Professional Documents
Culture Documents
DOI 10.1007/s10384-014-0344-2
CLINICAL INVESTIGATION
Received: 24 February 2014 / Accepted: 17 July 2014 / Published online: 5 September 2014
Ó Japanese Ophthalmological Society 2014
123
484 N. Maeda et al.
123
Biomechanics after keratoplasty 485
Fig. 1 Representative displays of the dynamic Scheimpflug analyzer corneal velocity. Note the DA in PK eyes (1.47 mm) is deeper than in
in a normal eye (left) and in an eye following penetrating keratoplasty normal eyes (1.06 mm) and that the radius at the highest concavity
(right). The Scheimpflug image at the highest concavity is shown in (Rhc) in the PK eyes (4.55 mm) is steeper than in normal eyes
addition to the IOP, pachymetry, and 10 biomechanical parameters (6.62 mm)
and graphs for deformation amplitude (DA), applanation length, and
Table 1 Output of dynamic Scheimpflug analyzer in normal eyes and eyes that underwent one of three corneal transplantation techniques
Control DSAEK DALK PK P value
IOP 14.63 ± 1.49 15.03 ± 3.74 14.14 ± 3.16 14.28 ± 1.23 0.6678
CCT 552 ± 31*, ** 652 ± 61*, # 628 ± 63**, ##
552 ± 56#, ## \0.0001
DA 1.07 ± 0.09*, ** 1.08 ± 0.12# 1.18 ± 0.18* 1.20 ± 0.13**, #
0.0004
, , # #
Rhc 7.57 ± 0.78* ** 6.44 ± 0.58* 6.04 ± 1.22** 6.34 ± 0.37 \0.0001
Thc 17.19 ± 0.75 17.31 ± 0.67 16.93 ± 0.62 17.17 ± 0.42 0.4644
PDhc 3.62 ± 1.30 2.76 ± 1.08* 3.56 ± 1.37 4.37 ± 1.19* 0.0043
Ti 7.67 ± 0.24 7.72 ± 0.55 7.46 ± 0.47 7.60 ± 0.22 0.1456
Li 1.81 ± 0.24 1.81 ± 0.21 1.80 ± 0.24 1.80 ± 0.23 0.9968
Vi 0.16 ± 0.03 0.14 ± 0.03 0.15 ± 0.05 0.15 ± 0.03 0.3990
To 22.30 ± 0.39* 20.07 ± 0.69 21.85 ± 1.02* 22.11 ± 0.26 0.0471
Lo 1.86 ± 0.39 1.89 ± 0.11 1.93 ± 0.45 2.00 ± 0.48 0.6712
Vo -0.33 ± 0.07 -0.34 ± 0.12 -0.34 ± 0.12 -0.35 ± 0.11 0.8297
DSAEK Descemet stripping automated endothelial keratoplasty, DALK deep anterior lamellar keratoplasty, PK penetrating keratoplasty, IOP
intraocular pressure (mmHg), CCT central corneal thickness (lm), DA deformation amplitude at the highest concavity (mm), Rhc radius at the
highest concavity (mm), Thc time at the highest concavity (ms), PDhc peak distance at the highest concavity (mm), Ti time at the inward
applanation (ms), Li length at the inward applanation (mm), Vi velocity at the inward applanation (m/s), To time at the outward applanation (ms),
Lo length at the outward applanation (mm), Vo velocity at the outward applanation (m/s)
P value: 1-way ANOVA; *, **, #, ##
: Tukey–Kramer test, P \ 0.05
values of the different parameters obtained by the dynamic a patient who had undergone PK are shown in Fig. 1.
Scheimpflug analyzer for the 4 groups of eyes. Probability Supplemental digital videos of these eyes during the
values below 0.05 were considered significant for all measurements are shown in Video 1. The DA at the highest
analyses. concavity for the PK eye was 1.47 mm, which was larger
than the 1.06 mm in the normal eye. Moreover, the Rhc in
the PK eye was 4.55 mm, which was steeper than the
Results 6.22 mm in the normal eye.
