PRESSURE GAUGE

INSTALLATION,
OPERATION AND
MAINTENANCE

I&M012-10098-5/02 (250-1353H)
ii
CONTENTS

1.0 Selection and Application Page


1.1 Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
1.2 Temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
1.3 Media . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
1.4 Oxidizing media . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
1.5 Pulsation/Vibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
1.6 Gauge fills . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
1.7 Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2

2.0 Temperature
2.1 Ratings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2
2.2 Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2
2.3 Steam service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2
2.4 Hot or very cold media . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2
2.5 Diaphragm seals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2
2.6 Autoclaving . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2

3.0 Installation
3.1 Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
3.2 Gauge reuse . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
3.3 Tightening of gauge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
3.4 Process isolation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
3.5 Surface mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
3.6 Flush mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3

4.0 Operation
4.1 Frequency of inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
4.2 In-service inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
4.3 When to check accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
4.4 When to recalibrate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
4.5 Other considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
4.6 Spare parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4

5.0 Gauge Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4

6.0 Accuracy: Procedures/Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-5


6.1 Calibration - Rotary movement gauges . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
6.2 Calibration - 1009 Duralife® Pressure Gauge . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
6.3 Positive Pressure Ranges . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-10

iii
CONTENTS
Page

7.0 Diaphragm Seals


7.1 General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12
7.2 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12
7.3 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12
7.4 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12
7.5 Failures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12

8.0 Dampening Devices


8.1 General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
8.2 Throttle Screws & Plugs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
8.3 Ashcroft Gauge Saver . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
8.4 Ashcroft Pulsation Dampener . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
8.5 Ashcroft Pressure Snubber . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
8.6 Campbell Micro-Bean® . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
8.7 Ashcroft Needle Valves . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
8.8 Chemiquip® Pressure Limiting Valves . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13

9.0 Resources
9.1 Training Videos . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14
9.2 Pressure Instrument Testing Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14
9.3 Tools & Tool Kits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14
9.4 Sales Bulletins . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14

Cover photo courtesy of Johnson/Yokogawa Co.

Campbell Micro-Bean® is a registered trademark of J.A. Campbell Co.

Chemiquip® is a registered trademark of Chemiquip Products Co. Inc.

iv
1.0 SELECTION & APPLICATION or Dresser Instrument Division. Use of a dia-
phragm seal with the gauge is recommended for
process media that (a) are corrosive to the
Users should become familiar with ASME
process sensing element; (b) contain heavy par-
B40.1 (Gauges – Pressure Indicating Dial
ticulates (slurries) or (c) are very viscous includ-
Type – Elastic Element) before specifying
ing those that harden at room temperature.
pressure measuring instruments. That docu-
ment – containing valuable information 1.4 Oxidizing media – Gauges for direct use
regarding gauge construction, accuracy, safety, on oxidizing media should be specially
selection and testing – may be ordered from: cleaned. Gauges for oxygen service should be
ordered to variation X6B and will carry the
ASME International
ASME required dial marking “USE NO OIL”
Three Park Avenue
in red letters. Gauges for direct use on other
New York, N.Y. 10016-5990
oxidizing media may be ordered to variation
800-843-2763 (US/Canada)
X6W. They will be cleaned but carry no dial
95-800-843-2763 (Mexico)
marking. Plus!® gauges or Halocarbon filled
973-882-1167 outside North America
gauge or diaphragm fill is required for use
Email: infocentral@asme.org
with oxidizing media; order variation XCF.
WARNING: To prevent misapplication, pres- See page 4 of the text of the Ashcroft
sure gauges should be selected considering Corrosion Guide mentioned in paragraph 1.4
media and ambient operating conditions. above for more information.
Improper application can be detrimental to the
gauge, causing failure and possible personal 1.5 Pulsation/Vibration – Pressure pulsation
injury, property damage or death. The infor- can be dampened by several mechanisms; the
mation contained in this manual is offered as patented PLUS! Performance gauge will han-
a guide in making the proper selection of a dle the vast majority of applications. One
pressure gauge. Additional information is exception to this is high frequency pulsation
available from Dresser Instrument Division. which is difficult to detect. The only indication
may be an upscale zero shift due to movement
The following is a highlight of some of the more wear. These applications should be addressed
important considerations: with a liquid filled gauge, or in extreme cases,
a remotely mounted liquid filled gauge con-
1.1 Range – The range of the instrument nected with a length of capillary line. The
should be approximately twice the maximum small diameter of the capillary provides excel-
operating pressure. Too low a range may lent dampening, but can be plugged. The
result in (a) low fatigue life of the elastic ele- Ashcroft 1106 pulsation dampener and 112
ment due to high operating stress and (b) sus- snubber are auxiliary devices which dampen
ceptibility to overpressure set due to pressure pulsation with less tendency to plug.
transients that exceed the normal operating
pressure. Too high a range may yield insuffi- 1.6 Gauge fills. – Once it has been determined
cient resolution for the application. that a liquid filled gauge is in order, the next
step is selecting the type of fill. Glycerin sat-
1.2 Temperature – Refer to page 2 of this isfies most applications. While being the least
manual for important information concerning expensive fill, its usable temperature range is
temperature related limitations of pressure 20/250°F. Silicone filled gauges have a
gauges, both dry and liquid filled. broader service range: –40/250°F. Oxidizing
1.3 Media – The material of the process sensing media require the use of Halocarbon, with a
element must be compatible with the process service range of –50/250°F. Pointer motion
media. Consult the Corrosion Guide available will be slowed at the low end of the low end of
on the website: www.dresserinstruments.com these temperature ranges.

