully quantitative calibration can be revised, see
Figure 9-11.
9-18
Thermo Scientific
Figure 9-12. Result Concentration view with context menu to Exclude entry
The Details option is activated either by double-clicking the calibration
graph or via the toolbar. The calibration graph for the selected
compound is displayed in a larger size with related information such as
Thermo Scientific
9-19
9-20
Thermo Scientific
Chapter 10
Data Evaluation
Data Evaluations are calculation methods that are applied within the
Qtegra software. These calculation or evaluation methods are selected
when creating a Template in Qtegra (see Evaluation Methods on
page 6-9).
The Qtegra data evaluation is handled by several evaluation modules
(evaluation plug-ins) called virtual evaluation (VE). The system
currently knows the following common evaluation modules for
quantification:
Contents
External Calibration
Standard Addition
Isotope Dilution
Thermo Scientific
10-1
Data Evaluation
Integration - Raw Data Handling
Average:
channel
Intensity i = i k channel
k=1
Integral:
channel
Intensity i = i k
k=i
10-2
Thermo Scientific
Data Evaluation
Integration - Raw Data Handling
Raw data are based on the number of Main Runs shown in the
corresponding line of the Samplelist in Figure 10-2 and are calculated
for every analyte for which data has been collected.
Thermo Scientific
10-3
Data Evaluation
External Calibration
External Calibration
The external calibration strategy is employed with the Evaluation
methods eQuant (steady state signals), tQuant (transient signals), and
trQuant (transient regions).
The External Calibration module is the most complex quantification
module. It currently offers seven sample types to specify measurement
blocks, see Table 10-1.
Table 10-1. Supported sample types
Name
Description
UNKNOWN
STD
BLK
AVERAGE BLK
ZERO STD
UPDATE CALIB
QC
Sample types can be defined for each sample line, see Sample
Definition for a Template on page 6-135.
10-4
Thermo Scientific
Data Evaluation
External Calibration
Thermo Scientific
10-5
Data Evaluation
External Calibration
10-6
Thermo Scientific
Data Evaluation
External Calibration
i rISC
= -------i tISC
where i tISC denotes the averaged intensity of the internal standard of all
main runs in the first line of the preceding block corresponding to a
certain chosen analyte, and i rISC denotes the averaged intensity of the
internal standard in each of the following sample lines.
NOTICE By definition, only one internal standard or internal standard
mixture can be used for the whole experiment. Overlapping of internal
standards and standards used for full-quantification is not allowed.
Blanks
Consecutive blank sample lines are handled as so-called blank blocks.
Each blank sample line generates one blank intensity value for each
measured isotope. This intensity is defined as the mean value of all runs
of the current sample. Depending on the sample type, each blank block
is used in a different manner:
i BlankjIsotope
j=1
-.
i Blank Isotope = ------------------------------------------------------# blank block lines
Thermo Scientific
BLK uses the blank intensity last measured as the blank block
intensity.
10-7
Data Evaluation
External Calibration
Standards
Standards consisting of elements with known concentrations present the
basis of any comparing quantification method. Consecutive standard
sample lines are handled as a standard block. Each standard sample line
generates one intensity value for each measured isotope. This value is
combined with the known element concentration from the standard and
forms a data point of the calibration curve.
Updating the Calibration Curve
With the sample type UPDATE CALIB, it is possible to recalculate your
concentrations without running the standard measurement again if you
realize, for example, drifts in the results. This sample type is used to
calculate a correction value which will be applied to the preceding
calibration block. Element concentration and dilution factor of the
update calibration sample line must be identical to the one of the
standard block.
The intensity of a measured isotope with the sample type UPDATE
CALIB
UPDATECALIB
i Isotope
10-8
Thermo Scientific
Data Evaluation
External Calibration
is used for scaling any intensity value of the calibration curve applied to
calculate the concentrations of the samples following the measurement
of the UPDATE CALIB sample, where s r and s t are the reference and
averaged target sensitivity in the preceding standard block of each
standard isotope, respectively. The sensitivities are calculated using the
equations:
UPDATECALIB
i Isotope
s r = --------------------------------STD
c Analyte
STD
i Isotope
s t = --------------STD
c Analyte
The sample type ZERO STD function can be combined with all other
available sample types in the External Calibration module. For details,
see Standard Addition on page 10-13.
