Professional Documents
Culture Documents
SPECTROSCOPY
February 1998
Comalco Smelting
Comalco Place
12 Creek Street
Brisbane Queensland
Australia 4000
GPO Box 153
Queensland Australia 4001
Phone: (07) 3867 1711
Fax: (07) 3867 1739
1. INTRODUCTION 3
2. OPERATION PRINCIPLES 3
3. CALIBRATION 4
3.1 Family of Curves Method 4
3.2 Master Curve Method 5
3.3 Drift Correction 5
3.4 Standards 5
4. SAMPLES 5
4.1 Sampling from Molten Metal 5
4.2 Preparation of Samples 6
4.3 Non-standard Samples 7
5. MEASUREMENT 7
5.1 Procedure 7
5.2 Accuracy and Precision 8
6. INSTRUMENT MAINTENANCE 8
7. TROUBLE SHOOTING 9
7.1 Bad Burn 9
7.2 Variation 9
8. GLOSSARY 10
9. REFERENCES 10
E1 E2
Figure 2. Schematic diagram illustrating light emission by generation of photons of characteristic wavelength.
Material is volatised from the surface of the Curves or Master Curve method which are outlined
sample resulting in emission of light. When atoms below. If further details are required the reader
are excited by an external energy source, some should refer to the ASTM standard, E1251 [3] or OES
electrons move into higher energy levels. As these equipment suppliers.
electrons drop back to their original levels, the
atoms emit light (photons) of characteristic 3.1 Family of Curves Method
wavelength. This is illustrated in Figure 2.
The spectrometer collects the light emitted from The Family of Curves method of calibration uses
the sample and splits it into its component a number of calibration standards with compositions
wavelengths using a diffraction grating. Fixed close to the alloy being tested. The calibration
detectors (photo multiplier tubes) simultaneously curve is a best fit line of intensity versus the
measure light intensity at wavelengths concentration of each element in a range of
characteristic of each element. This is illustrated in standards (Figure 4). In addition, a drift adjustment
Figure 3. is required which is based on a reference standard
Wavelengths used for OES range between 150 close to the alloy of interest.
800 nanometres. The intensity of the signals depends
on the number of photons produced per unit time.
The spectrometer is programmed for fixed sample phototube sample
types, elements and concentration ranges and only
gives results within the limits of calibration.
lens
3. CALIBRATION
mirror
Generally, prior to performing metal analysis,
primary slit
the spectrometer must be calibrated to interpret
the light spectrum emitted by the sample. This is
accomplished by inputting information relating
Binary
standards
High purity
standard
Certified reference
alloy materials
Aluminium Company 600 Block Freeport Rd. Ph: (412) 339 6702
of America New Kensington PA 15068 USA Fax: (412) 339 6740
Graham B. Jackson 125 Thomas St. Ph: (3) 9793 3322
Pty. Ltd. Dandenong Vic. 3175 Australia Fax: (3) 9794 7240
Measurement region
Figure 6. Example of chill-cast mould and disk sample with measurement region shaded on the schematic diagram.
Measurements can be performed around the disk, Previous work by Comalco has illustrated the
except at the edge (5 mm) and in the centre (up to effect of non-standard samples, as shown in Tables
12 mm radius) to avoid possible segregation. This is 2 and 3. Measurements on billet slices and extrusion
illustrated in Figure 6. The unknown sample and profiles generally gave lower values compared to
the cross-check reference samples should be the analysis in the original melt certificate. Re-
prepared by the same method. melting of an extruded profile and casting of a
standard chill-cast disk resulted in an analysis
4.3 Non-standard Samples similar to that obtained from the original chill-cast
disk from the melt.
The correct samples for composition analysis by
OES are chill-cast disks collected from the melt and 5. MEASUREMENT
measured according to the relevant standards [3-5].
Measurements on billet slices or wrought product 5.1 Procedure
samples are only approximate since alloy The analysis of aluminium alloys is performed
calibrations are based on chill-cast disk reference according to standard ASTM E 1251 (JIS 2611-
samples. This inaccuracy is due to matrix effects 1977 [15]). Prior to conducting measurements, the
caused by the differences in the metallurgical OES must be calibrated, and corrected for any drift,
structures of the standards and samples. If analyses as discussed in Section 3.
are required on cast billet, ingot or wrought samples, The machined chill-cast disk sample is sparked
they should be re-melted and sampled according to two to four times in the measurement region
the ASTM E716 [4,5]. However, great care is required shown in Figure 6. Excessive variation in the
to avoid losses of volatile elements (such as composition from repeated sparks (or burns) is
magnesium, Mg). Care must also be taken to avoid normally a sign of a problem in the sample or its
contamination during re-melting. preparation.
Table 3. Example of differences in OES measured composition (wt%) for chill-cast disk sample compared to billet slice (6063 alloy)
measured along the log length [13].
