You are on page 1of 18

OPTICAL EMISSION

SPECTROSCOPY

TECHNICAL REPORT NO. 5

February 1998

Comalco Smelting
Comalco Place
12 Creek Street
Brisbane Queensland
Australia 4000
GPO Box 153
Queensland Australia 4001
Phone: (07) 3867 1711
Fax: (07) 3867 1739

Comalco Technical Report, No. 51


CONTENTS

1. INTRODUCTION 3
2. OPERATION PRINCIPLES 3
3. CALIBRATION 4
3.1 Family of Curves Method 4
3.2 Master Curve Method 5
3.3 Drift Correction 5
3.4 Standards 5
4. SAMPLES 5
4.1 Sampling from Molten Metal 5
4.2 Preparation of Samples 6
4.3 Non-standard Samples 7
5. MEASUREMENT 7
5.1 Procedure 7
5.2 Accuracy and Precision 8
6. INSTRUMENT MAINTENANCE 8
7. TROUBLE SHOOTING 9
7.1 Bad Burn 9
7.2 Variation 9
8. GLOSSARY 10
9. REFERENCES 10

This report was prepared for Comalco Smelting by Comalco


Research & Technical Support. All enquiries should be
directed to Comalco Smeltings sales office or agents.

Comalco Smelting is a division of Comalco Aluminium Ltd.


ACN 009 679 127

2Comalco Technical Report, No. 5


OPTICAL EMISSION
SPECTROSCOPY
ABSTRACT wavelengths inside the spectrometer. Fixed
detectors simultaneously measure light intensity at
Optical Emission Spectroscopy (OES) is used to wavelengths characteristic of each element.
measure the composition of aluminium alloys for OES requires calibration of the spectrometer
control of molten metal composition. It is a rapid with certified chill-cast standards, which must have
test method with relative precision of 13%. the same matrix as the alloys to be tested. Therefore
Samples are collected from the melt according to quantitative OES can only be performed on chill-
standard ASTM E716-94 while analyses are made cast samples collected from the melt according to
according to ASTM E1251-94 by sparking of a standard methods [4,5]. Other techniques, such as
clean machined surface. The resulting emitted light Induction Coupled Plasma (ICP), can be used for
is measured using detectors at fixed wavelengths. quantitative analysis of alloys in other forms, for
The composition of the sample is determined by example billet slices, ingots, extrusions and forgings.
comparison of the measured intensity of each
element with that of a reference standard of known 2. OES PRINCIPLES
composition.
This technical report gives a basic introduction The OES system consists of four main parts [6]:
to Optical Emission Spectroscopy, as it is used in the sample stand;
measurement of the composition of aluminium alloy generator;
melts. spectrometer; and
Attention to sample collection, OES calibration, electronics assembly.
monitoring, and sample preparation are highlighted The clean sample is mounted in the stand and a
as essential for accurate and precise measurement spark is generated between the sample and a
of composition. tungsten electrode. An example of this set up is
illustrated in Figure 1. High purity argon is used as
1. INTRODUCTION a discharge atmosphere to prevent any interaction
between the atmosphere and the sample surface.
Optical Emission Spectrometry (OES) is a fast,
accurate and reliable method for quantitative
analysis of metals to about 13% relative precision.
sample
Optical emission spectroscopy methods are among
the most useful and flexible means of performing
elemental analysis of aluminium alloys [1].
effective
Accurate analysis of metals and alloys requires sample
that [2]:
the melt or bulk sample is homogeneous;
samples are representative of the melt or bulk discharge
sample; and spot
the metallurgical structure of the sample is similar discharge plasma
to that of the standards used. (instantaneous
The sample requirements are met by chill-cast exposure)
disk samples collected as described in ASTM E 716-
94 (or equivalents AS2612-83 or JISH 1305-1976).
For the point-to-plane method of spectrochemical
analysis [3] , samples are machined on a lathe to give
a smooth, clean surface just prior to analysis. counter
The sample is mounted in the spectrometer and
electrode
a spark is generated between the sample and a
tungsten electrode. Material is volatilised from the
surface of the sample resulting in emission of light.
[7]
This light is diffracted into its component Figure 1. Sample excitation by electrical discharge .

