Professional Documents
Culture Documents
THERMOCOUPLES
2). The advantages of 2, 3 & 4 wire pT100 measurements.
3). Introduction to Thermocople measurements.
4). Basic RTD measurements
Fig. 1.1
The contact potential is not measurable for a single connection (or junction), but
when two junctions are in a circuit with the junctions at different temperatures
then a voltage of a few millivolts can be detected (Fig. 1.1). This voltage will be
zero
if the junctions are at the same temperature, and will increase as the temperature
of one junction relative to the other is changed until a peak is reached.
Fig. 1.2
The shape of the typical characteristic is shown in Fig. 1.2, from which you can
see that the thermocouple is useful only over a limited range of temperature due
to the non-linear shape of the characteristic and the reversal that takes place at
temperatures higher than the turn-over point.
2
The output from a thermocouple is small, of the order of millivolts for a 10°C
temperature difference, and Fig. 1.3 shows typical sensitivity and useful range
for a variety of the common types. Of these, the copper/constantan type is used
mainly for the lower range of temperatures and the platinum! rhodium type for
the higher temperatures.
Fig.1.3
Because of the small voltage output, amplification is usually needed unless the
thermocouple is used for temperature measurement along with a sensitive milli
voit meter. If the output of the thermocouple is required to drive anything more
than a meter movement, then DC amplification will be needed, using an
operational amplifier or chopper amplifier.
The type of amplifier that is used needs to be carefully selected, because good
drift stability is necessary unless the device is recalibrated at frequent intervals.
This
makes the chopper type of amplifier preferable for most applications.
CONNECTION CONFIGURATION
Platinum resistors are most commonly 100 ohms at 0 °C and 138.51 ohms at
100 °C. They are available in different grades according to the accuracy
required and can be supplied as duplex sensors, two independent sensors on a
single former. The respective accuracies of the three main specification types,
BS.EN 60751 Class A, BS.EN 60751 Class B and 1/10th Class B. Pt100
Platinum Resistance sensors are as shown in the tolerance table below.
Resistance at 38.51
Tolerances for 100 ohms
ohms fundamental
Thermometers
interval
1/10 Class
Temperature Class A Class B
B
Nominal ± ± ±
°C °F ± °C ± °C ± °C
Value ohm ohm ohm
-200 -328 18.52 0.55 0.24 1.3 0.56 0.13 0.06
-100 -148 60.26 0.35 0.14 0.8 0.32 0.08 0.03
0 32 100.00 0.15 0.06 0.3 0.12 0.03 0.01
100 212 138.51 0.35 0.13 0.8 0.30 0.08 0.03
200 392 175.86 0.55 0.20 1.3 0.48 0.13 0.05
300 572 212.05 0.75 0.27 1.8 0.64 0.18 0.06
400 752 247.09 0.95 0.33 2.3 0.79 0.23 0.08
500 932 280.98 1.15 0.38 2.8 0.93 0.28 0.09
600 1112 313.71 1.35 0.43 3.3 1.06 0.33 0.10
700 1292 345.28 - - 3.8 1.17 - -
800 1472 375.70 - - 4.3 1.28 - -
5
LEADS
Pt100 sensors are supplied with 2, 3 or 4 wire connections and, unless otherwise
specified, will be supplied as 3 wire type 7 x 0.2 mm Cu PTFE insulated, with
two red wires indicating one end of the element and one white wire indicating
the other. Alternative types of wire insulation can be supplied.
Flying lead sensors can be supplied with an optional stainless steel over braid or
convolute sheathing for more arduous environments.
6
INTRODUCTION
Two, Three and Four wire measurement techniques have been developed for
measuring accurately the resistance of resistive temperature detectors (RTD).
This application note looks at the new techniques being adopted in ‘Smart’
instruments where, using the intelligence of a microprocessor, the traditional
drawbacks of 3 wire systems no longer apply.
