Operation Manual
Anyscan‐36
I
Content
Chapter I Overview ...................................................................................................................... 1
Chapter II Specifications & Features .............................................................................................. 2
2.1 Basic Specifications .......................................................................................................... 2
2.2 Main Functions ................................................................................................................. 3
Chapter III Structures ..................................................................................................................... 4
3.1 Buttons & Menu Settings ................................................................................................. 4
3.1.1 Power Key .............................................................................................................. 4
3.1.2 Gain & Gain Adjustment ....................................................................................... 4
3.1.3 Main Menu Page Keys ........................................................................................... 5
3.1.4 Parameter Adjustment .......................................................................................... 5
3.1.5 Freeze Key ............................................................................................................. 5
3.1.6 Main Menu & Submenu Buttons .......................................................................... 5
3.1.7 Submenu Functions & Menu Parameters ............................................................. 6
3.2 Introduction of Instrument Interface ............................................................................... 6
3.3 Battery Supply .................................................................................................................. 7
Chapter IV Menu & Parameter Settings ....................................................................................... 8
4.1 Display Interface ............................................................................................................... 8
4.2 Menu Structure ................................................................................................................ 8
4.2.1 BASIC ..................................................................................................................... 8
4.2.2 PLSRCVR ................................................................................................................ 9
4.2.3 GATE ...................................................................................................................... 9
4.2.4 PROBE .................................................................................................................. 10
4.2.5 MEMO ................................................................................................................. 10
4.2.6 DAC ...................................................................................................................... 11
4.2.7 AVG ...................................................................................................................... 12
4.2.8 CAL ...................................................................................................................... 12
4.2.9 MEASURE ............................................................................................................ 13
4.2.10 DISPLAY ............................................................................................................. 13
4.2.11 CONFIG .............................................................................................................. 14
4.2.12 TEST SETUP ........................................................................................................ 15
4.2.13 HOTKEY .............................................................................................................. 16
4.3 Selection of Menus & Actions ........................................................................................ 17
4.3.1 Main Menu Selection .......................................................................................... 17
4.3.2 Submenu Options ................................................................................................ 18
4.3.3 Main Menu Shortcuts ......................................................................................... 18
4.3.4 Submenu Operating Characteristics .................................................................... 18
4.4 Parameter Settings & Confirmation ............................................................................... 19
4.4.1 Parameter Settings .............................................................................................. 19
4.4.2 Parameter Confirm & Save .................................................................................. 19
4.4.3 "Preview" List ...................................................................................................... 19
Chapter V Instrument Main Function .......................................................................................... 21
II
5.1 Constraints Between the Detection Range and PRF ...................................................... 21
5.2 Square‐wave Emission and Damping, Band’s Setting .................................................... 21
5.2.1 Square‐wave Emission ......................................................................................... 21
5.2.2 The Damping and Band Mix ................................................................................ 22
5.3 Gate Function and Its Setting ......................................................................................... 22
5.3.1 Gate Basic Function ............................................................................................. 23
5.3.2 Gate New Function .............................................................................................. 23
5.4 Probe Parameter Setting ................................................................................................ 24
5.4.1 TYPE‐SPEC ............................................................................................................ 24
5.4.2 P‐DELAY ............................................................................................................... 24
5.4.3 ANGLE ................................................................................................................. 24
5.4.4 X‐VALUE ............................................................................................................... 25
5.5 Calibration Function ....................................................................................................... 25
5.5.1 Auto‐calibration of Longitude‐wave Incident Point and Velocity of Normal Probe
......................................................................................................................................... 25
5.5.2 X‐VALUE Test of Angle Probe ............................................................................... 27
5.5.3 Delay Auto‐Calibration of Angle Probe ............................................................... 27
5.5.4 Angle Auto‐Calibration of Angle Probe ............................................................... 28
5.6 Drawing and Application of DAC Curve .......................................................................... 29
5.6.1 DAC Curve Drawing. ............................................................................................ 29
5.6.2 DAC Curve Adjust, Modify, Store and Delete ...................................................... 30
5.6.3 Applications and Considerations for DAC Curves ................................................ 31
5.7 Storage of Inspection Data and Recall ............................................................................ 32
5.7.1 DATA ID ................................................................................................................ 32
5.7.2 Remarks ............................................................................................................... 33
5.7.3 Recall ................................................................................................................... 33
5.7.4 Record ................................................................................................................. 34
5.8 TEST SETUP ..................................................................................................................... 34
5.8.1 Auto Gain and Gain Amend................................................................................. 34
5.8.2 Envelope .............................................................................................................. 35
5.8.3 Surf Amend ......................................................................................................... 36
5.8.4 Weld Setting ........................................................................................................ 36
5.9 Measurement ................................................................................................................. 36
5.9.1 TOF ...................................................................................................................... 36
5.9.2 Peak Holding ....................................................................................................... 37
5.9.3 Echo Compare ..................................................................................................... 37
5.9.4 Readings .............................................................................................................. 37
5.9.5 FFT ....................................................................................................................... 38
5.10 Display and Configuration ............................................................................................ 38
5.10.1 Filled .................................................................................................................. 38
5.10.2 Scheme .............................................................................................................. 38
5.10.3 Brightness .......................................................................................................... 39
5.10.4 Freeze ................................................................................................................ 39
5.10.5 Nse Immn .......................................................................................................... 39
III
5.10.6 Buzzer ................................................................................................................ 39
5.10.7 Screen Copy & Data Copy .................................................................................. 39
5.10.8 Languages & Units ............................................................................................. 40
5.10.9 Date, Time ......................................................................................................... 40
5.10.10 Preview ............................................................................................................ 40
5.10.11 Help ................................................................................................................. 40
5.10.12 Software Update ............................................................................................. 40
5.10.13 Default Setting................................................................................................. 41
5.11 AVG Function ................................................................................................................ 41
5.11.1 Summarize ......................................................................................................... 41
5.11.2 AVG Making ....................................................................................................... 41
5.11.3 Notes ................................................................................................................. 42
5.11.4 AVG Application ................................................................................................ 42
Chapter VI Brief Introduction of UT Application ............................................................................. 43
6.1 Preparation of Flaw Detection .......................................................................................... 43
6.2 Weld Flaw Detection ...................................................................................................... 44
6.2.1 Probe Detect and Instrument Setting ................................................................. 44
6.2.2 Key Points for Field Inspection ............................................................................ 45
6.3 Flaw Detection of Axis Forging Piece ............................................................................. 45
6.4 Other Mechanical Parts Flaw Detection ......................................................................... 46
6.5 Attention Problems of Ultrasonic Test ........................................................................... 47
6.5.1 Probe Precision of Echo Wave Frequency ........................................................... 47
6.5.2 Mind Whether Frequency Range of the Instrument Meets Standard Value ...... 48
6.5.3 Instrument Settings When Use High Frequency Narrow Pulse Probe to Perform
High Resolution Ultrasonic Test ....................................................................................... 49
6.5.4 Mind Choosing the Measure Points .................................................................. 49
Chapter VII Communication Printing of Anyscan‐36 ....................................................................... 50
7.1 Port ................................................................................................................................. 51
7.2 Data Communication ..................................................................................................... 51
7.3 Use Instruction ............................................................................................................... 51
7.3.1 Menu Construction ............................................................................................. 51
7.3.2 Display Interface Structure .................................................................................. 53
7.4 Test Report ..................................................................................................................... 54
7.5 Software Update ............................................................................................................ 55
7.5.1 Software Updating .............................................................................................. 55
7.5.2 Instrument Parameters Default Settings ............................................................. 55
Chapter VIII Safe Operation Prompts ........................................................................................... 55
8.1 Use Manual and the Instruments ...................................................................................... 55
8.1.1 Instrument Power Work ......................................................................................... 56
8.1.2 Preview Function and Help for Flexible Menu .................................................... 57
8.1.3 Holding the Instrument in Filed Application ....................................................... 58
8.2 Emphasis on the Training and Application of Ultrasonic Testing Foundation ................... 58
8.2.1 Inspect with Qualified Certificate ........................................................................... 58
8.2.2 Periodic Inspect the Instrument Strictly Compliance with Industry Standards..... 58
IV
8.2.3 Reduce Misjudgment Rate by Expert Diagnosis .................................................. 59
8.3 Attention to Safe Electricity ........................................................................................... 59
Chapter IX Maintenance and Repair ............................................................................................ 59
9.1 Instrument Maintenance .............................................................................................. 59
9.2 Instrument Maintenance and Warranty ........................................................................ 60
9.3 User Notice ..................................................................................................................... 61
APPENDIX ........................................................................................................................................ 61
Appendix I Definition of Readings ...................................................................................... 61
Appendix II Range and Repetition Frequency ...................................................................... 62
V
Chapter I Overview
Anyscan-36 Digital Ultrasonic Flaw Detector is choicest model that Doppler devotes to
NDT industry, not only easy to use, concise & practical appearance, but also adhering to
the concept of excellent products, industry-leading, and great performance. And more
advanced technical indicators compared with any other similar products, also with
abundant & practical additional functions, which bring more convenient & efficient to
users.
