Sample Preparation For Metals in Food PDF
Sample Preparation For Metals in Food PDF
Abstract: The present article gives an overview of recent publications and modern
techniques of sample preparation for food analysis employing atomic and inorganic
mass spectrometric techniques, such as flame atomic absorption spectrometry,
chemical vapor generation atomic absorption and atomic fluorescence spectrometry,
graphite furnace atomic absorption spectrometry, inductively coupled plasma optical
emission spectrometry, and inductively coupled plasma mass spectrometry. Among
the most frequently applied sample preparation techniques for food analysis are dry
ashing, usually with the addition of an ashing aid, and acid digestion, preferably
with the assistance of microwave energy. Slurry preparation, particularly with the
assistance of ultrasound, is increasingly used to reduce acid consumption and sample
preparation time. Direct analysis of solid samples is gaining importance in the field
of food analysis as it offers the highest sensitivity, avoids the use of acids and other
67
Downloaded By: [Korn, Maria das Graas Andrade] At: 13:30 6 February 2008
aggressive reagents, makes possible the analysis of micro-samples, and can be applied
for fast screening analysis, e.g., of fresh meat.
INTRODUCTION
Elemental food composition data are important to both consumers and health
professionals, and recent food labeling legislation has highlighted this require-
ment. The determination of trace elements and contaminants in complex
matrices, such as food, often requires extensive sample preparation and/or
extraction regimes prior to instrumental analysis. Flame atomic absorption
spectrometry (FAAS), graphite furnace atomic absorption spectrometry (GF
AAS), and inductively coupled plasma optical emission spectrometry (ICP
OES) are the main techniques used for the determination of trace element
contents in food analysis laboratories. The traditional techniques for sample
preparation are time consuming and require large amounts of reagents,
which are expensive, generate hazardous waste, and might contaminate the
sample with the analytes. Advances in sample preparation over the last few
decades have been propelled by the advance of microwave-assisted acid
digestion (1 3), ultrasound-assisted, extraction and slurry preparation (4),
and direct solid sampling analysis (5).
Quality control and safety in the food supply chain demands reliable
methodology that is both rapid and easily transferable. In order to minimize
the uncertainty in sample preparation a number of factors need to be con-
sidered. As statistically the degree of uncertainty in a method is directly
related to the number of stages involved, a minimization of that number
should reduce the uncertainty proportionally. Automation and mechanization
of processes also leads to a reduction in uncertainty. Automated procedures
are generally more reproducible than manual methods and will also
decrease the staff time spent on sample preparation, which is often the bottle-
neck in analytical laboratories (4, 5).
This review will discuss recent development in procedures for sample
preparation of food samples particularly under the above-mentioned aspects.
The discussion emphasizes analytes, samples, and effects on measurement
conditions using atomic and mass spectrometric techniques.
DRY-ASHING TECHNIQUES
vessel and removing the organic matter from the samples by thermal
decomposition, normally in the presence of an ashing aid, using a muffle
furnace. Typical ashing temperatures are 450 to 5508C at atmospheric
pressure, and the ash residues are dissolved in an appropriate acid. The
degree of volatilization loss is a limiting factor and depends on (i) the
applied temperature, (ii) the form in which the analyte is present in the
sample, and (iii) the chemical environment in the ashing stage. Oxidizing
reagents may be used as ashing aids in order to prevent the volatilization of
analytes and also to speed up the ashing process. High-purity magnesium
nitrate and magnesium oxide are commonly used for that purpose (6).
Several papers have been published about dry ashing as a sample prep-
aration method for metal determination in food samples. Tuzen et al. investi-
gated the application of dry ashing to promote the decomposition of fish (7),
baby food (8), and honey (9). Approximately 1 g of sample was submitted to
dry ashing at 4508C for 4 16 h, depending on the matrix, and the residue was
dissolved in nitric acid. Aluminum, Cd, Co, Cr, Cu, Fe, Mn, Se, and Zn were
determined using GF AAS; recoveries were quantitative (95%) for all inves-
tigated elements.
Mindak et al. (10) developed a flow-injection hydride generation atomic
absorption spectrometry (FI-HG AAS) method for the determination of total
arsenic and selenium in food. A combination of microwave-assisted
digestion with nitric acid and dry ashing utilizing magnesium nitrate and
magnesium oxide as ashing aids was used to destroy the organic matrix
including refractory organometallic compounds present in many food
samples. Complete mineralization of these compounds is a pre-requirement
for the application of hydride generation for these elements. The resulting
ash was dissolved in hydrochloric acid and diluted to volume. The method
was validated using 21 food samples and nine reference materials.
The application of dry ashing methods is simple and large quantities of
food samples may be treated at the same time. This procedure permits the pre-
concentration of trace elements in the final solution, which is useful when very
low concentrations are to be determined. The ash is also completely free of
organic matter, which is a prerequisite for some analytical techniques. The
addition of an ashing aid, on the other hand, increases the content of
inorganic salts significantly, which might be a problem for the subsequent
determination of trace elements, and it might also contribute to contamination,
necessitating careful blank control.
