J 375 to 750C 0.4% of reading 333 to 750C 0.75% of reading
-40 to 375C +/-1.5C -40 to +333C +/- 2.5C
K-N 375 to 1200C 0.4% of reading 333 to 1200C 0.75% of reading
-40 to 125C +/-0.5C -40 to +133C +/- 1C
T 125 to 350C 0.4% of reading 133 to 350C 0.75% of reading
-40 to 375C +/-1.5C -40 to +333C +/- 2.5C
E 375 to 800C 0.4% of reading 333 to 800C 0.75% of reading
0 to 1100C +/-1.C -0 to 600C +/- 1.5C
R-S 1100 to 1600C +/-[1+0.003(t-1100)] 333 to 800C 0.25% of reading
Not applicable -0 to 600C EMK to small
B 600 to 1700C 0.25% of reading
Limits of Error ASTM E230-ANSI MC96.1
TC type Special limits Tolerance Standard limits Tolerance
0 to 750C +/-1.1C-or 0.4% of 0-750 +/- 2.2C or 0.75%
reading of reading J -200 to 0C +/- 2.2C or 2% of reading
0 to 750C +/-1.1C-or 0.4% of 0-750 +/- 2.2C or 0.75%
K reading of reading
0 to 350C +/-0.5C-or 0.4% of 0 to 350C +/- 1C or 0.75%
reading of reading T +/- 1.C or 1.5% of -200 to 0C reading
0 to 900C +/-1C-or 0.4% of 0 to 900C +/- 1.7C or 0.5%
reading of reading E +/- 1.7C or 1 % of -200 to 0C reading
All specifications are subject to change without notice 1
Rodax Antwerp Belgium F-D1992JAN14 E-mail: customersupport@rodax-europe.com 0 to 900C +/-1C-or 0.4% of 0 to 900C +/- 1.7C or 0.5% reading of reading N +/- 1.7C or 1 % of -200 to 0C reading
0 to 1450C +/-0.6 or 0.1% of 0 to 1450C +/-1.5 or 0.25%
R-S reading of reading
Not applicable -0 to 800C EMK to small
B 800 to 1700C +/-0.5%
All specifications are subject to change without notice 2