Comalco Technical Report, No. 53


Applied Photon,
energy, E2-E1 = hc

E

E1 E2

Initial energy E1 Electron excited, Return to initial energy


energy E2 with emission of a photon
of wavelength

Figure 2. Schematic diagram illustrating light emission by generation of photons of characteristic wavelength.

Material is volatised from the surface of the Curves or Master Curve method which are outlined
sample resulting in emission of light. When atoms below. If further details are required the reader
are excited by an external energy source, some should refer to the ASTM standard, E1251 [3] or OES
electrons move into higher energy levels. As these equipment suppliers.
electrons drop back to their original levels, the
atoms emit light (photons) of characteristic 3.1 Family of Curves Method
wavelength. This is illustrated in Figure 2.
The spectrometer collects the light emitted from The Family of Curves method of calibration uses
the sample and splits it into its component a number of calibration standards with compositions
wavelengths using a diffraction grating. Fixed close to the alloy being tested. The calibration
detectors (photo multiplier tubes) simultaneously curve is a best fit line of intensity versus the
measure light intensity at wavelengths concentration of each element in a range of
characteristic of each element. This is illustrated in standards (Figure 4). In addition, a drift adjustment
Figure 3. is required which is based on a reference standard
Wavelengths used for OES range between 150 close to the alloy of interest.
800 nanometres. The intensity of the signals depends
on the number of photons produced per unit time.
The spectrometer is programmed for fixed sample phototube sample
types, elements and concentration ranges and only
gives results within the limits of calibration.
lens
3. CALIBRATION
mirror
Generally, prior to performing metal analysis,
primary slit
the spectrometer must be calibrated to interpret
the light spectrum emitted by the sample. This is
accomplished by inputting information relating

signal intensities to chemical concentration for


particular elements. The relationship between


optical emission intensity and concentration must
be established using certified standards. Normally, grating
reference standards are certified by wet chemical
methods.
Comalcos Smelters and Research Centre have
calibrated their OES units via either a Family of Figure 3. Optical systems of a typical OES spectrometer.

4Comalco Technical Report, No. 5


standards to produce the Master Curve ranging
from 0.0001 to 6 wt% Cu which covers the range for
Outside calibration most aluminium alloys. Three certified 6063
Mn standards are used to adjust for matrix effects and
(wt%) range
establish the alloy working curve from 0.001 to 0.09
wt% Cu. These standards are also used for drift
Each point correction.
represents a
calibration
standard 3.3 Drift Correction
Drift correction may be required due to changes
in the optics, the excitation source, processing
electronics and even ambient room temperature or
humidity [8]. These changes can cause drifts in the
Intensity intensity ratios from those recorded during the
initial calibration. If a drift correction is not made,
errors will occur and false concentration readings
Figure 4. Example of a calibration curve for Family of Curves will result.
method. The drift may be corrected using a two-point
standardisation procedure. Standards are measured
at high and low concentrations for the elements of
This method of calibration is suited to analysis of
interest, and adjustments are made. For the Master
relatively few alloys of similar composition. An
Curve method of calibration, matrix correction
analytical curve is required for each element based
standards can also be used as drift correction
on 520 standards covering a limited concentration
standards.
range. Providing the standards are available, the
Family of Curves method can be used for in-house
calibration of a spectrometer.
3.4 Standards
Standards should be selected to cover the
3.2 Master Curve Method concentration ranges of all elements for which the
spectrometer is capable. The standards should also
The Master Curve method of calibration assumes
match the alloy type being evaluated.
that there is a relationship between the shape of the
Certified standards can be obtained from a
master curve and the alloy (working) curve for each
number of sources, as shown in Table 1.
element in a given alloy matrix. There may be a
change in the background, but no change to the
shape of the curve. Binary standards containing
only the element of interest in a pure aluminium
4. SAMPLES
base are used to establish the master curve and The point-to-plane method is dependent on the
then a number of certified reference alloys matrix and metallurgical structure of the samples.
(commercially available) are used to define the Hence, the samples must be carefully collected to
alloy (working) curve for a given alloy type. This is ensure compatibility with the certified standards.
illustrated in Figure 5. This involves collection of chill-cast disk samples.
Master Curve calibrations are generally set up by However, segregation profiles can also exist in
the equipment suppliers or an OES consultant using chill-cast samples [9,10], and thus the conditions of
certified binary reference standards. Fifty or more sample collection, as well as the surface preparation
binary standards are normally required to cover must be controlled.
the entire concentration range for common
aluminium alloy types. Up to four certified reference 4.1 Sampling from Molten Metal
standards of similar composition to the alloy of
interest are then used to adjust for the alloy matrix. Sample collection for OES analysis should be
These standards can also be used to correct for performed according to standard ASTM E716-94
instrument drift. (equivalents AS 2612-83 [5] or JISH 1305-1976 [14]).
Setting up a calibration curve for copper (Cu) in Different moulds may be used to collect
6063, for example, requires 36 Al-Cu binary representative samples of the melt, but these must

