Professional Documents
Culture Documents
The Xyzs of Using A Scope: Tektronix
The Xyzs of Using A Scope: Tektronix
MULTI-PURPOSE OSCILLOSCOPE
PRIMER
THE XYZs
OF USING A SCOPE
Tektronix
COMMITTED TO EXCELLENCE
CONTENTS
REPRODUCED BY PERMISSION
OF THE COPYRIGHT OWNER:
TEKTRONIX, INC.
(Letter dated: 11/20/85,
Gaithersbucp, Md.)
The oscilloscope draws a graph the path of the beam. A grid of Common controls for display Intensity
by moving an electron beam lines etched on the inside of the systems include intensity and An intensity control adjusts the
across a phosphor coating on faceplate serves as the refer- focus; less often, you will also brightness of the trace. It’s
the inside of the CRT (cathode- ence for your measurements; find beam finder and trace rota- necessary because you use a
ray tube). The result is a glow for this is the graticule shown in tion controls. On a Tektronix scope in different ambient light
a short time afterward tracing Figure 2. 2200 Series instrument they are conditions and with many kinds
all present, grouped next to and of signals. For instance, on
on the right of the CRT At the top square waves you might want to
of the group is the intensity con- change the trace brightness to
trol (labeled AUTO INTENSITY look at different parts of the
CENTER on a 2200 because these in- waveform because the slower
I
GWTICUE
LINES - struments automatically main-
tain the trace intensity once it is
set). The TRACE ROTATION ad-
horizontal components will al-
ways appear brighter than the
faster vertical components.
justment is under that, and then Intensity controls are also use-
the beam finder (BEAM FIND). ful because the intensity of a
Under the probe adjustment trace is a function of both how
jack is the focus control (labeled bright the beam is and how long
AUTO FOCUS because it’s also it’s on-screen. As you select dif-
automatic). The functions of ferent sweep speeds (a sweep
these controls are described is one movement of the electron
below and their positions on the beam across the scope screen)
Tektronix 2213 Portable Oscillo- with the SEC/DIV switch, the
scope are illustrated by the fold- beam ON and OFF times
out illustration at the rear of the change and the beam has more
booklet. or less time to excite the phos-
phor
Beam Finder On most scopes, you have to
The beam finder is a conven- turn the intensity up or down to
ience control that allows you to restore the initial brightness. On
locate the electron beam any- the 2200 Series scopes, how-
time it’s off-screen. When you ever the automatic intensity cir-
push the BEAM FIND button, cuit compensates for changes
you reduce the vertical and in sweep speed from 0.5 milli-
horizontal deflection voltages seconds (0.5 ms) to 0.5 micro-
(more about deflection voltages seconds (0.5 ,us). W ithin this
later) and over-ride the intensity range, the automatic circuit
control so that the beam always maintains the same trace inten-
appears within the 8 x 10- sity you initially set with the
centimeter screen. When you AUTO INTENSITY control.
see which quadrant of the
screen the beam appears in, Focus
you’ll know which directions to The scope’s electron beam is
turn the horizontal and vertical focused on the CRT faceplate
POSITION controls to reposition by an electrical grid within the
Figure 2. the trace on the screen for nor- tube. The focus control adjusts
THE GRATICLJLE is a grid of lines typically etched or silk-screened on the inside of mal operations. that grid for optimum trace
the CRT faceplate. Putting the graticule inside-on the same plane as the trace
drawn by the electron beam. and not on the outside of the glass - eliminates
focus. On a 2200 scope, the
measurement inaccuracies called parallax errors. Parallax errors occur when the AUTO FOCUS circuit maintains
trace and the graticule are on different planes and the observer is shifted slightly from your focus settings over most
the direct line of sight. Though different sized CRT’s may be used, graticules are intensity settings (0.5 ms to
usually laid out in an 8 x 10 pattern. Each of the eight vertical and ten horizontal lines
block off major divisions (or just divisions) of the screen. The labeling on scope
0.5 us).
controls always refers to major divisions. The tick marks on the center graticule lines
represent minor divisions or subdivisions. Since rise time measurements are very
common. 2200 Series scope graticules include rise time measurement markings:
dashed lines for 0 and 100% points, and labeled graticule lines for 10 and 90%
PART I
Trace Rotation
Another display control you’ll
find on the front panel of a 2200
Series instrument is TRACE RO-
TATION. The trace rotation ad-
justment allows you to electri-
cally align the horizontal deflec-
tion of the trace with the fixed
graticule. To avoid accidental
misalignments when the scope
is in use, the control is recessed
and must be adjusted with a
small screwdriver.
If this seems like a calibration
item that should be adjusted
once and then forgotten, you’re
right; that’s true for most oscillo-
scope applications. But the
earth’s magnetic field affects
Figure 3.
the trace alignment and when a THE DISPLAY SYSTEM of your scope
scope is used in many different consists of the cathode-ray tube and its
positions - as a service scope controls. To draw the graph of your mea-
will be - it’s very handy to have surements, the vertical system of the
scope supplies the Y, or vertical, coordi-
a front panel trace rotation ad- nates and the horizontal system
justment. supplies the X coordinates. There is also
a Z dimension in a scope; it determines
Using the Display Controls whether or not the electron beam is
The display system and its con- turned on, and how bright it is when it’s
l-x---/
trols are shown as functional on.
blocks in Figure 3. Use Exercise
2 to review the display controls.
