Professional Documents
Culture Documents
Multicomponent UV Analysis
Multicomponent spectral analysis is described in stan-
dard analytical texts (9, 10) and is a common experiment in
the undergraduate curriculum (11). In addition to accessing
the software in the Hewlett-Packard 8450A Spectrophotom-
eter, University of Florida students are required to analyze
the data manually using simultaneous equations, as
described in the references. Because the number of equations
must equal or exceed the number of components, manual
data reduction can be unwieldy for a three-component
system. To simplify the math, the analysis is limited to two
of the compounds, caffeine and benzoic acid.
* Corresponding author.
JChemEd.chem.wisc.edu Vol. 75 No. 5 May 1998 Journal of Chemical Education 1
In the La bor a tory
Also, benzoic acids long retention time is Figure 3. Liquid chromatogram of a caffeine / benzoic acid / aspar-
indicative of considerable interaction of the tame mixture obtained on a 15 cm 4.5 mm Higgins Analytical
column packed with 5 m Hiasil C18, using a flow rate of 1.0
phenyl group with the stationary phase, and
mL/ min and a mobile phase composition of 45% methanol / 55%
this ef- fect may also occur with aspartame. 0.025 M aqueous phosphate, pH 3.0. Other instrument compo-
The CE experiment is performed on a nents included a TSP P200 pump, a Valco injector with a 20-L
Hewlett-Packard 3DCapillary Electrophoresis loop, a TSP UV 100 detector set to 218 nm, and a SpectraPhysics
system, which has an automatic sample SP4270 integrator.
changer and a diode-array detector, using an
applied potential of 20 kV with the cathode at
the outlet. With this configuration, the
migration order is cations first in order of
decreasing electrophoretic mobility (ep), Figure 4. Electropherogram of a caffeine / benzoic acid /
neutrals next as a group, and anions last in aspartame mixture obtained on a Hewlett-Packard 3DCE system
order of increasing ep. To separate the three using a 0.025 M borate buffer, pH 9.4. The 33 cm 50 m
components, a 0.025 M borate buffer, pH 9.4, capillary was oper- ated at 20 kV. The detection wavelengths were
is used. At this high pH only Caf is neutral; 272 nm for caf- feine, 229 nm for benzoate, and 210 nm for
aspartame and ben- zoate are both anionic aspartame. Before the laboratory period, the capillary was flushed
with 0.10 M NaOH, water, and the borate buffer. Buffer flushes were
included after every third sample.
dents expect Caf to migrate fastest, followed by Asp, with Table 1 . Spectral Data for Caffeine, Benzoic Acid,
Benz last. As shown in Figure 4, this order is indeed and Aspartame
observed. a
Compound (n (L/ g (103L/ mol
In addition to explaining the pH effects and predicting m cm)
cm)
the elution/migration orders, students are asked to use the Caffeine 2 51.1 9.93
UV spectra to choose wavelengths for the LC and CE 1
2 25.9 5.02
analyses during the prelab quiz. For the LC analysis, the 2 48.9 9.50
7
chosen wave- length is 218 nm. As Figure 1 shows, Benzoic Acid 2 52.8 6.45
although the compo- nents all absorb appreciably at 218 1
2 90.1 11.00
nm, this wavelength does not correspond to max for any of 2 7.66 0.935
the three compounds. The diode-array capability of the CE 7
Aspartame 2 29.3 8.61
system allows an optimum wavelength to be used for each 1
2 11.09 3.26
component (210 for Asp, 272 for Caf, and 229 for Benz). 1
In the laboratory, students analyze the same solutions
on both instruments. First, individual solutions of each method (sample pretreatment, injection, peak quantitation).
com- ponent in water are injected to determine the elution This is considered to be a truer representation of the detec-
and migration times. On the CE, students also verify the tion limit than the value obtained according to IUPAC rec-
peak assignments from the spectra obtained by the diode- ommendations (17, 18), which utilizes only the
array detector. They next inject a series of mixed standards, fluctuations in the instrument signal. The full EPA protocol
each containing known concentrations of all three is slightly modified (see Appendix) to meet the time
components, for preparation of the three calibration plots. constraints of the student laboratory.
The standards, which are prepared by the instructor before
the laboratory period, have concentrations in the ranges Results
40200 mg/L for caffeine, 25150 mg/L for benzoic acid,
and 75600 mg/L for aspartame in water. The remaining The wavelengths chosen for the UV analysis are the
samples are unknowns: filtered soft drinks, including Mello maxs for the two components: 229 nm (benzoic acid) and
Yello and several diet drinks, a synthetic unknown, and a 272 nm (caffeine). Both compounds obey Beers law over
solution of one packet of Equal (aspartame) in 100 mL of the con- centration ranges used in the analysis. Table 1
water. As described fur- ther below, students determine the summarizes the absorptivity data, including values at 218
method detection limits for the three compounds. To nm and 210 nm (aspartame only), which are used in the
obtain the necessary data, they prepare seven replicate HPLC and CE experiments.
