You are on page 1of 37

TEK OSCILLOSCOPES

MULTI-PURPOSE OSCILLOSCOPE
PRIMER

THE XYZs
OF USING A SCOPE

Tektronix
COMMITTED TO EXCELLENCE
CONTENTS

INTRODUCTION 1 PART II. Making Measurements 19

PART I. Scopes, Controls, & Probes 2


Chapter 6. WAVEFORMS 19

Chapter 1. THE DISPLAY SYSTEM 4 Chapter 7. SAFETY 22


Beam Finder 4
Intensity 4 Chapter 8. GETTING STARTED 22
Focus 4
Compensating the Probe 22
Trace Rotation 5 Checking the Controls 22
Using the Display Controls 5 Handling Probe 22

Chapter 2. THE VERTICAL SYSTEM 6 Chapter 9. MEASUREMENT TECHNIQUES 24


Vertical Position 6 The Foundations: Amplitude and Time Measurements 24
Input Coupling 6 Frequency and Other Derived Measurements 25
Vertical Sensitivity 6 Pulse Measurements 25
Variable VOLTS/DIV 7 Phase Measurements 26
Channel 2 Inversion 7 X-Y Measurements 26
Vertical Operating Modes 7 Using the Z-Axis 27
Alternate Sweep Separation 8 Using TV Triggering 28
Using the Vertical Controls 8 Delayed Sweep Measurements 28
Single Time Base Scopes 28
Chapter 3. THE HORIZONTAL SYSTEM 10 Dual Time Base Scopes 29
Horizontal Position 10
Horizontal Operating Modes 10 Chapter 10. SCOPE PERFORMANCE 32
Sweep Speeds 10 Square Wave Response and High Frequency Response 32
Variable SEC/DIV 11 Instrument Rise Time and Measured Rise Times 32
Horizontal Magnification 11 Bandwidth and Rise Time 33
The DELAY TIME and MULTIPLIER Controls 11
The B DELAY TIME POSITION Control 11 INDEX 34
Using the Horizontal Controls 11

Chapter 4. THE TRIGGER SYSTEM 12


Trigger Level and Slope 13
Variable Trigger Holdoff 14
Trigger Sources 14
Trigger Operating Modes 15
Trigger Coupling 16
Using the Trigger Controls 16

Chapter 5. ALL ABOUT PROBES 17


Circuit Loading 17
Measurement System Bandwidth 17
Probe Types 18
Picking a Probe 18

REPRODUCED BY PERMISSION
OF THE COPYRIGHT OWNER:
TEKTRONIX, INC.
(Letter dated: 11/20/85,
Gaithersbucp, Md.)

Copyright @ 1981, Tektronix, Inc All rights reserved


INTRODUCTION

If you watch an electrical en- performance and its effects on


gineer tackling a tough design your measurements.
project, or a service engineer Having a scope in front of you
troubleshooting a stubborn while working through the chap-
problem, you’ll see them grab a ters is the best way to both learn
scope, fit probes or cables, and and practice applying your new
start turning knobs and setting knowledge. While the funda-
switches without ever seeming mentals will apply to almost any
to glance at the front panel. To scope, the exercises and illus-
these experienced users, the trations use two specific instru-
oscilloscope is their most impor- ments: the Tektronix 2213 and
tant tool but their minds are fo- 2215 Portable Oscilloscopes.
cused on solving the problem, The 2213 is a dual-channel, 60
not on using the scope. MHz portable designed as an
Making oscilloscope mea- easy-to-use, lightweight,
surements is second nature to general-purpose oscilloscope.
them. It can be for you too, but The 2215 is a dual time base
before you can duplicate the oscilloscope with more features
ease with which they use a and capabilities; it’s included so
scope, you will need to concen- you will understand dual time
trate on learning about the base scopes and appreciate
scope itself: both how it works the additional measurement
and how to make it work for you. capabilities they offer.
The purpose of this primer is If you have comments or
to help you learn enough about questions about the material in
oscilloscopes and oscilloscope this primer, please don’t hesitate
measurements that you will be to write.
able to use these measurement
tools quickly, easily, and accu- Chet Heyberger
rately. Portable Scopes Technical
The text is divided into two Communications Manager
parts: Marshall E. Pryor
The first four chapters of Part I 2200 Series Product Line
describe the functional parts of Marketing Manager
scopes and the controls asso- Primer - DS 391199
ciated with those parts. Then a Tektronix, Inc.
chapter on probes concludes PO. Box 500
the section. Beaverton, OR 97077
Part II allows you to build on
the knowledge and experience
you gained from Part I. The sig-
nals you’ll see on the screen of
an oscilloscope are identified by
waveshape and the terms for
parts of waveforms are dis-
cussed. The next two chapters
cover safety topics and instru-
ment set-up procedures.
Then Chapter 9 describes
measurement techniques.
Exercises there let you practice
some basic measurements, and
several examples of advanced
techniques that can help you
make more accurate and con-
venient measurements are also
included. The last chapter in this
primer discusses oscilloscope
PART I. SCOPES, CONTROLS, & PROBES

You can measure almost any-


thing with the two-dimensional
graph drawn by an oscillo-
scope. In most applications the m
scope shows you a graph of
voltage (on the vertical axis)
versus time (on the horizontal
axis). This general-purpose
display presents far more infor-
mation than is available from VfZgWAL
other test and measurement in- %- SECTION
struments like frequency coun-
ters or multimeters. For exam-
ple, with a scope you can find
out how much of a signal is di-
rect, how much is alternating,
how much is noise (and whether
or not the noise is changing with
time), and what the frequency of
the signal is as well. Using a
scope lets you see everything at
once rather than requiring you to
make many separate tests.
Most electrical signals can be
easily connected to the scope
with either probes or cables.
And then for measuring non- Figure 1.
electrical phenomena, trans- THE BASIC OSCILLOSCOPE in its most general form has only four functional blocks:
ducers are available. Transduc- vertical, horizontal, trigger, and display systems (and sometimes, sections). The
ers change one kind of energy display system is also sometimes called the CRT (for cathode-ray tube) section.
into another. Speakers and
microphones are two examples;
the first changes electrical en- Each is named for its function. two-page, fold-out illustration of plug the power cord of the
ergy to sound waves and the The vertical system controls the a Tektronix 2213 front panel at scope into a properly-wired re-
second converts sound into vertical axis of the graph; any- the back of the primer to locate ceptacle before connecting
electricity. Other typical trans- time the electron beam that them on your scope. Next the your probes or turning on the
ducers can transform tempera- draws the graph moves up or controls and their functions are scope; use the proper power
ture, mechanical stress, pres- down, it does so under control of described, and at the end of cord for your scope, and use
sure, light, or heat into electrical the vertical system. The horizon- each chapter there are hands- only the correct fuse. Don’t re-
signals. You can see that given tal system controls the left to on exercises using those con- move the covers and panels of
the proper transducer, your test right movement of the beam. trols. your scope.
and measurement capabilities The trigger system determines The last chapter in this section Now fold out the front panel
are almost endless with an oscil- when the oscilloscope draws; it describes probes. When you illustration at the back of the
loscope. triggers the beginning of the finish reading these five chap- primer, so that it is visible as you
Making measurements is horizontal sweep across the ters, you’ll be ready to make fast read. Use the foldout and follow
easier if you understand the screen. And the display system and accurate oscilloscope Exercise 1 to initialize (set in
basics of how a scope works. contains the cathode-ray tube, measurements. standard positions) the scope
You can think of the instrument in where the graph is drawn. But before you turn on your controls. These standard set-
terms of the functional blocks il- This part of the primer is di- scope, remember that you tings are necessary so that as
lustrated in Figure 1: vertical sys- vided into chapters for each of should always be careful when you follow the directions on
tern, trigger system, horizontal those functional blocks. The you work with electrical equip- these pages, you’ll see the
system, and display system. controls for each block are ment. Observe a// safety pre- same thing on your scope’s CRT
named first, and you can use a cautions in your test and mea- as is pictured or described here.
surement operations. Always
PART I

Exercise 1. INITIALIZING THE SCOPE


Use the foldout figure and call- 3. HORIZONTAL SYSTEM After following the steps in
outs to locate the controls men- CONTROLS: Make sure the Exercise 7, you should plug your
tioned here. HORIZONTAL MODE switch is scope into a properly-grounded
1. DISPLAY SYSTEM CON- set to NO DLY for no delay. (If outlet and turn it on. With a Tek-
TROLS: Set the AUTO INTEN- you ’re using a 2275, move the tronix 2200 scope, there ’s no
SITY control at midrange (about switch to the A sweep position.) need to change the scope's
halfway from either stop). Turn Rotate the SEC/DIV switch to power supply settings to match
the AUTO FOCUS knob com- 0.5 millisecond (0.5 ms). Make the local power line; the scopes
pletely clockwise. sure the red VAR (variable) operate on main power from 90
2. VERTICAL SYSTEM CON- switch in the center of the knob to 250 Vat at 48 to 62 Hz.
TROLS: Turn the channel 7 PO- is in its detent position by mov-
SITION control completely ing it completely clockwise. And
counterclockwise. Make sure push in on the VAR switch to
the lefthand VERTICAL MODE make sure the scope is not in a
switch is set to CH 7. Move both magnified mode.
channel VOLTS/D/V switches to 4. TRIGGER SYSTEM CON-
the least sensitive setting by TROLS: Make sure the VAR
rotating them completely coun- HOLDOFF controlis set to its full
terclockwise. And make sure counterclockwise position. Set
the center, red VAR controls are the trigger MODE switch (2275:
locked in their detents at the ex- A TRIGGER MODE) on AUTO.
treme clockwise position. Input And move the trigger SOURCE
coupling switches should be set switch (A SOURCE on a 2275) to
to GND. /NT (internal) and the INT selec-
tion switch (A&B INT on a 2275)
to CH 7.
CHAPTER 1. THE DISPLAY SYSTEM

The oscilloscope draws a graph the path of the beam. A grid of Common controls for display Intensity
by moving an electron beam lines etched on the inside of the systems include intensity and An intensity control adjusts the
across a phosphor coating on faceplate serves as the refer- focus; less often, you will also brightness of the trace. It’s
the inside of the CRT (cathode- ence for your measurements; find beam finder and trace rota- necessary because you use a
ray tube). The result is a glow for this is the graticule shown in tion controls. On a Tektronix scope in different ambient light
a short time afterward tracing Figure 2. 2200 Series instrument they are conditions and with many kinds
all present, grouped next to and of signals. For instance, on
on the right of the CRT At the top square waves you might want to
of the group is the intensity con- change the trace brightness to
trol (labeled AUTO INTENSITY look at different parts of the
CENTER on a 2200 because these in- waveform because the slower

I
GWTICUE
LINES - struments automatically main-
tain the trace intensity once it is
set). The TRACE ROTATION ad-
horizontal components will al-
ways appear brighter than the
faster vertical components.
justment is under that, and then Intensity controls are also use-
the beam finder (BEAM FIND). ful because the intensity of a
Under the probe adjustment trace is a function of both how
jack is the focus control (labeled bright the beam is and how long
AUTO FOCUS because it’s also it’s on-screen. As you select dif-
automatic). The functions of ferent sweep speeds (a sweep
these controls are described is one movement of the electron
below and their positions on the beam across the scope screen)
Tektronix 2213 Portable Oscillo- with the SEC/DIV switch, the
scope are illustrated by the fold- beam ON and OFF times
out illustration at the rear of the change and the beam has more
booklet. or less time to excite the phos-
phor
Beam Finder On most scopes, you have to
The beam finder is a conven- turn the intensity up or down to
ience control that allows you to restore the initial brightness. On
locate the electron beam any- the 2200 Series scopes, how-
time it’s off-screen. When you ever the automatic intensity cir-
push the BEAM FIND button, cuit compensates for changes
you reduce the vertical and in sweep speed from 0.5 milli-
horizontal deflection voltages seconds (0.5 ms) to 0.5 micro-
(more about deflection voltages seconds (0.5 ,us). W ithin this
later) and over-ride the intensity range, the automatic circuit
control so that the beam always maintains the same trace inten-
appears within the 8 x 10- sity you initially set with the
centimeter screen. When you AUTO INTENSITY control.
see which quadrant of the
screen the beam appears in, Focus
you’ll know which directions to The scope’s electron beam is
turn the horizontal and vertical focused on the CRT faceplate
POSITION controls to reposition by an electrical grid within the
Figure 2. the trace on the screen for nor- tube. The focus control adjusts
THE GRATICLJLE is a grid of lines typically etched or silk-screened on the inside of mal operations. that grid for optimum trace
the CRT faceplate. Putting the graticule inside-on the same plane as the trace
drawn by the electron beam. and not on the outside of the glass - eliminates
focus. On a 2200 scope, the
measurement inaccuracies called parallax errors. Parallax errors occur when the AUTO FOCUS circuit maintains
trace and the graticule are on different planes and the observer is shifted slightly from your focus settings over most
the direct line of sight. Though different sized CRT’s may be used, graticules are intensity settings (0.5 ms to
usually laid out in an 8 x 10 pattern. Each of the eight vertical and ten horizontal lines
block off major divisions (or just divisions) of the screen. The labeling on scope
0.5 us).
controls always refers to major divisions. The tick marks on the center graticule lines
represent minor divisions or subdivisions. Since rise time measurements are very
common. 2200 Series scope graticules include rise time measurement markings:
dashed lines for 0 and 100% points, and labeled graticule lines for 10 and 90%
PART I

Trace Rotation
Another display control you’ll
find on the front panel of a 2200
Series instrument is TRACE RO-
TATION. The trace rotation ad-
justment allows you to electri-
cally align the horizontal deflec-
tion of the trace with the fixed
graticule. To avoid accidental
misalignments when the scope
is in use, the control is recessed
and must be adjusted with a
small screwdriver.
If this seems like a calibration
item that should be adjusted
once and then forgotten, you’re
right; that’s true for most oscillo-
scope applications. But the
earth’s magnetic field affects
Figure 3.
the trace alignment and when a THE DISPLAY SYSTEM of your scope
scope is used in many different consists of the cathode-ray tube and its
positions - as a service scope controls. To draw the graph of your mea-
will be - it’s very handy to have surements, the vertical system of the
scope supplies the Y, or vertical, coordi-
a front panel trace rotation ad- nates and the horizontal system
justment. supplies the X coordinates. There is also
a Z dimension in a scope; it determines
Using the Display Controls whether or not the electron beam is
The display system and its con- turned on, and how bright it is when it’s

l-x---/
trols are shown as functional on.
blocks in Figure 3. Use Exercise
2 to review the display controls.

