You are on page 1of 6

Environmental Protection Agency Pt. 50, App.

J
computing whether the minimum data com- tion if the average annual fourth maximum
pleteness requirements have been met, mete- 8-hour concentration is greater than the
orological or ambient data may be sufficient level of the standard.
to demonstrate that meteorological condi- (d) Comparisons with the primary and sec-
tions on missing days were not conducive to ondary ozone standards are demonstrated by
concentrations above the level of the stand- examples 1 and 2 in paragraphs (d)(1) and (d)
ard. Missing days assumed less than the level (2) respectively as follows:
of the standard are counted for the purpose (1) As shown in example 1, the primary and
of meeting the data completeness require- secondary standards are met at this moni-
ment, subject to the approval of the appro- toring site because the 3-year average of the
priate Regional Administrator. annual fourth-highest daily maximum 8-hour
(c) Years with concentrations greater than average ozone concentrations (i.e., 0.084 ppm)
the level of the standard shall not be ignored is less than or equal to 0.08 ppm. The data
on the ground that they have less than com- completeness requirement is also met be-
plete data. Thus, in computing the 3-year av- cause the average percent of days with valid
erage fourth maximum concentration, cal- ambient monitoring data is greater than
endar years with less than 75% data com- 90%, and no single year has less than 75%
pleteness shall be included in the computa- data completeness.
EXAMPLE 1. AMBIENT MONITORING SITE ATTAINING THE PRIMARY AND SECONDARY OZONE STANDARDS
1st Highest 2nd Highest 3rd Highest 4th Highest 5th Highest
Percent Daily Max Daily Max Daily Max Daily Max Daily Max
Year Valid Days 8-hour 8-hour 8-hour 8-hour 8-hour
Conc. (ppm) Conc. (ppm) Conc. (ppm) Conc. (ppm) Conc. (ppm)

1993 ...................................................... 100% 0.092 0.091 0.090 0.088 0.085

1994 ...................................................... 96% 0.090 0.089 0.086 0.084 0.080

1995 ...................................................... 98% 0.087 0.085 0.083 0.080 0.075

Average .......................................... 98%

(2) As shown in example 2, the primary and concentration data for 1994 is used in these
secondary standards are not met at this computations, even though the data capture
monitoring site because the 3-year average is less than 75%, because the average fourth-
of the fourth-highest daily maximum 8-hour highest daily maximum 8-hour average con-
average ozone concentrations (i.e., 0.093 ppm) centration is greater than 0.08 ppm.
is greater than 0.08 ppm. Note that the ozone
EXAMPLE 2. AMBIENT MONITORING SITE FAILING TO MEET THE PRIMARY AND SECONDARY OZONE
STANDARDS
1st Highest 2nd Highest 3rd Highest 4th Highest 5th Highest
Percent Daily Max Daily Max Daily Max Daily Max Daily Max
Year Valid Days 8-hour 8-hour 8-hour 8-hour 8-hour
Conc. (ppm) Conc. (ppm) Conc. (ppm) Conc. (ppm) Conc. (ppm)

1993 ...................................................... 96% 0.105 0.103 0.103 0.102 0.102

1994 ...................................................... 74% 0.090 0.085 0.082 0.080 0.078

1995 ...................................................... 98% 0.103 0.101 0.101 0.097 0.095

Average .......................................... 89%

3. Design Values for Primary and Secondary average ozone concentration is also the air
Ambient Air Quality Standards for Ozone. The quality design value for the site.
air quality design value at a monitoring site
is defined as that concentration that when [62 FR 38895, July 18, 1997]
reduced to the level of the standard ensures
that the site meets the standard. For a con- APPENDIX J TO PART 50—REFERENCE
centration-based standard, the air quality METHOD FOR THE DETERMINATION OF
design value is simply the standard-related PARTICULATE MATTER AS PM10 IN
test statistic. Thus, for the primary and sec- THE ATMOSPHERE
ondary ozone standards, the 3-year average
annual fourth-highest daily maximum 8-hour 1.0 Applicability.

