Professional Documents
Culture Documents
Revised Flammability Test Methods For Aerosol Products
Revised Flammability Test Methods For Aerosol Products
Discussion
Because of the difficulty in relating the results of these
tests to the degree of hazard involved in the transportation,
warehousing or end uses of these foam type products, the
two procedures are not to be used in order to determine if a
product is to be called "Flammable" as distinguished from
"Non-flammable." They are presented simply as the means
NOTt: by which the relative flammability of foam products may
W£i£> ALL WM* be assessed.
$e£AX AU fHjer £OM££$
The procedures have been most thoroughly evaluated on
Figure 2. Apparatus for Trough Method simple shaving cream aerosols, and are certified only for
application to this particular product type. In addition, the
methods have been applied to a variety of stable foam
items, such as upholstery cleaner, charcoal lighter, metal
time; and the third, after a five-minute holding time. The polish and hair shampoo, with good results. The procedures
trough is left undisturbed and in a draft-free area for the have not been evaluated with thermal foams, foaming gels,
designated holding times between tests.
quick-breaking foams, anhydrous foams and reticulated
7. At the end of the holding time, a small lighted gas gels; so that any conclusions made concerning the relative
flame is brought to the surface of the foam in the area to flammability of these products must be made according to
one-half inch from the end of the trough. the investigator's own recognizance.
8. A positive test is indicated when the flame is
sustained or propagated by the propellent gas or foam
material after withdrawal of the flame.
Outline of Method
2. The sample is placed in the cup of a Tag Open Tester,
and heated at a slow but constant rate. A small test flame is
passed at a uniform rate across the cup at specified
intervals. The flashpoint is taken as the lowest temperature
at which application of the test flame causes the vapor at Figure 1
the surface of the liquid to flash, that is, ignite but not
continue to burn.
(d) Glass test cup. Glass test cup (Fig. 2), of molded
Apparatus clear glass, annealed, heat-resistant, and free from surface
defects.
3. The Tag open-cup tester is illustrated in Fig. 1. It
consists of the following parts, which must conform to the (e) Leveling device. Leveling device or guide, for proper
dimensions shown, and have the additional characteristics adjustment of the liquid level in the cup (Fig. 3). This shall
as noted: be made of No. 18-gage polished aluminum, with a
(a) Copper bath, preferably equipped with a constant projection for adjusting the liquid level when the sample is
level overflow so placed as to maintain the bath liquid level
1/8 inch below the rim of the glass cup.
(b) Thermometer holder. Support firmly with ringstand
and clamp.
(c) Thermometer. For flashpoints above 40° F., use the
ASTM Tag Closed Tester Thermometer, range of +20 to
+230° F., in 1° F. divisions, and conforming to
thermometer 9F. of ASTM Standard E 1. For flashpoints
below 20° F., use ASTM Thermometer 33F. The original
Tag Open-Cut (Paper Scale) Thermometer will be a
permissible alternate unit January 1, 1962. It is calibrated
to -20° F.
INTRODUCTION
A simple device is presented for the routine
measurement of the "A-D" valve dimension. This distance Figure 1. "A-D" Valve Dimension
is defined as the standard for stipulating dip tube depth. It
is the centerline dimension from the top plane of the affected by relative dip iuoe curvatures and by variations in
mounting cup curl to the far end of the diptube; (Figure 1). the ratio of can diameter to total height — factors which
For quality control purposes it is desirable to have a cannot be anticipated by any simple gauging device. The
method available for the rapid measurement of dip tube present measurement compensates for variations in cup
depth. This is especially important where large numbers of depth, cup contour, body length and upper reference
measurements must be taken. The present device eliminates position of the dip tube. Operations such as crimping the
cumbersome hand manipulations, parallax errors and valve and pressurizing the aerosol container will shorten the
similar difficulties. It provides a means for measuring dip "A-D" dimension slightly, but these effects are fairly
tube depths with excellent accuracy at speeds of from reproduceable and compensations could be made if
about 10 to 14 per minute. necessary.
