Century CC Bed
for support when getting out of bed and could be injured if the
bed unexpectedly moves. After setting the brakes, push and added precaution.
pull the bed sideways to insure stability. INSTANT CPR RELEASE ®
FLUID SPILLS The emergency head and knee release is to be used by
When massive spills occur in the Siderail area or foot end of the healthcare professionals only.The mode selector lever must be
bed, immediately: continually held down in the CPR position until the bed reaches a
flat position.This will insure a smooth operation and avoid delay.
• Clean the fluid from the bed.
• Check the bed controls (i.e. Head, Knee, Hi-Lo). CONVERTIBLE FOOTBOARD
• Have the maintenance department check the internal When auxiliary equipment is placed on the footboard, secure the
electronics. equipment with the straps provided.The weight limit of the
footboard in the transport shelf position is 45 lbs.Always return the
Fluids remaining on the electronic controls may cause corrosion,
footboard to its original position after use. Check to insure proper
which may cause the electronic components to fail. These
latching by tugging on the footboard.
component failures may cause the bed to move or operate on
its own at a time that may be injurious to the patient or staff. FOOT EXTENDER
When using the extender, always give a tug on the extender to
LOCKOUT CONTROLS
insure locking. Do not place any heavy weight on the foot
Whenever a patient should be restricted from operating the
extender and keep in mind the additional length of the bed.
patient controls, activate the appropriate lockout controls
located in the nurse control panel at the foot end of the bed. NURSE CALL SYSTEM INTERFACE
This is especially true for patients in traction since it can prevent If proper interface between the bed and the nurse station is
a patient or visitor from accidentally activating a control which provided, the scale can be set to place a REMOTE call to the
would cause the bed to move and possibly result in an injury. To nurse station. With interface, the call for a scale monitoring
insure the patient’s safety when unattended, place the bed in condition is momentary and repeats until turned off at the bed
its lowest position and lockout the Hi-Lo control. Also, the mode scale. A call indicating a low battery is repeated every ten
selector dial should not be in the Hi-Lo position. minutes until the bed is plugged in. Without Nurse Call
interface, no call will occur in REMOTE. Consult the Nurse Call
SIDERAILS/RESTRAINTS/PATIENT MONITORING
manual to verify compatibility.
The Siderails should always be in a full upright position and
latched when a patient is unattended. When raising the BED POSITION CHANGES
Siderails, an audible “click” should indicate that the Siderails are Be certain that feet and hands are well clear of the lift arm and
completely raised and locked in place. frame assemblies of the bed when changing bed positions,
both manually or electrically.
Hill-Rom recognizes that certain healthcare situations may
indicate the need for specialized Siderail configurations. In MATTRESSES
response to this need, we offer, upon request, several Siderail The use of mattresses that are not sold by Hill-Rom may reduce
accessories. the effectiveness of the safety features and systems
incorporated into Hill-Rom beds.
Siderails are intended to be a reminder, not a patient restraining
device. Hill-Rom recommends the appropriate medical X-RAY CASSETTE HOLDER
personnel determine the level of restraint necessary to ensure a If using the x-ray cassette holder accessory, always secure the
patient will remain safely in bed. Consult the restraint manufac- holder to the bed when not in use. This will avoid possible
turer’s instructions for use to verify the correct damage to the bed and holder.
application of each restraining device. PREVENTIVE MAINTENANCE
Whenever “high profile” patients (typically, the frail, elderly and Annual preventive maintenance must be performed to insure all
medicated or confused) are involved, Hill-Rom recommends bed features are functioning as originally designed.
the following minimum actions: Particular attention must be addressed on safety features,
including but not limited to:
1. Develop guidelines for all high profile patients that indicate:
• Siderail latching mechanisms
• Which patients may need to be restrained and the
• Caster braking systems
appropriate restraint to utilize.
• Frayed electrical cords and components
• The proper method to monitor a patient, whether
• All controls return to off or neutral position when released.
restrained or not, including time interval, visual check of
• Controls or cabling entanglement of bed mechanisms in
restraint, etc.
Siderails.
