Current Transfemoral Sockets
CCUINICAL ORTHOPAEDICS AND RELATED RESEARCH
Numosr 3, pp 48-54
(01999 Lippingat Willams & Wikies, Ine
C. Michael Schuch, CPO; and Charles H. Pritham, CPO
Clinicians have two types of prosthetic sockets
for patients with a transfemoral amputation:
the quadrilateral socket introduced in the
1950s and the ischial containment socket intro-
duced in the 1980s, For many years, the
quadrilateral socket was acceptable to clini
cians and patients alike. With the introduction
of the ischial containment socket claims were
made that the quadrilateral socket had been
inadequate from its inception. It also was
claimed that the ischial containment design
was the answer to all problems that patients
with transfemoral amputations may have, Un-
fortunately, there is only one scientific study
that lends any support to claims of the ischial
containment socket’s superiority and its find-
ings are limited. Still, clinicians must make
recommendation and prescription decisions
daily regarding what is best for their patients.
There are two acceptable socket designs for
patients with transfemoral amputations, Expe-
rience with both supports the concept that
they have more similarities than differences
and that each has a legitimate place in the
treatment of patients with transfemoral ampu-
tations. A comprehensive understanding of
‘cach socket design and its biomechanical in
tentions is essential for successful clinical ap-
plication and treatment of patients with ampu-
tations,
From the *Depantent of Prosthetics and Onhoiies,
Duke University Medical Center, Dutham, NC,
Reprint requests to C. Michael Schuch, CPO, Depart:
‘ment of Prosthetics and Orthotics, M04 Davison Build-
ing, Duke University Medical Center, Durham, NC
270.
48
The basic goals for fitting and aligning pros-
theses for patients with a transfemoral ampu-
tation seem simple enough: comfort, func-
tion, stability, and cosmesis. Obtaining these
goals is significantly more challenging than
might be expected. One reason is the opinion
that many practitioners have taken too many
shortcuts in trying to treat the patient with a
tansfemoral amputation, It is disturbing to
note the number of new practitioners who
have never read the publications of Rad.
liffe, who is considered worldwide to be the
founding expert of transfemoral (above
knee) biomechanics and socket design.* It is
‘equally disturbing to consider the number of,
experienced practitioners who do not follow
these time proven principles. Simple formu-
las for the quadrilateral socket design have
been ignored, forgotten, or replaced with
shortcuts, New transfemoral socket designs
that incorporate the principle of ischial con
tainment have been developed and Michael*
reports that “these new designs represent
evolutionary rather than revolutionary ad-
vances”. Pritham? has stated his belief that
the principle of ischial containment “is fully
compatible with Radcliffe's biomechanical
analysis of the function of the quadrilateral
socket and that the varying socket configura-
tions are not at odds, but rather, are separate
but related entities in @ continuum labeled
above knee sockets”
The biomechanical requirements of the
patient with a transfemoral amputation func-
tioning with a prosthesis do not vary because
of socket design: the outcomes may vary, but
the needs remain the same. SuspensionNumber 351
Apa 1999
should not be influenced to any substantial
degree by socket design. Alignment in the
sagittal plane should be the same from
socket to socket, with variations arising only
from the stability afforded by different knee
and foot components and/or the patient's de-
sire for more or less alignment stability.
There are no documented or stated sagittal
plane alignment variations attributable to
socket design. Coronal plane alignment
might vary with mediolateral socket stabil-
ity, but the variations ultimately are not sub-
stantial
The goals of this paper are for the reader
to: (I) recognize and acknowledge that there
are two acceptable socket designs for trans-
femoral amputations: (2) understand the sim-
ilarities and differences of these wo designs;
(3) understand the indications for each; and
(4) be able to relate accepted transfemoral
biomechanical principles to cach.
ASSESSMENT AND EVALUATION
OF THE RESIDUAL LIMB RANGE
OF MOTION
Careful measurement and evaluation of the
patient’s anatomy and kinesiology are essen-
tial for correct socket design and initial
socket alignment. Important and less under-
stood is the need for accurate measurement
and evaluation of the range of motion of the
residual limb in the sagittal and coronal
planes, that is, flexion and adduction analy-
sis, Proper planning and incorporation of
these angular measurements into the socket
‘and overall prosthesis design allows for cer-
tain biomechanical and alignment principles
that are advantageous to the patient during
the various phases of gait. {tis the authors’
contention that this is the most neglected
area of evaluation and treatment of a patient
with a transfemoral amputation
BIOMECHANICS
Four key phases of the gait eycle are consid-
eed when meeting the biomechanical objec-
Current Transtemoral Sockets 49
tives for a transfemoral prosthesis. The bio-
mechanical objectives may be met or as
sisted by any or all of the following: socket
esign, prosthetic alignment, and component
selection. The relevance of socket design to
each of these four key phases of the gait cy-
cle is discussed below.
