Instruction Manual Dear Customer, Specifications Display Code Guide

Thank you for choosing a HANNA product. This manual


HI 93733 will provide you with the necessary information for the This indicates that the meter is in a
correct operation of the meter. Please read it carefully before
Ammonia ISM using the meter. If you need additional technical information,
ready state and zeroing can be
do not hesitate to e-mail us at tech@hannainst.com. performed.
This instrument is in compliance with directives. Sampling in Progress. This prompt
appears each time the meter is
performing a measurement.
This indicates that the meter is in a
Preliminary Examination zeroed state and measurement can be
Remove the instrument from the packing material and examine performed.
it carefully to make sure that no damage has occurred during A zero reading was not taken. Insert
shipment. If there is any damage, notify your Dealer. a sample before adding reagent and
Each Ion Specific Meter is supplied complete with press ZERO.
• 9V Battery Under range. A blinking "0.0"
• Two Sample Cuvets and Caps indicates that the sample absorbs less
• One Transport Cap light than the zero reference. Check
Note: Conserve all packing material until the instrument has the procedure and make sure you use
been observed to function correctly. Any defective item the same cuvet for reference (zero)
must be returned in its original packing. and measurement.
SPECIFICATIONS Over range. A flashing value higher
This Instrument is in
www.hannainst.com Compliance with the Directives Range 0.0 to 50.0 mg/L than the maximum concentration
Resolution 0.1 mg/L readable (see specifications) indicates
Accuracy ±0.5 mg/L ±5% of reading that the sample absorbs too much
Typical EMC ±0.1 mg/L light, meaning that the concentration
Warranty General Description Deviation is too high. Dilute the sample.
The HI 93733 meter measures the ammonium ion (NH4+) Light Source Light Emitting Diode @ 470 nm
HI 93733 is warranted for two years against defects in
Light over range. The cuvet is not
content in water, wastewater and seawater in the 0.0 to 50.0 Method Adaptation of the ASTM Manual of
workmanship and materials when used for its intended purpose and Water and Environmental Technology, inserted correctly and an excess
maintained according to instructions. mg/L (ppm) range. ambient light is reaching the detector.
D1426-92, Nessler method. The
This warranty is limited to repair or replacement free of charge. The meter uses an exclusive positive-locking system to ensure reaction between ammonia and If the cover is properly installed, then
Damages due to accident, misuse, tampering or lack of prescribed that the cuvet is in the same place every time it is placed into reagents causes a yellow tint in the contact your dealer or the nearest
maintenance are not covered. the measurement cell. sample. Hanna Customer Service Center.
If service is required, contact the dealer from whom you purchased Light Detector Silicon Photocell
the instrument. If under warranty, report the model number, date of The reagents are in liquid form and are supplied in bottles. Light under range. The zero sample is
purchase, serial number and the nature of the failure. If the repair The amount of reagent is precisely dosed to ensure maximum Environment 0 to 50°C (32 to 122°F); too dark for proper zeroing. If this is
is not covered by the warranty, you will be notified of the charges repeatability. max 95% RH non-condensing not the case, contact your dealer or
incurred. If the instrument is to be returned to Hanna Instruments, Battery Type/Life 1 x 9 volt/40 hours
first obtain a Returned Goods Authorization Number from the Display codes aid the user in routine operations. the nearest Hanna Customer Service
Customer Service department and then send it with shipment costs
Auto-Shut off After 10' of non-use Center.
The meters have an auto-shut off feature that will turn the
prepaid. When shipping any instrument, make sure it is properly Dimensions 180 x 83 x 46 mm
instrument off after 10 minutes of non-use. The "V" indicates that the battery
packaged for complete protection. (7.1 x 3.3 x 1.8")
voltage is getting low and the battery
To validate your warranty, fill out and return the enclosed warranty Weight 290 g (10 oz.)
card within 14 days from the date of purchase. needs to be replaced.
All rights are reserved. Reproduction in whole or in part is prohibited
REQUIRED REAGENTS This indicates that the battery is dead
without the written consent of the copyright owner, Hanna Instruments Code Description Quantity and must be replaced.