2. Develop training programs for all caregivers concerning the
• Proper operation of the lockout function controls.
proper use and application of restraints.
3. Maintain the bed at its lowest position whenever a caregiver
is not in the room.
4. Clarify the need for restraint devices to families or guardians.
1
TABLE OF CONTENTS FEATURES
FEATURES...................................................................................................1
DIMENSIONS.............................................................................................1 SideCom ® Siderail Patient
Siderail Communtication System Controls
BED CONTROLS Nurse Controls (Optional) (Both Sides) Convertible
(Both Sides)
Hi-Lo.......................................................................................................2 Footboard
Head With Automatic Contour .........................................................2 Restraint Strap Holders;
Head, Thigh, Foot End DynamicAire™
Head Only ............................................................................................2 (Both Sides) Sleep Surface
(Optional)
Knee......................................................................................................3
Removable
Trendelenburg/Reverse Trendelenburg ...........................................3 Headboard
Vascular ................................................................................................3 Patient Scale
(Optional)
Cardiac Chair......................................................................................3
IV Pole Sockets
Electric Control Lockouts ...................................................................3 (One At
FEATURES Each Corner)
Trendelenburg/
Siderails .................................................................................................4 Reverse Trendelenburg
Controls
Central Brake and Steer.....................................................................4 Permanent
Permanent IV Pole ..............................................................................4 IV Pole
Instant CPR ..........................................................................................5 Electric Controls
Headboard Removal .........................................................................5 Lockout Panel
Convertible Footboard ......................................................................5
Drainage Bag Hooks Integrated Air Support
C-arm Interfacing................................................................................6 (Both Sides System Controls
Other Features .....................................................................................6 And Foot End) (Optional)
OPTIONS
Head Elevation
Integrated Air Support System...........................................................7 Indicator Motor Power
(Both Sides) Switch
SideCom® Communication System..................................................7
IV Pole
Low Leakage .......................................................................................7 Storage Area Pump/Release Pedals Central Brake And Steer
Scale .................................................................................................8–9 For Manual Control (Both Ends)
(Both Sides)
ACCESSORIES Manual Control
For Bed Positioning 8" Polyurethane Carpet Casters
O2 Tank Holder......................................................................................9 Including Instant CPR (6" Polyurethane
X-Ray Cassette Holder................................................................10–11 Carpet Casters Optional)
TuckAway
ISS System ...........................................................................................12 Siderails
Transducer Holder .............................................................................12
IV Pole .................................................................................................12 DIMENSIONS
Trapeze Support Bracket..................................................................12
Cane Bumpers...................................................................................12 Measurements: Inches (centimeters)
Siderail Pads .......................................................................................12
Chartholder .......................................................................................12
91.0 (231)
CLEANING/REMOVAL OF DYNAMICAIRE™ SLEEP SURFACE .............13 93.5 With Roller Bumpers (243) 38.5 (98)
SAFETY TIPS.............................................................................Back Cover 34.0 (86)
15.0
NOTE An In-Service videotape is available. Contact your (38) 8" Casters
Hill-Rom representative or call 1-800-445-3720. 55.5 (140) High
37.5 (97) Low
6" Casters
54.3 (137) High
36.3 (94) Low
8" Casters
FOR SPECIAL ASSISTANCE:
40.5 (102) High/22.5 (57) Low
6" Casters
Hill-Rom Account Manager ________________________________________ Phone ______________________________
39.3 (99) High/21.3 (54) Low
8" Casters 6.5 (17)
Clinical Consultant ________________________________________________ Phone ______________________________ 6" Casters 4.6 (12)
Field Service Technician __________________________________________ Phone ______________________________ 57.0 (145)
23.5 (60)
Customer Service Representative __________________________________ Phone ______________________________
1
BED CONTROLS BED CONTROLS
Introduction: The Century CC Bed is designed to use electric and/or manual controls. KNEE For Electric: For Manual:
• Locate the knee control button on the inside of • Turn the mode selector dial to the “Knee”position…
ELECTRIC CONTROLS MANUAL CONTROLS either head end siderail… • To raise: pump the green pedal…
• To raise: push the G button then release… • To lower: push the orange pedal.