NOTICE UPDATE CALIB samples can only compensate drifts affecting
the sensitivity of the instrument. To compensate for increased or
decreased blank values it is recommended to insert BLK or AVERAGE
BLK sample types before the UPDATE CALIB sample and to use the
QA/QC functionality of Qtegra. If a ZERO STD is used, all blank
concentration is calculated by subtraction intensity values.
Calculating the Calibration Curve
Based on the calibration curve, the original concentration c of the
analyte in an UNKNOWN sample is calculated with
1
Thermo Scientific
10-9
Data Evaluation
External Calibration
The available options for calculating the calibration curve are listed in
Table 10-2.
Table 10-2. Options for calibration curve calculation
Option
Description
Fit Type
Linear
2nd order
f ( x ) = a2 x + a1 x + a0
will be used.
Forcing
No
Zero
Blank
None
Absolute SD
Weight k = 1 k2 .
Each point is weighted by the standard
deviation k of the analyte over the
runs in the sample.
Relative SD
Weight k = 1 ( k i k ) .
Each point is weighted by the standard
deviation k of the analyte over the
runs in the sample relative to the mean
value i k .
10-10
Thermo Scientific
Data Evaluation
External Calibration
Semi-Quant
The semi-quant feature is used for isotopes that cannot be quantified by
using a calibration curve based on standards. For this type of analysis it
is assumed that each isotope on a standard has a defined
instrument-specific response. By default, any isotope calibration curve
of the previous standard block will be used as input for the semi-quant
methods.
A semi-quantitative response curve is produced for every standard block
in the experiment. A semi-quantitative curve is a 2nd order line fit of
sensitivity against mass. Each calibration curve in a standard block that
has been selected for use in the response curve is included in the line fit.
The sensitivity of each isotope is taken from the slope of its calibration
curve and is corrected for relative sensitivity using the defined analyte
RSF (relative sensitivity factor). Only linear curves may be used for
semi-quantitative analysis and at least three calibration curves are
needed to solve a 2nd order line fit problem. The semi-quant sensitivity
ssq is given by
s sq = a 1isotope ( RSF analyte A isotope )
where Aisotope defines the isotopic abundance (semi-quant curves are
always defined in terms of isotopic sensitivity) and a 1isotope defines the
slope given by the calibration curve.
The slope results lead directly to the quantification of a non-standard
isotope with the measure intensity inon-standard isotope
c sq = i non-standard isotope a i .
If final quantity q, amount a, and dilution d are specified in the sample
list, the concentration value is corrected by c corr = c d q a .
Unspecified values are set to be 1.
NOTICE In case of semi-quantitative analysis units are chosen
automatically for best representation.
Thermo Scientific
10-11
Data Evaluation
External Calibration
Isotope Quantification
To fully quantify an isotope the preceding methods are applied. A full
isotope quantification consists of the following steps:
10-12
Calculation of blanks.
Thermo Scientific
Data Evaluation
Standard Addition
Standard Addition
The Standard Addition module (aQuant) uses the sample type ZERO
STD (defining the zero spike) and STD (for standards) to set up a
measurement. For details on sample types see Sample Definition for a
Template on page 6-135.
A valid sample list for standard addition consists of blocks of one ZERO
STD measured at the beginning followed by a sequence of standards
(STD). For each block, a first-order calibration curve is produced for
every analyte. The curve is constructed from the mean results of given
runs using the given concentrations.
The first-order calibration curve is calculated using a linear least square
fit. The available options are listed in Table 10-3:
Table 10-3. Options for calibration curve calculation
Option
Description
Forcing
No
Zero
None
Absolute SD
Weight k = 1 k2 .
Weighting
Weight k = 1 ( k i k ) .
Each point is weighted by the standard
deviation k of the analyte over the
runs in the sample relative to the mean
value i k .
c = f (x)
where x is the measured intensity.
Thermo Scientific
10-13
Data Evaluation
Standard Addition
With
f ( x ) = a1 x + a0
the measured concentration of the zero standard is calculated as
i ZEROSTD = a 0
c ZEROSTD = i ZEROSTD a 1 .