Exercise 2. THE DISPLAY SYSTEM CONTROLS


In Exercise 7 you initialized your 2. AUTO FOCUS: The trace you to use the vertical POSITION
scope and turned on the power. have on the screen now should control again to align the trace
Now you can find the display be out of focus. Make it as sharp on the graticule line.)
system controls labeled on the as possible with the AUTO You have used all the scope’s
foldout front panel illustration FOCUS control. display sys tern controls. If at the
and use them as you follow 3. AUTO INTENSITY Set the end of one of these chapters
these instructions. brightness at a level you like, you’re not going to go on imme-
1. BEAM FIND: Locate theposi- 4. Vertical POSITION: Now you diately, be sure to turn your
tion of the electron beam by can use the vertical POSITION scope off.
pushing and holding in the control to line up the trace with
BEAM FIND button; then use the the first major division line above
channel 7 vertical POSITION the center of the graticule.
knob to position the trace on the
center horizontal graticule line. 5. TRACE ROTATION: Use a
Keep BEAM FIND depressed small screwdriver and the
and use the horizontal POSI- TRACE ROTATION control to ro-
TION control to center the trace. tate the trace in both directions.
Release the beam finder. When you finish, align the trace
parallel to the horizontal division
line closest to it. (After setting
the trace rotation, you may have

5
CHAPTER 2. THE VERTICAL SYSTEM

The vertical system of your Because all 2200s are two- Input Coupling back is a handy way to measure
scope supplies the display sys- channel scopes, you will have The input coupling switch for signal voltage levels with re-
tem with the Y axis - or vertical one set of these switches for each vertical channel lets you spect to chassis ground. (Using
- information for the graph on each channel. There are also control how the input signal is the GND position does not
the CRT screen. To do this, the two switches for choosing the coupled to the vertical channel. ground the signal in the circuit
vertical system takes the input scope’s vertical display mode DC (the abbreviation normally you’re probing.)
signals and develops deflection and one control that allows you stands for direct current) input
voltages. The display system to invert the polarity of the chan- coupling lets you see all of an Vertical Sensitivity
then uses the deflection volt- nel 2 signal. input signal. AC (alternating cur- A volts/division rotary switch
ages to control - deflect-the For the exercises in this chap- rent) coupling blocks the con- controls the sensitivity of each
electron beam. ter, you’ll need a 10X probe like stant signal components and vertical channel. Having differ-
The vertical system also gives the Tektronix P6120 10X Probes permits only the alternating ent sensitivities extends the
you a choice of how you connect supplied with every 2200 Series components of the input signal range of the scope’s applica-
the input signals (called cou- scope. to reach the channel. An illustra- tions; with a VOLTS/DIV switch,
pling and described below). tion of the differences is shown a multipurpose scope is capa-
And the vertical system pro- Vertical Position in Figure 5. ble of accurately displaying sig-
vides internal signals for the Your scope’s POSITION controls The middle position of the nal levels from millivolts to many
trigger circuit (described in let you place the trace exactly coupling switches is marked volts.
Chapter 4). Figure 4 illustrates where you want it on the screen. GND for ground. Choosing this Using the volts/division
the vertical system schemati- The two vertical POSITION con- position disconnects the input switch to change sensitivity also
cally. trols (there’s one for each chan- signal from the vertical system changes the scale factor, the
nel) change the vertical place- and makes a triggered display value of each major division on
Some of the vertical system ment of the traces from each show the scope’s chassis the screen. Each setting of the
controls - see the foldout front vertical channel; the horizontal ground. The position of the trace control knob is marked with a
panel illustration for their loca- POSITION control changes the on the screen in this mode is the number that represents the
tions - are: vertical position, horizontal position of both ground reference level. Switch- scale factor for that channel. For
sensitivity, and input coupling. channels at once. ing from AC or DC to GND and example, with a setting of 10 V,

N’lENJAl~
“L-

Figure 4.
THE VERTICAL SYSTEM of a Tektronix 2200 Series scope consists of two identical
channels though only one is shown in the drawing. Each channel has circuits to
couple an input signal to that channel, attenuate (reduce) the input signal when
necessary, preamplify it, delay it, and finally amplify the signal for use by the display
system. The delay line lets you see the beginning of a waveform even when the
scope is triggering on it.

6
PART I

each of the eight vertical major


divisions represents 10 volts and
the entire screen can show 80
volts from bottom to top. With a
VOLTSDIV setting of 2 milli-
volts, the screen can display 16
mV from top to bottom.
If you pronounce the "/ ” in
VOLTS/DIV as “per” when you
read the setting, then you’ll re-
member the setting is a scale
factor; for example, read a 20
mV setting as “20 millivolts per
division."
The probe you use influences
the scale factor. Note that there
are two unshaded areas under
the skirts of the VOLTSDIV
switches. The right-hand area
shows the scale factor when you
use the standard 1 OX probe. The
left area shows the factor for a IX
probe.
Variable VOLTS/DIV
The red VAR (for variable) con-
trol in the center of the VOLTS/
DIV switch provides a continu-
ously variable change in the
scale factor to a maximum
greater than 2.5 times the
VOLTS/DIV setting. Figure 5.
A variable sensitivity control is VERTICAL CHANNEL INPUT COUPLING CONTROLS let you choose AC and DC
input coupling and ground. DC coupling connects the entire input signal to the
useful when you want to make vertical channel. AC coupling blocks constant signal components and only con-
quick amplitude comparisons nects alternating components to the vertical channel. The GND position disconnects
on a series of signals. You could, the input signal and shows you the scope’s chassis ground level. AC coupling is
for example, take a known sig- handy when the entire signal (alternating plus constant components) might be too
large for the VOLTSDIV switch settings you want. In a case like this, you might see
nal of almost any amplitude and something like the first photo. But eliminating the direct component allows you to look
use the VAR control to make at the alternating signal with a VOLTSDIV setting that is more convenient as in the
sure the waveform fits exactly on second photo.
major division graticule lines.
Then as you used the same ver-
tical channel to look at other sig- Vertical Operating Modes To make the scope display a sweep on channel 1, then a
nals, you could quickly see Scopes are more useful if they only channel 1, use the CH 1 po- sweep on channel 2, and so on.
whether or not the later signals have more than one vertical dis- sition on the left-hand switch. To display both channels in
had the same amplitude. play mode, and with your Tek- To display only channel 2, use the chop mode, you move the
tronix 2200, you have several the CH 2 position on the left- left-hand switch to BOTH and
Channel 2 Inversion
controlled by two VERTICAL hand switch. the right-hand one to CHOP In
To make differential measure-
MODE switches: channel 1 To see both channels in the the chop mode, the scope
ments (described in Part II), you
alone; channel 2 alone; both alternate vertical mode, move draws small parts of both sig-
have to invert the polarity of one
channels in either the alternate the left-hand switch to BOTH nals by switching back and forth
of your input channels. The IN-
or chopped mode; and both (which enables the right-hand at a fast fixed rate while your
VERT control on the vertical
channels algebraically switch) and then move the eyes fill in the gaps.
amplifier for channel 2 provides
summed. right-hand switch to ALT Now
this facility. When you push it in,
you can see both channels
the signal on channel 2 is in-
since the signals are drawn al-
verted. When the switch is out,
ternately. The scope completes
both channels have the same
polarity.
THE VERTICAL SYSTEM CONT.

Both chop and alternate are


provided so that you can look at
two signals at any sweep speed.
The alternate mode draws first
one trace and then the other, but
not both at the same time. This
works great at the faster sweep
speeds when your eyes can’t
see the alternating. To see two
signals at the slower sweeps,
you need the chop mode.
If you want to see the two input
signals combined into one
waveform on the screen, use
BOTH on the left-hand and ADD
on the right-hand switch. This
gives you an algebraically-
combined signal: either channel
1 and 2 added, (CH 1) + (CH 2);
or channel 1 minus channel 2
when channel 2 is inverted,
(+CH 1) + (-CH 2).
Alternate Sweep Separation
On the 2215 dual time base
scope, there is also a sweep
separation control: A/B SWP
SEP. It’s used to change the po-
sition of the scope’s B sweep Figure 6.
traces with respect to the A THE TEKTRONIX P6120 10X PROBE connects to the BNC connector of either chan-
sweeps. Using the A/B sweep nel 1 (shown) or 2; unlike the photo, the probe’s ground strap is usually connected to
the ground of the circuit you are working on. The probe adjustment jack is labeled
separation in conjunction with PROBE ADJUST and is located near the CRT controls on the front panel.
the vertical POSITION controls
lets you place all four traces (two
channels and two time bases)
Using the Vertical Controls l both VAR VOLTSDIV switches collar of the channel 2 BNC
on the screen so that they don’t
overlap. (Dual time base scope Before using the vertical system in their detents at the extreme connector as shown in Figure 6.
measurements are described in controls, make sure all the con- clockwise position; Use the callouts on the foldout
Chapter 9.) trols are positioned where you _ l input coupling levers in GND; figure to remind yourself of the
left them at the end of the last l VERTICAL MODE is CH 1; control locations and follow the
chapter: directions in Exercises 3 to re-
l and HORIZONTAL MODE is
l AUTO INTENSITY and AUTO
NO DLY (2215 mode is A). view the vertical system con-
FOCUS set for a bright, crisp trols.
trace; Now connect your 10X probe
on the channel 1 BNC connector
l trigger SOURCE (A SOURCE
on the front panel of your scope.
on the 2215) switch on INT and (BNC means “bayonet Neill-
the INT (2215: A&B INT) switch Concelman”; named for Paul
on CH 1; Neill, who developed the N
l trigger MODE (2215: A TRIG- Series connector at Bell Labs,
GER MODE) switch on AUTO; and Carl Concelman, who
l trigger VAR HOLDOFF control developed the C Series con-
in its extreme counterclock- nector.)
wise position; Put the tip of the probe into the
l SEC/DIV (2215: A and B
PROBE ADJ jack. Probes come
SECDIV) switch to 0.5 ms; with an alligator-clip ground
l both channel VOLTSDIV strap that’s used to ground the
switches on 100 V (10X probe probe to the circuit-under-test.
reading); Clip the ground lead onto the
8
PART I

Exercise 3. VERTICAL SYSTEM CONTROLS


Compensating Your Probe 2. Turn the VOLTS/D/V switch was all negative-going with re- 3. Turn the SEC/DIV switch back
1. Turn on the scope and move two more click stops to the right. spect to ground, because in DC, to 0.5 ms and switch to CHOP as
the CH 7 VOLTS/D/V switch The channel 7 scale factor is all signal components are con- your vertical mode. The display
clockwise to 0.5 V; remember now 0.7 volts /division, and the nected to the vertical channel. looks a lot like the alternate
the P6720 is a 10X probe, so use signal -still half a volt -is now mode, but the way it’s achieved
about five major divisions in The Vertical Mode Controls is entirely different. In alternate,
the VOLTS/D/V readout to the
amplitude. 1. So far you’ve been using your you saw that one channel’s sig-
right.
scope to see what channel 7 can nal was completely written be-
2. Switch the channel 7 input 3. Turn the VAR VOLTS/D/V con- tell you, but that’s only one of
coupling to AC. trol to the left. That will take it out fore the other started. When
many possible vertical modes. you’re looking at slow signals
3. If the signal on the screen isn’t of its calibrated detent position Look at the trace for channel 2
and let you see its effect, Since it with your scope, that can be a
steady, turn the trigger LEVEL (A by moving the scope’s left-most bother because only one trace
TRIGGER LEVEL on the 2275) reduces the scale factor 22%~ VERTICAL MODE switch to CH
times, the signal should be less at a time will be on-screen. In the
control until the signal stops 2. The input coupling for chan- chop mode, however, the scope
moving and the TRIG’D light is than two major divisions in nel 2 should still be GND at this
amplitude with this control all the switches back and forth very
on. (Use the AUTO FOCUS con- point, so what you’ll see is the quickly between the two traces
trol if you think you can get the way to the left. If it isn’t exactly ground reference line. Line up
that, don’t worry The variable so that a little part of each is
signal sharper, and AUTO IN- this trace with the graticule line drawn before going on to the
TENSITY to adjust the bright- volts/division controls are used second from the top of the
to compare signals, not make next. When you look at the
ness .) screen with the channel2 screen, both signals seem con-
amplitude measurements, and POSITION control.
4. Next, compensate your consequently the exact range of tinuous because the scope is
probe. There’s a screwdriver ad- variation isn’t critical. Return the 2. Now move the lever on the “chopping”back and forth at a
justment on the compensation variable control to its detent. left-hand vertical mode switch to very fast rate - approximately
box at the base of the probe; BOTH. That lets you pick one of 250 kHz in the 2200 Series. You
turn it until the tops and bottoms Coupling The Signal the vertical modes controlled by can see the chopping if you pick
of the square wave on the 1. Switch your channel 7 input the right-hand side vertical a very fast sweep speed. Move
screen are flat. (There’s more in- coupling to GND and position mode switch; move it to ALT the SECIDIV switch to 70 p.
formation about probes and the trace on the Center graticule. You’ve just selected the alter- Now the display shows broken
compensation in Chapter 5.) Switch back to the AC coupling nate vertical mode. In this mode, lines because of the chopping.
position. Note that the waveform your scope alternates between CHOP is most useful for slow
Controlling Vertical Sensitivity is centered on the screen, Move the signals on channel 7 and 2, sweep speeds, and ALT for
1. The probe adjustment signal the CH 7 VOLTS/D/V switch drawing one complete sweep faster sweeps.
is a square wave of approxi- back to 0.5 volts and note that on channel 7 first, and then 4. Move the SECIDIV switch
mately 0.5 volts, and the scale the waveform is still centered drawing a complete sweep on back to 0.5 ms. There’s one
factor for channel 7 is now a around the zero reference line. channel 2. You can see this more verticalmode: ADD. In the
half-volt per division. At this set- 2. Switch to DC coupling. The happening when you slow down add mode, the two signals are
ting every major division on the top of the probe adjustment sig- the sweep speed, so move the algebraically summed (either
screen represents half a volt. nal should be on the center SEC/DIV swtich left to 0.7 sec- CH 1 + CH2, or CH1 - CH2 H2
Use the channel 7 vertical POSI- graticule line and the signal onds per division. Now you can when channel 2 is inverted). To
TION control to line up the bot- should reach to the next lower see the two dots from the AC- see it in operation, move the
tom edge of the waveform with major division. Now you can see coupled channel 7 move across right-hand VERTICAL MODE
the center graticule line. The the difference between AC and the screen for one sweep. Then switch to ADD. Now you can see
tops of the square wave should DC coupling. AC coupling the single dot from channel 2 will the combined signal roughly
be just touching the next major blocked the constant part of the move across the screen. The halfway between where the two
division line, proving the probe signal and just showed you a point is that in the alternate separate signals were.
adjustment signal is approxi- half-volt, peak-to-peak, square mode each channel is drawn
mately 0.5 volts. (Note that the wave centered on the zero ref- completely before the scope
probe adjustment signal is not a erence you set at the center of switches to the other channel.
critical circuit in the scope; this the screen. But the DC coupling
is why the square wave is ap- showed you that the constant
proximately 0.5 volts.) component of the square wave
CHAPTER 3. THE HORIZONTAL SYSTEM