1
1.7 Mounting – Users should predetermine gauges is 0.5% of full scale change for every
how the gauge will be mounted in service: 40°F change from 75°F. Double that allowance
stem (pipe), wall (surface) or panel (flush). for gauges with hermetically sealed or liquid
Ashcroft wall or panel mounting kits should be filled cases, except for Duragauge® gauges
ordered with the gauge. See paragraph 3 where no extra allowance is required due to
Installation. the elastomeric, compensating back. Above
250°F there may exist very significant errors
in indication.
2.0 TEMPERATURE
2.3 Steam service – In order to prevent live
2.1 Ambient Temperature – To ensure long steam from entering the bourdon tube, a
life and accuracy, pressure gauges should siphon filled with water should be installed
preferably be used at an ambient tempera- between the gauge and the process line.
ture between –20 and +150F (–30 to +65C). At Siphons can be supplied with ratings up to
very low temperatures, standard gauges may 4,000 psi. If freezing of the condensate in the
exhibit slow pointer response. Above 150F, the loop of the siphon is a possibility, a diaphragm
accuracy will be affected by approximately seal should be used to isolate the gauge from
1.5% per 100F. Other than discoloration of the the process steam. Siphons should also be
dial and hardening of the gasketing, non-liq- used whenever condensing, hot vapors (not
uid filled type 1279 (phenolic case) and 1379 just steam) are present. Super heated steam
(aluminum case) Duragauges, with standard should have enough piping or capillary line
glass windows, and Duralife gauges types ahead of the siphon to maintain liquid water
1008 and 1009, can withstand continuous in the siphon loop.
operating temperatures up to 250F. Liquid
filled gauges can withstand 200F but glycerin 2.4 Hot or very cold media – A five foot capil-
fill and the acrylic window of Duragauges will lary line assembly will bring most hot or cold
tend to yellow. Silicone fill will have much less process media within the recommended gauge
tendency to yellow. Low pressure, liquid filled ambient temperature range. For media above
types 1008 and 1009 gauges may have some 750F (400C) the customer should use his own
downscale errors caused by liquid fill expan- small diameter piping to avoid possible corro-
sion. This can be alleviated by “burping” the sion of the stainless steel. The five foot capillary
gauge by gently pushing the top fill plug to will protect the gauges used on the common
one side to admit air to the case. cryogenic (less than –300F (200C) gases, liquid
argon, nitrogen, and oxygen. The capillary and
Although the gauge may be destroyed and cal- gauge must be cleaned for oxygen service. The
ibration lost, gauges can withstand short media must not be corrosive to stainless steel,
times at the following temperatures: gauges and must not plug the small bore of the capillary.
with all welded pressure boundary joints,
750F (400C); gauges with silver brazed joints, 2.5 Diaphragm seals – As mentioned above, a
450F (232C) and gauges with soft soldered diaphragm seal should be used to protect
joints, 250F (121C). For expected long term gauges from corrosive media, or media that
service below –20F (–30C) Duragauges and will plug the instrument. Diaphragm seals
41⁄2˝ 1009 gauges should be hermetically are offered in a wide variety of designs and
sealed and specially lubricated; add “H” to the corrosion resistant materials to accommodate
product code for hermetic sealing. Add varia- almost any application and most connections.
tion XVY for special lubricant. Standard Request bulletin OH-1 for details. Recom-
Duralife gauges may be used to –50F (–45C) mended materials for corrosive service may
without modification. be found in the Corrosion Guide available at
www.dresserinstruments.com.
2.2 Accuracy – Heat and cold affect accuracy
of indication. A general rule of thumb for dry 2.6 Autoclaving – Sanitary gauges with
2
clamp type connections are frequently steam and subsequent inaccurate readings. Always
sterilized in an autoclave. Gauges equipped apply torque only to the wrench flats on the
with polysulfone windows will withstand lower seal housing when installing filled,
more autoclave cycles than those equipped diaphragm seal assemblies or removing same
with polycarbonate windows. Gauges from process lines.
equipped with plain glass or laminated safety
glass should not be autoclaved. Gauge 3.4 Process isolation – A shut-off valve
cases should be vented to atmosphere (remov- should be installed between the gauge and
ing the rubber fill/safety plug if necessary) the process in order to be able to isolate the
before autoclaving to prevent the plastic gauge for inspection or replacement without
window from cracking or excessively distort- shutting down the process.
ing. If the gauge is liquid filled, the fill should 3.5 Surface mounting – Also known as wall
also be drained from the case and the front ring mounting. Gauges should be kept free of piping
loosened before autoclaving. strains. The gauge case mounting feet, if applic-
able, will ensure clearance between the pres-
3.0 INSTALLATION sure relieving back and the mounting surface.
3.6 Flush mounting – Also known as panel
3.1 Location – Whenever possible, gauges mounting. The applicable panel mounting
should be located to minimize the effects of cutout dimensions can be found in Ashcroft
vibration, extreme ambient temperatures and sales bulletins – see item 9.4 RESOURCES
moisture. Dry locations away from very high on page 14 of this manual. These dimensions
thermal sources (ovens, boilers etc.) are pre- are also on Ashcroft® gauge general dimension
ferred. If the mechanical vibration level is drawings which can be obtained from the
extreme, the gauge should be remotely locat- Customer Service department in Stratford,
ed (usually on a wall) and connected to the Connecticut.
pressure source via flexible tubing.
3.2 Gauge reuse – ASME B40.1 recommends
that gauges not be moved indiscriminately 4.0 OPERATION
from one application to another. The cumula-
4.1 Frequency of inspection – This is quite
tive number of pressure cycles on an in-service
subjective and depends upon the severity of
or previously used gauge is generally
the service and how critical the accuracy of
unknown, so it is generally safer to install a
the indicated pressure is. For example, a
new gauge whenever and wherever possible.
monthly inspection frequency may be in order
This will also minimize the possibility of a
for critical, severe service applications.
reaction with previous media.
Annual inspections, or even less frequent
3.3 Tightening of gauge – Torque should schedules, are often employed in non-critical
never be applied to the gauge case. Instead, applications.
an open end or adjustable wrench should 4.2 In-service inspection – If the accuracy of
always be used on the wrench flats of the the gauge cannot be checked in place, the user
gauge socket to tighten the gauge into the fit- can at least look for (a) erratic or random
ting or pipe. NPT threads require the use of a pointer motion; (b) readings that are suspect
suitable thread sealant, such as pipe dope or – especially indications of pressure when the
teflon tape, and must be tightened very user believes the true pressure is 0 psig. Any
securely to ensure a leak tight seal. gauge which is obviously not working or indi-
CAUTION: Torque applied to a diaphragm cating erroneously, should be immediately
seal or its attached gauge, that tends to loosen valved-off or removed from service to avoid a
one relative to the other, can cause loss of fill possible pressure boundary failure.