If the steel ruler is attached to the gauge bar with a high
METHOD degree of precision, the device may be expected to function
The measuring guage is produced from an aluminum with an accuracy of about ±0.012". Where the ruler is
cylinder machined to a length of 8.717" and an outside attached with a metal adhesive, the inset end of the gauge
diameter of 1.216". A complete description is given in the bar should end at 0.033" by extension of the cut offend of
accompanying drawing; (Figure II). This is for a left-handed the ruler. If it is too long the end should be machined
guage. A right-handed one may be produced by machining a down. Where the ruler is attached by means of several
mirror-image of the gauge bar itself and locating the ruler screws, it is desirable to use roundhead machine screws, set
on the opposite quadrant flat. through holes in the ruler large enough to allow for precise
positioning of the scale before final tightening.
For individual accurate "A-D" measurements the valve is
inserted so that the dip tube slides down the centerbore of Only one assumption is made in the construction of the
the gauge. With the mounting cup held snugly in place gauge: that the vertical distance through the curl of the
against the gauge end, the exact dip tube depth is read off mounting cup, including the thickness of the cup gasket,
the steel rule. totals 0.0333". Studies made with an Ames No. S-4469 dial
For repetitive or quality control measurements the valve micrometer or comparable equipment will show that this
is inserted as before, using a gauge where the steel rule has crown dimension usually holds true to within about
been temporarily marked in pencil to show the tolerance or ±0.004" on a 2-sigma or 95% confidence level. The
range of acceptance. Each valve may then be rapidly variation is due mainly to differences in cup gasket
checked to show compliance with the "A-D" specification. thickness. It is not possible for the average operator to read
The pencil marks may then be removed by hand rubbing. the. ruler this accurately; normal reproduceability being
Do not use ball point pens or other marking devices where about ±0.0065" with a perfectly sheared dip tube. The
the lines will have to be removed with solvent or abrasive. variation may be neglected for all practical purposes.
This can also remove the black lettering and line fillers of
the ruler itself, rendering it much harder to read. With only
a slight loss in accuracy the tolerance marks may be made
upon the opposite quadrant flat. For some operations
permanent marks may be scribed on this flat with a suitable
tool.
DISCUSSION
The true measure of dip tube depth is a function of the
depth of protrusion into the aerosol can. This will be Figure 2. Measuring Gauge
The average dip tube is not cut off at precisely a right
angle, nor is the cut end smooth and free from
imperfections. Often the center-line distance from the short
sector to the long sector is as much as 0.006". The slight
residual curvature of the dip tube, as it reposes within the
gauge bar, may require the averaging of high and low PROCEDURE FOR EVALUATION
readings on the oblique surface. Polyethylenes stretch easily
and should never be straightened with any method which
involves tension. Polypropylenes and nylon dip tubes are OF COMBING CHARACTERISTICS OF
less easily stretched.
The distortional effects of valve crimping and can HAIR SPRAY FORMULATIONS
pressurization have been mentioned. Hot tanking also (Editor's Method)
produces an irreversible distortion of about 0.0025". To
these vectors are added the swelling effect of the aerosol
formula and the consideration that the cut end of the dip
tube is never horizontal, but slanted towards a point on the SCOPE
bottom chime of the container. From all this it is apparent
that dip tube measurements with accuracies better than This test evaluates, subjectively, the adhesion, flaking
±0.016" serve no real purpose in quality control operations. and general effect on combing characteristics of aerosol
The usual specification for dip tube depth is supplied by formulations on hair.
valve manufacturers at L±0.063". Two undesirable effects seen in some formulations on
A gauge bar has several applicational limitations. It hair:
cannot be used to provide an accurate "A-D" measurement 1. Flaking
for ferrule type valves*, co-dispensing valves, extra-large This condition manifests itself in the appearance of
dosage metering valves, or those with unusual mounting cup small polymer particles distributed throughout the
designs, such as the turret-top styles. Double dip-tube valves
and those with dip tubes having an outside diameter greater hair.