65

VerDate 11<MAY>2000 14:19 Jul 18, 2000 Jkt 190137 PO 00000 Frm 00065 Fmt 8010 Sfmt 8002 Y:\SGML\190137T.XXX pfrm07 PsN: 190137T
Pt. 50, App. J 40 CFR Ch. I (7–1–99 Edition)
1.1 This method provides for the measure- mechanism, there may be no upper limit.
ment of the mass concentration of particu- For samplers that do not have an automatic
late matter with an aerodynamic diameter filter-changing mechanism, the upper limit
less than or equal to a nominal 10 microm- is determined by the filter mass loading be-
eters (PM1O) in ambient air over a 24-hour yond which the sampler no longer maintains
period for purposes of determining attain- the operating flow rate within specified lim-
ment and maintenance of the primary and its due to increased pressure drop across the
secondary national ambient air quality loaded filter. This upper limit cannot be
standards for particulate matter specified in specified precisely because it is a complex
§ 50.6 of this chapter. The measurement proc- function of the ambient particle size dis-
ess is nondestructive, and the PM10 sample tribution and type, humidity, filter type, and
can be subjected to subsequent physical or perhaps other factors. Nevertheless, all sam-
chemical analyses. Quality assurance proce- plers should be capable of measuring 24-hour
dures and guidance are provided in part 58, PM10 mass concentrations of at least 300 µg/
appendices A and B, of this chapter and in std m3 while maintaining the operating flow
References 1 and 2. rate within the specified limits.
2.0 Principle. 4.0 Precision.
2.1 An air sampler draws ambient air at a 4.1 The precision of PM10 samplers must be
constant flow rate into a specially shaped 5 µg/m3 for PM10 concentrations below 80 µg/
inlet where the suspended particulate matter m3 and 7 percent for PM10 concentrations
is inertially separated into one or more size above 80 µg/m3, as required by part 53 of this
fractions within the PM10 size range. Each chapter, which prescribes a test procedure
size fraction in the PM1O size range is then that determines the variation in the PM10
collected on a separate filter over the speci- concentration measurements of identical
fied sampling period. The particle size dis- samplers under typical sampling conditions.
crimination characteristics (sampling effec- Continual assessment of precision via collo-
tiveness and 50 percent cutpoint) of the sam- cated samplers is required by part 58 of this
pler inlet are prescribed as performance chapter for PM10 samplers used in certain
specifications in part 53 of this chapter. monitoring networks.
2.2 Each filter is weighed (after moisture 5.0 Accuracy.
equilibration) before and after use to deter- 5.1 Because the size of the particles making
mine the net weight (mass) gain due to col- up ambient particulate matter varies over a
lected PM10. The total volume of air sampled, wide range and the concentration of par-
corrected to EPA reference conditions (25 C, ticles varies with particle size, it is difficult
101.3 kPa), is determined from the measured to define the absolute accuracy of PM10 sam-
flow rate and the sampling time. The mass plers. Part 53 of this chapter provides a spec-
concentration of PM10 in the ambient air is ification for the sampling effectiveness of
computed as the total mass of collected par- PM10 samplers. This specification requires
ticles in the PM10 size range divided by the that the expected mass concentration cal-
volume of air sampled, and is expressed in culated for a candidate PM10 sampler, when
micrograms per standard cubic meter (µg/std sampling a specified particle size distribu-
m3). For PM10 samples collected at tempera- tion, be within ±10 percent of that calculated
tures and pressures significantly different for an ideal sampler whose sampling effec-
from EPA reference conditions, these cor- tiveness is explicitly specified. Also, the par-
rected concentrations sometimes differ sub- ticle size for 50 percent sampling
stantially from actual concentrations (in effectivensss is required to be 10±0.5 microm-
micrograms per actual cubic meter), particu- eters. Other specifications related to accu-
larly at high elevations. Although not re- racy apply to flow measurement and calibra-
quired, the actual PM10 concentration can be tion, filter media, analytical (weighing) pro-
calculated from the corrected concentration, cedures, and artifact. The flow rate accuracy
using the average ambient temperature and of PM10 samplers used in certain monitoring
barometric pressure during the sampling pe- networks is required by part 58 of this chap-
riod. ter to be assessed periodically via flow rate
2.3 A method based on this principle will be audits.
considered a reference method only if (a) the 6.0 Potential Sources of Error.
associated sampler meets the requirements 6.1 Volatile Particles. Volatile particles col-
specified in this appendix and the require- lected on filters are often lost during ship-
ments in part 53 of this chapter, and (b) the ment and/or storage of the filters prior to
method has been designated as a reference the post-sampling weighing 3. Although ship-
method in accordance with part 53 of this ment or storage of loaded filters is some-
chapter. times unavoidable, filters should be re-
3.0 Range. weighed as soon as practical to minimize
3.1 The lower limit of the mass concentra- these losses.
tion range is determined by the repeatability 6.2 Artifacts. Positive errors in PM10 con-
of filter tare weights, assuming the nominal centration measurements may result from
air sample volume for the sampler. For sam- retention of gaseous species on filters 4, 5.
plers having an automatic filter-changing Such errors include the retention of sulfur

