BENTLY.
soos Shaft CenterLINES
Monitoring versus diagnostics
Why do peak readings differ?
Randy Chitwood
systoms Bngmecring Manacer
‘Bently Nenada Corporation
onitoring and diagnostic
Mies st wll sometimes
red efferent when measur
ing the peak value of complex wave
forms associated with machinery
vibration. Knowing how the vibration
signal's peak amplitude is measured can
sive you a betier understanding of why
there are differences and how they are
apple for your benefit
lux's first define the baste vibration
measurement units we will he discussing
Peak-to-peak, zero-to-peak and
true peak
When monitoring cidkal vibration
swith proximity probes, peaketo-peak
defined as the toual movement of the
shalt in a single plane. When seismic
transducers are used to measure
acceleration or velocity, Bently Ney-
ada strongly recommends use ofthe
zero-to-peak measurement because
it is more effective and easier to
March 1994
interpret, Bently Nevada defines 20:
topeak as the peak-to-peak value of
vibration signal divided by two, At Berl
Nevada, reroto-peak is synonymous
with ewe peak (Figure 1
Derived peak
Some instruments divide a measured
[RMS value by 0.707 in order to peta peak
Coc ulimately peak-to-peak) valve, The
result, called “derived peak,” is
misleading and is not recom.
mended because itis omly accurate
for a sine wave. There
relationship between pe
salves that applies wo all waveforms. For
insane, dhe RMS vale ofa square wave
110, and of a tangle wave is 0.577
Typically a vibration signal is nota sine
wave, bucrathera complex combination
Cofseveral dtferent waveforms,
Peak detectors, a closer look
Allzeto-to peak (true peak and peal.
torpeak detectors have fundan
* Valid for sine
‘wave signals only
performance factors that aflect their
ity to respon to various input wave
ms. These performance factors are
the same whether the circuit isanalogor
digital in nature and regardless of
wwherher the circuit displays peaketo-
peak or zeo-10peak,
ently Nevada has developed ‘wo
basic pes of peak-to-peak detectors
during the 38 years i has measured
vibration. One type of peak-ospeak
tector is used in monitoring systems,
such asthe 3300 System. A second type
(of detector is used in diagnostic instru
mens, such ashe Digital Vecwor Fikes 2
4, 108 Data Acquisition Instrument
and the 208 Data Acquistion Inerface
Unit. These diferent detectors are used
because of the different uses of the
“There ts nodiference inthe measure
ment of peak-co peak between a Bently
Nevada Monitor and a diagnostic ins
ment (within the instrument's fre =
orbit 11‘quency response) when the tn peso
instruments are measuring waveforms,
such as square, sine and triangle wave
forms. In fat, this characteristic ean be
used to determine whether an insieu
iment displays a ue or a derived peak
amplitude value. Input cach waveform
with the same peak-torpeak value froma
signal generator (check it with an
oscilloscope.) Make sure the frequency
content of the signal falls within the
bandwidth of the instrument. IF the
strument’s peak (oF peak-to-peak)
value does not change as you switch
waveforms, then iC mezsuring a true
peak value (Gee Note 1. The ehiet ler:
ences are exhibited when the two types
Of instruments process waveforms with
transient noise oF complex frequency
‘Monitor peak-to-peak detectors
Bently Nevada monitors have evohed
to provide extremely relkable machine
monitoring. This has led to a design that
minimizes susceptibility 1 noise, whi
providing the peak-to-peak function.
‘Originally achieved by using dode-based
analog circuits, these circuits now use
‘combination of analog and digital tech
nology. Although the technology has
‘evolved, the ercut's response has been
‘maintained so itis consistent with older
Bently Nevara monitors (Figure 2).
Bently Nevada 7200 and 3300 Moni-
tors continually process the incoming
signal. The peakcco-peak circu output
1s constantly adjusted based on the
instantaneous value ofthe incoming si
nal and the circuit's memory of past
peaks. When the input signal excceds
Diode type peak to peak detector
Waveform simulated from pump spectrum
Velocity, 0.02 in/s/div
0.0 40
8.0 12.0
Milliseconds:
Figure 2
Monitor type peakpeak detector
the circul’s memory of past peaks, the
peak-to-peak detector increases’ the
peak-to-peak value. We describe this as
Charging, When the input signal is less
than the current memory of the past
peaks, the peak value decreases, We call
this discharging,
Charging
When the inpuc signal exceeds the
prior peak value, the peak-to-peak
detcetor charges the peak capacitor.
arly analog circuits Limited the rate at
which the peak eapacitor was charged.