3
4.3 When to check accuracy – Obviously any than 25%. It is likely the bourdon tube has
suspicious behavior of the gauge pointer war- seen significant overpressure leaving
rants a full accuracy check be performed. residual stresses that may be detrimental
Even if the gauge is not showing any symp- to the application.
toms of abnormal performance, the user may
want to establish a frequency of bench type c. Gauges which have accumulated over
inspection. 1,000,000 pressure cycles with significant
pointer excursion.
4.4 When to recalibrate – This depends on the
criticality of the application. If the accuracy of d. Gauges showing any signs of corrosion
a 3-2-3% commercial type gauge is only 0.5% and/or leakage of the pressure system.
beyond specification, the user must decide e. Gauges which have been exposed to high
whether it’s worth the time and expense to temperature or simply exhibit signs of
bring the gauge back into specification. having been exposed to high temperature
Conversely if the accuracy of a 0.25% test – specifically 250°F or greater for soft sol-
gauge is found to be 0.1% out of specification dered systems; 450°F or greater for brazed
then obviously the gauge should be recalibrat- systems; and 750°F or greater for welded
ed. systems.
4.5 Other considerations – These include (a)
f. Gauges showing significant friction error
bent or unattached pointers due to extreme
and/or wear of the movement and linkage.
pressure pulsation; (b) broken windows which
should be replaced to keep dirt out of the g. Gauges having damaged sockets, especially
internals; (c) leakage of gauge fill; (d) case damaged threads.
damage – dents and/or cracks; (e) any signs of
service media leakage through the gauge h. Liquid filled gauges showing loss of case
including its connection; (f) discoloration of fill.
gauge fill that impedes readability. NOTE: ASME B40.1 does not recommend
4.6 Spare parts – As a general rule it is rec- moving gauges from one application to anoth-
ommended that the user maintain in his er. This policy is prudent in that it encourages
parts inventory one complete Ashcroft instru- the user to procure a new gauge, properly tai-
ment for every ten (or fraction thereof) of that lored by specification, to each application that
instrument type in service. arises.