than about 0.242" cannot be handled. The "A-D" 2. Snarling
dimension itself is often found to be inapplicable to these This appears to be due to general tackiness or,
unique valves. conversely, extreme brittleness. In the case of brittle
CREDITS polymers, the resin ruptures during combing and
produces small barbs of polymer protruding from the
The Modern Machine Shop, Inc., at 123 North Hazel hair shafts. Both the tacky and brittle polymers resist
Street, Danville, Illinois 61832, has had experience in the combing and increase the possibility of entanglement
manufacture of a number of these gauge bars and may, of individual hair shafts.
therefore, be considered as a reliable source.
We are indebted to the British Aerosol Manufacturers PROCEDURE
Association (BAMA) and the Federation of European 1. Ten inch hair swatches are prepared according to the
Aerosol Associations (FEA) for providing the original idea description given in the "Curl Retention Procedure".
for the construction of this instrument. 2. All testing; i.e., spraying, drying and combing is
performed at 72°F./50RH and run on three swatches
ALTERNATE DEVICE FOR MEASURING THE for each formulation.
DIP TUBE DEPTH OF AEROSOL VALVES 3. The test formulation is applied to a rotating (20
This device consists basically of 3 machined parts made RPM) hair swatch by a 5-second aerosol discharge at a
of plexiglas. In addition, there is a scale glued to the "scale distance of 7 inches.
block"; a foam bottom pad may be added for non-slip 4. The swatch is allowed to dry, undisturbed, for 30
characteristics. minutes.
The valve to be measured is inserted into the device, 5. Comb the length of the hair piece 5 times (at a rate of
placing the actuator/stem end to the higher end of the 3"/second) with the fine teeth of a comb (13-7/8"
device. The slide bar is then pushed firmly to the left to teeth/inch): Record evaluations.
"seat" the mounting cup against base end. The reading may 6. Repeat steps 3 to 5 until the formulations under test
then be taken off the scale at the end of the dip tube, giving show severe flaking or are difficult to comb.
the "A-D dimension" from top of mounting cup to dip
tube end. 7. Perform test on 3 swatches for each formulation
variable.
Note that this measuring device can be used to measure
valves of various types, including ferrule-type valves. It EVALUATION
would not be applicable to co-dispensing or large-tank
metered valves; undue distortion would occur with double A. Combing Rating
dip tube valves or large O.D. dip tubes. None to slight resistance to combing —1
Moderate resistance distributed at different points
*Reference is made to an attached alternative device for measuring along length of hair -2
dip tube depths. It may be used for "A-D" measurements of ferrule
type valves as well as for aerosol can valves. Moderate to severe resistance throughout length of
of hair -3
Definite snarling of hair, resisting combing at Motor and rheostat
different points along length of hair ~4 Shampoo (J. H. Breck)
Definite snarling of hair throughout length of Interval timer
hair, completely resisting combing —5
B. Microscopic Examination Rating PROCEDURE
Good adhesion —1 A. Preparation of Curl Swatches
Small amount of resin particles -2 1. Separate hair into small swatches of approximately 2
Appearance of polymer fragments protruding from grams in weight.
resin film - increased flaking (moderate) -3 2. Secure hair sample 1 inch from root end by winding
A more general loss of adhesion from hair shafts —4 with cotton thread.
Severe Flaking -5 3. Fold over at root end and secure the hair into a small
Note - With microscope ratings of 3-5 the flaking is usually loop with thread (Figure 1). Leave approximately 6
apparent upon visual examination of the hair swatch inches of loose thread to secure identification tag.
without magnification. 4. Wash hair swatch in a 10% solution of shampoo. Then
rinse thoroughly with distilled water.
5. Comb to untangle the hair shafts.
C. Flaking Rating - Visual Inspection 6. Cut the hair swatch to measure 10 inches in length
None - very slight visible -1 from the looped root end. Dry 1/2 hour at 14O0F.