WET-ASHING TECHNIQUES
0.5 10% for all analytes, and the quantification limits were lower than
2.5 mg g21. Although acceptable results were obtained with all procedures,
wet digestion with HNO3/H2SO4/H2O2 was recommended because of the
better recovery. The good agreement between measured and certified concen-
trations for IAEA-331 and IAEA-359 CRM indicates that the developed
analytical method is well suited for the determination of toxic and nutrient
elements in legumes and possibly similar matrices.
ICP OES with continuous hydride generation was used for the determi-
nation of As in seafood samples (14). The lyophilized samples were
digested with concentrated nitric and sulfuric acid. The reliability of the
developed method was checked by analyzing several CRM. Complete miner-
alization was obtained for an arsenobetaine-containing CRM with a mixture of
nitric and sulfuric acids followed by adding hydrogen peroxide in an open
digestion system and a digestion time of 4 h.
Studies on the transfer of chemical contaminants through the food chain
provide useful information for the development of surveillance programs
aimed at ensuring the safety of the food supply and minimizing human
exposure to toxic agents. Tinggi et al. (15) investigated two wet digestion pro-
cedures using acid mixtures of HNO3/H2SO4/HClO4 and HNO3/H2SO4 for
decomposition of food samples in Australian diet. The addition of hydrofluo-
ric acid to the mixture of HNO3/HSO4 was also investigated for the determi-
nation of Cr. All the acid mixtures tested were found to be satisfactory but, for
safety reasons, HNO3/H2SO4 was the method of choice. Olivares et al. (16)
employed a wet digestion procedure using a mixture of nitric, perchloric,
and sulfuric acids for sample preparation of common Chilean foods for the
determination of Fe, Zn, and Cu by FAAS and assessed the intake of these
elements in a population living in Santiago, Chile. In another study, the
levels of essential elements, such as Cu, Cr, Fe, and Zn, and toxic elements
such as Al, Ni, Pb, and Cd were evaluated in a total of 40 samples of
legumes and 56 samples of nuts that are widely consumed in Spain (17).
These samples were mineralized in a digestion block with HNO3 and V2O5
and determined using GF AAS as the analytical technique. The reliability of
the procedure was checked by the analysis of a CRM; no matrix effects
were observed and aqueous standard solutions were used for calibration.
Kira et al. (18) developed a fast procedure for the determination of Ca, Cr,
Cu, Fe, K, Mg, Mn, Na, P, and Zn in milk samples by ICP OES. This procedure
consisted of a partial digestion with hydrochloric acid on a hot plate. The results
were compared with two digestion procedures (dry ashing and microwave-
assisted acid digestion). All the procedures showed similar levels of
precision, with coefficients of variation ,10% for the majority of the
elements. The accuracy was evaluated using a CRM, and the values were
within the confidence intervals for these products. Rodriguez et al. (19) used
a mixture of HNO3 and HClO4 (9:1 v/v) for the preparation of bovine milk
samples. Calcium, Cu, Fe, K, Mg, Na, Se, and Zn were determined by
FAAS, FAES, and fluorimetry. In another paper, Cava-Montesinos et al. (20)
Downloaded By: [Korn, Maria das Graas Andrade] At: 13:30 6 February 2008
compared two sample preparation procedures for bovine milk samples for the
determination of Se and Te by HG AFS. The first digestion was in a
microwave oven using HNO3 and 30% v/v H2O2. The other procedure
consisted of many stages using a muffle furnace and a hot plate. A suspension
with 10% (m/v) Mg(NO3)2 . 6H2O and 1% (m/v), MgO as an ashing aid was
used for dry ashing of the sample in the muffle furnace. The ash was treated with
KBr and HCl before the quantification of the analytes using HG AFS. The
proposed method involved the use of a low-cost instrumentation, and
microwave-assisted sample pretreatment provides fast and accurate results.
Ferreira et al. (21) developed an acid digestion procedure for the determi-
nation of Cu in various food samples of animal and plant origin. A (3:1 v/v)
mixture of HNO3:HClO4 was used for digestion of the samples on a hot plate
until the total oxidation of the organic material. Copper was determined by
FAAS.
Santos et al. (22) proposed a fast and inexpensive wet digestion procedure
for beans samples. Essential (Ca, Cu, Fe, K, Mg, Mn, Ni, P, Zn) and non-
essential (Al, Ba, Sr) elements were determined in bean digestates by ICP
OES. Experimental designs for five factors (HNO3 and H2O2 volume,
digestion time, block temperature, and particle size) were used for optimiz-
ation of the digestion procedure, adopting a factorial experiment with 2521
design. The factor block temperature was found to be the most important
parameter and Doehlert designs were applied in order to determine the
optimum conditions. Digestion conditions were attained using 3.5 mL of con-
centrated HNO3 for 45 min. The accuracy of the results was demonstrated
using one CRM (spinach leaves NIST 1570a) and comparison with the
recommended official method.
Microwave-Assisted Digestion
different digestion methods for biscuits prior to the determination of iron and
zinc using FAAS. In an initial study, MW-assisted digestion with hydrochloric
and nitric acids 3:1 v/v, 1808C, and 600 W provided an accuracy (spike
recovery 96 102%), precision and digestion time comparable to dry ashing,
wet digestion (using different acid mixtures), and also to a simple acid
treatment at room temperature. The latter technique was further investigated
because of its simplicity and to reduce the digestion time. The addition of
ethanol was found necessary to digest the organic residue at room tempera-
ture. The method was validated by comparison of the data found for commer-
cial biscuit samples using the proposed procedure and the AOAC official
spectrophotometric reference method. Jalbani et al. (24) investigated the
dietary intake of aluminum in bakery products consumed in the urban areas
of Hyderabad, Pakistan. Samples of different branded and non-branded
bread and biscuits were dissolved using MW-assisted and conventional wet
acid digestion prior to GF AAS analysis.