• To lower: push the H button then release.
6 7
FEATURES OPTIONS
C-ARM INTERFACING STEP ONE: BED SET-UP INTEGRATED AIR
Introduction: The radiolucent window of the Century CC Bed Prior To Positioning The C-arm: SUPPORT SYSTEM
is accessed from either side of the bed. It measures 21" x 38" • If attached, remove the X-ray cassette holder from the
without the scale and 20" x 37" with the scale. bed…(See page 10)
• Place the head end siderails in the intermediate position…
• Insure that the sleep surface is flat…
• Raise the bed to a medium high position.
NOTE Some smaller C-arms may result in reduced coverage
especially if used with a large patient. If this occurs, either
slide the mattress pad toward the head of the bed or adjust Prevention Mode Comfort Mode
the patient toward the bedside with the C-arm. To be used when it has been clinically determined that the To be used when it has been clinically determined that the
patient is at risk of developing pressure ulcers. mattress can be manually controlled for comfort purposes.
STEP TWO: C-ARM POSITIONING
After Bed Is Set Up: • Insure the air system power switch is turned ON… • Turn the mode switch to COMFORT…
• Move the C-arm to the bed…(For proper penetration, • Turn the mode switch to PREVENTION. The surface will • Press either the FIRM or SOFT comfort switch to adjust to
insure the X-ray tube is placed under the bed.) automatically adjust to patient movements. The comfort the desired comfort level.
• Using the Hi-Lo control on the outside of either head-end controls will not function when the PREVENTION switch is If the compressor runs excessively, the yellow AIR LOSS light
siderail, while looking under the surface, slowly lower the activated. will flash on the mattress control panel. To correct, turn the
bed to a comfortable working height… mattress power off, then on again to reset the system. If the
• Sweep the C-arm over the patient to check coverage… AIR LOSS light still flashes, call maintenance for assistance.
• In some cases, raising the siderail will increase penetration.
NOTE The integrated air support system upgrade kit
To Remove C-arm: operates in the same manner as described above.
• Raise the bed to high position using the Hi-Lo control on
the outside of the head-end siderail…
• Insure the siderails are in the intermediate position…
• Insure there is adequate clearance to remove the
C-arm from under the bed…
• Remove the C-arm from the bed.
6 7
OPTIONS OPTIONS
SCALE CHANGING ITEMS ON THE BED To Adjust Empty Bed Weight:
Each control button is color coded and graphically represented. A caption appears below each button describing its During the patient’s stay, For example, the bed has been Zeroed with the patient on
function. Additional operating instructions are located on the footboard. items may be added or the bed. If the empty bed weight is known,
removed from the bed. The you can adjust the empty bed weight to
ZERO DISPLAY button will account for obtain the patient weight.
The ZERO Scale display for the weight of the items • Touch button until a single beep
button resets the weight readings. without affecting the patient’s weight.You will need to sounds, then release…
scale to 000.0 prior account for the items each time they are added or • During the empty bed weight display, press
to placing a patient removed from the bed. and hold the appropriate or button
ADJUST WEIGHT
on the bed. • Touch to obtain most current patient weight… until the correct empty bed weight is
The ADJUST
WEIGHT buttons are • Within 10 seconds of the new patient weight display, displayed, then release…
BED EXIT touch the button until one beep sounds, then • The display will automatically change
used to manually
The BED release… back to the patient’s weight.
adjust patient or
button activates • Place or remove items on the bed, a plus or minus will flash
empty bed weight. ACTIVATING BED EXIT MONITORING
the monitoring for indicating the scale is waiting for further instructions…
patient egress. The bed exit feature will sound an alarm at the bed and, as
• Touch the until one beep sounds, then release… an option, at the nurse station when a patient attempts to
LBS/KGS WEIGH CHANGE ITEMS • The new empty bed weight will display for 10 seconds, leave the bed.