Main Features:
• Optional Square / Spike transmit mode, suitable for welding & forging detections.
• Realistic welding diagrams combined with 400MHz hardware real-time sampling,
and recording playback.
• 22MHz wide bandwidth & low noise amplifier fitted with a variety of probes for high
sensitivity and high resolution detections.
• New software interface, multiple shortcut functions, convenient on-site operations.
• Compact design, high IP rating, particularly well-suited for field & aerial work
detections.
1
2
· Echo Spectrum Auto-test Function: To test the center frequency & spectrum of echo
of probes.
· Auto-gain Function: automatic adjust gain to achieve quick adjust of echo amplitude.
· Welding Illustration Function: Direct display of weld diagrams, easy & clear to see
defect location, diagrams can be stored into reports, very convenient to playback & print.
· Continuous Save & Playback Functions: dynamic waveform recording, more detailed
data records; files can be saved to the USB storage device directly.
· Zoom Function: Waveform display area can be enlarged to full screen, can effectively
prevent mis-operation during on-site detection
· Gate Area Zoom-in Function: Zoom in the gate zone into whole waveform display zone,
easy to observe the wave details.
3
Gain Adjustment:
: Gain Step Control Key, press to switch gain step value: 0.0, 0.1, 0.5, 1.0, 2.0, 6.0dB;
(Following as "↔")
4
: Gain Increase Key, each press to increase the gain value; (following as "dB+")
: Gain Decrease Key, each press to decrease the gain value; (Following as "dB-")
Note: when step value is "0.0dB", that the gain cannot be adjusted to prevent
misoperation during on-site operation.
: Press to enter previous page of menu, can be used as Save Key when "Freeze"
status is applied.
: Press to enter the next page of menu, the keys can also be used to view the "Help"
menu.
: Parameter Increase Key, press to increase step value of the submenu parameter,
keep holding the key for fast increase value; When Screen Tips appear, the key is used to
confirm the operation.
: Parameter Decrease Key, press to decrease step value of the submenu parameter,
keep holding the key for fast decrease value; when the Screen Tips appear, the key is
used to cancel the operation.
: Press one time to freeze screen waveform and data, and press again to defreeze.
When "Freeze", the right corner of the screen displays "※" symbol.
. Current selected main menu text item would be in yellow, on deep blue background;
Color of text item which not selected would be in white, on light blue background. Press
one time, the selected main menu (blue on yellow background), right of the screen will
display the corresponding submenu; if the submenu has several pages, then press the ▲
key corresponding to main menu, the submenu pages will loop around.
5
: Press one time, select the corresponding submenu (yellow on blue background); If
the submenu labeled ">" symbol, which indicates that the submenu is switchable, every
press can cycle select the submenu; for the adjustable parameter values of submenu,
cycle switches between steps on every press.
①LEMO‐01 Probe Sockets (C9)
②Ethernet Interface
③USB‐B interface
④USB‐A Interfaces
⑤DC power socket
⑥Buckle
⑦Battery Cover ⑧Back Bracket
6
(1) Insert Battery: Battery case is at the backside of instrument, use a screwdriver to
open the battery cover, insert the battery plug into the machine socket and place the
battery well, then close the battery cover and tighten the screws.
(2) Battery Power Indicator: The right corner with five-cell symbol is the power indicator
of battery, if battery is less than two cell, it should stop testing immediately, and save
parameters and inspection data, then replace the battery or recharge.
(3) Battery Charging: Please use the manufacturer-specified power adapter to charge
battery; please read the 9th chapter of this manual carefully before use the adapter. The
battery charging time will take about 4.5 hours on this instrument.
Note: Please bring spare batteries with, if you require a long period of time of on-site
operation, especially in low outdoor temperatures winter, when the battery discharge
efficiency decreased attention to more.
7
Gain‐Step Value
Submenu
Gate A
Measurement
Data
Main Menu
4.2.1 BASIC
The main menu contains one page of submenu: "BASIC".
8
4.2.2 PLSRCVR
The main menu contains two pages of submenus: "PLUSE 1>/PLUSE 2>" and "FREQ
BAND", "PULSE 1>" and "PLUSE 2>" display in one page.
(1) Select " PLUSE 1>/PLUSE 2>", display the following submenu (1/2):
① PLUSE 1>/PLUSE 2>: Press the key to scroll
• Select "PLUSE 1>" for "Square Wave", Pulse Width Range is 40~1000ns,
press《+》,《-》keys to select pulse width; Select " PLUSE 2>" for "Spike Wave"
• When “PROBE”-- "TYPE SPEC” 2.5 MHz is selected, Square Wave pulse
width automatically defaults to "200ns".
• When “PROBE”-- "TYPE SPEC” 5.0 MHz is selected, Square Wave pulse
width automatically defaults to "100ns"
• When “PROBE”-- "TYPE SPEC” 10.0 MHz is selected, Square Wave pulse
width automatically defaults to "50ns".
② DAMPING: 50/100/200/500Ω; default value is "200Ω".
③ DUAL: I/O; default value is "off".
4.2.3 GATE
The main menu contains 2 pages of submenus: "GATE A", "GATE B".
9
"40%".
4.2.4 PROBE
The main menu contains 1 page of submenu: "PROBE".
4.2.5 MEMO
The main menu contains 2 pages of submenus: "MEMO", "RECORD".
10
③ U DISK SAVE: ; will save the contents to a USB drive, press 《+》 button to
confirm, press 《-》 key to exit.
④ REPLAY: REPLAY/STOP recording, press 《》 key to switch, press 《+》 key to
confirm.
4.2.6 DAC
The main menu contains 4 pages of submenus: "record", "amend", "dac" and "off".
11
4.2.7 AVG
The main menu contains 4 pages of submenus: ">3N PLOT", "<3N PLOT", "AVG" and
"OFF".
(2) Select "<3N PLOT" appears when the following submenu (2/4):
① ECHO: Display the current echo number of reference points (up to 10), press 《+》
《-》 to increase or decrease points.
② A START: Enter starting point of GATE A, Range: 0~5000mm.
③ Save: Save the AVG Curve.
4.2.8 CAL
The main menu contains submenus 2 page: “VELOC/DELAY", " ANGLE CAL".
(1) Select “VELOC/DELAY”, display the following submenu (1/2):
① VELOC/DELAY: Display value after calibration and save "VELOC/DELAY" at the
same time.