Slight -2 7. In order to prevent loss of individual hair shafts
Moderate -3 during subsequent combing, glue the root end at the
thread windings with Duco cement
Moderate - severe —4
Severe -5 B. Preparation and exposure of Test Samples
1. Hang a hair swatch from a motor chuck. Secure the
Note - Also report differences in gloss among the end of the tress with a Swingline binder clip to
formulations under test. prevent motion of the hair during spraying. The
motor should be adjusted with a rheostat so that the
shaft rotates at 20 RPM.
2. With the hair swatch rotating, hold the aerosol
DETERMINATION OF THE container at a distance of 7 inches and spray for 10
seconds. Direct the spray uniformly throughout the
HIGH HUMIDITY CURL RETENTION length of the tress.
3. Remove the binder clip and bring fingers gently down
OF AEROSOL HAIR SPRAYS the full length of the hair, squeezing off any excess
resin solution.
(Editor's Method)
4. Comb hair twice, using the wide teeth (9-7/8"
teeth/inch) of a 7 1/2" nylon comb. Roll curl on a
1/2" diameter Teflon mandrel. Remove curl from
mandrel and secure with a "Klippee".
The curl retention properties of polymeric hair spray 5. Dry in a 14O0F oven for 30 minutes and then allow
resins are measured at 90% RH over a period of three hours. curl to condition at 720F at 50% RH overnight.
The change in percentage retention versus time may be 6. Remove "Klippee" and gently bring the curl down by
plotted to illustrate differences among formulations. The unwinding into a spiral with a glass rod.
procedure lends itself to statistical analysis of formulation 7. Place the unwound curls at random on Plexiglass
variables. retention boards (Figure 2). Take initial reading.
8. Set retention boards into humidity chamber and
record length of curl every 30 minutes over a period
of three hours (Figure 3).
EQUIPMENT
12" Remi, Blue String, European, brown hair C. Calculations
Barber shears Calculate percentage curl retention by:
7 1/2" nylon combs
Alligator and paper clips L-Lt
White cotton thread, size # 8 Curl Retention (%) = x 100
Identification tags L - LO
"Klippees" hair clips
Where L = Length of hair fully extended
3" x 1/2" Teflon mandrel Where L0 = Length of hair before exposure
140° oven - Forced air Where Lt = Length of hair after exposure
Swingline binder clips
Plexiglass retention board Note - Determine retention of 10 replicates of each
Humidity Cabinet (precision to ± 2% RH at 7O0F.) variable.
DETERMINATION OF THE PRESSURE (OR VACUUM) IN AN AEROSOL CONTAINER
(Editor's Method)
SCOPE PROCEDURE
The method is satisfactory for all aerosols having a 1. Condition test container by upright immersion in water
pressure of 180 psi-g. or less, provided they have a valve to bath for 30 minutes.
which the gauge assembly can be connected. 2. For pre-pressurized gauges, pressurize to 2 or 3 psi.
greater than anticipated test container pressure.
APPARATUS 3. Remove container from water bath, actuate briefly,
remove button or foam spout, shake for five seconds,
Apparatus requirements will vary according to the and apply the gauge assembly firmly to the valve to get a
pressure (or vacuum) in the container, the design of the good seal.
aerosol valve, and the accuracy desired. Gauge assemblies
should be selected in accordance with the following table:
*General Purpose Gauges may be used with or without the pre-pressurization assembly.
**Inspector's Gauge is used without the pre-pressurization assembly.
For the accurate readings the U. S. Inspector's Gauge is 4. Actuate the valve by moving the gauge assembly against
threaded to a coupling, then a toggle valve and finally to an it. Open toggle valve and agitate assembly to bring gauge
adapter made to match the valve of the test unit. For less pointer to maximum reading. Tap gauge face lightly.
accurate readings the Inspector's Gauge may be used Re-actuate several times and take the average of the
without the pre-pressurization assembly. readings.
