User Guide: Blood Glucose Monitoring System
User Guide: Blood Glucose Monitoring System
Blood Glucose
Monitoring System
User Guide
Testing
Getting the Blood Drop 9
• Read your CONTOUR XT user guide, the MICROLET®2 Testing Your Blood 10
lancing device insert and all instructional materials provided in Cleaning Your Meter 13
your meter kit before testing. Please follow all instructions for use Control Solution Testing 14
and care exactly as described to help avoid inaccurate results.
• Wash and dry your hands well before testing.
• Used test strips and lancets are possible biohazards and must be
disposed of as medical waste.
Technical,
investigation.1 Caring for Your Meter 37
For complete cleaning instructions, see page 13. Symptoms of High or Low Blood Glucose 38
Specifications 39
• The MICROLET 2 lancing device is not suitable for use by healthcare Service Information 44
professionals. It must not be used on more than one person due to Warranty 45
the risk of infection. Supplies 47
Getting Started
Your CONTOUR®XT Meter Your Meter Display
The picture below shows all the symbols that appear on
your meter display. To view, with the meter off, press
and hold ▲ or ▼. All display segments will appear for
10 seconds. If you need more than 10 seconds to check
the display, press and hold ▲ or ▼ again.
Display
While all characters are displayed, it is very impor-
tant to verify that is fully displayed. If there
Up/Down are missing segments, see Error Codes and Symbols
Meter Overview
Meter Overview
Memory Button
Scroll Buttons (pages 32 – 34). This may affect the way you see your
(On/Off, Memory results.
Recall, Enter) Indicates sound
Test Strip Port or will
is turned on. appear if the meter
Top area on
the display is in 12 hour setup.
Your CONTOUR®NEXT Test Strip
presenting
Grey End date and time.
Insert this end into
The batteries are
Sample Tip the test strip port. low and need to be
Blood sample changed, and will
pulled in here. Pre-meal remain on display
result marker. until batteries are
Your MICROLET ®2 Lancing Device changed.
Center area on the
Re-setting Handle display that shows either
your test result or any Day/month format.
Post-meal
Release Button result marker. error codes.
The meter is
Lancet Holder ready to test.
Grip Indent Results are
Identifies a displayed as
Lancet Diary entry.
Round Control solution mmol/L or mg/dL.
result marker. The 7 appears when
Endcap Protective Cap viewing 7 day HI–
Dial LO summary.
Indicates it is too
Indicates a cold or too hot for
reminder has your meter to test
Adjustable Endcap been set. accurately. Indicates Average.
2 3
Getting Started
Thank you for choosing the CONTOUR®XT blood Information For Safety
glucose monitoring system! We are proud to be your
partner in helping you manage your diabetes. • Always keep the test strips in the
original bottle. Tightly close the
Your Meter Features bottle immediately after removing
a test strip. The bottle is designed to
Easy: The CONTOUR XT blood glucose monitoring system keep the test strips dry. Exposure to room
is about simplicity. You will notice its simple approach the humidity from leaving the bottle open or
first time you see it. not storing them in the original bottle can
damage your test strips. This could lead
Automatic: Your CONTOUR XT meter is designed with No
Your Meter Features
Programmed: Your CONTOUR XT meter has been preset • Your test strip is designed to easily “sip” the
at the factory with the standard date and time for your blood into the sample tip. Do not drop blood directly
convenience. on the flat surface of the test strip.
• Your meter has been preset and locked to display results • Do not press the test strip against your finger
in mmol/L (millimoles of glucose per liter). In some when testing. This may block the sample tip.
countries, the meters are preset to display units in mg/dL • Let the meter and test strips adjust to the temper-
(milligrams of glucose per deciliter); ature where you are testing. Whenever the meter is
• Results in mmol/L will always have a decimal point; moved from one location to another, allow approximately
20 minutes for the meter to adjust to the temperature of
• Results in mg/dL will never have a decimal point. a new location before performing a blood glucose test.
The meter has been designed to give accurate results at
Example: or temperatures between 5°C and 45°C.
Check your display to ensure that results are being • Check the expiry dates on your test strips and
displayed in mmol/L. If they are not, contact Bayer control solution. It is important to not use the test
Diabetes Support at 0845 600 6030 (UK) / strips or control solution if the expiry date printed on
1 890 920 111 (Ireland). the bottle label and carton has passed. For the control
solution do not use if it has been six months since you
Customizable: The CONTOUR XT meter allows you to set first opened the bottle. It will help to write the six month
personal choices for some features. For more information, discard date on the control solution label.
please see pages 21 – 27.
4 5
• Examine product for missing, damaged, or broken parts. 3. Hold the test strip with the
If the test strip bottle is open inside a new box of strips, grey end facing up.
Testing
Testing
do not use those strips. Contact Bayer Diabetes Support 4. Insert the grey end into the
at 0845 600 6030 (UK) / 1 890 920 111 (Ireland) for test strip port on the meter.
replacement parts.