5
CHAPTER 2. THE VERTICAL SYSTEM
The vertical system of your Because all 2200s are two- Input Coupling back is a handy way to measure
scope supplies the display sys- channel scopes, you will have The input coupling switch for signal voltage levels with re-
tem with the Y axis - or vertical one set of these switches for each vertical channel lets you spect to chassis ground. (Using
- information for the graph on each channel. There are also control how the input signal is the GND position does not
the CRT screen. To do this, the two switches for choosing the coupled to the vertical channel. ground the signal in the circuit
vertical system takes the input scope’s vertical display mode DC (the abbreviation normally you’re probing.)
signals and develops deflection and one control that allows you stands for direct current) input
voltages. The display system to invert the polarity of the chan- coupling lets you see all of an Vertical Sensitivity
then uses the deflection volt- nel 2 signal. input signal. AC (alternating cur- A volts/division rotary switch
ages to control - deflect-the For the exercises in this chap- rent) coupling blocks the con- controls the sensitivity of each
electron beam. ter, you’ll need a 10X probe like stant signal components and vertical channel. Having differ-
The vertical system also gives the Tektronix P6120 10X Probes permits only the alternating ent sensitivities extends the
you a choice of how you connect supplied with every 2200 Series components of the input signal range of the scope’s applica-
the input signals (called cou- scope. to reach the channel. An illustra- tions; with a VOLTS/DIV switch,
pling and described below). tion of the differences is shown a multipurpose scope is capa-
And the vertical system pro- Vertical Position in Figure 5. ble of accurately displaying sig-
vides internal signals for the Your scope’s POSITION controls The middle position of the nal levels from millivolts to many
trigger circuit (described in let you place the trace exactly coupling switches is marked volts.
Chapter 4). Figure 4 illustrates where you want it on the screen. GND for ground. Choosing this Using the volts/division
the vertical system schemati- The two vertical POSITION con- position disconnects the input switch to change sensitivity also
cally. trols (there’s one for each chan- signal from the vertical system changes the scale factor, the
nel) change the vertical place- and makes a triggered display value of each major division on
Some of the vertical system ment of the traces from each show the scope’s chassis the screen. Each setting of the
controls - see the foldout front vertical channel; the horizontal ground. The position of the trace control knob is marked with a
panel illustration for their loca- POSITION control changes the on the screen in this mode is the number that represents the
tions - are: vertical position, horizontal position of both ground reference level. Switch- scale factor for that channel. For
sensitivity, and input coupling. channels at once. ing from AC or DC to GND and example, with a setting of 10 V,
N’lENJAl~
“L-
Figure 4.
THE VERTICAL SYSTEM of a Tektronix 2200 Series scope consists of two identical
channels though only one is shown in the drawing. Each channel has circuits to
couple an input signal to that channel, attenuate (reduce) the input signal when
necessary, preamplify it, delay it, and finally amplify the signal for use by the display
system. The delay line lets you see the beginning of a waveform even when the
scope is triggering on it.
6
PART I
All the instruments of the Tek- Horizontal Magnification The 10X magnification is use- The B DELAY TIME POSITION
tronix 2200 Series offer sweep Most scopes offer some means ful when you want to look at sig- Control
speeds from a half-second for of horizontally magnifying the nals and see details that occur This calibrated IO-turn dial is
each division to 0.05 us / waveforms on the screen. The very closely together in time. used to position the beginning
division. The markings appear- effect of magnification is to mul- of the B sweep relative to the A
ing on the scopes are: tiply the sweep speed by the The DELAY TIME and sweep in a 2215. Its uses are
amount of magnification. On MULTIPLIER Controls described under “Delayed
.5s half a second
.2s 0.2 second 2200 Series scopes there is a This switch and dial are used in Sweep Measurements” in
.l s 0.1 second
10X horizontal magnification conjunction with either the in- Chapter 9.
50 ms 50 milliseconds
(0.05 second) that you engage by pulling out tensified or delayed-sweep
20 ms 20 milliseconds on the red VAR switch. The 10X horizontal operating modes in Using the Horizontal Controls
(0.02 second)
horizontal magnification gives the 2213. These features are de- As you can see in Figure 8, the
10 ms 10 milliseconds
(0.01 second) you a sweep speed ten times scribed later under “Delayed horizontal system can be di-
5 ms 5 milliseconds
faster than the SEC/DIV switch Sweep Measurement” in vided into two functional blocks:
(0.005 second)
2 ms 2 milliseconds setting; for example, 0.05 psi Chapter 9. the horizontal amplifier and the
(0.002 second)
division magnified is a very fast sweep generator.
1 ms 1 millisecond
(0.001 second) 5-nanosecond/division sweep.
.5 ms half a millisecond
(0.0005 second)
.2 ms 0.2 millisecond
(0.0002 second)
.l ms 0.1 millisecond
(0.0001 second)
50 I_CS 50 microseconds
(0.00005 second)
20 ps 20 microseconds
(0.00002 second)
10 ps 10 microseconds
(0.00001 second)
5 ps 5 microseconds
(0.000005 second)
2 ps 2 microseconds
(0.000002 second)
1 ps 1 microsecond
(0.000001 second)
.5 ps half a microsecond
(0.0000005 second)
.2 ps 0.2 microsecond
(0.0000002 second)
.1 ps 0.1 microsecond
(0.0000001 second)
.05 ps 0.05 microseconds
(0.00000005 second)
11
PART I
All the instruments of the Tek- Horizontal Magnification The 10X magnification is use- The B DELAY TIME POSITION
tronix 2200 Series offer sweep Most scopes offer some means ful when you want to look at sig- Control
speeds from a half-second for of horizontally magnifying the nals and see details that occur This calibrated IO-turn dial is
each division to 0.05 us / waveforms on the screen. The very closely together in time. used to position the beginning
division. The markings appear- effect of magnification is to mul- of the B sweep relative to the A
ing on the scopes are: tiply the sweep speed by the The DELAY TIME and sweep in a 2215. Its uses are
amount of magnification. On MULTIPLIER Controls described under “Delayed
.5s half a second
.2s 0.2 second 2200 Series scopes there is a This switch and dial are used in Sweep Measurements” in
.l s 0.1 second
10X horizontal magnification conjunction with either the in- Chapter 9.