Costa et al. (25) used a factorial design for optimization of a focused-
MW assisted digestion of bean samples for the determination of Ca, Fe,
Mg, Mn, and Zn. A closed-vessel MW-assisted digestion was used to
certify the elemental compositions obtained after open digestion. The
accuracy was checked using the NIST SRM 8433 Corn Bran CRM. Results
were in agreement with certified values at the 95% confidence limit using a
Students t-test. Volumes of nitric and sulfuric acid, temperature, and the
interaction between the initial volumes of HNO3 and H2SO4 were significant
variables according to the P-values in the analysis of variance (ANOVA).
Santelli et al. (26) also used a Doehlert matrix response surface method to
optimize a focused-MW assisted digestion of various food samples for the
determination of Fe, Mn, and Zn by FAAS. Three variables, irradiation
power, irradiation time, and composition of the oxidant solution
(HNO3 H2O2), were considered as factors in the optimization study. The
working conditions were established as a compromise between optimum
values found for each analyte, taking into consideration the robustness of
the procedure. These values were 260 W, 12 min, and 42% (v/v) for
irradiation power, irradiation time, and percentage of H2O2 in solution,
respectively. The accuracy of the optimized procedure was evaluated by the
analysis of CRM and by comparison with a well-established closed-vessel
MW-assisted digestion method.
Nookabkaew et al. (27) used MW-assisted digestion for three types of
popular herbal tea products, Gynostemma pentaphyllum, Camellia sinensis,
and Morus alba, which are widely consumed in Thailand and in the rest of
the world. The contents of Mg, Al, Ca, V, Cr, Mn, Fe, Co, Ni, Cu, Zn, Se,
Sr, Sb, Ba, As, Cd, Hg, and Pb were determined by ICP-MS. Costa et al.
(28) investigated a focused-MW assisted extraction of Al, Ca, Mg, and Mn
in tea leaves. The efficiency of extraction was evaluated using diluted acid
and alkaline solution of a tertiary amine in water. The extraction procedure
was implemented in 5 min. A hot plate digestion procedure was developed
Downloaded By: [Korn, Maria das Graas Andrade] At: 13:30 6 February 2008
for comparison. A colorless digest was obtained for all the tea samples and the
total content for each analyte was used to calculate the efficiency of the extrac-
tion. The determination was carried out by FAAS and ICP OES; quantitative
recovery (80 100%) was obtained for Ca, Mg, and Mn, but for Al recoveries
were only 10 80%, which is probably due to the strong interaction of this
element with the matrix. The results indicated that the percentages of extrac-
tion of the elements are directly related with the chemical form in which the
metal is present in the tea leaf.
Cubadda et al. (29, 30) used closed-vessel MW-assisted digestion for
determination of Cd, Pb, Fe, Ni, and Cr in selected food matrices by plasma
spectrometric techniques and to investigate the transfer of metal contaminants
through the food chain and the effect of food processing. Cadmium, Pb, Fe,
Ni, and Cr, were accurately determined in durum wheat grain and
derived products, wheat-based reference materials, and drinking water, used
both as an ingredient and for technology purposes in the industrial process.
The analytical determination was performed using ICP-MS. Changes in
the sample introduction system and complementary use of ICP OES
overcame the difficulties in determining the analytes in the food matrixes.
The benefits of ultrasonic nebulization in reducing spectral interferences
were demonstrated. Overall, a robust analytical method with high sample
throughput was developed.
The determination of trace elements in edible oils is important because of
both the metabolic role of metals and possibilities for adulteration detection
and oil characterization. Zeiner et al. (31) used an MW-assisted digestion of
the olive oil in closed vessels with a mixture of nitric acid and hydrogen
peroxide; the trace element content of olive oils was determined by ICP
OES and GF AAS. Recently, the quantification of selected metals in various
oils (olive, pumpkin seed, sunflower, sesame seed, hazelnut, grape, soy, and
rice oil) was carried out by ICP OES and GF AAS after MW-assisted
digestion (32). Differences in the metal concentrations for edible oils
obtained in this preliminary study were the basis for the development of an
additional analytical procedure applicable for oil characterization.
Santos et al. (33) developed a method for the determination of Ba, Ca, Cu,
K, Mg, Na, P, and Zn in whole and non-fat bovine milk after digestion in a
focused MW oven, using an alternate procedure based on gradual sample
addition to hot and concentrated acids. A two-level 23 full factorial design
experiment with eight runs was carried out to evaluate the optimum exper-
imental conditions for reducing both the RCC and the final acidity of the
digestates. The best conditions were attained by adding small aliquots of
milk (ten additions of a volume of 0.5 mL during 5 min) to a digestion
mixture containing 3.0 mL nitric acid plus 1.0 mL sulfuric acid heated at
1058C. It was demonstrated that the digestion efficiency of the alternative
procedure was better than the conventional procedure; i.e., 98% compared
to 80% for the latter one. The accuracy was checked using two CRM,
whole and non-fat milk powder. This strategy expands the application of
Downloaded By: [Korn, Maria das Graas Andrade] At: 13:30 6 February 2008
advantages are the low RCC of less than 1.4% without reflux and less than
0.3% with a reflux step and the possibility of using diluted acid as the
absorbing solution.
