78
CARDIAC PACING
JACC Vol, 14, No.3
Septeiber189728-33
A Prospective Randomized Evaluation of Biphasic Versus Monophasic
Waveform Pulses on Defibrillation Efficacy in Humans
GUST H. BARDY, MD, FACC, TOM D. IVEY, MD, MARGARET D. ALLEN, MD,
GEORGE JOHNSON, BSEE, RAHUL MEHRA, PxD, H. LEON GREENE, MD, FACC
Seattle, Washington
Biphasic waveforms have been suggested as a superior
waveform for ventricular defibrillation. To test this
premise, a prospective randomized intraoperative evalua-
tion of defibrillation efficacy of monophasic and biphasic
waveform pulses was performed in 22 survivors of out of
hospital ventricular fibrillation who were undergoing car-
diac surgery for implantation of an automatic dfibrilator.
‘The initial waveform used in a patient for defibrillation
testing, ether monophasic or biphasic, was randomly se-
lected. Subsequently, each patient served as his or her own
‘contro for defibrillation testing ofthe other waveform. The
defibrillation threshold was defined as the lowest pulse
amplitude that would effectively terminate ventricular fi
brillation with a single discharge delivered 10 s after
initiation of an episode of ventricular fibrillation induced
alternating current. Esch defibrillation pulse was
recorded osclloscopically, and defibrillation pulse voltage,
current, resistance and stored energy were measured.
Fifteen (68%) of the 22 patients had a lower defibrilla-
tion threshold with the biphasic pulse, 3 (14%) had a lower
threshold with the monophasle pulse and 4 (18%) had equal
Aefbeilation thresholds (within 1.0 J) regardless of wave-
form. The mean lading edge defibrillation threshold volt-
age was 317 & 105 V when the monophasic pulse was used
and 267 £ 102 V (16% less) when the biphasic pulse was
used (p = 0,008). Mean leading edge defibrillation thresh-
ald current was7.9 # 3.7 A when the monophasic puse was
used and 6.8 + 3.8 A (14% less) when the biphasic pulse
was used (p = 0.081). There were no differences in system
resistance with the two waveforms: the leading edge ress-
tance for the monophasic pulse was 45.3 + 17.8 and 44.1 +
144 0 for the biphasic pulse (p = 0.183). The stored
nergy defbrilation threshold was 8.5 6.1 J when the
‘monophasic pulse was used and 6.3 + §.2J (26% less) when
the biphasic pulse was used (p = 0.028). These findings
show that in the majority of patients, biphasie pulse
defibrillation is more efficient than monophasic pulse
defibrillation,
(J Am Coll Cardiol 1989;14:728-33)
In 1947, Gurvich and Yuniev (1) demonstrated improved
defibrillation efficacy of underdamped condenser discharges
over standard condenser discharges in animals. In 1962,
Kouwenhoven and Knickerbocker (2) proposed the use of
sequential capacitor discharges of opposite polarity as a
‘means of facilitating the development of a portable defibril-
lator, Subsequently, through the initial and later work of
Schuder et al. (3-6) and Negovsky etal. (7), animal data had
From the Division of Cardiology, Department of Medicine and the
Division of Cardiothoracic Surgery, Department of Surgery from the Unive
sity of Washington, Seattle, Washington. This work was supported in prt by
(Grants R23-L 3617001 apd ROL-HLST4724O1A3 fom the National Insitutes
of Heath, Bethesda, Maryland and by a grant ffom the American Hear
Associaton, Washiegton Aflate, Seat
‘Manuscipt received October 17, 1988; revised manuscript rectved
March 17, 193, accepted Apel 1,199.
‘Addcess for sepunts: Gast H. Bardy, MD, Universiy of Washington
Medial Center, RO-2, 1959 Pacic Avene, Nonthes, Sete, Washingon
SaI88
(01989 bythe American Collegeof Cardoogy
accumulated to suggest that biphasic rectangular wave-
forms, atleast of certain types, provided a higher margin of
defibrillation efficacy compared with monophasic wave-
forms. Preliminary work in humans by Winkle et al. (8)
suggested that biphasic pulses may decrease energy require-
‘ments for defibrillation by automatic antiarrhythmic devices.