6.0 ACCURACY:
5.0 GAUGE REPLACEMENT PROCEDURES/DEFINITIONS
It is recommended that the user stock one Accuracy inspection – Readings at approxi-
complete Ashcroft instrument for every ten mately five points equally spaced over the
(or fraction thereof) of that instrument type in dial should be taken, both upscale and down-
service. With regard to gauges having a ser- scale, before and after lightly rapping the
vice history, consideration should be given to gauge to remove friction. A pressure standard
discarding rather than repairing them. with accuracy at least 4 times greater than
Gauges in this category include the following: the accuracy of the gauge being tested is rec-
a. Gauges that exhibit a span shift greater ommended.
than 10%. It is possible the bourdon tube Equipment – A finely regulated pressure sup-
has suffered thinning of its walls by corro- ply will be required. It is critical that the pip-
sion. ing system associated with the test setup be
leaktight. The gauge under test should be
b. Gauges that exhibit a zero shift greater
4
positioned as it will be in service to eliminate Span – A span error exists when the error at
positional errors due to gravity. full scale pressure is different from the error
at zero pressure. This error is often propor-
Method – ASME B40.1 recommends that
tional to the applied pressure. Most Ashcroft
known pressure (based on the reading from
gauges are equipped with an internal, adjust-
the pressure standard used) be applied to the
ing mechanism with which the user can cor-
gauge under test. Readings including any
rect any span errors which have developed in
error from the nominal input pressure, are
service.
then taken from the gauge under test. The
practice of aligning the pointer of the gauge Linearity – A gauge that has been properly
under test with a dial graduation and then spanned can still be out of specification at
reading the error from the master gauge intermediate points if the response of the
(“reverse reading”) can result in inconsistent gauge as seen in Figure 1 on page 6 is not lin-
and misleading data and should NOT be used. ear. The Ashcroft Duragauge® is equipped
with a rotary movement feature which per-
Calibration chart – After recording all of the
mits the user to minimize this class of error.
readings it is necessary to calculate the errors
Other Ashcroft gauge designs (e.g., 1009
associated with each test point using the fol-
Duralife®) require that the dial be moved left
lowing formula: ERROR in percent = 100
or right prior to tightening the dial screws.
times (TRUE VALUE minus READING) ÷
RANGE. Plotting the individual errors Hysteresis – Some bourdon tubes have a
(Figure 1 on page 6) makes it possible to visu- material property known as hysteresis. This
alize the total gauge characteristic. The plot material characteristic results in differences
should contain all four curves: upscale – between the upscale and downscale curves.
before rap; upscale – after rap; downscale – This class of error can not be eliminated
before rap; downscale – after rap. Rap means by adjusting the gauge movement or dial
lightly tapping the gauge before reading to position.
remove friction as described in ASME B40.1.
Friction – This error is defined as the differ-
Referring to Figure 1 on page 6, several class- ence in readings before and after lightly tap-
es of error may be seen: ping the gauge case at a check point. Possible
causes of friction are burrs or foreign materi-
Zero – An error which is approximately equal
al in the movement gearing, “bound” linkages
over the entire scale. This error can be mani-
between the movement and the bourdon tube,
fested when either the gauge is dropped or
or an improperly tensioned hairspring. If cor-
overpressured and the bourdon tube takes a
recting these potential causes of friction does
permanent set. This error may often be cor-
not eliminate excessive friction error, the
rected by simply repositioning the pointer.
movement should be replaced.
Except for test gauges, it is recommended
that the pointer be set at midscale pressure to (Continued on page 7)
“split” the errors.

5
TYPICAL CALIBRATION CHART
INDICATED VALUE (PSI)
True Value – Increasing – Increasing – Decreasing – Decreasing –
PSI Without RAP With RAP Without RAP With RAP
0 –.4 0 –.4 0
40 +.8 +1.0 +1.4 +1.1
80 +.4 +.5 +1.2 +1.0
120 –.4 –1.0 +.8 +.6
160 –.8 –.5 +.6 +.4
200 +.4 +.8 +.4 +.4

ERROR (% OF FULL SCALE)


True Value – Increasing – Increasing – Decreasing – Decreasing –
% of Range Without RAP With RAP Without RAP With RAP
0 –.20 0 –.20 0
20 +.40 +.50 +.70 +.55
40 +.20 +.25 +.60 +.50
60 –.20 –.05 +.40 +.30
80 –.40 –.25 +.30 +.20
100 +.20 +.40 +.20 +.20

1.0
Fig. 1




0.5  
  
 
Error (% of Full Scale)


  

0.0 


 



  upscale – without rap
–0.5
  upscale – with rap
  downscale – without rap
  downscale – with rap