For the less accurate readings a General Purpose Gauge For best readings at elevated temperatures, leave test
may also be used, either with or without the container in water bath while taking readings.
pre-pressurization assembly. If pre-pressurization is not 5. Clean out any product forced into valve system by
desired, connect the gauge directly to the valve adapter. In applying gauge to a can containing a P-12/ethanol
this case the gauge should be either two-inch or three-inch mixture.
diameter, to minimize the size of the Bourdon tube.
Individual gauges or complete assemblies may be REPORTING
purchased from several sources, including Builder's Sheet Accurate gauge readings should be reported to the
Metal Works, Inc., 108 Wooster Street, New York, N. Y. nearest + 0.5 psi. Less accurate readings may be reported to
10012; or General Kinetics Division, 451 Stephens Street, the nearest + 1.0 psi.
SW, Atlanta, Ga. 30302. The Inspector's Gauge with
pre-pressurizing assembly cost about $69.29 in 1972. The
General Purpose Gauge with adapter cost about $9.70
during 1972.
Test methods in this chapter supplement discussions
Pressure gauges should be checked against standards on a elsewhere in this text. The reader is referred also to the
daily basis, using pure propellent as the pressure source.
CSMA Aerosol Guide, Seventh Edition, 1981.
A water bath, normally maintained at 70° ± 0.5 0 F., is
required for accurate work. For less accurate studies the
water bath may have a tolerance of ± 1.00F.
FORMI
FURNITURE POLISH EVALUATION
INDIVIDUAL RATINGS FOR 4.3
TEST PANEL APPLICA TION NO. 1 TEST PANEL APPLICA TION NO. 2
Properties Left Center Right Left I Center Right
6. Report
6.1 All properties should be rated 0-5. A value of 5
equals excellent and O equals complete failure. Values in These standards have undergone several revisions since
between are various degrees between these extremes. they were first adopted by the Department of Defense
in September 1950. The present revision, MIL-STD-
6.2 Form I should be used to record the raw data. 105D, 29 April 1963, has been approved by the Depart-
Form II should be used to summarize and compare the raw ment of Defense and is mandatory for use by the Depart-
ments of the Army, the Navy, the Air Force and the when 5 consecutive lots or batches have been considered
Defense Supply Agency. The large number of defense acceptable on original inspection.
contracts awarded since the original adoption may,
in large part, account for the wide acceptance of the N O R M A L TO REDUCED. When normal inspection
standards by industry as well as the Government. is in effect, reduced inspection shall be instituted provid-
ing that all of the following conditions are satisfied:
a. The preceding 10 lots or batches (or more, as indicated
A. Preliminary Considerations by the note to Table V I I I ) have been on normal
inspection and none has been rejected on original
As with any sampling plan, one of the first considerations inspection; and
is to select the characteristic to be checked, and to provide
an adequate test procedure. 6. The total number of defectives (or defects) in the
samples from the preceding 10 lots or batches (or such
other number as was used for condition -a*' above)
B. Selection of an AQL Value is equal to or less than the applicable number given
in Table VIII. If double or multiple sampling is
After the characteristic has been defined, MIL-STD in use, all samples inspected should be included,
105D requires the selection of an AQL. This necessitates not "'first" samples only; and
a classification of particular characteristic as critical,
major, or minor. The usage of the particular character- c. Production is at a steady rate; and
istic must be known and considered with respect to the d. Reduced inspection is considered desirable by the
possibility and consequences of failure in assembly and responsible authority.
use. If the characteristic is of a critical nature, a low
AQL must be selected. If it is of a non-critical nature, REDUCED TO NORMAL. When reduced inspection
a higher AQL would be indicated. Where a classification is in effect, normal inspection shall be instituted if any
of characteristics has been made by specifying them of the following occur on original inspection:
as critical, major, minor, the selection of an AQL value a. A lot or batch is rejected: or
can be greatly simplified.
b. A lot or batch is considered acceptable under certain
conditions of reduced inspection when the sampling
C. Normal, Tightened, and Reduced Inspection procedure may terminate without either acceptance
or rejection criteria having been met. In these
MIL-STD-105D provides a specific set of rules as to circumstances, the lot or batch will be considered
the use of nermal, tightened, and reduced inspection, acceptable, but normal inspection will be reinstated
including continuing one and switching from one to starting with the next lot or batch.
another. c. Production becomes irregular or delayed; or
The following instructions are taken from MIL-STD- d. Other conditions warrant that normal inspection
105D. shall be instituted.
