Professional Documents
Culture Documents
Incubator 1 PDF
Incubator 1 PDF
T
he following article describes ber 31487. This is a gravity convec-
the various steps regarding T his validation tion unit model number 6M with an
the Installation Qualification was designed operating temperature range of 25C
(IQ), Operational Qualification (OQ) to 70C. The incubator has the fol-
and Performance Qualification (PQ) and per for med lowing electrical characteristics:
that were executed to complete the fr om the
validation of a pharmaceutical micro- Volts- 120
biology laboratory incubator. This viewpoint of a Hertz- 50/60
validation was designed and per- micr obiologist Watts- 300
formed from the viewpoint of a
microbiologist with assistance from with assistance The overall dimensions of the
validation experts. The protocol used fr om validation unit are 42 X 21 X 39 with cham-
for this validation was previously ber dimensions 36 X 18 X 27.The
published in the Journal of Valida- exper ts. incubator contains a dual hydraulic
tion Technology.1 thermostat, a mercury-in-glass ther-
mometer, two shelves, two heaters, double inner
Descr iption of the I ncubator Unit doors, and double outer doors.
LC 7 20713 5219 In the diagrams seen for the top and lower shelf,
LC 8 20703 4521 each number corresponds to the thermocouple and
LC 9 11542 5679 the logger placed at the location. The two devices
LC 10 11537 4553 were side-by-side for each one of the locations indi-
Windows-based PC or laptop computer to oper- cated. The box in the lower shelf with an R in the
ate data loggers center indicates the location of the chart recorder, as
External Temperature Device shown in Figure 2.
Timer
Figure 2
E quipment Picture R elative P os ition of the
See Figure 1. T hermoc ouples and
Figure 1 Wireles s S ens ors
Top S helf
E mpty C hamber S et-Up
1 2
4 5
Front
L ower S helf
6 R 7
9 10
Front
E mpty
Figure 3 Figure 4
R es ults of L o w R es ults of High
Temperature S etting Temperature S etting
40.0 70.00
35.0 60.00
30.0 50.00
25.0 40.00
20.0 30.00
00.32 00:37 00:42
S etting
TC 1 Top Shelf Back Left TC 8 Bottom Shelf Center 20.00
TC 2 Top Shelf Back Right TC 9 Bottom Shelf Left Front
TC 3 Top Shelf Center TC 10 Bottom Shelf Front Right
TC 4 Top Shelf Front Left TC 13 Description
TC 5 Top Shelf Front Right TC 14 Description 10.00
TC 6 Bottom Shelf Back Left TC 15 Description
TC 7 Bottom Shelf Back Right TC 16 Description
Figure 5 Figure 6
R es ults of 24-Hour E mpty R es ults of 24-Hour E mpty
C hamber Mapping C hamber Mapping
T hermoc ouples Wireles s Trac ks ens e
40.0 40.0
30.0 30.0
20.0 20.0
10.0 10.0
0.0 0.0
02:40 07:25 12:10 16:55 21:40 08:43-1 13:28-1 18:13-1 22:58-1 03:43-2
S etting S etting
TC 1 Top Shelf Back Left TC 8 Bottom Shelf Center LC 1 Top Shelf Back Left LC 6 Bottom Shelf Back Left
TC 2 Top Shelf Back Right TC 9 Bottom Shelf Left Front LC 2 Top Shelf Back Right LC 7 Bottom Shelf Back Right
TC 3 Top Shelf Center TC 10 Bottom Shelf Front Right LC 3 Top Shelf Center LC 8 Bottom Shelf Center
TC 4 Top Shelf Front Left TC 13 Description LC 4 Top Shelf Front Left LC 9 Bottom Shelf Left Front
TC 5 Top Shelf Front Right TC 14 Description LC 5 Top Shelf Front Right LC 10 Bottom Shelf Front Right
TC 6 Bottom Shelf Back Left TC 15 Description
TC 7 Bottom Shelf Back Right TC 16 Description
temperature setting test was completed immediately
In order for this to be a 24-hour run, the readings prior to the empty chamber mapping. Future mapping
needed to last until 10:07:00 on day number two. The studies should allow for more time between tests
last reading on day number two was made at either by adding two extra hours, or by running the
08:41:00, which completes the 24-hour run, but does mapping study for 30 hours in order to compensate
not map the temperature within specified ranges for for the adjustment period. There are companies that
24 hours. This is one hour and 20 minutes short of the change temperature ranges of their incubators on a
requirement. This is due to the fact of not being able regular basis, due to the variety of tests conducted and
to see the data until the study is complete. Note that the lack of incubators. In this situation, the chamber
the wireless system is being used to determine its should be allowed several hours to equilibrate prior to
suitability for validating incubators by comparing it loading it. However, this is not the case of this partic-
to the real time system. This validation study is based ular piece of equipment. Therefore, this phenomenon
on the data obtained from traditional thermocouples. observed during the 24-hour empty chamber test
should not occur during day-to-day use.