Comalco Technical Report, No. 55


Master curve - Alloy (working) curve -
Concentration defined using defined using matrix
binary standards correction standards

Binary
standards

High purity
standard

Spectral Intensity ratio


background

Certified reference
alloy materials

Figure 5. Relationship between master and alloy curves.

be capable of producing homogeneous chill-cast 4.2 Preparation


disks having smooth surfaces free of surface pockets
and porosity. One type of chill cast disk is shown in Surface preparation is critical, and particular
Figure 6. care is required in sample preparation to ensure a
The molten metal from which the sample is clean flat surface is presented for OES measurement.
collected must be well above the liquidus (melting) The sprue of the chill-cast disk is removed, and
temperature. Dross must also be removed from the the bottom surface is machined using a lathe or
surface of the melt prior to sampling. A ladle is milling machine to a specified depth (1422% of
preheated and then dipped into the metal and disk thickness) below the original surface, according
stirred below the surface before quickly to ASTM E716-94. This depth is important since it
withdrawing. The disk mould is carefully filled at corresponds to the composition that best represents
an even rate allowing air to escape. The metal must the average composition of the whole disk [9].
freeze without disturbance. The initial cast disk is The surface finish should be smooth, with only
used for preheating the mould and should be fine tool marks, and should be the same as the
discarded. An additional sample is collected for the reference sample. Ethanol or isopropanol, may be
analysis disk. Collecting an additional sample may used as the machining lubricant to avoid
not be required if the mould is already hot. contamination of the surface.

Table 1. Some suppliers of OES Chill-cast Reference Standards.

Company Address Contact Numbers


Analytical Reference PO Box 2246 Ph: (303) 670 1300
Materials International Inc. Evergreen Colorado 80439 USA Fax: (303) 670 1392

Aluminium Company 600 Block Freeport Rd. Ph: (412) 339 6702
of America New Kensington PA 15068 USA Fax: (412) 339 6740
Graham B. Jackson 125 Thomas St. Ph: (3) 9793 3322
Pty. Ltd. Dandenong Vic. 3175 Australia Fax: (3) 9794 7240

6Comalco Technical Report, No. 5


64mm 613mm

Measurement region

Figure 6. Example of chill-cast mould and disk sample with measurement region shaded on the schematic diagram.