–1.0
0 20 40 60 80 100
% of Range

6
6.1 Calibration – Rotary Movement Gauges movement should be rotated counter clock-
– Inspect gauge for accuracy. Many times wise. Rotating the movement one degree
gauges are simply “off zero” and a simple will change this error by approximately
pointer adjustment using the micrometer 0.25%. Rotating the movement often
pointer is adequate. If inspection shows the affects span and it should be subsequently
gauge warrants recalibration to correct span rechecked and readjusted if necessary
and/or linearity errors, proceed as follows: according to step 6.1e and 6.1f.
a. Remove ring, window and, if solid front h. While recalibrating the gauge, the friction
case, the rear closure assembly. error – difference in readings taken with
b. Pressurize the gauge once to full scale and and without rap – should be noted. This
back to zero. error should not exceed the basic accuracy
of the gauge. If the friction error is exces-
c. Refer to Figure 2 on page 8 for a view of a sive, the movement should be replaced.
typical Ashcroft system assembly with One possible cause of excessive friction is
component parts identified. improper adjustment of the hairspring.
The hairspring torque,or tension, must be
d. For solid front gauges, adjust the microme- adequate without being excessive. The
ter pointer so that it rests at the true zero hairspring should also be level, unwind
position. For open front gauges the pointer evenly (no turns rubbing) and it should
and dial must also be disassembled and never tangle.
the pointer should then be lightly pressed
onto the pinion at the 9:00 o’clock position. NOTES:
1 For operation of test gauge external zero
e. Apply full scale pressure and note the reset, refer to Figure 3 on page 8.
magnitude of the span error. With open 2 For test gauge calibration procedure,
front gauges, ideal span (270 degrees) will refer to Figure 4 on page 9.
exist when at full scale pressure the point-
er rests exactly at the 6:00 o’clock position. 6.2 Calibration – 1009 Duralife® Gauge –
Inspect gauge for accuracy. Many times
f. If the span has shifted significantly (span gauges are simply “off zero” and a simple
error greater than 10%), the gauge should pointer adjustment using the adjustable
be replaced because there may be some pointer is adequate. If the inspection shows
partial corrosion inside the bourdon tube that the gauge warrants recalibration to cor-
which could lead to ultimate failure. If the rect span and/or linearity errors, proceed as
span error exceeds 0.25%, loosen the lower follows:
link screw and move the lower end of the
link toward the movement to increase Remove ring, window, gasket and pointer
span or away to decrease span. An adjust- using Ashcroft tool kits 1205T and 1206T.
ment of 0.004 inch will change the span by
6.3 Positive Pressure Ranges –
approximately 1%. This is a repetitive pro-
cedure which often requires more than one a. Remove dial and lightly press pointer onto
adjustment of the link position and the pinion at 9:00 o’clock position.
subsequent rechecking of the errors at zero
b. Apply full scale pressure and rotate span
and full scale pressure.
block as shown in Figure 5a on page 11
g. Apply midscale pressure and note error in until pointer rests at 6:00 o’clock position.
reading. Even though the gauge is accu-
c. Fully exhaust pressure and check that
rate at zero and full scale, it may be inac-
pointer still is at 9:00 o’clock position. If
curate at the midpoint. This is called lin-
not repeat steps 1 and 2 until span is
earity error and is minimized by rotating
correct.
the movement. If the error is positive, the
(Continued on page 10)
7
Fig. 2 ASHCROFT SYSTEM ASSEMBLY

SEGMENT BOURDON TUBE

TIP HAIRSPRING

LINK PINION

ROTARY MOVEMENT
BACKPLATE

SOCKET

Fig. 3 TYPE 1082 EXTERNAL ZERO ADJUST FEATURE*

B RING
A

INSTRUCTIONS FOR USE:


LOOSEN RING LOCKING SCREW “A’’
OBTAIN REQUIRED ADJUSTMENT BY ROTATING
KNOB “B’’ CLOCKWISE OR COUNTER-CLOCKWISE.
TIGHTEN SCREW “A’’ DOWN ON KNOB “B.’’

*Applicable only for test gauge with hinged ring design.


8
Fig. 4 TYPE 1082 TEST GAUGE CALIBRATION PROCEDURE

9
d. Remove pointer and reassemble dial and h. Repeat steps 4 through 8 until span is
dial screws (finger tight). correct.
e. Lightly press pointer onto pinion. i. Remove the pointer.
f. Check accuracy at full scale. If error exceeds j. With 25 inches of Hg vacuum applied,
1% return to step 1, otherwise proceed. reassemble dial, dial screws (finger tight)
and pointer.
g. Check accuracy at midscale. If error exceeds
1% slide dial left or right to compensate. k. Apply 15 inches of Hg vacuum and note
accuracy of indication. If required, slide dial
h. Continue at * below. left or right to reduce error to 1% maximum.
Vacuum range – l. Continue at * below.
a. Using a pencil, refer to dial and mark the
0 and 25 inch of Hg positions on the case
*Now complete calibration of the gauge as
follows:
flange.
a. Firmly tighten dial screws.
b. Remove the dial.
b. Firmly tap pointer onto pinion, using brass
c. Apply 25 inches of Hg vacuum. back-up tool from Ashcroft kit 1205T if
d. Lightly press pointer onto pinion carefully gauge has rear blow-out plug. If gauge has
aligning it with the 25 inch of Hg vacuum top fill hole no back-up is required.
mark on case flange. c. Recheck accuracy at zero, midscale and full
e. Release vacuum fully. scale points (Figures 5a & 5b on page 11).

f. Note agreement of pointer to zero mark on d. Reassemble window, gasket and ring.
case flange.
g. If span is high or low, turn span block as
shown in Figure 5b on page 11.