Statistical comparisons for the 4 groups of the parame-
Typical Scheimpflug images at the time of highest con- ters obtained by the dynamic Scheimpflug analyzer are
cavity due to the air pulse in a normal eye and in an eye of shown in Table 1. One-way ANOVA showed significant
123
486 N. Maeda et al.
Fig. 2 Scatterplot of deformation amplitude (DA) (left) and radius at normal zone, some of the PK and DALK eyes have higher DA and
the highest concavity (Rhc) (right) as a function of age. Although lower Rhc values than those of the normal eyes of the same age
some of the DSAEK, PK, and DALK eyes are distributed inside the
differences among the 4 groups in CCT, DA, Rhc, peak found in all eyes (control eyes: R2 = 0.5519, P \ 0.0001;
distance at the highest concavity, and time at the outward DSAEK eyes: R2 = 0.8437, P \ 0.0001; PK eyes:
applanation. R2 = 0.2923, P = 0.0138) except the DALK eyes
The CCT of the DSAEK or DALK eyes was signifi- (P = 0.1644). A significant positive correlation between
cantly thicker than that of the control or PK eyes. IOP and Rhc was found in the control eyes (R2 = 0.1051,
The PK eyes had significantly larger DA than did the P = 0.0217) and the DALK eyes (R2 = 0.3877,
control and DSAEK eyes, and the DALK eyes also had P = 0.0174), but not in the DSAEK eyes (P = 0.4768) or
significantly larger DA than did the control eyes. The the PK eyes (P = 0.6801).
DSAEK, DALK, and PK eyes had shorter Rhc than did the The scatterplots in Fig. 4 show the relationships
control eyes. The DSAEK eyes had a significantly shorter between CCT and DA (left) and between CCT and Rhc
peak distance at the highest concavity than did the PK eyes. (right). A significant negative correlation between CCT and
The DALK eyes had significantly faster time at the out- DA was found in the control eyes (R2 = 0.1080,
ward applanation than did the control eyes. P = 0.0197), but not in the other eyes (DSAEK eyes:
The scatterplots in Fig. 2 show the relationships 0.9842; DALK eyes: 0.3306; and PK eyes: 0.9305). A
between age and DA (left) and between age and Rhc (right). significant positive correlation between CCT and Rhc was
Although most of the DSAEK eyes and some of the DALK found in the control eyes (R2 = 0.2319, P = 0.0004), but
and PK eyes were located close to the distribution of the not in the other eyes (DSAEK eyes: P = 0.0720; DALK
normal eyes, all of the eyes that had DA greater than 1.29 eyes: P = 0.4008; and PK eyes: P = 0.4929).
were PK or DALK eyes. The DA in all of the normal and
the DSAEK eyes was 1.29 or less. Significant correlations
between age and DA were not found in any of the eyes:
normal eyes: P = 0.9996; DSAEK eyes: P = 0.1223; Discussion
DALK eyes: P = 0.6638; and PK eyes: P = 0.2106.