50 ms 50 milliseconds
(0.05 second) that you engage by pulling out tensified or delayed-sweep
20 ms 20 milliseconds on the red VAR switch. The 10X horizontal operating modes in Using the Horizontal Controls
(0.02 second)
horizontal magnification gives the 2213. These features are de- As you can see in Figure 8, the
10 ms 10 milliseconds
(0.01 second) you a sweep speed ten times scribed later under “Delayed horizontal system can be di-
5 ms 5 milliseconds
faster than the SEC/DIV switch Sweep Measurement” in vided into two functional blocks:
(0.005 second)
2 ms 2 milliseconds setting; for example, 0.05 psi Chapter 9. the horizontal amplifier and the
(0.002 second)
division magnified is a very fast sweep generator.
1 ms 1 millisecond
(0.001 second) 5-nanosecond/division sweep.
.5 ms half a millisecond
(0.0005 second)
.2 ms 0.2 millisecond
(0.0002 second)
.l ms 0.1 millisecond
(0.0001 second)
50 I_CS 50 microseconds
(0.00005 second)
20 ps 20 microseconds
(0.00002 second)
10 ps 10 microseconds
(0.00001 second)
5 ps 5 microseconds
(0.000005 second)
2 ps 2 microseconds
(0.000002 second)
1 ps 1 microsecond
(0.000001 second)
.5 ps half a microsecond
(0.0000005 second)
.2 ps 0.2 microsecond
(0.0000002 second)
.1 ps 0.1 microsecond
(0.0000001 second)
.05 ps 0.05 microseconds
(0.00000005 second)
11
.
12
.
12
THE TRIGGER SYSTEM PART I
Figure 9.
TRIGGERING GIVES YOU A STABLE DISPLAY because the same trigger point starts
the sweep each time. Slope and level controls define the trigger points on the trigger
signal. When you look at a waveform on the screen, you’re seeing all those sweeps
overlaid into what appears to be one picture.
Instruments like those in the trols the trigger sources and the
Tektronix 2200 Portable Oscillo- external trigger coupling. At the
scope family offer a variety of bottom of the column of trigger
trigger controls. Besides those controls is the external trigger
already mentioned, you also input BNC connector.
have controls that determine On dual time base 2215
how the trigger system operates scopes, there is a slightly differ-
(trigger operating mode) and ent control panel layout be-
how long the scope waits be- cause you can have a separate
tween triggers (holdoff). trigger for the B sweep.
The control positions are illus-
Trigger Level and Slope
trated by the foldout at the end
of the primer. All are located on These controls define the trigger
point. The SLOPE control de-
the far right of the front panel. On
termines whether the trigger
the 2213, the variable trigger
point is found on the rising or the
holdoff (VAR HOLDOFF) is at the
top, and immediately below it is falling edge of a signal. The
LEVEL control determines Figure 10.
the trigger MODE switch. Below
where on that edge the trigger SLOPE AND LEVEL CONTROLS determine where on the trigger signal the trigger
that the trigger SLOPE and actually occurs. The SLOPE control specifies either a positive (also called the rising
LEVEL controls are grouped. point occurs. See Figure IO.
orpositive-going) edge or on a negative (falling or negative-going) edge. The LEVEL
Then a set of three switches con- control allows you to pick where on the selected edge the trigger event will take
place.
13
THE TRIGGER SYSTEM CONT.
Trigger Sources
Trigger sources are grouped
into two categories that depend
on whether the trigger signal is
provided internally or externally.
The source makes no difference
in how the trigger circuit oper-
ates’but internal triggering usu-
ally means your scope is trigger-
ing on the same signal that it is
14
PART I
displaying. That has the obvious on the front panel. Occasions on the channel. In the 2200 the scope triggers alternately on
advantage of letting you see when external triggering is use- Series, the automatic mode is a the two vertical channels. That
where you’re triggering. ful often occur in digital design signal-seeking auto mode. This means you can look at two com-
Two switches on the front and repair; there you might want means that for most of the sig- pletely unrelated signals. Most
panel (labeled SOURCE and to look at a long train of very nals you’ll be measuring, the scopes only trigger on one
INT) determine the trigger similar pulses while triggering auto mode will match the trigger channel or the other when the
source. The internal triggering with an external clock or with a level control to the trigger signal. two signals are not synchro-
sources are enabled when you signal from another part of the That makes it most unlikely that nous.
move the SOURCE lever to INT. circuit. you will set the trigger level con- Here’s a review of the 2200
In this position, you can trigger The LINE position on the trol outside of the signal range. trigger modes:
the scope on the signal from SOURCE switch gives you The auto mode lets you trigger
Trigger
either channel, or you can another triggering possibility: on signals with changing volt- Operating Mode Switch Settings
switch to VERT MODE. the power line. Line triggering is age amplitudes or waveshapes normal NORM on the MODE switch
Triggering on one of the useful anytime you’re looking at without making an adjustment of automatic AUTO on the MODE switch
channels works just like it circuits that are dependent on the LEVEL control. television field TV FIELD on the MODE
switch
sounds: you’ve set the scope to the power line frequency. Another useful operating television line NORM or AUTO on the
trigger on some part of the Examples include devices like mode is television triggering. MODE switch
vertical mode VERT MODE on the INT
waveform present on that light dimmers and power Most scopes with this mode let switch
channel. supplies. you trigger on tv fields at
Using the VERTICAL MODE These are all the trigger sweeps of 100 &division and
setting on the internal source source possibilities on a 2200 slower, and tv lines at 50 psldiv
switch means that the scope’s Series scope: or faster. With a 2200 Series
VERTICAL MODE switches de- Switch Positions
scope, you can trigger on either
termine what signal is used for Trigger Source SOURCE INT
fields or lines at any sweep
triggering. If the VERTICAL channel 1 only INT CH1
speed; for tv field triggering, use
MODE switches are set at CH 1, channel 2 only INT CH2 the TV FIELD switch position,
external EXT disabled and for television line triggering,
then the signal on channel 1 line LINE disabled
triggers the scope. If you’re look- vertical mode INT VERT MODE use the NORM or AUTO
ing at channel 2, then that chan- (either channel settings.