Heel Strike and Initial Stance Phase
For the patient with a transfemoral amputa-
tion, dissipation of forces at heel strike is not
possible in the normal manner. Thus. heel
strike and initial stance phase are a period of
potential knee instability for these patients.
‘The biomechanical requirement is that of
knee stability, maintaining the prosthetic
knee in complete extension, and overcoming
the tendency of the shank to rotate forward
and cause the knee to buckle. The transfemoral
socket must be designed in a position of flex-
ion at least 5° greater than the resting flexion
angie of the patient’s residual limb. This posi-
tions their hip extensors on stretch allowing,
them to more effectively control the knee flex-
ion moment at heel strike.
Midstance
Saunders et al!® described the six determinants
of gait that provide efficiency in locomotion.
‘The hip abductors, primarily the gluteus
medius, maintain pelvic control during mid-
stance through eccentric and isometric con-
traction. This determinant of gait allows for a
narrow based gait and upper trunk control, and
minimizes excursion of the center of gravity.
However, in the case of the patient with a
transfemoral amputation, the femur is not part
of a structure that terminates in a foot firmly
planted on the ground. The residual femur is
only a lever of less than half of the normal
length of the entire lower leg and floats in a
mass of muscle, tissue, and fluid. As a result,
in midstance, the residual femur tends to dis
place faterally in the mass of residual muscle
and tissue, making it difficult to. maintain
horizontal stability of the pelvis and trunk.
Effective pelvis and trunk stabilization and
the resultant narrow based gait only can be80 Schuch and Pritham
achieved in a transfemoral prosthesis by pro-
viding adequate lateral support to the entire
femoral shaft, maintaining normal or maxi-
‘mum possible adduction of the femur, and dis-
tributing the proximal medial component of
the resulting midstance force couple evenly
against the adductor muscle mass of the upper
thigh of, as is the case of the ischial contain-
ment socket, against the medial aspect of the
ischium and ramus (Fig 1). This is accom-
plished by appropriate socket design and
alignment.
Heel Off and Terminal Stance Phase
Effective heel off and terminal stance
propulsion are provided in a transfemoral
prosthesis via properly assessed hip flexion
analysis and proper initial hip flexion de-
signed in the socket. By designing and align-
ing the socket in a position of initial flexion,
the patient is allowed to more easily reach
the 15° of extension required on the pros-
thetic side to provide a normal stride length
on the sound side.
Swing Phase
Swing phase tracking addresses the smooth-
ness of the pathway of the prosthetic limb
during the swing phase of the gait cycle.
Goals are lack of vertical displacement of the
prosthesis with respect to the residual limb,
and lack of deviation in the sagittal plane as
the prosthetic limb advances forward during
the swing phase. Problems regarding vertical
displacement are the result of poor suspen-
sion and resulting piston action, Deviations
in the sagittal plane include whips during the
swing phase, which may be caused by im-
proper socket shape, Socket design geometry
and volume are critical for smooth swing
phase tracking
AMPUTATION TECHNIQUE AND.
SOCKET DESIGN
In their study published in 1989, Gottschalk
et al} reported: “We believe that socket con-
Clinical Ornapaedics
and Folated Research
alters
|
Fig 1. Transtemoral midstance force couplo di-
gram. The force couple creates significant sta:
bilization pressure at the proximal medial
socket, The adluctor muscle mass of the upper
thigh must be capable of withstanding and sta-
bilzing these forces in the quadrilateral socket
design. (Reprinted with permission from Bray J,
Staats T: University of California, Los Angeles,
Above-knee Teaching Manual 1977.)
figuration does not influence the position of
the residual fermur and that a proper surgical
procedure for above knee (transfemoral) am-
putations is essential for functional restora-
tion and success of prosthetic fitting”. This
point is reiterated in their most recent paper.*
‘The present authors do not disagree with the
premise of Gottschalk et al; in fact, it would
bbe ideal if all transfemoral amputations were
performed by surgeons with such a sure
knowledge of, and passion for, surgical bio-
mechanics. However, this is mot the case.