The LBS/KGS The button The CHANGE ITEMS then revert back to the patient’s weight.
button is used to weighs the patient. button maintains patient • Press and hold button until a single beep sounds…
MANUALLY ADJUSTING DISPLAYED WEIGHT • The display will show “BED EXIT,”“LOCAL,” and if available,
switch to lbs.or kgs. weight when items are added
The weight of the patient or empty bed can be manually “REMOTE” to indicate the monitor and alarms are
or removed from the bed.
overridden, if necessary by using the adjust buttons. activated. In addition, the yellow LED bed exit monitor
BEFORE OPERATING THE SCALE Weight Interferences Use these buttons if an incorrect weight needs to be located below the scale controls will illuminate…
For the most accuracy, follow these procedures: Insure that items such as pillows, linens, equipment on corrected or if the exact weight of an item is known. • To deactivate bed exit monitoring, press and hold the
Sleep Surface Preparation stands, etc. are not touching the head/foot board or
To Adjust Patient Weight: button until a single beep sounds and the
Before placing the patient on the bed, the sleep surface permanent IV pole. It is important that staff or visitors do not
For example, if a patient has had a cast put on, the added display clears…
must be prepared. This includes placing articles on the bed touch the siderails or sleep surface during the weighing
weight of the cast must be accounted for. To do this, you • To cancel the alarm, press any button.
such as pillows, linens, etc. that the patient will be using. It procedures.
must first record the patient’s weight before the cast was LOW BATTERY CALL
can be helpful to place a list of Continuous Display put on. When the procedure is completed and the patient When the scale’s battery
these items at the bedside or on The last patient weight reading is back in bed, touch and release to reweigh the power becomes low, the
the footboard for future reference. will appear continuously on the patient. word “BAT” will appear in the
Bed Position display until a new weight reading
• During the weight display, press and hold the appropriate scale’s display. If the bed is
The bed should be in the same is taken.
adjust button, to increase the weight reading, and interfaced to the nurse
position each time the patient is
to decrease the weight reading… station, a nurse call
weighed. The recommended bed
• Once the correct weight is displayed, release the adjust will also be placed. The
position is:
button… tone stops when the bed is
• level sleep surface
• Press to verify the patient weight. plugged in.
• medium bed height
• siderails raised
• brake set
ACCESSORIES
PATIENT WEIGHING
Insure the sleep surface is prepared, the bed is in the correct position, and there are no weight interferences. OXYGEN TANK HOLDER To Horizontally Mount/Store:
The O2 tank holder will hold an • Locate the support
Step One – Zero Scale “E” sized tank and can be brackets underneath the
To recall empty bed weight: Touch until one beep
Before placing the patient mounted either horizontally head of the bed…
sounds then release.
on the surface, the scale underneath the head of the • Place the holder’s
must be cleared of the Step Two – Weigh Patient
bed or vertically in any of the mounting bar into the
empty bed weight. For most accurate weight readings, the patient should be
four IV pole sockets. support brackets…
centered on the bed and lying still.
• Press and hold , one • Push in the mounting bar…
To Vertically Mount:
beep will sound and the • Touch until one beep • The O2 tank holder should
• Place the holder’s mounting
empty bed weight will sounds, then release… click in place…
bar into one of the four IV
appear on the display. Continue to hold the button until • The patient weight will • Place the O2 tank into the holder with the regulator facing up…
sockets located at the four
two consecutive beeps sound, then release… appear in the display. For • Secure the O2 tank in the holder by turning the thumb screw to
corners of the bed…
• The display will flash 000.0 signaling the process has been LBS. to KGS., press the the right or until it stops.
• Place the O2 tank in the holder…
activated… button during weight display.
• Secure the O2 tank in the To Remove The Holder:
• The empty bed weight will appear in the display. It is holder by turning the thumb • Loosen thumb screw…
important to record the empty bed weight for future screw to the right or until it stops. • Remove the O2 tank from the holder…
references… • Grasp the holder and move it to the right…
• The display will flash 000.0 and two beeps will sound NOTE The vertical O2 tank holder will extend the
• Lift the holder from the support brackets.
signaling that the bed has been zeroed. length of the bed.