66

VerDate 11<MAY>2000 14:19 Jul 18, 2000 Jkt 190137 PO 00000 Frm 00066 Fmt 8010 Sfmt 8002 Y:\SGML\190137T.XXX pfrm07 PsN: 190137T
Environmental Protection Agency Pt. 50, App. J
dioxide and nitric acid. Retention of sulfur d. Protect the filter and sampler from pre-
dioxide on filters, followed by oxidation to cipitation and prevent insects and other de-
sulfate, is referred to as artifact sulfate for- bris from being sampled.
mation, a phenomenon which increases with e. Minimize air leaks that would cause
increasing filter alkalinity 6. Little or no ar- error in the measurement of the air volume
tifact sulfate formation should occur using passing through the filter.
filters that meet the alkalinity specification f. Discharge exhaust air at a sufficient dis-
in section 7.2.4. Artifact nitrate formation, tance from the sampler inlet to minimize the
resulting primarily from retention of nitric sampling of exhaust air.
acid, occurs to varying degrees on many fil- g. Minimize the collection of dust from the
ter types, including glass fiber, cellulose supporting surface.
ester, and many quartz fiber filters 5, 7, 8, 9, 10. 7.1.2 The sampler shall have a sample air
Loss of true atmospheric particulate nitrate inlet system that, when operated within a
during or following sampling may also occur specified flow rate range, provides particle
due to dissociation or chemical reaction. size discrimination characteristics meeting
This phenomenon has been observed on Tef- all of the applicable performance specifica-
tions prescribed in part 53 of this chapter.
lon filters 8 and inferred for quartz fiber fil-
The sampler inlet shall show no significant
ters 11, 12. The magnitude of nitrate artifact
wind direction dependence. The latter re-
errors in PM10 mass concentration measure-
quirement can generally be satisfied by an
ments will vary with location and ambient
inlet shape that is circularly symmetrical
temperature; however, for most sampling lo-
about a vertical axis.
cations, these errors are expected to be 7.1.3 The sampler shall have a flow control
small. device capable of maintaining the sampler’s
6.3 Humidity. The effects of ambient humid- operating flow rate within the flow rate lim-
ity on the sample are unavoidable. The filter its specified for the sampler inlet over nor-
equilibration procedure in section 9.0 is de- mal variations in line voltage and filter pres-
signed to minimize the effects of moisture on sure drop.
the filter medium. 7.1.4 The sampler shall provide a means to
6.4 Filter Handling. Careful handling of fil- measure the total flow rate during the sam-
ters between presampling and postsampling pling period. A continuous flow recorder is
weighings is necessary to avoid errors due to recommended but not required. The flow
damaged filters or loss of collected particles measurement device shall be accurate to ±2
from the filters. Use of a filter cartridge or percent.
cassette may reduce the magnitude of these 7.1.5 A timing/control device capable of
errors. Filters must also meet the integrity starting and stopping the sampler shall be
specification in section 7.2.3. used to obtain a sample collection period of
6.5 Flow Rate Variation. Variations in the 24 ±1 hr (1,440 ±60 min). An elapsed time
sampler’s operating flow rate may alter the meter, accurate to within ±15 minutes, shall
particle size discrimination characteristics be used to measure sampling time. This
of the sampler inlet. The magnitude of this meter is optional for samplers with contin-
error will depend on the sensitivity of the uous flow recorders if the sampling time
inlet to variations in flow rate and on the measurement obtained by means of the re-
particle distribution in the atmosphere dur- corder meets the ±15 minute accuracy speci-
ing the sampling period. The use of a flow fication.
control device (section 7.1.3) is required to 7.1.6 The sampler shall have an associated
minimize this error. operation or instruction manual as required
6.6 Air Volume Determination. Errors in the by part 53 of this chapter which includes de-
air volume determination may result from tailed instructions on the calibration, oper-
errors in the flow rate and/or sampling time ation, and maintenance of the sampler.
measurements. The flow control device 7.2 Filters.
serves to minimize errors in the flow rate de- 7.2.1 Filter Medium. No commercially avail-
termination, and an elapsed time meter (sec- able filter medium is ideal in all respects for
tion 7.1.5) is required to minimize the error all samplers. The user’s goals in sampling de-
in the sampling time measurement. termine the relative importance of various
filter characteristics (e.g., cost, ease of han-
7.0 Apparatus.
dling, physical and chemical characteristics,
7.1 PM10 Sampler. etc.) and, consequently, determine the
7.1.1 The sampler shall be designed to: choice among acceptable filters. Further-
a. Draw the air sample into the sampler more, certain types of filters may not be
inlet and through the particle collection fil- suitable for use with some samplers, particu-
ter at a uniform face velocity. larly under heavy loading conditions (high
b. Hold and seal the filter in a horizontal mass concentrations), because of high or
position so that sample air is drawn down- rapid increase in the filter flow resistance
ward through the filter. that would exceed the capability of the sam-
c. Allow the filter to be installed and re- pler’s flow control device. However, samplers
moved conveniently. equipped with automatic filter-changing

