CERTIFICATE OF ACCREDITATION

ANSI-ASQ National Accreditation Board


500 Montgomery Street, Suite 625, Alexandria, VA 22314, 877-344-3044
This is to certify that
ARREGA TECNOLOGIA, SAPI, DE CV/ARREGA INDUSTRIAL
Blvd. Tercera Oeste No. 17524, Fracc. Garita de Otay
Tijuana, Baja California Mexico
has been assessed by ANAB
and meets the requirements of international standard
ISO/IEC 17025:2005
and national standard
ANSI/NCSL Z540-1-1994
while demonstrating technical competence in the fields of
CALIBRATION AND TESTING
Refer to the accompanying Scope of Accreditation for information regarding the types of
calibrations and/or tests to which this accreditation applies.

ACT-2077
Certificate Number

Certificate Valid: 09/07/2016-09/07/2018


Version No. 001 Issued: 09/07/2016

This laboratory is accredited in accordance with the recognized International Standard ISO/IEC 17025:2005.
This accreditation demonstrates technical competence for a defined scope and the operation of a laboratory
quality management system (refer to joint ISO-ILAC-IAF Communiqu dated January 2009).
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005
& ANSI/NCSL Z540-1-1994

ARREGA TECNOLOGA, SAPI, DE CV/ ARREGA INDUSTRIAL


Blvd. Tercera Oeste No. 17524, Fracc. Garita de Otay CP 22430
Tijuana, Baja California, Mxico
Gilberto Escandn Phone: +52 664-608-9263 / 619-391-7925
gtescandon@arregaindustrial.com arregaindustrial.com

CALIBRATION and TESTING

Valid to: September 7, 2018 Certificate Number: ACT-2077

I. Electromagnetic - DC/Low Frequency


Calibration and
Measurement
Parameter/ Reference Standard
Range Capability Method(s)
Equipment or Equipment
[Expressed as
Uncertainty()]
(3.3 to 330) mV 5.4 V/V + 0.19 V
330 mV to 3.3 V 9.1 V/V + 0.4 V OEM, FLUKE
DC Voltage - Source (3.3 to 33) V 10 V/V + 3.9 V Fluke 5522A MET/CAL and
(30 to 330) V 15 V/V + 29 V Internal Procedures
100 V to 1.1 kV 15 V/V + 0.3 mV
(10 to 100) mV 2.8 V/V + 0.08 V
(0.1 to 1) V 2.0 V/V + 0.08 V
(1 to 10) V 2.0 V/V + 0.13 V Agilent-3458A
(10 to 100) V 2.5 V/V + 7.5 V Opt. 002
100 V to 1 kV 2.5 V/V + 25 V
OEM, FLUKE
DC Voltage - Measure MET/CAL And
(Up to 5) kV 0.11 mV/V + 0.3 V
Internal Procedures
(5 to 10) kV 0.11 mV/V + 0.3 V
(10 to 20) kV 0.11 mV/V + 0.3 V Vitrek 4700
(20 to 30) KV 0.11 mV/V + 1.6 V
(30 to 40) kV 0.13 mV/V + 3.5 V
(40 to 50) kV 0.13 mV/V + 6.1 V
(0 to 330 mA) 29 A/A + 4 nA
(0.33 to 3.3) mA 22 A/A + 10 nA
OEM, FLUKE
(3.3 to 33) mA 34 A/A + 49 nA
DC Current - Source MET/CAL and
(33 to 330) mA 33 A/A + 0.49 A Fluke 5522A
Internal Procedures
330 mA to 1.1 A 54 A/A + 8 A
(1.1 to 3) A 81 A/A + 8 A
Version 001 Issued: 09/07/2016 Page 1 of 15

500 Montgomery St. Suite 625 Alexandria, VA 22314703-836-0025 www.anab.org


Calibration and
Measurement
Parameter/ Reference Standard
Range Capability Method(s)
Equipment or Equipment
[Expressed as
Uncertainty()]
(1.1 to 3) A 81 A/A + 8 A
(3 to 11) A 0.16 mA/A + 0.10 mA Fluke 5522A
OEM, FLUKE
DC Current Source (11 to 20.5) A 0.40 mA/A + 0.15 mA
MET/CAL and
(cont.)
Internal Procedures
(10 to 550) A 2.0 mA/A + 0.13 A Fluke
(550 to 1025) A 2.0 mA/A + 0.13 A 50 Turn Coil
Up to 100 A 20 A/A + 0.13 nA
(0.1 to 1) mA 17 A/A + 0.83 nA
(1 to 10) mA 17 A/A + 8.3 A Agilent-3458A OEM, FLUKE
DC Current - Measure (10 to 100) mA 24 A/A + 0.08 A Opt. 002 MET/CAL and
(0.1 to 1) A 54 A/A + 1.7 A Internal Procedures