8 9
OPTIONS OPTIONS
SCALE CHANGING ITEMS ON THE BED To Adjust Empty Bed Weight:
Each control button is color coded and graphically represented. A caption appears below each button describing its During the patient’s stay, For example, the bed has been Zeroed with the patient on
function. Additional operating instructions are located on the footboard. items may be added or the bed. If the empty bed weight is known,
removed from the bed. The you can adjust the empty bed weight to
ZERO DISPLAY button will account for obtain the patient weight.
The ZERO Scale display for the weight of the items • Touch button until a single beep
button resets the weight readings. without affecting the patient’s weight.You will need to sounds, then release…
scale to 000.0 prior account for the items each time they are added or • During the empty bed weight display, press
to placing a patient removed from the bed. and hold the appropriate or button
ADJUST WEIGHT
on the bed. • Touch to obtain most current patient weight… until the correct empty bed weight is
The ADJUST
WEIGHT buttons are • Within 10 seconds of the new patient weight display, displayed, then release…
BED EXIT touch the button until one beep sounds, then • The display will automatically change
used to manually
The BED release… back to the patient’s weight.
adjust patient or
button activates • Place or remove items on the bed, a plus or minus will flash
empty bed weight. ACTIVATING BED EXIT MONITORING
the monitoring for indicating the scale is waiting for further instructions…
patient egress. The bed exit feature will sound an alarm at the bed and, as
• Touch the until one beep sounds, then release… an option, at the nurse station when a patient attempts to
LBS/KGS WEIGH CHANGE ITEMS • The new empty bed weight will display for 10 seconds, leave the bed.
The LBS/KGS The button The CHANGE ITEMS then revert back to the patient’s weight.
button is used to weighs the patient. button maintains patient • Press and hold button until a single beep sounds…
MANUALLY ADJUSTING DISPLAYED WEIGHT • The display will show “BED EXIT,”“LOCAL,” and if available,
switch to lbs.or kgs. weight when items are added
The weight of the patient or empty bed can be manually “REMOTE” to indicate the monitor and alarms are
or removed from the bed.
overridden, if necessary by using the adjust buttons. activated. In addition, the yellow LED bed exit monitor
BEFORE OPERATING THE SCALE Weight Interferences Use these buttons if an incorrect weight needs to be located below the scale controls will illuminate…
For the most accuracy, follow these procedures: Insure that items such as pillows, linens, equipment on corrected or if the exact weight of an item is known. • To deactivate bed exit monitoring, press and hold the
Sleep Surface Preparation stands, etc. are not touching the head/foot board or
To Adjust Patient Weight: button until a single beep sounds and the
Before placing the patient on the bed, the sleep surface permanent IV pole. It is important that staff or visitors do not
For example, if a patient has had a cast put on, the added display clears…
must be prepared. This includes placing articles on the bed touch the siderails or sleep surface during the weighing
weight of the cast must be accounted for. To do this, you • To cancel the alarm, press any button.
such as pillows, linens, etc. that the patient will be using. It procedures.
must first record the patient’s weight before the cast was LOW BATTERY CALL
can be helpful to place a list of Continuous Display put on. When the procedure is completed and the patient When the scale’s battery
these items at the bedside or on The last patient weight reading is back in bed, touch and release to reweigh the power becomes low, the
the footboard for future reference. will appear continuously on the patient. word “BAT” will appear in the
Bed Position display until a new weight reading
• During the weight display, press and hold the appropriate scale’s display. If the bed is
The bed should be in the same is taken.
adjust button, to increase the weight reading, and interfaced to the nurse
position each time the patient is
to decrease the weight reading… station, a nurse call
weighed. The recommended bed
• Once the correct weight is displayed, release the adjust will also be placed. The
position is:
button… tone stops when the bed is
• level sleep surface
• Press to verify the patient weight. plugged in.