67

VerDate 11<MAY>2000 14:19 Jul 18, 2000 Jkt 190137 PO 00000 Frm 00067 Fmt 8010 Sfmt 8002 Y:\SGML\190137T.XXX pfrm07 PsN: 190137T
Pt. 50, App. J 40 CFR Ch. I (7–1–99 Edition)
mechanisms may allow use of these types of 8.1.2 Particle size discrimination by iner-
filters. The specifications given below are tial separation requires that specific air ve-
minimum requirements to ensure accept- locities be maintained in the sampler’s air
ability of the filter medium for measurement inlet system. Therefore, the flow rate
of PM10 mass concentrations. Other filter through the sampler’s inlet must be main-
evaluation criteria should be considered to tained throughout the sampling period with-
meet individual sampling and analysis objec- in the design flow rate range specified by the
tives. manufacturer. Design flow rates are speci-
7.2.2 Collection Efficiency. ≥99 percent, as fied as actual volumetric flow rates, meas-
measured by the DOP test (ASTM–2986) with ured at existing conditions of temperature
0.3 µm particles at the sampler’s operating and pressure (Qa). In contrast, mass con-
face velocity. centrations of PM10 are computed using flow
7.2.3 Integrity. ±5 µg/m3 (assuming sampler’s rates corrected to EPA reference conditions
nominal 24-hour air sample volume). Integ- of temperature and pressure (Qstd).
rity is measured as the PM10 concentration 8.2 Flow Rate Calibration Procedure.
equivalent corresponding to the average dif- 8.2.1 PM10 samplers employ various types
ference between the initial and the final of flow control and flow measurement de-
weights of a random sample of test filters vices. The specific procedure used for flow
that are weighed and handled under actual rate calibration or verification will vary de-
or simulated sampling conditions, but have pending on the type of flow controller and
no air sample passed through them (i.e., fil- flow indicator employed. Calibration in
ter blanks). As a minimum, the test proce-
terms of actual volumetric flow rates (Qa) is
dure must include initial equilibration and
generally recommended, but other measures
weighing, installation on an inoperative
of flow rate (e.g., Qstd) may be used provided
sampler, removal from the sampler, and final
the requirements of section 8.1 are met. The
equilibration and weighing.
general procedure given here is based on ac-
7.2.4 Alkalinity. <25 microequivalents/gram
tual volumetric flow units (Qa) and serves to
of filter, as measured by the procedure given
illustrate the steps involved in the calibra-
in Reference 13 following at least two
tion of a PM10 sampler. Consult the sampler
months storage in a clean environment (free
manufacturer’s instruction manual and Ref-
from contamination by acidic gases) at room
erence 2 for specific guidance on calibration.
temperature and humidity.
Reference 14 provides additional information
7.3 Flow Rate Transfer Standard. The flow
on the use of the commonly used measures of
rate transfer standard must be suitable for
flow rate and their interrelationships.
the sampler’s operating flow rate and must
be calibrated against a primary flow or vol- 8.2.2 Calibrate the flow rate transfer stand-
ume standard that is traceable to the Na- ard against a primary flow or volume stand-
tional Bureau of Standards (NBS). The flow ard traceable to NBS. Establish a calibration
rate transfer standard must be capable of relationship (e.g., an equation or family of
measuring the sampler’s operating flow rate curves) such that traceability to the primary
with an accuracy of ±2 percent. standard is accurate to within 2 percent over
7.4 Filter Conditioning Environment. the expected range of ambient conditions
7.4.1 Temperature range: 15 to 30 C. (i.e., temperatures and pressures) under
7.4.2 Temperature control: ±3 C. which the transfer standard will be used. Re-
7.4.3 Humidity range: 20% to 45% RH. calibrate the transfer standard periodically.
7.4.4 Humidity control: ±5% RH. 8.2.3 Following the sampler manufacturer’s
7.5 Analytical Balance. The analytical bal- instruction manual, remove the sampler
ance must be suitable for weighing the type inlet and connect the flow rate transfer
and size of filters required by the sampler. standard to the sampler such that the trans-
The range and sensitivity required will de- fer standard accurately measures the sam-
pend on the filter tare weights and mass pler’s flow rate. Make sure there are no leaks
loadings. Typically, an analytical balance between the transfer standard and the sam-
with a sensitivity of 0.1 mg is required for pler.
high volume samplers (flow rates >0.5 m3/ 8.2.4 Choose a minimum of three flow rates
min). Lower volume samplers (flow rates <0.5 (actual m3/min), spaced over the acceptable
m3/min) will require a more sensitive bal- flow rate range specified for the inlet (see
ance. 7.1.2) that can be obtained by suitable adjust-
8.0 Calibration. ment of the sampler flow rate. In accordance
8.1 General Requirements. with the sampler manufacturer’s instruction
8.1.1 Calibration of the sampler’s flow manual, obtain or verify the calibration re-
measurement device is required to establish lationship between the flow rate (actual m3/
traceability of subsequent flow measure- min) as indicated by the transfer standard
ments to a primary standard. A flow rate and the sampler’s flow indicator response.
transfer standard calibrated against a pri- Record the ambient temperature and baro-
mary flow or volume standard shall be used metric pressure. Temperature and pressure
to calibrate or verify the accuracy of the corrections to subsequent flow indicator
sampler’s flow measurement device. readings may be required for certain types of