The meter will turn on. An image of a test strip with a flashing
• The MICROLET 2 lancing device is intended blood drop will appear letting you know the meter is
for self testing by a single patient. It must not be ready to test.
used on more than one person due to the risk of
infection.
Preparing the Lancing Device
Information For Safety
Need Help? 0845 600 6030 (UK) / 1 890 920 111 (Ireland)
6 CONTOUR®XT meter CONTOUR®NEXT test strips www.bayerdiabetes.co.uk 7
Testing
drop of blood. Do not
squeeze around the
5. Replace the grey endcap puncture site.
prior to getting your blood
drop.
Need Help? 0845 600 6030 (UK) / 1 890 920 111 (Ireland)
8 CONTOUR®XT meter CONTOUR®NEXT test strips www.bayerdiabetes.co.uk 9
Testing
If the first blood drop is Fingertip
not enough, the meter may Testing Blood glucose values will vary depending on food intake,
beep twice and display a strip medication dosages, health, stress, or activity. Non diabetic
underfill screen image. plasma glucose concentrations are normally maintained
within a relatively narrow range, roughly 3.9 - 6.1 mmol/L in
You have 20 seconds to apply more blood to the same strip. a fasting state.2 You should consult with your healthcare
If you are not able to apply enough blood within that time, professional for expected values specific to your needs.
an E2 error code will be displayed. Dispose of the used test
Testing Your Blood
Testing
down.
3. With the used lancet still
in the lancing device,
push the lancet needle
completely into the middle
of the exposed side of the
cap.
Testing Your Blood
Need Help? 0845 600 6030 (UK) / 1 890 920 111 (Ireland)
12 CONTOUR®XT meter CONTOUR®NEXT test strips www.bayerdiabetes.co.uk 13
Testing
• If you have repeated unexpected blood glucose results
Normal, Low or High control solutions are available 6. Squeeze a small drop
separately. Call Bayer Diabetes Support at of control solution on
0845 600 6030 (UK) / 1 890 920 111 (Ireland). a clean nonabsorbent
Always use Bayer’s CONTOUR NEXT control solutions. surface. Do not apply
Other brands could present incorrect results. control solution to your
fingertip or to the test
strip directly from the
Control Solution Testing
Need Help? 0845 600 6030 (UK) / 1 890 920 111 (Ireland)
14 CONTOUR®XT meter CONTOUR®NEXT test strips www.bayerdiabetes.co.uk 15
Testing
control result for you.
Control results will not
be included in your blood
glucose averages or in the
7 day HI-LO summary.
Need Help? 0845 600 6030 (UK) / 1 890 920 111 (Ireland)
16 CONTOUR®XT meter CONTOUR®NEXT test strips www.bayerdiabetes.co.uk 17
Reminder Features
you track these HI and LO
Reminder Settings
Reminder Features
any button will silence the
Using Marker and
Press M to cancel the reminder. Support at 0845 600 6030 (UK) / 1 890 920 111 (Ireland).
Using Marker and
Pre-Meal
Setup and Use
Need Help? 0845 600 6030 (UK) / 1 890 920 111 (Ireland)
28 CONTOUR®XT meter CONTOUR®NEXT test strips www.bayerdiabetes.co.uk 29
Technical,
below 0.6 mmol/L. test strip.
E4 Test strip not Remove and reinsert the test strip
• If your result still flashes “LO”,
inserted correctly. properly (see page 7).
follow medical advice immediately.
Technical,
Need Help? 0845 600 6030 (UK) / 1 890 920 111 (Ireland)
34 CONTOUR®XT meter CONTOUR®NEXT test strips www.bayerdiabetes.co.uk 35
Technical,
Need Help? 0845 600 6030 (UK) / 1 890 920 111 (Ireland)
36 CONTOUR®XT meter CONTOUR®NEXT test strips www.bayerdiabetes.co.uk 37
(Hypoglycaemia): (Hyperglycaemia):
Battery Type: Two 3-volt lithium batteries, 225 mAh capacity
• shakiness • frequent urination (DL2032 or CR2032)
• sweating • excessive thirst Battery Life: Approximately 1000 tests (1 yr. average use)
• fast heartbeat • blurred vision Operating Temperature Range:
• blurred vision • increased fatigue
Humidity: 10 – 93% RH
• confusion • hunger
Dimensions: 77 mm (H) x 57 mm (W) x 19 mm (T)
• passing out
• seizure Weight: 47.5 grams (1.7 oz.)
• irritability Sound: A beep sounds when the meter is turned on, a test strip is
System Specifications
inserted, when a test strip is filled with blood, or when a test
• extreme hunger
result appears on the display. Two beeps sound when the
• dizziness meter turns off or to indicate an error. You will hear twenty
Ketones (Ketoacidosis): beeps when a programmed reminder sounds.