1 or 2 or both)
nel triggers it. If you switch to the You’ll probably use the normal
alternate vertical mode, then the Trigger Operating Modes and automatic modes the most
scope looks for triggers alter- The 2200 Series trigger circuits often. The AUTO because it’s
nately on the two channels. If the can operate in four modes: nor- essentially totally automatic,
vertical mode is ADD, then CH 1 mal, automatic, television, and and normal because it’s the
+ CH 2 is the triggering signal. vertical mode. most versatile. For example, it’s
And in the CHOP vertical mode, One of the most useful is the possible to have a low fre-
the scope triggers the same as normal trigger mode (marked quency signal with a repetition
in ADD, which prevents the in- NORM on the MODE switch) rate that is mismatched to the
strument from triggering on the because it can handle a wider run-out of an automatic mode
chop frequency instead of your range of trigger signals than any timer; when that happens the
signals. other triggering mode. The nor- signal will not be steady in the
You can see that vertical mal mode does not permit a auto mode. Moreover, the auto-
mode triggering is a kind of au- trace to be drawn on the screen matic signal-seeking mode
tomatic source selection that if there’s no trigger. The normal can’t trigger on very low fre-
you can use when you must mode gives you the widest quency trigger signals. The
switch back and forth between range of triggering signals: from normal mode, however, will give
vertical modes to look at differ- DC to 60 MHz. you a steady signal at any rep
ent signals. In the automatic (or “bright rate.
But triggering on the dis- baseline”) mode (labeled AUTO The last 2200 Series trigger
played signal isn’t always what on the front panel): a trigger operating mode, the vertical
you need, so external triggering starts a sweep; the sweep ends mode, is unique in its advan-
is also available. It often gives and the holdoff period expires. tages. Selecting the VERT
you more control over the dis- At that point a timer begins to MODE position on the INT
play. To use an external trigger, run; if another trigger isn’t found switch automatically selects the
you set the SOURCE switch to before the timer runs out, a trig- trigger source as you read in
its EXT position and connect the ger is generated anyway caus- “Trigger Source”above. It also
triggering signal to the BNC ing the bright baseline to appear makes alternate triggering pos-
connector marked EXT INPUT even when there is no waveform sible. In this operating mode,
15
THE TRIGGER SYSTEM CONT.
Triggering Coupling
Just as you may pick either al-
ternating or direct coupling
when you connect an input sig-
nal to your scope’s vertical sys-
tem, you can select the kind of
coupling you need when you
connect a trigger signal to the
trigger system’s circuits. For
internal triggers, the vertical
input coupling selects the trig-
ger coupling. For external trig-
ger signals, however, you must
select the coupling you want:
Figure 13.
Coupling Applications
THE TRIGGER CIRCUIT AND ITS CON-
TROLS are shown in the diagram above.
DC DC couples all elements of the
Trigger source describes whether or not
triggering signal (both AC and
DC) to the trigger circuit.
the trigger signal is internal or external to
the scope. Coupling controls the con-
DC with If you want DC coupling and nection of an external trigger to the trig-
attenuation the external trigger is too large ger circuit. The level and slope controls
for the trigger system, move determine where the trigger point will be
the TRIGGER COUPLING on the trigger signal. And the mode con-
switch to its DC+10 setting. trol determines the operations of the
AC This coupling blocks DC
trigger circuit.
components of the trigger
signal and couples only the AC
components.
16
CHAPTER 5. ALL ABOUT PROBES PART I
Connecting all the measure- plus the cable plus the scope’s
ment test points you’ll need to own input, the 10X attenuator
the inputs of your oscilloscope is probe introduces about ten
best done with a probe like the times less capacitance, as little
one illustrated in Figure 14. as IO-14 picofarads (pF). The
Though you could connect the penalty is the reduction in signal
scope and circuit-under-test amplitude from the 1O:1
with just a wire, this simplest of attenuation.
all possible connections would These probes are adjustable
not let you realize the full to compensate for variations in
capacities of your scope. The oscilloscope input capacitance
connection would probably load and your scope has a reference
the circuit and the wire would signal available at the front
act as an antenna and pick up panel. Making this adjustment is
stray signals - 60 Hz power, calledprobe compensation and
CBers, radio and tv stations - you did it as the first step in
and these would be displayed Exercise 3 of Chapter 2.
on the screen along with the Remember when you are
signal of interest. measuring high frequencies,
that the probe’s impedance (re-
Circuit Loading sistance and reactance)
Using a probe instead of a bare changes with frequency. The
wire minimizes stray signals, but probe’s specification sheet or
there’s still an effect from putting manual will contain a chart like
a probe in a circuit called circuit that in Figure 15 that shows this
loading. Circuit loading modi- change. Another point to re-
fies the environment of the sig- member when making high fre-
nals in the circuit you want to quency measurements is to be
measure; it changes the signals sure to securely ground your
in the circuit-under-test, either a probe with as short a ground
little or lot, depending on how clip as possible. As a matter of
great the loading is. fact, in some very high fre- Figure 14.