If final quantity q, amount a, and Dilution d are specified in the sample
list, the zero concentration value is corrected by
c ZEROSTDcorr = c ZEROSTD d q a .
Unspecified values are set to be 1. The dilution factor is in a comparable
way also handled for standards.
NOTICE The units used to display the calculation results for an analyte
depend on the first appearance of that given analyte in a standard. Unit
selections are set for the complete experiment.
NOTICE Using weighting for the calibration curve is only available if all
standard measurements in the current block of the sample list consist of
at least two runs. In all other cases the option is disabled. Any leading
standard samples before measuring the first zero standard will be
ignored.
10-14
Thermo Scientific
Data Evaluation
Isotope Dilution
Isotope Dilution
Due to the methodical structure of Isotope Dilution, this module has
more complexity than the Standard Addition module. To quantify an
element in a Isotope Dilution experiment, the standards for spiking
must first be defined. The known concentration of an element as well as
the known information about the specifying isotopes, their abundance
values as well as atomic weight have to be entered.
A global Isotope Dilution Standard can be created in the applet
Standard Editor of the Configurator tool, see Figure 10-6.
Thermo Scientific
10-15
Data Evaluation
Isotope Dilution
Isotope1
Isotope2
Sample
measured
)a
Sample
where
cSpike = element concentration in spike solution
MSpike = amount of added spike
MSample = sample amount
Isotope k
a Spike
Isotope k
10-16
Thermo Scientific
Data Evaluation
Isotope Dilution
and
ATW ElementSample
R ATWElement = ------------------------------------ATW ElementSpike
with ATW as the atomic weight of the specified element.
If final quantity q, amount a, and dilution d are specified in the sample
list, the element concentration value is corrected by
c Elementcorr = c Element d q a .
Unspecified values are set to be 1.
NOTICE This measurement method quantifies the elements defined in
the Evaluation Parameters tab control. Because of the structure of the
Evaluation Parameters list, each element can be specified only once,
even if there is theoretically the possibility to have different spiking
standards including the same element.
Thermo Scientific
10-17
Data Evaluation
Isotope Dilution
10-18
Thermo Scientific
Glossary
This section lists and defines terms used in this manual. It also includes acronyms, metric prefixes, symbols, and
abbreviations.
A
C
C degrees Celsius.
CE European conformity. Mandatory European
marking for certain product groups to indicate
conformity with essential health and safety
requirements set out in European Directives.
cool gas serves to prevent the glass torch from melting.
Thermo Scientific
G-1
Glossary: DSPP/N
f femto (10-15).
F degrees Fahrenheit.
lb pound.
h hour.
h height.
M+ molecular ion.
O
OD outside diameter.
G-2
P
PCB printed circuit board.
PCL abbr. for Process Control Language.
P/N part number.
Thermo Scientific
Glossary: ppbWEEE
T
TCP/IP transmission control protocol / Internet
protocol.
U
u symbol for atomic mass unit.
R
V
regression types are used in the creation of calibration
curves during a sequence of analyses: the software
offers four types: linear, thru zero, weighted, and
square fit.
RF radio frequency.
s second.
square fit regression type: the fit is performed with a
second order (quadratic) function.