• When Straight Probe N is selected, default is "5920m/s/0.00 μs"
• When Angle Probe A is selected, default is "3230m/s/0.00 μs"
② S-REF1> / S-REF2>: "S-REF1>" defaults value is "50.0mm",” S-REF2>" defaults
value is "100.0mm" .
③ A START: 0.00~15000mm; GATE A default logic is "positive", starting from
20.00mm, width 40.00mm, threshold 40%.
④ CAL: Press 《》 key after calibration, and value will be saved.
12
4.2.9 MEASURE
The main menu contains 2 pages of submenus: "TOF" and "FFT".
(2) Select "FFT", display the when the following submenus (2/2):
① FFT: ON/OFF; default value is "OFF", press 《+》,《-》 keys to switch; When set
"ON", the default values are set as below:
• Straight Probe N default: Range 10, velocity 5920, Pulse Shift 20, RF, automatic
amplitude 80% (applies to CSK-IA test block, thickness 25mm).
• Angle Probe A default: Range 10, velocity 3230, Pulse Shift 95, RF ,automatic
amplitude 80% (applies to CSK-IA test block, R100).
Note: (a) When "FFT " set "on" and the echo displays normal, press 《+》 key to check
echo spectrum, after finished, display center frequency and -6dB spectrum of
echo under freeze state, and "FFT" automatic reset "OFF";
(b) at this time, press 《※》key for one time to unfreeze, instrument back to state
4.2.10 DISPLAY
The main menu contains 2 pages of submenus: "FILLED" and "FREEZE".
13
④ BASELINE: sound path / horizontal / vertical / div; default value is "sound path"
4.2.11 CONFIG
The main menu includes three pages of submenus:
“NSE IMMN", " LANGUAGE " and " PREVIEW ".
If the instrument cannot be started for some reason, press and hold the 《PgUp》
and 《PgDn》 key under shutdown state, and then press the power key to turn into
backup systems, in accordance with the method(3)-③ can be normal after you
14
(1) select " AUTO GAIN ", display the following submenu (1/3):
① AUTO GAIN: 20%~80%/OFF; automatically adjusts the gain to get desired echo
amplitude, the gain adjustment associated with "base gain"
② GAIN AMEND: 0.0~20.0dB, gain correction associated setting value "additional
gains"
③ ENVELOPE: ON/OFF, default to "OFF"
④ SURF AMEND: ¢200.0~2000mm/plane; Default is "Flat"
Note: The main menu HOTKEY "AUTO GAIN" shortcut menu, only switch within
"OFF/80%"
(3) Select "A INDIC" appears when the following submenu (3/3), was selected into the
weld rating feature.
① A INDIC
② B REF
③ C ATTEN
④ D RATING
Note: Here enclosed weld rating method:
(1) In order to ensure accurate reading of Sa, should first entry the correct parameters of
probes, and then perform delay and angle calibration;
(2) enter "AWS D1.1" menu;
(3) The angle probe identified maximum echo of IIW block or CSK-IA block 1.5 diameter
side-drilled hole, adjust gate A to capture it, makes the echo amplitude to 50% by
adjust the gain (instrument tolerance of ± 5%), select "B REF", press again and the
current gain value is automatically inputted into the instrument for B reference value;
(4) After founding flaw echo, adjust gate A to capture it, adjust the gain to make the flaws
echo to 50% (instrument tolerance of ± 5%), select "A INDIC", and then press the
15
corresponding key on the right side again, the current gain value is automatically
inputted into the instrument for A indication value;
(5) After getting A/B value, the " C ATTEN " is automatically calculated through sound
path Sa, the formula is as follows:
(a) The unit is "millimeter (mm)": C= (Sa/25.4-1) x 2, result rounded to keep one
decimal.
(b) Units to "inches (inch)": C= (Sa-1) x 2, result rounded to keep one decimal.
(6) "D rating" generated according to the following formula: D=A – B – C.
(7) According to the "AWS D1.1" specifications, weld defect level is obtained by checking
reference table identification.
4.2.13 HOTKEY
Anyscan-36 type added HOTKEY in main menu, which contains: "AUTO GAIN",
“ENVELOPE", "ZOOM", "SAVE", "A MAGNIFY", "B MAGNIFY", "A START/A WIDTH, B
START/B WIDTH”, and "PEAK HOLDING" total 8 sections, select and press to implement
the function, very convenient for application.
(2) ENVELOPE: record echo peak change of trajectory with a purple line; ON/OFF,
default is "OFF".
16
Press " A MAGNIFY " 《》 to display full screen, convenient to observe echo details; At
this time, if adjust appropriate "Start " and "Width" of GATE A, can make any point of echo
within detection range broadening to full screen.
(7) A START/A WIDTH, B START/B WIDTH: Gate related shortcut function keys
The menu comprises "A START>", "A WIDTH>", "B START>", "B WIDHT> ". If you
choose "A MAGNIFY", then the menu automatically displays "A START>", you can use
the 《+》《-》 keys to adjust the parameter; if you select "B MAGNIFY", then the menu
automatically displays "B START>".
(2) when you select main menu, press 1 time under their corresponding to the 《▲》 key,
the main menu has changed from light blue background, white text blue yellow at the end
of a Word is "selected", and at the same time on the right side of the screen displays the
main menu the submenu;
(3) in this case, select a different main menu as described above, the master menu is quit
(back light blue background, white text), the corresponding submenu set value is saved;
(4) the instrument shuts down and start up after, the screen displays the last instrument
parameter settings for normal shutdown.
17
(2) HOTKEY main menu consists of 3 pages, field testing commonly used functions on the
Home;
(3) For the purposes of quick action, "automatic amplitude" switch "off/80%", other
associated submenu menu with the same name.
Note: (a) In the "Freeze" State, the 《PgUp》 key is set to "SAVE" shortcut. Press 1 time
during testing the 《※》freeze screen displays the graphic, and then press the
《PgUp》 key, the screen graphics are deposited in the ID with tips.
(b) the submenu "save" under HOTKEY is different from《PgUp》key under "freeze"
state, which does not limited to "freeze" and with higher degree of freedom.
(1) Function switch submenu (> symbols, such as "square-wave emission >"), press the
《》 key can switch functions;
(2) Parameter with large range (as "range", and "velocity", and gate "start"), by press 《》
key can rough adjust:
(3) The rough adjust of "VELOCITY" ("BASIC" ) and "ANGLE" (PROBE) are their common
used values, such as: 3,230, 5920m/s or 45 °, 63.4°. Field testing applications very
convenient;
18
(4) To the gate and correction type submenu (such as "velocity/delay"), has a default
value, initial screening only slightly.
For example, select the "Range" (Main Menu) and set to 300.0mm, press "VELOCITY"
corresponding key《》,the selected menu change from "Range" (back to light blue
background white text) into "VELOCITY" (dark blue background yellow text), formerly
"Range" set value are recognized and preserved.
When "Preview" submenu (under CONFIG main menu) set on, the screen pop-up menu
list and the default select first row/column. At this point, press the《dB+》, 《dB-》 key to
move the cursor left and right, press the 《PgUp》,《PgDn》keys cursor loop page turning,
you can quickly browse through menu settings or select to amend menu settings.
In the "Preview" list, menu parameter setting method is same with the "menu settings"
interface, press CONFIG to quit "Preview" interface.
19
Figure 4-2 “Preview” menu interface page1 Figure 4-3 “Preview” menu interface page 2
20
Anyscan-36 with compact design in superb performance and powerful function, not only
suitable for large flowability and high overhand weld inspection operation, but also can be
used in forging, aircrafts parts and other high sensibility and resolution ultrasonic testing.
This chapter focus on introducing characteristics of instrument’s function.
