N
Tae
Instrumentation & Technologies
Cement, Lime
Plaster & Mortar
Testing EquipmentTo determine the fineness of cements, pozzolanas
and other powdery materials, various international
specifications recommend the use of “specific area”
method. In the instrument described below, air
permeability method is used to determine the specific
surface as total surface area in om*/g of material.
Crs
Ref. Standards :18:1727, 4031,4825
Itis a variable flow type AirPermeabilty Apparatus.
The equipment consist of the following
replaceable parts :
Gees
Bg
AIM 39001 Permeability Cell.
‘AIM 39002 ‘U'Tube Manometer, mounted on stand.
‘AIM 39003 Perforated Metal Disc.
‘AIM 39004 Plunger.
AIM 39005 Rubber Stopper.
‘AIM.39006 Rubber Tube, 20 cm long.
‘AIM 39007 Filter Paper Discs (Twelve Nos.)
AIM 39008 —_Dibutylphthalate Liquid, 100 ml bottle.
‘AIM 39009 Punch.
AIM 39010 Non Perforated Disc.
AIM 39011 Suction Bulb.
2
‘AIM 390
Ordering Information :
‘AIM390. Air Permeabilty Apparatus (Blaine Type)
with SI Certification Mark, IS:5516
As per BS/ASTM is also available on request
Ref, Standards: 1S:1727, 4031, 4825
Blaine apparatus is used to determine the fineness
of coment
The Aimil - DT Automatic Blaine has the
following features:
* Single touch operation:
* Automatic control of the movement of fuid
unti!the upper mark.
+ Automatic sensing of the upper & lower
marks ensuring error free repeatable
measurement of the lime taken for the
liquid column to fall
* Automatic measurement of temperature
during the testusing a Pt 100 probe
+ Automatic correction of formula for
calculation of the Blaine value per 1S:5516
with variation in temperature.
* Facility to operate in either Standalone Mode or
through a Computer Controlled system complete
with software,
* Login Facility as Supervisor & Operator
* Report Generation based on Date/Time, User type
& typeof cement
* Facility tomeasure Cement Sample Load
* Facility to Monitor and Configure various
Cement types
* Timing accuracy up to 200mSec.
* Power Consumption|ess than 5W
I
AIM 391-1
Ordering Information :
‘AIM391-1 Automatic Blaine Apparatus with
standalone data recording & controlling
system having inbuilt temperature
‘monitoring facility
Automatic Blaine Apparatus with
PC based data recording & monitoring
system
‘AIM 391-2
Mineo et azi
Tal)
Cel
a °
‘Apparatus for determining the normal consistency,
‘standard consistency and time of setting of cement and
lime in accordance with IS, ASTM, BS and AASHTO
‘specifications,
Vicat Method :
Ref. Standards - 1S:2542 (Part-2), 2645, 1727, 5513,
BS:12, 146, 915, 1370, 4027, 4248, 4248, ASTM C-01,
(C-141,C-187, C-308, C-359, C-472& AASHTOT-129, E-134
‘The procedure, as recommended in various standards, is
for determining the quantity of water required to produce
‘a cement paste of standard consistency. The standard
‘consistency is attained when the 10 mm plunger of the
apparatus penetrates the material to a pre-determined
depth under freefall. A new sample is prepared and
tested with initial and final needles in accordance with the
procedure described in various specifications.
The equipment consist of the following
replaceable parts :
‘AIM 39301 Vicat Mould
‘AIM 39302 Glass Base Plate
‘AIM 39303 _ Initial Needle (in Plastic Case)
‘AIM 39304 Final Needle (in Plastic Case)
‘AIM 39305. Consistency Plunger (in Plastic Case)
‘AIM 39306 Mild Steel Base Plate
‘AIM 39307 —_Vicat Mould SpiitType, with Clamping Ring
‘AIM 394
Ordering Information :
‘AIM 394° Vicat Apparatus,
with IS! Certification Mark, 1$:5513
‘As per BS/ASTM Is also available on request.
setting ime of hydraulic
cement. it consists of a base with a vertical shaft and
‘wo horizontal arms, the lower arm being adjustable for height.
‘Supplied complete with one initial needle, one final
needle anda glass base plate,
The equipment consist of the following
replaceable parts :
‘AIM 39601 Initial Needle 1 /12inch dia. 1/4 b.wt
‘AIM39602__FinalNeedle 1 /24inchdia. 1/41b.wt
‘AIM39603 Giassbase plate.
AIM 396
Ordering Information:
Gillmore Needle Apparatus
AIM396
‘The soundness of ordinary and rapid hardening portiand
cement, low heat porttand cement and class-A lime can
be determined by an expansion test using Le-Chatelier
moulds as specified in 1S:5514, 1727, 2645, 4031, 6932
(Part-9), BS:890, 915, 1370, 4027, 4226 and 4248,
oes
Ref. Standard - 1S:5514
consists ofa small spit cyinder forming a mould of 30mm,
internal dia. and 30 mm high. Oneiher side ote spl oyinde,
‘wo parallelindicating arms with pointedends arefixed.
‘Two loops of suitable material and strength soldered to
the upper half of the mould on each side ofthe central split
is provided to facilitate demoulding of the hardened paste:
specimen after test
‘The resilience of the mould shall be such that the action of.
‘a mass of 300g applied shall increase the distance
between the indicator ends of the needle by 17.5mm +
2.5mm without permanent deformation.
‘The mould is supplied complete with two glass plates and
_- Bge
‘AIM 400