Circuit loading is resistive, PROBES CONNECTTHE SCOPE AND THE CIRCUIT-UNDER-TEST. Tektronix probes
quency applications a special consist of a patented resistive cable and a grounded shield. Two P6120 probes and
capacitive, and inductive. For socket is provided in the circuit the accessories pictured above are supplied with every 2200 Series scope. The
signal frequencies under 5 kHz, and the probe is plugged into probe is a high impedance, minimum loading 1O X passive probe. The accessories
the most important component that. for each probe (from left to right) are: a grabber tip for ICs and small diameter leads; a
of loading is resistance. To avoid retractable hook tip; and IC tester tip cover; an insulating ground cover; marker
significant circuit loading here, Measurement System bands; and (in the center) the ground lead.
all you need is a probe with a Bandwidth
resistance at least two orders of Then there is one more probe
magnitude greater than the cir- characteristic to consider: For example, if both probe and you use the particular probe de-
cuit impedance (100 Ma probes bandwidth. Like scopes, probes scope have rise times of 5.83 signed for that instrument. For
for 1 Ma sources; 1 MSZ probes have bandwidth limitations; nanoseconds: example, in the case of the 2200
for IO k0 sources, and so on). each has a specified range Series scopes and the P6120
When you are making mea- within which it does not at- Tr (system) = v Tr21scof3el + Tr2(probe) 10X Passive Probe, the probe
surements on a circuit that con- tenuate the signal’s amplitude and the scope have been de-
tains high frequency signals, in- more than -3 dB (0.707 of the T,=d34+34 signed to function together and
ductance and capacitance be- original value). But don’t as- you have the full 60 MHz band-
come important. You can’t avoid sume that a 60 MHz probe and a That works out to 8.25 width at the probe tip..
adding capacitance when you 60 MHz scope give you a 60 nanoseconds, the equivalent to
make connections, but you can MHz measurement capability. a bandwidth of 42.43 MHz be-
avoid adding more capacitance The combination will approxi- cause:
than necessary. mately equal the square root of
the sum of the squares of the 350
One way to do that is to use an B wlmegahertz) =
attenuator probe; its design rise times (also see Chapter 10). Tr (nanoseconds)
greatly reduces loading. In-
stead of loading the circuit with To get the full bandwidth from
capacitance from the probe tip your scope, you need more
bandwidth from the probe. Or
17
ALL ABOUT PROBES CONT.
The first five chapters described discussions in Chapter 7 start Chapter 9 discusses mea-
how to select the exact oscillo- with safety because you should surement techniques, begin-
scope functions you need to always observe safety precau- ning with fundamental time and
make the measurements you tions when working on electrical amplitude measurements and
want. Now you can put what equipment. ending with delayed sweep
you’ve learned into practice with The first step in ensuring ac- measurements.
this section of the primer. curate measurements is making The last chapter in the primer
It begins with a review of sure your scope is set up prop- describes oscilloscope per-
waveform shapes and charac- erly, and this subject is dis- formance and how it affects your
teristics in Chapter 6. Then the cussed in Chapter 8. measurements.
CHAPTER 6. WAVEFORMS
The definition of a wave is “a
disturbance traveling through a
Figure 16.
medium” while the definition of a BASIC WAVESHAPES include sine
waveform is “a graphic repre- waves, and various non-sinusoidal
SINE WAVE
sentation of a wave.” waves such as triangle waves, square
Like a wave, a waveform is waves, and sawtooth waves. A square
wave has equal amounts of time for its
dependent on two things: two states. Triangle and sawtooth waves
movement and time. The ripple are usually the result of circuits de-
on the surface of a pond exists signed to control voltage with respect to
as a movement of water in time. time, like the sweep of an oscilloscope
and some television circuits. In these
The waveform on your scope’s waveforms, one (or both) transitions
screen is the movement of an from state to state are made with a
electron beam during time. steady variation at a constant rate, a
The changes in the waveform ramp. (Changes from one state to
another on all waveforms except sine
with time form the waveshape, waves are called transitions.) The last
the most readily identifiable
-fRIAIWLE WAVE
two drawings represent aperiodic,
characteristic of a waveform. single-shot waveforms. The first is a
Figure 16 illustrates some com- pulse; all pulses are marked by a rise, a
finite duration, and a decay. The second
mon waveshapes. one is a step, which is actually a single
transition.
PUISE STEP
19
MAKING MEASUREMENT CONT.
Figure 18.
PERIOD IS THE TIME REQUIRED FOR ONE CYCLE OF A SIGNAL if the signal
repeats itself. Period is a parameter whether the signal is symmetrically shaped like
the sine and square waves above, or whether it has a more complex and asymmetri-
cal shape like the rectangular wave and damped sine wave. Period is always
expressed in units of time. Naturally, one-time signals like the step or uncorrelated
signals (without a time relation) like noise have no period.
20
PART II
Figure 19.
IF A SIGNAL IS PERIODIC, IT HAS A FREQUENCY. Frequency is the number of times
a signal repeats itself in a second; frequency is measured in Hertz: 1 Hz = 1 cycle per
second; 1 kHz (kilohertz) = 1000 cycles/second; and 1 MHz (megahertz) =
1 ,OOO,OOO cycles/second. Period and frequency are reciprocal: l/period = fre-
quency, and l/frequency = period. For example, a 7 Hz signal has a period of 0.143
seconds: 1/7 Hz = 0.143 s, and 1/0.143 s = 7 Hz.
Figure 22.