But considering some unknown echo needs to be judged by reducing repetition frequency
under forging detection, it is still in the main menu under the "PLSRCVR" main-menu with
the “PRF”.
The choice of “PRF” range from 40~1000HZ in 40/250/1000 optional, default value 250HZ
according to 250mm “Range” .
While “PRF” setting is beyond the restrict in table 2 of Appendix 2,system will respond with
red words as a clue, at that time, follow the screen tips.
In view of the characteristics of the spectral distribution of the square wave pulse ,frontier
steeper square wave pulse suitable for 2 ~ 5MHz probe flaw detection applications, it’s the
main reason for using square-wave emission in weld and forging parts inspection.
This instrument with pulse width auto adjust function for convenience of on-site
applications, enter the nominal frequency first at “TYPE SPEC” of PROBE menu, the
pulse width will auto change to accordingly value in “Square-wave emission”. For example:
enter the probe frequency as 2.5MHZ, pulse width auto changed to 200ns, enter the
probe frequency as 5MHZ,pulse width auto changed to 100ns;while probe
frequency≥10MHz,pulse width as 50ns.Just slightly check in initial use.
21
Although the transmit system of ultrasonic flaw detector and probe won’t be included in
the scope of "power transfer" and pay attention to the match, but the square-wave
transmitter circuit load capacity is still a very important indicator, so “the effective output
impedance of transmitter circuit” be used in so many standards(same in spike pulse
transmit system). Anyscan-36’s effective output impedance approximately 10Ω, load
capacity is quite strong; For example: The transmit pulse amplitude remains ≥ 300V while
connect with 2.5P20N probe (the impedance approximately 50Ω),with much highly
sensitivity in flaw detection.
The receive system of Anyscan-36 adopt with a new design of high gain, wideband and
low noise amplifier, divided into 1~6/0.5~22/3~22 three different gear according to
different requirement. The -3dB band width≥20MHz in 0.5 ~ 22 and 3 ~ 22 MHz two
wideband gears, it can be totally used in high sensitivity inspections with narrow pulse
probe under 20 MHz.
22
functions of the alarm, echo distance and amplitude measurements, they also increase “A
magnify” and “B magnify” functions, which provide a convenient conditions for further
expansion of application.
(2)Gate A/B can measure the distance and amplitude of the echo inside the gate. A
large font displayed in the upper right corner of the screen from measure value
display; Conventional measured value of gate A is sound path “Sa”, horizontal
distance “Pa”, vertical distance “Da”, echo amplitude “Ha” etc.
(3)Used for material velocity amend, probe delay and angle amend etc. See 5.5.
(4)Gate A used for monitoring echo amplitude and distance point by point in make DAC
curve.
(5)In Auto Gain mode, the highest echo in gate A auto adjust to setting amplitude, gain
displayed in “Basic gain”.
(a)For more accurate test result by using Gate measuring, select Peak or Flank in
Note:
Peak/ Flank (menu of MEASURE) before testing. This instrument set a default
value according to different requirement of various measurements for
convenience, slightly check in initial use.
measurement; chose the same acoustic point by using secondary echo test
echo distance.
Adjust the Gate width to capture flaw echo for observation. When opening "Magnify"
function, the width of the gate will be widened to full screen, the echo is amplified, and its
details are clearly displayed.
Anyscan-36 not only widened 1mm small "range" to full screen, but also by adjusting the
23
gate "starting point" to enlarge the observed detection range at any point within the echo
details.
5.4.1 TYPE-SPEC
Enter straight/angle probe specification, N means straight probe, A means angle probe.
After input probe type specification, all probe feature and measurement reserve default
value, much easier for application. For example:
Chose straight probe, setting value of material velocity in submenu auto change
to 5920m/s, other submenu parameters like “thickness”, “weld set”, “weld width”
all unajustable.
Pulse width of “Square-wave emission” will auto adjust according to probe
frequency. See 5.2.
and its angle;element size of the probe is not essential, if element size isn’t
5.4.2 P-DELAY
Make sure the probe delay will facilitate the measurement of defect distance, and
therefore should enter the known value of the probe delay. Failure to determine the delay
of the probe, should process “delay calibration” based on fellows 5.5 method, the delay
value will auto saved in the menu.
5.4.3 ANGLE
Enter the known value of “angle” to facilitate the defect in flaw detection with angle probe,
if “angle” unknown or need to re-confirmed, go for calibration according to fellow 5.5
method, the values will auto saved in the menu.
24
Note: there is coarse/fine-tune setting mode in “ANGLE” menu, press《》at first time
appear with “*” for fine-tune mode, press again without “*” appear for coarse mode.
In coarse setting mode, all universal angle value have been pre-set, if still for
fine-tune, you can select the approximately value in coarse setting mode, then
5.4.4 X-VALUE
Enter the known value of X-VALUE to facilitate the defect in flaw detection with angle
probe, especially the instrument opened weld illustration (set sub-menu" weld type "on)
function, use this X-VALUE for calculate and display the defect position, see 5.8.4.
Calibration principle: the instrument makes sound velocity and probe delay calculations
through arrival time (T1 and T2) of two echoes (B1 and B2) with known sound path (S1
and S2).
Taking the CSK-ⅠA test block as an example, we will introduce four kinds of calibration
methods as follows.
25
(3) Enter the CAL main menu, select: CAL SET: speed/delay; S-REF 1 >: 100.0mm;
S-REF 2 >: 200.0mm.
(4) Move Gate A to capture the first bottom echo (refer to Figure 5-2). Enter the CAL
submenu, and press [+] key to confirm echo 1, then "0" is turned to "1" in CAL
submenu.
(5) Move Gate A to capture the second bottom echo (refer to Figure 5-3). Enter the CAL
submenu, press [+] key to confirm echo 2 (after "calibration completed" being
displayed in the screen, "1" is turned to "0" in CAL submenu), then the delay calibration
value will display in CAL submenu and also in P-DELAY submenu.
26
(1) Put the angle probe on B position of the CSK-ⅠA test block, and make sure the
coupling is good. Refer to Figure 5-1.Adjust the gain to a appropriate value and find
the highest reflected echo of R100 arc surface.
(2) You can get the X-VALUE by reading the scale value on the probe housing which is
aligned to the center groove of the R100 arc. Moreover, you can get the X-VALUE by
calculating the formula "X-VALUE = 100-L", and you can get "L" by measuring with a
ruler (refer to Figure 5-1).
(1) Put the angle probe on B position of the CSK-ⅠA test block, and make sure the
coupling is good. Refer to Figure 5-1. Adjust the test range to more than 100.0 mm,
sound velocity 3250m/s. Adjust gain appropriately. Move the probe to find the highest
reflected echo of R50 and R100 arc surface. Enter the CAL main menu, select
"speed/delay" calibration, S-REF 1 >: 50.0mm; S-REF 2 >: 100.0mm.
(2) Move Gate A to capture the echo of R50 arc (refer to Figure 5-4). Enter the CAL
submenu and press [+] key to confirm, then echo1 is recorded.
27
(3) Move Gate A to capture the echo of R100 arc (refer to Figure 5-5). Enter the CAL
submenu and press [+] key to confirm, then echo 2 is recorded. Meantime, the actual
velocity is calculated and calibrated "speed/delay" is set automatically. After that, the
calibration is finished.
(1) Put the angle probe on C position of the CSK-ⅠA test block, and make sure the
coupling is good. Refer to Figure 5-1.
(2) Enter the CAL main menu, choose: CAL SET: angle; HOLE DEPTH: 30.0mm;
APERTURE: 50.0mm. Adjust gain appropriately and move the probe to find the
highest echo of ¢50.0mm hole.
(3) Move Gate A to capture the highest reflected echo of ¢50mm hole (refer to Figure 5-6).