A voltage excites the bridge and the voltage across the bridge is proportional to
the resistance of the RTD.A problem occurs when we introduce lead resistances
(See figure below). It is apparent that any resistance in the lead looks as though
there is additional resistance in the element to be measured.
To minimize these errors the three wires compensated bridge was introduced
(see figure below).
7
This has the effect of removing the error introduced by the lead resistance as
long as lead resistances RL1 and RL3 are matched.
However the effect of the lead resistance can be to cause less current to flow in
the detector leg and hence introduce a small but possibly significant span error.
This can be eliminated by exciting the bridge from a constant current source
rather than a constant voltage and so whatever the lead resistance, the same
current always flows through the detector. With this method there are no lead
resistance errors introduced as long as the lead resistances are equally matched.
In practice they are very closely matched as long as the wire used is part of the
same multi-core cable.
The exception to this is when the sensor is used in a Hazardous area and
connected to the bridge circuit via a Zener Barrier. Here any mis-match in the
resistance of the two legs of the Zener barrier can appear as a sensor error.
Although still small, this error can be as much as 0.15 ohm or approx.
0.3°C.(MTL 155 Barrier).For analogue transmitters Status Instruments Ltd.
have traditionally used a variation of this technique using an in house active
bridge circuit. The exception being the new ‘Smart’ series of instruments which
use a different technique which will be explained later. Another way of
measuring Pt100 elements is to use a 4 wire current and voltage method (see
figure below).
Here the detector is excited by a constant current and the voltage across the
detector measured by an amplifier with a high impedance input. If the current
source is perfect and the input impedance of the voltage measuring circuit is
infinite, then there is no error whatsoever introduced by the lead resistances
even if they are mismatched.
8
In addition, the microprocessor can determine which if any, of the RTD inputs
has become disconnected and detect other errors such as RTD short circuit. This
is an improvement over both conventional three and four wire circuits because
you can now have a predictable failure mode which does not depend upon
which of the three wires has become disconnected.
This technique removes lead resistance effects as long as they are equal. Again,
we have the problem when using Zener barriers, in that if the legs of the barrier
are not accurately matched, then a small error could be introduced.
Introduction
The most common devices used for sensing temperature include thermocouples,
resistance temperature detectors (RTDs), and thermistors. Each has unique
characteristics and properties that make one more suitable than another for a
certain application.
Thermocouples are the most widely used device for sensing temperature, and
probably the least understood. They are simple and efficient, and provide a
small voltage signal proportional to the temperature difference between two
junctions in a closed thermoelectric circuit. In its most basic configuration, one
9
This medium can be gas, liquid, or solid, but in all cases, the medium shall not
be allowed to chemically, electrically, or physically contaminate or alter the
thermocouple junction. For special applications or to protect them from the
environment, thermocouples are available with protective coatings and shields
or sheaths. RTDs are composed of metals with a high positive temperature
coefficient of resistance. Most RTDs are simply wire-wound or thin-film
resistors made of wire with a known resistance vs. temperature relationship.
Platinum is one of the most widely used materials for RTDs. They come in a
broad range of accuracies, and the most accurate are also used as NIST
(National Institute of Standards and Technology) temperature standards.
Thermistors are similar to RTDs in that they also change resistance between
their terminals with a change in temperature. However, they can be made with
either a positive or negative temperature coefficient. In addition, they have a
much higher ratio of resistance change per degree C (several %) than RTDs,
which make them more sensitive.
In principle, a TC can be made from any two dissimilar metals such as nickel
and iron. In practice, however, only a few TC types have become standard
because their temperature coefficients are highly repeatable, they are rugged,
and they generate relatively large output voltages. The most common
thermocouple types are called J, K, T, and E, followed by N28, N14, S, R, and
B. The junction temperature could be inferred from the See beck voltage by
consulting standard tables. However, this voltage cannot be used directly
because the thermocouple wire connection to the copper terminal at the
measurement device itself constitutes a thermocouple junction (unless the TC
lead is also copper) and generates another emf that must be compensated.