(Up to 1 000) A 6.0 mA/A + 0.06 A Empro B-1000-100


2 0.27 m
11 0.44 m
12 1.8 m
19 0.49 m
30 0.64 m
33 0.57 m
110 1.2 m
120 1.3 m
190 1.7 m
300 2.4 m
330 2.7 m
1.1 k 7.7 m
1.2 k 12 m
1.9 k 16 m
Resistance - Source 3.0 k 24 m OEM, FLUKE
Fluke 5522A
(Fixed Values) 3.3 k 38 m MET/CAL and
11 k 85 m Internal Procedures
12 k 120 m
19 k 180 m
30 k 260 m
33 k 270 m
110 k 0.79
120 k 1.3
190 k 1.9
300 k 2.9
330 k 3
1.1 M 12
1.2 M 22
1.90 M 38
3 M 55
Version 001 Issued: 09/07/2016 Page 2 of 15

500 Montgomery St. Suite 625 Alexandria, VA 22314703-836-0025 www.anab.org


Calibration and
Measurement
Parameter/ Reference Standard
Range Capability Method(s)
Equipment or Equipment
[Expressed as
Uncertainty()]
3.3 M 61
11 M 0.3 k
12 M 1.1 k
19 M 1.6 k
30 M 2.3 k
Resistance - Source 33 M 4.0 k OEM, FLUKE
Fluke 5522A
(Fixed Values) (cont.) 110 M 25 k MET/CAL and
120 M 92 k Internal Procedures
290 M 0.2 M
400 M 1.3 M
640 M 2.0 M
1.1 G 3.5 M
Up to 10 9.6 / + 9.7
(10 to 100) 9.2 / + 97.0
(0.1 to 10) k 8.0 / + 97.0
OEM, FLUKE
(1 to 10) k 8.0 / + 0.97 m Agilent-3458A
Resistance - Measure MET/CAL and
(10 to 100) k 8.0 / + 9.7 m Opt. 002
Internal Procedures
(0.1 to 1) M 8.7 / + 0.39
(1 to 10) M 18 / + 0.19 k
(10 to 100) M 0.27 m/ + 0.19 k
Up to 33 mV
(10 to 45) Hz 180 V/V + 1.2 V
45 Hz to 10 kHz 91 V/V + 1.2 V
(10 to 20) kHz 91 V/V + 1.2 V
(20 to 50) kHz 0.23 mV/V + 1.2 V
(50 to 100) kHz 0.7 mV/V + 2.3 V
(100 to 500) kHz 1.8 mV/V + 9.7 V
(33 to 330) mV
(10 to 45) Hz 66 V/V + 1.6 V
45 Hz to 10 kHz 38 V/V + 1.6 V OEM, FLUKE
AC Voltage - Source (10 to 20) kHz 45 V/V + 1.6 V Fluke 5522A MET/CAL and
(20 to 50) kHz 80 V/V + 1.6 V Internal Procedures
(50 to 100) kHz 170 V/V + 6.2 V
(100 to 500) kHz 700 V/V + 14 V
(0.33 to 3.3) V
(10 to 45) Hz 66 V/V + 9.7 V
45 Hz to 10 kHz 37 V/V + 12 V
(10 to 20) kHz 44 V/V + 12 V
(20 to 50) kHz 71 V/V + 9.7 V
(50 to 100) kHz 1500 V/V + 24 V
(100 to 500) kHz 0.6 mV/V + 120 V

