Professional Documents
Culture Documents
bisphenoI-A polycarbonate
C. Bailly,* D. Daoust, R. Legras and J. P. Mercier
Laboratoire des Hauts Polym~res, Universit~ Catholique de Louvain,
B- 1348 Louvain-La- Neuve, Belgium
In order to assess the reliability of absolute molecular weights of bisphenol-A polycarbonate samples
measured by size exclusionchromatography, the results obtained from two experimental systemsand two
calibration methods are compared with each other as well as with data provided by light scattering
determinations. The interest of coupling a light scattering detector to the chromatographic system is
discussed.
(Keywords: bisphenoI-A polycarbonate; size exclusion chromatography; light scattering; viscometry; universal calibration)
using a Fica 50 instrument. The solutions were annealed Table ! Fractionation of sample PC1
either for 15 h at 36°C or for 48 to 64 h at 50°C before Cumulative
centrifugation at 25 000 g for 90 rain. Weight weightfract. [v/]
Fraction (g) (%) (cm 3 g 1) Mya
Size exclusion chromatography. Two experimental
systems were used for measuring the MW distributions by PCI F6 0.3 1.90 109 82200
SEC. The first system comprised a set of two Shodex PCI F5 1.9 16.00 152 130400
ABOM columns (10#m particle size) providing a good PC1 F4 2.1 41.65 193 183 500
PC1 F3 2.0 70.60 218 218 500
separation between 3 x 10 3 and 3 x 10 6 (PS equivalent PC1 F2 1.3 89.05 255 273000
MW). Distilled and filtered T H F was used as the mobile PC1 F1 0.2 98.70 295 340000
phase at a 1 cm 3 min-1 flow rate. The detector was a
Shodex SE-11 differential refractometer. Preliminary PC 1 (unfracL) 7.9 1O0 230 235 000
calibration of the columns was obtained with 12 narrow PCI (after fract.) 7.8 98.7 200 195000
PS standards (MW ranging between 1500 and 3.6 x 106). "Viscometric average M W (My) calculated from the viscosity relation
The viscosity relation for linear PS in T H F at 25°C, LS proposed by Schnell 2 for PC in T H F
is 6 :
A
2.5
/ / "/-/"
o / 111/../I
~ 2.0
1.5
f.
t I ! I
4.0, 4.5 50 5.5
Log M
Figure 4 M a r k - H o u w i n k relations for PC in T H F at 25'C: A,
proposed by Schnell; C, proposed by Moore; C, this work. O ,
fractionated samples; O , unfractioned samples
?
p I. ,~ t i
4.0 4.5 5.0 5.5 6.0 SEC in THF. A similar study was carried out, using
Log M T H F as the mobile phase. The Shodex column set chosen
Figure 3 Molecular weight distribution of sample PCI F5 as obtained was found to be less selective in the low molecular weight
from SEC-LS coupling (Mw= 124000; Mn = l l4000) region than the Ultrastyragel set, as expected from the
specifications of the columns. The universal and PC
calibration curve are shown in Figure 7. The MW values
experimental error, confirming the validity of the recalculated using them are summarized in Table 6. It is
assumption made to obtain the PC calibration curve. The observed that the two calibration curves diverge at low
two calibration curves diverge for MW lower than 1500. molecular weights (below 1500) but also, slightly, in the
Although this discrepancy has no influence on the M,~ high molecular weight range (above 8 x 104). This
values of our sa_mples, it can lead to significant differences behaviour is confirmed by the Mw and M, values obtained
by both calibrations (Table 6).
in ~ t when Mw is below 10000. This highlights the
interest of including PC oligomers i_nthe calibration curve Comparison between SEC results in T H F and methylene
when precise determinations of M, and the molecular chloride. The comparison between Tables 5 and 6 and
weight distributions are required. Figure 6 illustrates this between Figures 6 and 7 shows that, independently of the
point by showing the number and weight average MW nature of the mobile phase, the universal and PC
distributions of sample PC3. The high resolution of the calibrationsgive very similar average MW for samples
column set used shows up oligomers with MW below having a Mw below l0 s. Moreover, the h4tw values
2000. Moreover, the precision of the PC calibration curve obtained by LS and SEC are close together in the same
allows a positive identification of the species 11 MW region. For samples of higher MW, the PC
Weight distribution
Number distribution . . . . . .
/
iI I', ,it
I
/ L_
;i :i,1~LI j
*l
11
:', ;j
% A
\
23 3', 39 43 ,i7 i ........ z . . . .