If biphasic pulsing is superior to monophasic pulsing,
incorporation of this methodology into an automatic defibril-
lator might facilitate the development of smaller, more
efficient devices. Therefore, to address the question of
biphasic defibrillation efficacy in humans, we pursued a
prospective randomized study in survivors of out of hospital
ventricular fibrillation at the time of automatic defibrillator
insertion, comparing the standard monophasic waveform
pulse now used in implantable defibrillators with a biphasic
waveform pulse of equal tlt. In addition, the study was
designed such thatthe biphasic pulse chosen for evaluation
could be generated from a single capacitor and easily imple-
73s. 10979989.50BARDY ET AL.
70
BIPHASIC VERSUS MONOPHASIC DEFIBRILLATION
JACC Vol, 14, No.3
September 180728-33
‘Table 1. Electrical Variables for Defibrlation Thresholds as a Function of Defibrillation Waveform in 22 Patients
Voliage* Caren Resistance? ‘Stored Enereyé
wo (amps) (ohms) CGoues)
Patient No. __Momophasic Biphasic _Menophasic ‘Biphasic ——Monaphasc——=iasic —=Monophaic ‘Biphasic
1 38 T 98 o 369 382 2 Py
2 ™m as 50 18 Ma aL 23 49
3 Be co 69 6s M9 wa so 50
4 63 m 16 163 ME x09 ws ma
s 1% ‘0 3A 40 ay 486 2 a
6 26 18 st 38 20 8 so as
1 om 21 86 39 as a9 99 st
8 wi ” 3 4a a8 368 4g a
9 an 22 62 a3 as 15 SL
0 w 2 32 ros sw 1 53
4 23 1% 40 “1 48 si as
n n 36 61 48 0 189 a
B m ns 3 389 356 2 2a
“ a4 aM 94 37 BG 9 18
15 ae 2 3S a9 319 99 to
6 au 338 03 ue Mr W so
0 35 10 Iss ua ne 00 203
8 ot 39 60 a8 5.1 no 2
” we 268 4s oy a 16 su
» 31 2» m2 m3 Ba 96 1
2 1s 9 i 2. 47 2 to
2 30 70 2 so S19 19 sa
Mean #SD 3721S 67S M—TYEAT BRASS ASGTINS LEM RS EOI 3282
p= 0.08; fp = 0.051; tp = 0.188; bp = 0.028
where the value of the energy (E,) stored on a capacitor is
half the product ofthe capacitance (C) and the square of the
voltage across the capacitor (V.) (equation 1). In this ease,
the capacitor voltage is equal to the sum of the load voltage
across the defibrillation electrode (V,) and the voltage
across the internal resistance of the pulse generator (V))
(equation 2). If the value of the internal resistance (R,) is
known, the voltage drop across the internal resistance can be
calculated (equation 3). Because V,_ rather than V. was
being measured, the voltage decrease (V,) across the internal
resistance (R,) had to be taken into account when calculating
the stored energy (equation 4). The values for V, and fy in
the equation for stored energy are the leading edge values of
the load voltage and current waveforms. The internal resis-
tance of the pulse generator was approximately 5. and the
capacitance was 120 uF. Knowledge ofthese factors allowed
for calculation of stored energy values of each defibrillation
pulse.
Statistical considerations. Statistical analysis of the data
was performed with use of paired Student's f tests to
compare the efficacy of the two waveforms with respect to
leading edge voltage, leading edge current, resistance and
stored energy at defibrillation threshold values. The defibril-
Tation threshold was defined as the lowest stored energy that
could successfully terminate ventricular irillation with only
‘one discharge delivered 10s after the initiation of ventricular
fibrillation. After the defibrillation threshold was determined
with the intial monophasic or biphasic waveform, the de-
fibrillation threshold was compared with that determined
with the alternate waveform so that each patient served as
his or her own control.
