REFLEX EZ-GYRO™
Your REFLEX EZ-GYRO™ system has been Calibration Test Data
manufactured using pre-tested components. Calibration completed on 03-Apr-2018
Following assembly approval, the REFLEX Recalibration Due 03-Apr-2019
EZ-GYRO™ probe has been subjected to the Tool serial number EG0318
REFLEX standard testing procedures. Temperature range (°C) +2.85 to +59.72
Following satisfactory completion of the
Tool Error
assembly tests, the REFLEX EZ-GYRO™
probe has been calibrated in accordance with 1σ Error Max 1σ Error Max
Azimuth Dip VTF GTF
REFLEX standards.
(Dip -80° to -60°) (Dip -90° to - (Dip -90° to - (Dip -80°
60°) 60°) to -60°)
Actual calibration results are shown in this IMU 0.55 0.04 0.53 0.21
certificate for your information and reference. Limits 0.7 0.25 0.7 1
In addition, your REFLEX EZ-GYRO™ system
is covered by the Worldwide Warranty of TOOL STATUS: PASSED
reflex@imdexlimited.com
www.reflexinstruments.com