• medium bed height
• siderails raised
• brake set
ACCESSORIES
PATIENT WEIGHING
Insure the sleep surface is prepared, the bed is in the correct position, and there are no weight interferences. OXYGEN TANK HOLDER To Horizontally Mount/Store:
The O2 tank holder will hold an • Locate the support
Step One – Zero Scale “E” sized tank and can be brackets underneath the
To recall empty bed weight: Touch until one beep
Before placing the patient mounted either horizontally head of the bed…
sounds then release.
on the surface, the scale underneath the head of the • Place the holder’s
must be cleared of the Step Two – Weigh Patient
bed or vertically in any of the mounting bar into the
empty bed weight. For most accurate weight readings, the patient should be
four IV pole sockets. support brackets…
centered on the bed and lying still.
• Press and hold , one • Push in the mounting bar…
To Vertically Mount:
beep will sound and the • Touch until one beep • The O2 tank holder should
• Place the holder’s mounting
empty bed weight will sounds, then release… click in place…
bar into one of the four IV
appear on the display. Continue to hold the button until • The patient weight will • Place the O2 tank into the holder with the regulator facing up…
sockets located at the four
two consecutive beeps sound, then release… appear in the display. For • Secure the O2 tank in the holder by turning the thumb screw to
corners of the bed…
• The display will flash 000.0 signaling the process has been LBS. to KGS., press the the right or until it stops.
• Place the O2 tank in the holder…
activated… button during weight display.
• Secure the O2 tank in the To Remove The Holder:
• The empty bed weight will appear in the display. It is holder by turning the thumb • Loosen thumb screw…
important to record the empty bed weight for future screw to the right or until it stops. • Remove the O2 tank from the holder…
references… • Grasp the holder and move it to the right…
• The display will flash 000.0 and two beeps will sound NOTE The vertical O2 tank holder will extend the
• Lift the holder from the support brackets.
signaling that the bed has been zeroed. length of the bed.
8 9
ACCESSORIES ACCESSORIES
X-RAY CASSETTE HOLDER X-RAY CASSETTE HOLDER:
Release Tabs LOADING FILM – 2 METHODS
Introduction: The x-ray cassette
holder allows for lower abdominal
to upper chest x-rays with the head
of the bed in a flat or upright
position. The cassette holder Thumb
accepts a 14" x 17" film cassette Release
which can be placed in the holder Film Caddy
either vertically or horizontally.
Positioning
Handle
12 13
ACCESSORIES CLEANING/REMOVAL OF DYNAMICAIRE SLEEP SUFACE
REMOVAL CLEANING MEDICAL FLUID SPILLS
ISS SYSTEM INFUSION SUPPORT SYSTEM TRANSDUCER HOLDER Fluid spills should be wiped up as soon as possible. Some
To remove the sleep surface, raise the foot end of the bed
The Century CC bed comes with four Use of the ISS system is required if the fluids used in the hospital environment, such as betadine,
frame manually and disconnect all four quick disconnect
standard ISS mounts located on the four Hill-Rom transducer is to be used. will leave a permanent stain, but most will not.
hoses and detach the strap.
corners of the bed. To utilize the ISS To Install:
System, the following equipment is To reconnect, replace mattress onto the frame with the hoses Temporary stains can be removed by wiping vigorously with
• After installing the ISS transfer pole a rag or sponge that is wet with a standard household
needed: at the foot end of the bed. Insert the hoses and the strap
and offset bar, place the holder’s cleaning solution. Whatever cleaning solution is used, it must
through the bed frame and re-attach the strap and hoses.
• A transfer pole metal clip directly onto the ISS transfer be diluted according to the manufacturer’s instructions.