10
The classical method used to compensate the emf at the instrument terminals is
a thermocouple immersed in an actual ice-water bath which in turn connects in
series with the measuring thermocouple. The ice and water combination holds
the temperature bath to a constant and accurate 0°C (32°F). NIST’s
thermocouple emf tables list the emf output of a thermocouple based on a
corresponding reference thermocouple junction held at 0°C.
Software Compensation
Ice baths and multiple reference junctions in large test fixtures are nuisances to
set up and maintain, and fortunately they all can be eliminated. The ice bath can
be ignored when the temperature of the lead wires and the reference junction
points (isothermal terminal block at the instrument) are the same. The emf
correction needed at the terminals can be referenced and compensated to the
NIST standards through computer software. When ice baths are eliminated, cold
junction compensation (CJC) is still necessary in order to obtain accurate
thermocouple measurements. The software has to read the isothermal block
temperature. One common technique uses a thermistor, mounted close to the
isothermal terminal block that connects to the external thermocouple leads. No
temperature gradients are allowed in the region containing the thermistor and
terminals.
The type of thermocouple employed is pre-programmed for its respective
channel, and the dynamic input data for the software includes the isothermal
block temperature and the measured environmental temperature. The software
uses the isothermal block temperature and type of thermocouple to look up the
value of the measured temperature corresponding to its voltage in a table, or it
calculates the temperature with a polynomial equation. The latter method allows
numerous channels of thermocouples of various types to be connected
simultaneously while the computer handles all the conversions automatically.
Hardware Compensation
Although a polynomial approach is faster than a look-up table, a hardware
method is even faster, because the correct voltage is immediately available to be
scanned. One method uses a battery in the circuit to null the offset voltage from
the reference junction so the net effect equals a 0°C junction. A more practical
approach is an "electronic ice point reference," which generates a compensating
voltage as a function of the temperature sensing circuit powered by a battery or
similar voltage source. The voltage then corresponds to an equivalent reference
junction at 0°C.
11
Type Mixing
Thermocouple test systems often measure tens to hundreds of points
simultaneously. In order to conveniently handle such large numbers of channels
without the complication of separate, unique compensation TCs for each,
thermocouple-scanning modules come with multiple input channels and can
accept any of the various types of thermocouples on any channel,
simultaneously. They contain special copper-based input terminal blocks with
numerous cold junction compensation sensors to ensure accurate readings,
regardless of the sensor type used. Moreover, the module contains a built-in
automatic zeroing channel as well as the cold-junction compensation channel.
Although measurement speed is relatively slower than most other types of
scanning modules, the readings are captured in ms, they contain less noise, and
they are more accurate and stable. For example, one TC channel can be
measured in 3 ms, 14 channels in 16 ms, and 56 channels in 61 ms. Typical
measurement accuracies are better than 0.7°C, with channel-to-channel
variation typically less than 0.5°C.
Linearization
After setting up the equivalent ice point reference emf in either hardware or
software, the measured thermocouple voltage must be converted to a
temperature reading. Thermocouple output voltage is proportional to the
temperature of the TC junction, but it is not perfectly linear over a very wide
range.
The standard method for obtaining high conversion accuracy for any
temperature uses the value of the measured thermocouple voltage plugged into a
characteristic equation for that particular type thermocouple. The equation is a
polynomial with an order of six to ten. The computer automatically handles the
calculation, but high-order polynomials take considerable time to process. In
order to accelerate the calculation, the thermocouple characteristic curve is
divided into several segments. Each segment is then approximated by a lower
order polynomial.
Analogue circuits are employed occasionally to linearize the curves, but when
the polynomial method is not used, the thermocouple output voltage frequently
connects to the input of an analogue to digital converter (ADC) where the
correct voltage to temperature match is obtained from a table stored in the
computer’s memory. For example, one data acquisition system TC card includes
a software driver that contains a temperature conversion library that changes
raw binary TC channels and CJC information into temperature readings. Some
software packages supply CJC information and automatically linearize the
thermocouples connected to the system.