Similarly, in all of the control eyes, the Rhc was 6.2 mm The in situ measurement of corneal biomechanics with the
or larger. However, some of the DSAEK, DALK, and PK Ocular Response Analyzer (ORA) uses 2 biomechanical
eyes had a Rhc less than 6.2 mm. Significant correlations parameters: corneal hysteresis (CH) and corneal resistance
between age and Rhc were not found in any of the eyes: factor (CRF). These values are derived by measuring the
control eyes: P = 0.8587; DSAEK eyes: P = 0.6059; time of the 2 peak signals that correspond to the inward and
DALK eyes: P = 0.9972; and PK eyes: P = 0.5341. outward times of applanation of the cornea in response to a
The scatterplots in Fig. 3 show the relationships variable air pulse. In contrast, the dynamic Scheimpflug
between IOP and DA and between IOP and Rhc. A sig- analyzer, the Corvis ST, records the dynamic deformation
nificant positive correlation between DA and IOP was of the horizontal Scheimpflug images in response to
123
Biomechanics after keratoplasty 487
Fig. 3 Scatterplot of deformation amplitude (DA) (left) or radius at PK eyes: R2 = 0.2923, P = 0.0138). A significant positive correla-
the highest concavity (Rhc) (right) as a function of IOP. A significant tion between IOP and Rhc was found in the control eyes
positive correlation was found between DA and IOP (control eyes: (R2 = 0.1051, P = 0.0217) and the DALK eyes (R2 = 0.3877,
R2 = 0.5519, P \ 0.0001; DSAEK eyes: R2 = 0.8437, P \ 0.0001; P = 0.0174)
Fig. 4 Scatterplot of deformation amplitude (DA) (left) or radius at (CCT) and DA (R2 = 0.1080, P = 0.0197) or between CCT and Rhc
highest concavity (Rhc) (right) as a function of central corneal (R2 = 0.2319, P = 0.0004) was found only in the control eyes
thickness. A significant correlation between central corneal thickness
pressure by a constant air pulse [22] In addition to the Under the constant strain of the air pulse, eyes with
inward and outward applanation times, the dynamic higher ocular rigidity will show lower DA and larger Rhc.
deformation of the cornea in response to the air pulse is In contrast, eyes with lower ocular rigidity will have higher
recorded, as shown in the supplemental digital content DA and smaller Rhc.
(Video 1). The biomechanical parameters such as the DA Our findings showed that the DA in the PK eyes was
at the highest concavity and the radius at the highest con- significantly greater than in the control eyes and the
cavity can be obtained from these images [25]. To the best DSAEK eyes, and the DA in the DALK eyes was also
of our knowledge, this is the first study to use a dynamic significantly higher than in the control eyes. No significant
Scheimpflug analyzer to investigate the corneal biome- difference in DA was found between the DSAEK and the
chanics following corneal transplantation. control eyes. That eyes after PK or DALK have lower
123
488 N. Maeda et al.
rigidity than do normal eyes is reasonable because the The relationship between the parameters obtained with
stromal wound healing at the host-graft junction and/or the ORA and those obtained with the Corvis ST should be
pathologic peripheral host cornea is not strong enough to clarified in the future. Although both instruments use air
keep the structural integrity of the cornea. Lower DA in pulse pressure to induce strain, the air-pulse pressure is
DSAEK eyes might suggest better corneal integrity than in variable with the ORA and constant with the Corvis ST.
PK or DALK eyes. The Corvis ST may have potential advantages for deter-
On the other hand, the Rhc in all postkeratoplasty groups mining the stress–strain relationship of the eye. In addition,
was significantly smaller than in the control group. This the stress is evaluated by different principles, i.e., corneal
may suggest that the ocular integrity after these corneal light reflex and Scheimpflug images. Therefore, it might be
transplantations is not as good as that of normal eyes. The difficult to develop interchangeable parameters for both
discrepancy between DA and Rhc in DSAEK eyes should instruments.
be investigated further. This study has some limitations. The numbers of par-
Significant difference in the peak distance at the highest ticipants was small, and there were many variables among
concavity between DSAEK and DALK eyes may indicate the participants and procedures. The mean age of the
differences among the procedures in centrifugal deforma- DSAEK eye group was significantly higher than those of
tion to the same load. The time of outward applanation also the other groups. Such variations might lead to a scattering
behaved differently in the DALK eyes. The shorter time of of data. However, this preliminary study with a dynamic
outward applanation in DALK eyes may suggest that a Scheimpflug analyzer showed that differences in corneal
thicker cornea will return to normal shape more easily. The biomechanical properties exist between eyes with normal
relationships among the biomechanical parameters corneas and eyes after corneal transplantation, or among
obtained with the Corvis ST are unknown. At least, we can eyes that have undergone different types of corneal
say that these indices following keratoplasty show some transplantation.