68

VerDate 11<MAY>2000 14:19 Jul 18, 2000 Jkt 190137 PO 00000 Frm 00068 Fmt 8010 Sfmt 8002 Y:\SGML\190137T.XXX pfrm07 PsN: 190137T
Environmental Protection Agency Pt. 50, App. J
flow measurement devices. When such cor- 9.8 Set the timer to start and stop the sam-
rections are necessary, correction on an indi- pler at appropriate times. Set the elapsed
vidual or daily basis is preferable. However, time meter to zero or record the initial
seasonal average temperature and average meter reading.
barometric pressure for the sampling site 9.9 Record the sample information (site lo-
may be incorporated into the sampler cali- cation or identification number, sample
bration to avoid daily corrections. Consult date, filter identification number, and sam-
the sampler manufacturer’s instruction man- pler model and serial number).
ual and Reference 2 for additional guidance. 9.10 Sample for 24±1 hours.
8.2.5 Following calibration, verify that the 9.11 Determine
¯ and record the average flow
sampler is operating at its design flow rate rate (Qa) in actual m3/min for the sampling
(actual m3/min) with a clean filter in place. period in accordance with the instructions
8.2.6 Replace the sampler inlet. provided in the sampler manufacturer’s in-
9.0 Procedure. struction manual. Record the elapsed time
9.1 The sampler shall be operated in ac- meter final reading and, if needed, the aver-
cordance with the specific guidance provided age ambient temperature and barometric
in the sampler manufacturer’s instruction pressure for the sampling period (see note
manual and in Reference 2. The general pro- following step 9.6).
cedure given here assumes that the sampler’s 9.12 Carefully remove the filter from the
flow rate calibration is based on flow rates sampler, following the sampler manufactur-
at ambient conditions (Qa) and serves to il- er’s instruction manual. Touch only the
lustrate the steps involved in the operation outer edges of the filter.
of a PM10 sampler. 9.13 Place the filter in a protective holder
9.2 Inspect each filter for pinholes, par- or container (e.g., petri dish, glassine enve-
ticles, and other imperfections. Establish a lope, or manila folder).
filter information record and assign an iden- 9.14 Record any factors such as meteoro-
tification number to each filter. logical conditions, construction activity,
9.3 Equilibrate each filter in the condi- fires or dust storms, etc., that might be per-
tioning environment (see 7.4) for at least 24 tinent to the measurement on the filter in-
hours. formation record.
9.4 Following equilibration, weigh each fil- 9.15 Transport the exposed sample filter to
ter and record the presampling weight with the filter conditioning environment as soon
the filter identification number. as possible for equilibration and subsequent
9.5 Install a preweighed filter in the sam- weighing.
pler following the instructions provided in 9.16 Equilibrate the exposed filter in the
the sampler manufacturer’s instruction man- conditioning environment for at least 24
ual. hours under the same temperature and hu-
9.6 Turn on the sampler and allow it to es- midity conditions used for presampling filter
tablish run-temperature conditions. Record equilibration (see 9.3).
the flow indicator reading and, if needed, the 9.17 Immediately after equilibration, re-
ambient temperature and barometric pres- weigh the filter and record the postsampling
sure. Determine the sampler flow rate (ac- weight with the filter identification number.
tual m3/min) in accordance with the instruc- 10.0 Sampler Maintenance.
tions provided in the sampler manufacturer’s 10.1 The PM10 sampler shall be maintained
instruction manual. NOTE.—No onsite tem- in strict accordance with the maintenance
perature or pressure measurements are nec- procedures specified in the sampler manufac-
essary if the sampler’s flow indicator does turer’s instruction manual.
not require temperature or pressure correc- 11.0 Calculations.
tions or if seasonal average temperature and 11.1 Calculate the average flow rate over
average barometric pressure for the sam- the sampling period ¯corrected to EPA ref-
pling site are incorporated into the sampler erence conditions as Qstd. When the sampler’s
calibration (see step 8.2.4). If individual or flow indicator is ¯calibrated in actual volu-
daily temperature and pressure corrections metric units (Qa), Qstd is calculated as:
¯ ¯
are required, ambient temperature and baro- Qstd=Qa×(Pav/Tav)(Tstd/Pstd)
metric pressure can be obtained by on-site
measurements or from a nearby weather sta- where
¯
tion. Barometric pressure readings obtained Qstd = average flow rate at EPA reference
from airports must be station pressure, not 3
¯ conditions, std m /min;
corrected to sea level, and may need to be Qa = average flow rate at ambient conditions,
corrected for differences in elevation be- m3/min;
tween the sampling site and the airport. Pav = average barometric pressure during the
9.7 If the flow rate is outside the accept- sampling period or average barometric
able range specified by the manufacturer, pressure for the sampling site, kPa (or mm
check for leaks, and if necessary, adjust the Hg);
flow rate to the specified setpoint. Stop the Tav = average ambient temperature during
sampler. the sampling period or seasonal average