PHASE is best explained with a sine
wave. Remember that this waveform is
based on the sine of all the angles from 0
through 360. The result is a plot that
Figure 21. changes from 0 to 00,l at 900,O again at
DUTY CYCLE, DUTY FACTOR, AND 1800, -1 at 270”, and finally 0 again at
REPETITION RATE are parameters of all 360”. Consequently, it is useful to refer to
rectangular waves. They are particularly the phase angle (or simply phase, when
important in digital circuitry. Duty cycle there is no ambiquity) of a sine wave
is the ratio of pulse width to signal period when you want to describe how much of
expressed as a percentage. For square the period has elapsed. Another use of
waves, it’s always 50% as you can see; phase is found when you want to de-
for the pulse wave in the second draw- scribe a relationship between two sig-
ing, it’s 30%. Duty factor is the same nals. Picture two clocks with their sec-
thing as duty cycle except it is ex- ond hands sweeping the dial every 60
pressed as a decimal, not a percentage. seconds. If the second hands touch the
A repetition rate describes how often a twelve at the same time, the clocks are in
pulse train occurs and is used instead of phase; if they don’t, then they’re out of
frequency to describe waveforms like phase. To express how far out of phase
that in the second drawing. they are, you use phase shift in degrees.
To illustrate, the waveform labeled
CURRENT in the drawing above is said
to be 90” out of phase with the voltage
waveform. Other ways of reporting the
Figure 20. same information are “the current
THE PARAMETERS OF A PULSE can be important in a number of different applica- waveform has a 90 degree phase angle
tions. Digital circuitry, X-ray equipment, and data communications are examples. with respect to the voltage waveform”or
Pulse specifications include transition times measured on the leading edge of a “the current waveform lags the voltage
positive-going transition; this is the rise time. Fall time is the transition time on a waveform by 9OO.” Note that there is al-
negative-going trailing edge. Pulse width is measured at the 50% points and ways a reference to another waveform;
amplitude from 0 to 100%. Any displacement from 0 volts for the base of the pulse is in this case, between the voltage and
the baseline offset. current waveforms of an inductor.
21
CHAPTER 7. SAFETY
Before you make any oscillo- service electrical devices alone. under-test. Remember that if Then there are few rules about
scope measurement, remember Know the symbols for danger- you lose the ground, all ac- the scope itself: To avoid a
that you must be careful when ous circuits and observe the cessible conductive parts - in- shock, plug the power cord of
you work with electrical equip- safety instructions for the cluding knobs that appear to be the scope into a properly-wired
ment. Always observe all safety equipment you’re working on. insulated - can give you a receptacle before connecting
precautions described in the Don’t operate an electrical de- shock. To avoid personal injury, your probes; only use the power
operators or service manual for vice in an explosive atmos- don’t touch exposed connec- cord for your scope, and don’t
the equipment you’re working phere. Always ground the scope tions and components in the use one that isn’t in good condi-
on. to the circuit, and ground both circuit-under-test when the tion (cracked, broken, missing
Some general rules about your scope and the circuit- power is on. And remember to ground pin, etc.). Use the right
servicing electrical equipment consult the service manual for fuse to avoid fire hazards. Don’t
are worth repeating here. Don’t the equipment you’re working remove covers and panels on
on. your scope.
controls: variable controls (CH l Then check your trigger sys- and plug. Check the circuit
measurement with an attenuator
1 and CH 2 variable VOLTS/ tem controls to make sure your ground by touching the probe
probe, you should make sure it’s
DIV) should be in their cali- scope will pick the right slope tip to the point you think is
compensated. Figure 23 illus-
brated detent positions; make on the trigger signal, that the ground before you make a hard
trates what can happen to the
sure CH 2 isn’t inverted (unless right coupling is selected, and ground by attaching the ground
waveforms you’ll see when the
you want it to be); check the that the correct operating strap of your probe.
probe is not properly compen-
vertical mode switches to mode will be used. Also make If you’re going to be probing a
sated.
make sure the signal from the sure that the trigger variable lot of different points in the same
Note that you should com- holdoff control is at its mini- circuit and measuring frequen-
pensate your probe as it will be proper channel(s) will be dis-
played; check the two vertical mum position. cies less than 5 MHz, you can
used when you make the mea- ground that circuit to your scope
surement. Compensate it with system VOLTS/DIV switches to
see if their settings are right once instead of each time you
the accessory tip you’ll be using move the probe. Connect the
and don’t compensate the (and don’t forget to use the
VOLTS/DIV readout that circuit ground to the jack
probe in one vertical channel marked GND on the front panel.
and then use it on another. matches the probe, either IX or
10X); check the input coupling
levers too.
22
GETTING STARTED CONT. PART II
Figure 23.
IMPROPERLY COMPENSATED PROBES can distort the waveforms you see on the
screen of your scope. In the photographs the probe adjustment signal and a 1 MHz
square wave are shown as they will appear with proper and improper compensa-
tions. Notice the amplitude and ringing changes on the square wave with the
differences in compensation.
23
CHAPTER 9. MEASUREMENT TECHNIQUES
24
PART II
Use the directions in Exercise Phase Measurements Note that general purpose surement as long as your known
9 to make a pulse measurement You know that a waveform has oscilloscope Lissajous pattern signal is accurate and both sig-
on the probe adjustment square phase, the amount of time that phase measurements are usu- nals are sine waves. The pat-
wave. has passed since the cycle be- ally limited by the frequency re- terns you can see are illustrated
gan, measured in degrees. sponse of the horizontal ampli- in Figure 26, where the effects of
There is also a phase relation- fier (typically designed with far both frequency and phase dif-
Exercise 9. ship between two or more less bandwidth than vertical ferences are shown.