Thermocouple Test: Started at 2:40:40 and ended
at 26:52:40. Tracksense: All wireless sensors were within
All TCs reached the target temperature range at specified temperature range at 9:43:00. During the
2:40:40. During the first hour and eight minutes, there first 24 minutes, three sensors reached 37.5C. This
were 10 temperature excursions: one on TC 5 of was not observed at any other point in time during
0.1C and the other seven on TC 8. Of the excursions the 24-hour study.
on TC 8, six were of 0.1C, and one was of 0.2C. During the review of TC temperatures, the excur-
This was not observed at any other point in time dur- sions stated above were not found. The highest tem-
ing the 24-hour study. This indicates that there were perature found during this review using the comput-
still a few hot spots in the chamber after the highest er screen was 37.1C
Laura Valdes-Mora & Pamela Hardt-English
All temperature devices were checked, and all were Acceptance Criterion
within the specified limit of + 2C from each other. All sites measured are within 32-37C
Note that the temperature loggers recorded different
actual clock times. The loggers were not synchronized PQ Observations
for this study, neither were they set to local time. Each A maximum load pattern was selected for the
one was set for a different time zone. Areal time elapsed PQ. In addition, the effect of opening doors for two
adjustment was made to account for the difference. different lengths of time, and the effect of a power
failure on the temperature profile were evaluated in
Performance Qualification: (PQ) this phase. Typically, these evaluations are con-
In this section it was verified that the loaded incu- ducted during the OQ, however, we chose the
bator chamber was capable of reaching and main- loaded chamber for these tests as we consider this
taining the target temperature range. scenario more realistic, therefore, more meaning-
ful.
Purpose
The PQ confirms that the loaded chamber works Loaded Chamber Description
within the specified limits of temperature through- Top Shelf:
out the chamber. The effect of opening doors for two 4* 2000 mL screw cap bottles of Lactose
different lengths of time, and the effect of a power Broth
failure on the incubators ability to get back into the 4 1000 mL screw cap bottles contain-
temperature profile were also evaluated. ing 500 mL each of Lactose Broth
1 Rack of 40 screw cap test tubes with
E quipment and Materials 10 mL each of Nutrient Broth
Precision Thelco Incubator (32-37C) 2 Baskets each containing 11 Whirlpak
Real Time Data Logger Ellab Model TM9616 bags each with 100 mL TSB**
Real Time Data Logger
Serial Number 00270110141 Lower Shelf:
Thermocouples Type T (Cu/Con) 1 Chart recorder
Wireless Data Loggers 1 Water bottle (used by laboratory for
Ellab Model TSM9803 temperature determinations)
Serial Number 2056 15 Stacks of plates- total of 72 plates- 42
Temperature Sensor Serial Number Blood Agar and 30 TSA***
LC 1 20022 6422 12 Whirlpak bags with 250 mL Lactose
LC 2 20753 5757 Broth each (in a basket)
LC 3 20770 5747 4 API strips
LC 4 20764 5770
LC 5 20804 5766 * Two (2) 2000 mL bottles were removed from
LC 6 20717 5309 the study. See observations for explanation.