Version 001 Issued: 09/07/2016 Page 3 of 15

500 Montgomery St. Suite 625 Alexandria, VA 22314703-836-0025 www.anab.org


Calibration and
Measurement
Parameter/ Reference Standard
Range Capability Method(s)
Equipment or Equipment
[Expressed as
Uncertainty()]
(3.3 to 33) V
(10 to 45) Hz 65 V/V + 0.13 mV
45 Hz to 10 kHz 38 V/V + 0.12 mV
(10 to 20) kHz 56 V/V + 0.12 mV
(20 to 50) kHz 92 V/V + 0.12 mV
(50 to 100) kHz 200 V/V + 0.31 mV
(33 to 330) V
OEM, FLUKE
AC Voltage Source 45 Hz to 1 kHz 46 V/V + 0.39 mV
Fluke 5522A MET/CAL and
(cont.) (1 to 10) kHz 52 V/V + 1.2 mV
Internal Procedures
(10 to 20) kHz 93 V/V + 1.2 mV
(20 to 50) kHz 150 V/V + 1.2 mV
(50 to 100) kHz 540 V/V + 9.7 mV
(330 to 1.02) kV
45 Hz to 1 kHz 69 V/V + 1.9 mV
(1 to 5) kHz 64 V/V + 1.9 mV
(5 to 10) kHz 78 V/V + 1.9 mV
Up to 10 mV
40 Hz to 1 kHz 290 V/V + 0.3 V
(1 to 20) kHz 310 V/V + 0.3 V
(20 to 100) kHz 2.0 mV/V + 0.3 V
(100 to 300) kHz 12 mV/V + 0.5 V
(10 to 100) mV
40 Hz to 1 kHz 61 V/V + 0.5 V
(1 to 20) kHz 88 V/V + 0.5 V
(20 to 100) kHz 430 V/V + 0.5 V
(100 to 300) kHz 1.0 mV/V + 2.5 V
(0.1 to 1) V
40 Hz to 1 kHz 54 V/V + 5.0 V
OEM, FLUKE
(1 to 20) kHz 79 V/V + 5.0 V
AC Voltage Measure Agilent-3458A
MET/CAL and
(20 to 50) kHz 150 V/V + 5.0 V
Internal Procedures
(50 to 100) kHz 300 V/V + 5.0 V Opt. 002
(100 to 300) kHz 1.0 mV/V + 25.0 V
(300 to 500) kHz 3.0 mV/V + 25.0 V
(1 to 10) V
(1 to 40) Hz 56 V/V + 100 V
40 Hz to 1 kHz 53 V/V + 50 V
(1 to 20) kHz 78 V/V + 50 V
(20 to 50) kHz 150 V/V + 50 V
(50 to 100) kHz 260 V/V + 15 V
(100 to 300) kHz 1.0 mV/V + 0.25 mV/V
(300 to 500) kHz 2.0 mV/V + 0.25 mV/V
500 kHz to 1 MHz 3.0 mV/V + 0.25 mV/V

Version 001 Issued: 09/07/2016 Page 4 of 15

500 Montgomery St. Suite 625 Alexandria, VA 22314703-836-0025 www.anab.org


Calibration and
Measurement
Parameter/ Reference Standard
Range Capability Method(s)
Equipment or Equipment
[Expressed as
Uncertainty()]
(10 to 100) V
40 Hz to 1 kHz 100 V/V + 0.5 mV/V
(1 to 20) kHz 130 V/V + 0.5 mV/V
(20 to 50) kHz 170 V/V + 0.5 mV/V
0.46 mV/V + 0.5 mV/V Agilent-3458A
(50 to 100) kHz OEM, FLUKE
AC Voltage Measure Opt. 002
(100 to 1 000) V MET/CAL and
(cont.) 0.25 mV/V + 0.5 V
(50 to 60) Hz Internal Procedures
(100 to 1 000) V
(100 to 450) Hz 6.3 mV/V + 0.6 V
(1 to 9) kV
(50 to 60) Hz 0.38 mV/V + 0.5 V Vitrek 4700
(33 to 330) A
(10 to 20) Hz 0.40 mA/A + 19 nA
(20 to 45) Hz 0.30 mA/A + 19 nA
45 Hz to 1 kHz 0.26 mA/A + 19 nA
(1 to 5) kHz 0.59 mA/A + 0.029 A
(5 to 10) kHz 1.6 mA/A + 0.039 A
(10 to 30) kHz 3.1 mA/A + 0.078 A
(0.33 to 3.3) mA
(10 to 20) Hz 0.40 mA/A + 0.029 A
(20 to 45) Hz 0.25 mA/A + 0.029 A
45 Hz to 1 kHz 0.21 mA/A + 0.029 A
(1 to 5) kHz 0.40 mA/A + 0.039 A
(5 to 10) kHz 0.98 mA/A + 0.058 A
(10 to 30) kHz 2.0 mA/A + 0.12 A OEM, FLUKE
AC Current - Source (3.3 to 33) mA Fluke 5522A MET/CAL and
(10 to 20) Hz 0.38 mA/A + 0.39 A Internal Procedures
(20 to 45) Hz 0.20 mA/A + 0. 0.39A
45 Hz to 1 kHz 0.19 mA/A + 0. 0.39A
(1 to 5) kHz 0.23 mA/A + 0. 0.39A
(5 to 10) kHz 0.50 mA/A + 0.58 A
(10 to 30) kHz 0.81 mA/A + 0.78 A
(33 to 330) mA
(10 to 20) Hz 0.38 mA/A + 3.9 A
(20 to 45) Hz 0.20 mA/A + 3.9 A
45 Hz to 1 kHz 0.11 mA/A + 3.9 A
(1 to 5) kHz 0.22 mA/A + 10 A
(5 to 10) kHz 0.41 mA/A + 19 A
(10 to 30) kHz 0.81 mA/A + 39 A