PULSE WIDTH waveforms: the phase shift (if channels). Specialized X-Y Component checking in ser-
MEASUREMENTS any). There are two ways to scopes or monitors will have al- vice or production situations is
To measure the pulse width of measure the phase shift be- most identical vertical and hori- another X-Y application; it re-
the probe adjustment square tween two waveforms. One is by zontal systems. quires only a simple transistor
wave quickly and easily, set your putting one waveform on each checker like that shown in Fig-
channel of a dual-channel X-Y Measurements
scope to trigger on and display ure 27
scope and viewing them directly Finding the phase shift of two
channel 7. Your probe should sinusoidal signals with a Lissa- There are many other applica-
still be connected to the channel in the chop or alternate vertical
mode; trigger on either channel. jous pattern is one example of tions for X-Y measurements in
7 BNC connector and the probe an X-Y measurement. The X-Y television servicing, in engine
adjustment jack from the previ- Adjust the trigger LEVEL control
for a stable display and mea- capability can be used for other analysis, and in 2-way radio ser-
ous exercises. Use 0.7 ms / measurements as well. The Lis- vicing, for examples. In fact, any
division and the no delay hori- sure the period of the wave-
forms. Then increase the sweep sajous patterns can also be time you have physical phe-
zontal mode (A sweep if you’re used to determine the fre- nomena that are interdependent
using a 2275). Use AUTO trig- speed so that you have a dis-
play something like the second quency of an unknown signal and not time-dependent, X-Y
gering on the positive slope and when you have a known signal measurements are the answer.
adjust the trigger LEVEL control drawing back in Figure 22. Then
measure the horizontal distance on the other channel. This is a Aerodynamic lift and drag,
to get as much of the leading very accurate frequency mea- motor speed and torque, or
edge as possible on your between the same points on the
screen. Switch the coupling on two waveforms. The phase shift
channel 7 to ground and center is the difference in time divided
the baseline on the center hori- by the period and multiplied by
zontal graticule. Now use AC 360 to give you degrees.
coupling because that will cen- Displaying the two waveforms
ter the signal on the screen and and measuring when one starts
you make pulse width mea- with respect to another is possi-
surements at the 50 % point of ble with any dual trace scope,
the waveform. Use your horizon- but that isn’t the only way to
tal POSITION control to line up make a phase measurement.
the 50 % point with the first major Look at the front panel and you’ll
see that the vertical channel
oooo/\oovoo
graticule from the left side of the
screen. Now you can count divi- BNC connectors are labeled X
sions and subdivisions across and Y. The last position on the
the center horizontal and multi- SEC/DlV switch is XY, and when +
ply by the SEC/D/V switch set- you use it, the scope’s time base o”zo30f45 90 135” 700’
ting to find the pulse width. is bypassed. The channel 1
input signal is still the horizontal
axis of the scope’s display, but
now the signal on channel 2 be-
comes the vertical axis. In the
3 v\MDoooNooov\
0” 15O 30° 60” TO” lZO”
X-Y mode, you can input one
sinusoidal on each channel and
your screen will display a Lissa-
jous pattern. (They are named
for Jules Antoine Lissajous, a
French physicist; say “LEE-za-
shu”). The shape of the pattern Figure 26.
will indicate the phase differ- FREQUENCY MEASUREMENTS WITH LISSAJOUS PATTERNS require a known sine
ence between the two signal. wave on one channel. If there is no phase shift, the ratio between the known and
Some examples of Lissajous unknown signals will correspond to the ratio of horizontal and vertical lobes of the
pattern. When the frequencies are the same, only the shifts in phase will affect the
patterns are shown in Figure 26. pattern. In the drawings above, both phase and frequency differences are shown.
26
PART II
Figure 27.
X-Y COMPONENT CHECKING requires the transistor checker shown above. With it
connected to your scope and the scope in the X-Y mode, patterns like those
illustrated indicate the component’s condition. The waveforms shown are found
when the components are not in a circuit; in-circuit component patterns will differ
because of resistors and capacitors associated with the component.
can simply switch the vertical delay out to where you want the
operating mode to display both sweep to start.
channels. That makes the total But the delayed sweep fea-
holdoff time for one channel ture you’ll probably use the most
greater than one field period. often is the intensified sweep; it
Then just position the unused lets you use the delayed sweep
vertical channel off-screen to as a positionable magnifier. You
avoid confusion. trigger normally and then use
c SCOPE
It is also important to select
the trigger slope that corre-
sponds to the edge of the
the scope’s intensified horizon-
tal mode. Now the signal on the
screen will show a brighter zone
waveform where the sync after the delay time. Run the
pulses are located. Picking a delay time (and the intensified
negative slope for pulses at the zone) out to the part of the signal
bottom of the waveform allows that interests you. Then switch to
you to see as many sync pulses the delayed mode and increase
as possible. the sweep speed to magnify the
When you want to observe the selected waveform portion so
TV line portion of the composite that you can examine it in detail.
video signal, use the NORM Since the 2200 Series has two
trigger mode and trigger on the types of delayed sweep, read
horizontal synchronization the paragraphs and use the de-
pulses for a stable display. It is layed sweep measurement
usually best to select the blank- exercise below that applies to
ing level of the sync waveform your scope: “Single Time Base
so that the vertical field rate will Scopes” and Exercise 10 for de-
not cause double triggering. layed sweep measurements
with single time base scopes
Delayed Sweep like the Tektronix 2213; or “Dual
Measurements Time Base Scopes”and Exer-
Delayed sweep is a technique cise 11 for delayed sweep on
that adds a precise amount of dual time base scopes like the
time between the trigger point
2215.
and the beginning of a scope
sweep. Often delayed sweep is Single Time Base Scopes
used as a convenient way to Very few single time base
Figure 29.
make a measurement (the rise scopes offer delayed sweep
USING THE Z-AXIS can provide additional information on the scope screen. In the time measurement in Exercise measurements. Those that do
set-up drawn above, a function generator sweeps through the frequencies of interest 10 is a good example). To make may have measurement
during the product testing-20 to 20,OOO Hz, in this case. Then an adjustable notch a rise time measurement without capabilities similar to those of
filter is used to generate a marker, at 15 kHz, for instance, and this signal is applied to
the Z-axis input to brighten the trace. This allows the tester to evaluate the product’s
delayed sweep, you must trig- the Tektronix 2213 which has
performance with a glance. ger on the edge occurring be- three possible horizontal opera-
fore the desired transition, With ting modes annotated on the
delayed sweep, you may front panel as NO DLY, INTENS,
Using TV Triggering mode. This mode allows the choose to trigger anywhere and DLY’D.