Araujo et al. (42) investigated the efficiency of MW-assisted acid digestion
of plants using different concentrations of nitric acid with hydrogen peroxide
(30% v/v) by measuring the RCC using ICP OES with axial viewing. Two
CRM, spinach leaves (NIST 1570a) and corn bran (NIST 8433), were used
for evaluating the accuracy attained when 2 mol L21 HNO3 was employed
for digestion. Under all experimental conditions RCC values were lower than
13% w/v, and even the highest concentration did not cause any interference
with element recovery. It has been observed that the high pressure that is
attained in closed-vessel operation improved the oxidative action of nitric
acid due to consequent temperature increase, even when this reagent was not
used at high concentration. The residual acid present in the digestates varied
from 1.2 to 4.0 mol L21, depending on the initial acid concentration. Hence,
for plant materials, MW-assisted acid digestion can be carried out under mild
conditions, which implies that digestates do not need extensive dilution
before introduction by pneumatic nebulization into an ICP OES. An additional
advantage is the lower amount of residue generated when working with less
concentrated acid solutions.
Ultrasonic Extraction
Nascentes et al. (44) proposed a fast and accurate method for the extrac-
tion of Ca, Mg, Mn, and Zn from vegetables using ultrasonic energy and dilute
acid. Optimized conditions for the ultrasonic bath were 1 L of water, 258C,
and 2% v/v detergent concentration. The best conditions for extraction
were 0.14 mol L21 HNO3, 10 min of sonication time, and a particle size
,75 mm. The accuracy of the proposed ultrasound-assisted extraction
method was assessed by using certified reference materials, as well as wet
digestion.
An ultrasound-assisted leaching procedure using diluted mixed acid
solution was developed for the determination of cadmium, copper, and zinc
in fish and mussel samples by FAAS (45). The effects of several parameters
such as nitric acid, hydrochloric acid, and hydrogen peroxide concentration,
volume of leaching solution, and sonication time were investigated. A 30-min
sonication, 568C operating temperature and 6 mL of a 1:1:1 mixture of 4 mol
L21 HNO3, 4 mol L21 HCl, and 0.5 mol L21 H2O2 were used for 0.5 g of
dried sample. The results from the proposed procedure were compared with
those obtained by microwave-assisted digestion, and the recovery obtained
with the leaching technique ranged from 92 to 114% for fish and from 88 to
103% for mussel samples. The accuracy of the developed method was investi-
gated by analyzing a certified reference material (DORM-2). Melo et al. (46)
developed an ultrasound-assisted metal extraction with an aqueous solution of
tertiary amines (CFA-C reagent) in the presence of hydrogen peroxide for
improving metal solubilization in fish samples. The optimization was carried
out using a factorial design. The proposed procedure made possible the quanti-
tative extraction of Ca, Cu, Fe, Mg, and Zn. Accuracy was evaluated by com-
parison with total acid digestion of the sample in a Parr bomb and using a CRM
(fish homogenate, MA-A-2, IAEA). All results were in agreement at a 95% con-
fidence level according to a paired-t test.
Three different ultrasonic-based sample treatment approaches, automated
ultrasonic slurry sampling, ultrasound-assisted acid solid-liquid extraction
(ASLE), and enzymatic probe sonication (EPS), were compared for the determi-
nation of Cd and Pb by GF AAS in biological reference materials (47). The
sample mass chosen to perform the analysis was 10 mg and the liquid volume
was 1 mL of 1 mol L21 nitric acid. Optimum performance (total metal extrac-
tion) of ultrasound-assisted ASLE for Cd was only achieved in two of the four
materials investigated, and total Pb recovery was only possible in three of the
five samples. Total extraction with the enzymatic probe sonication was only
obtained for Cd in oyster tissue. Neither ASLE nor EPS were able to extract
Cd or Pb from spruce needles. Pb concentration obtained after EPS was
found to be highly dependent on sample centrifugation speed and time.
Krishna and Arunachalam (48) investigated the application of an ultra-
sound-assisted extraction procedure for the determination of major, minor,
and trace elements in lichen and mussel tissue as a possible alternative to con-
ventional digestion methods. ICP-MS and ICP OES were used for the quanti-
fication of the elements. Parameters affecting extraction, such as extractant
Downloaded By: [Korn, Maria das Graas Andrade] At: 13:30 6 February 2008
Slurry sampling was considered to have certain advantages over direct solid
sampling, since it is possible to change the slurry concentration by simple
dilution, hence combining the advantages of solid and liquid sampling.
Another advantage that has been claimed is that aqueous standards may be
used for calibration. However, the stabilization of the slurry, its homogeneity,
particle size, and sedimentation also have to be considered.