Version 001 Issued: 09/07/2016 Page 5 of 15

500 Montgomery St. Suite 625 Alexandria, VA 22314703-836-0025 www.anab.org


Calibration and
Measurement
Parameter/ Reference Standard
Range Capability Method(s)
Equipment or Equipment
[Expressed as
Uncertainty()]
(0.33 to 1.1) A
(10 to 45) Hz 0.37 mA/A + 19 A
45 Hz to 1 kHz 0.16 mA/A + 19 A
(1 to 5) kHz 1.4 mA/A + 190 A
(5 to 10) kHz 4.9 mA/A + 0.97 mA
(1.1 to 3) A
(10 to 45) Hz 0.36 mA/A + 19 A
45 Hz to 1 kHz 0.16 mA/A + 19 A
(1 to 5) kHz 1.3 mA/A + 0.19 mA
(5 to 10) kHz 5.0 mA/A + 0.97 mA
(3 to 11) A
(45 to 100) Hz 0.20 mA/A + 0.39 mA
100 Hz to 1 kHz 0.25 mA/A + 0.39 mA
(1 to 5) kHz 2.3 mA/A + 0.39 mA
OEM, FLUKE
AC Current Source (11 to 20.5) A
MET/CAL and
(Cont.) (45 to 100) Hz 0.3 mA/A + 0.97 mA
Fluke 5522A Internal Procedures
100 Hz to 1 kHz 0.4 mA/A + 0.97 mA
(1 to 5) kHz 5.9 mA/A + 0.97 mA
(16.5 to 55) A
(45 to 65) Hz 2.1 mA/A + 0.13 mA
(65 to 440 Hz) 2.1 mA/A + 0.13 mA
(55 to 150) A
(45 to 65) Hz 2.1 mA/A + 0.13 mA
(65 to 440 Hz) 2.1 mA/A + 0.13 mA
(150 to 550) A
(45 to 65) Hz 1.9 mA/A + 0.13 mA
(65 to 440 Hz) 2.1 mA/A + 0.13 mA
(550 to 1 000) A
(45 to 65) Hz 1.9 mA/A + 0.14 mA Fluke 50 Turn Coil
(65 to 440 Hz) 1.9 mA/A + 0.14 mA
Up to 10 A
1 kHz 0.78 mA/A + 1 nA
(10 to 100) A
1 kHz 0.35 mA/A + 6 nA
(0.1 to 1) mA
1 kHz 0.14 mA/A + 0.04 A
Agilent-3458A OEM, FLUKE
AC Current Measure (1 to 10) mA
Opt. 002 MET/CAL and
1 kHz 0.20 mA/A + 0.4 A
Internal Procedures
(10 to 100) mA
1kHz 0.20 mA/A + 4 A
(0.1 to 1) A
1 kHz 0.29 mA/A + 40 A
Up to 1 000 A Empro B1000-100
(60 to 100) Hz 6.1 mA/A + 0.6 A
Version 001 Issued: 09/07/2016 Page 6 of 15

500 Montgomery St. Suite 625 Alexandria, VA 22314703-836-0025 www.anab.org


Calibration and
Measurement
Parameter/ Reference Standard
Range Capability Method(s)
Equipment or Equipment
[Expressed as
Uncertainty()]
Electrical Simulation
of Thermo-Couple
Indicators (600 to 800) C 0.54 C
(800 to 1 000) C 0.49 C
Type B (1 000 to 1 550) C 0.38 C
(1 550 to 18 20) C 0.39 C

(0 to 150) C 0.38 C
Type C
(150 to 650) C 0.25 C
(650 to 1 000) C 0.28 C
(1 000 to 1 800) C 0.44 C
(1 800 to 2 316) C 0.70 C

Type E (-250 to -100) C 0.43 C


(-100 to -25) C 0.19 C
(-25 to 350) C 0.18 C
(350 to 650) C 0.19 C
(650 to 1 000) C 0.22 C