LC 7 20713 5219 ** TSB= Tryptic Soy Broth
LC 8 20703 4521 *** TSA= Tryptic Soy Agar
LC 9 11542 5679
LC 10 11537 4553 Loading Pattern Diagrams: The purpose of the
diagrams in Figure 7 is to show the relative position
Windows-based PC or laptop computer to of the various items in the load. The diagrams are
operate data loggers. not to scale.
External Temperature Device (chart recorder) Numbers on each of the above rectangles denote
Timer number of plates on each stack. If the number is fol-
Agar plates and culture broths to create a maxi- lowed by mL, it denotes the volume of medium on
mum loaded chamber pattern. each container.
Laura Valdes-Mora & Pamela Hardt-English
Figure 7 Figure 8
R elative P os ition of the Various P ic ture of F ully L oaded C hamber
Items in the L oad
Top S helf
T1 Test Tube T2
1 1
2000 Rack
0 0
mL Basket
0 0
#1
0 0
mL mL
T3
1 1 Basket
0 0 #2
2000 0 0
mL F ull
0 0
T4 mL mL T5
the loaded chamber by two 2-liter flasks, based on
B ottom S helf previously performed work.
Following this action, it took the chamber two
T6 T7
R C 3 5 hours and 45 minutes to reach the set temperature.
A A
10 10 8 Basket Five-Minute Opened Door Test
3B 2B
A B A This incubator consists of two double doors.
4 2B During this test all doors were opened for five-min-
B T8 4 API utes. The purpose of the test is to simulate a worse case
4 3B B scenario when laboratory analysts are loading inocu-
B
6 4 lated samples and their controls into the incubator.
4
A B B During this test time, the temperature of the incu-
T9 T10 bator decreased. Temperature ranged from 22.9 to
R= Chart recorder T= Thermocouple number (a 30.2C across the loaded chamber as shown inFig-
A= TSA plates wireless sensor is next to ure 9. The chart recorder reading was 30.5C. It took
C= Water Control each thermocouple)
B= Blood Agar Plates 50.5 minutes for the loaded chamber to return to the
set temperature.
Loaded Chamber Picture
An example of a loaded chamber can be seen in Three-Hour Power Outage Test
Figure 8. The purpose of the test was to simulate the effect
of a power outage on the temperature profile of a
Maximum Load Pattern fully loaded incubator chamber. The TCs were
After loading the chamber and waiting one hour checked to ensure all sites were at the set temperature
and 45 minutes, it was observed that only 2 TCs were and proceeded to disconnect the incubator. After
within the target temperature range. We analyzed the three hours, the temperature across the chamber was
content of the load, and laboratory personnel re- reviewed and the incubator was reconnected to its
viewed the loading pattern. It was concluded that the power supply.
incubator was loaded above the worst-case scenario. There was a decrease in temperature that took the
Four 2-liter flasks are never placed into this incubator loaded chamber outside the set temperature and out-
by laboratory personnel. This was done only once side the target temperature. Temperatures ranged
during the validation of a microbial limit test, but as it from 27.7 to 29.4C as shown inFi gure 10. The chart
is not a regular procedure, it was chosen to decrease recorder was reading 28.7C.
Laura Valdes-Mora & Pamela Hardt-English
Figure 9 Figure 10
R es ults of F ive-Minute R es ults of T hree-Hour
Opened Door Tes t Power Outage Tes t
70.00 70.00
60.00 60.00
50.00 50.00
40.00 40.00
30.00 30.00
20.00 20.00
10.00 10.00
0.00 0.00
02:34 02:44 02:54 03:04 03:14 03:24 03:34 03:36 04:31 05:26 06:21 07:16
S etting S etting
TC 1 Top Back Left TC 6 Bottom Back Left
TA 1 Top Back Left TA 5 Top Front Right TC 2 Top Back Right TC 7 Bottom Back Right
TA 2 Top Back Right TA 6 Bottom Back Left TC 3 Top Center TC 8 Bottom Center
TA 3 Top Center TA 7 Bottom Back Right TC 4 Top Front Left TC 9 Bottom Front Left
TA 4 Top Front Left TA 8 Bottom Center TC 5 Top Front Right TC 10 Bottom Front Right
Following the power start-up, it took one hour of the double doors was opened, not both, for the
and 35.5 minutes for the temperature to reach the set same reason of mimicking actual use.
point of 35-37C. This unexpectedly long ramp time During the one-minute opened door test, the tem-
signals that the load also dropped in temperature, perature decreased and was outside the set tempera-
thus requiring a longer time to return to the set tem- ture in some areas of the chamber. The temperatures
perature. ranged from 29.3 to 35.2C as shown inFi gure 11.