• An offset bar pole… NOTE Never remove the mattress from the bed frame
without disconnecting the hoses and straps at the foot of CLEANING BLOOD AND EXCRETA
To Install: • Depending on specific monitoring
the bed. If possible, excess blood and excreta should be wiped up
needs, adjust for distance in relation
• Place the offset bar into any of the when wet, since the cleaning process is more difficult when
to patient by sliding the transducer NOTE Before the bed is cleaned, and especially before
four ISS mounts… these substances are dry and caked onto the ticking.
holder along the ISS transfer pole… touching any electrical components, it should be
• Push the transfer pole into the offset disconnected from the electrical outlet. Since the bed is Cleaning of blood and excreta from the ticking can be
• When desired position is reached,
bar… IV POLE electric, the excess use of cleaning fluids should be avoided. accomplished by wiping vigorously with rag or sponge that
firmly tighten the butterfly screws
• To secure the system, tighten the located on the metal clip…
3
⁄4" telescopic IV pole is wet with a standard household cleaning solution.
butterfly screw on the bed and on the NOTE Care must be taken to secure the foot section to the
that can be used in any The rag or sponge must be rinsed periodically until, after
• Place transducer onto the holder’s head section during any of the above procedures to ensure
offset bar. of the four corner
vertical bar…firmly tighten the that the foot section does not drop back onto the frame wiping, the rag or sponge shows no evidence of the
NOTE Insure the transfer pole and offset sockets of the bed. The substance being cleaned. Rinsing may be required five to
butterfly screws located on the while cleaning.
bar are firmly secured to each other pole can be stored on ten times for puddled, dried blood.
extended bar.
before removing them for transfers. the base of the bed. CLEANING
To clean under the frame, manually lift the foot end frame A final wipe should be done using clean disinfecting
until it’s secure. solution, and a clean rag or sponge. Allow 30 to 60 minutes
for the ticking to dry completely.
To clean directly beneath the sleep surface, lift the mattress
from the head end of the bed. Whatever cleaning or disinfecting solution is used, it must be
diluted according to the manufacturer’s instructions.
GENERAL CLEANING
General cleaning of the sleep surface mattress cover should DAMAGE REPAIR
be done by wiping down with a sponge or rag that has Repair of tears or holes in the upper mattress cover section
been dampened in a cleaning solution. is not recommended due to leak potential of the repair.
The cleaning solution can be any standard household Tears or punctures in the air chambers can usually be
cleaner. For disinfecting purposes, any phenolic or repaired. A repair kit is available from Hill-Rom’s Service
quaternary type cleaners may be used. Whatever cleaning Organization. An air chamber can withstand up to 15
solution is used it must be used, it must and diluted typical needle punctures without significantly affecting
according to the manufacturer’s instructions. performance.
Clorox® brand bleach can be used for disinfecting
purposes, but should not be used in concentrations greater
TRAPEZE SUPPORT BRACKET CANE BUMPERS than 1 part bleach to 10 parts water.
The trapeze support bracket is mounted at the head- May be used with Hill-Rom Envirocare System or the
end frame of the bed and will accommodate the Hill-Rom Horizon System to increase the bed’s distance Care should be taken to wipe up excess cleaning solutions.
installation of trapeze equipment. from the headwall. The cane bumpers insert into Damp areas can take 15 to 30 minutes to dry.
brackets located on the head-end base of the bed. LINENS
Although the bottom sheet can be flat, fitted sheets are
recommended. Proper sizing of fitted sheets is important to
insure good retention. The best retention will be achieved if
the fitted sheet extends 2 inches or more under the corners
of the surface.
Magnets are used on the underside of the surface to assist
in the retention of additional sheets and blankets. Better
retention will be achieved if the number of layers folded
under the magnets is minimized.
12 13
SAFETY TIPS
BED POSITION ELECTRICAL SAFETY
Always leave the bed in the low position when the patient is Policies and procedures must be established to train and
unattended. This could reduce the possibility of patient falls and educate your staff on the risks associated with electric
the severity of resultant injury. equipment. At any time, it is not prudent or necessary for
BRAKES personnel to have their entire body within the confines of the
Brakes should always be set when the bed is occupied and bed. Whenever a bed is being cleaned or serviced it should be
especially during patient transfers. Patients often use the bed unplugged from its power source. If service personnel need to
get under the bed, the Hi-Lo portion must be blocked up as an
Century CC Bed
for support when getting out of bed and could be injured if the
bed unexpectedly moves. After setting the brakes, push and added precaution.
pull the bed sideways to insure stability. INSTANT CPR RELEASE ®
FLUID SPILLS The emergency head and knee release is to be used by
When massive spills occur in the Siderail area or foot end of the healthcare professionals only.The mode selector lever must be
bed, immediately: continually held down in the CPR position until the bed reaches a
flat position.This will insure a smooth operation and avoid delay.