Type J OEM, FLUKE


(-210 to -100) C 0.26 C
Fluke 5522A MET/CAL and
(-100 to -30) C 0.19 C
Internal Procedures
(-30 to 150) C 0.18 C
(150 to 760) C 0.20 C
(760 to 1 200) C 0.24 C

Type K (-200 to -100) C 0.31 C


(-100 to -25) C 0.22 C
(-25 to 120) C 0.21 C
(120 to 1 000) C 0.27 C
(1 000 to 1 372) C 0.36 C

Type L (-200 to -100) C 0.33 C


(-100 to 800) C 0.25 C
(800 to 900) C 0.20 C

Type N (-200 to -100) C 0.41 C


(-100 to -25) C 0.31 C
(-25 to 120) C 0.30 C
(120 to 410) C 0.29 C
(410 to 1 300) C 0.34 C

Version 001 Issued: 09/07/2016 Page 7 of 15

500 Montgomery St. Suite 625 Alexandria, VA 22314703-836-0025 www.anab.org


Calibration and
Measurement
Parameter/ Reference Standard
Range Capability Method(s)
Equipment or Equipment
[Expressed as
Uncertainty()]

Type R (0 to 250) C 0.62 C


(250 to 400) C 0.37 C
(400 to 1 000) C 0.36 C
(1 000 to 1767) C 0.41 C

Type S (0 to 250) C 0.45 C


(250 to 1 000) C 0.38 C
OEM, FLUKE
(1 000 to 1400) C 0.39 C Fluke 5522A
MET/CAL and
(1 400 to 1767) C 0.44 C
Internal Procedures
(-250 to -150) C 0.54 C
Type T
(-150 to 0) C 0.27 C
(0 to 120) C 0.17 C
(120 to 400) C 0.16 C

Type U (-200 to 0) C 0.49 C


(0 to 600) C 0.24 C
Electrical Simulation
of RTDs
Pt 385, 100 (-200 to -80) C 0.08 C
(-80 to 0) C 0.08 C
(0 to 100) C 0.09 C
(100 to 300) C 0.11 C
(300 to 400) C 0.11 C
(400 to 630) C 0.13 C
(630 to 800) C 0.21 C
OEM, FLUKE
Pt 385, 200 Fluke 5522 A MET/CAL and
(-200 to -80) C 0.08 C
Internal Procedures
(-80 to 0) C 0.08 C
(0 to 100) C 0.08 C
(100 to 260) C 0.08 C
(260 to 300) C 0.13 C
(300 to 400) C 0.14 C
(400 to 600) C 0.16 C
(600 to 630) C 0.17 C

Version 001 Issued: 09/07/2016 Page 8 of 15

500 Montgomery St. Suite 625 Alexandria, VA 22314703-836-0025 www.anab.org


Calibration and
Measurement
Parameter/ Reference Standard
Range Capability Method(s)
Equipment or Equipment
[Expressed as
Uncertainty()]
Electrical Simulation
of RTDs (-200 to -80) C 0.08 C
(cont.) (-80 to 100) C 0.08 C
Pt 385, 500 (0 to 100) C 0.09 C
(100 to 260) C 0.09 C
(260 to 300) C 0.10 C
(300 to 400) C 0.10 C
(400 to 600) C 0.11 C
(600 to 630) C 0.12 C

Pt 385, 1 000 (-200 to -80) C 0.08 C


(-80 to 100) C 0.08 C OEM, FLUKE
(0 to 100) C 0.08 C Fluke 5522 A MET/CAL and
(100 to 260) C 0.09 C Internal Procedures
(260 to 300) C 0.09 C
(300 to 400) C 0.10 C
(400 to 600) C 0.10 C
(600 to 630) C 0.20 C

Pt 3926, 100 (-200 to -80) C 0.08 C


(-80 to 0) C 0.08 C
(0 to 100) C 0.09 C
(100 to 300) C 0.11 C
(300 to 400) C 0.11 C
(400 to 630) C 0.13 C
Capacitance
Source
10 Hz to 10 kHz (220 to 400) pF 14.8 mF/F + 1.9 pF
10 Hz to 10 kHz 400 pF to 1.1 nF 4.5 mF/F + 1.9 pF
10 Hz to 3 kHz (1.1 to 3.3) nF 2.4 mF/F + 1.9 pF
10 Hz to 1 kHz (3.3 to 11) nF 0.90 mF/F + 1.9 pF
OEM, FLUKE
10 Hz to 1 kHz (11 to 33) nF 0.87 mF/F + 1.9 pF
Fluke 5522A MET/CAL and
10 Hz to 1 kHz (33 to 110) nF 0.88 mF/F + 1.9 pF
Internal Procedures
10 Hz to 1 kHz (110 to 330) nF 0.90 mF/F + 5.8 pF
(10 to 600) Hz 330 nF to 1.1 F 0.9 mF/F + 0.2 pF
(10 to 300) Hz (1.1 to 3.3) F 0.9 mF/F + 0.6 pF
(10 to 150) Hz (3.3 to 11) F 0.9 mF/F + 1.9 pF
(10 to 120) Hz (11 to 33) F 1.1 mF/F + 5.8 pF
(10 to 80) Hz
(33 to 110) F 1.6 mF/F + 19.4 pF