Measurements can be performed around the disk, Previous work by Comalco has illustrated the
except at the edge (5 mm) and in the centre (up to effect of non-standard samples, as shown in Tables
12 mm radius) to avoid possible segregation. This is 2 and 3. Measurements on billet slices and extrusion
illustrated in Figure 6. The unknown sample and profiles generally gave lower values compared to
the cross-check reference samples should be the analysis in the original melt certificate. Re-
prepared by the same method. melting of an extruded profile and casting of a
standard chill-cast disk resulted in an analysis
4.3 Non-standard Samples similar to that obtained from the original chill-cast
disk from the melt.
The correct samples for composition analysis by
OES are chill-cast disks collected from the melt and 5. MEASUREMENT
measured according to the relevant standards [3-5].
Measurements on billet slices or wrought product 5.1 Procedure
samples are only approximate since alloy The analysis of aluminium alloys is performed
calibrations are based on chill-cast disk reference according to standard ASTM E 1251 (JIS 2611-
samples. This inaccuracy is due to matrix effects 1977 [15]). Prior to conducting measurements, the
caused by the differences in the metallurgical OES must be calibrated, and corrected for any drift,
structures of the standards and samples. If analyses as discussed in Section 3.
are required on cast billet, ingot or wrought samples, The machined chill-cast disk sample is sparked
they should be re-melted and sampled according to two to four times in the measurement region
the ASTM E716 [4,5]. However, great care is required shown in Figure 6. Excessive variation in the
to avoid losses of volatile elements (such as composition from repeated sparks (or burns) is
magnesium, Mg). Care must also be taken to avoid normally a sign of a problem in the sample or its
contamination during re-melting. preparation.

Comalco Technical Report, No. 57


Table 2. Example of changes in OES measured composition (wt%) due to non-standard sampling of a 6000 alloy [12] .
Melt Cast billet Homogenised Extruded Extrusion re-melted
certificate slice billet slice profile into chill-cast disk
Si 0.77 0.78 0.73 0.73 0.80
Fe 0.22 0.21 0.19 0.19 0.22
Cu 0.18 0.20 0.19 0.19 0.20
Mg 0.85 0.81 0.78 0.78 0.83
Zn 0.003 0.003 0.002 0.002 0.003
Ni 0.009 0.009 0.008 0.008 0.009
Mn 0.11 0.11 0.11 0.11 0.11
Cr 0.08 0.07 0.07 0.07 0.08
Ti 0.013 0.014 0.015 0.015 0.014

Table 3. Example of differences in OES measured composition (wt%) for chill-cast disk sample compared to billet slice (6063 alloy)
measured along the log length [13].
Chill-cast Disk from Melt Billet Slice Analysis
0.2 m 3.5 m 6.8 m 0.2 m 4m 7m
Si 0.44 0.47 0.47 0.41 0.43 0.43
Fe 0.16 0.17 0.17 0.15 0.15 0.16
Cu 0.03 0.04 0.04 0.03 0.03 0.03
Mg 0.50 0.52 0.52 0.51 0.49 0.49
Mn 0.03 0.03 0.03 0.03 0.03 0.03
Ti 0.014 0.015 0.016 0.015 0.015 0.016

To ensure that the OES instrument is delivering The general precision of OES testing is given in
accurate results, certified reference standards Table 4, while Table 5 shows the precision for
(similar alloy to that being tested) are routinely particular elements in the range of interest for
cross-checked. These standards are measured prior aluminium alloys [3].
to the unknown alloy analysis. The cross check The 3 sigma values define the 99.73% confidence
results should be recorded and monitored. limits (or the range which covers 99.73% of results).
Appropriate action is required if the precision falls
outside acceptable criteria, as discussed below.
Accurate results can only be assured by tracing 6. INSTRUMENT MAINTENANCE
calibrations to primary standards/methods, and if
outside established control chart limits, by applying The cleaning and general maintenance
a drift correction. The cross-check analysis should procedures used by Comalco (on an ARL 3560 OES)
be repeated after making the drift correction. are given below as a guide [11]. The instrument
instruction manual should be referred to for specific
5.2 Accuracy and Precision maintenance and cleaning practices.
The sample stand, table, glass bowl and o-rings
Accuracy is the closeness of the measured value should be cleaned with tissue paper before each
to the real value, while precision is a measure of analysis. Thorough cleaning of the sample stand is
repeatability. The accuracy and precision of the also required after a number of samples have been
spectrochemical measurement are a function of analysed (up to about 30), or when changing alloys.
concentration and may deteriorate at low levels of Solvent should not be used to clean the table cavity
elements. and stand, but ethanol can be used to clean the glass
The accuracy of spectrochemical measurements bowl and the electrode tip.
is assessed by determining the deviation from a When deposits form on the electrode, it should be
reported elemental concentration for an accepted cleaned and and sharpened as required. A wire
reference standard. brush can be used for cleaning of the electrode tip
OES may be used for quantitative elemental prior to sample analysis, although care must be
analysis of metals to about 13% relative precision. taken to avoid contamination.