To draw a graph, your scope


needs horizontal as well as ver-
tical data. The horizontal system
of your scope supplies the sec-
ond dimension by providing the
deflection voltages to move the
electron beam horizontally. And
the horizontal system contains a
sweep generator which pro-
duces a sawtooth waveform, or
ramp (see Figure 7), that is used
to control the scope’s sweep
rates.
It’s the sweep generator that
makes the unique functions of
the modern oscilloscope possi-
ble. The circuit that made the
rate of rise in the ramp linear- a
refinement pioneered by Tek-
tronix- this was one of the most Figure 7.
important advances in oscil- THE SAWTOOTH WAVEFORM is a voltage ramp produced by the sweep generator.
The rising portion of the waveform is called the ramp; the falling edge is the retrace;
Iography. It meant that the hori- and the time between ramps is the holdoff time. The sweep of the electron beam
zontal beam movement could across the screen of a scope is controlled by the ramp and the return of the beam to
be calibrated directly in units of the left side of the screen takes place during the retrace.
time. That advance made it pos-
sible for you to measure time be-
tween events much more accu- The horizontal system con- Horizontal Operating Modes Sweep Speeds
rately on the scope screen. trols of a Tektronix 2213 scope Single time base scopes usu- The seconds/division switch
Because it is calibrated in are shown in the foldout figure: ally have only one horizontal op- lets you select the rate at which
time, the sweep generator is the horizontal POSlTION control erating mode, but the 2213 the beam sweeps across the
often called the time base. It lets is near the top of the panel, and offers normal, intensified, or screen; changing SEC/DIV
you pick the time units, observ- the HORIZONTAL MODE con- delayed-sweep operating switch settings allows you to
ing the signal for either very trol is below it; the magnification modes. Dual time base scopes look at longer or shorter time
short times measured in and variable sweep speed con- like the 2215 usually let you intervals of the input signal. Like
nanoseconds or microseconds, trol is a red knob in the center of select either of two sweeps. The the vertical system VOLTS/DIV
or relatively long times of several the SEC/DIV switch; at the bot- A sweep is undelayed (like the switch, the control’s markings
seconds. tom of the column of horizontal sweep of a single time base in- refer to the screen’s scale fac-
system controls are the DELAY strument), while the B sweep is tors. If the SEC/DIV setting is 1
TIME switch and the delay time started after a delay time. Addi- ms, that means that each hori-
MULTIPLIER. The dual time tionally, some scopes with two zontal major division represents
base 2215 has two concentric time bases- and the 2215 is an 1 ms and the total screen will
SEC/DIV controls, and a B example again - let you see the show you IO ms.
DELAY TIME POSITION control two sweeps at once: the A On the 2215, which has two
instead of the delay time switch sweep intensified by the B time bases, there are two SEC/
and multiplier. (The scope con- sweep; and the B sweep itself. DIV controls. The A sweep offers
trols that you use to position the. This is called an alternate hori- all the settings described below;
start of a delayed sweep are zontal operating mode. the SEC/DIV switch for the de-
also often called delay time mul- Only the normal horizontal layed B sweep has settings for
tipliers or DTMs.) operating mode is used in these 0.05 psldiv to 50 ms/div.
Horizontal Position first few chapters, so leave your
Like the vertical POSITION con- scope’s HORIZONTAL OPERA-
trols, you use the horizontal PO- TING MODE switch in NO DLY
SITION control to change the lo- (no delay) on the 2213 and A (for
cation of the waveforms on the A sweep only) on the 2215.
screen. Chapter 9, in the second section
of this primer, describes how to
make delayed sweep mea-
surements.
PART I

All the instruments of the Tek- Horizontal Magnification The 10X magnification is use- The B DELAY TIME POSITION
tronix 2200 Series offer sweep Most scopes offer some means ful when you want to look at sig- Control
speeds from a half-second for of horizontally magnifying the nals and see details that occur This calibrated IO-turn dial is
each division to 0.05 us / waveforms on the screen. The very closely together in time. used to position the beginning
division. The markings appear- effect of magnification is to mul- of the B sweep relative to the A
ing on the scopes are: tiply the sweep speed by the The DELAY TIME and sweep in a 2215. Its uses are
amount of magnification. On MULTIPLIER Controls described under “Delayed
.5s half a second
.2s 0.2 second 2200 Series scopes there is a This switch and dial are used in Sweep Measurements” in
.l s 0.1 second
10X horizontal magnification conjunction with either the in- Chapter 9.
50 ms 50 milliseconds
(0.05 second) that you engage by pulling out tensified or delayed-sweep
20 ms 20 milliseconds on the red VAR switch. The 10X horizontal operating modes in Using the Horizontal Controls
(0.02 second)
horizontal magnification gives the 2213. These features are de- As you can see in Figure 8, the
10 ms 10 milliseconds
(0.01 second) you a sweep speed ten times scribed later under “Delayed horizontal system can be di-
5 ms 5 milliseconds
faster than the SEC/DIV switch Sweep Measurement” in vided into two functional blocks:
(0.005 second)
2 ms 2 milliseconds setting; for example, 0.05 psi Chapter 9. the horizontal amplifier and the
(0.002 second)
division magnified is a very fast sweep generator.
1 ms 1 millisecond
(0.001 second) 5-nanosecond/division sweep.
.5 ms half a millisecond
(0.0005 second)
.2 ms 0.2 millisecond
(0.0002 second)
.l ms 0.1 millisecond
(0.0001 second)
50 I_CS 50 microseconds
(0.00005 second)
20 ps 20 microseconds
(0.00002 second)
10 ps 10 microseconds
(0.00001 second)
5 ps 5 microseconds
(0.000005 second)
2 ps 2 microseconds
(0.000002 second)
1 ps 1 microsecond
(0.000001 second)
.5 ps half a microsecond
(0.0000005 second)
.2 ps 0.2 microsecond
(0.0000002 second)
.1 ps 0.1 microsecond
(0.0000001 second)
.05 ps 0.05 microseconds
(0.00000005 second)

Scopes also have an XY set-


ting on the SEC/DIV switch for
making the X-Y measurements
described in Chapter 9.
Variable SEC/DIV
Besides the calibrated speeds,
you can change any sweep
speed by turning the red VAR
control in the center of the
SEC/DIV switch counterclock-
wise. This control slows the
sweep speed by at least 2.5:1,
making the slowest sweep you
have 0.5 seconds x 2.5, or 1.25
seconds/division. Remember
that the detent in the extreme
clockwise direction is the cali-
Figure 8.
brated position. HORIZONTAL SYSTEM components include the sweep generator and the horizontal amplifier. The sweep generator produces a
sawtooth waveform that is processed by the amplifier and applied to the horizontal deflection plates of the CRT The horizontal
system also provides the Z axis of the scope; the Z axis determines whether or not the electron beam is turned on -and how bright
it is when it’s on.

11
PART I

All the instruments of the Tek- Horizontal Magnification The 10X magnification is use- The B DELAY TIME POSITION
tronix 2200 Series offer sweep Most scopes offer some means ful when you want to look at sig- Control
speeds from a half-second for of horizontally magnifying the nals and see details that occur This calibrated IO-turn dial is
each division to 0.05 us / waveforms on the screen. The very closely together in time. used to position the beginning
division. The markings appear- effect of magnification is to mul- of the B sweep relative to the A
ing on the scopes are: tiply the sweep speed by the The DELAY TIME and sweep in a 2215. Its uses are
amount of magnification. On MULTIPLIER Controls described under “Delayed
.5s half a second
.2s 0.2 second 2200 Series scopes there is a This switch and dial are used in Sweep Measurements” in
.l s 0.1 second
10X horizontal magnification conjunction with either the in- Chapter 9.
50 ms 50 milliseconds
(0.05 second) that you engage by pulling out tensified or delayed-sweep
20 ms 20 milliseconds on the red VAR switch. The 10X horizontal operating modes in Using the Horizontal Controls
(0.02 second)
horizontal magnification gives the 2213. These features are de- As you can see in Figure 8, the
10 ms 10 milliseconds
(0.01 second) you a sweep speed ten times scribed later under “Delayed horizontal system can be di-
5 ms 5 milliseconds
faster than the SEC/DIV switch Sweep Measurement” in vided into two functional blocks:
(0.005 second)
2 ms 2 milliseconds setting; for example, 0.05 psi Chapter 9. the horizontal amplifier and the
(0.002 second)
division magnified is a very fast sweep generator.
1 ms 1 millisecond
(0.001 second) 5-nanosecond/division sweep.
.5 ms half a millisecond
(0.0005 second)
.2 ms 0.2 millisecond
(0.0002 second)
.l ms 0.1 millisecond
(0.0001 second)
50 I_CS 50 microseconds
(0.00005 second)
20 ps 20 microseconds
(0.00002 second)
10 ps 10 microseconds
(0.00001 second)
5 ps 5 microseconds
(0.000005 second)
2 ps 2 microseconds
(0.000002 second)
1 ps 1 microsecond
(0.000001 second)
.5 ps half a microsecond
(0.0000005 second)
.2 ps 0.2 microsecond
(0.0000002 second)
.1 ps 0.1 microsecond
(0.0000001 second)
.05 ps 0.05 microseconds
(0.00000005 second)

Scopes also have an XY set-


ting on the SEC/DIV switch for
making the X-Y measurements
described in Chapter 9.
Variable SEC/DIV
Besides the calibrated speeds,
you can change any sweep
speed by turning the red VAR
control in the center of the
SEC/DIV switch counterclock-
wise. This control slows the
sweep speed by at least 2.5:1,
making the slowest sweep you
have 0.5 seconds x 2.5, or 1.25
seconds/division. Remember
that the detent in the extreme
clockwise direction is the cali-
Figure 8.
brated position. HORIZONTAL SYSTEM components include the sweep generator and the horizontal amplifier. The sweep generator produces a
sawtooth waveform that is processed by the amplifier and applied to the horizontal deflection plates of the CRT The horizontal
system also provides the Z axis of the scope; the Z axis determines whether or not the electron beam is turned on -and how bright
it is when it’s on.

11
.

THE HORIZONTAL SYSTEM CONT.

horizontal system controls, fol-


Exercise 4. THE HORIZONTAL SYSTEM CONTROLS
low the directions in Exercise 4
and refer to the foldout for con- 1. Switch the VERTICAL MODE 3. Change sweeps to 0.2 ms, setting on the SEC/DIV switch
trol locations. First, make sure to CH 7 and the CH 7 VOLTS/D/V line up a rising edge with the will result in a sweep that’s ten
the front panel controls have setting to 0.5 volt. Be sure your vertical graticule on the left times faster; for example, the
these settings: probe is connected to channel 7 edge of the screen and count to sweep now is 0.05 msldivision,
and the PROBE ADJ jack. Turn the next rising edge. Because not 0.5 ms.
l the SEC/DIV switch is on 0.5 the switch was changed from 6. While your scope is magnify-
on vour scope and move the
ms; channel1 input coupling lever to 0.5 to 0.2 ms, the waveform will ing the probe adjustment signal,
l the trigger SOURCE (A TRIG-
GND and center the signal on look 2.5 times as long as before. use the horizontal POSITION
GER SOURCE on the 2215) is Of course, the signal hasn’t control. Its range is now mag-
the screen with the POSITION
INT; INT (2215: A&B INT) is on control. Switch to AC coupling. changed, only the scale factor. nified as well, and the combina-
CH 1; 2. Now you can use the horizon- 4. In the middle of the SEC/D/V tion of magnified signal and
l the trigger MODE (2215: A
tal system of your scope to look switch is the red variable con- POSITION control gives you the
TRIGGER MODE) is AUTO; at the probe adjustment signal. trol; in its counterclockwise de- ability to examine small parts of
l the channel 2 INVERT switch is
Move the waveform with the tent, the settings of the SEC/DIV a waveform in great detail. Re-
out (no signal inverting); switch are calibrated. Move the turn your scope to its normal
horizontal POSITION control
l and HORIZONTAL MODE is
until one rising edge of the control from its detent to see its sweep speed range by pushing
NO DLY (A on the 2215) effect on the sweep speed. Note the VAR switch in.
waveform is lined up with the
center vertical gra ticule. that now the cycles of the wave-
Examine the screen to see form are approximately two-
where the leading edge of the and-a-half times smaller. Return
next pulse crosses the horizon- the VAR control to its detent.
tal center line of the graticule. 5. Move the SEC/DIVswitch to
Count major and minor gra ticule 0.5 ms and then pull out the red
markings along the center hori- VAR control. This gives you a
zontal graticule and remember 10X magnification of the sweep
the number. speed. In other words, every

CHAPTER 4. THE TRIGGER SYSTEM


So far you’ve found that the dis- The when is the trigger and it’s ing 1 graph every 0.5 ps (0.05 ger signal to select with the
play system draws the wave- important for a number of rea- psldivision times ten screen di- source switches. With an exter-
forms on the screen, the vertical sons. First, because getting visions). That’s 2,000,OOO nal signal, you connect the trig-
system supplies the vertical in- time-related information is one graphs every minute (not count- ger signal to the trigger system
formation for the drawing, and of the reasons you use a scope. ing retrace and holdoff times, circuit with the external coupling
the horizontal system provides Equally important is that each which we’ll get to shortly). Imag- controls. Next you set the trigger
the time axis. In other words, you drawing start with the same ine the jumble on the screen if circuit to recognize a particular
know how the oscilloscope “when." each sweep started at a differ- voltage level on the trigger sig-
draws a graph; the only thing Obviously the graph drawn on ent place on the signal. nal with the slope and level con-
missing is the “when”: when the screen isn’t the same one all But each sweep does start at trols Then everytime that level
should the other circuits of your the time you’re watching. If the right time - if you make the occurs, the sweep generator is
scope start drawing the signal, you’re using the 0.05 ps SEC/ right trigger system control set- turned on. The process is dia-
and when shouldn’t they? DIV setting, the scope is draw- tings. Here’s how it’s done. You grammed in Figure 9.
tell the trigger circuit which trig-

12
.