Inc., Woonsocket, Rhode Island, 02895 , USA. HI 93733A-0 Nessler Reagent 4 drops (in fresh and Note: once this indication is
seawater) displayed, the meter will lockup.
HANNA instruments® reserves the right to modify the design, HI 93733B-0 Reagent B 9 mL (in fresh and
construction and appearance of its products without advance notice. Change the battery to restart.
IST93733R4 04/07 seawater)
Operational Guide • Remove the cuvet. Tips for an Accurate Measurement Accessories
MEASUREMENT PROCEDURE
• Add 4 drops of HI 93733A-0 The instruction listed below should be carefully followed REAGENT SETS
ON
• Turn the meter on by Nessler Reagent. during testing to ensure best accuracy. HI 93733-01 Reagents for 100 tests Ammonia
OFF • Do not touch the cuvet walls with hands.
pressing ON/OFF. HI 93733-03 Reagents for 300 tests Ammonia
• In order to maintain the same conditions during the
• When the LCD displays OTHER ACCESSORIES
zeroing and the measuring phases, it is necessary to close
“- - -”, it is ready. the cuvet to prevent any contamination. HI 710009 Blue rubber boot
• Replace the cap and swirl the
• Do not let the test sample stand too long after reagent is HI 710010 Orange rubber boot
• Fill a cuvet with 1 mL of solution.
unreacted sample, by means added or accuracy will be lost. HI 721310 9V battery (10 pcs)
of the syringe. • Whenever the cuvet is placed into the measurement cell, HI 731318 Tissue for wiping cuvets (4 pcs)
it must be completely free of fingerprints, oil or dirt. HI 731321 Glass cuvets (4 pcs)
Wipe it thoroughly with HI 731318 or a lint-free cloth HI 731325 Caps for cuvets (4 pcs)
• Reinsert the cuvet into the
prior to insertion. HI 93703-50 Cuvets cleaning solution (230 mL).
instrument.
• It is important that the sample does not contain any
debris. This would corrupt the readings. CE Declaration of Conformity
• It is possible to take multiple readings in a row, but it is
• Press READ TIMED and the display will show the recommended that a zero reading be taken for each
countdown prior to the measurement or, alternatively, sample and that the same cuvet is used for zeroing and
wait for 3 minutes and 30 seconds and press READ measurement.
• Using the plastic pipette, add DIRECT. In both cases “SIP” will appear during • It is important to discard the sample immediately after
9 mL of HI 93733B-0 measurement. the reading is taken because the glass might become
Ammonia Reagent B, up to permanently stained.
the 10 mL mark. Place the • Shaking the cuvet can generate bubbles in the sample,
cap and swirl the solution READ causing higher readings. To obtain accurate measurements,
to mix.
TIMED
remove such bubbles by swirling or by gently tapping the
vial.
• All the reaction times reported in this manual are referred
to 20°C (68°F). As a general rule of thumb, they should
• Place the cuvet into the • The instrument directly displays concentration in mg/L of
be doubled at 10°C (50°F) and halved at 30°C (86°F).
holder and ensure that the ammonium ion (NH4+).
notch on the cap is • To convert the reading to mg/L of ammonia (NH3),
positioned securely into the multiply by the factor 0.944.
groove. • To convert the reading to mg/L of ammonia nitrogen
(NH3-N), multiply by the factor 0.776. Battery Replacement Recommendations for Users
• Press ZERO and “SIP” will appear on the display. Before using these products, make sure that they are entirely suitable for the environment
Battery replacement must only take place in a non-hazardous in which they are used.
Operation of these instruments in residential area could cause unacceptable interferences
area using a 9V alkaline battery.
ZERO INTERFERENCES to radio and TV equipments, requiring the operator to take all necessary steps to correct
Simply slide off the battery cover on the back of the meter. interferences.
Interference may be caused by: Detach the battery from the terminals and attach a fresh 9V Any variation introduced by the user to the supplied equipment may degrade the
instruments' EMC performance.
acetone, alcohols, aldehydes, glycine, hardness above 1 g/L, battery while paying attention to the correct polarity. Replace To avoid damages or burns, do not perform any measurement in microwave ovens.
• Wait for a few seconds and iron, organic chloramines, sulfide, various aliphatic and the battery and the cover.
the display will show aromatic amines.
“-0.0-”. Now the meter is Safety Data Sheets
ACCESSORIES
zeroed and ready for Read the relevant SDS sheets before performing the test.
9V

• 1 mL syringe with tip


measurement. SLIDE
• 1 plastic pipette OFF Safety Data Sheets are available at: www.hannainst.com