‘Although itis recognized that defibrillation isto a degree
a statistical phenomenon, the definition of defibrillation
threshold chosen for this study adjusts tothe clinical realities,
and limitations of the repetitive induction of ventricular
fibrillation in patients. A strength-efficacy curve would be
preferred, but it would require large numbers of fibrillation
defibrillation episodes to acquire, which is clinically unreal-
istic. Therefore, we relied on the statistical method of the
two-tailed test to accommodate forthe inherent limitations
of the “defibrillation threshold.”
Results
Clinical and defibrillation characteristics (Table 2). Of the
22 patients studied, 16 were male and 6 were female; their
mean age was 57 + 13 years. Twelve patients principally had
coronary artery disease, six principally had idiopathic car-
iomyopathy, one had primary electrical disease, one had
long QT syndrome and two had a combination of coronary
disease and cardiomyopathy. The mean ejection fraction
was 0.39 + 0.18. Amiodarone had been administered over aJACC Vol 14, No.3 BARDY ET AL Bi
September 199°738-33 BIPHASIC VERSUS MONOPHASIC DEFIBRILLATION
Table 2. Clinical Characteristics of 22 Patents
Defrilation
Pest Age (yn) & Clinical Structural Heart Surpcal Hleciroge
No, Genter Ahythna Disease Drugs Procedure System
H 2M VF CAD. DM 0 = ‘CABG. AICD uo
2 “6F ve we 03 = ay UL
3 aM VF«G Primary clerical 089 Aniodirone ICD u
sisease
4 on ve CAD, AN om = ARG, AICD uu.
5 ar VE.NSVT —ICM,MR,CAD 0 = Alc UL
5 oan ve CAD. DMI 065 = CABG, AICD cL
1 ont VE,NSVT IDC, CAD. on = Aled uo
8 Ste ve we on = leo LL
¥ aM vF lew oa = ‘led u
0 ‘tt VE.VT we 0x = ‘led LL
u oM VE Iew 03 = arc LL
R 0M ve CAD, OM on Amiodarone alcD
B oF vex Long OT 038 - Aled
syndrome
or ve CAD, SEAME 099 = ARG. ale
is 13M ve CAD, AML, DMI 0x ICD
16 ant ve CAD, AME 035 = CABG. patches
7 oF ve CAD, AMI 0% Alc
rn ™ ved CAD. AM 025 = alco
% oon ve CAD. AMI 018 CABG. AICD
» ot vE CAD, AME os Amiodarone ——_AlCD:
1 xm ve Hew ase Disopyramise ICD
n 70M ve CAD, DMI. LM os Amiodarone AICD
Mea=S0 97213 039 = 018
‘A= small (or average) epicardial AICD deflation path; AICD = automatic implantable cardioverter defibrillator; AMI anterior myocardial infarction:
‘CABG = coroeary
patch; LY ~ left ventiolig; M = me; MR = mitral regurgitation; NSV'
‘SD = standard deviation: SEAME
Tong term up to the time of defibrillation testing in four;
disopyramide had been given to one of these patients. Total
pulse duration for monophasic pulses was 6.2 + 1.8 s. Pulse
duration was 6.1 + 1.6 ms for the intial positive phase ofthe
biphasic pulse and 5.9 = 1.6 s for the negative phase of the
biphasic pulse. In 19 of the 22 patients, defibrillation was
performed with use of two large CPI defibrillation electrodes
(one over the anterior right ventricle and one over the
posterolateral left ventricle). One patient underwent defiril-
lation with two small CPI defibrillation electrodes and two
patients with a superior vena cava-left ventricular small
patch electrode system.