THE HORIZONTAL SYSTEM CONT.

horizontal system controls, fol-


Exercise 4. THE HORIZONTAL SYSTEM CONTROLS
low the directions in Exercise 4
and refer to the foldout for con- 1. Switch the VERTICAL MODE 3. Change sweeps to 0.2 ms, setting on the SEC/DIV switch
trol locations. First, make sure to CH 7 and the CH 7 VOLTS/D/V line up a rising edge with the will result in a sweep that’s ten
the front panel controls have setting to 0.5 volt. Be sure your vertical graticule on the left times faster; for example, the
these settings: probe is connected to channel 7 edge of the screen and count to sweep now is 0.05 msldivision,
and the PROBE ADJ jack. Turn the next rising edge. Because not 0.5 ms.
l the SEC/DIV switch is on 0.5 the switch was changed from 6. While your scope is magnify-
on vour scope and move the
ms; channel1 input coupling lever to 0.5 to 0.2 ms, the waveform will ing the probe adjustment signal,
l the trigger SOURCE (A TRIG-
GND and center the signal on look 2.5 times as long as before. use the horizontal POSITION
GER SOURCE on the 2215) is Of course, the signal hasn’t control. Its range is now mag-
the screen with the POSITION
INT; INT (2215: A&B INT) is on control. Switch to AC coupling. changed, only the scale factor. nified as well, and the combina-
CH 1; 2. Now you can use the horizon- 4. In the middle of the SEC/D/V tion of magnified signal and
l the trigger MODE (2215: A
tal system of your scope to look switch is the red variable con- POSITION control gives you the
TRIGGER MODE) is AUTO; at the probe adjustment signal. trol; in its counterclockwise de- ability to examine small parts of
l the channel 2 INVERT switch is
Move the waveform with the tent, the settings of the SEC/DIV a waveform in great detail. Re-
out (no signal inverting); switch are calibrated. Move the turn your scope to its normal
horizontal POSITION control
l and HORIZONTAL MODE is
until one rising edge of the control from its detent to see its sweep speed range by pushing
NO DLY (A on the 2215) effect on the sweep speed. Note the VAR switch in.
waveform is lined up with the
center vertical gra ticule. that now the cycles of the wave-
Examine the screen to see form are approximately two-
where the leading edge of the and-a-half times smaller. Return
next pulse crosses the horizon- the VAR control to its detent.
tal center line of the graticule. 5. Move the SEC/DIVswitch to
Count major and minor gra ticule 0.5 ms and then pull out the red
markings along the center hori- VAR control. This gives you a
zontal graticule and remember 10X magnification of the sweep
the number. speed. In other words, every

CHAPTER 4. THE TRIGGER SYSTEM


So far you’ve found that the dis- The when is the trigger and it’s ing 1 graph every 0.5 ps (0.05 ger signal to select with the
play system draws the wave- important for a number of rea- psldivision times ten screen di- source switches. With an exter-
forms on the screen, the vertical sons. First, because getting visions). That’s 2,000,OOO nal signal, you connect the trig-
system supplies the vertical in- time-related information is one graphs every minute (not count- ger signal to the trigger system
formation for the drawing, and of the reasons you use a scope. ing retrace and holdoff times, circuit with the external coupling
the horizontal system provides Equally important is that each which we’ll get to shortly). Imag- controls. Next you set the trigger
the time axis. In other words, you drawing start with the same ine the jumble on the screen if circuit to recognize a particular
know how the oscilloscope “when." each sweep started at a differ- voltage level on the trigger sig-
draws a graph; the only thing Obviously the graph drawn on ent place on the signal. nal with the slope and level con-
missing is the “when”: when the screen isn’t the same one all But each sweep does start at trols Then everytime that level
should the other circuits of your the time you’re watching. If the right time - if you make the occurs, the sweep generator is
scope start drawing the signal, you’re using the 0.05 ps SEC/ right trigger system control set- turned on. The process is dia-
and when shouldn’t they? DIV setting, the scope is draw- tings. Here’s how it’s done. You grammed in Figure 9.
tell the trigger circuit which trig-

12
THE TRIGGER SYSTEM PART I

Figure 9.
TRIGGERING GIVES YOU A STABLE DISPLAY because the same trigger point starts
the sweep each time. Slope and level controls define the trigger points on the trigger
signal. When you look at a waveform on the screen, you’re seeing all those sweeps
overlaid into what appears to be one picture.

Instruments like those in the trols the trigger sources and the
Tektronix 2200 Portable Oscillo- external trigger coupling. At the
scope family offer a variety of bottom of the column of trigger
trigger controls. Besides those controls is the external trigger
already mentioned, you also input BNC connector.
have controls that determine On dual time base 2215
how the trigger system operates scopes, there is a slightly differ-
(trigger operating mode) and ent control panel layout be-
how long the scope waits be- cause you can have a separate
tween triggers (holdoff). trigger for the B sweep.
The control positions are illus-
Trigger Level and Slope
trated by the foldout at the end
of the primer. All are located on These controls define the trigger
point. The SLOPE control de-
the far right of the front panel. On
termines whether the trigger
the 2213, the variable trigger
point is found on the rising or the
holdoff (VAR HOLDOFF) is at the
top, and immediately below it is falling edge of a signal. The
LEVEL control determines Figure 10.
the trigger MODE switch. Below
where on that edge the trigger SLOPE AND LEVEL CONTROLS determine where on the trigger signal the trigger
that the trigger SLOPE and actually occurs. The SLOPE control specifies either a positive (also called the rising
LEVEL controls are grouped. point occurs. See Figure IO.
orpositive-going) edge or on a negative (falling or negative-going) edge. The LEVEL
Then a set of three switches con- control allows you to pick where on the selected edge the trigger event will take
place.

13
THE TRIGGER SYSTEM CONT.

Variable Trigger Holdoff


Not every trigger event can be
accepted as a trigger. The trig-
ger system will not recognize a
trigger during the sweep or the
retrace, and for a short time af-
terward called the ho/doff
period. The retrace, as you re-
member from the last chapter, is
the time it takes the electron
beam to return to the left side of
the screen to start another
sweep. The holdoff period pro-
vides additional time beyond
the retrace that is used to ensure
that your display is stable, as
illustrated by Figure 11.
Sometimes the normal holdoff
period isn’t long enough to en-
sure that you get a stable dis-
play; this possibility exists when Figure 11.
TRIGGER HOLDOFF TIME ensures valid triggering. In the drawing only the labeled
the trigger signal is a complex points start the display because no trigger can be recognized during the sweep or
waveform with many possible the retrace and holdoff period. The retrace and holdoff times are necessary because
trigger points on it. Though the the electron beam must be returned to the left side of the screen after the sweep, and
waveform is repetitive’a simple because the sweep generator needs reset time. The CRT Z axis is blanked between
sweeps and unblanked during sweeps.
trigger might get you a series of
patterns on the screen instead
of the same pattern each time.
Digital pulse trains are a good
example; each pulse is very
much like any other, so there are Figure 12.
THE VARIABLE HOLDOFF CONTROL
many possible trigger points, lets you make the scope ignore some
not all of which result in the same potential trigger points. In the example,
display. all the possible trigger points in the input
What you need now is some signal would result in an unstable dis-
INW-f play. Changing the holdoff time to make
way to control when a trigger 5l6NA.L sure that the trigger point appears on
point is accepted. The variable the same pulse in each repetition of the
trigger holdoff control provides input signal is the only way to ensure a
the capability. (The control is ac- stable waveform.
tually part of the horizontal sys-
tem - because it adjusts the
holdoff time of the sweep
generator - but its function
interacts with the trigger con-
trols.) Figure 12 diagrams a situ-
ation where the variable holdoff
is useful.

Trigger Sources
Trigger sources are grouped
into two categories that depend
on whether the trigger signal is
provided internally or externally.
The source makes no difference
in how the trigger circuit oper-
ates’but internal triggering usu-
ally means your scope is trigger-
ing on the same signal that it is
14
PART I

displaying. That has the obvious on the front panel. Occasions on the channel. In the 2200 the scope triggers alternately on
advantage of letting you see when external triggering is use- Series, the automatic mode is a the two vertical channels. That
where you’re triggering. ful often occur in digital design signal-seeking auto mode. This means you can look at two com-
Two switches on the front and repair; there you might want means that for most of the sig- pletely unrelated signals. Most
panel (labeled SOURCE and to look at a long train of very nals you’ll be measuring, the scopes only trigger on one
INT) determine the trigger similar pulses while triggering auto mode will match the trigger channel or the other when the
source. The internal triggering with an external clock or with a level control to the trigger signal. two signals are not synchro-
sources are enabled when you signal from another part of the That makes it most unlikely that nous.
move the SOURCE lever to INT. circuit. you will set the trigger level con- Here’s a review of the 2200
In this position, you can trigger The LINE position on the trol outside of the signal range. trigger modes:
the scope on the signal from SOURCE switch gives you The auto mode lets you trigger
Trigger
either channel, or you can another triggering possibility: on signals with changing volt- Operating Mode Switch Settings
switch to VERT MODE. the power line. Line triggering is age amplitudes or waveshapes normal NORM on the MODE switch
Triggering on one of the useful anytime you’re looking at without making an adjustment of automatic AUTO on the MODE switch
channels works just like it circuits that are dependent on the LEVEL control. television field TV FIELD on the MODE
switch
sounds: you’ve set the scope to the power line frequency. Another useful operating television line NORM or AUTO on the
trigger on some part of the Examples include devices like mode is television triggering. MODE switch
vertical mode VERT MODE on the INT
waveform present on that light dimmers and power Most scopes with this mode let switch
channel. supplies. you trigger on tv fields at
Using the VERTICAL MODE These are all the trigger sweeps of 100 &division and
setting on the internal source source possibilities on a 2200 slower, and tv lines at 50 psldiv
switch means that the scope’s Series scope: or faster. With a 2200 Series
VERTICAL MODE switches de- Switch Positions
scope, you can trigger on either
termine what signal is used for Trigger Source SOURCE INT
fields or lines at any sweep
triggering. If the VERTICAL channel 1 only INT CH1
speed; for tv field triggering, use
MODE switches are set at CH 1, channel 2 only INT CH2 the TV FIELD switch position,
external EXT disabled and for television line triggering,
then the signal on channel 1 line LINE disabled
triggers the scope. If you’re look- vertical mode INT VERT MODE use the NORM or AUTO
ing at channel 2, then that chan- (either channel settings.
1 or 2 or both)
nel triggers it. If you switch to the You’ll probably use the normal
alternate vertical mode, then the Trigger Operating Modes and automatic modes the most
scope looks for triggers alter- The 2200 Series trigger circuits often. The AUTO because it’s
nately on the two channels. If the can operate in four modes: nor- essentially totally automatic,
vertical mode is ADD, then CH 1 mal, automatic, television, and and normal because it’s the
+ CH 2 is the triggering signal. vertical mode. most versatile. For example, it’s
And in the CHOP vertical mode, One of the most useful is the possible to have a low fre-
the scope triggers the same as normal trigger mode (marked quency signal with a repetition
in ADD, which prevents the in- NORM on the MODE switch) rate that is mismatched to the
strument from triggering on the because it can handle a wider run-out of an automatic mode
chop frequency instead of your range of trigger signals than any timer; when that happens the
signals. other triggering mode. The nor- signal will not be steady in the
You can see that vertical mal mode does not permit a auto mode. Moreover, the auto-
mode triggering is a kind of au- trace to be drawn on the screen matic signal-seeking mode
tomatic source selection that if there’s no trigger. The normal can’t trigger on very low fre-
you can use when you must mode gives you the widest quency trigger signals. The
switch back and forth between range of triggering signals: from normal mode, however, will give
vertical modes to look at differ- DC to 60 MHz. you a steady signal at any rep
ent signals. In the automatic (or “bright rate.
But triggering on the dis- baseline”) mode (labeled AUTO The last 2200 Series trigger
played signal isn’t always what on the front panel): a trigger operating mode, the vertical
you need, so external triggering starts a sweep; the sweep ends mode, is unique in its advan-
is also available. It often gives and the holdoff period expires. tages. Selecting the VERT
you more control over the dis- At that point a timer begins to MODE position on the INT
play. To use an external trigger, run; if another trigger isn’t found switch automatically selects the
you set the SOURCE switch to before the timer runs out, a trig- trigger source as you read in
its EXT position and connect the ger is generated anyway caus- “Trigger Source”above. It also
triggering signal to the BNC ing the bright baseline to appear makes alternate triggering pos-
connector marked EXT INPUT even when there is no waveform sible. In this operating mode,

15
THE TRIGGER SYSTEM CONT.

Triggering Coupling
Just as you may pick either al-
ternating or direct coupling
when you connect an input sig-
nal to your scope’s vertical sys-
tem, you can select the kind of
coupling you need when you
connect a trigger signal to the
trigger system’s circuits. For
internal triggers, the vertical
input coupling selects the trig-
ger coupling. For external trig-
ger signals, however, you must
select the coupling you want:
Figure 13.
Coupling Applications
THE TRIGGER CIRCUIT AND ITS CON-
TROLS are shown in the diagram above.
DC DC couples all elements of the
Trigger source describes whether or not
triggering signal (both AC and
DC) to the trigger circuit.
the trigger signal is internal or external to
the scope. Coupling controls the con-
DC with If you want DC coupling and nection of an external trigger to the trig-
attenuation the external trigger is too large ger circuit. The level and slope controls
for the trigger system, move determine where the trigger point will be
the TRIGGER COUPLING on the trigger signal. And the mode con-
switch to its DC+10 setting. trol determines the operations of the
AC This coupling blocks DC
trigger circuit.
components of the trigger
signal and couples only the AC
components.