This rac allows the peak-to-peak detec:
or 1 “respond” 10 the new peak at a
rite of approximately 5% of the instru
rment’s Fill scale value per millisecond
This controlled charge rate has the
added! benefit of redveing the peakto-
peak cletector’s response to transient
16.0
Note 1 The response of instruments may vary with square and triangle waves when the entire sgnal conditioning path is
consitered For example, bfgh ana lowpass ilters can disiont square and tangle waves, causing changes inthe peake-to- peas
‘waveform and readings
12 orbit ____
‘March 1994and high frequency noise. Because of
the design's success, we haven't
changed the circuit's charging rate even
though advancements in electronics
make ic echniclly feasible
A result of this controlled charging
rate is that roquencies with petiods of
Jess than J millisecond require multiple
cycles to charge the detector 10 the
Input peakecoupeak value. For example,
the peak-to-peak detector would
require approximately 20 cycles to
Charge to within 1% of ts final value IF
the inputwerea 1 ki sine wave
Discharging
If the input signal goes 10 zero, the
2 value wil discharge co zero after 2
cerain period of time. The time it takes
‘Sample & Hold peak to peak detector
Waveform simulated from pump spectrum
0.020 in/s/div
Velocity,
00 40
March 1994.
the circuit to discharge to within 37% oF
lis final value is described as the dis
charge time constant. Monitoring cir:
‘cuts use two dliferent discharge ume
‘constants. The fst discharge time can.
stant s the more common. Its used on
all monitors optioned for use ith sei
mic transducers and on radial vibration
monitors with a low pass frequency
response of 240 ¢pm, This ime constant
Is approudmately 1 second.
‘Alonger time constant of# seconds is
fnly used on radia vibration monitors
swith alow frequency response of 6 cpm
for very low speed machines. This time
constant allows the peak to peak detec
tor to remember past peaks fora longer
time period.
8.0 12.0
Milliseconds
Figure
Diagnostic type peak-to-peak detector
Diagnostic instrument peak-to-
peak detector response
Bently Nevada's diagnostic instew
‘ments calculate peak-to-peak with results
‘equal to what woukl be seen on an
‘oseiloscope. Ths ype of detectors per
formance has evolved with technology,
producing faster and more accurate
Serectors. The goat sto trackall informa.
tion ina transducer signal, egardlies of
iwsorgin
This «ype of instcument samples the
Input for a specific time or for a set
number of shaft rotations, It takes a
‘smapshor” ofthe signal during this ume
pesod, in contrast to the monitor which
{s constantly providing a valle propor
‘ional to the Current input based on the
past inputs. The diagnostic instrument
16.0
onit 13works very diferent ch
and discharge eyes.
Duringitscharge-cycte, dhe dagnestc
instrument's peak-to-peak detect
‘changes much faster than the monitor's
detector. Typically, a diagnostic instru:
_ment’s peak peak detector charges at
arate of 5% of fullscale In 50 micro
seconds, 20 times faster than a monitor's
Lever. This makes the detector vee)
responsive to-most signals.
During the discharge cycle, the dis
charge of is peakcorpeak detector hat
set interval oF 5 tiggered by the Key
phasor®. The peak-to-peak value dis
charges ina few miceasceonds, prior 10
teh new snapshot of the input signal
igure 3)
ing ts charge
Conclusion
When reading complex waveforms,
{te monitorand dagnostictnsteument’s
peak-tospeak values will difler by sn
Amounts. These readings vary because
‘ofthe different applications each inst
‘meats intended for. The monitor peak
lorpeak detecor’s reduced response 10
higherorder frequency content and
transient noise makes it very usefl for
reliable machinery monigoring. Dagnos-
tic instruments need a broadband high:
speed response to capture as much
information from the signal as possible,
toprovidea machinery diagnostics eng
ect with the data necessary todiagnose
‘machinery and instrumentation faults.
14 omit
March 1994