TECHNICAL DATA SHEET No.

SPECS-GAS-124

Rev. 3
INFRARED TOTAL
HYDROCARBON ANALYZER Page 1 of 1

TITLE: INFRARED TOTAL HYDROCARBON ANALYZER

DEPARTMENT: TECHNICAL SUPPORT

DESCRIPTION:

The main advantage of infrared sensors


over catalytic and flame ionization
detectors (FID) is that it is unaffected by
catalytic inhibitors and poisons. It also
does not consume the sample gas
(unlike other types of detectors which
consume themselves in the process).

The advantages above make the


infrared-based system suitable for
stand-alone applications as the sensor
will practically last indefinitely and will
retain calibration for months at a time.

Finally, to address the inherent lower


sensitivity disadvantage of IR compared
to combustion-based sensors, the IR
Total Gas system is designed with both
low-range and full-range sensor-transmitter sets to achieve a comparable and respectable
0.1% sensitivity and repeatability.

SPECIFICATIONS:
RANGE: 0-100% Volume (Methane equivalent)
RESOLUTION: 0.1% @ C1 < 5% Volume
1.0% @ C1 5% Volume
OUTPUT: WITS Level 0 over Serial
0-5 volts for 0-5% Volume
0-5 volts for 0-100% Volume
FEATURES: Compatible with all Electronic Drilling Recorders (EDR); Onboard
LCD Display (field read out available); dilution of gas sample is
unnecessary.
CERTIFICATIONS: CSA/FM/UL/CENELEC Explosion Proof - Class I Div 1 (Zone 1),
Groups B, C, and D

APPROVAL:

Technical Supervisor :____LANCE WHEAT________ Date: 3rd April 2009

Commercialization Mgr :____VICTOR SAET_________ Date: 3rd April 2009