10
1009 DURALIFE® PRESSURE GAUGE CALIBRATION

1009 DURALIFE CALIBRA


TION

PRESSURE VACUUM

Pencil marks on case flange

Start at 9:00 o’clock

25
25
Span Block

7/64 Open End Wrench


0
Decrease Increase Increase
Decrease

End at 6:00 o’clock

Mid-scale mark

15

25

Full scale mark


Zero box
Dial screws

Fig. 5a Fig. 5b

11
7.0 DIAPHRAGM SEALS filled prior to being placed in service. Ashcroft
diaphragm seal assemblies should only be
7.1 General – A diaphragm seal (isolator) is a filled by a seal assembler certified by Dresser
device which is attached to the inlet connec- Industries. Refer to section 3.3 for a caution-
tion of a pressure instrument to isolate its ary note about not applying torque on either
measuring element from the process media. the instrument or seal relative to the other.
The space between the diaphragm and the 7.3 Operation – All Ashcroft diaphragm seals,
instrument’s pressure sensing element is with the exception of Type 310 mini-seals, are
solidly filled with a suitable liquid. continuous duty. Should the pressure instru-
Displacement of the liquid fill in the pressure ment fail, or be removed accidentally or delib-
element, through movement of the erately, the diaphragm will seat against a
diaphragm, transmits process pressure matching surface preventing damage to the
changes directly to a gauge, switch or any diaphragm or leakage of the process fluid.
other pressure instrument. When diaphragm
seals are used with pressure gauges, an addi- 7.4 Maintenance – Clamp type diaphragm
tional 0.5% tolerance must be added to the seals – Types 100, 200 and 300 – allow for
gauge accuracy because of the diaphragm replacement of the diaphragm or diaphragm
spring rate. capsule, if that ever becomes necessary. The
Type 200 top housing must also be replaced
Used in a variety of process applications with the diaphragm. With all three types the
where corrosives, slurries or viscous fluids clamping arrangement allows field disassem-
may be encountered, the diaphragm seal bly to permit cleaning of the seal interior.
affords protection to the instrument where:
7.5 Failures – Diaphragm failures are gener-
• The process fluid being measured would ally caused by either corrosion, high tempera-
normally clog the pressure element. tures or fill leakage. Process media build-up
• Pressure element materials capable of on the process side of the diaphragm can also
withstanding corrosive effects of certain require seal cleaning or replacement. Consult
fluids are not available. Customer Service, Stratford CT for advice on
seal failures and/or replacement. Refer also to
• The process fluid might freeze due to Product Information page ASH/PI-14C con-
changes in ambient temperature and dam- taining drawing 96A121 Corrosion Data
age the element. Guide.
7.2 Installation – Refer to sales bulletin OH-1 WARNING: All seal components should be
for information regarding (a) seal configura- selected considering process and ambient
tions; (b) filling fluids; (c) temperature range operating conditions to prevent misapplica-
of filling fluids; (d) diaphragm material pres- tion. Improper application could result in fail-
sure and temperature limits; (e) bottom hous- ure, possible personal injury, property
ing material pressure and temperature lim- damage or death.
its; (f) pressure rating of seal assembly; (g)
accuracy/temperature errors of seal assembly;
(h) diaphragm seal displacement. The volu-
8.0 DAMPENING DEVICES
metric displacement of the diaphragm must
8.1 General – Some type of dampening device
at least equal the volumetric displacement of
should be used whenever the pressure gauge
the measuring element in the pressure
may be exposed to repetitive pressure fluctu-
instrument to which the seal is to be attached.
ations that are fairly rapid, high in magni-
It is imperative that the pressure instru- tude and especially when transitory pressure
ment/diaphragm seal assembly be properly spikes exceeding the gauge range are present
(as with starting and stopping action of
12
valves and pumps). A restricted orifice of restrict the flow of glycerin through an orifice
some kind is employed through which pres- and thus provide a degree of dampening. The
sure fluctuations must pass before they reach greater the compression of the felts the
the bourdon tube. The dampener reduces the greater the degree of dampening.
magnitude of the pressure pulse thus extend-
ing the life of the bourdon tube and move- 8.4 Ashcroft Pulsation Dampener – Type
ment. This reduction of the pressure pulsa- 1106 Ashcroft Pulsation Dampener is a mov-
tion as “seen” by the pressure gauge is gener- ing pin type in which the restricted orifice is
ally evidenced by a reduction in the pointer the clearance between the pin and any one of
travel. If the orifice is very small the pointer five preselected hole diameters. Unlike a sim-
may indicate the average service pressure, ple throttle screw/plug, this device has a self-
with little or no indication of the time varying cleaning action in that the pin moves up and
component of the process pressure. down under the influence of pressure fluc-
tuations.
Commonly encountered media (e.g. – water
and hydraulic oil) often carry impurities 8.5 Ashcroft Pressure Snubber – The heart
which can plug the orifice over time thus ren- of the Type 1112 pressure snubber is a thick
dering the gauge inoperative until the damp- porous metal filter disc. The disc is available
ener is cleaned or replaced. in four standard porosity grades.