abnormalities. Further studies will be necessary to clarify Future studies will be necessary to clarify the effects of
the significance of each parameter. keratoplasty on the integrity of the eye and to evaluate the
The rigidity of the eye is generally thought to be accuracy of IOP measurements in such eyes with different
affected by age, corneal thickness, IOP, and corneal bio- corneal biomechanics. As shown in the video images, only
mechanics. In this study, significant correlations were not limited data are currently available for the dynamic Sche-
found between age and DA or between age and Rhc in any impflug analyzer. In other words, this technology may have
of the groups. This may be due to the limited age range of the potential to elucidate the detailed biomechanical
the patients studied. Significant correlations between IOP behavior of the eye in future investigations.
and DA were found in normal, DSAEK, and PK eyes, and
significant differences were also found between IOP and Acknowledgments Publication of this article was supported in part
by a Grant-in-Aid for Scientific Research (no. 24592669; to Naoyuki
Rhc in normal and DALK eyes. These results suggest that Maeda) from the Japanese Ministry of Education, Culture, Sports,
IOP should be matched when comparing the corneal bio- Science and Technology. The authors would like to thank Duco
mechanics among groups. A significant correlation Hamasaki, PhD, for the English language review.
between the CCT and biomechanical parameters was found
Conflicts of interest N. Maeda, Research Grant (Topcon), Instru-
only in the control group. Although higher ocular rigidity ment (CorvisÒ ST: Oculus); R. Ueki, None; M. Fuchihata, None;
can be expected in normal eyes with thicker corneas than in H. Fujimoto, None; S. Koh, None; K. Nishida, None.
eyes with thinner corneas, the integrity of the thicker cor-
nea after keratoplasty might not be similar to that of the
normal thicker cornea.
As shown in Figs. 2, 3, and 4, the variations in the
distribution of DA and Rhc for the postkeratoplasty groups References
were more prominent than those for the control group.
Because the stress–strain relationship is the basis of IOP 1. Archila EA. Deep lamellar keratoplasty dissection of host tissue
with intrastromal air injection. Cornea. 1984–1985;3:217–8.
measurements, careful attention should be paid during the 2. Sugita J, Kondo J. Deep lamellar keratoplasty with complete
interpretation of measured IOPs in eyes with different removal of pathological stroma for vision improvement. Br J
corneal biomechanical parameters. In addition, some Ophthalmol. 1997;81:184–8.
postkeratoplasty eyes might behave differently in corneal 3. Anwar M, Teichmann KD. Big-bubble technique to bare Desc-
emet&s membrane in anterior lamellar keratoplasty. J Cataract
refractive surgery from eyes with normal corneas owing to Refract Surg. 2002;28:398–403.
these different corneal biomechanical properties. 4. Melles GR, Eggink FA, Lander F, Pels E, Rietveld FJ, Beekhuis
Abnormal biomechanical parameters in postkeratoplasty WH, et al. A surgical technique for posterior lamellar keratopl-
eyes may indicate the vulnerability of the host-graft wound. asty. Cornea. 1998;17:618–26.
123
Biomechanics after keratoplasty 489
5. Terry MA, Ousley PJ. Replacing the endothelium without corneal 16. Fontes BM, Ambrósio R Jr, Jardim D, Velarde GC, Nosé W.
surface incisions or sutures: the first United States clinical series Corneal biomechanical metrics and anterior segment parameters
using the deep lamellar endothelial keratoplasty procedure. in mild keratoconus. Ophthalmology. 2010;117:673–9.
Ophthalmology. 2003;110:755–64. 17. Vinciguerra P, Albè E, Mahmoud AM, Trazza S, Hafezi F,
6. Price FW Jr, Price MO. Descemet’s stripping with endothelial Roberts CJ. Intra- and postoperative variation in ocular response
keratoplasty in 200 eyes: early challenges and techniques to analyzer parameters in keratoconic eyes after corneal cross-
enhance donor adherence. J Cataract Refract Surg. linking. J Refract Surg. 2010;26:669–76.