Defibrillation threshold data (Table 2). Fifteen (68%) of
the 22 patients had a lower defibrillation threshold (> 1.0 J)
with the biphasic pulse. Three patients (14%) had a lower
defibrillation threshold with the monophasic pulse and four
(18%) had equal defibrillation thresholds (within 1.0 J)
regardless of waveform. The mean leading edge defibrillation
threshold voltage was 317 + 105 V when the monophasi
pulse was used and 267 + 102 V (a 16% decrease) when the
biphasic pulse was used (p = 0.008) (Fig. 2)
bypass surgery: CAD = coronary atery disease: DM
HM » hypertrophic cardiomyopathy: ICM = idiopathic cardiomyopathy: IDC ~ idiopathic lated cardiomyopathy: L
onsustaned ventricular tachycardia: S
subendocardalaneiormyocarda infarction; VF = ventricular Rbilaion;
iphragmatic myocardial infarction: BF = ejection faction; F = female
are epicardial AICD deftriliion
superie vent cava AICD coi eletrde:
entcula achycadi,
Mean leading edge defibrillation threshold current was
7.9 = 3.7 A when the monophasic pulse was used and 6.8 +
3.8 A (14% less) when the biphasic pulse was used (p
0.051), There were no differences in system resistances with
the two waveforms. The leading edge resistance was 45.3
17.5.0 for the monophasic pulse and 44.1 + 14.4 0 for the
Figure 2. Delivered leading edge voltage atthe defibrillation thresh-
ld for both monophasic and biphasic pulses
Yensge()
o 8888 8 8732 BARDY ETAL,
BIPHASIC VERSUS MONOPHASIC DEFIBRILLATION
Monoonesic — Bohasic
Figure3. Stored energy (voltage) fr both monophasic and biphasic
pulses atthe defibrillation threshold
biphasic pulse (p = 0.183). The stored energy defibrillation
threshold was 8.5 + 6.1 J when the monophasic pulse was
used and 6.3 + 5.2 J (a 26% decrease) when the biphasic
pulse was used (p = 0.028) (Fig. 3).
Discussion
Clinical implications and technologie considerations. Our
findings show a 26% decrease in the amount of energy
needed for implantable defibrillators if biphasic pulses rather
than monophasic pulses are used for defibrillation. ‘This is
consistent with recent animal and human investigations
8.11.12) of biphasic pulses on defibrillation efficacy, even
though the waveforms used in our study are different in
shape. Although many of the recent animal studies investi-
gating biphasic waveforms have used rectangular pulses
rather than truncated exponential waveforms, use of rectan-
gular waveforms in implantable defibrillators is impractical
at this time. The capacitance necessary to deliver a rectan-
gular pulse is probitive given limitations on capacitor size.
Therefore, with currently available capacitors, it is more
practical to use truncated pulses.
We chose a 65% tilt, 120 wF, biphasic truncated expo-
nential waveform to keep our investigation referrable to the
monophasic pulsing techniques already in use with the
automatic implantable cardioverter-defibrillator (AICD by
CPI), In addition, we maintained constant waveform tlt for
the monophasic pulse as well as for both phases of the
biphasic pulse to control for the effect of patient to patient
variability in tissue resistance on defibrillation efficacy
‘Also, the waveform tilt chosen for this study (65%) is based
on previous animal data (13) showing that optimal tilt is a
compromise between improved defibrillation efficacy and
post-shock myocardial dysfunction at low tilts. Finally, we
tried to use a biphasic pulse that could be generated by &
single capacitor. The need for two capacitors can be circum-
vented if the characteristics of the biphasic waveform are
properly delineated. Thus, a biphasic waveform should not
significantly alter device size if the pulse is created by
‘truncating the positive phase, reversing the output potarity
JACE Vol, 4, No.3
September 1980-728-33
and delivering the remaining charge on the capacitor as the
negative phase of the pubs.