Li and Jiang (50) used an electrothermal vaporization dynamic reaction
cell ICP-MS to determine trace elements in rice slurry samples. The
influence of instrument operating conditions and slurry preparation on the
ion signals was investigated. Since the sensitivities of Cr, Cu, Cd, Hg, and
Pb in the rice flour slurry and aqueous solution were quite different,
standard addition and isotope dilution methods were used for the determi-
nation of these elements in NIST SRM 1568a rice flour CRM and two rice
samples purchased from the market. The analytical results for the CRM
agreed with the certified values. The results for the rice samples, for which
no reference values were available, were also found to be in good
agreement between the isotope dilution and standard addition methods.
Vinas et al. (51) developed a rapid and accurate procedure for the deter-
mination of Se, Cd, and Pb in different types of baby food using slurry
Downloaded By: [Korn, Maria das Graas Andrade] At: 13:30 6 February 2008
magnetic stirrer was used for on-line slurry formation in a dispersant solution of
0.5% v/v Triton X-100 and 1% v/v HNO3. A Babington-type nebulizer
combined with cyclonic-type spray chamber was adopted for on-line slurry neb-
ulization. The recommended particle size was ,70 mm and the slurry concen-
tration was 0.6% m/v, while the working slurry concentration could range from
0.3 to 3.3% m/v with proportional sensitivity. Excellent agreement was found
between the standard addition calibration procedure and calibration against
aqueous standard solutions for almost all of the investigated elements. The
reliability of the proposed method was confirmed by comparing it with FAAS
and GF AAS after wet digestion and no significant differences were observed
between the two methods.
A simple method combining slurry sampling after cryogenic grinding and
the use of a permanent modifier was proposed for the determination of Cd and
Pb in foods by GF AAS (56). The potentialities of cryogenic grinding were
evaluated for different materials that are difficult to homogenize, such as
high-fat and high-fiber tissues. Animal and vegetal samples were cut into
small pieces and ground in liquid nitrogen for 2 min. Slurries were prepared
directly in the autosampler cup by transferring an exact amount of ground
material (5 20 mg) to the cup, followed by 1.00 mL of 0.2% (v/v) HNO3
containing 0.04% (v/v) Triton X-100 and sonication for 30 s, before transfer-
ring onto the platform that has previously been coated with 250 mg W and
200 mg Rh. No statistical differences were found by the paired t-test at the
95% level between the results for Cd and Pb in foods slurries and those
obtained with digested samples.
Santos et al. (57) tested five different slurry preparation procedures for
fish tissue samples after grinding the solid samples to a particle size of
53 mm: (1) using aqua regia plus HF, 30 min of sonication, standing time of
24 h followed by another 30 min of sonication; (2) same as the previous
one, except that the standing time and the second ultrasound treatment were
omitted; (3) same as the previous one, except that HF was not used; (4)
same as the previous one, except that the aqua regia was replaced by nitric
acid; (5) same as the previous one, except that the nitric acid was replaced
by tetramethylammonium hydroxide (TMAH). The Hg vapor was generated
in a continuous-flow system and the emission signal intensity measured on-
line at 253.652 nm by axial view ICP OES. The first three procedures
produced results in agreement with the certified values. The two last pro-
cedures using nitric acid or TMHA could not be used for quantitative determi-
nation. For practical reasons, Procedure 3, with a detection limit (3 s, n 10)
of 0.06 microgram per gram for a sample mass of 20 mg in a final volume of
15 mL was recommended, because it was simple, rapid, and robust.
Bugallo et al. (58) developed a slurry sampling method for the determi-
nation of Ca, Cu, Fe, Mg, and Zn in fish tissue samples by FAAS. In compari-
son with microwave-assisted digestion, the proposed method was simple,
required only a short time, and eliminated total sample dissolution before
analysis. The suspension medium was optimized for each analyte to obtain
Downloaded By: [Korn, Maria das Graas Andrade] At: 13:30 6 February 2008
Direct solid sampling (SS) analysis is the oldest technique for the determi-
nation of metals by spectrometric techniques using arc or spark emission
and, together with X-ray fluorescence spectrometry, it is still the most
widely used technique in metallurgical laboratories nowadays. Among the
techniques that can be used for direct SS in combination with AAS, ICP
OES, and ICP-MS are laser ablation and electrothermal atomization or vapor-
ization. From these alternatives, GF AAS has been shown to be the most
attractive technique for the direct analysis of solid samples, mainly because
of the absence of a nebulizer system, which simplifies the introduction of
the solid material into the atomizer. Direct SS analysis offers a number of
advantages, such as the reduced sample preparation time and hence a faster
analysis; higher accuracy, as errors due to analyte loss and/or contamination
are dramatically reduced; higher sensitivity due to the absence of any dilution;
and the absence of any corrosive or toxic waste. Another advantage is the long
residence time of the sample in the GF AAS atomizer, which usually makes
possible complete volatilization of the particles independent of their size
and complete atomization of the analyte. Moreover, it shows quite low
limits of detection, which is highly desirable in trace analysis. Most of the dis-
advantages that have been mentioned for direct SS analysis using GF AAS are
actually no longer valid. There are reliable tools available nowadays both for
manual and automatic introduction of solid samples into the graphite furnace,
and it has been shown that in most cases aqueous standards can be used for
calibration also in direct SS analysis. The only limitations that have to be
mentioned are the relatively short linear working range of AAS, which
usually limits direct SS analysis to the determination of low trace concen-
trations, and the imprecision of the results, which is typically of the order
of 10% due to the inhomogeneity of natural samples (59).