12
Additional Concerns
Thermocouple Assembly
Thermocouples are twisted pairs of dissimilar wires and soldered or welded
together at the junction. When not assembled properly, they can produce a
variety of errors. For example, wires should not be twisted together to form a
junction; they should be soldered or welded. But solder is sufficient only at
relatively low temperatures, usually less than 200°C. And although soldering
also introduces a third metal, such as a lead/tin alloy, it will not likely introduce
errors if both sides of the junction are at the same temperature.
Welding the junction is preferred, but it must be done without changing the
wires’ characteristics. Commercially manufactured thermocouple junctions are
typically joined with capacitive discharge welders that ensure uniformity and
prevent contamination. Thermocouples can become un-calibrated and indicate
the wrong temperature when the physical makeup of the wire is altered. Then it
cannot meet the NIST standards. The change can come from a variety of
sources, including exposure to temperature extremes, cold working the metal,
stress placed on the cable when installed, vibration, or temperature gradients.
The output of the thermocouple also can change when its insulation resistance
decreases as the temperature increases. The change is exponential and can
produce a leakage resistance so low that it bypasses an open-thermocouple wire
detector circuit. In high-temperature applications using thin thermocouple wire,
the insulation can degrade to the point of forming a virtual junction. The data
acquisition system will then measure the output voltage of the virtual junction
instead of the true junction.
In addition, high temperatures can release impurities and chemicals within the
thermocouple wire insulation that diffuse into the thermocouple metal and
change its characteristics. Then, the temperature vs. voltage relationship
13
deviates from the published values. Choose protective insulation intended for
high-temperature operation to minimize these problems.
Isolation
Thermocouple isolation reduces noise and errors typically introduced by ground
loops. This is especially troublesome where numerous thermocouples with long
leads fasten directly between an engine block (or another large metal object)
and the thermocouple-measurement instrument. They may reference different
grounds, and without isolation, the ground loop can introduce relatively large
errors in the readings.
Auto-Zero Correction
Subtracting the output of a shorted channel from the measurement channel’s
readings can minimize the effects of time and temperature drift on the system’s
analogue circuitry. Although extremely small, this drift can become a
significant part of the low-level voltage supplied by a thermocouple. One
effective method of subtracting the offset due to drift is done in two steps. First,
the internal channel sequencer switches to a reference node and stores the offset
error voltage on a capacitor. Next, as the thermocouple channel switches onto
the analogue path, the stored error voltage is applied to the offset correction
input of a differential amplifier and automatically nulls out the offset. See
Figure 9.
Thermal Shunting
An ideal thermocouple does not affect the temperature of the device being
measured, but a real thermocouple comprises a mass that when added to the
14
device under test can alter the temperature measurement. Thermocouple mass
can be minimized with small diameter wires, but smaller wire is more
susceptible to contamination, annealing, strain, and shunt
impedance. One solution to help ease this problem is to use the small
thermocouple wire at the junction but add special, heavier thermocouple
extension wire to cover long distances. The material used in these extension
wires has net open-circuit voltage coefficients similar to specific thermocouple
types. Its series resistance is relatively low over long distances, and it can be
pulled through conduit more easily than premium grade
thermocouple wire. In addition to its practical size advantage, extension wire is
less expensive than standard thermocouple wire, especially platinum.
Despite these advantages, extension wire generally operates over a much
narrower temperature range and is more likely to receive mechanical stress. For
these reasons, temperature gradients across the extension wire should be kept to
a minimum to ensure accurate temperature measurements.
Temperature Sensors
Temperature can be measured via a diverse array of sensors. All of them infer
temperature by sensing some change in a physical characteristic. The two main
categories of these sensors are contact and noncontact style temperature sensors.
There are three methods of contact style temperature measurement, which can
generally be referred to as temperature probes: thermal resistors (RTD,
thermistors), thermocouples and IC sensors.