Enter the CAL submenu and press [+] key to finish calibration, and then the result will
28
Figure 5-6 Move Gate A to Capture the Highest Reflected Echo of ¢50.0mm Hole
(1) select the "DAC" main menu to enter the "DAC mode record" submenu.
Figure 5-7 Schematic diagram of connected probes coupling on CSK-IIIA
29
(2) As shown in Figure 5-7; move probe to get highest echo ofΦ1×6 hole in depth 10mm,
amplitude about 80% (if necessary, use the main menu shortcut feature "AUTO
GAIN").
(3) Adjust the "GATE A Starting" catch the ECHO, press 1 time 《》key for confirmation in
“echo” submenu, then digital No in “echo” increased from 0 to 1, appears on the
screen the 1th DAC curve, as shown in Figure 5-8.
(4) According to the same method to find out-deep 20/30/40/50/... Short transverse holes
of maximum echo and identify them, the screen displays the full set of DAC curve, as
shown in Figure 5-9.
Figure 5-8 The 1st DAC curve interface Figure 5-9 Full set of DAC curve interface
(5) After finish drawing of DAC curve, press ※ key and then press the PgUp key, the
completed DAC curve will be store orderly in the smallest No of “Store” menu (screen
display the reality store ID No.); at this point, the DAC curves do not display, you can
start to make a new DAC curve.
Note: Because of the hole on CSK-IIIA be arranged in two side by odd and even numbers
of the depth, for easy application, the instrument have automatic recognition
function to the echo path length and therefore does not require a full point-by-point
in order to drawing the (that is, you can stagger) DAC curve. Such as you can make
DAC curve for hole in 10mm depth and 50mm depth at first, then make the holes in
30mm and 70mm depth, last make the holes in 20mm and 40mm depth, the
instrument will automatically combined into a single completed set of DAC curve
while all the points completed..
30
to different standards to a curved space is necessary to make adjustment, can enter the
"EL >" submenu to modify one by one.
Note: Because of Generate and SL is 0.0dB default, the both line overlap and appear as a
yellow which is equivalent to " GENERAT HIDE " -three-line status.
31
③ Set the DAC evaluation line alarming and a variety of "measured value display"
project, see Figure 5.9.
④ Inspection results and DAC curve stored to facilitate diagnosis and inspection reports
to print.
(2) Considerations
Pay attention to fellow points in using DAC curve:
① DAC curves are made with a probe on the Reference test block, it only valid for the
probe in general. In view of the difference of probe performance, even with similar
probes, should first be authenticated (particular within the probe near-field).
② The angle probe will abrasion after several times using, its angle and performance
may changed. In compliance with the requirements of the completed curve, it should
be trimmed before using the DAC curve.
③ Most surface condition of work piece on-site won’t be smooth than reference test block,
so must go on with gain compensation while loading the DAC curve (or called as gain
amendment), the specific compensation methods see 5.8.1.
5.7.1 DATA ID
The instrument have 500 data ID which can store 500 group of screen graphics and
associated parameters; for apply purpose of end user, this inside data ID divided into
1~50 and 51~500 two groups, the store contents of each data ID are same and can be
loading for apply.
(2) Use the main menu HOTKEY store orderly (storage of non-freeze)
After entering the main menu HOTKEY, select "SAVE" submenu, press 《》or《+》 key
32
and screen graphics and instrument associated parameters will store in the minimum
empty data ID from 51 to 500 orderly (convenient for instrument standard and fast storage
in on-site flaw detection.).
5.7.2 Remarks
The “Remarks” menu can enter 19 digital numbers or capital letters for simple introduction
of save contents, such as standard No., user or weld thickness, easy for memory and
loading choice.
Method of writing digital numbers and letters: enter the “REMARKS” menu, the first
position in the left side will flashing with “1”, then holding the 《+》key, the numbers will
increase from 0 to 9,then return to A~Z with same method; after confirmed the first
position, press menu related 《》key to move the cursor right shift, the first position be
saved, then writing the next accordingly.
5.7.3 Recall
In the menu of “DATA ID”, there is “recall/gain amend” submenu, press related “”key for
switch. Select the data ID in “data ID” menu, then select “Recall” submenu, press one time
“+” key for confirm, the screen will freeze display the echo wave and related instrument
setting of selected data ID, used for flaw detection after unfreeze. In application of loading
stored DAC curve or inspection standard, if need for calibrate inspection sensitivity(such
as surface compensation),enter “Gain Amend” submenu for adjustment; the dB values of
Gain amend will sorely stored in submenu of “Gain Amend” and display in “gain display
area” with additional gain; beside, if the loading data with DAC curve, the gain amend
value won’t change the previous DAC position (means echo amplitude changes, DAC
stay in same).
33
5.7.4 Record
For recording, playback, saving dynamic echo-graphic, convenient for analysis or
diagnosis of the defects. Select "RECORD" menu page, there is: "RECORD",
"STOP/BEGIN" ,"U DISK SAVE", "REPLAY" four submenu, functions as follows:
(1) “RECORD”: this menu is used to adjust the recording speed; the instrument input
speed is 5~15fps, default is 8fps.
(2) STOP/BEGIN: control entry of "BEGIN" and “STOP", default is "STOP"; select this
menu and press + key to start with recording; the recording will continue even quit
from STOP/BEGIN menu(turn to slight blue background and letter in white), stop only
in BEGIN mode or storage space out of capacity.
(3) U DISK SAVE: save the recording contents to U disk, select this menu and plug with U
disk, press + key, the saved recording contents will save to U disk, there is prompt in
screen once finished or abnormal in U disk.
(4) REPLAY: controlling replay of " BEGIN" and "STOP", defaulting to "BEGIN";after stop
recording, select the "REPLAY" menu and press + key, the screen continuous replay
the recording dynamic echo graphics; freeze display its wave picture when replay is
over, at this point, press ※ (freeze) key to exit .
Note: Due to "replay" primarily for dynamic echo records when testing, data consumption
is quite large, so if you want to save, be sure resave to U disk, otherwise the
instrument will delete all the testing data after shutdown!
34
are shown in the upper left corner of the screen "gain display area" such as
"36.0+0.0dB", "36.0dB".
② Under main menu HOTKEY, the "Auto Gain" is actually "Auto Gain =80%" quick
switch menu, selected and press one time its related key, the screen will display
“Auto 80%” (disappear after 3 second), the echo amplitude in Gate A will auto adjust
to 80%
③ Application instance: the instrument equipped with probe 2.5P10x10A60 ,after getting
the highest amplitude echo of Φ1×6 hole in depth 30mm on CSK-IIIA test block,
adjust Gate A to capture the echo, enter “Auto Gain” menu to set at 40% from “off”
mode, then whatever its previous amplitude is, the “Gain” will auto adjust to make it
40%(more convenient for laboratory instrument sensitivity calibration). Similarly, if
want to set “Auto Gain” at 80%,enter main menu HOTKEY, press one time “Auto Gain”
related key, operation more convenient.
Note: When "gain step =0.0", the "Auto Gain" cannot adjust to avoid misoperation.
5.8.2 Envelope
The “Envelope” function used for observe echo amplitude changes in probe performance
testing and measuring flaw size, display the tracks of dynamic echo peak completely on
the screen. Set the “Envelope” on, the echo envelope color change to purple, much
clearly.
35
When using, enter “TEST SETUP” menu, select “Surf Amend”, the menu display “flat” in
default, it means “Surf Amend” in “off” mode, press one time “-“ key, the menu display
Φ2000mm,hold the “-“ key, surface diameter will decrease little by little until to Φ100.0mm
The related key in “Surf Amend” for step setting, press “+” key for number increasing.
Note: (a) When the "probe specifications" (main menu PROBE) is set to normal probe N,
the “Surf Amend” fixed to “flat”, unchangeable.