Using the Trigger Controls


To review what you’ve learned Exercise 5. TRIGGER CONTROLS
about the trigger circuit and its 1. Move the trace to the right 4. You can also see the differ- a difference. Now change the
controls (shown schematically with the horizontal POSITION ence between the two triggering VERTICAL MODE and INT
in Figure 13), first make sure all control until you can see the modes by using channel 2, even switches back to CH 7, and
your controls are in these posi- beginning of the signal (you’ll with that channel coupled to switch to the NORM mode. Use
tions: probably have to increase the GND for ground. Change both the LEVEL control and notice
l 0.5 VOLTS/DIV on channel 1 intensity to see the faster vertical the vertical display mode and how much control range there
and VAR in its detent position; part of the waveform). Watch the the INT (2275:A&B INT) is. Now change the CH 7
l AC vertical coupling;
signal while you operate the switches to CH 2. With NORM VOLTS/D/V switch to 0.7 V and
l CH 1 on the VERTICAL MODE
SLOPE control. If you pick + , the triggering, there’s no signal; use the LEVEL control. There’s
switch; signal on the screen starts with a with AUTO, you’ll see the more control range now.
l 0.5 ms sweep speed and no
rising edge; the other SLOPE baseline. Try it. 6. The alternate-channel trigger-
magnification or variable control position makes the 5. Without a trigger signal ing with vertical mode triggering
SEC/DIV; scope trigger on a falling edge. applied to the EXT INPUT BNC can’t be demonstrated without
l your trigger settings should be 2. Now move the LEVEL control connector, it’s impossible to two unrelated signals on the
AUTO for MODE, INT for back and forth through all its show you the use of this trigger channels, but you’ll find it
SOURCE, and CH for INT travel; you’ll see the leading source, but the trigger MODE, useful the first time such an oc-
edge climb up and down the SLOPE, and LEVEL controls will casion comes up. You can take
Turn your scope on with the signal. The scope remains all operate the same for either another look at the difference
probe connected to the channel triggered because you are internal or external triggers. One between the normal and auto
1 BNC connector and the probe using the AUTO setting. difference between internal and trigger operating modes. Move
adjustment jack. Use the foldout external sources, however, is the LEVEL control slowly in the
3. Turn the MODE switch to
figure to remind yourself of the the sensitivity of the trigger cir- NORM mode until the scope is
NORM. Now when you use the
control locations and follow the cuit. All external sources are untriggered. Now switch the
LEVEL control to move the trig-
directions in Exercise 5. measured in voltage (say 750 trigger operating mode to AU TO
ger point, you’ll find places
where the scope is untriggered. millivolts) while the in ternal and note that the waveform is
This is an illustration of the es- sources are rated in divisions. In automatically triggered.
sential difference between nor- other words, for internal signals,
mal and automatic triggering. the displayed amplitude makes

16
CHAPTER 5. ALL ABOUT PROBES PART I

Connecting all the measure- plus the cable plus the scope’s
ment test points you’ll need to own input, the 10X attenuator
the inputs of your oscilloscope is probe introduces about ten
best done with a probe like the times less capacitance, as little
one illustrated in Figure 14. as IO-14 picofarads (pF). The
Though you could connect the penalty is the reduction in signal
scope and circuit-under-test amplitude from the 1O:1
with just a wire, this simplest of attenuation.
all possible connections would These probes are adjustable
not let you realize the full to compensate for variations in
capacities of your scope. The oscilloscope input capacitance
connection would probably load and your scope has a reference
the circuit and the wire would signal available at the front
act as an antenna and pick up panel. Making this adjustment is
stray signals - 60 Hz power, calledprobe compensation and
CBers, radio and tv stations - you did it as the first step in
and these would be displayed Exercise 3 of Chapter 2.
on the screen along with the Remember when you are
signal of interest. measuring high frequencies,
that the probe’s impedance (re-
Circuit Loading sistance and reactance)
Using a probe instead of a bare changes with frequency. The
wire minimizes stray signals, but probe’s specification sheet or
there’s still an effect from putting manual will contain a chart like
a probe in a circuit called circuit that in Figure 15 that shows this
loading. Circuit loading modi- change. Another point to re-
fies the environment of the sig- member when making high fre-
nals in the circuit you want to quency measurements is to be
measure; it changes the signals sure to securely ground your
in the circuit-under-test, either a probe with as short a ground
little or lot, depending on how clip as possible. As a matter of
great the loading is. fact, in some very high fre- Figure 14.
Circuit loading is resistive, PROBES CONNECTTHE SCOPE AND THE CIRCUIT-UNDER-TEST. Tektronix probes
quency applications a special consist of a patented resistive cable and a grounded shield. Two P6120 probes and
capacitive, and inductive. For socket is provided in the circuit the accessories pictured above are supplied with every 2200 Series scope. The
signal frequencies under 5 kHz, and the probe is plugged into probe is a high impedance, minimum loading 1O X passive probe. The accessories
the most important component that. for each probe (from left to right) are: a grabber tip for ICs and small diameter leads; a
of loading is resistance. To avoid retractable hook tip; and IC tester tip cover; an insulating ground cover; marker
significant circuit loading here, Measurement System bands; and (in the center) the ground lead.
all you need is a probe with a Bandwidth
resistance at least two orders of Then there is one more probe
magnitude greater than the cir- characteristic to consider: For example, if both probe and you use the particular probe de-
cuit impedance (100 Ma probes bandwidth. Like scopes, probes scope have rise times of 5.83 signed for that instrument. For
for 1 Ma sources; 1 MSZ probes have bandwidth limitations; nanoseconds: example, in the case of the 2200
for IO k0 sources, and so on). each has a specified range Series scopes and the P6120
When you are making mea- within which it does not at- Tr (system) = v Tr21scof3el + Tr2(probe) 10X Passive Probe, the probe
surements on a circuit that con- tenuate the signal’s amplitude and the scope have been de-
tains high frequency signals, in- more than -3 dB (0.707 of the T,=d34+34 signed to function together and
ductance and capacitance be- original value). But don’t as- you have the full 60 MHz band-
come important. You can’t avoid sume that a 60 MHz probe and a That works out to 8.25 width at the probe tip..
adding capacitance when you 60 MHz scope give you a 60 nanoseconds, the equivalent to
make connections, but you can MHz measurement capability. a bandwidth of 42.43 MHz be-
avoid adding more capacitance The combination will approxi- cause:
than necessary. mately equal the square root of
the sum of the squares of the 350
One way to do that is to use an B wlmegahertz) =
attenuator probe; its design rise times (also see Chapter 10). Tr (nanoseconds)
greatly reduces loading. In-
stead of loading the circuit with To get the full bandwidth from
capacitance from the probe tip your scope, you need more
bandwidth from the probe. Or

17
ALL ABOUT PROBES CONT.

Probe Types PROBE TYPES CHARACTERISTICS


Proper termination is impor-
Generally you can divide probes No signal reduction, which allows the maximum sensitivity
tant to avoid unwanted reflec-
1 X passive,
by function, into voltage- voltage-sensing at the probe tip; limited bandwidths: 4-34 MHz; high capacitance: tions of the signal you want to
32-l 12 pF; signal handling to 500 V
sensing and current-sensing measure within the cable.
types. Then voltage probes can 10X/100X/1000X Attenuates signals; bandwidths to 300
MHz; adjustable capacitance; signal
Probe/cable combinations de-
passive,
be further divided into passive voltage-sensing, handling to 500 V (10X), 1.5 kV (100X), signed to drive 1 megohm
attenuator or 20 kV (1000X)
and active types. One of these (1 MCI) inputs are engineered to
should meet your measurement active, Switchable attenuation; capacitance as low as 1.5 pF; more
expensive, less rugged than other types; limited dynamic range; but
suppress these reflections. But
voltage-sensing,
requirements. FET bandwidths to 900 MHz; minimum circuit loading for 50 fl scopes, 50 fl probes
current-sensing Measure currents from 1 mA to 1000 A; DC to 50 MHz; very low loading should be used. The proper
Signal handling to 40 kV
termination is also necessary
high voltaae
when you use a coaxial cable
instead of a probe. If you use a
50 a cable and a 1 MSZ scope,
be sure you also use a 50 fl
terminator at the scope input.
The probe’s ruggedness, its
flexibility, and the length of the
cable can also be important (but
remember, the more cable
length, the more capacitance at
the probe tip). And check the
specifications to see if the
bandwidth of the probe is suffi-
cient, and make sure you have
the adapters and tips you’ll
need. Most modern probes fea-
ture interchangeable tips and
adaptors for many applications.
Retractable hook tips let you at-
tach the probe to most circuit
components. Other adaptors
connect probe leads to coaxial
connectors or slip over square
pins. Alligator clips for contact-
ing large diameter test points
are another possibility.
But for the reasons already
mentioned (probe bandwidth,
Figure 15. loading, termination), the best
PROBE IMPEDANCE IS RELATEDTO FREQUENCY as shown in the table above. way to ensure that your scope
The curves plot both resistance (R) and reactance (X) in ohms against frequency in
megahertz. The plot shown is for the Tektronix P6120 probe on a l-meter cable.
and probe measurement sys-
tem has the least effect on your
measurements is to use the
probe recommended for your
scope. And always make sure
Picking A Probe
it’s compensated.
For most applications, the based on the voltage you intend
probes that were supplied with to measure. For example, if
your scope are the ones you you’re going to be looking at a
should use. These will usually 50 volt signal and your largest
be attenuator probes. Then, to vertical sensitivity is 5 volts, that
make sure that the probe can signal will take up ten major divi-
faithfully reproduce the signal sions of the screen. This is a
for your scope, the compensa- situation where you need at-
tion of the probe should be ad- tenuation; a 10X probe would
justable. If you’re not going to reduce the amplitude of your
use the probes that came with signal to reasonable propor-
your scope, pick your probe tions.
18
PART Il. MAKING MEASUREMENTS

The first five chapters described discussions in Chapter 7 start Chapter 9 discusses mea-
how to select the exact oscillo- with safety because you should surement techniques, begin-
scope functions you need to always observe safety precau- ning with fundamental time and
make the measurements you tions when working on electrical amplitude measurements and
want. Now you can put what equipment. ending with delayed sweep
you’ve learned into practice with The first step in ensuring ac- measurements.
this section of the primer. curate measurements is making The last chapter in the primer
It begins with a review of sure your scope is set up prop- describes oscilloscope per-
waveform shapes and charac- erly, and this subject is dis- formance and how it affects your
teristics in Chapter 6. Then the cussed in Chapter 8. measurements.

CHAPTER 6. WAVEFORMS
The definition of a wave is “a
disturbance traveling through a
Figure 16.
medium” while the definition of a BASIC WAVESHAPES include sine
waveform is “a graphic repre- waves, and various non-sinusoidal
SINE WAVE
sentation of a wave.” waves such as triangle waves, square
Like a wave, a waveform is waves, and sawtooth waves. A square
wave has equal amounts of time for its
dependent on two things: two states. Triangle and sawtooth waves
movement and time. The ripple are usually the result of circuits de-
on the surface of a pond exists signed to control voltage with respect to
as a movement of water in time. time, like the sweep of an oscilloscope
and some television circuits. In these
The waveform on your scope’s waveforms, one (or both) transitions
screen is the movement of an from state to state are made with a
electron beam during time. steady variation at a constant rate, a
The changes in the waveform ramp. (Changes from one state to
another on all waveforms except sine
with time form the waveshape, waves are called transitions.) The last
the most readily identifiable
-fRIAIWLE WAVE
two drawings represent aperiodic,
characteristic of a waveform. single-shot waveforms. The first is a
Figure 16 illustrates some com- pulse; all pulses are marked by a rise, a
finite duration, and a decay. The second
mon waveshapes. one is a step, which is actually a single
transition.

PUISE STEP

19
MAKING MEASUREMENT CONT.

Waveshapes tell you a great


deal about the signal. Anytime
you see a change in the vertical
dimension of a signal, you know
that this amplitude change rep-
resents a change in voltage.
Anytime there’s a flat horizontal
line, there was no change for
that length of time. Straight
diagonal lines mean a linear
change, equal rise (or fall) of
voltage for equal amounts of
time. Sharp angles on a
waveform mean a sudden SINE WA= STW
change. But waveshapes alone TIME-
are not the whole story. When
you want to completely describe
a waveform, you’ll want to find
the parameters of that particular
waveform. Depending on the
signal, these parameters might
be amplitude, period, fre-
quency, width, rise time, or
phase. You can review these
signal parameters with Figures
17 through 22.