Highly viscous media and media that tend to 8.6 Campbell Micro-Bean® – Type 1110
periodically harden (e.g., asphalt) require a Micro-Bean is a precision, stainless steel, nee-
diaphragm seal be fitted to the gauge. The dle valve instantly adjustable to changing
seal contains an internal orifice which damp- conditions of flow and viscosity. A very slight
ens the pressure fluctuation within the fill taper on the valve stem fits into tapered hole
fluid. in the body. The degree of dampening is easi-
ly adjusted by turning the valve handle. A fil-
8.2 Throttle Screws & Plugs – These acces- ter is built into the Micro-Bean to help pre-
sories provide dampening for the least cost. vent plugging.
They have the advantage of fitting completely
within the gauge socket and come in three 8.7 Ashcroft Needle Valves – Type 7001 thru
types: (a) a screwed-in type which permits 7004 steel needle valves provide varying
easy removal for cleaning or replacement; (b) degrees of dampening similar to the Campbell
a pressed in, non-threaded design and (c) a Micro-Bean but with a less precise and less
pressed in, threaded design which provides a costly adjustability. Like the Micro-Bean
highly restrictive, helical flow path. Not all these devices, in the event of plugging, can
styles are available on all gauge types. easily be opened to allow the pressure fluid to
clear away the obstruction.
8.3 Ashcroft Gauge Saver – Type 1073
Ashcroft Gauge Saver features an elastomer- 8.8 Chemiquip® Pressure Limiting Valves –
ic bulb that fully isolates the process media Model PLV-255, PLV-2550, PLV-5460, PLV-
from the bourdon tube. In addition to provid- 5500 and PLV-6430, available with and with-
ing dampening of pressure pulses, the bour- out built-in snubbers, automatically “shut off”
don tube is protected from plugging and cor- at adjustable preset values of pressure to pro-
rosion. The space between the bulb and bour- tect the gauge from damage to overpressure.
don tube is completely filled with glycerin. They are especially useful on hydraulic sys-
Felt plugs located between the bulb and bour- tems wherein hydraulic transients (spikes)
don tube are first compressed some amount to are common.

13
9.0 RESOURCES
9.1 Training Videos
9.1.1 Test gauge calibration
9.1.2 1009 Duralife® calibration
9.1.3 Duragauge® calibration
9.1.4 Diaphragm seal filling

9.2 Pressure Instrument Testing Equipment


9.2.1 Type 1305D Deadweight Tester
9.2.2 Type 1327D Pressure Gauge Comparator
9.2.3 Type 1327CM “Precision” Gauge Comparator

9.3 Tools & Tool Kits


9.3.1 Type 2505 universal carrying case for 1082 test gauge
9.3.2 Type 266A132-01 span wrench for 1082 test gauge
9.3.3 Type 1280 conversion kit for 41⁄2˝ lower connect 1279/1379
9.3.4 Type 1283 conversion kit for 41⁄2˝ back connect 1279/1379
9.3.5 Type 1284 conversion kit for 6˝ lower & back connect
9.3.6 Type 1281 socket O-Ring kit for 1279/1379 lower connect
9.3.7 Type 1285 41⁄2˝ ring wrench for 1279/1379 lower & back connect
9.3.8 Type 1286 6˝ ring wrench for 1379 lower & back connect
9.3.9 Type 1287 cone tool for installing diaphragm & spring on 1279/1379 back connect
9.3.10 Type 1105T calibration tool kit (all gauges except 1009 Duralife®)
9.3.11 Type 3220 pointer puller (all gauges except 1009 Duralife®)
9.3.12 Type. 3530 pinion back-up tool for 1009 Duralife®
9.3.13 Type 1230 throttle plug insertion (1⁄4 NPT) for 1009 Duralife®
9.3.14 Type 1231 throttle plug insertion (1⁄2 NPT) for 1009 Duralife® (body only)
9.3.15 Type 1205T calibration hand tools for 1009 Duralife®
9.3.16 Type 1206T ring removal & assembly tools for 1009 Duralife®

9.4 Sales Bulletins


9.4.1 Pressure Instrument Testing Equipment – Bulletin TE-1
9.4.2 Type 1327CM Portable Precision Gauge Comparator – Bulletin TE-2
9.4.3 Test Gauges – Bulletin TG-2
9.4.4 Duragauge® Pressure Gauges – Bulletin DU-1
9.4.5 General Service Gauges – Bulletin IG-1
9.4.6 Type 1009 Duralife® Industrial Gauges – Bulletin SS-1
9.4.7 Type 1008 Metric Case Gauges – Bulletin SS-1
9.4.8 Duralife® Metric Process Gauges – Bulletin SS-1
9.4.9 Type 1032 Sanitary Pressure Gauges – Bulletin SG-2
9.4.10 Special Service Gauges – Bulletin IG-1
9.4.11 Diaphragm Seals – Bulletin OH-1

