Microplate Reader
Service Manual
copyright
© 2005 Shenzhen Mindray Bio-medical Electronics Co., Ltd. All rights Reserved.
For this Service Manual, the issued Date is 2005-12 (Version: 1.0).
Mindray) owns the intellectual property rights to this Mindray product and this manual. This
manual may refer to information protected by copyrights or patents and does not convey any
license under the patent rights of Mindray, nor the rights of others. Mindray does not assume
any liability arising out of any infringements of patents or other rights of third parties.
Disclosure of the information in this manual in any manner whatsoever without the written
manual in any manner whatsoever without the written permission of Mindray is strictly
forbidden.
and other countries. All other trademarks that appear in this manual are used only for
editorial purposes without the intention of improperly using them. They are the property of
All information contained in this manual is believed to be correct. Mindray shall not be liable
for errors contained herein nor for incidental or consequential damages in connection with the
furnishing, performance, or use of this manual.
Mindray is responsible for safety, reliability and performance of this product only in the
condition that:
-I-
• all installation operations, expansions, changes, modifications and repairs of this product
are conducted by Mindray authorized personnel;
• the electrical installation of the relevant room complies with the applicable national and
local requirements;
• the product is used in accordance with the instructions for use.
Upon request, Mindray may provide, with compensation, necessary circuit diagrams,
calibration illustration list and other information to help qualified technician to maintain and
Note
This equipment is to be operated only by medical professionals trained and authorized by
Warning
It is important for the hospital or organization that employs this equipment to carry out a
reasonable service/maintenance plan. Neglect of this may result in machine breakdown or
injury of human health.
Warranty
THIS WARRANTY IS EXCLUSIVE AND IS IN LIEU OF ALL OTHER WARRANTIES,
EXPRESSED OR IMPLIED, INCLUDING WARRANTIES OF MERCHANTABILITY OR
FITNESS FOR ANY PARTICULAR PURPOSE.
Exemptions
Mindray's obligation or liability under this warranty does not include any transportation or
other charges or liability for direct, indirect or consequential damages or delay resulting from
the improper use or application of the product or the use of parts or accessories not approved
by Mindray or repairs by people other than Mindray authorized personnel.
- II -
Return Policy
Return Procedure
In the event that it becomes necessary to return this product or part of this product to Mindray,
the following procedure should be followed:
1. Obtain return authorization: Contact the Mindray Service Department and obtain a
Customer Service Authorization (Mindray) number. The Mindray number must appear on
the outside of the shipping container. Returned shipments will not be accepted if the
Mindray number is not clearly visible. Please provide the model number, serial number,
and a brief description of the reason for return.
2. Freight policy: The customer is responsible for freight charges when this product is
shipped to Mindray for service (this includes customs charges).
3. Return address: Please send the part(s) or equipment to the address offered by
Company Contact
Manufacture: Shenzhen Mindray Bio-Medical Electronics Co., Ltd.
Address: Mindray Building, Keji 12th Road South, Hi-tech Industrial
Park, Nanshan, Shenzhen, P.R.China,518057
Phone: +86 755 26582479 26582888
- III -
Safety Precautions
1. Meaning of Signal Words
NOTE are used regarding safety and other important instructions. The signal words and
their meanings are defined as follows. Please understand their meanings clearly before
The labels attached to some MR-96 panels use symbols with the text to clarify the meaning of
the text. The table below explains the symbols on the labels.
Serial Number
Manufacturer
Date of Manufacture
Graphics
All graphics, including screens and printout, are for illustration purposes only and must not be
- IV -
Table of Contents
I. Overview .......................................................................................................................................... 1
IV. Troubleshooting.............................................................................................................................. 4
1. System Description:
The MR-96 Microplate Reader is an 8-channel optical measuring system, which can be used
to measure the density of vertical light. It is a specialized analyzer to take readings and
With the advanced high speed microprocessor and man-machine conversation design on
the software, the MR-96 makes the operation easier and more convenient, result calculation
2. Technical Specifications
2) Wavelength range: 405 nm, 450 nm, 492 nm, 630nm. Additional 4 free positions for optional
filters
8) Stability: ≤ 0.005A/10min
9) Reading Speed: Single wavelength <20s/plate (96-well plate); dual wavelength <35s/plate
(96-well plate).
10) Vibrancy: Vibrant Speed (High, Normal, Low); Vibrant Time (0 - 60s).
11) Power: 100VA
Photoelectric Sample
Test Platform 8 Fibers Conversion Motor
Compartment
The optical coupler is with dust Remove the dust and replace the
blocked
Sound is abnormal during Foreign matter exists Remove the foreign matter
local agency
1) After replacing the lamp, you should clean it, lock the retaining nuts and adjust W201 to set the
2) When adjusting the reading range, touch the Date and Time button and enter “985”, then
exit. Open the “985” menu and adjust W103 to set the voltage of four channels to 0.4V - 1.24V.
When replacing the filter, you should proceed with a pair of tweezers to prevent from dust and
moisture. Open the “985” menu and adjust the attenuator to set the voltage of four channels to
0.4V - 1.24V.
2) Open the “985” menu and adjust W103 to set the voltage of four channels to 0.4V - 1.24V.
After replacing the photoelectric conversion board, you should adjust the reading range. Open
the “985” menu and adjust W103 to set the voltage of four channels to 0.4V - 1.24V, or
reconfigure a filter.
1. The power supply should comply with requirements of the device. A regulated power supply or
an UPS should be added as you need to avoid unstable results and system down.
2. The device should be free of direct sunlight and dust. Keep good ventilation and add a dust
shield as needed.
3. For the sake of the lamp’s lifespan, you should wait for more than 10 minutes to avoid
unreliable results after turning on the lamp. Moreover, the interval between turning on/off the
lamp should be longer than 5 minutes, otherwise the lamp will probably be damaged.
4. When starting up the device everyday, you should check the filters before measuring. The
MR-96 reads the air blank absorbance of each filter in the 8 channels, which is used to check
the photometric system. The default absorbance of the 405 - 630 filters is less than 0.500.
However, it may increase gradually in the result of dust, humidification and moldiness when
the device has been used for a long time. If the absorbance is greater than 1.000, you should
clean the photometric system or replace the accessories like the filters.
5. During operation, pay much attention to the patient serum and waste. Every department
6. You are obligated to observe your local or national guidelines for biohazard waste disposal.
The MR-96 Microplate Reader is a precision optical measuring system. Therefore, it’s
1) Keeping the operating environment dry and clean. Covering the sample compartment after
2) Keeping the mechanical and electrical parts of the device away from water and reagent.
3) Switching off the power supply while cleaning the surface and screen of the device. Be sure