(b) Particularly pay attention to pipe wall thickness/diameter ratio(t/d value) while
process with circumference flaw detection for pipes, according to related
standard, when t/d ≤ 0.2 should take other effective methods of testing defects
to avoid a miscarriage of justice.
Enter the main menu “Flaw Setup” ,select "Weld Setting" ,there is “Weld Set”, “Weld
Width”, “P-Position, “Refresh” four submenus , you can set weld parameters and probe
position for marking defects separately, or refresh weld graphics and probe position
together and display the next defects, very simple and clear.
Note: (a) “P-Position” is distance from probe Front to weld center.
(b) When the "probe specifications" (main menu PROBE) is set to normal probe N,
"weld illustrations" can’t be entered.
5.9 Measurement
Main menu “Measure” contain “TOF”, “PEAK HOLDING”, “ECHO COMPARE”,
“READINGS”, “FFT” five submenu, their characteristics as follow:
5.9.1 TOF
The measurement accuracy of digital ultrasonic flaw detector associated with selection of
measure point in using Gate, so correct choice before measurement is needed to make.
Followed with explanation of “peak” and “flank” in “TOF” menu:
• Peak: means the measure point is the highest echo peak point in the gate;
36
• Flank: means the measure point is the cross point of first echo within the gate and
gate.
Note: The location of the measure point, marked on the gate.
In General, use “peak” in test echo amplitude, such as echo amplitude Ha%, making DAC
curve etc, use ”flank” in test echo distance, such as sound path of defects inspection with
angle probe, auto velocity calibration. The instrument have set its related default value for
those need to chose peak or flank in testing, check in the first time using.
Note: Default value of the instrument set as “peak”, in thickness testing with echo in the
gate, firstly set “TOF” as “flank”, it’s really important for surplus wall thickness of corroded
pipelines and thin workpieces inspection.
Active the” PEAK HOLDING” function, peak of dynamic wave and its position will display
on the screen marked as “×”,refreshing once there is new peak appears, it’s the normal
way to confirm echo peak in probe performance testing and on-site inspection scanning.
Application instance: press 《※》 key to freeze the selected defect echo, then select
“ECHO COMPAR” menu( under “Measure” main menu) and press one time “+” key ,then
find other direction echo for the same flaw, the screen will display all details of these two
selected echo with two different color(include amplitude, position, rise and declined trends
of front and back edge),the experienced inspector can distinguished some characteristics
of defects.
5.9.4 Readings
As long as improvement of electronic technology, the ultrasonic digital flaw detector can
get more and more digital information from echoes. For convenient display and
observation, besides fixed Sa/Ha% display of straight probes and Sa/Da/Pa display of
angle probes, there is “READINGS” menu for user choice.
37
See “READINGS” in table 1 of "Appendix one" Selected items will be displayed in large
print in the top right corner of the screen.
5.9.5 FFT
Wave length of ultrasonic wave propagation in workpiece is one of most important
elements of equivalent effect defect judgment, so the probes echo frequency accuracy is
very important, echo frequency error of standards at domestic and abroad will be limited
to within ± 10%. Anyscan-36 using FFT technique set "FFT" menu, can easily measure
and display probe signal -6dB bandwidth and center frequency “fo”. Followed with
example of using normal probe 2.5P20N to explain its using method:
(1) Enter “probe” menu, select “TYPE-SPEC” and set probe type as “N2.5P20”.
(2) Connect the probe and coupling on 25mm depth of CSK-IA test block, get a moderate
and stable echo for observation.
(3) Select "FFT" (under “Measure” menu) and press one time “+”key, the screen display
with red letter for measure tips, the finished FFT curve convenient for observation.
Note: The required default value of straight/angle probe have been preset in
“TYPE-SPEC” while enter “FFT” menu for echo spectrum testing(see
“Measure”/”FFT” menu in section 4.2.9).
5.10.1 Filled
Anyscan-36 use with new high density, high brightness, high contrast TFT panel for
display, although under normal light to have enough brightness and clarity of the echoes
of the details appear to be exhaustive, but in contrast strong sunlight also slightly lacking,
thus adding "FILLED" feature. While active “FILLED” menu, the hollow echo will filled into
solid, which can greatly increase the contrast in such a case, selecting "SCHEME"
configuration, the result will be better.
5.10.2 Scheme
The instrument set six kinds of scheme configuration for choice based on characteristics
of high quality TFT panel, normally chose 1 or 2 under normal lighting(default with
38
5.10.3 Brightness
The backlighting of the instruments in "high/medium/low" third gear (default in "medium"),
convenience for different brightness selection. Normally chose “medium” under indoor
environment, chose another under strong light or darkness environment.
5.10.4 Freeze
“Freeze” means the screen display the echo graphic in stationary, it usually used in
instrument setting and on-site inspection, press “※” key in the panel of the instrument to
save the defect echo wave on the screen.
While loading the DAC curve or setting parameters, the screen will display its “freeze”
status, press one time “unfreeze” key to active it, so, the “unfreeze” is more important
application way of “Freeze”.
5.10.6 Buzzer
“Buzzer” menu of the instrument is switch of inside buzzer, default set “off”, set “on” for
continuous alarm of Gate.
39
(1) Select “DATE” menu and press one time its related key, the yellow cursor will display
the first group number, press “+” or “-” key to adjust.
(2) After confirm the first group number, press one time its related key move the cursor
to the right, then set the other group numbers accordingly(month/date).
(3) After finished the last numbers setting, press one time key, the cursor disappeared,
date setting finished.
5.10.10 Preview
Menu function of the UFD (ultrasonic flaw detector) become more and more sophisticated
in recent years, of which bring inconvenience on setting, check and modification.
Therefore, new function menu such as “preview” and “help” was designed for Anyscan-36.
Main menu and other items will be shown in the screen when press “PREVIEW”, making
complex menu items easier to see. Moreover, modification for any item in the menu is
available, which makes management and application of the instrument more convenient.
5.10.11 Help
Main function and operation methods are listed in the “HELP” menu, which enables
primary UT inspectors to use this product correctly free of “User Manual”. Open “HELP”
menu, table of contents show firstly, press to loop, selected item is yellow, press
to check the contents of the item through “PgUp”,”PgDn”, press to return to “Table of
content”.
40
“Software update” and press “+” once, “Updating” will be shown in the screen in red, after
vanish, withdraw the U-disk, reboot the instrument and run the updated software.
5.11.1 Summarize
In ultrasonic flaw detection, the shape, nature and direction of natural flaws are different
from each other, flaws with same echoes amplitude actually differ greatly, thus introduce
"equivalent size" to evaluate flaws size. Under same detect condition, if the echo
amplitude of natural flaw equal to that of regularly shaped artificial flaw, the artificial flaw
size is defined as the natural flaw equivalent size.
The curve used for describing relations among distance, amplitude and equivalent size of
regular reflector is called distance-amplitude-equivalent size curve, in German called AVG,
in English called DGS.
(3) Make the probe tightly coupled on the CS-1-5 block, move probe to seek the highest
echo of flat bottom hole, refer to Figure 5-10;
41
225mm
25mm
Figure 5-10 The Making Diagram of AVG Figure 5-11 AVG Waveform Display
(4) Enter AVG menu, adjust aperture to 2mm, equivalent of reference curve can be set
according to requirement (here take 1mm for example). Move gate A to capture flat
bottom hole echo, select POLT submenu, then press [+] key to confirm AVG making.
Refer to Figure 5-11.
5.11.3 Notes
Before Making AVG , you should pay attention to whether probe frequency and crystal
size is suitable, whether the probe parameter settings agree with the actual and reference
curve range is "0.5~23.5mm,∞" or not.
The attenuation coefficient between test block and work piece is not considered when
using the AVG.