14
Instrument Division Sales and Customer Service Locations

U.S. & International U.S. Sales Offices International Operations Japan


Headquarters Dresser Japan Ltd.
Midwest Region Brazil Room 818, Shin Tokyo Building
Stratford, Connecticut Chicago, Illinois Dresser Industria e 3-1 Marunouchi 3-Chome,
250 E. Main Street 400 W. Lake Street Comercio Ltda. Chiyoda-ku, Tokyo, Japan
Stratford, CT 06614-5145 Suite 318 Rua Senador Vergueiro #433 Tel: 813-3201-1501
Tel: (203) 378-8281 Roselle, IL 60172-3573 09521-320 Sao Caetano do Sul Fax: 813-3213-6567
Fax: (203) 381-9042 Tel: (630) 980-9030 Sao Paulo, Brazil E-Mail: yuichi.yamamoto@
Fax: (630) 980-9440 Tel: 55-11-4224-7400 dresserjapan.co.jp
Fax: 55-11-4224-7477
Southeast Region E-Mail: vendas.instrumentos@ Korea
Mobile, Alabama dresser.com Dresser International, S.A
605 Bel Air Blvd. Room 2107
Suite 10 Brazil (Jacarei) #2015 Kuk Dong Bldg.
Mobile, AL 36606 Dresser Industria e Comercio Ltda. 60-1, 3-KA, Choongmu-Ro,
Tel: (251) 473-1692 Divisao Masoneilan Chung-ku, Seoul, Korea 100-705
Fax: (251) 473-1782 Rua Particular– Estrada Tel: 82-2-2274-0792
Velha Rio De Janeiro – Fax: 82-2-2274-0794
Northeast Region Sao Paulo, KM 101Jacarei, E-Mail: dkisjlee@chollian.net
Stratford, Connecticut Sao Paulo Caixa
250 E. Main Street Postal 167, CEP 12305-330 Mexico
Stratford, CT 06614-5145 Tel: 55-11-3958-2011 Dresser Instruments S.A. De C.V.
Tel: (203) 385-0670 Fax: 55-11-3958-2670 Mexico Operations
Fax: (203) 385-0756 E-Mail: dresserjac@uol.com.br Henry Ford No. 114
Esq. Foulton Fracc.
Pacific Region Canada Industrial San Nicolas
Stratford, Connecticut Dresser Canada, Inc. 54030 Tlalnepantla,
250 E. Main Street 2135 Meadowpine Blvd. Edo De Mexico
Stratford, CT 06614-5145 Mississauga, Tel: (52)55-53-10-72-17
Tel: (203) 385-0399 Ontario L5N 6L5 Canada (52)55-53-10-89-83
Fax: (203) 385-0402 Tel: 905-335-3529 (52)55-53-10-28-29
Fax: 905-826-9106 (52)55-53-10-28-75
E-Mail: Lance_Barette@dresser.com Fax: (52)55-53-10-26-08
E-Mail: mendiet@avantel.net
China
Dresser Industries, Inc. Saudi Arabia
Room #2404, Capital Mansion Dresser Al Rushaid Valve &
No. 6 Xin Yuan Nan Road Beijing, Instrument Co. (DARVICO)
People’s Republic of China 100004 P.O. Box 10145
Tel: 86-10-84862440/1/2/3/4 Jubail Industrial City
Fax: 86-10-84862445 Saudi Arabia 31961
E-Mail: dresser@public3.bta.net.cn Tel: 966-3-341-0278
Fax: 966-3-341-7624
France E-Mail: bill_dumasia@darvico.com
Dresser Europe GmbH E-Mail: sam_dastur@darvico.com
Dresser Instrumentation
206 ZA Le Mandinet Singapore
Rue des Campanules Dresser Singapore
F 77185 Lognes, France Instrument Operations
Tel: 33-1-60372530 Block 1004 Toa Payoh North
Fax: 33-1-60372539 #07-15/17 Singapore 318995
E-Mail: europe@wanadoo.fr Tel: 65-6252-6602
Fax: 65-6252-6603
Germany E-Mail: john.wong@dresser.com.sg
Dresser Europe GmbH
Postfach 11 20 Max-Planck-Str. 1 United Kingdom
D-52499 Baesweiler, Germany Dresser Europe GmbH
Tel: 49-24-01-8080 East Gillibrands, Skelmersdale
Fax: 49-24-01-7027 Lancashire, WN8 9TU
E-Mail: jbiermans@dresserbae.de United Kingdom
Tel: 44-16-95-52600
Germany Fax: 44-16-95-52693
Ebro Electronic GmbH E-Mail: sales@dresser-instrument.co.uk
Peringerstr 10D-85055
Ingolstadt, Germany Venezuela
Tel: 49-84-1-95478-0 Manufactures Petroleras
Fax: 49-84-1-95478-80 Venezolanas (M.P.V.)
E-Mail: info@ebro.de KM 7 Carretera A El Mojan Calle 18,
#15B355 ZONA Ind. Norte Sector
Canchancha Maracaibo Edo
Zulia Venezuela
Tel: 58-261-757-9070
Visit our web site www.ashcroft.com Fax: 58-261-757-9561
E-Mail: contactenos@mapvensa.com
© Dresser, Inc., Dresser Instruments, 250 East Main Street, Stratford, CT 06614, E-Mail: ventasmpv@telcel.net.ve
203-378-8281, 203-378-0499 Fax, www.ashcroft.com
All sales subject to standard terms and conditions of sale. I&M012-10098-5/02 (250-1353H) GBE 5M DPS 4P7/2003

You might also like