Figure 17. SQUARE WAVE


AMPLITUDE IS A CHARACTERISTIC OF
ALL WAVEFORMS. It is the amount of
displacement from equilibrium at a par-
ticular point in time. Note that without a
modifier, the word means the maximum
change from a reference without regard
to the direction of the change. In the first
two drawings above (sine wave and
square wave), the amplitude is the same
even though the sine wave is larger from ERIOQ
peak to peak. In the third drawing, an I - l
alternating current waveform is shown
with peak (or maximum) amplitude and
peak-to-peak amplitude parameters
annotated. In oscilloscope measure-
ments, amplitude usually means peak-
to-peak amplitude.

Figure 18.
PERIOD IS THE TIME REQUIRED FOR ONE CYCLE OF A SIGNAL if the signal
repeats itself. Period is a parameter whether the signal is symmetrically shaped like
the sine and square waves above, or whether it has a more complex and asymmetri-
cal shape like the rectangular wave and damped sine wave. Period is always
expressed in units of time. Naturally, one-time signals like the step or uncorrelated
signals (without a time relation) like noise have no period.

20
PART II

Figure 19.
IF A SIGNAL IS PERIODIC, IT HAS A FREQUENCY. Frequency is the number of times
a signal repeats itself in a second; frequency is measured in Hertz: 1 Hz = 1 cycle per
second; 1 kHz (kilohertz) = 1000 cycles/second; and 1 MHz (megahertz) =
1 ,OOO,OOO cycles/second. Period and frequency are reciprocal: l/period = fre-
quency, and l/frequency = period. For example, a 7 Hz signal has a period of 0.143
seconds: 1/7 Hz = 0.143 s, and 1/0.143 s = 7 Hz.

Figure 22.
PHASE is best explained with a sine
wave. Remember that this waveform is
based on the sine of all the angles from 0
through 360. The result is a plot that
Figure 21. changes from 0 to 00,l at 900,O again at
DUTY CYCLE, DUTY FACTOR, AND 1800, -1 at 270”, and finally 0 again at
REPETITION RATE are parameters of all 360”. Consequently, it is useful to refer to
rectangular waves. They are particularly the phase angle (or simply phase, when
important in digital circuitry. Duty cycle there is no ambiquity) of a sine wave
is the ratio of pulse width to signal period when you want to describe how much of
expressed as a percentage. For square the period has elapsed. Another use of
waves, it’s always 50% as you can see; phase is found when you want to de-
for the pulse wave in the second draw- scribe a relationship between two sig-
ing, it’s 30%. Duty factor is the same nals. Picture two clocks with their sec-
thing as duty cycle except it is ex- ond hands sweeping the dial every 60
pressed as a decimal, not a percentage. seconds. If the second hands touch the
A repetition rate describes how often a twelve at the same time, the clocks are in
pulse train occurs and is used instead of phase; if they don’t, then they’re out of
frequency to describe waveforms like phase. To express how far out of phase
that in the second drawing. they are, you use phase shift in degrees.
To illustrate, the waveform labeled
CURRENT in the drawing above is said
to be 90” out of phase with the voltage
waveform. Other ways of reporting the
Figure 20. same information are “the current
THE PARAMETERS OF A PULSE can be important in a number of different applica- waveform has a 90 degree phase angle
tions. Digital circuitry, X-ray equipment, and data communications are examples. with respect to the voltage waveform”or
Pulse specifications include transition times measured on the leading edge of a “the current waveform lags the voltage
positive-going transition; this is the rise time. Fall time is the transition time on a waveform by 9OO.” Note that there is al-
negative-going trailing edge. Pulse width is measured at the 50% points and ways a reference to another waveform;
amplitude from 0 to 100%. Any displacement from 0 volts for the base of the pulse is in this case, between the voltage and
the baseline offset. current waveforms of an inductor.

21
CHAPTER 7. SAFETY

Before you make any oscillo- service electrical devices alone. under-test. Remember that if Then there are few rules about
scope measurement, remember Know the symbols for danger- you lose the ground, all ac- the scope itself: To avoid a
that you must be careful when ous circuits and observe the cessible conductive parts - in- shock, plug the power cord of
you work with electrical equip- safety instructions for the cluding knobs that appear to be the scope into a properly-wired
ment. Always observe all safety equipment you’re working on. insulated - can give you a receptacle before connecting
precautions described in the Don’t operate an electrical de- shock. To avoid personal injury, your probes; only use the power
operators or service manual for vice in an explosive atmos- don’t touch exposed connec- cord for your scope, and don’t
the equipment you’re working phere. Always ground the scope tions and components in the use one that isn’t in good condi-
on. to the circuit, and ground both circuit-under-test when the tion (cracked, broken, missing
Some general rules about your scope and the circuit- power is on. And remember to ground pin, etc.). Use the right
servicing electrical equipment consult the service manual for fuse to avoid fire hazards. Don’t
are worth repeating here. Don’t the equipment you’re working remove covers and panels on
on. your scope.

CHAPTER 8. GETTlNG STARTED


Accurate oscilloscope mea- Checking the Controls l Check the horizontal system Handling a Probe
surements require that you The most common mistake in control settings: magnification Before you probe a circuit, you
make sure your system is prop- making oscilloscope measure- is off (push in the red VAR should make sure you have the
erly setup each time you begin ments is forgetting to compen- switch in the middle of the right probe tips and adaptors for
to use your scope. sate the probe. The second SEC/DIV switch); variable the circuits you will be working
most frequent source of inac- SEC/DIV is in its calibrated de- on. (Tips available for the Tek-
Compensating the Probe curacies is forgetting to check tent position. Make sure the tronix P6120 10X probes were
Most measurements you make the controls to make sure they’re horizontal mode switch is shown in Figure 14, Chapter 5.)
with an oscilloscope require an where you think they are. Here where you need it: NO DLY Then make sure that the
attenuator probe, which is any are some things to check on when you’re not making de- ground in the circuit-under-test
probe that reduces voltage. The your Tektronix 2200 Series layed sweep measurements, is the same as the scope ground
most common are 10X (“times scope (arranged according to INTENS for making measure- - don’t just assume it is. The
ten”) passive probes which re- the functional blocks of your ments with an intensified zone, scope ground will always be
duce the amplitude of the signal scope): or DLY’D if you want a delayed earth ground as long as you’re
and the circuit loading by 10:1. sweep (A, ALT or B on a 2215).
Check all the vertical system using the proper power cord
But before you make any l

controls: variable controls (CH l Then check your trigger sys- and plug. Check the circuit
measurement with an attenuator
1 and CH 2 variable VOLTS/ tem controls to make sure your ground by touching the probe
probe, you should make sure it’s
DIV) should be in their cali- scope will pick the right slope tip to the point you think is
compensated. Figure 23 illus-
brated detent positions; make on the trigger signal, that the ground before you make a hard
trates what can happen to the
sure CH 2 isn’t inverted (unless right coupling is selected, and ground by attaching the ground
waveforms you’ll see when the
you want it to be); check the that the correct operating strap of your probe.
probe is not properly compen-
vertical mode switches to mode will be used. Also make If you’re going to be probing a
sated.
make sure the signal from the sure that the trigger variable lot of different points in the same
Note that you should com- holdoff control is at its mini- circuit and measuring frequen-
pensate your probe as it will be proper channel(s) will be dis-
played; check the two vertical mum position. cies less than 5 MHz, you can
used when you make the mea- ground that circuit to your scope
surement. Compensate it with system VOLTS/DIV switches to
see if their settings are right once instead of each time you
the accessory tip you’ll be using move the probe. Connect the
and don’t compensate the (and don’t forget to use the
VOLTS/DIV readout that circuit ground to the jack
probe in one vertical channel marked GND on the front panel.
and then use it on another. matches the probe, either IX or
10X); check the input coupling
levers too.

22
GETTING STARTED CONT. PART II

Figure 23.
IMPROPERLY COMPENSATED PROBES can distort the waveforms you see on the
screen of your scope. In the photographs the probe adjustment signal and a 1 MHz
square wave are shown as they will appear with proper and improper compensa-
tions. Notice the amplitude and ringing changes on the square wave with the
differences in compensation.

23
CHAPTER 9. MEASUREMENT TECHNIQUES

Rather than attempt to describe


how to make every possible Exercise 6. AMPLITUDE MEASUREMENTS
measurement, this chapter de-
scribes common measurement .1 .Connect your probe to the
techniques you can use in many channe 7 lBNC connector and to
applications. the probe adjustment jack. At-
tach the probe ground strap to
The Foundations: Amplitude the collar of the channel 2 BNC.
and Time Measurements Make sure your probe is com-
The two most basic measure- pensated and that all the vari-
ments you can make are able controls are set in their de-
amplitude and time; almost tent positions.
every other measurement you’ll 2. The trigger MODE switch
make is based on one of these should be set to NORM for nor-
two fundamental techniques. mal triggering. The HORIZON-
Since the oscilloscope is a TAL MODE should be NO DLY (A screen. (The settings should be 5. Now you can count major and
voltage-measuring device, volt- on the 2275). Make sure the 0.7 V on the VOLTS/D/V and 0.2 minor divisions down the center
age is shown as amplitude on channel 7 coupling switch is set ms on the SEC/D/ switches.)
V vertical graticule line and multi-
your scope screen. Of course, to AC and that the trigger 4. Now use the CH 7 vertical ply by the VOLTSIDI setting V to
voltage, current, resistance, and SOURCE switch is on internal POSITION control to move the make the measurement. For
power are related: and the /N Tswitch on CH 7. Set square wave so that its top is on example, 5.0 divisions times 0.7
the VERTICAL MODE switch to the second horizontal gra ticule volts equals 0.5 volts. (If the
CH 7 as well. line from the top edge of the voltage of the probe adjustment
current = -!!GF- 3. Use the trigger LEVEL control screen. Use the horizontal posi- square wave in your scope is
resistance
to obtain a stable trace and tion control to move the signal so different from this example,
resistance = voltage move the volts /division switch that the bottom of one cycle in- that ’s because this signal is not
current
until the probe adjust square tersects the center vertical a critical part of your scope and
power = current x voltage wave is about five divisions gra ticule line. tight tolerances and exact cali-
high. Now turn the seconds/ bration are not required.)
division switch until two cycles
Amplitude measurements are of the waveform are on your
best made with a signal that
covers most of the screen verti-
cally. Use Exercise 6 to practice
making amplitude measure-
ments. Exercise 7. TIME MEASUREMENTS
Time measurements are also Time measurements are best
more accurate when the signal made with the center horizontal
covers a large area of the graticule line. Use the instru-
screen. Continue with the set-up ment settings from Exercise 6
you had for the amplitude mea- and center the square wave ver-
surement, but now use Exercise tically with the vertical POSI-
7 to make a period measure- TION control. Then line up one
ment. rising edge of the square wave
with the gra ticule line that ’s sec-
ond from the lefhdan sidedof the
screen with the HORIZONTAL
position control. Make sure the
next rising edge intersects the
center horizontal graticule.
Count major and minor divisions
across the center horizontal
graticule line from left to right as milliseconds equals 7.74 milli- this example, remember that
shown in the photo above. Mul- seconds. (If the period of the this signal is not a critical part of
tiply by the SEC/D/ setting;
V for probe adjustment square wave the calibration of your scope.)
example, 5.7 divisions times 0.2 in your scope is different from

24
PART II

Frequency and Other Derived Pulse Measurements


Measurements Pulse measurements are impor-
The voltage and time measure- tant when you work with digital
ments you just made are two equipment and data communi-
examples of direct measure- cations devices. Some of the
ments. Once you’ve made a di- signal parameters of a pulse
rect measurement, there are de- were shown in Figure 20, but
rived measurements you can that was an illustration of an
calculate. Frequency is one ideal pulse, not one that exists in
example; it’s derived from the real world. The most impor-
period measurements. While tant parameters of a real pulse
period is the length of time re- are shown in Figure 25.
quired to complete one cycle of Use Exercise 8 to make de-
a periodic waveform, frequency rived measurements with the
is the number of cycles that take probe adjustment square wave.
place in a second. The mea-
surement unit is a hertz (1
cycle/second) and it’s the recip-
rocal of the period. So a period
of 0.00114 second (or 1.14 mil-
liseconds) means a frequency
of 877 Hz.
More examples of derived
measurements are the alternat-
ing current measurements illus- Figure 25.
trated by Figure 24. REAL PULSE MEASUREMENTS include a few more parameters than those for an
ideal pulse. In the diagram above, several are shown. Preshooting is a change of
amplitude in the opposite direction that precedes the pulse. Overshooting and
Figure 24. rounding are changes that occur after the initial transition. Ringing is a set of
DERIVED MEASUREMENTS are the re- amplitude changes - usually a damped sinusoid -that follows overshooting. All
sult of calculations made after direct are expressed as percentages of amplitude. Settling time expresses how long it
measurements. For example, alternat- l+--fEpIoP - I takes the pulse to reach it maximum amplitude. Droop is a decrease in the maximum
ing current measurements require an amplitude with time. And nonlinearity is any variation from a straight line drawn
amplitude measurement first. The through the 10 and 90% points of a transition.
easiest place to start is with a peak-to-
peak amplitude measurement of the
voltage - in this case, 330 volts be-
cause peak-to-peak measurements ig-
nore positive and negative signs. The
peak voltage is one-half that (when there
Exercise 8. DERIVED MEASUREMENTS
is no DC offset), and is also called a
maximum value; it’s 165 V in this case.
With the period measurement in this case because repetition
Theaverage value is the total area under you just made in Exercise 7, cal- rate and frequency are equal for
the voltage curves divided by the period culate the frequency of the square waves. Your probe ad-
in radians; in the case of a sine wave, the probe adjustment square wave. justment tsignal might differ
average value is 0 because the positive
and negative values are equal. The RMS
For example, if the period is 7 slightly from this example; cal-
(root mean square) voltage for this sine millisecond, then the frequency culate the derived measure-
wave - which represents the line volt- is the reciprocal, 7 IO.007 or 7000 ments for it. You can also calcu-
age in the United States- is equal to Hz. Other derived measure- late the peak, peak-to-peak,
the maximum value divided by the
square root of 2: 165 + 1.414 = 117 volts.
ments you can make are duty and average values of the probe
You get from peak-to-peak to RMS volt- cycle, duty factor, and repetition adjustment square wave in your
age with: peak-to-peak + 2 x the square rate. Duty cycle is the ratio of scope, Don’t forget that you
root of 2. pulse width to signal period ex- need both the alternating and
pressed as a percentage: 0.5 direct components of the signal
ms t 7 ms, or 50%. But you to make these measurements,
knew that because for square so be sure to use direct coupling
waves, it’s always 50%. Duty (DC) on the vertical channel
factor is 0.5. And the repetition you’re using.
rate (describing how often a
pulse train occurs) is 7 /second
MEASUREMENT TECHNIQUES CONT.