Version 001 Issued: 09/07/2016 Page 9 of 15

500 Montgomery St. Suite 625 Alexandria, VA 22314703-836-0025 www.anab.org


Calibration and
Measurement
Parameter/ Reference Standard
Range Capability Method(s)
Equipment or Equipment
[Expressed as
Uncertainty()]
Capacitance (10 to 100) pF 3.2 mH/H + 0.020 pF
Measure (10 to 100) pF 3.2 mH/H + 0.02 pF OEM, FLUKE
(100 to 1000) pF 2.6 mH/H + 0.020 nF Fluke PM6303 MET/CAL and
(1 to 100) nF 2.6 mH/H + 0.20 pF Internal Procedures
1 kHz to 1 mF 2.7 mH/H + 2.04 nF
100 H 385 nH Genrad 1482B OEM, FLUKE
Inductance - Source
1 mH 2.3 H Genrad 1482E MET/CAL and
100 Hz to 10 kHz
100 mH 104 H Genrad 1482L Internal Procedures
DC Power Source Up to 3060 W 310 W/W + 0.03 W
(3.06 to 20.9) kW 0.89 mW/W + 0.62 W
OEM, FLUKE
Up to 336.6 W 0.14 mW/W + 4 mW
AC Power Source Fluke 5522 A MET/CAL and
(0.336 to 1.12) kW 0.18 mW/W + 0.02 W
(45 to 65 Hz) Internal Procedures
(1.12 to 3.06) kW 0.18 mW/W + 0.02 W
(3.06 to 11.22) kW 0.21 mW/W + 0.41 W
(11.2 to 20.91) kW 0.31 mW/W + 1.02 W
(65 to 500) Hz 1.5
(500 to 1) kHz 1.5 OEM, FLUKE
Fluke 5522A
Phase (1 to 5) kHz 1.6 MET/CAL and
(5 to 10) kHz 1.9 Internal Procedures
(10 to 30) kHz 2.7
Oscilloscopes
1 kHz
Amplitude Square Up to 5 mV 1.8 mV/V + 6.2 V
Wave (50 ) (5 to 11) mV 2.0 mV/V + 14 V
(11 to 45) mV 2.0 mV/V + 56 V
(45 to 110) mV 2.0 mV/V + 140 V
(0.11 to 0.45) V 2.0 mV/V + 0.56 mV
(0.45 to 1.09) V 2.0 mV/V + 1.4 mV
Leveled Sine Wave 10 Hz
(1.09 to 2.2) V 2.0 mV/V + 5.5 mV OEM, FLUKE
100 Hz Fluke 5502A MET/CAL and
(1.09 to 2.2) V 2.0 mV/V + 2.8 mV Internal Procedures
1 kHz
(1.09 to 2.2) V 2.0 mV/V + 2.8 mV
Amplitude Square 1 kHz
Wave (1 M ) Up to 5 mV 2.2 mV/V + 0.01 V
(5 to 20) mV 2 mV/V + 0.03 V
(20 to 89) mV 2 mV/V + 110 mV
(89 to 219) mV 2 mV/V + 0.27 mV
(219 to 890) mV 2 mV/V + 1.1 mV
(0.89 to 6.5) V 2 mV/V + 8 mV
Version 001 Issued: 09/07/2016 Page 10 of 15

500 Montgomery St. Suite 625 Alexandria, VA 22314703-836-0025 www.anab.org


Calibration and
Measurement
Parameter/ Reference Standard
Range Capability Method(s)
Equipment or Equipment
[Expressed as
Uncertainty()]
Oscilloscopes (cont.)
10 Hz
Amplitude Square (6.5 to 55) V 2.0 mV/V + 140 mV
Wave (1 M) 100 Hz
(6.5 to 55) V 2.0 mV/V + 69 mV
1 kHz
(6.5 to 55) V 2.0 mV/V + 69 mV
OEM, FLUKE
10 kHz
Fluke 5502A MET/CAL and
(6.5 to 55) V 4.0 mV/V + 69 mV
Internal Procedures