The composite video waveform scope to trigger at the field rate along the displayed waveform When you set the HORIZON-
consists of two fields, each of of the composite video signal on and use the delay time control to TAL MODE switch to NO DLY (no
which contains 262 lines. Many either field one or field two. start the sweep exactly where delay), only the normal sweep
scopes offer television trigger- Since the trigger system cannot you want. functions.
ing to simplify looking at video recognize the difference be- Sometimes, however, de- When you choose INTENS (in-
signals. Usually, however, the tween field one and field two, it layed sweep is the only way to tensified sweep), your scope
scope will only trigger on fields will trigger alternately on the two make a measurement. Suppose will display the normal sweep
at some sweep speeds and fields and the display will be that the part of the waveform you and the trace will also be inten-
lines at others. The 2200 Series confusing if you look at one line want to measure is so far from sified after a delay time. The
scopes allow you to trigger on at a time. the only available trigger point amount of delay is determined
either lines or fields at any To prevent this, you add more that it will not show on the by both the DELAY TIME switch
sweep speed. holdoff time, and there are two screen The problem can be (you can use 0.5 ps, 10 ps, or
To look at tv fields with a 2200 ways to do that. You can use the solved with delayed sweep: 0.2 ms) and the DELAY TIME
Series scope, use the TV FIELD variable holdoff control, or you trigger where you have to. and MULTIPLIER control. The multi-
28
PART II
30
PART II
31
INDEX
A D F
A sweep 29 DC coupling 6 Frequency measurements 25 Initialization 3
A/B SWP SEP 8 Deflection voltages 6 Focus control 4 Input coupling control 6
AC coupling 6 Delay time 29 Internal triggering 14
Alternate horizontal operating Delay time control 11 G Intensified sweep 28
mode 10 Delay time multiplier (DTM) GND input coupling 6 Intensity control 4
Alternate sweep separation control 10 Graticule 4
control 8 Delayed sweep 28 L
Amplitude 20 Delayed sweep measurements
H Line triggering 15
Holdoff 14 Lissajous figures 26
Amplitude measurements 24 28
Horizontal magnification
Attenuation 6 Delaying sweep 29
control 11 M
Derived measurements 25 Magnification 11
B Differential measurements 27
Horizontal operating mode
Major divisions 4
B delay time position control 11 control 10
Display system 4 Measurement system band-
B sweep 29 Horizontal position control 10
Divisions 4 width 17
BNC 8 Horizontal system 10
Dual time base scopes 29 Measurement techniques,
Bandwidth and rise time 33
Duty cycle 21 Amplitude 24
Bandwidth specifications 33
Duty factor 21 Differential 27
Beam finder control 4
Delayed sweep 28
Blanking 14 E Derived 25
Electron beam 4
C Exercises, 24 Fall time 25
Capacitance 17 Frequency 25
Amplitude measurement 24
Cathode ray tube (CRT) 2 Period 24
Controlling vertical
Channel 2 inversion control 7 Phase 26
sensitivity 9
Circuit loading 17 Compensating the probe 9 Pulse 25
Compensating the probe 9 Pulse width 26
Delayed sweep 29
Compensation box 9 Rise time 29
Scope initialization 3
Coupling 6 Time 24
Signal coupling 9
video 28
Time measurement 24
X-Y 26
Using the display controls 5
Minor divisions 4
Using the horizontal
controls 12 0
Using the trigger controls 16 Oscilloscope performance 32
Using the vertical controls 9
Vertical operating mode 9
External triaaerina 6
34
PART II
P T U
Parallax errors 4 Television triggering 28 Unblanking 14
Period 24 Termination 18
Phase 21 Time base 10 V
Picking a probe 18 Time measurements 24 Variable SEC/DlV control 11
Phase measurements 26 Trace rotation control 5 Variable trigger holdoff
Phosphor 4 Transducers 2 control 14
Position controls, Transitions 19 Variable VOLTS/DIV control 7
horizontal 10 Triangle waves 19 Vertical operating mode
vertical 6 Trigger, control 7
Probes, automatic mode (AUTO) 15 Vertical operating modes, 7
accessories 17 coupling control 16 ADD 8
attenuator 18 external 14 ALT (alternate) 7
adjustment jack 4 holdoff control 14 BOTH 7
bandwidth 17 internal 14 CHOP 7
compensation 17 line 15 CH1 7
current 18 level control 12 CH27
handling 22 normal mode (NORM) 15 Vertical mode triggering 15
loading 17 operating modes control 15 Vertical position control 6
passive 18 point 13 Vertical sensitivity control 6
Pulse measurements 25 slope control 12 Vertical system 6
Vertical system coupling 6
R
source control 14
system 12 VOLTS/DIV control 6
Ramp 10
Repetition rate 21
TV FIELD mode 15
W
Retrace 10 Waveforms,
Rise time and bandwidth 33 Pulse 19
Rise times, scope and mea- Sawtooth 19
surement 32 Sine 19
Square 19
S Step 19
Safety 22 Triangle 19
Sawtooth waves 19
Scale factors 6 X
SEC/DIV control 10 X axis 5
Screen 4 X-Y measurements 26
Sine waves 19
Single time base scopes 28
Y
Y axis 5
Square waves 19
Square wave and high fre- Z
quency response 32 Z axis 5
Steps 19
Subdivisions 4
Sweep generator 10
Sweep speeds 10
35