Transfemoral amputation is morbid and morNurroer 361
foe 1909)
tal surgery, and more frequently than not is
performed by surgeons other than or-
thopaedists, Who are muich more concerned
with life saving measures than postamputa-
tion rehabilitation potential."? The authors
contend that even in the best transfemoral
amputations. proper detail to socket design
amd biomechanics is crucial to transfemoral
outcomes,
SOCKET ANALYSES
ilarities and Differences
General Overview
Similarities between the quadrilateral and is-
chial containment sockets are biomechanical
principles, socket volume, distal socket de-
sigh. and socket alignment in initial flexion
and adduction. There is one striking differ-
cence between the sockets, which is direct is-
chial bearing in the quadrilateral socket, as
‘opposed to ischial containment in that so
hamed socket design. AS a result, the con-
touts of the proximal brim of the ischial con-
tainment socket are different from the proxi-
mal brim contours of the quadrilateral
socket.
Ischial containment may be defined as a
Proximal extension of the posteromedial
brim of the socket so that it bears against the
pelvis, specifically the medial aspects of the
ischial tuberosity and the ramus of the is-
chium (Fig 2), Ischial beating has been ac-
cepted as ditect inferior support to the ischial
tuberosity by means of a flat, relatively hori-
Zontal surface of the posterior brim that pro-
Vides vertically directed upward forces. This,
is opposed to the rather oblique and sloping
contour of the proximal most posteromedial
brim of the ischial containment socket. Dis
tal fo this oblique and sloping contour of the
ischial containment socket, at a point ap-
proximately 3 or 4 em below the edge of the
brim, contouring beneath the ischial tuberos-
ity may occur and thereby provide the same
"measure of ischial bearing as is present in the
‘quadrilateral socket? (Fig 3)
Current ransfemoral Sockets 81
Fig 2. Schematic sagittal plane view of a right
ischial containment socket showing the concept
of ischial containment and the significantly low-
ered medial wall allowing the inferior ramus of
the pubis to exit the socket comfortably.
(Reprinted with permission from Pritham CH:
Biomechanics and shape of the above-knee
‘socket considered in light of the ischial contain-
ment concept. Prosthet Orthot Int 14:9-21,
1990.)
Distribution of the proximal and medial
concentration of forces to the ischium helps
reduce trauma to the soft tissue of the per-
ineum. Obtaining rotational stability in the
quadrilateral socket is dependent on muscle
channel contours and generally is unsuecess-
ful in an unstable residual limb. Rotational
forces are better controlled by purchase
against the medial border of the ischium and
balanced by the accurate diameter between
the medial ischial surface and the greater
trochanter.52__Schuch and Pritham
Fig 3. Schematic sketch of a coronal view of
the ischial containment concept. The high pres-
sures from the medical component of the mid-
stance force couple aro stabilized against the
pelvis, specifically the ischial tuberosity and ra-
mus. (Reprinted with permission from Pritham
CH: Biomechanics and shape of tho abovo-
knee socket considered in light of the ischial
containment concept. Prosthet Orthot Int
14:9-21, 1990.)
Anteroposterior and Mediolateral
Diameters
‘The anteroposterior (AP) dimension of the
quadrilateral socket will be smaller than that
of an ischial containment socket, because the
ischium rests proximal and posterior to the
inner AP dimension of the quadrilateral
socket. Conversely, the ischium is within the
Glinioal Onthopacdies
and Felaied Resuarch
ischial containment socket, necessitating 2
greater inner AP diameter.
Measurement of the mediolateral diame.
ter is referenced differently by various prac-
joners. There is no consensus as to the best
‘anatomic landmarks, The term narrow medi-
olateral socket is misleading because this di-
mension varies between males and females
because of anatomic differences between the
male and female pelvis® (Fig 4).
Fig 4. Schematic transverse views of the vary-
ing pelve configurations ft with ischial contain-
ment brim shapes. Note the medial wall
variations based on pelvic variations. In the top
view the ischial walls are wide, as in a fomeale
The lowest view shows ischii that are closer (0
cone another, and the midline of the body, as in a
male. (Reprinted with permission from Hoyt CP.
LitigD, Lunct J, Staais TB: The UCLA GatCam
Above Knee Prosthesis. Los Angoles, University
of Caliomia, Los Angeles Prosthetics Education
‘and Research Program 14-49, 72, 1987.)‘umber 364
‘ont 1999)
INDICATIONS,
CONTRAINDICATIONS, AND
RECOMMENDATIONS FOR
SOCKET DESIGN
The indications and contraindications for
transfemoral socket design were considered
by a panel of experts in 1987." A sum-
mary of their conclusions follow. There were
no specific contraindications reported for
any socket design, Some advocated not at-
tempting to change successful quadrilateral
socket users to the newer ischial containment
socket design, Successful fitting of quadri-
lateral sockets is more likely to be attained
on long, firm residual limbs with intact ad-
ductor musculature, whereas ischial contain-
ment sockets are more successful than
‘quadrilateral sockets on short, fleshy, unsta-
ble residwal limbs, [schial containment sock-
ets are the better recommendation for high
activity sports participation and ronning. Fi-
nally, there was lack of agreement on the
best recommendation for patients with bilat-
eral transfemoral amputations.