Leveled Sine Wave [10 mV to 5.5 V]


50 kHz to 100 MHz 28 mV/V + 58 mV/V
100 kHz to 300 MHz 33 mV/V + 58 mV/V

Time Marker into 50 2 ns to 50 ms 4.9 s/s


50 ms to 5 s 14 s/s

II. Time & Frequency


Calibration and
Measurement
Parameter/ Reference Standard
Range Capability Method(s)
Equipment or Equipment
[Expressed as
Uncertainty()]
OEM, FLUKE
Stop-Watch (1 to 86 400) s 0.07 s Agilent 53132A MET/CAL and
Internal Procedures

1 MHz to 225 MHz (6.4 10-05) Hz + 2 R OEM, FLUKE


Frequency Measure -3 Agilent 53132A MET/CAL and
DC to 3 GHz (6.2 10 ) Hz + 2 R
Internal Procedures
50 kHz @ 5.5 V 0.21 Hz
500 kHz @ 5.5 V 0.32 Hz OEM, FLUKE
Frequency - Source 5 MHz @ 5.5 V 3.1 Hz Fluke 5502A MET/CAL and
50 MHz @ 5.5 V 31 Hz Internal Procedures
300 MHz @ 2 Vpp 190 Hz

Version 001 Issued: 09/07/2016 Page 11 of 15

500 Montgomery St. Suite 625 Alexandria, VA 22314703-836-0025 www.anab.org


III. Thermodynamic
Calibration and
Measurement
Parameter/ Reference Standard
Range Capability Method(s)
Equipment or Equipment
[Expressed as
Uncertainty()]
OEM, FLUKE
(0 to 100) % RH 1.2 % RH Vaisala MET/CAL and
Humidity
MI70/HMP76B Internal Procedures

35 C 0.65 C
100 C 0.67 C OEM, FLUKE
Temperature Source Fluke 4181/ IR MET/CAL and
200 C 0.72 C
(Black Body) Calibrator Internal Procedures
350 C 0.84 C
500 C 1.0 C
-200 to 0 C 0.03 C Fluke 1523
0.03 C OEM, FLUKE
0 to 100 C W/ 5628 PRT
Temperature MET/CAL and
100 to 300 C 0.05 C Fluke 9144 Field
Measurement-Source Internal Procedures
300 to 420 C 0.11 C Metrology Well
429 to 650 C 0.12 C

IV. Mechanical
Calibration and
Measurement
Parameter/ Reference Standard
Range Capability Method(s)
Equipment or Equipment
[Expressed as
Uncertainty()]
5 to 50 in-oz 0.13 % of reading
5 to 50 in-lb 0.27 % of reading OEM, FLUKE
Torque - Measure 40 to 400 in-lb 0.27 % of reading CDI Torque Transducers MET/CAL and
100 to 1000 in-lb 0.26 % of reading
25 to 250 lb-ft 0.15 % of reading Internal Procedures
100 to 1 000 lb-ft 0.15 % of reading
Force - Measure OEM, FLUKE
(100 to 500) lb 0.05 % of reading FUTEK,
Tension and (200 to 1 000) lb 0.37 % of reading MET/CAL and
Compression (1 000 to 5 000) lb 0.28 % of reading Load Cell Internal Procedures
Up to 50 g 0.04 mg + 0.6 R OEM, FLUKE
Scales / Balances (50 to 200) g 2.3 mg + 0.6 R
Class F Weights MET/CAL and
200 g to 10 kg 8.50 mg + 0.6 R
7 600 mg+ 0.6 R Internal Procedures
(25 to 500) kg
(-12 to 300) psi 0.06 psi OEM, FLUKE
Pressure - Measure Fluke 3130 / Fluke 750
(-1 to 1) psig 0.000 2 psig MET/CAL and
R8 / Fluke 750PD2
(Up to 1 000) psig 0.44 psig Internal Procedures