INITIATION OF INSPECTION. Normal inspection DISCONTINUATION OF INSPECTION. In the event
will be used at the start of inspection unless otherwise that 10 consecutive lots or batches remain on tightened
directed by the responsible authority. inspection (or'such other number as may be designated
CONTINUATION OF INSPECTION. Normal, tight- by the responsible authority), inspection under the pro-
ened or reduced inspection shall continue unchanged for visions of this document should be discontinued pending
each class of defects or defectives on successive lots or action to improve the quality of submitted material.
batches except where the switching procedures given It should be noted here that many of the rules estab-
below require a change. The switching procedures shall lished for the use of MIL-STD-105D by the Government
be applied to each class of defects or defectives inde- have been modified for use in industry. The foregoing
pendently. rules have often been modified or changed, but the pro-
spective user of these plans should benefit by restating
the above conditions with his own modifications to insure
SWITCHING PROCEDURES
a definite plan for changing from one type of plan to
another.
NORMAL TO TIGHTENED. When normal inspec-
t.on is in effect, tightened inspection shall be instituted
when 2 out of 5 consecutive lots or batches have been
D. Selection of an Inspection Level
rejected on original inspection (i.e., ignoring resubmitted
lots or batches for this procedure).
The MIL-STD-105D tables provide for varying the
TIGHTENED TO NORMAL. When tightened inspec- discriminatory power of the sampling plan by selecting
tion is in effect, normal inspection shall be instituted one of four special inspection levels, S-I, S-2, S-3, S-4,
or one of three general inspection levels, I, II, III. In are accumulated. If the accumulated number of defec-
each case the higher the number the more discriminatory tives is equal to or less than the second acceptance number,
the plan. The more discriminatory the plan the larger the lot should be accepted. If the accumulated number
the sample size. The OC Curves and certain tabulated of defectives is equal to or greater than the second rejection
values for them are given. number, the lot should be rejected. In multiple sampling
Inspection level II is usually selected. this process is repeated until a decision is reached.
It is observed that a decision is reached with one sample
in single sampling, with two or less samples in double
E. Single, Double, and Multiple Sampling
sampling, and with seven or less samples in multiple
sampling. However, the size of each sample decreases
For most AQL values and inspection levels, the user
as one goes from single to multiple sampling, so that,
has the opportunity to select single, double, or multiple
in general, a smaller average amount of inspection is
sampling. The selection of the type of sampling plan necessary for multiple than for single sampling.
will depend upon:
The disposition of any defectives found in the samples
will depend upon the nature and magnitude of the defect
1. Psychological likes and dislikes, and upon the agreement made between the producer
2. Difficulty of administering and using the various and consumer.
kinds of sampling,
3. Average amount of inspection required by using
the various kinds of sampling, and
4. Difficulty in selecting samples. G. An Illustrative Example
However, for any given AQL and inspection level, the OC Suppose we have agreed on an AQL of 1.5°0 for a
curves will be nearly the same regardless of the kind of certain characteristic, a shipment of 1500 pieces is received,
sampling. and we are to determine appropriate sampling plans
based on general inspection level II, normal inspection.
From Table I, we obtain the sample size code letter K.