• Clean the fluid from the bed.
• Check the bed controls (i.e. Head, Knee, Hi-Lo). CONVERTIBLE FOOTBOARD
• Have the maintenance department check the internal When auxiliary equipment is placed on the footboard, secure the
electronics. equipment with the straps provided.The weight limit of the
footboard in the transport shelf position is 45 lbs.Always return the
Fluids remaining on the electronic controls may cause corrosion,
footboard to its original position after use. Check to insure proper
which may cause the electronic components to fail. These
latching by tugging on the footboard.
component failures may cause the bed to move or operate on
its own at a time that may be injurious to the patient or staff. FOOT EXTENDER
When using the extender, always give a tug on the extender to
LOCKOUT CONTROLS
insure locking. Do not place any heavy weight on the foot
Whenever a patient should be restricted from operating the
extender and keep in mind the additional length of the bed.
patient controls, activate the appropriate lockout controls
located in the nurse control panel at the foot end of the bed. NURSE CALL SYSTEM INTERFACE
This is especially true for patients in traction since it can prevent If proper interface between the bed and the nurse station is
a patient or visitor from accidentally activating a control which provided, the scale can be set to place a REMOTE call to the
would cause the bed to move and possibly result in an injury. To nurse station. With interface, the call for a scale monitoring
insure the patient’s safety when unattended, place the bed in condition is momentary and repeats until turned off at the bed
its lowest position and lockout the Hi-Lo control. Also, the mode scale. A call indicating a low battery is repeated every ten
selector dial should not be in the Hi-Lo position. minutes until the bed is plugged in. Without Nurse Call
interface, no call will occur in REMOTE. Consult the Nurse Call
SIDERAILS/RESTRAINTS/PATIENT MONITORING
manual to verify compatibility.
The Siderails should always be in a full upright position and
latched when a patient is unattended. When raising the BED POSITION CHANGES
Siderails, an audible “click” should indicate that the Siderails are Be certain that feet and hands are well clear of the lift arm and
completely raised and locked in place. frame assemblies of the bed when changing bed positions,
both manually or electrically.
Hill-Rom recognizes that certain healthcare situations may
indicate the need for specialized Siderail configurations. In MATTRESSES
response to this need, we offer, upon request, several Siderail The use of mattresses that are not sold by Hill-Rom may reduce
accessories. the effectiveness of the safety features and systems
incorporated into Hill-Rom beds.
Siderails are intended to be a reminder, not a patient restraining
device. Hill-Rom recommends the appropriate medical X-RAY CASSETTE HOLDER
personnel determine the level of restraint necessary to ensure a If using the x-ray cassette holder accessory, always secure the
patient will remain safely in bed. Consult the restraint manufac- holder to the bed when not in use. This will avoid possible
turer’s instructions for use to verify the correct damage to the bed and holder.
application of each restraining device. PREVENTIVE MAINTENANCE
Whenever “high profile” patients (typically, the frail, elderly and Annual preventive maintenance must be performed to insure all
medicated or confused) are involved, Hill-Rom recommends bed features are functioning as originally designed.
the following minimum actions: Particular attention must be addressed on safety features,
including but not limited to:
1. Develop guidelines for all high profile patients that indicate:
• Siderail latching mechanisms
• Which patients may need to be restrained and the
• Caster braking systems
appropriate restraint to utilize.
• Frayed electrical cords and components
• The proper method to monitor a patient, whether
• All controls return to off or neutral position when released.
restrained or not, including time interval, visual check of
• Controls or cabling entanglement of bed mechanisms in
restraint, etc.
Siderails.
2. Develop training programs for all caregivers concerning the
• Proper operation of the lockout function controls.
proper use and application of restraints.
3. Maintain the bed at its lowest position whenever a caregiver
is not in the room.
4. Clarify the need for restraint devices to families or guardians.