Version 001 Issued: 09/07/2016 Page 12 of 15

500 Montgomery St. Suite 625 Alexandria, VA 22314703-836-0025 www.anab.org


Calibration and
Measurement
Parameter/ Reference Standard
Range Capability Method(s)
Equipment or Equipment
[Expressed as
Uncertainty()]
1 mg 0.17 mg
2 mg 0.18 mg
5 mg 0.23 mg
10 mg 0.27 mg
20 mg 0.33 mg
50 mg 0.43 mg
100 mg 0.52 mg
200 mg 0.64 mg
500 mg 0.84 mg
1g 1.1 mg
2g 1.3 mg
5g 1.8 mg
10 g 2.3 mg OEM, Sourced
20 g 4.7 mg Rice Lake Class F Procedures
Mass 50 g 12 mg NIST 44-15
Weights and Balances
100 g 23 mg Handbook
200 g 46 mg Up to Class F only
500 g 81 mg
1 kg 120 mg
2 kg 230 mg
5 kg 580 mg
25 kg 2.9 g

1/32 oz 1.0 mg
1/16 oz 1.3 mg
1/8 oz 1.5 mg
1/4 oz 2.0 mg
1/2 oz 3.3 mg
1 oz 6.3 mg
2 oz 13 mg
4 oz 27 mg
OEM, Sourced
8 oz 52 mg
Procedures
Mass 1 lb 81 mg Rice Lake Class F
NIST 44-15
(Continue) 2 lb 110 mg Weights and Balances
Handbook
5 lb 270 mg
Up to Class F only
10 lb 520 mg
20 lb 60 mg

Version 001 Issued: 09/07/2016 Page 13 of 15

500 Montgomery St. Suite 625 Alexandria, VA 22314703-836-0025 www.anab.org


Calibration and
Measurement
Parameter/ Reference Standard
Range Capability Method(s)
Equipment or Equipment
[Expressed as
Uncertainty()]
22.9 HRC 0.43 HRC
33.1 HRC 0.42 HRC
43.6 HRC 0.42 HRC
54.4 HRC 0.42 HRC
60.7 HRC 0.36 HRC OEM and ASTM
Block Hardness
Hardness 62.9 HRC 0.36 HRC E18 Sourced
Standards
43.6 HR30N 0.38 HR30N Procedures
53.7 HR30N 0.38 HR30N
62.6 HR30N 0.36 HR30N
76.1 HR30N 0.38 HR30N
79.3 HR30N 0.37 HR30N

V. Dimensional
Calibration and
Measurement
Parameter/ Reference Standard
Range Capability Method(s)
Equipment or Equipment
[Expressed as
Uncertainty()]
Up to 12 in (290 + 2.9 L) in
Calipers (12 to 40) in (200 + 9.6 L) in
(40 to 80) in (110 + 12 L) in
OD Micrometers Up to 12 in (35 + 9.75 L) in
Up to 12 in (35 + 9.75 L) in
ID Micrometers Grade 0 Gage Blocks
(12 to 40) in (5.4 + 12.79 L) in
and Bars
Dial Indicators Up to 4 in (27 + 7.4 L) in
OEM and Internal
Test Indicators Up to 0.05 in (29 + 0.12 L) in Procedures.
Up to 12 in (110 + 6 L) in
Height Gages
(12 to 40) in (43 + 12 L) in
Optical Comparator/ (0 to 50) mm (2.5 + 0.004 L) m 172-116 / 172-117 Glass
Microscope (0 to 50.8) mm (1.7 + 0.023 L) m Scale
Protractor/Angle (30 to 90) 1.8 arc min Angle Block Set

Outside Diameter (1 to 60) mm (1.0 + 0.0043 D) m Micrometer Laser

Version 001 Issued: 09/07/2016 Page 14 of 15

500 Montgomery St. Suite 625 Alexandria, VA 22314703-836-0025 www.anab.org


VI. Dimensional Inspection
Calibration and
Measurement
Parameter/ Reference Standard
Range Capability Method(s)
Equipment or Equipment
[Expressed as
Uncertainty()]
Dimensional Inspection

Linear (0 to 1 500) mm 41 m Articulated Arm Customer


Coordinate Measuring Drawings
Volumetric Machines (AACMM)
Up 1.8 m 52 m
Performance Faro
Notes:
1. Calibration and Measurement Capabilities (CMC) (Expanded Uncertainties) are based on approximately a 95% confidence interval, using a
coverage of k=2.
2. This laboratory offers calibration services at on-site at customer-designated locations. Since on-site conditions are typically more variable than
those in the laboratory, larger measurement uncertainties are expected on-site than what is reported on the accredited scope.
3. On-site calibration is offered for all parameters.
4. The use of (R) stands for resolution.
5. The use of (L) stands for Length in inches.
6. The use of (t) is an expression of Time in seconds.
7. This scope is formatted as part of a single document including the Certificate of Accreditation No. ACT-2077.

Version 001 Issued: 09/07/2016 Page 15 of 15