Professional Documents
Culture Documents
Excellence in
Cataract Surgery
A Step-by-Step Approach
Edited by
D. Michael Colvard, MD, FACS
Clinical Professor of Ophthalmology
Doheny Eye Institute
Keck School of Medicine
University of Southern California
Los Angeles, CA
Medical Director of Colvard Eye Center
Encino, CA
This project made possible through a grant from AMO.
Cover image of the Advanced Medical Optics ZCB00 single-piece acrylic IOL and the MicroSurgical
Technologies bimanual irrigation/aspiration system courtesy of D. Michael Colvard, MD, FACS.
All rights reserved. No part of this book may be reproduced, stored in a retrieval system, or transmitted
in any form or by any means, electronic, mechanical, photocopying, recording, or otherwise, without
written permission from the editor, D. Michael Colvard, MD, FACS, except for brief quotations em-
bodied in critical articles and reviews.
The procedures and practices described in this book and the companion DVD should be implemented
in a manner consistent with the professional standards set for the circumstances that apply in each
specific situation. Every effort has been made to confirm the accuracy of the information presented and
to correctly relate generally accepted practices. The authors and editor cannot accept responsibility
for errors or exclusions or for the outcome of the material presented herein. There is no expressed or
implied warranty of this book or information imparted by it. Care has been taken to ensure that surgi-
cal techniques, drug selection, and dosages are in accordance with currently accepted/recommended
practice. Due to continuing research, changes in government policy and regulations, and various effects
of drug reactions and interactions, it is recommended that the reader carefully review all materials and
literature provided for each drug, especially those that are new or not frequently used.
Chapter 2: Incisions......................................................................................................................................... 11
D. Michael Colvard, MD, FACS
Chapter 3: Capsulorrhexis............................................................................................................................... 19
Howard V. Gimbel, MD, MPH, FACS, FRCSC
operating microscope and improvements in hand-held Topical anesthesia was reintroduced to modern
aspiration-irrigation systems, standard extracapsular cataract surgery by Richard Fichman in the late 1990s6
surgery became much “cleaner.” This reduction in and, to the surprise of most ophthalmologists, both
postoperative inflammation with the new standard retrobulbar and peribulbar injections of anesthetics
extracapsular methods and the maintenance of a pos- were found to be largely unnecessary with the new
terior capsule that provided support for Shearing’s new surgical techniques.
posterior chamber IOL suddenly made extracapsular Innovations by Michael McFarland,7 Paul Ernest,8
surgery much more appealing to surgeons. and Howard Fine9 in incision construction have allowed
Shearing’s innovation was a change that made a us to design better, stronger incisions, some of which
huge difference in ophthalmology. Cataract surgery no longer require sutures.
skyrocketed in popularity with the introduction of the The can opener capsulotomy, the age-old main-
posterior chamber IOL. Like phacoemulsification, pos- stay of extracapsular surgery, was replaced by Howard
terior chamber lenses were initially rejected by many as Gimbel and Thomas Neuhann with the continuous
inherently dangerous. The great fear was that posterior curvilinear capsulotomy in the early 1990s.10 This
chamber IOLs would ultimately float in the vitreous innovation resulted in greater stability of the capsular
cavity, but as positive experience with the new lenses bag during phacoemulsification and improved the cen-
increased and long-term success became established, tration of IOLs postoperatively, but its introduction
this new IOL technology was universally embraced. led to other challenges. The nucleus of a dense cataract
By the mid-1980s, there were two camps of poste- could no longer be tipped easily into the pupillary plane
rior chamber lens users: those who performed standard for emulsification. New techniques had to be devel-
extracapsular surgery and inserted the lenses through oped for disassembling the nucleus within the capsular
an 8- to 11-mm incision, and those who performed bag. Howard Gimbel11 and John Shepherd12developed
phacoemulsification and then enlarged the phaco inci- “divide and conquer” techniques in the early 1990s,
sion to 6 mm in order to implant the IOL. All IOLs followed soon by Kunihiro Nagahara13 who intro-
at this time were made of polymethylmethacrylate duced the first of the many chopping techniques now
(PMMA). Conventional extracapsular surgeons in the used widely by surgeons all over the world. These
mid-1980s were very much in the majority. They saw techniques, which all require the ability to move the
little advantage in adopting phacoemulsification with nucleus within the capsular bag without placing undue
all of its inherent difficulties and challenges as long as stress on the zonular support of the capsule, were made
the incision needed to be enlarged for IOL insertion. possible by the development of hydrodissection and
Thomas Mazzocco changed this. The “Mazzocco hydrodelineation techniques pioneered by Aziz Anis14
Taco,” a plate-shaped IOL made of silicone, was the and Howard Fine.15
first IOL designed to be rolled and inserted through an IOL technology has continued to advance with
incision smaller than 6 mm.4 Many surgeons believed lens edge configurations that delay the onset of capsu-
only in the sanctity of PMMA and doubted that a lar opacification and lens optics that improve the qual-
foldable material such as silicone would remain bio- ity of vision through aspheric design. Capsular tension
logically inert or that it would remain clear over the rings now reduce the risks of capsular destabilization
course of time. Fortunately, the detractors, once again, during our most challenging cases.