One-Minute Opened Door Test: The chart recorder was reading 33.6C. It took 15 min-
The protocol was originally written for a five- utes for the chamber to return to the set temperature
minute and a 15-minute opened door test. After per- once the door was closed.
forming the five-minute opened door test, it was
determined that most of the time the incubator doors 24-Hour Loaded Chamber Temperature Mapping
do not remain open longer than just a few minutes, The loaded chamber test was conducted for 24
typically not even reaching the five-minute mark. hours. Temperature readings were recorded every
Based on this, it was decided to change the 15- two minutes. The study started at 08:32:00 and ended
minute test to a one-minute test in order to represent at 32:16:00. This is 16 seconds short of precisely 24-
a true scenario. In addition, for this test, only one side hours, however, due to rounding rules, this small
Laura Valdes-Mora & Pamela Hardt-English
Figure 11 Figure 12
R es ults of One-Minute R es ults of 24-Hour L oaded
Open Door Tes t C hamber Temperature Mapping
T hermoc ouples
40.0
45.0
30.0
40.0
20.0
35.0
10.0
30.0
0.0
08:32 13:17 18:02 22:47 27:32
25.0 S etting
TC 1 Top Back Left TC 6 Bottom Back Left
TC 2 Top Back Right TC 7 Bottom Back Right
20.0 TC 3 Top Center TC 8 Bottom Center
TC 4 Top Front Left TC 9 Bottom Left Front
07:58 08:08 08:18 08:28 TC 5 Top Front Right TC 10 Bottom Front Right
S etting
The lowest temperature recorded was 34.3C found on
TC 1 Top Back Left TC 6 Bottom Back Left
TC 2 Top Back Right TC 7 Bottom Back Right LC 8, while the highest was 37.0C found on LC 6.
TC 3 Top Center TC 8 Bottom Center All temperature devices were checked and all
TC 4 Top Front Left TC 9 Bottom Front Left
TC 5 Top Front Right TC 10 Bottom Front Right
were within the specified limit of + 2C from each
other.
shortage is considered acceptable as this is a 24-hour Note that the temperature loggers recorded differ-
run, not a 24.0-hour run. ent actual clock times. The loggers were not syn-
All thermocouples were within temperature range chronized for this study, neither were they set to local
at the start of the mapping, as shown in Fi gure 12. time. Each one was set for a different time zone.
No excursions were found at any point in time,
thus corroborating that the excursions seen during C ompar isons B etween Data C ollected by
the empty chamber test were due to hot pockets left T her mocouples and by W ir eless L ogger s
from the highest temperature setting test.
The loaded chamber maintained a tight temperature The thermocouples used are type T, copper-con-
range throughout the 24-hour mapping study on a per stantan thermocouples. The quality of wire and con-
TC basis. The lowest temperature recorded was struction can significantly affect the accuracy of the
34.5C found on TC 8, while the highest was 36.9C thermocouples. The thermocouples used in these
found on TC 9. tests are Ellabs sealed thermocouples with a stain-
less steel tip. The Ellab Tracksense logger sensor is
Tracksense: All loggers were within temperature a thermister made of a Silicon chip. Each logger and
range at the start of the mapping as shown in Fi gure sensor has a separate ID number, and is factory cal-
13. The study started at 09:50:00, and ended at ibrated at five points in the range of -40 to 140C.
33:52:00, for a total of 24 hours and two seconds. The accuracy of the Ellab thermocouples and the
There were no excursions found at any point in time. Ellab Tracksense sensors is +/-0.2C.