42
This chapter shows specific application of weld and forging piece of Anyscan-36 briefly, to
enable inspectors to be in full use of this instrument, to find out the flaws precisely and
make do with flaw data and report rapidly.
(1)Prepare specific probe, make or detect DAC curve, to see whether it meets the
requirement;
(2)To know about the features of the test piece and its surface status, calculate and set
43
(2) Connect the probe, place it on CSK-IA, test the point of incidence, X-value, and
calibrate delay and angle of it (Refer to 5.5), check whether relative parameters
that being entered to corresponding submenu correctly.
(3) Choose TEST SETUP in the main menu and enter submenu of “Weld set”, select
“V” in “Weld set”, enter “10” in “Weld width”.
(4) Choose BASIC in the main menu, set “Range” as “100”, “Thickness” as “20”,
check whether “Velocity” is as default “3230”; then check submenus of PULSVR
in the main menu, whether as default set (See 4.2.2).
(5) Confirm the gain amend value which is based on surface status of the test piece,
to reserve enough allowance during the process of calibrating the sensitivity of
flaw detection.
44
amend.
③ Means of use “Auto-gain” and “GAIN AMEND” in main menu HOTKEY: entangle
the peak ofΦ1×6 hole, enter HOTKEY and select “Auto-gain” submenu, echo
wave amplitude is being adjusted to 80% automatically(Gain is shown as
“40.0+0.0dB”), If 6dB gain amend is needed for field inspection, enter “Gain
amend” in “TEST SETUP”, set as “6.0dB”, 40.0+6.0dB would be shown on right
top corner of the screen, 40.0 is “Initial sensitivity”, +6.0 is added as “Additional
gain” for gain amend.
(a)Check “Preview” fully after above settings finished, save flaw detection interface,
Note:
(b)Make DAC curve as 5.6 shows, check echo wave frequency of the probe as
5.10, if necessary.
(2) When peak is being discovered, choose “Gate A”(GATE main menu), press “+”
again, gate A would entangle the echo wave automatically. Sa/Da/Pa value of the
flaw would be shown on right of the screen; Select “Ha%” at “Readings” to show
echo wave amplitude of the flaw.
(3) Measure the distance between probe front edge and weld center, enter it to
“Probe position” (Main menu TEST SETUP), weld image down the screen
indicates the flaw position.
(4) Press “※” (Freeze), then press PgUp once to save screen graphics and current
instrument parameters orderly, set “Refresh” (Main menu TEST SETUP) as “ON”
to clear previous flaw position, easy to be used for the next flaw.
(5) Transfer above saved data to PC via U-disk or cable, to print flaw detection
reports.
45
Application of flaw detection: Choose DC2.5P20N with membrane probe to test Φ800
forging piece according to the requirement of flaw detection.
(1) Go to “TYPE SPEC” (Main menu Probe) and check whether default is N2.5P20.
(2) Select main menu BASIC, set “RANGE” as 850mm, check whether velocity is as
default value “5920”.
(3) Choose “PLSRCVR”, check whether it is with default set(see 4.2.2). Usually,
mid-size forging piece should be set as: Pulse width of square wave=200,
damping: 200, if necessary, make DAC curve as 5.6, check echo frequency as
5.10.
Note: Better to open “Preview” list and check them all and then save after completing
(5) Confirm the compensation value (Gain amend) according to surface status of the
test piece, curvature and scan requirement during the process of field inspection.
If compensation is 18dB, go to “Gain amend” (Main menu TEST SETUP).
(6) If single or multiple fixed or mobile unknown reflected wave appear during the
process of scanning, to lower down repetitive frequency to judge (Go to main
menu PULSVR for “PRF”); To judge the trend and length of flaws of non-metallic
inclusions, select “Peak holding” (Main menu MEASURE) or go to “Envelope”
(Main menu TEST SETUP) submenu, to determine echo peak and -6dB(or -12dB)
scanning length.
(7) After scanning, save echo graphics of each flaw and instrument data, if necessary,
transfer the data to PC to print flaw detection reports.
46
mechanical parts flaw detection besides above weld and forging. This chapter takes
thickness 160mm square aluminum as an example to introduce the application features of
ultrasonic test.
(1) Instrument set: Adopt 5P10N narrow pulse probe to detect, Set “PLSRCVR” menu
as below(Others are set as conventional)
① Square wave transmit:Pulse width 100ns;
② Damping:50Ω;
③ Frequency band:0.5~22MHz.
(2) Judging difficulty: When set “Range” as 170mm, bottom resolution of the
instrument could distinguish 5mm hole clearly, but near surface 5mmΦ1.2 is stuck
with initial wave, which makes it difficult to distinguish whether near flaw exists,
detect its depth, of which make it hard to judge the flaws.
(3) Solution: If initial wave becomes wider during the process of scanning, use “A
magnify” to observe, detailed operation methods are as below:
①After set of “Range”, open gate A and set as: Start=3mm, width=10mm,
threshold=35%.
②Select “A magnify” (Main menu HOTKEY), press “+”,enter expandable status,
whether 5mm Φ1.2 hole and initial wave are totally separated or not and the
beam path Sa value would be shown in the screen.
③Select “Save” (Main menu HOTKEY),press “+” to save and check, after
completion, set “A magnify” as OFF.
Note:Means of flaw detection for die steel(before opening mould) is similar to above, be
47
“Frequency spectrum” function was specially designed to facilitate users to check echo
wave frequency of Anyscan-36, which can be used to test feature curve of frequency
spectrum with probe and instrument.
Besides, current conventional narrow band probes like 2.5P20,5P14,etc, all of them are
equipped with resonance coil, therefore 2m cable(Material: SYV-50-2-1) is required, or
resonance point of the probe may shift. For some cases, if cable is required to be more
than 3m for some occasion, to avoid waveform distortion of the echo wave, change coil
inside the probe or take some other compensation means seems to be a must.
Therefore, users who use 10MHz or above to perform high resolution tests should
carefully check the instrument during purchasing and using, if necessary, hand it to
authoritative organization to be identified, to avoid abnormality like, sensitivity and
resolution of ultrasonic test from imported high frequency narrow pulse probes are not
as good as low frequency ones.
Besides, Compared with sharp pulse of the same amplitude, square wave transmission
could increase sensitivity up to 4~6dB, but, it is suitable for the excitation of rated 2~5MHz
probe only; If, frequency above 10MHz, especially for high frequency narrow pulse probe,
it is useless for square wave transmission, other problems may occur by wrong settings,
users should be alert of it.
Note:They are not the same concept between amplifying 86dB and “Gain” regulating
variability 110dB.
48
Test
High frequency narrow pulse probes, like 10C6N (or 10P6N), 15P6Y, are widely used for
high resolution test of aircraft engines and auto parts due to that they have only 1.5~2 ring
circle, and the negative echo wave amplitude is high enough, for example: Paste quality
check of composite material (overall thickness 0.8mm), auto detect for spot welding and
small caliber thin pipe, etc.
High frequency narrow pulse probe is also high damping probe, it adopts special crystal
and being made by complicated process, relative bandwidth of its frequency spectrum of
echo wave -6dB is above 60%,for high level, it is even up to 80%~100%(Some probes
with composite material may be even higher ).Therefore, use probes correctly and ensure
its wide frequency band could be displayed in high fidelity, extreme high requirement for
the matching instrument is listed below:
(1) Rising time of pulser front edge tr should be small enough, usually requires
tr≤10ns, otherwise, sensitivity test may be effected.
(2) Working frequency range for receiver amplifier should be wide enough, usually
requires -6dB width ≥15MHz (Better ≥20MHz), of which can assure high fidelity
transmission of probe echo wave(Say, display 1.5~2 ring circle after amplification,
waveform would not be distorted), to obtain high resolution.
