Cite National
‘CIPM MRA
Metrology Centre
AYSTAR i
CALIBRATION REPORT
Report No. i
Customer
‘Subject Details
Subject x
Manufacturer
Model t
Serial Number
Sales Order No.
Calibration Date
Ambient Conditions
Temperature
Relative Humidity:
ipa
Bay Swee Pheng
Calibration Officer
01000711 Page 4 of 2
: Hi-Tech NDT Inspection Services (S) Pte Lid
Bik 40, Toh Guan Road East,
+#01-64 Enterprice Hub
Singapore 608582
llluminance Meter
: TES
1335
: 060202820
2 2028010260
+ 2013-08-01
: (23#2)°C
(60 + 10) % relative humidity
Yuaniie
Approving Officer
Optical Metrology
For futher enquires, please contact the calibration offer at Te: +85 62791955, Fax: +85 92791992 or Ema:
Day_swee_pheng@nme.a-star.edu.sg
National Metrology Centre [Note : Ths Report i issued subject ro the
| Science Park Drive Singapore 118221 “Tees and Condon for Sevies"avalable
‘Tek 68) 6279 1900 Fax (65) 6279 1992 at wwwnmea-saredusg and on request rom
Website: wwwinmea-staredusg Naor Metology Centre. The Report isnot
2 Cerificate of Qual. on apis tothe
"ample ofthe specific productlequipment
Science and Engineering Institutes (Co.Reg No, 2007208292) ven a che time ofits esting calibration,
TRogare the way of ech oars calraon ind maureen certs fr the quanta ange ond mestrenent unre: specied happend
(or deete see hepsivwnbipm re)Report No. : 01000711 Page 2 of 2
Method of Calibration
The subject has been calibrated at the National Metrology Centre (NMC) according to the NMC
calibration procedure (PS/EVI001) under the ambient conditions stated above.
The test meter was mounted with the plane of its diffuser vertical and normal to the direction of
measurement. The test meter head was illuminated normally by the luminous intensity working standard
lamp. The distance between the working standard lamp and the test meter was adjusted to give the test
meter readings as recorded.
The calibration was carried out using the following working standards traceable to national reference
standards maintained at the National Metrology Centre:
1) Luminous intensity working standard lamp operated at a colour temperature of 2856 K (serial no.
0s01)
Results of Calibration
The results of calibration are shown on table below.
The expanded uncertainty of measurement is 1.4 %, estimated at a level of confidence of approximately
95 % with a coverage factor k = 2.
‘The user should determine the suitability of this instrument for its intended use.
‘The actual recalibration interval should be determined based on the user's requirements.
Calibration results of TES 1335 ( serial no: 060202820 ) illuminance meter:
Description Applied Value Of Indicated Value Of Deviation
Working Standard Test Meter
(ix) (x) (%)
Range :
400 Ix 102.9 86.0 “16.4
389.1 323.7 168
4.000 x 389.1 324 “16.7
2043.6 1669 17.0
yar
ration Officer