69

VerDate 11<MAY>2000 14:19 Jul 18, 2000 Jkt 190137 PO 00000 Frm 00069 Fmt 8010 Sfmt 8002 Y:\SGML\190137T.XXX pfrm07 PsN: 190137T
Pt. 50, App. K 40 CFR Ch. I (7–1–99 Edition)
ambient temperature for the sampling site, ies of Major Sampling Interferences. Atmos.
K; Environ., 13:543, 1979.
Tstd = standard temperature, defined as 298 K; 10. Appel, B.R. Letter to Larry Purdue,
Pstd = standard pressure, defined as 101.3 kPa U.S. EPA, Environmental Monitoring and
(or 760 mm Hg). Support Laboratory. March 18, 1982, Docket
No. A–82–37, II–I–1.
11.2 Calculate the total volume of air sam- 11. Pierson, W.R., W.W. Brachaczek, T.J.
pled as: Korniski, T.J. Truex, and J.W. Butler. Arti-
¯
Vstd = Qstd×t fact Formation of Sulfate, Nitrate, and Hy-
drogen Ion on Backup Filters: Allegheny
where Mountain Experiment. J. Air Pollut. Control
Vstd = total air sampled in standard volume Assoc., 30:30, 1980.
units, std m3; 12. Dunwoody, C.L. Rapid Nitrate Loss
t = sampling time, min. From PM10 Filters. J. Air Pollut. Control
Assoc., 36:817, 1986.
11.3 Calculate the PM10 concentration as:
13. Harrell, R.M. Measuring the Alkalinity
PM10 = (Wf¥Wi)×106/Vstd of Hi-Vol Air Filters. EMSL/RTP–SOP–QAD–
534, October 1985. Available from the U.S. En-
where
vironmental Protection Agency, EMSL/QAD,
PM10 = mass concentration of PM10, µg/std Research Triangle Park, NC 27711.
m3; 14. Smith, F., P.S. Wohlschlegel, R.S.C.
Wf, Wi = final and initial weights of filter col- Rogers, and D.J. Mulligan. Investigation of
lecting PM1O particles, g; Flow Rate Calibration Procedures Associ-
106 = conversion of g to µg. ated With the High Volume Method for De-
NOTE: If more than one size fraction in the termination of Suspended Particulates.
PM10 size range is collected by the sampler, EPA–600/4–78–047, U.S. Environmental Pro-
the sum of the net weight gain by each col- tection Agency, Research Triangle Park, NC
lection filter [Σ(Wf¥Wi)] is used to calculate 27711, 1978.
the PM10 mass concentration. [52 FR 24664, July 1, 1987; 52 FR 29467, Aug. 7,
12.0 References. 1987]
1. Quality Assurance Handbook for Air Pol-
lution Measurement Systems, Volume I, APPENDIX K TO PART 50—INTERPRETA-
Principles. EPA–600/9–76–005, March 1976. TION OF THE NATIONAL AMBIENT AIR
Available from CERI, ORD Publications, QUALITY STANDARDS FOR PARTICU-
U.S. Environmental Protection Agency, 26 LATE MATTER
West St. Clair Street, Cincinnati, OH 45268.
2. Quality Assurance Handbook for Air Pol- 1.0 General.