were wrong. With the eventual development of a fold- Last, but certainly not least, phacoemulsification
able IOL that could be placed through an unenlarged technology has constantly improved with innova-
3-mm phaco incision, the advantages of small incision tions in fluidics, power control, and duty cycles.
surgery were finally realized. These improvements in phacoemulsification provide
The foundations were now in place for the steady the surgeon with a level of control and safety that even
evolution in materials and techniques that have made Charles Kelman could not have imagined. What comes
phacoemulsification one of the safest and most elegant next? What will be the direction of change that makes
procedures in medicine today. surgery safer, easier, more reliable, and more efficient?
David Miller and Roger Stegmann introduced Thirty-five years ago, it would have been impos-
sodium hyaluronate to ophthalmology in the late sible for any one person to guess where the collective
1970s5 and over the past three decades continuous genius of a generation of eye surgeons was about to take
improvements in ophthalmic viscoelastic devices have us. No one knew then what the future would bring, and
added greatly to the safety of cataract surgery. no one knows now. Only one thing is certain. Those of
Preface xv
you who are just beginning to learn phacoemulsifica- Intraocular Implantation; Cannes, France; 1979.
tion today will be part of that future. You are the next 6. Fichman RA. Use of topical anesthesia alone in cataract sur-
gery. J Cataract Refract Surg. 1996;22:612-614.
generation of innovators. Your challenge is to resist the
7. McFarland MS. Surgeon undertakes phaco, foldable IOL se-
notion that everything worthwhile has been discov- ries sans sutures. Ocular Surg News. 1990;8(5):1,15.
ered, that all the obstacles have been surmounted, and 8. Ernest PH, Lavery KT, Kiessling LA. Relative strength of
that there is nothing left to do. You are our future and scleral corneal and clear corneal incisions constructed in ca-
there will be much left for you to do. daver eyes. J Cataract Refract Surg. 1994;20:626-629.
9. Fine IH. Self-sealing corneal scleral tunnel incision for small-
incision cataract surgery. Ocular Surg News. 1992;May 1.
References 10. Gimbel HV, Neuhann T. Development, advantages, and
methods of the continuous curvilinear capsulorhexis. J Cata-
1. Kelman CD. Phacoemulsification and aspiration—a new
technique of cataract removal: a preliminary report. Am J ract Refract Surg. 1991;17:110-111.
Ophthalmol. 1967;64:23-35. 11. Gimbel HV. Trough and crater divide and conquer nucleo-
2. Kratz RP, Colvard DM. Kelman phacoemulsification in the fractis techniques. Euro J Implant Refract Surg. 1991;3:123-126.
posterior chamber. J Cataract Refract Surg. 1979;86:1983-1984. 12. Shepherd JR. In situ fracture. J Cataract Refract Surg.
3. Shearing SP. Mechanism of fixation of the Shearing poste- 1990;16:436-440.
rior chamber intra-ocular lens. Contact Intraocular Lens Med J. 13. Nagahara K. Phaco chop film. Presented at: International
1979;5:74-77. Congress on Cataract, IOL, and Refractive Surgery. Ameri-
4. Mazzocco T. 6 mm optic for a 3 mm wound. American Intra- can Society of Cataract and Refractive Surgeons; Seattle,
ocular Lens Society, US Intraocular Lens Symposium; New WA; May 1993.
Orleans, LA; March 1983. 14. Anis A. Understanding hydrodelineation: the term and re-
5. Balazs EA, Miller D, Stegmann R. Viscosurgery and the use lated procedures. Ocular Surg News. 1991;9:134-137.
of Na Hyaluronate in intraocular lens implantation. Present- 15. Fine IH. Cortical cleaving hydrodissection. J Cataract Refract
ed at: the International Congress and First Film Festival on Surg. 1992;18:508-512.