(b)In the concept of “high damping” for the probe, it requires not only enough high
for the “electric damping”, but also includes that nice coupling “sound
damping” should be good enough. In other words: the high resolution result
would be low for the narrow pulse probe if coupling is not good, special
attention should be paid for field use.
49
equipped with the instrument: “PEAK” and “FLANK”, set it as default set to avoid trouble
for field operation, but for the first time to use this instrument, check it and know it better
should be performed.
Usually, below two principles should be followed for the choosing of measuring points:
(1) Adopt “PEAK” if measurement is related with amplitude of echo wave.
(2) Adopt “FLANK” if measurement if related with distance of echo wave.
Note:(a)“FLANK” ensuring points refer to the intersection between the front edge of the
first echo wave and gate, it relates to the “height” of the gate. Strictly speaking,
when measure times for the distance between two echo waves by adopting two
gates A/B, be sure that the “measuring points” are the same “sound point”,
which means, the height of gate A/B is not in the same ground level contour.
(b)When measure surplus wall thickness of the chemical pipes, adopt “FLANK” is a
must; If set as “PEAK” by mistake, as front edge of echo wave is too slanted
and variable, major deviation display, which may cause misjudge.
Anyscan-36
There are overall 3 types of A/B ports in the instrument, which could communicate with PC
and printer to print flaw detection report. Connect the instrument with PC, and use the
50
7.1 Port
Definition of port:
PIN-NO Description Signal direction Level
1 USB- USB Low Signal USB
Restart the instrument once abnormality occurs.
(2) Open “Data View”, double click “setup.exe”, complete installation, an icon of “Data
View” will be seen in the desktop.
(3) Start the instrument and connect it with PC: With a specific USB cable to connect
both.
(4) Double click “Data View” on the desktop to start this software.
(5) The instrument could transport saved pictures and data to PC via USB. Software
on PC could be used to edit, store or print
Note: If work with PC software, after starting the instrument, connect it to the PC and
51
(5) "Video" menu, refer to Figure 7‐5.
52
(3) Config
It displays configuration values corresponding to current data ID.
(4) Gate A
It displays start point, width and threshold of Gate A.
(5) Gate B
It displays start point, width and threshold of Gate B.
53
After input, click "OK" button to confirm, then the information will be saved to the file; if you
close the window before confirming, it will not be saved.
(7) Unit
Select the unit and change the display unit.
(8) SCALE
Select the display mode of the baseline.
(1) Through "DATA COPY” submenu (main menu "CONFIG "), copy instrument
parameters configuration to U disk, and then view them via online software.
(2) In "U DISK SAVE" submenu (main menu "STORE"), transfer the recording dynamic
echo in the “replay” to U disk, and then view them via online software.
54
(3) After connecting to PC with matched USB cable, detection information interface
opened by software can be part of test report.
Note : (a) Implement the " back to the factory state " , the instrument will revert to the
initial settings to the factory , all data and modifications stored after factory will
be removed, so it should be used with caution !
(b) The instrument can automatically perform a self-test after switching on, and
then enter the normal operating state. If the self-test is not normal, should shut
down first and then restart, If still cannot pass self-test (software crashes), then
use the “default settings” to re- update the system to return to normal .
This chapter focuses on the safe operation prompts, and expect users to use properly
instrument to avoid safety hazards.
55
Anyscan-36 can work safely in more harsh field or high overhead environment , especially
expand the working environment temperature at -10 ~ +50 ℃, and has waterproof of IP54
standard level , so no adverse effects caused on the test results even work in the summer
and winter or micro rain weather.
(2) Battery-powered
Anyscan 36 specially use (10.8V 5200mAh) lithium battery-powered. A fully charged
battery can work around 6h.
Pressing the panel "Power" button for about 2s and turn on the instrument, and enter
detection working screen after displaying boot interface, menu settings is the one before
normal shutdown last time.
Instrument is in working condition , press the panel " power " key for about 2s and will shut
off the power supply , it’s a normal shutdown state , the original parameters will be
recognized and saved . If the instrument suddenly loses power (battery depletion or AC
adapter disconnected), it’s a non- normal shutdown, the parameters set before the power
failure may not been saved, so should be checked.
The instrument features screen battery level indicator and low voltage alarm shutdown
function, favorable to remind users to timely charge.
Note: CD-30A of the instrument is the preferred high-capacity dedicated adapter; it can
simultaneously change battery and power to the instrument, but do not mix with other
instruments or batteries.
The optimum use temperature Lithium batteries is about 25 ℃, the temperature is too high
56
or too low will affect the effect of the charge and discharge, so should pay attention to the
efficiency of the battery when the ambient temperature deviation is too large.
Batteries generally have a requirements of storage period, even if a fully charged battery
doesn’t use for too long time, it should recharge again before use in order to ensure a
sufficient supply time.
Lithium batteries with electrical storage for too long time and easily cause internal damage
so should be charged and discharged as soon as possible after a period of inactivity.
57
The standard echo probe of Anyscan-36 named as PE-50 made off with special material
and technology, the center frequency is 5MHZ. The most important feature of PE-50
probe is the distance between the echo amplitude and multiple echoes can be saved in a
long time ( several years ) , stable and unchanging , therefore it can be calibration tools for
instrument detection sensitivity , vertical linearity , horizontal linearity and detection
range ,simple and convenient.
For example:In check the acceptance of instrument, recording the gain value of first or
second echo B1 or B2 amplitude at 80%,check every time while connect with echo probe,
58
the little change of sensitivity can be found out. In addition, save the echo wave of PE-50
in check the acceptance, the above inspection is more convenient.
59
interference; good grounding device should be installed to strengthen the shield facilities
when necessary, which can effectively prevent the interference signal string into the
instrument and ensure that the work of instrument is stable and reliable.
The instruments should be covered by dust cover are use. In Maidu season, should
always energize the instrument 1 to 2 hours to disperse its internal moisture.
If it is found that the instrument works abnormal , you should first read the user manual to
check if the operation method is correct, and can also consulted to the experienced
personnel or factory , but do not disassemble the instrument to overhaul , to avoid
unnecessary trouble .
If the user have questions or need help during use, please write to us or call us, we are
committed to reply you as quickly as possible.
Doppler provides one - year warranty. When customers comply with the custody
installation and use rules of the user manual, then the warranty time is from the shipping
date from factory, if it is found that manufacturing quality is poor or work abnormal ,
Doppler is responsible for free repair or replacement if the product seal is integrated . After
one year warranty expires, Doppler still continues to provide maintenance services.
Must use the original box and packaging liner to pack the instrument if need to return for
faulty repair, to minimize the possibility of damage in transit, in order to ensure the
instrument can work normal after repair. For this reason, we strongly required not to use
substandard shipping container. If use unqualified box and caused the quality problems
for the repaired instrument, Doppler takes no responsibility. Therefore, after open and use
the newly purchased instruments, plus take good care of the original box and packaging
liner for facilitating future use.
When the instrument returns to factory for repair, plus list your shipping unit Name,
department, Correspondence, contacts, contact phone and delivery stations in the box,
preferably contact us by phone or e-mail if necessary , print out the above information and
60
fault conditions and requirements, and put into the box for our rapid processing .
The company's products, starting from the date of purchase by the user, quality failure
(non- warranty are excluded) occurred within one year since the date of purchase, please
contact the Doppler Customer Service Department with the “warranty card” (User retained
joint) to repair the instrument. If can’t present the warranty card within warranty period,
Doppler calculated warranty period according to the date of departure from factory and the
warranty period is one year .
APPENDIX
Measured
Definition
values
Sa Sound path of first echo inside Gate A
61
Sb-a Sound path difference between first echo inside Gate A and Gate B
62