Flores et al. (60) developed a new device to introduce solid biological test
samples directly into the flame of an AAS instead of the traditional
Downloaded By: [Korn, Maria das Graas Andrade] At: 13:30 6 February 2008
used a number of botanical CRM and preanalyzed samples of green coffee for
method validation, and ICP OES after MW-assisted acid digestion of the
samples was used as reference method. Manganese and cobalt could be deter-
mined using aqueous standards for calibration, but calibration with solid CRM
was necessary for the determination of copper. No significant difference was
found between the results obtained with the proposed method and certified or
independently determined values. Seven samples of Brazilian green coffee
were analyzed, and there was no significant difference between the values
obtained with SS-GF AAS and ICP OES for Mn and Cu. Hence, a single-
element technique that does not require any sample preparation besides
grinding of the coffee beans appears to be an attractive alternative to the
multi-element techniques that have been used up to now. The much better sen-
sitivity of this technique is an additional advantage in the determination of
trace elements such as cobalt and others that might be of importance.
Oliveira et al. (64) investigated systematically sample preparation and
micro-homogeneity for the determination of Cd and Pb in bovine liver using
direct SS-GF AAS. Two different procedures for sample preparation have been
investigated: (a) drying in a household microwave oven followed by drying in
a stove at 608C to constant mass and (b) freeze drying; ball and cryogenic
mills were used for grinding. Particle size, sample size, and microsample hom-
ogeneity were investigated. All samples showed good homogeneity (He , 10)
even for low sample mass, but samples dried in a microwave oven/stove and
ground in a ball mill presented the best homogeneity. The results obtained
with both methods of sample preparation indicated the possibility to produce
bovine liver of reference for determination of Cd and Pb by SS-GF AAS.
A very interesting series of studies about direct SS-GF AAS has been
carried out by Lucker et al. (65 70) between 1987 and 1999, investigating
the possibility of analyzing fresh meat for contaminants as kind of a
screening method to be carried out directly in the slaughterhouse. This idea
has been picked up recently by Damin et al. (71) in order to find out if this
technique could be used within the Brazilian program of residue control in
products of animal origin. The authors investigated the determination of Cd
and Pb in fresh meat, which was weighed directly onto the SS platform
using palladium and magnesium nitrates as a mixed modifier. The results
were in good agreement with those obtained after acid digestion, taking into
account the average humidity of 27 + 2% of fresh meat. Aqueous standards
could be used for calibration and the limits of detection of 0.13 mg kg21 for
Cd and 1.9 mg kg21 for Pb as well as the average RSD of 14% were more
than adequate for the purpose.
The recently introduced technique of high-resolution continuum source
(HR-CS) AAS (7274) appears to offer even greater advantages for direct
SS-GF AAS, as the entire spectral environment of the analytical line becomes
visible at high resolution. This feature makes it possible to detect and avoid
spectral interferences and the system also offers new possibilities to correct
for spectral interferences and is greatly facilitating method development (72).
Downloaded By: [Korn, Maria das Graas Andrade] At: 13:30 6 February 2008
Ribeiro et al. (75) investigated the determination of Co in fish and other biologi-
cal samples comparing direct SS-GF AAS and alkaline treatment with TMAH
and conventional line source GF AAS with HR-CS GF AAS. Method develop-
ment was found to be much easier using HR-CS AAS, and the best LOD of 5 ng
g21 was obtained with direct SS and HR-CS GF AAS.
Borges et al. (76) used HR-CS GF AAS for the determination of Pb in fish
and meat CRM using direct SS analysis. Ruthenium was used as a permanent
modifier, and the 217.001-nm resonance line was used for the determination
because of its better signal-to-noise ratio with HR-CS AAS. Under
optimized conditions the electron excitation spectrum of the PO molecule
with rotational fine structure could be separated in time from the Pb absorption
signal, avoiding any spectral interference. A limit of detection of 10 ng g21
could be obtained and the precision was typically better than 10% RSD.
The values obtained in seven CRM were in agreement with the certified
values according to the t-test for a 95% confidence level.
Da Silva et al. (77) investigated the determination of Hg in fish and meat
CRM using direct SS analysis with HR-CS GF AAS. Initial experiments
indicated that it was not possible to use a chemical modifier for this kind of
analysis, as in this case the Hg absorption peak would coincide with the
excessive background absorption caused by the organic matrix. It was also
found that without a modifier Hg from fish samples was already lost at temp-
eratures around 1008C, as it is mostly present as methyl mercury in this matrix,
which is much more volatile than inorganic Hg. The authors finally used a
temperature program without a pyrolysis stage, using only a drying stage of
3 s at 1008C, followed directly by the atomization stage at 11008C. Under
these conditions the Hg signal appeared before the background and could
be separated because of the superior background correction capabilities of
HR-CS AAS. Aqueous standards were used for calibration, which had to be
stabilized with potassium permanganate in order to avoid losses of Hg in
the drying stage. Good agreement was found between determined and
certified values for six CRM according to the t-test for a 95% confidence
level. The precision, expressed as RSD, was typically around 5% and the
detection limit was determined as 0.1 mg g21 Hg in the solid sample.