Explanations for biphasic waveform superiority. There is,
no clear explanation why defibrillation with biphasic pulsing
generally proves superior to defibrillation with monophasic
pulsing. However, insight into the general effect of biphasic
pulses on cardiac mechanical function has been provided by
Jones etal. (14-17). They have shown that biphasic wave-
forms with modest degrees of undershoot cause less post:
shock mechanical myocardial dysfunction than monophasic
pulses. Their observations are derived from a model of
cultured chick embryo layered myocardial cells in which
electric shocks were shown to produce transient microle-
sions in the sarcolemma, possibly as a consequence of
‘membrane compression by the electric field. The size of the
microlesions and the fraction of the membrane surface over
which they occur increase with shock intensity. During and
after the shock, an indiscriminate ionic exchange occurs
through the microlesions, which results in membrane depo-
larization and low resting membrane potentials (17). If,
however, a small reverse polarity component is added to the
biphasic pulse, Jones et al. (14-17) postulated that the
phospholipids of the sarcolemma are reoriented, thereby
decreasing the magnitude and duration of microlesion for-
mation and, thus, membrane depolarization, Although these
experiments satisfactorily account for the mechanical dys-
function seen after a shock, the relation between enhanced
sarcolemmal microlesion closure and enhanced defibrillation
efficacy with biphasic pulses is unclear.
One possible mechanism to explain why biphasic pulse
defibrillation may be superior to monophasic pulse defibrit-
lation is break excitation. In intracellular recordings from
cells that are proximal to an extracellular stimulating elee-
trode during monophasic stimulation, the cells closest to the
cathode become depolarized and those underneath the an-
ode become hyperpolarized (18,19). Because excitation is
normally associated with a reduction in the transmembrane
Potential, excitation would not be expected to occur in the
cells that are hyperpolarized. However, it might be possible
to explain excitation in hyperpolarized tissue by considering
the phenomenon of break excitation (that is, excitation
elicited at the “break” or at the end of the stimulus). This
phenomenon has been observed most frequently in nerve
fibers subjected to high extracellular potassium. The first
explanation for this phenomenon was given by Hill (20) and
subsequently by Hodgkin and Huxley (21,22), who observed
that break excitation was facilitated not only in a high
extracellular potassium milieu, but also in cells with low
resting transmembrane potentials. In the fibrillating heart,
both high extracellular potassium and low transmembrane
potentials are present, thereby providing the optimal envi-
ronment for break excitation to occur in the hyperpolarized
tissue underneath the anode (22,23). In contrast, withJACC Vol. 16, No.3
‘September 196072835
monophasic pulses, excitation of the hyperpolarized tissue
underneath the anode may be less likely to occur.
‘An alternative explanation to break excitation for the
improved defibrillation efficacy with biphasic pulsing may be
associated with the two pulses contained in a “single
biphasic pulse. During the positive phase of the biphasic
pulse, excitation or defibrillation may occur more easily in
the tissue underneath the anode (10). With the onset of the
second phase of the pulse, the tissue that had been under the
cathode is then under the anode. Consequently, during the
second phase of the pulse, this tissue under the ‘new’
anode (that is, the old cathode) may now undergo excitation.
Thus, this use of two pulses and, in effect, two anodes may
facilitate depolarization of more tissue than is possible with
‘a monophasic pulse of equal leading edge voltage.
‘Conclusions. Whatever the mechanism by which biphasic
pulsing facilitates defibrillation, the difference in voltage and
energy requirements for defibrillation is sufficient to warrant
inclusion of this waveform in future generation automat
defibrillators. Together with other pulsing methods, biphasi
pulsing should help decrease device size and improve de-
fibrillation efficiency.