Use the directions in Exercise Phase Measurements Note that general purpose surement as long as your known
9 to make a pulse measurement You know that a waveform has oscilloscope Lissajous pattern signal is accurate and both sig-
on the probe adjustment square phase, the amount of time that phase measurements are usu- nals are sine waves. The pat-
wave. has passed since the cycle be- ally limited by the frequency re- terns you can see are illustrated
gan, measured in degrees. sponse of the horizontal ampli- in Figure 26, where the effects of
There is also a phase relation- fier (typically designed with far both frequency and phase dif-
Exercise 9. ship between two or more less bandwidth than vertical ferences are shown.
PULSE WIDTH waveforms: the phase shift (if channels). Specialized X-Y Component checking in ser-
MEASUREMENTS any). There are two ways to scopes or monitors will have al- vice or production situations is
To measure the pulse width of measure the phase shift be- most identical vertical and hori- another X-Y application; it re-
the probe adjustment square tween two waveforms. One is by zontal systems. quires only a simple transistor
wave quickly and easily, set your putting one waveform on each checker like that shown in Fig-
channel of a dual-channel X-Y Measurements
scope to trigger on and display ure 27
scope and viewing them directly Finding the phase shift of two
channel 7. Your probe should sinusoidal signals with a Lissa- There are many other applica-
still be connected to the channel in the chop or alternate vertical
mode; trigger on either channel. jous pattern is one example of tions for X-Y measurements in
7 BNC connector and the probe an X-Y measurement. The X-Y television servicing, in engine
adjustment jack from the previ- Adjust the trigger LEVEL control
for a stable display and mea- capability can be used for other analysis, and in 2-way radio ser-
ous exercises. Use 0.7 ms / measurements as well. The Lis- vicing, for examples. In fact, any
division and the no delay hori- sure the period of the wave-
forms. Then increase the sweep sajous patterns can also be time you have physical phe-
zontal mode (A sweep if you’re used to determine the fre- nomena that are interdependent
using a 2275). Use AUTO trig- speed so that you have a dis-
play something like the second quency of an unknown signal and not time-dependent, X-Y
gering on the positive slope and when you have a known signal measurements are the answer.
adjust the trigger LEVEL control drawing back in Figure 22. Then
measure the horizontal distance on the other channel. This is a Aerodynamic lift and drag,
to get as much of the leading very accurate frequency mea- motor speed and torque, or
edge as possible on your between the same points on the
screen. Switch the coupling on two waveforms. The phase shift
channel 7 to ground and center is the difference in time divided
the baseline on the center hori- by the period and multiplied by
zontal graticule. Now use AC 360 to give you degrees.
coupling because that will cen- Displaying the two waveforms
ter the signal on the screen and and measuring when one starts
you make pulse width mea- with respect to another is possi-
surements at the 50 % point of ble with any dual trace scope,
the waveform. Use your horizon- but that isn’t the only way to
tal POSITION control to line up make a phase measurement.
the 50 % point with the first major Look at the front panel and you’ll
see that the vertical channel

oooo/\oovoo
graticule from the left side of the
screen. Now you can count divi- BNC connectors are labeled X
sions and subdivisions across and Y. The last position on the
the center horizontal and multi- SEC/DlV switch is XY, and when +
ply by the SEC/D/V switch set- you use it, the scope’s time base o”zo30f45 90 135” 700’
ting to find the pulse width. is bypassed. The channel 1
input signal is still the horizontal
axis of the scope’s display, but
now the signal on channel 2 be-
comes the vertical axis. In the
3 v\MDoooNooov\
0” 15O 30° 60” TO” lZO”
X-Y mode, you can input one
sinusoidal on each channel and
your screen will display a Lissa-
jous pattern. (They are named
for Jules Antoine Lissajous, a
French physicist; say “LEE-za-
shu”). The shape of the pattern Figure 26.
will indicate the phase differ- FREQUENCY MEASUREMENTS WITH LISSAJOUS PATTERNS require a known sine
ence between the two signal. wave on one channel. If there is no phase shift, the ratio between the known and
Some examples of Lissajous unknown signals will correspond to the ratio of horizontal and vertical lobes of the
pattern. When the frequencies are the same, only the shifts in phase will affect the
patterns are shown in Figure 26. pattern. In the drawings above, both phase and frequency differences are shown.

26
PART II

Now set your volts/division If you use the channel 2


switches so that the two signals VOLTS/DIV switch and VAR con-
are about equal in amplitude. trol for maximum cancellation of
Then you can move the right- the common signal, the signal
hand VERTICAL MODE switch that remains on-screen will only
to ADD and press the INVERT contain the desired part of the
button so that the common channel 1input signal. The two
mode signals have opposite common mode signals have
polarities. cancelled out leaving only the
difference between the two.

Figure 27.
X-Y COMPONENT CHECKING requires the transistor checker shown above. With it
connected to your scope and the scope in the X-Y mode, patterns like those
illustrated indicate the component’s condition. The waveforms shown are found
when the components are not in a circuit; in-circuit component patterns will differ
because of resistors and capacitors associated with the component.

pressure and volume of liquids Differential Measurements Figure 28.


and gasses are more examples. The ADD vertical mode and the DIFFERENTIAL MEASUREMENTS allow you to remove unwanted information from a
With the proper transducer, you channel 2 INVERT button of your signal anytime you have another signal that closely resembles the unwanted com-
can use your scope to make any 2200 Series scope let you make ponents. For example, the first photo shows a 1 kHz square contaminated by a 60 Hz
sine wave. Once the common-mode component (the sine wave) is input to channel 2
of these measurements. differential measurements. and that channel is inverted, the signals can be added with the ADD vertical mode.
Often differential measurements The result is shown in the second photo.
let you eliminate undesirable
components from a signal that
you’re trying to measure. If you Using the Z Axis Z-axis input will accept a signal
have a signal that’s very similar Remember from Part I that the of up to 30 V through a usable
to the unnecessary noise, the CRT in your scope has three frequency range of DC to 5 MHz.
set up is simple. Put the signal axes of information: X is the hori- Positive voltages decrease the
with the spurious information on zontal component of the graph, brightness and negative volt-
channel 1. Connect the signal Y is the vertical, and Z is the ages increase it; 5 volts will
that is like unwanted compo- brightness or darkness of the cause a noticeable change.
nents to channel 2. Set both electron beam. The 2200 Series The Z-axis input is an advan-
input coupling switches to DC scopes all have an external tage to users that have their in-
(use AC if the DC components of Z-axis input BNC connector on struments set up for a long
the signal are too large), and the back of the instrument. This series of tests. One example is
select the alternate vertical input lets you change the the testing of high fidelity
mode by moving the VERTICAL brightness (modulate the inten- equipment illustrated by
MODE switches to BOTH and sity) of the signal on the screen Figure 29.
ALT with an external signal. The
27
MEASUREMENT TECHNIQUES CONT.

can simply switch the vertical delay out to where you want the
operating mode to display both sweep to start.
channels. That makes the total But the delayed sweep fea-
holdoff time for one channel ture you’ll probably use the most
greater than one field period. often is the intensified sweep; it
Then just position the unused lets you use the delayed sweep
vertical channel off-screen to as a positionable magnifier. You
avoid confusion. trigger normally and then use

c SCOPE
It is also important to select
the trigger slope that corre-
sponds to the edge of the
the scope’s intensified horizon-
tal mode. Now the signal on the
screen will show a brighter zone
waveform where the sync after the delay time. Run the
pulses are located. Picking a delay time (and the intensified
negative slope for pulses at the zone) out to the part of the signal
bottom of the waveform allows that interests you. Then switch to
you to see as many sync pulses the delayed mode and increase
as possible. the sweep speed to magnify the
When you want to observe the selected waveform portion so
TV line portion of the composite that you can examine it in detail.
video signal, use the NORM Since the 2200 Series has two
trigger mode and trigger on the types of delayed sweep, read
horizontal synchronization the paragraphs and use the de-
pulses for a stable display. It is layed sweep measurement
usually best to select the blank- exercise below that applies to
ing level of the sync waveform your scope: “Single Time Base
so that the vertical field rate will Scopes” and Exercise 10 for de-
not cause double triggering. layed sweep measurements
with single time base scopes
Delayed Sweep like the Tektronix 2213; or “Dual
Measurements Time Base Scopes”and Exer-
Delayed sweep is a technique cise 11 for delayed sweep on
that adds a precise amount of dual time base scopes like the
time between the trigger point
2215.
and the beginning of a scope
sweep. Often delayed sweep is Single Time Base Scopes
used as a convenient way to Very few single time base
Figure 29.
make a measurement (the rise scopes offer delayed sweep
USING THE Z-AXIS can provide additional information on the scope screen. In the time measurement in Exercise measurements. Those that do
set-up drawn above, a function generator sweeps through the frequencies of interest 10 is a good example). To make may have measurement
during the product testing-20 to 20,OOO Hz, in this case. Then an adjustable notch a rise time measurement without capabilities similar to those of
filter is used to generate a marker, at 15 kHz, for instance, and this signal is applied to
the Z-axis input to brighten the trace. This allows the tester to evaluate the product’s
delayed sweep, you must trig- the Tektronix 2213 which has
performance with a glance. ger on the edge occurring be- three possible horizontal opera-
fore the desired transition, With ting modes annotated on the
delayed sweep, you may front panel as NO DLY, INTENS,
Using TV Triggering mode. This mode allows the choose to trigger anywhere and DLY’D.
The composite video waveform scope to trigger at the field rate along the displayed waveform When you set the HORIZON-
consists of two fields, each of of the composite video signal on and use the delay time control to TAL MODE switch to NO DLY (no
which contains 262 lines. Many either field one or field two. start the sweep exactly where delay), only the normal sweep
scopes offer television trigger- Since the trigger system cannot you want. functions.
ing to simplify looking at video recognize the difference be- Sometimes, however, de- When you choose INTENS (in-
signals. Usually, however, the tween field one and field two, it layed sweep is the only way to tensified sweep), your scope
scope will only trigger on fields will trigger alternately on the two make a measurement. Suppose will display the normal sweep
at some sweep speeds and fields and the display will be that the part of the waveform you and the trace will also be inten-
lines at others. The 2200 Series confusing if you look at one line want to measure is so far from sified after a delay time. The
scopes allow you to trigger on at a time. the only available trigger point amount of delay is determined
either lines or fields at any To prevent this, you add more that it will not show on the by both the DELAY TIME switch
sweep speed. holdoff time, and there are two screen The problem can be (you can use 0.5 ps, 10 ps, or
To look at tv fields with a 2200 ways to do that. You can use the solved with delayed sweep: 0.2 ms) and the DELAY TIME
Series scope, use the TV FIELD variable holdoff control, or you trigger where you have to. and MULTIPLIER control. The multi-
28
PART II