F. Selecting a Sampling Plan
From Tables II, III, and IV, we obtain:
After the preceding decisions have been made, a sampl- Single Double
ing plan can be selected for any given lot when the lot Sampling Sampling
size is known. A sample size code letter, dependent
upon the lot size and the inspection level, is obtained Sample Sample Sample
from Table I. (See page 342.) Size Ac Re Number Size Ac Re
202 Diameter
202 x 200 202 x 406 202 x 700
202 x 214 202 x 509 202 x 708 METHOD FOR DETERMINING DELIVERY RATE
202x314 202x608
207.5 Diameter OF AEROSOL PRODUCTS
207.5x413 207.5x605 207.5x708
207.5x509 207.5x701 207.5x713
211 Diameter
211 x407.5 211 x 5 1 0 211 x 6 1 2
211 x 4 1 3 211 x604 211 x713 Introduction
The delivery rate of an aerosol dispenser is determinable
300 Diameter by measuring the quantity of material expelled through the
300 x 709
valve in a given time. The exact duration of discharge and
the temperature of the dispenser must be carefully
Dimensions included in Industry Specifications to Date: controlled for good reproduceability.
A. Inside Diameter of 1" Cup Opening. Delivery rate tests assist in evaluating one aspect of valve
B. Outside Diameter of 1" Cup Opening. performance and are considered a prerequisite to both
C. Height of Curl Opening above Double Seam. biological and storage testing. Biological and clinical tests
D. Height over Double Seam. are often made with one or two dispensers selected from a
group similiar in all respects except for delivery rate. When
E. Overall Height of Container. it is desirable to select dispensers with equal delivery rates,
F. Height Between Double Seams. two or three tests should be performed on each dispenser.
G. Thickness of Curl Around the One Inch Opening. In the case of storage tests a single delivery rate test is
normally performed at each examination period to conserve
the contents and extend the life of the dispenser.
Equipment
Water-bath, maintained at 70° ±0.5°F. (80° ±0.5° F.
for insecticides, to match Peet-Grady test requirements),
with a screen or perforated metal shelf one inch above the
bottom of the bath.
Stirrer — air or electric.
Balance, one-tenth gram scale.
Stop watch or electric timer.
Procedure
Remove the protective cover, paper label and other
detachable materials from the dispenser, except the button.
If a foam spout is used, remove, cut away all non-essential
plastic, and replace.
Activate the valve for a few seconds, remove any valve
cup impingements, and weigh the dispenser to the nearest
0.05 gram. Place the dispenser on the shelf in the water
bath, which is at the test temperature of 70° ± 0.1° F.
Keep the dispensers in an upright position, spaced one
inch apart and covered with one inch of water. Circulate
the water with the mechanical stirrer. Hold for 15 minutes
in the case of essentially anhydrous products and for 25 to
30 minutes for viscous or water-based formulations.
Remove the dispenser. Actuate the valve for a given time
(preferably 10 seconds), preferably into an exhaust hood.
Dry the dispenser with a cloth or towel and use a blast of
compressed air to remove moisture from the mounting cup
and can seams, in the case of metal containers. Reweigh the
dispenser and compute the difference in weight.
The results of each test are reported as delivery rate in occur within the dispenser as it is used up, and these will
grams per second and calculated as follows: have a small depressing effect upon delivery rates.
Weight loss in grams The variation in valve orifice diameters may be as high as
Delivery Rate = ±10%. Because of this, delivery rates may vary as much as
10 Seconds ±15% from can to can.
Repeat the above procedure if duplicate tests are made Generally, the product should be shaken before any
on dispensers. actuation. Reference should be made to label instructions.
In the case of thixotropic, rheopectic and certain other
Discussion emulsion types the degree of agitation will have a
considerable effect upon delivery rates.
The reproduceability of results by this method is
considered to be about ±1%. Precision is about ±2%. The delivery rate of freshly prepared aerosols will not
usually be the same as that after aging. This is due to gasket
Any air, trapped in the product and outage space, may swelling, changes in elastomer hardness and other factors.
be expected to have an elevating effect upon delivery rates.
The effect decreases as the can is emptied. The rate of
decrease is much more rapid for dispensers with vapor-tap NOTH: About 0.03% of sodium nitrite may be added to
valves. Even without air, minor distillation effects will the water bath to prevent the minor rusting of
some tinplate containers.