2006;32:411–8. 18. Yenerel NM, Kucumen RB, Gorgun E. Changes in corneal bio-
7. Lee WB, Jacobs DS, Musch DC, Kaufman SC, Reinhart WJ, mechanics in patients with keratoconus after penetrating kera-
Shtein RM. Descemet’s stripping endothelial keratoplasty: safety toplasty. Cornea. 2010;29:1247–51.
and outcomes: a report by the American Academy of Ophthal- 19. Jafarinasab MR, Feizi S, Javadi MA, Hashemloo A. Graft bio-
mology. Ophthalmology. 2009;116:1818–30. mechanical properties after penetrating keratoplasty versus deep
8. Melles GR, Ong TS, Ververs B, van der Wees J. Descemet anterior lamellar keratoplasty. Curr Eye Res. 2011;36:417–21.
membrane endothelial keratoplasty (DMEK). Cornea. 20. Hosny M, Hassaballa MA, Shalaby A. Changes in corneal bio-
2006;25:987–90. mechanics following different keratoplasty techniques. Clin
9. Price MO, Giebel AW, Fairchild KM, Price FW Jr. Descemet&s Ophthalmol. 2011;5:767–70.
membrane endothelial keratoplasty: prospective multicenter 21. Acar BT, Akdemir MO, Acar S. Corneal biomechanical proper-
study of visual and refractive outcomes and endothelial survival. ties in eyes with no previous surgery, with previous penetrating
Ophthalmology. 2009;116:2361–8. keratoplasty and with deep anterior lamellar keratoplasty. Jpn J
10. Koh S, Maeda N, Nakagawa T, Higashiura R, Saika M, Mihashi Ophthalmol. 2013;57:85–9.
T, et al. Characteristic higher-order aberrations of the anterior and 22. Hon Y, Lam AK. Corneal deformation measurement using
posterior corneal surfaces in 3 corneal transplantation techniques. Scheimpflug noncontact tonometry. Optom Vis Sci.
Am J Ophthalmol. 2012;153:284–90. 2013;90:e1–8.
11. Rudolph M, Laaser K, Bachmann BO, Cursiefen C, Epstein D, 23. Hong J, Xu J, Wei A, Deng SX, Cui X, Yu X, et al. A new
Kruse FE. Corneal higher-order aberrations after Descemet&s tonometer—the Corvis ST tonometer: clinical comparison with
membrane endothelial keratoplasty. Ophthalmology. noncontact and Goldmann applanation tonometers. Invest Oph-
2012;119:528–35. thalmol Vis Sci. 2013;54:659–65.
12. Kawashima M, Kawakita T, Shimmura S, Tsubota K, Shimazaki 24. Faria-Correia F, Ramos I, Valbon B, Luz A, Roberts CJ, Am-
J. Characteristics of traumatic globe rupture after keratoplasty. brósio R Jr. Scheimpflug-based tomography and biomechanical
Ophthalmology. 2009;116:2072–6. assessment in pressure-induced stromal keratopathy. J Refract
13. Sari ES, Koytak A, Kubaloglu A, Culfa S, Erol MK, Ermis SS, Surg. 2013;29:356–8.
et al. Traumatic wound dehiscence after deep anterior lamellar 25. Nemeth G, Hassan Z, Csutak A, Szalai E, Berta A, Modis L Jr.
keratoplasty. Am J Ophthalmol. 2013;156:767–72. Repeatability of ocular biomechanical data measurements with a
14. Luce DA. Determining in vivo biomechanical properties of the Scheimpflug-based noncontact device on normal corneas.
cornea with an ocular response analyzer. J Cataract Refract Surg. J Refract Surg. 2013;29:558–63.
2005;31:156–62. 26. Reznicek L, Muth D, Kampik A, Neubauer AS, Hirneiss C.
15. Feigenbaum SK, Qazi MA, Sanderson JP, Roberts CJ, Pepose JS. Evaluation of a novel Scheimpflug-based non-contact tonometer
Changes in corneal biomechanics and intraocular pressure fol- in healthy subjects and patients with ocular hypertension and
lowing LASIK using static, dynamic, and noncontact tonometry. glaucoma. Br J Ophthalmol. 2013;97:1410–4.
Am J Ophthalmol. 2007;143:39–47.
123
Reproduced with permission of the copyright owner. Further reproduction prohibited without
permission.