Chill-cast Disk from Melt Billet Slice Analysis
0.2 m 3.5 m 6.8 m 0.2 m 4m 7m
Si 0.44 0.47 0.47 0.41 0.43 0.43
Fe 0.16 0.17 0.17 0.15 0.15 0.16
Cu 0.03 0.04 0.04 0.03 0.03 0.03
Mg 0.50 0.52 0.52 0.51 0.49 0.49
Mn 0.03 0.03 0.03 0.03 0.03 0.03
Ti 0.014 0.015 0.016 0.015 0.015 0.016
To ensure that the OES instrument is delivering The general precision of OES testing is given in
accurate results, certified reference standards Table 4, while Table 5 shows the precision for
(similar alloy to that being tested) are routinely particular elements in the range of interest for
cross-checked. These standards are measured prior aluminium alloys [3].
to the unknown alloy analysis. The cross check The 3 sigma values define the 99.73% confidence
results should be recorded and monitored. limits (or the range which covers 99.73% of results).
Appropriate action is required if the precision falls
outside acceptable criteria, as discussed below.
Accurate results can only be assured by tracing 6. INSTRUMENT MAINTENANCE
calibrations to primary standards/methods, and if
outside established control chart limits, by applying The cleaning and general maintenance
a drift correction. The cross-check analysis should procedures used by Comalco (on an ARL 3560 OES)
be repeated after making the drift correction. are given below as a guide [11]. The instrument
instruction manual should be referred to for specific
5.2 Accuracy and Precision maintenance and cleaning practices.
The sample stand, table, glass bowl and o-rings
Accuracy is the closeness of the measured value should be cleaned with tissue paper before each
to the real value, while precision is a measure of analysis. Thorough cleaning of the sample stand is
repeatability. The accuracy and precision of the also required after a number of samples have been
spectrochemical measurement are a function of analysed (up to about 30), or when changing alloys.
concentration and may deteriorate at low levels of Solvent should not be used to clean the table cavity
elements. and stand, but ethanol can be used to clean the glass
The accuracy of spectrochemical measurements bowl and the electrode tip.
is assessed by determining the deviation from a When deposits form on the electrode, it should be
reported elemental concentration for an accepted cleaned and and sharpened as required. A wire
reference standard. brush can be used for cleaning of the electrode tip
OES may be used for quantitative elemental prior to sample analysis, although care must be
analysis of metals to about 13% relative precision. taken to avoid contamination.
[3]
Table 5. OES Precision for Some Elements in Aluminium Alloys .
Induction Coupled Plasma analysis (ICP)A wet (2) J-L. Fortier and F.M. Kimmerle, 1987 O.E.S.
chemical analysis technique where a liquid Quality Control of Aluminium Alloys
sample is drawn up into a nebulizer, atomised Composition, Quality and Process Control,
in a stream of argon, and vented into an Light Metals, p155.
induced argon plasma. Sample atoms are
excited, and the emitted light intensity at (3) ASTM E 1251-94, Standard Test Method For
each wavelength is measured. Optical Emission Spectrometric Analysis Of
Aluminium And Aluminium Alloys By The
PhotonA massless particle carrying energy. Argon Atmosphere, Point-To-Plane, Unipolar
Self-Initiating Capacitor Discharge, (1994),
Point-To-Plane MethodA self-initiating oscillatory Vol 03.06.
capacitor discharge or triggered capacitor
discharge is produced between a prepared (4) ASTM E 716-94 Standard Practices For
flat surface of the sample and the tip of a Sampling Aluminium And Aluminium Alloys
shaped graphite electrode. For Spectrochemical Analysis, (1994), Vol.
03.06.
Quantitative AnalysisAnalysis to determine the
elemental composition of the sample. (5) AS 2612-83, Aluminium And Aluminium
Alloys Sampling And Preparation Of Solid
Relative PrecisionA measure of reproducibility Samples For Optical Emission Spectrometry,
expressed as a percentage of the concentration (1983).
of the element being measured.
(6) K. Slickers, 1983Automatic Emission
SegregationNon-uniform distribution of alloying Spectroscopy, ARL.
elements or phases resulting in localised
concentrations. (7) Optical Emission Spectrometers Systems
Description, 1982Bausch and Lomb ARL.
SparkA short duration electric discharge causing
volatilisation of material accompanied by (8) D.E. Groteke, 1994Improving Your
emission of light. Spectrometric Analysis of Aluminium,
Modern Casting, September, p36.
SpectrometerA spectroscope calibrated for the
measurement of light of specific wavelengths. (9) J-L. Fortier and C. Tremblay, 1984
Segregation Effects in Aluminium Metal
SpectrumThe ordered arrangement of Samples, Light Metals, p1,373.
electromagnetic radiation according to
wavelength.
IMPORTANT DISCLAIMER
This brochure was prepared with no
particular reader in mind and therefore,
although we believe that the information
herein is accurate and reliable, no warranty
of accuracy, reliability or completeness is
given and (except insofar as liability under
any statute cannot be excluded). No
responsibility arising in any other way for
errors or omissions or negligence is accepted
by the company or any director, employee, or
agent of the company.