You are on page 1of 2

Vaisala Dew point GE Panametrics VeriDri

Transmitters
Polymer sensors Aluminium Oxide

Automatic Pressure compensation No automatic Pressure compensation


to give the most accurate reading which makes it non suitable for direct
and direct mounting in process, line mounting
hence no sampling system

Sensors have linear response Sensors have non linear response

Have a Patented Autocalibration sometimes require individual


Feature calibration curves

-recommended calibration after - need frequent calibration after every


2yrs 6- 12 months

-excellent stability - poor stability

-negligible hysteresis - considerable hysteresis

- no effect of ageing- - suffers from ageing: dynamic


dynamic range is stable range decreases

-no effect of contaminants - suffers from contamination


,purging option to remove
Contaminants
- hefty programming process requiring
-programmable on site to be exported to Shanghai,Tokyo

COMPARISON OF DEW POINT Transmitters