The most important specification to consider when searching for temperature
probes is sensed temperature range. This is the temperature range the device is
rated for sensing. The number of elements contained in the temperature probe
is another important consideration. Typically the more elements in the probe the
more accurate and expensive the probe is. Probes usually have one or two
elements for sensing, but some can be
configured to have greater than two
elements.
Basic sensing technologies for
temperature probes include thermocouple
(shown on right), RTD, thermistor, solid
state, and fiber optic. A thermocouple
consists of two wires, each made of a
different homogeneous metal or alloy. The
wires are joined at one end to form a
15
Piezoelectric sensor
16
One disadvantage of piezoelectric sensors is that they cannot be used for truly
static measurements. A static force will result in a fixed amount of charges on
the piezoelectric material. While working with conventional readout electronics,
imperfect insulating materials, and reduction in internal sensor resistance will
result in a constant loss of electrons, and yield a decreasing signal. Elevated
temperatures cause an additional drop in internal resistance and sensitivity. The
main effect on the piezoelectric effect is that with increasing pressure loads and
temperature, the sensitivity is reduced due to twin-formation. While quartz
sensors need to be cooled during measurements at temperatures above 300 °C,
special types of crystals like GaPO4 gallium phosphate do not show any twin
formation up to the melting point of the material itself.
However, it is not true that piezoelectric sensors can only be used for very fast
processes or at ambient conditions. In fact, there are numerous applications that
show quasi-static measurements, while there are other applications with
temperatures higher than 500 °C.
Piezoelectric sensors are also seen in nature. The collagen in bone is
piezoelectric, and is thought by some to act as a biological force sensor.
18
Principle of operation
Depending on how a piezoelectric material is cut, three main modes of
operation can be distinguished: transverse, longitudinal, and shear.[4]
Transverse effect
A force is applied along a neutral axis (y) and the charges are generated along
the (x) direction, perpendicular to the line of force. The amount of charge
depends on the geometrical dimensions of the respective piezoelectric element.
When dimensions apply,
,
where is the dimension in line with the neutral axis, is in line with the
charge generating axis and is the corresponding piezoelectric coefficient.
Longitudinal effect
The amount of charge produced is strictly proportional to the applied force and
is independent of size and shape of the piezoelectric element. Using several
elements that are mechanically in series and electrically in parallel is the only
way to increase the charge output. The resulting charge is
,
where is the piezoelectric coefficient for a charge in x-direction released by
forces applied along x-direction (in pC/N). is the applied Force in x-
direction [N] and corresponds to the number of stacked elements .
Force applied and the element dimension.
Electrical properties
the source is directly proportional to the applied force, pressure, or strain. The
output signal is then related to this mechanical force as if it had passed through
the equivalent circuit.
In the flat region, the sensor can be modeled as a voltage source in series with
the sensor's capacitance or a charge source in parallel with the capacitance
For use as a sensor, the flat region of the frequency response plot is typically
used, between the high-pass cutoff and the resonant peak. The load and leakage
resistance need to be large enough that low frequencies of interest are not lost.
A simplified equivalent circuit model can be used in this region, in
20
Two main groups of materials are used for piezoelectric sensors: piezoelectric
ceramics and single crystal materials. The ceramic materials (such
as PZT ceramic) have a piezoelectric constant / sensitivity that is roughly
two orders of magnitude higher than those of the natural single crystal materials
and can be produced by inexpensive sintering processes. The piezo effect in
piezo ceramics is "trained", so unfortunately their high sensitivity degrades over
time. The degradation is highly correlated with temperature. The less sensitive
'natural' single crystal materials (gallium phosphate, quartz, tourmaline) have a
much higher – when carefully handled, almost infinite – long term stability.
There are also new single crystal materials commercially available such as Lead
Magnesium Niobate-Lead Titanate (PMN-PT). These materials offer greatly
improved sensitivity (compared with PZT) but suffer from a lower maximum
operating temperature and are currently much more expensive to manufacture.