Thermo Scientific
G-3
Glossary: WEEEWEEE
G-4
Thermo Scientific
Index
Symbols
*.csv 3-24
*.imhwd 3-24
*.imrep 3-313-32
*.lic 5-70
*.licreq 5-68
*.panel 3-26
A
Accela LC autosampler 6-128
Accela LC pump 6-130
access control editor
access rights 3-7
layout 3-4
open 3-6
summary 2-3
user levels 3-6
acquisition
restart 4-14
start 4-14
stop 4-14
acquisition parameters 6-18
export analytes table 6-23
number of sweeps 6-22
template 6-18
add
comments of sample list 7-17
labBook to scheduler 5-76
measurement mode in edit mode 4-23
monitored analyte 6-25
region 6-60
report heading 7-51
row to sample list 7-16
scan region 6-28
add report table 7-51
adjust
Accela LC autosampler settings 6-128
Accela LC pump settings 6-130
Cetac ASX 260 autosampler settings 6-113
Cetac ASX 520 autosampler settings 6-111
chromeleon settings 6-115
ESI SC-4S autosampler settings 6-120
SpectraSystem LC autosampler settings 6-125
SpectraSystem LC pump settings 6-126
Thermo Scientific
analyte
open view 4-6
select for interference correction 6-31
analytes
method parameter 6-15
template 6-15
applets 2-2
aQuant
evaluation method 6-10
quantification 6-62
assistance 1-i
audit trail labBook 7-9
audit trail template 6-6
automatic export
csv export settings 6-143
labBook 7-19
report export settings 6-146
template 6-143
autosampler
Accela 6-128
Cetac ASX-260 6-112
Cetac ASX-520 6-111
chromeleon 6-114
ESI SC-4Q 6-119
SpectraSystem 6-124
autotune
create sequence 4-66
edit sequence 4-57
run sequence 4-74
view report 4-129
views group report 4-126
wizard 4-56
average intensities 4-5
B
blank
define QC test 6-78
external calibration 10-7
verification 6-77
C
calibration
define QC test 6-82
external, verification 6-81
I-1
Index: CC
CCB 6-77
CCT 4-15, 4-21
CCTS 4-15, 4-21
CCV 6-81
Cetac ASX-260 autosampler 6-112
Cetac ASX-520 autosampler 6-111
change
color scheme of the periodic table 6-12
configuration 5-13
default concentration 3-37
layout of instrument control 4-136
order of labBook in scheduler 5-79
tune settings of all measurement modes 4-20
check
intensity of instrument 5-15
system status 5-15
chromeleon
adjust settings 6-115
autosampler 6-114
pump 6-114
system 6-114
close
edit mode 4-25
iCAP Q system 5-12
labBook 5-25, 7-2
template 5-34, 6-2
color scheme
block 6-12
electron affinity 6-12
electronegativity 6-12
individual 6-12
ionization potential 6-12
periodic table 6-11, 7-12
series 6-11
uncolored 6-11
compare
labBook history 7-7
template history 6-4
completed labBooks
generate report 7-39
move 5-3
open 5-81
compliance FDA 2-9
compound standard 6-40
compounds
activate internal standardization 6-49
define 6-48
delete 6-49
method parameter 6-46
open view 6-48
template 6-46
tQuant evaluation results view 9-11
I-2
concentration
detection limits 6-101
eQuant evaluation results view 8-10
standard elements 6-33
tQuant evaluation results view 9-18
concentration ratios 8-13
configuration
change 5-13
delete 3-22
load 4-11
new configuration 3-14
select in instrument control 4-11
configurator
applets 2-2
open 2-4, 3-1
open access control editor 3-6
open element editor 3-9
open settings 3-34
overview 2-2
user interface 3-2
continuous tests 6-92
define QC test 6-93
QC tests 6-92
control group 4-12
control panel
define order and display 4-138
open 4-138
overview 4-137
create
autotune sequence with wizard 4-66
compound standard in template 6-40
elemental standard in template 6-34
internal standard in template 6-36
isotope dilution standard in template 6-38
labBook 5-21
labBook for eQuant analysis 8-7
labBook in labBook view 7-3
labBook in template view 6-3
molecules 3-29
new performance report 4-37
new quality control test 6-74
preset configuration 3-22
subfolder 5-51
template 5-30
template in labBook view 7-3
template in template view 6-3
tQuant labBook 9-7
workspace folder 5-50
create new configuration 3-14
create new internal standard file 3-40
