PERHITUNGAN
2. Daerah Plato
V1 = 760 volt
V2 = 800 volt
N1 = 22,85 cps
22,85 60
= x
1 1
= 1371 cpm
N2 = 23,32 cps
23,32 60
= x
1 1
= 1399 cpm
2 1
= (2 1 )1
100%
1399 1371
= (800 760 )1371 100%
= 0,051 % / volt
6.2 Penentuan kestabilan alat
Cacah background rata-rata = 0,64 cps
Cacah Cbackground N rata-rata N-Nrata- (N-Nrata-
(cps) (cps) N (cps) (cps) rata (cps) rata)2
22,44
22.91667 0,64 22.27666667 -0.16333 0.026678
22.33333 21.69333333 22,44 -0.74667 0.557511
0,64
23.83333 23.19333333 22,44 0.753333 0.567511
0,64
23.08333 22.44333333 22,44 0.003333 1.11E-05
0,64
23.41667 22.77666667 22,44 0.336667 0.113344
0,64
23.3 22.66 22,44 0.22 0.0484
0,64
22.73333 22.09333333 22,44 -0.34667 0.120178
0,64
22.96667 22.32666667 22,44 -0.11333 0.012844
0,64
23.73333 23.09333333 22,44 0.653333 0.426844
0,64
22.51667 21.87666667 22,44 -0.56333 0.317344
0,64
)2
( = 2,190667
( )2
X2 =
2,190667
= 22,44
= 0.097623292 cps
Untuk pengujian dengan melakukan 10 kali pengukuran berulang (N=10),
sistem pencacah masih dapat dikatakan stabil bila nilai chi squarenya berkisar
antara 3,33-16,9 atau 3,33< X2<16,9
6.3 Penentuan Waktu Mati detektor
N1 N1,2 Nbackground
No. (cps) N2 (cps) (cps) (cps)
1 23,30 41,46 64,19 0,64
1 + 2 12
=
12 2 1 2 2 2
(23,30+41,4664,190,64)
= (64,19^223,30^241,46^2)2
= 23,28 cps
Jadi, Nsebenarnya adalah 23,28 cps.
Persentase kesalahan = x 100%
23,2823,30
= X 100%
23,28
= 0,08591 %
6.4 Penentuan efisiensi detektor
Sumber standar = Ti-204
Ao = 1 x 10-6 Ci
1
t2 = 3,78 tahun
0,693
= 3,78
= 0,183/tahun
At = 0
= 1 x 10-6 Ci( 0,1836 )
= 5 x 10-6 Ci ( 1,098 )
= 0,33 x 10-6 Ci
= 12210 dps
Nbackground :
Nbackground (cps)
0,64
0,64
N1 (cps)
600
600
= 600 dps 0,64 dps
= 599,36 dps
Efisiensi () = 100%
599,36
= x 100%
12210
= 4,91 %
At Ti-204 = 0
= 0,33 Ci
N Ti-204 rata-rata = 600 dps
0,676 0,64
= 0,33
600 0,64
0,036
= 599,36 0,33
= 1,98211 x 10^-5
= 0,7333807 Bq