dilutions of one of the standards and obtain the peak areas Typical analytical results are presented in Table 2. Results
using the same separation conditions as the drink samples. for the synthetic unknown show excellent agreement with
Data analysis for both experiments starts with preparation the instructors values, although the caffeine result for Mello
of six calibration plots of peak area versus concentration for Yello is somewhat high. As stated above, this is probably
each component by each method. The least-squares due to a small absorbance for Mello Yello at 300 nm, where
equations are used to evaluate the concentrations of the the individual components do not absorb. This
components in the unknowns and the aspartame content of a undoubtedly ex- tends into the UV and accounts for the
packet of Equal. high caffeine result. In the conclusion section of their report,
To emphasize the relationship of the detection systems students are expected
to the UV spectrophotometer, students are also asked to
cal- culate the ratios (Caf:Asp and HBenz:Asp)
of the slopes of the LC calibration plots.
They observe that these ratios are close to EPA method takes into consideration the statis- tical fluctuations of
the corresponding ratios of the absorptivi- the entire analytical
ties at the LC wavelength (218 nm)
obtained from the UV spectra. The CE
report in- cludes comparisons of the LC
and CE re- sults for the same samples and
results for Mello Yello by all three
methods.
Analysis of the same compounds by
more than one method provides an excel-
lent opportunity to evaluate detection lim-
its and compare the values for different in-
struments. For this exercise, students are
asked to determine the method detection
limit according to the method specified by
the Environmental Protection Agency (15,
16 ) for caffeine, aspartame, and benzoic
acid by both LC and CE. As explained
further in the Appendix to this paper, the
Table 2 . Results of Caffeine / Benzoic Acid / Asp 156.9 1 153
Aspartame Analyses 7
Diet Coke Caf 127 1 133
Valuea Experimental Benz 3
1 149
Sample Compou Result
nd (mg /L) UVb Asp 521 5 496
0
UV Syn Unk Caf 11.12 11.09 / Diet Pepsi Caf 101 1 89
Benz 4.15 11.38/ 4.24
4.15 Benz 0
1 162
Mello Yello Caf 148 167 / 176 Asp 6
4 475
Benz 208 / 219 d 6
Equal Asp 367 3 317
LC / CE Syn Caf 101.8 a
Instructors value for synthetic unknowns; manufacturers value 5for commercial prod-
Unk Benz 78.6 ucts, if available. bResults by HP software / simultaneous equations. cAverage of 2 deter-
d
minations. Solution prepared by dissolving 1 packet in 100 mL of water.
to comment on how well the Mello Yello system meets the reinforce students knowledge of acid/base behavior and stress the
requirements for multicomponent analysis, and the extra importance of fundamental chemistry in the design of an analytical
ab- sorption is an obvious point to address. Thus, this method.
analytical interference is actually very useful instructionally.
There is also some concern about possible interactions
between the two components, because the solubility of
caffeine in pharmaceutical preparations is known to increase
in the presence of sodium benzoate (19). However, the
additivity test described above produces spectra that match
almost perfectly, and a published UV analysis for
pharmaceutical mixtures of caf- feine, sodium benzoate,
phenacetin, and Pyramidon showed good agreement, with
relative errors of less than 3% (14 ).
The LC and CE calibration plots are also very linear,
and the results for the synthetic unknown are generally
quite good, with relative errors less than 10% (less than
4%, if aspartame by LC and caffeine by CE are excluded).
For the most part, LC and CE analyses of the drinks agree
with each other, the major exception being caffeine in Mello
Yello, which produces a significantly higher value by CE. Of
the Mello Yello analyses, the LC result is closest to the
manufacturers value. The high CE result may be due to
another neutral co- migrating with the caffeine, but the
spectral profile shows no obvious deviation from that of
pure caffeine.
The LC and CE values for caffeine in Diet Coke may
also be compared to those reported previously in this Journal:
92 mg/L (3) and 134 mg/L (5 ) by LC; 124 (7 ) by CE.
Agree- ment with the latter two values is very satisfactory.
The MDLs (mg/L, M) by LC are 1.2, 6.1 for
caffeine; 0.57, 4.7 for benzoic acid; and 1.4, 4.9 for
aspartame. The
MDLs by CE are 1.7, 8.5 for caffeine; 0.29, 2.4 for benzoic
acid; and 2.2, 7.5 for aspartame. The LC and CE values are
quite close to each other, and all six values are within 2 and
8.5 on a micromolar basis. The two caffeine MDLs are also
remarkably close to the previously reported value of 1.9 mg/L
by CE (7 ).
Conclusions
In the conclusion to the CE report (i.e., after all three
experiments have been performed), students are asked to
com- pare the advantages and limitations of the three
methods. Factors such as the number of analyzable
components, MDLs, sample size, ease of performance, and
accuracy (compared to manufacturers values) are discussed.