cross calibration
factors in views group 4-126
result 4-131
Thermo Scientific
Index: DD
customize
appearance of columns 6-139
home page settings 5-58
instrument control layout 4-134
log messages filter 7-27
scheduler settings 5-59
D
dashboard page 5-55-6
data display tab 4-9
data evaluation
eQuant 8-10
external calibration 10-4
integration raw data handling 10-2
isotope dilution 10-15
standard addition 10-13
tQuant 9-10
data view region
average intensities 4-5
instrument settings 4-5
debug 4-152
default
change concentration 3-37
Thermo Scientific
I-3
Index: EF
detector setup
perform 4-88, 4-97, 4-105
wizard 4-87
detector setup report
view 4-130
views group 4-126
display
instrument settings in data view region 4-5
list of interferences 6-20
result data 5-39
display group 4-25
display software information 4-136
DUP 6-86
duplicate rows 6-21
E
edit
autotune sequence with wizard 4-57
existing standard file 6-44
Instrument settings 3-18
labBook 5-23
measurement mode 4-21
measurement mode in edit mode 4-22
performance report 4-29
script 4-152
template 5-32
edit mode for measurement mode
close 4-25
open 4-22
element editor
add isotope 3-10
change default isotope 3-11
change element properties 3-10
change isotope properties 3-10
layout 3-9
open 3-9
summary 2-3
elemental standard 3-38, 6-34
eQuant
analysis 8-1
create labBook 8-7
evaluation 8-1
evaluation method 6-10
quality control 6-72
quantification 6-62
results view concentration 8-10
results view concentration ratios 8-13
results view intensities 8-13
results view intensities ratios 8-17
results view spectra view 8-20
results view survey concentrations 8-19
results view survey intensities 8-18
setting up template 8-2
template detection limits 6-101
I-4
F
failure rules 6-99
FDA compliance 2-9
file manager 5-53
copy 5-53
create subfolder 5-51
create workspace folder 5-50
cut 5-52
delete 5-54
open labBook 5-50
open template 5-49
open view 5-49
overview 5-49
page 5-49
rename 5-55
filter
customize for log messages 7-27
log messages 7-27
Thermo Scientific
Index: GI
generate
report in completed labBook 7-39
report in labBook query 5-46
grant access 3-7
group
control 4-12
display 4-25
measurement mode 4-15
report entries 7-57
views 4-126
wizard 4-27
iCAP Q ribbon
control group 4-12
display group 4-25
measurement mode group 4-15
open 4-12
views group 4-126
wizards group 4-27
iCAP Q ribbon tab 4-12
iCAP Q system
close down 5-12
prepare for measurement 5-10
prepare for measurement with manual sampling 5-6
switch off 4-14
ICB 6-77
ICV 6-81
import
detection limit files 6-103
detection limits 6-102
instrument control
experiment configuration 4-11
layout 4-134
load instrument controls 4-11
minimize ribbon 4-3
open 2-7, 4-1
overview 2-5
restart acquisition 4-14
start acquisition 4-14
stop acquisition 4-14
user interface 4-2
integration raw data handling 10-2
intensities
eQuant evaluation results view 8-13
intensities ratios
eQuant evaluation results view 8-17
interference correction
method parameter 6-30
template 6-30
internal standard
define QC test 6-97
QC test 6-96
H
hardware configurator
commands 3-24
layout 3-24
open 3-25
summary 2-3
hardware panel configurator
commands 3-26
layout 3-26
open 3-27
summary 2-3
help page
open 5-57
overview 5-57
registering 5-61
hide
comments of sample list 7-17
Content pane 6-7, 7-11
history
labBook 7-6
template 6-3
home page 5-2
customize settings 5-58
dashboard 5-5
Thermo Scientific
I-5
Index: KM
K
KED 4-15, 4-21
KEDS 4-15, 4-21
L
labBook
automatic export 7-19
close 5-25, 7-2
compare history 7-7
create 5-21
delete 5-24
export audit trail 7-9
labBooks page 5-18
open 5-19, 5-50, 7-1
qtegra 7-1
run 7-3, 7-20
save 7-2
sign 7-31
toolbar 7-2
view history 7-6
labBook query
delete report layout 7-40
query page 5-38
reports page 5-45
labBook query page 5-36
labBook view
create labBook 7-3
create template 7-3
labBooks page 5-18