Ic has been the authors” experience that is
chial containment sockets are preferred by
‘most patients with bifateral transfemoral am-
purations. Correctly designed quadritateral
sockets function fine in the authors’ experi-
ence when used on Ionger amputations that
reveal principles of surgical biomechanics.>4
‘The more of the adductor magnus intact, the
more successful the quadrilateral fit. Quadri-
lateral sockets ate recommended for the geri-
atric or debilitated patient. These patients
ambulate with canes and walkers for assis-
tance, thereby canceling out or greatly de-
creasing the typical medial lateral pelvic and
trunk stability demands of midstance.
For the very short fleshy, and rotationally
unstable residual limb, ischial containment
sockets have provided suecess where quadri-
lateral sockets have failed previously. In ad-
dition to enhanced stability and perineal
Comfort for these patients with transfemoral
amputations, the additional socket volume
afforded by the higher brims enhances the
Current Tanstemoral Sockets 63
ability to obtain and maintain suction sus-
pension,
Flexible brim sockets are recommended for
use with ischial containment sockets. The
socket frame or retainer should provide rigid
support in the posteromedial comer of the
proximal brim, under and beside the ischium:
however, the remainder of the proximal frame
‘can be lowered below the level of the flexible
socket brim. Use of thermoplastic flexible
sockets is common practice.
Discussion has focused on the tans-
femoral socket designs of quacrilateral and
ischial containment, including. similarities
and differences, indications and recommen-
dations. The authors strongly propose the
case for recognizing and clinically using
both socket designs, depending on the indi
vidual patient and his or her prognosis, level
of amputation, and residual limb characteri
tics. A comprehensive understanding of each.
socket design and their biomechanical inten-
tions is essential for successful clinical ap-
plication and treatment of patients with am-
putations,
References
1. Donovan RG. Prtham CH, Wilson AB (eds): Report
(of ISPO Werkshops: International Workshop on
‘Above-Knee Fitting and Alignment, Gasgow, Iner-
national Society for Prostetes and Orthotics 31-37,
‘Appendices AF. 1987
2, Gailey RS, Lawrence D, BuntitC, etal The eatcam
socket and quadrilateral socker ‘comparison of en-
ergy cost during ambulation. Prosthet Orthot Ink
1798-100, 1998,
3. Gowchalk PA, Kourosh S, Stills M, MeClellan B,
Robens J: Does socket configuration influence he
position ofthe femur in above-kace amputation? J
Prosthet Onthot 2:94-102, 1989.
4. Gowtschalk FA, Stills M: The biomechanics of trans
femoral amputation. Prosthet Onthot Int 18: 12-17,
19s
5. Hoyt CP, Litig D, Lundk J, Staats TB: The UCLA
CatCam Above-Knee Prosthesis, Los Angeles, Uni
versity of California, Los Angeles Prosthetics Educe
tion and Research Program 14-49, 72, 1987
6. Michael JW: Courrem concepts in above-knee
Socket design. Instr Course Lect 30:373-378,
190.
7. Pritham CH: Biomechanics and. shape of the
above-knee socket considered in light ofthe ischial
containment concept, Prosthet Ortot {nt 49-21
1990.54 Schuch and Fritham
8
Radelife CW: Functional considerations in the ft-
ting of above-knee prostheses. Af Limbs 2:35.60,
1935,
Sabolich J: Contoured adducted trochanteric = con
twoed alignment method (ct-cam): Introduction and
basic principles. Clin Prosthet Onhot 9:15~26, 1985.
Saunders 18, Inman VT, Eberhart HD: The major de=
terminanis in eormal and pathologie gnit. 1 Bone
Joint Surg 35A°543-558, 1953,
‘Schuch CM: Repor from interational workshop on
Clinical Othopaedics
and Related Research
above-knee fiting and slignment techniques. Clin
Prosthet Onhot 12:81-98, 1988,
‘Sehuch CM: Modern abovesknee fitting. practice.
‘Prosthet Onhot ln 12-77-90, 1988,
Schuch CM: Above-Knee Amputation: Literature
Review and Prosthetic Experience. In Murdoch G.
Jacobs NA, Wilson Jr AB (eds). Report of ISRO
‘Consensus ‘Conference on Amputation Surgery:
Copentiagen, Inernational Society for Prosthetics
sind Onhoties 66-68, 1992