8Comalco Technical Report, No. 5


[8]
Table 4. OES Precision capability .

Element Range Tolerance


(wt%) ( wt%)
0.001 - 0.009 0.002
0.041 - 0.10 0.010
0.76 - 1.00 0.05
1.51 - 2.00 0.07
5.51 - 6.00 0.17
> 7.01 3%

Typically the lens should be cleaned every three


months while replacement of the filter which collects
metal concentrates is generally required on a
monthly basis.
Check the argon supply frequently. Also regularly
check the oil level in the vacuum pump and replace
Figure 7. Photograph illustrating good burns on a chillcast
the oil every six months.
disk sample.
Preventative maintenance of the OES instrument
should be performed (by the equipment supplier or
agent) on a six-monthly basis. 7.2 Variation
Variation in results may be due to a number of
7. TROUBLE SHOOTING possible causes. These include:
incorrectly collected chill-cast disk which contains
7.1 Bad Burn inclusions or porosity, or a chill-cast disk collected
from a melt with low metal temperature;
A bad burn is usually due to an air leak between inhomogeneous sample, or incorrect machined
the sample and the top of the excitation chamber. depth;
This may be caused by a crack in the sample, or contamination during sample machining and
samples which have not been machined flat. preparation;
A bad burn has shallow pitting in the centre area drift correction not performed; and
of the spark, and is surrounded by a white/brown/ contamination of OES instrument.
silver ring. If the aluminium intensity is measured, Discrepancies may also arise between sites or
the value will be low in the case of a bad burn, and instruments due to the use of different standards,
the results should be rejected. In contrast, a good or use of the wrong standards.
burn, shown in Figure 7, features a deeply pitted
area in the centre of the spark, and is surrounded by
a black ring. Glossary and References over page

[3]
Table 5. OES Precision for Some Elements in Aluminium Alloys .

Element Wavelength Low 3* High 3*


(nm) Concentration ( wt%) Concentration ( wt%)
(wt%) (wt%)

Na 588.995 0.0006 0.0003 0.015 0.003


Ti 363.545 0.10 0.0007
Cr 425.435 0.0012 0.0002 0.3 0.003
Cu 327.396 0.001 0.0003 4.8 0.11
Mn 403.076 0.0015 0.0002
Si 288.158 0.005 0.002 1.2 0.03
Si 212.415 0.052 0.005 11.3 0.16
Mg 285.213 0.15 0.002 1.6 0.07
Mg 518.362 1.6 0.02 3.0 0.08
* standard deviation

Comalco Technical Report, No. 59


8. GLOSSARY SprueAn opening through which molten metal is
poured into a mould.
AccuracyThe closeness of the measured value to
the true value. VolatiliseConversion of a substance from a liquid
or solid state to a gaseous or vapour state.
Binary AlloyAn alloy composed of two principal
metallic components, for example Al-Mg. WavelengthThe distance between two points
having the same phase in two consecutive
Diffraction GratingAn optical device which cycles of a periodic wave.
consists of an assembly of narrow slits or
grooves which produce a large number of Wrought AlloyAn alloy that is mechanically
light beams of different wavelengths. worked after casting.