CONCLUSION
ACKNOWLEDGMENTS
REFERENCES
42. Araujo, G.C.L., Gonzalez, M.H., Ferreira, A.G., Nogueira, A.R.A., and
Nobrega, J.A. (2002) Effect of acid concentration on closed-vessel microwave-
assisted digestion of plant materials. Spectrochimica Acta, 57B (12): 2121 2132.
43. Korn, M., Santos Junior, D., and Krug, F.J. (2006) Currents on ultrasound-assisted
extraction for sample preparation and spectroscopic analytes determination.
Applied Spectroscopy Reviews, 41: 305 321.
44. Nascentes, C.C., Korn, M., and Arruda, M.A.Z. (2001) A fast ultrasound-assisted
extraction of Ca, Mg, Mn and Zn from vegetables. Microchemical Journal, 69:
37 43.
45. Manutsewee, N., Aeungmaitrepirom, W., Varanusupakul, P., and Imyim, A.
(2007) Determination of Cd, Cu, and Zn in fish and mussel by AAS after ultra-
sound-assisted acid leaching extraction. Food Chemistry, 101 (2): 817 824.
46. Melo, M.H.A., Costa, A.C.S., Nobrega, J.A., and Ferreira, S.L.C. (2005) The use
of water soluble tertiary amine reagent for solubilization and metal determination
in fish muscle tissue. Journal of the Brazilian Chemical Society, 16 (1): 69 73.
47. Maduro, C., Vale, G., Alves, S., Galesio, M., da Silva, M.D.R.G., Fernandez, C.,
Catarino, S., Rivas, M.G., Mota, A.M., and Capelo, J.L. (2006) Determination of
Cd and Pb in biological reference materials by electrothermal atomic absorption
spectrometry: A comparison of three ultrasonic-based sample treatment pro-
cedures. Talanta, 68 (4): 1156 1161.
48. Balarama Krishna, M.V. and Arunachalam, J. (2004) Ultrasound-assisted extrac-
tion procedure for the fast estimation of major, minor and trace elements in
lichen and mussel samples by ICP-MS and ICP-AES. Analytica Chimica Acta,
522 (2): 179 187.
49. Alvarez, J., Marco, L.M., Arroyo, J., Greaves, E.D., and Rivas, R. (2003) Determi-
nation of calcium, potassium, manganese, iron, copper and zinc levels in represen-
tative samples of two onion cultivars using total reflection X-ray fluorescence and
ultrasound extraction procedure. Spectrochimica Acta B, 58 (12): 2183 2189.
50. Li, P.C. and Jiang, S.J. (2003) Electrothermal vaporization inductively coupled
plasma-mass spectrometry for the determination of Cr, Cu, Cd, Hg and Pb in
rice flour. Analytica Chimica Acta, 495: 143 150.
51. Vinas, P., Martnez, M.P., and Cordoba, M.H. (2000) Rapid determination of
selenium, lead and cadmium in baby food samples using electrothermal atomic
absorption spectrometry and slurry atomization. Analytica Chimica Acta, 412:
121 130.
52. Silva, E.G.P., Santos, A.C.N., Costa, A.C.S., Fortunato, D.M.N., Jose, N.M.,
Korn, M.G.A., Santos, W.N.L., and Ferreira, S.L.C. (2006) Determination of
manganese and zinc in powdered chocolate samples by slurry sampling using
sequential multi-element flame atomic absorption spectrometry. Microchemical
Journal, 82 (2): 159 162.
53. Lopez-Garca, I., Vinas, P., Romero-Romero, R., and Hernandez-Cordoba, M.
(2007) Fast determination of phosphorus in honey, milk and infant formulas by
electrothermal atomic absorption spectrometry using a slurry sampling
procedure. Spectrochimica Acta B, 62: 48 55.
54. Cava-Montesinos, P., Cervera, M.L., Pastor, A., and de la Guardia, M. (2004)
Determination of As, Sb, Se, Te and Bi in milk by slurry sampling hydride gener-
ation atomic fluorescence spectrometry. Talanta, 62: 175 184.
55. Anthemidis, A.N. and Pliatsika, V.G. (2005) On-line slurry formation and nebuli-
zation for inductively coupled plasma atomic emission spectrometry. Multi-
element analysis of cocoa and coffee powder samples. Journal of Analytical
Atomic Spectrometry, 20: 1280 1286.
Downloaded By: [Korn, Maria das Graas Andrade] At: 13:30 6 February 2008
56. Santos, D. Jr., Barbosa, F. Jr., Tomazelli, A.C., Krug, F.J., Nobrega, J.A., and
Arruda, M.A.Z. (2002) Determination of Cd and Pb in food slurries by GFAAS
using cryogenic for sample preparation. Analytical and Bioanalytical Chemistry,
373: 183 189.
57. dos Santos, E.J., Herrmann, A.B., Vieira, M.A., Frescura, V.L.A., and Curtius, A.J.
(2005) Evaluation of slurry preparation procedures for the determination of
mercury by axial view inductively coupled plasma optical emission spectrometry
using on-line cold vapor generation. Spectrochimica Acta B, 60: 659 665.
58. Bugallo, R.A., Segade, S.R., and Gomez, E.F. (2007) Comparison of slurry
sampling and microwave-assisted digestion for calcium, magnesium, iron,
copper and zinc determination in fish tissue samples by flame atomic absorption
spectrometry. Talanta, 72: 60 65.