layout
access control editor 3-4
element editor 3-9
experiment configurator 3-13
hardware configurator 3-24
hardware panel configurator 3-26
instrument control 4-134
molecule editor 3-28
report editor 3-31
script editor 3-33
settings 3-34
standard editor 3-35
I-6
M
mass calibration
perform 4-117
view result 4-133
views group 4-126
wizard 4-116
measurement mode
CCT 4-15, 4-21
CCTS 4-15, 4-21
change tune settings 4-20
edit 4-21
iCAP Q ribbon 4-15
KED 4-15, 4-21
KEDS 4-15, 4-21
load 4-16
STD 4-15, 4-21
STDS 4-15, 4-21
measurement mode in edit mode
add 4-23
delete 4-24
edit 4-22
Thermo Scientific
Index: NP
N
number of sweeps
acquisition parameters 6-22
survey scan settings 6-29
O
open
acquisition parameters view 6-20
analytes view 4-6, 6-16
completed labBooks 5-81
compounds view 6-48
control panel 4-138
dashboard of qtegra 5-6
data display tab 4-9
edit mode for measurement mode 4-22
ESI FAST autosampler settings 6-123
file manager page of qtegra 5-49
help page of qtegra 5-57
iCAP Q ribbon 4-12
instrument control 4-1
interference correction view 6-31
labBook 5-19
labBook from file manager page 5-50
labBooks page 5-19
log view qtegra 5-82
log viewer 3-3
monitor analytes view 6-24
peak detection view 6-51
peripheral data view 4-9
qtegra 5-1
quality control view 6-73
quantification view 6-64
query page 5-38
ratios view 6-68
recent labBook 5-20
recent template 5-29
regions view 6-60
reports page 5-45
sample definition view 6-138
scheduler 5-76
signing view 7-30
standard editor 3-36
standards view 6-34
status panel 4-149
survey scan settings view 6-28
template 5-28, 6-1
template from file manager page 5-49
templates page 5-28
open page 5-28
overview
configurator 2-2
instrument control 2-5
qtegra 2-8
P
paired sample
define QC test 6-86
verification 6-85
Thermo Scientific
I-7
Index: QQ
Q
QC
activate 6-73
blank verification 6-77
continuous test 6-92
copy values to grid 6-73
create new test 6-74
delete test 6-75
external calibration verification 6-81
failure rules 6-99
paired sample 6-85
spike recovery 6-89
I-8
QC test types
CCB 6-77
CCV 6-81
DUP 6-86
ICB 6-77
ICV 6-81
LCS 6-82
LFB 6-89
MTB 6-78
MXS 6-89
PDS 6-89
PRB 6-78
QCS 6-82
RCV 6-93
RSV 6-93
SER 6-86
qtegra
create template 5-30
dashboard page 5-5
help page 5-57
labBook 7-1
labBook query page 5-36
labBooks page 5-18
open 2-8, 5-1
open labBooks page 5-19
open templates page 5-28
overview 2-8
sample definition 6-135
templates page 5-27
user interface 5-2
quality control
activate 6-73
create new 6-74
open view 6-73
quantification
internal standardization 6-62
method parameter 6-62
open 6-64
standard 6-32
template 6-62
query 7-39
open view 7-32
overview 7-32
query page
labBook query 5-38
open 5-38
query preset
delete 7-38
save 7-34
Thermo Scientific
Index: RS
R
ratios 9-17
define analytes 6-69
define compounds 6-70
define isotopes 6-69
method parameter 6-67
open view 6-68
template 6-67
RCV 6-93
readback plot
open view 4-127
views group 4-126
real-time display 5-15
recent labBook 5-20
recent template 5-29
reference documentation 1-4
regions
add 6-60
delete 6-61
method parameter 6-60
open view 6-60
template 6-60
registering 5-61
remove script from script list 4-153
rename
labBook 5-55
template 5-55
report
add heading 7-51
add table 7-51
define conditions 7-64
delete layout 7-40
generate 5-46
group entries 7-57
labBook query 5-46
open 7-39
view 7-67
report editor
commands 3-31
layout 3-31
open 3-32
summary 2-3
report template
create 3-32
edit 3-32
reports page
labBook query 5-45
open 5-45
reports view 7-39
restart data acquisition in the real-time display 4-14
result data
define query display 7-33
display 5-39
generate report 5-46
save as preset 5-43
Thermo Scientific
S
sample definition
actions 6-140
customize 6-138
template 6-135
sample list 7-15
save
labBook 7-2
query preset 7-34
result data preset 5-435-44
standards to standard database 3-36
template 6-2