Electromagnetic Compatibility (EMC): The CONTOUR®XT meter
• shortness of breath complies with the electromagnetic requirements specified in ISO
• nausea or vomiting 15197: 2003. Electromagnetic emissions are low and unlikely to
• very dry mouth interfere with other nearby electronic equipment, nor are emissions
from nearby electronic equipment likely to interfere with the
CONTOUR XT. Immunity to electrostatic discharge meets the
Information For Safety requirements of IEC 61000-4-2: 2008. It is advisable to avoid use of
electronic devices in very dry environments especially if synthetic
If you are experiencing any of these symptoms,
materials are present. The CONTOUR XT meter has been tested for
test your blood glucose. If your test result is under radio frequency interference at the frequency range and test levels
Service & Care
Technical,
healthcare professional immediately. ence do not use the CONTOUR XT meter near cellular or cordless
telephones, walkie talkies, garage door openers, radio transmitters or
For additional information and a complete list of symptoms, other electrical or electronic equipment that are sources of electro-
contact your healthcare professional. magnetic radiation, as these may interfere with the proper operation
of the meter.
Specifications
laws regarding disposal and
results. recycling in your area.
Peritoneal dialysis solutions: Icodextrin does not interfere with Caution The meter should be treated
CONTOUR®NEXT test strips. as contaminated and
Xylose: Do not use during or soon after xylose absorption Catalogue number disposed of according to
testing. Xylose in the blood will cause an interference. local safety rules. It should
Contraindications: Capillary blood glucose testing may not be not be disposed of with
Control Range Normal waste electronic equipment.
clinically appropriate for persons with reduced peripheral blood flow.
Shock, severe hypotension, hyperosmolar hyperglycaemia and Contact your healthcare
severe dehydration are examples of clinical conditions that may Control Range Low professional or local waste
Service & Care
Technical,
Interference: Reducing substances occurring in the blood naturally Control Range High waste disposal guidelines.
(uric acid, bilirubin) or from therapeutic treatments (ascorbic acid,
paracetamol) will not significantly affect results. The limiting concen-
trations of these compounds are listed on your test strip insert.
Need Help? 0845 600 6030 (UK) / 1 890 920 111 (Ireland)
40 CONTOUR®XT meter CONTOUR®NEXT test strips www.bayerdiabetes.co.uk 41
Technical Information
Number (and percent) 48 of 54 54 of 54 54 of 54
of samples within 6.50 0.08 1.3
specified range (88.9%) (100.0%) (100.0%)
10.59 0.12 1.2
Table 2 - System accuracy results for glucose
17.05 0.29 1.7
concentration ≥ 4.2 mmol/L
Difference range
in values between Within ± Within ± Within ± Within ±
YSI laboratory
reference method and 5% 10% 15% 20%*
CONTOUR XT meter
Number (and percent) 439 of 546 544 of 546 546 of 546 546 of 546
Service & Care
of samples within
Technical,
*Acceptance criteria in ISO 15197: 2003 are that 95% of all differ-
ences in glucose values (i.e., between reference method and meter)
should be within 0.8 mmol/L for glucose values less than 4.2 mmol/L,
and within 20% for glucose values greater than or equal to 4.2 mmol/L.
Need Help? 0845 600 6030 (UK) / 1 890 920 111 (Ireland)
42 CONTOUR®XT meter CONTOUR®NEXT test strips www.bayerdiabetes.co.uk 43
This check list may be helpful when speaking with Bayer instrument in accordance with instructions. Further,
Diabetes Support: Bayer HealthCare assumes no liability for malfunction
1. Locate the serial number 6HULDO1R
or damage to Bayer HealthCare instruments caused by
1000001 7262 the use of reagents other than the appropriate Bayer
(A) and model number (B)
on the back of the meter. Serial No.
reagents (e.g., CONTOUR®NEXT test strips) manufactured
Warranty
A 1000001
_______________________ or recommended by Bayer HealthCare.
_______________________ ®
3. Bayer HealthCare reserves the right to make changes in
B
2. Locate the CONTOUR NEXT design of this instrument without obligation to incorporate
test strips’ expiry date on such changes into previously manufactured instruments.
the bottle. 4. Bayer HealthCare has no knowledge of the performance
_______________________ of the CONTOUR®XT blood glucose meter when used
Service & Care
Technical,
Changing the Batteries.) and therefore makes no warranty of the performance
of the CONTOUR XT meter when used with any test
strips other than CONTOUR NEXT test strips or when
the CONTOUR NEXT test strip is altered or modified in
any manner.
44 CONTOUR®XT meter CONTOUR®NEXT test strips 45
Supplies
3
American Diabetes Association, www.diabetes.org
4
Tietz Fundamentals of Clinical Chemistry, 5th Edition, Edited by
Burtis CA and Ashwood ER, W. B. Saunders Co., Philadelphia,
PA, 2001, p. 444.
5
Atkin, S. et al.: Fingerstick Glucose Determination in Shock.
Ann. Int. Med. 114:1020-24; 1991.
Service & Care
Technical,
Need Help? 0845 600 6030 (UK) / 1 890 920 111 (Ireland)
46 CONTOUR®XT meter CONTOUR®NEXT test strips www.bayerdiabetes.co.uk 47
1534497_CntrXT_UG_ENuk_FpBp_v0.indd