59. Vale, M.G.R., Oleszczuk, N., and Santos, W.N.L. (2006) Current status of direct
solid sampling for electrothermal atomic absorption spectrometryA critical
review of the development between 1995 and 2005. Applied Spectroscopy
Reviews, 41: 377 400.
60. Flores, E.M.M., Costa, A.B., Barin, J.S., Dressler, V.L., Paniz, J.N.G., and
Martins, A.F. (2001) Direct flame solid sampling for atomic absorption spec-
trometry: Determination of copper in bovine liver. Spectrochimica Acta B, 56:
1875 1882.
61. Detcheva, A. and Grobecker, K.H. (2006) Determination of Hg, Cd, Mn, Pb and Sn
in seafood by solid sampling Zeeman atomic absorption spectrometry. Spectrochi-
mica Acta B, 61: 454 459.
62. Grobecker, K.H. and Detcheva, A. (2006) Validation of mercury determination by
solid sampling Zeeman atomic absorption spectrometry and a specially designed
furnace. Talanta, 70: 962 965.
63. Oleszczuk, N., Castro, J.T., Silva, M.M., Korn, M.G.A., Welz, B., and
Vale, M.G.R. Method development for the determination of manganese, cobalt
and copper in green coffee comparing direct solid sampling electrothermal
atomic absorption spectrometry and inductively coupled plasma optical emission
spectrometry Talanta, 73: 862 869.
64. Nomura, C.S. and Oliveira, P.V. (2006) Preparacao de amostra de fgado bovino e
estudo de micro homogeneidade para a determinacao de Cd e Pb por espectrome-
tria de absorcao atomica com atomizacao eletrotermica e amostragem direta de
solido. Quimica Nova, 29 (2): 234 239.
65. Lucker, E., Rosopulo, A., and Kreuzer, W. (1987) Analyses of the distribution of
lead and cadmium in fresh renal tissue by means of solid sampling Zeeman-AAS.
Fresenius Zeitschrift fur Analytische Chemie, 328: 370 377.
66. Lucker, E., Gerbig, C., and Kreuzer, W. (1993) Distribution of Pb and Cd in the
liver of the mallardDirect determination by means of solid sampling ZAAS.
Fresenius Zeitschrift fur Analytische Chemie, 346: 1062 1067.
67. Lucker, E., Meuthen, J., and Kreuzer, W. (1993) Distribution of Pb and Cd in
equine liverDirect determination by means of solid sampling ZAAS.
Fresenius Zeitschrift fur Analytische Chemie, 346: 1068 1071.
68. Lucker, E. and Schuierer, O. (1996) Sources of error in direct solid sampling
Zeeman atomic absorption spectrometry analyses of biological samples with
high water content. Spectrochimica Acta B, 51: 201 210.
69. Lucker, E. (1999) Direct solid sampling ET AAS determination of cadmium in
equine muscle. Journal of Analytical Atomic Spectrometry, 14: 583 587.
Downloaded By: [Korn, Maria das Graas Andrade] At: 13:30 6 February 2008
70. Lucker, E. (1999) Direct solid sampling ET AAS determination of lead in muscle
tissue contaminated by gun-shot residues. Journal of Analytical Atomic Spec-
trometry, 14: 1731 1735.
71. Damin, I.C.F., Silva, M.M., Vale, M.G.R., and Welz, B. (2007) Feasibility of using
direct determination of cadmium and lead in fresh meat by electrothermal atomic
absorption spectrometry for screening purposes. Spectrochimica Acta B, 62:
1037 1045.
72. Welz, B., Becker-Ross, H., Florek, S., and Heitmann, U. (2005) High-Resolution
Continuum Source AAS; Wiley-VCH: Weinheim.
73. Welz, B. (2005) High-resolution continuum source AAS: the better way to perform
atomic absorption spectrometry. Analytical and Bioanalytical Chemistry, 381:
69 71.
74. Welz, B., Borges, D.L.G., Lepri, F.G., Vale, M.G.R., and Heitmann, U. (2007)
High-resolution continuum source atomic absorption spectrometryAn analytical
and diagnostic tool for trace analysis. Spectrochimica Acta B, 62: 873 883.
75. Ribeiro, A.S., Vieira, M.A., da Silva, A.F., Borges, D.L.G., Welz, B.,
Heitmann, U., and Curtius, A.J. (2005) Determination of cobalt in biological
samples by line-source and high-resolution continuum source graphite furnace
atomic absorption spectrometry using solid sampling or alkaline treatment. Spec-
trochimica Acta B, 60: 693 698.
76. Borges, D.L.G., da Silva, A.F., Welz, B., Curtius, A.J., and Heitmann, U. (2006)
Determination of lead in biological samples by high-resolution continuum
source graphite furnace atomic absorption spectrometry with direct solid
sampling. Journal of Analytical Atomic Spectrometry, 21: 763 769.
77. da Silva, A.F., Lepri, F.G., Borges, D.L.G., Welz, B., Curtius, A.J., and
Heitmann, U. (2006) Determination of mercury in biological samples using
solid sampling high-resolution continuum source electrothermal atomization
atomic absorption spectrometry with calibration against aqueous standards.
Journal of Analytical Atomic Spectrometry, 21: 1321 1326.