References
1, Gurvieh NL, Yuniev GS. Restoration of hear rhythm during iritation
bya condenser discharge. Am Rev Soviet Med 194742526,
Kouventoven WB, Knickerbocker GO. The development ofa portable
debate. Trans Ar lst Elec Engr 196281 (part 32831,
3, Schuder JC, Stoeckle H, Dolan AM. Transthoracic defibrillation with
square wave simul one-half cycle, ore eyse, and mutcycle wave-
Terms. Cite Res 19641525568
4. Schuder JC, God JH, McDaniel WC. Untrahiphbenereybyope tyre:
lrom'9CR bidirectional waveform defi. Med 3:0 Eng Comput
1982:20:419-26
5. Schoder JC, Gold JH, Stoeckle H, MeDaniel WC, Cheung KN, Transtho
race venticulrdefiiatin inthe 100 kg calf with syrmetcl Leycle
hidiretional rectangular weave stimuli IEEE Trans Am Biomed Eng
19883041522,
6 Schuder JG, MeDaniel WC, Stoeckle H. Defibration of 100 kg calves,
‘ith asyrmesrial, bidirectional, rectangular pulses. Cardiovasc Res
foes
17. Negovsky VA, Smerdov AA, Tabak VY, Venin IV, Bogushevich MS.
Chien of effciency and safety of defibilating impulse. Resuscitation
05367
u
23
BIPHASIC VERSUS MONOPHASIC DEFIBRILLATION
Winkle RA, Mead RH, Ruder MA, e al. Improved low enerey defi
lation efficacy in man using a biphasic iruncated exponental wavelort
(abst) J Am Coll Cail 1967 8:124,
Lardy GH, Ivey TD, Greene HL, Evaluation of eecrode polarity on
efiratin effcsy. Am J Carl I98563433-7
Barly GH, Ivey TD, Allen MD, Jobason G, Greene HL. Prospective
comparison of sequenal pulse and single pulse defibation using |
liferenc clinically avaiable systems in man. J Am Coll Cardia 198
M1657,
‘Warton JM, Fravier DW, Tang ASL, Ideker RE. Flecrphysiloic
effects in vivo of biphasic and monophasic simul (abst). 1 Am Call,
Crgit 19681114.
Dixon EG, Tang ASL, Wolf PD, eal. Improved defiristion thresholds
wth age contoured pian ees in biphae waveforms. Circo
Tan 187.76: 17-84
Wesssle JL. Bourland 3D, Tacket WA, Geddes LA. Bipolar catheter
efron in dogs using trapezokal waveforms of various ti
1 Blecroardial 1980; 1335466
Jones JL, Lepeschkin E, Jones RE, Rush S. Response of cured
ryocaral cell to countershocktype elects eld stimulation. Am
Physiol IESE
Joes JL Jones RE. Improved defiriat waveform safety factor with
biphasic waveforms, Am J Physio [983238 HO0-S
Jones IL, Jones RE. Decreased defieillatr induced dysfunction with
biphasic rectangular waveforms. Am J Physiol 1984247: H792-6.
2 Jones JL, Jones RE, Balasky G. Microlsion formation in myocardial
calls by bighintensy elec fi stimolation. Am J Physiol 1987
sae
Goto M. Brooks C McC. Membrane eacitabiliy of the fog ventricle
examined by lng pss. Am I Physiol 1969;27: 1286-48
Hoshi T, Matsuda K. Excitability cycle of cardiac muscle examined by
‘racelluta stimulation, Jpn Physiol 1962124356,
i AV, Excitation and secommodaton in nerve. Proe Rey Soe London
15362198 05-8,
Hodgkin AL. The ionic basis of electrical activity in neve and muscle,
Bl Rev I9S1:26319-409
Hodgkin AL, Huxley AF. A quantiative description of membrane
curren anit application fo condition and excitation a neve] Physik
I9Sts17-50-4
Janse MJ, Mosik H, Van Capelie FL, Kleber AG, Wilms Schopman F.
Durer D: Ventriuar arrhythmias nthe fst 15 minutes of acute regional
‘myocarl schema in the isolated pig heart possible role of ury
farens, In Kulberus HE, Welles HI. eS. Sudden Death. The
Hague: Martinis Not 1889-103,
Packer JO, Chong MA, West RO, Case RB. The effect of ischemia and
eaton in ear ate on myocardial potassium Balance in man. Cuca
[ation 197042:205-12,