scan regions 6-28
scheduler
add labBook 5-76
change order of labBook 5-79
customize settings 5-59
move 5-3
open 5-76
pause execution of labBook 5-78
stop execution of labBook 5-77
scheduling a labBook 7-20
script
debug 4-152
edit 4-152
load 4-149
remove from script list 4-153
reset 4-152
run 4-151
stop 4-151
script editor
layout 3-33
summary 2-3
I-9
Index: TT
select
default isotope 6-17
isotope 6-17
semi-quant
external calibration 10-11
quantification 6-62
SER 6-86
set
default concentration 6-43
real-time display 4-27
settings
layout 3-34
summary 2-3
show
comments of sample list 7-16
summary of labBook 7-14
signing
labBook 7-31
open view 7-30
overview 7-30
source autotune sequence 4-82
spectra view
eQuant evaluation results view 8-20
SpectraSystem LC autosampler 6-124
SpectraSystem LC pump 6-126
spike
define QC test 6-90
QC test 6-89
spike recovery 6-89
standard
elemental standard 3-38
internal (elemental) 3-39
internal (isotopical) 3-39
isotope dilution 3-42
standard addition 10-13
standard editor
commands 3-35
create new internal standard (elemental) 3-41
create new internal standard (isotopical) 3-40
create new isotope dilution standard 3-43
create new standard 3-38
layout 3-35
summary 2-3
standard elements
concentration 6-33
unit 6-33
standards of template
create 6-34
delete 6-45
editing standard file 6-44
load global from database 6-42
method parameter 6-32
overview 6-32
set default concentration 6-43
I-10
start
data acquisition in the real-time display 4-14
plasma 4-13
script 4-151
status panel 4-149
STD 4-15, 4-21
STDS 4-15, 4-21
stop
data acquisition in the real-time display 4-14
script 4-151
stop execution of labBook in scheduler 5-77
survey concentrations
eQuant evaluation results view 8-19
survey intensities
eQuant evaluation results view 8-18
survey scan settings
define number of sweeps 6-29
method parameter 6-26
template 6-26
switch between different detector mode 4-27
switch off system 4-14
system status 5-15
T
template
acquisition parameters 6-18
analytes 6-15
automatic export 6-143
close 5-34, 6-2
compare history 6-4
compounds 6-46
create 5-30
create standards 6-34
delete 5-33
detection limits 6-101
edit 5-32
export analytes table 6-23
export audit trail 6-6
interference correction 6-30
monitor analytes 6-24
open 5-28, 5-49, 6-1
peak detection 6-50
quality control 6-72
quantification 6-62
ratios 6-67
regions 6-60
sample definition 6-135
save 6-2
select default isotope 6-17
select isotope 6-17
standards 6-32
survey scan settings 6-26
templates page 5-27
toolbar 6-2
view history 6-3
Thermo Scientific
Index: UW
template view
create labBook 6-3
create template 6-3
templates 5-28
templates page 5-27
toolbar
labBook 7-2
template 6-2
tQuant
analysis 9-1
compounds 6-46
create labBook 9-7
evaluation 9-1
evaluation method 6-10
peak detection 6-50
results 9-10
results view compounds 9-11
results view concentration 9-18
results view peak 9-15
results view ratios 9-17
run labBook 9-9
setting up template 9-2
trQuant
regions 6-60
tune settings 4-17
typographical conventions 1-2
data input 1-2
signal words 1-2
topic headings 1-3
U
unit
detection limits 6-101
standard elements 6-33
user interface
configurator 3-2
instrument control 4-2
qtegra 5-2
Thermo Scientific
user levels
application 3-6
configurations 3-6
using this manual 1-1
V
view
autotune report 4-129
completed labBook 7-22
cross calibration result 4-131
detector setup report 4-130
evaluation results 7-23
instrument state 7-24
labBook history 7-6
log messages 7-26
mass calibration result 4-133
performance report 4-128
query 7-32
readback plot 4-127
reports 7-67
result of measurement 7-22
sample list 7-15
template history 6-3
tune settings 4-17
views group 4-126
W
window
ribbon tab 4-134
software information 4-134
wizard
autotune 4-56
detector setup 4-87
mass calibration 4-116
performance report 4-28
wizards group 4-27
I-11