The most notable consideration is the number of
components, which is clearly a limitation in UV
multicomponent analysis. Good students note that sample
size is another important factor. Although the MDLs are
about equal for LC versus CE, the 20-L injection loop on
the LC can be rinsed and filled with about 50 L of filtered
drink, whereas the standard CE autosampler vials hold
about 500 L. Thus, even though most of the sample can be
recovered from the autosampler vial, the CE analy- sis requires
a total sample volume about ten times that for LC.
The analysis of soft drink components stimulates student
interest and is instructionally useful. The experiments dem-
onstrate the analytical use of the three instruments, as well
as their advantages and limitations. The MDL determinations
teach an important procedure frequently used in the work-
place but not included in most texts. The experiments also
Acknowledgment 10. Harris, D. C. Quantitative Chemical Analysis, 4th ed.; Freeman:
New York, 1995; p 526.
Purchase of the capillary electrophoresis 11. Williams, K. R.; Cole, S. R.; Boyette, S. E.; Schulman, S. G. J.
system was facilitated by grant #DUE Chem. Educ. 1990, 67, 535.
9650497 from the National Science 12. Martell, A. E.; Smith, R. M. Critical Stability Constants;
Foundation. Plenum: New York, 1974; Vol. 3, p 16.
13. Martell, A. E.; Smith, R. M. Critical Stability Constants;
Note Plenum: New York, 1974; Vol 5, p 111.
1. Presented at the annual meeting of the 14. The Merck Index, 11th ed; Budavari, S., Ed.; Merck: Rahway, NJ,
Southeast Association of Analytical Chemists, Athens, 1989; p 248.
GA, October 1996, and the Annual Meeting of the 15. U.S. Environmental Protection Agency. In Code of Federal Regu-
Florida Sections of the American Chemical Society, lations; Part 136, Title 40, Appendix B, Revision 1.11, U.S.
Orlando, FL, May 1997. Gov- ernment Printing Office: Washington, DC, 1990; pp 537
539.
Literature Cited 16. Harris, D. C. Quantitative Chemical Analysis, 4th ed.; Freeman:
New York, 1995; p 84.
1. Smyly, D. S.; Woodward, B. B.; Conrad, E. C. J.A.O.A.C. 1976, 17. Winefordner, J. D.; Long, G. L. Anal. Chem. 1983, 55, 712A
5 724A.
9 18. Skoog, D. A.; Leary, J. J. Principles of Instrumental Analysis, 4th
, ed.; Saunders: Fort Worth, 1992; pp 78.
19. Machek, G.; Lorenz, F. Scientia Pharmaceutica 1966, 34, 213231.
1
4 Appendix
1 The EPA defines the MDL as the minimum
9 concentrationthat can be measured and reported with 99% con-
. fidence that the analyte concentration is greater than zero and is
2. Gillyon, E. C. P. Chromatogr. Newslett. 1980, 8, 5051. determined from an analysis of a sample in a given matrix
3. Delaney, M. F.; Pasko, K. M.; Mauro, D. M.; contain- ing the analyte (15 ). The latter part of the definition
Gsell, D. S.; Korologos, P. C.; Morawski, J.;
indicates that the MDL must be determined by an actual analysis
Krolikowski, L. J.; Warren, F. V., Jr. J. Chem.
Educ. 1985, 62, 618620. and is strictly valid only for the particular sample conditions. The
4. DiNunzio, J. E. J. Chem. Educ. 1985, 62, 446447. code also gives the complete protocol, but, for the readers
5. Strohl, A. N. J. Chem. Educ. 1985, 62, 447448. convenience, a brief explanation is included here.
6. Mabrouk, P. A.; Marzilli, L. A. Presented at the First, the analyst must have an estimate of the MDL. The
212th National ACS Meeting, Orlando, August code lists several types of estimates, but the two most applicable to
1996; Paper No. 276; see CHED Newslett. 1996, these analyses are (i) the concentration giving a signal roughly 2.5
Fall.
to 5 times the baseline noise (method usually used by the
7. Conte, E. D.; Barry, E. F.; Rubinstein, H. J. Chem. Educ. 1996,
73, 11691170. students), and
8. Schuster, R.; Gratzfeld-Hsgen, A. Hewlett- (ii) the lowest concentration in the linear response range (i.e., the
Packard Publ. #12- 5963-1122E; Hewlett- concentration at which the calibration plot shows a noticeable
Packard: Palo Alto, 1994. change in slope). Next, a standard containing 1 to 5 times the
9. Skoog, D. A.; Leary, J. J. Principles of approxi-
Instrumental Analysis, 4th ed.; Saunders: Fort
Worth, 1992; p 164.
mate MDL is prepared, and seven or more aliquots are analyzed The MDL is given by: MDL = t99 s
by the usual laboratory procedure. Because of the time required
to dissolve the solid compounds, the procedure is modified to
have
students prepare and inject seven replicate dilutions of one of the where s is the standard deviation of the replicate concentration mea-
mixed calibration standards. surements and t99 is Students t (one-sided) at the 99% confidence
The analyte concentration is evaluated from the least-squares level (98% confidence level if a table of two-sided ts is used) for
response equation (and suitable additional calculations, if applicable). N
1 degrees of freedom (N = the number of replicates).