Laura Valdes-Mora & Pamela Hardt-English
Figure 14
Data C ollection C omparis on B etween Thermocouples and Wireles s Dataloge rs
A ll Values are in Degrees C entigrade
Type of Data F ile T ime Mi n i mu m Ma x i mu m Av e r age Av e r age Delta T Av e r age
Te s t N a m e /S y s te m I n te rva l for all for all of Minimu m of Maximum ( R a n ge of all
(10 Mi n i mu m Ma x i mu m Va l u e s Va l u e s f ro m Va l u e s
H o u rs ) Va l u e s Va l u e s Av e r age )
Empty Eval1.dt 6:00:00- 36.3 37.4 36.5 37.1 0.3 36.8
Chamber 4/EVAL 16:00:00
Track1.dt 6:00:00- 36.3 37.4 36.7 37.1 0.2 36.8
4/Tracksense 16:00:00
Loaded Eval 22.dt 12:00:00- 35.5 36.9 35.9 36.3 0.3 36.1
Chamber 4 22:00:00
Track2.dt 12:00:00- 35.4 36.8 35.9 36.3 0.5 36.2
4/Tracksense
Laura Valdes-Mora & Pamela Hardt-English
the original protocol were justified and approved. mal laboratory practices are acceptable, and that
There was no reference available for the Kaye extending the incubation time for the same length of
dry bath. Equipment was sent out to the manufactur- time of the power failure is an acceptable solution.
er for routine calibration. The internal bath reference Note that time should be calculated from the time
was used for all calibrations of the dry bath per- the incubator reaches the temperature range (see rec-
formed for this validation. ommendations).
The protocol indicated that the chamber should be
allowed to be at temperature range for at least 30 C onclusion
minutes prior to conducting the five-minute opened
door test. Based on OQ data, it was shown that the The incubator was tested and verified to operate
incubator is capable of maintaining the temperature properly according to the manufacturer and protocol
within the range for extended periods; therefore wait- requirements. This unit is therefore considered vali-
ing 30 minutes would not have made a difference. dated.
A 15-minute Opened Door Test was described on
step 15 of the PQ protocol. This was changed to a one- Recommendations
minute Opened Door Test, as typically incubator doors In order to have enough data collected, one should
are opened for not more than a couple of minutes. The consider running the temperature mappings with a
new test length was considered to be reflective of a safety factor of six hours, that is, for a 24-hour run,
real situation. conduct a 30-hour mapping study, especially if wire-
The final post calibration of the thermocouples had less sensors will be used.
to be repeated because there was not enough data col- The validation presented here was completed in
lected to confirm the high temperature calibration at three days of continuous work. Scheduling and exe-
40C.The EVAL Datalogger collects and displays cuting the work in this manner is taxing and compli-
data every two seconds, but the program was set to cated. At times, one will be working very early or
save data every two minutes, and the system was shut very late hours. Careful pre-scheduling is highly rec-
down before the final data were stored. Therefore, the ommended. Consider adding extra time to the sched-
calibration was repeated at Star Services, Hayward, ule for unforeseen situations (example: chamber may
California, using a Kaye Liquid bath at two tempera- not get to temperature in the time you have predict-
tures, 30 and 40C to confirm that the thermocouples ed).
were within specifications. Older incubators that do not have a digital temper-
ature display can benefit from having one installed.
C onclusions Digital temperature displays are available; they have
IQ a Real Time Datalogger (RTD) which is placed in the
It is concluded that the unit was installed proper- incubator either through a port (if available) or
ly. All necessary components were found to be pre- through the top vent.
sent, and all the required SOPs were found to be in Microbiologists should consider that after open-
place. The utilities (power) supplied to the unit, as ing the incubator doors, the equipment loses heat,
per testing, were found to meet the requirements. causing a decrease in temperature, which typically
falls outside the teststemperature range. Based on
OQ this, incubation times should be recorded from the
All test results met the protocol acceptance crite- time the incubator is at temperature range, not from
ria. The unit is capable of maintaining the target the time samples are placed in the incubator.
temperature. Caution must be exercised when reading temper-
ature from an external chart recorder. Data from this
PQ validation clearly showed that chart recorders are
All test results met the protocol acceptance crite- slow in capturing accurate temperature.
ria. Based on data obtained from the opened door
tests and power failure, it can be concluded that nor-
Laura Valdes-Mora & Pamela Hardt-English