Notebook 11
Quality Control
Quality Control is the aspect of quality assurance that monitors technical equipment to
maintain quality standards. Rooted in the need to stabilize the various equipment components
of the radiographic imaging chain.
Quality Assurance consists of activities that provide adequate confidence that a radiology
service will render consistently high-quality images and services.
Tests done semiannually for Quality Control
= Itis the technologists job to understand and pick up if there is a problem with the
‘equipment
- Iti the physicists job to measure the equipment to the appropriate setting and
the service engineers do the calibrating.
Focal Spot Size Estimation
Line pair resolution tool, star test patterns,
and pinhole cameras are used. All are used
to analyze the focal spot size. The variance
tolerated to 50 percent.
Half-Value Layer
Dosimetry equipment is used to measure the
quantity and quality of the aluminium filtration
that will reduce the beam intensity to half the
original value.
Collimator, central ray, and bucky tray
accuracy
A collimator test tube is used to ensure
accuracy. An SID difference of 2 percent or
less is the acceptable amount between the
primary beam image and the light field size
Distance and centering indicators’ accuracy
Assurance that distance and centering are
accurate through testing with a distance
indicator. Distance indicators are tested to
ensure they are close to 10 percent from the
ideal spot, and centering indicators must be
around 2 percent from where the center
should be.
‘Angulator or protractor accuracy
A large protractor is used for angle
Measurements and a level to verify that locks,
‘stops, and detents are set to establish
horizontal and perpendicular surfacesKilovoltage Accuracy
‘Tend to drift from accuracy over time
- Computerized dosimeters provide
digital readouts or printouts of average
and single-pulse kVp, if settings drift
beyond 5 of what was set it needs to
be adjusted
- Tested annually by physicist
Timer Accuracy
- Also drifts over time
- Verification and calibration of the time
‘ensure technique exposure charts will
produce diagnostic-quality images
- Exposure time settings must be
maintained to 5 percent of what the
label reads to ensure accuracy-
= Checked annually by the physicist
Protective Apparel
-Technologists fluoro lead
aprons,gloves, thyroid shields for
holes/cracks
- Photometer used to measure light
intensity
- 10 % variance allowedmR/mAs and milamperage linearity
mA settings drift over time
Comparative measurements are made
after time and kVp accuracy have
been verified
Computerized dosimeters provide
digital readouts of mR/mAS
measurements
MA stations should read out to 10
percent of each other
Exposure reproducibility
Checked because generators must be
capable of repeating exposures
accurately
Test is done by taking several
exposures with the same techniques
and the computerized dosimeter
provides digital readouts or printouts
of mR/mAS
‘Must be maintained within 5 percent
of each other, if not the entire series of
tube and generator tests should be
analyzed to attempt to isolate the
problem
Other technologists responsibilities include weekly cleaning and inspecting receptors, cleaning
air intakes of the imaging plate reader, clean display screen, and clean computer keyboard and
mouse.