lution Measurement Systems, Volume II, (a) This appendix explains the computa-
Ambient Air Specific Methods. EPA–600/4–77– tions necessary for analyzing particulate
027a, May 1977. Available from CERI, ORD matter data to determine attainment of the
Publications, U.S. Environmental Protection 24-hour and annual standards specified in 40
Agency, 26 West St. Clair Street, Cincinnati, CFR 50.6. For the primary and secondary
OH 45268. standards, particulate matter is measured in
3. Clement, R.E., and F.W. Karasek. Sam- the ambient air as PM10 (particles with an
ple Composition Changes in Sampling and aerodynamic diameter less than or equal to
Analysis of Organic Compounds in Aerosols. a nominal 10 micrometers) by a reference
Int. J. Environ. Analyt. Chem., 7:109, 1979. method based on appendix J of this part and
4. Lee, R.E., Jr., and J. Wagman. A Sam- designated in accordance with part 53 of this
pling Anomaly in the Determination of At- chapter, or by an equivalent method des-
mospheric Sulfate Concentration. Amer. Ind. ignated in accordance with part 53 of this
Hyg. Assoc. J., 27:266, 1966. chapter. The required frequency of measure-
5. Appel, B.R., S.M. Wall, Y. Tokiwa, and ments is specified in part 58 of this chapter.
M. Haik. Interference Effects in Sampling (b) The terms used in this appendix are de-
Particulate Nitrate in Ambient Air. Atmos. fined as follows:
Environ., 13:319, 1979. Average refers to an arithmetic mean. All
6. Coutant, R.W. Effect of Environmental particulate matter standards are expressed
Variables on Collection of Atmospheric Sul- in terms of expected annual values: Expected
fate. Environ. Sci. Technol., 11:873, 1977. number of exceedances per year for the 24-
7. Spicer, C.W., and P. Schumacher. Inter- hour standards and expected annual arith-
ference in Sampling Atmospheric Particu- metic mean for the annual standards.
late Nitrate. Atmos. Environ., 11:873, 1977. Daily value for PM10 refers to the 24-hour
8. Appel, B.R., Y. Tokiwa, and M. Haik. average concentration of PM10 calculated or
Sampling of Nitrates in Ambient Air. Atmos. measured from midnight to midnight (local
Environ., 15:283, 1981. time).
9. Spicer, C.W., and P.M. Schumacher. Par- Exceedance means a daily value that is
ticulate Nitrate: Laboratory and Field Stud- above the level of the 24-hour standard after

70

VerDate 11<MAY>2000 14:19 Jul 18, 2000 Jkt 190137 PO 00000 Frm 00070 Fmt 8010 Sfmt 8002 Y:\SGML\190137T.XXX pfrm07 PsN: 190137T

You might also like