ExcitationThe addition of energy to a particle or


substance at ground state to produce an 9. REFERENCES
excited state.
(1) H.J. Heine, 1996Analysing Elemental
HomogeneousSubstance having uniform Analysers, Foundry Management and
composition or structure. Technology, October, p30.

Induction Coupled Plasma analysis (ICP)A wet (2) J-L. Fortier and F.M. Kimmerle, 1987 O.E.S.
chemical analysis technique where a liquid Quality Control of Aluminium Alloys
sample is drawn up into a nebulizer, atomised Composition, Quality and Process Control,
in a stream of argon, and vented into an Light Metals, p155.
induced argon plasma. Sample atoms are
excited, and the emitted light intensity at (3) ASTM E 1251-94, Standard Test Method For
each wavelength is measured. Optical Emission Spectrometric Analysis Of
Aluminium And Aluminium Alloys By The
PhotonA massless particle carrying energy. Argon Atmosphere, Point-To-Plane, Unipolar
Self-Initiating Capacitor Discharge, (1994),
Point-To-Plane MethodA self-initiating oscillatory Vol 03.06.
capacitor discharge or triggered capacitor
discharge is produced between a prepared (4) ASTM E 716-94 Standard Practices For
flat surface of the sample and the tip of a Sampling Aluminium And Aluminium Alloys
shaped graphite electrode. For Spectrochemical Analysis, (1994), Vol.
03.06.
Quantitative AnalysisAnalysis to determine the
elemental composition of the sample. (5) AS 2612-83, Aluminium And Aluminium
Alloys Sampling And Preparation Of Solid
Relative PrecisionA measure of reproducibility Samples For Optical Emission Spectrometry,
expressed as a percentage of the concentration (1983).
of the element being measured.
(6) K. Slickers, 1983Automatic Emission
SegregationNon-uniform distribution of alloying Spectroscopy, ARL.
elements or phases resulting in localised
concentrations. (7) Optical Emission Spectrometers Systems
Description, 1982Bausch and Lomb ARL.
SparkA short duration electric discharge causing
volatilisation of material accompanied by (8) D.E. Groteke, 1994Improving Your
emission of light. Spectrometric Analysis of Aluminium,
Modern Casting, September, p36.
SpectrometerA spectroscope calibrated for the
measurement of light of specific wavelengths. (9) J-L. Fortier and C. Tremblay, 1984
Segregation Effects in Aluminium Metal
SpectrumThe ordered arrangement of Samples, Light Metals, p1,373.
electromagnetic radiation according to
wavelength.

10Comalco Technical Report, No. 5


(10) G. Blancfene and F. Dugain, 1986Quality
Control of Spectrographic Standards for
Aluminium and Aluminium Alloys,
International Molten Aluminium Processing
Conference, p449.

(11) P. Napolitano, Brief Description of CRCs


OES Analysis, CRC:TN:96:0289.

(12) C. Taylor, 1995Internal Report, 22


November.

(13) D. Irwin, 1997Internal Report, 17


September.

(14) Japanese Industrial Standard JISH 1305


1976Method for Photoelectric Emission
Spectrochemical Analysis of Aluminium and
Aluminium Alloy.

(15) Japanese Industrial Standard JISZ 2611


1977General Rules for Photoelectric
Emission Spectrochemical Analysis of Metal
Materials.

IMPORTANT DISCLAIMER
This brochure was prepared with no
particular reader in mind and therefore,
although we believe that the information
herein is accurate and reliable, no warranty
of accuracy, reliability or completeness is
given and (except insofar as liability under
any statute cannot be excluded). No
responsibility arising in any other way for
errors or omissions or negligence is accepted
by the company or any director, employee, or
agent of the company.

Comalco Technical Report, No. 511


12Comalco Technical Report, No. 5
Comalco Technical Report, No. 513
14Comalco Technical Report, No. 5
Comalco Technical Report, No. 515
16Comalco Technical Report, No. 5
Comalco Technical Report, No. 517
18Comalco Technical Report, No. 5

You might also like