plier lets you pick from 1 to 20


times the switch setting. Exercise 10.2213 DELAYED SWEEP MEASUREMENTS
The third position, DLY ’D (de-
layed), makes the sweep start
after the delay time you ’ve cho-
sen. After selecting this position
you can move the SEC/DIV to a
faster sweep speed and
examine the waveform in
greater detail.
This list of horizontal modes
should begin to give ideas of
how useful these delayed
sweep features are. Start by
1. Connect your probe to the 4. Because a rise time mea- 7. Change to 0.7 Vldiv and line
making the rise time measure-
channel 7 BNC connector and surement is best made at faster up the signal with the 0 and
ment described below. (Note
the probe adjustmentjack, hook sweep speeds, turn the SEC/ 700% dotted lines of the
that when making rise time
the ground strap onto the collar DIV control to 2 w. Use the trig- graticule. (If you have a signal
measurements, you must take
of the channel2 BNC, and make ger LEVEL control to try to get all that doesn’t fit between the 0
the rise time of the measuring
sure the probe is compensated. of the positive transition on the and 700% lines of the graticule,
instrument into account. Be sure
2. Use these control settings: screen. You can’t; you lose your you have to count major and
to read Chapter IO.)
CH 7 VOLTS/D/V on 0.2 using trigger when you get off the minor divisions and estimate the
Dual Time Base Scopes the 10X probe VOLTS/DIV read- slope of the signal. rise time while ignoring the first
Delayed sweep is normally out; CH 7 input coupling on AC; 5. Turn back to 0.5 msldiv and and last 70% of the transition.)
found on dual time base scopes VERTICAL MODE is CH 7; switch to the intensified display
like the 2215 with two totally TRIGGER MODE is AUTO; with the HORIZONTAL MODE 8. Use the horizontal POSITION
separate horizontal sweep TRIGGER SLOPE is negative switch. Switch the DELAY TIME control to move the waveform
generators. In dual time base (-); trigger SOURCE is INT (for to 0.2 ms and use the DELAY until it crosses a vertical
instruments, one sweep is internal) and INT trigger switch TIME MULTIPLIER to move the graticule line at the 70% mark-
triggered in the normal fashion is either CH 7 or VERT MODE; intensified zone on the ing. Adjust the FOCUS control
and the start of the second HORIZONTAL MODE is NO DLY; waveform to a point before the for a sharp waveform and make
sweep is delayed. To keep these SEC/DIV is 0.5 ms. Check all the first complete positive-going your rise time measurement
two sweeps distinct when de- variable controls to make sure transition of the square wave. from that vertical line to where
scribing them, the delaying they’re in their calibrated detent The intensified zone now shows the step crosses the 90% line.
sweep is called the A sweep; the positions. you where the delayed sweep Now you can make a rise time
delayed sweep is called the B 3. Set the input coupling to G ND will start, like the second photo. measurement on a waveform
sweep. The length of time be- and center the trace. Switch 6. Switch the horizontal mode to like that in the third photograph.
tween the start of the A sweep back to AC and set the trigger DLY’D and the SEC/DIV switch For example, for 7 major division
and the start of the B sweep is LEVEL control for a stable dis- to 5 p. Now you can use the and 4 minor divisions: 7.8 times
called the delay time. play. The waveform should look horizontal POSITION and the SEC/DIV setting of 5 w
Dual time base scopes offer like the first photo above. DELAY TIME MULTIPLIER con- is9*.
you all the measurement trols to get a single transition on
capabilities of single time base the screen.
instruments, plus:
l convenient comparisons of
A sweep only, B sweep only, or A sweep and the B sweep de- This sounds more compli-
signals at two different sweep
intensified by B as well as B de- layed. As you set faster sweeps cated in words than it is in prac-
speeds
layed. The HORIZONTAL with the B SEC/DlV switch, you’ll tice. As you use the scope in
l jitter-free triggering of delayed
MODE switch controls the oper- see the intensified zone on the A Exercise 11, you’ll find that the
sweeps ating mode and two SEC/DIV trace get smaller and the B procedure is very easy. You will
l and timing measurement ac- switches - concentrically always see exactly where the B
sweep expanded by the new
curacy of 1.5%. mounted on a 2215 -control speed setting. As you move the sweep starts. And you can use
Most of this increase in mea- the sweep speeds. See Figure B DELAY TIME POSITION dial the size of the intensified zone to
surement performance is avail- 30. and change where the B sweep judge which B sweep speed you
able because you can sepa- When you use the ALT (for al- starts, you’ll see the intensified need to make the measurement
rately control the two sweep ternate horizontal mode) posi- zone move across the A trace you want.
speeds and use them in three tion the HORIZONTAL MODE and see the B waveform
horizontal operating modes. switch, the scope will display change.
These modes - in a 2215 - are the A sweep intensified by the B
29
MEASUREMENT TECHNIQUES CONT.

Measurements at Two Sweep On a 2215, the B TRIGGER


Speeds LEVEL control does double duty.
Looking at a signal with two dif- In its full clockwise position, it
ferent sweep speeds makes selects the run-after-delay
complicated timing measure- mode. At any other position, it
ments easy. The A sweep gives functions as a trigger level con-
you a large slice of time on the trol for the B trigger. The B
signal to examine. The inten- TRIGGER SLOPE control lets
sified zone will show you where you pick positive or negative
the B sweep is positioned. And transitions for the B trigger.
the faster B sweep speeds With these two controls you
magnify the smaller portions of can trigger a stable B sweep
the signal in great detail. You’ll even when the A sweep has
find this capability useful in jitter.
many measurement applica-
Increased Timing
tions; see Figure 31 for two
Measurement Accuracy
illustrations.
Besides examining signals at
Because you can use the two different sweep speeds and
scope to show A and B sweeps seeing a jitter-free B sweep, you
from both channel 1 and chan- get increased timing measure-
nel 2, you can display four ment accuracy with a dual time
traces. To prevent overlapping base scope.
traces, most dual time base Note that the B DELAY TIME
scopes offer an additional posi- POSITION dial is a measuring
tion control. On the 2215, it’s indicator as well as a positioning
labeled ALT SWP SEP for alter- device. The numbers in the win-
nate sweep separation. With it dow at the top of the dial are
and the two vertical channel calibrated to the major divisions
POSITION controls, you can of the scope screen. The num- Figure 30.
place all four traces on-screen bers around the circumference THE DELAYED SWEEP CONTROLS of the dual time base 2215 are shown on the
without confusion. photograph above. They include: HORIZONTAL MODE (under the horizontal POSI-
divide the major division into TION control); B TRIGGER SLOPE and LEVEL; ALT SWP SEP (alternate sweep
Separate B Trigger hundreds. separation, between the two vertical POSITION controls - not shown), and a
Jitter can prevent an accurate To make timing measure- concentric A and B SEC/DlV control. The B DELAY TIME POSITION dial is at the
bottom of the column of horizontal system controls.
measurement anytime you want ments accurate to 1.5% with the
to look at a signal that isn’t per- B DELAY TIME POSITION dial:
fectly periodic. But with two time l use the B runs-after-delay

bases and delayed sweep, you mode.


can solve the problem with the l place the intensified zone (or

separate trigger available for use the B sweep waveform)


the B sweep. You trigger the A where the timing measurement
sweep normally and move the begins, and note the B DELAY
intensified zone out to the por- TIME POSITION dial setting
tion of the waveform you want to l dial back to where the mea-
measure. Then you set the
surement ends and note the
scope up for a triggered B reading there
sweep, rather than letting the B
l subtract the first reading from
sweep simply run after the delay
time. the second and multiply by the Figure 31.
A sweep SEC/DlV setting. ALTERNATE DELAYED SWEEP MEASUREMENTS are fast and accurate. One use,
examining timing in a digital circuit, is demonstrated in the firs photograph.
t Suppose
You’ll find an example of this you need to check the width of one pulse in a pulse train like the one shown. To make
accurate - and easy - timing sure which pulse you are measuring, you want to look at a large portion of a signal.
measurement in Exercise 11. But to measure the one pulse accurately, you need a faster sweep speed. Looking at
both the big picture and a small enlarged portion of the signal is easy with alternate
delayed sweep. Another example is shown in the second photo. Here one field of a
composite video signal is shown in the first waveform. The intensified portion of that
field is the lines magnified by the faster B sweep. With a dual time base scope, you
can walk through the field with the B DELAY TIME POSITION dial and look at each line
individually.

30
PART II

Exercise 11.2215 DELAYED SWEEP MEASUREMENTS


Rise Time Measurement
waveform and count major and GER SOURCE is INT (for inter-
minor divisions across the nal) and INT trigger switch is
screen to where the step either CH 7 or VERT MODE;
crosses the 90% line. If there are HORIZONTAL MODE is A; A
4 major and 8 minor divisions, 4 SEC/DIV is 0.2 ms while B
and 8/10 times the SEC/DIV set- SEC/DIV is 0.05 w. Check the
ting of 7 p is 4.8 p. The third variable controls.
photo shows how the screen 2. Center the first complete
should look now. (Note: Any jitter pulse of the waveform horizon-
you see in the B sweep is from tally. Switch to the A LT display
the probe adjustment circuit, not with the HORIZONTAL MODE
the time base.) switch and move the B
1. Connect your probe to the 5. Pull out on the SEC/DIV knob
8. One last word on rise time waveform to the bottom of the
channel 7 BNC connector and and rotate it clockwise to
measurements: the accuracy of screen with the ALT SWP SEP
the probe adjustmentjack, hook change the B sweep speed to 2
the measurement you make de- control.
the ground strap onto the collar ,us /division. This will make the
pends on both the signal you’re 3. Center A sweep waveform
of the channel2 BNC, and make intensified zone smaller; move it
examining and the performance vertically. Turn down the inten-
sure the probe is compensated. to the first rising edge of the
of your scope. In Chapter 70, sity so that it’s easier to see the
2. Use these control settings: waveform as in the second
you’ll find a description of how small intensified zone.
CH 7 VOLTS/DIV on 0.2 (re- photograph.
the scope’s own rise time affects 4. Move the intensified zone to
member to use the 10X probe 6. Switch the horizontal mode to
your measurement results. the 50% point of the rising edge
readout); CH 7 input coupling on B, the channel 7 vertical sensitiv-
of the waveform with the DELAY
AC; VERTICAL MODE is CH 7; A ity to 0.7 volts/division, and the
Pulse Width Measurement TIME POSITION controlas in the
TRIGGER MODE is NORM; A sweep speed to 7 ,usldivision.
first photo above. Note the delay
TRIGGER SLOPE is negative Use the horizontal and vertical
time reading (the number in the
(-); A SOURCE is INT and the POSITION controls and the B
window first, for example: 3.7).
A&B INT trigger switch is either DELAY TIME POSITION control
Move the intensified zone to the
CH 7 or VERT MODE; HORI- to line up the waveform with the
50% point of the trailing edge as
ZONTAL MODE is A; A and B 0 and 700% dotted lines of the
in the second photo and note
SEC/DIV is 0.2 ms. Check the graticule. (If you have a signal
the reading.
variable controls to make sure that doesn’t fit between the 0
5. The time measurement, a
they’re in their calibrated detent and 700% lines of the graticule,
pulse width in this case, is equal
positions. you have to count major and
to the second dial reading
3. Set the A TRIGGER LEVEL minor divisions and estimate the
minus the first times the A sweep
control for a stable display and rise time while ignoring the first
speed: 5.77 - 3.73 x 0.2 ms =
position the waveform in the top and last 10%.)
0.528 ms. In other words, the B
half of the screen. Switch to the
DELAY TIME POSITION dial in-
ALT (for alternate A and B
dicates screen divisions for you,
sweeps) display with the HORI-
7 complete turn for every major
ZONTAL MODE switch. Use the
division.
channel 7 POSITION and ALT
SWP SEP (alternate sweep sep-
aration) controls to position the
two sweeps so that they don’t
overlap.
4. Use the B DELAY TIME POSI-
TION dial to move the beginning 1. Use these control settings:
of the intensified zone a point CH 7 VOLTS/D/V on 0.7; CH 7
before the first complete posi- 7. Position the waveform so that input coupling on AC; VERTI-
tive transition. Your screen it crosses a vertical graticule line CAL MODE is CH 7; A TRIGGER
should look like the first photo at the 70% marking. Adjust the MODE is NORM; A TRIGGER
above. FOCUS control for a sharp SLOPE is negative (-); A TRIG-

31
INDEX

A D F
A sweep 29 DC coupling 6 Frequency measurements 25 Initialization 3
A/B SWP SEP 8 Deflection voltages 6 Focus control 4 Input coupling control 6
AC coupling 6 Delay time 29 Internal triggering 14
Alternate horizontal operating Delay time control 11 G Intensified sweep 28
mode 10 Delay time multiplier (DTM) GND input coupling 6 Intensity control 4
Alternate sweep separation control 10 Graticule 4
control 8 Delayed sweep 28 L
Amplitude 20 Delayed sweep measurements
H Line triggering 15
Holdoff 14 Lissajous figures 26
Amplitude measurements 24 28
Horizontal magnification
Attenuation 6 Delaying sweep 29
control 11 M
Derived measurements 25 Magnification 11
B Differential measurements 27
Horizontal operating mode
Major divisions 4
B delay time position control 11 control 10
Display system 4 Measurement system band-
B sweep 29 Horizontal position control 10
Divisions 4 width 17
BNC 8 Horizontal system 10
Dual time base scopes 29 Measurement techniques,
Bandwidth and rise time 33
Duty cycle 21 Amplitude 24
Bandwidth specifications 33
Duty factor 21 Differential 27
Beam finder control 4
Delayed sweep 28
Blanking 14 E Derived 25
Electron beam 4
C Exercises, 24 Fall time 25
Capacitance 17 Frequency 25
Amplitude measurement 24
Cathode ray tube (CRT) 2 Period 24
Controlling vertical
Channel 2 inversion control 7 Phase 26
sensitivity 9
Circuit loading 17 Compensating the probe 9 Pulse 25
Compensating the probe 9 Pulse width 26
Delayed sweep 29
Compensation box 9 Rise time 29
Scope initialization 3
Coupling 6 Time 24
Signal coupling 9
video 28
Time measurement 24
X-Y 26
Using the display controls 5
Minor divisions 4
Using the horizontal
controls 12 0
Using the trigger controls 16 Oscilloscope performance 32
Using the vertical controls 9
Vertical operating mode 9
External triaaerina 6

34
PART II

P T U
Parallax errors 4 Television triggering 28 Unblanking 14
Period 24 Termination 18
Phase 21 Time base 10 V
Picking a probe 18 Time measurements 24 Variable SEC/DlV control 11
Phase measurements 26 Trace rotation control 5 Variable trigger holdoff
Phosphor 4 Transducers 2 control 14
Position controls, Transitions 19 Variable VOLTS/DIV control 7
horizontal 10 Triangle waves 19 Vertical operating mode
vertical 6 Trigger, control 7
Probes, automatic mode (AUTO) 15 Vertical operating modes, 7
accessories 17 coupling control 16 ADD 8
attenuator 18 external 14 ALT (alternate) 7
adjustment jack 4 holdoff control 14 BOTH 7
bandwidth 17 internal 14 CHOP 7
compensation 17 line 15 CH1 7
current 18 level control 12 CH27
handling 22 normal mode (NORM) 15 Vertical mode triggering 15
loading 17 operating modes control 15 Vertical position control 6
passive 18 point 13 Vertical sensitivity control 6
Pulse measurements 25 slope control 12 Vertical system 6
Vertical system coupling 6
R
source control 14
system 12 VOLTS/DIV control 6
Ramp 10
Repetition rate 21
TV FIELD mode 15
W
Retrace 10 Waveforms,
Rise time and bandwidth 33 Pulse 19
Rise times, scope and mea- Sawtooth 19
surement 32 Sine 19
Square 19
S Step 19
Safety 22 Triangle 19
Sawtooth waves 19
Scale factors 6 X
SEC/DIV control 10 X axis 5
Screen 4 X-Y measurements 26
Sine waves 19
Single time base scopes 28
Y
Y axis 5
Square waves 19
Square wave and high fre- Z
quency response 32 Z axis 5
Steps 19
Subdivisions 4
Sweep generator 10
Sweep speeds 10

35

You might also like