Foreword
This excellent text serves an important need in tury were typically beautifully illustrated testaments to
furthering the “science” of cataract surgery. The his- the surgeon-author’s skill and dexterity, but references
tory of progress in surgery, in general, contrasts in to the peer-reviewed literature, data, and statistical
many ways with the rest of medicine. The initial “bar- analysis supporting the author’s assertions were com-
ber surgeons” of England were looked down upon by monly minimal or altogether absent.
the elitist and self-declared medical intellectuals who Fortunately the field has evolved, and the term
called themselves physicians. In contrast to the per- surgical science is no longer an oxymoron. Prospective
ception of surgery as a crude assault on the body, the controlled trials comparing surgical interventions and
tools of the physicians included observation, dietary devices are no longer rare, and the claims from the
manipulation, pharmacologic therapy, and scientific podium of charismatic surgical “thought leaders” are
study. Surgeons learned their crafts via apprenticeships no longer routinely accepted as valid substitutes for
and accumulated anecdotal experience, but “medicine” objective data.
was a science. At the same time, our society is tasking surgeons in
This legacy persisted for quite a while. Prospective, general and ophthalmologists in particular with figur-
randomized, controlled clinical trials became routine in ing out how to do more surgery, with better outcomes
the evaluation of new proposed medical therapies. But and at lower costs. The looming demographic tidal
rarely was this methodology embraced by surgeons, wave of the baby boomer generation has led ophthal-
who would declare preeminence of their own surgical mic manpower studies to predict a 30% undersupply
techniques after reporting small case series in which of ophthalmologists within a decade or two in the
no control group was included. In the United States, a United States. The prevalence of cataracts and other
group of ophthalmologists actually sued in an effort to age-related eye diseases will increase dramatically; the
prevent the National Institutes of Health from carry- number of ophthalmic surgeons will not change appre-
ing out a prospective study of one eye operation. ciably. Not content to simply see our profession deal
Early in my own career, it was common to hear with this volume, our society demands that we reduce
interesting expressions from surgeons such as “in my the cost of this care, improve the results (eg, eliminate
hands.” In at least some cases, this was a mechanism the need for corrective eye wear for distance, near,
for explaining away a lack of replication of claimed and intermediate vision postoperatively after cataract
outcomes by other surgeons or medical centers. I have surgery), and reduce the risks of endophthalmitis and
witnessed surgeon innovators ridiculing surgeons in other complications. In short, ophthalmic surgeons will
the audience who described complications after try- need to do more with less.
ing the new surgical procedure, complications that We are also asked to change how we transform
the innovator claimed could never occur. “Perhaps you new ophthalmology residents into capable surgeons.
should go back and repeat your residency if you cannot The apprenticeship model of “see one, do one, teach
perform a simple operation,” said one guest lecturer to one” is being replaced by a more rigorous approach
a skilled local ophthalmic surgeon in California who of communicating the underlying scientific principles
did not see the same uniformly wonderful results in his of surgery, breakdown of multistep procedures into
patients. With the passage of time, it has become clear their component parts, and “certification” of trainees
to me that in every case the observant practitioner was as having mastered each of these steps. Pedagogical
correct, and the indignant surgeon-innovator was too scientists tell us that this will accelerate the progress
personally invested in his or her work to be objective. of new surgeons, more quickly identify strengths and
Could we give credence to an internist claiming that weaknesses of budding surgeons so that deficiencies
a drug works better “in my hands” than in those of can be quickly corrected, reduce the likelihood of
another internist, or that an internist in the audience complications during the early part of the learning
who reports an adverse event from a drug must be curve, and ensure society that the new surgeons we
incompetent? Surgical texts for most of the 20th cen- train possess the required competencies.
xviii Foreword
This text reflects the positive trends in how we are details of surgical technique before performing those
coming to embrace the science of ophthalmic surgery. techniques on their first patients, as well as for more
The physics that drives our cataract surgical instru- experienced surgeons looking to continuously improve
ments, the detailed exploration of techniques for each the outcomes for their patients.
step of the procedure, the optics of vision correction,
and the outcomes data that speak to the quality of our Peter J. McDonnell, MD
interventions are all beautifully illustrated. I believe Director of The Wilmer Eye Institute
this will prove a valuable resource for beginning sur- Johns Hopkins University School of Medicine
geons who will want to immerse themselves in the Baltimore, MD
Introduction
Modern cataract surgery is one of medicine’s fin- Phacoemulsification is unforgiving. If there is a stumble
est achievements. No procedure today is more gentle, on any one step, the next step becomes more difficult
safe, and successful; more important to the quality of and things begin to go badly. When each sequential
life and well-being of patients; or more beneficial to step is performed well, however, the procedure seems
society as whole than is phacoemulsification. The pro- to glide effortlessly and a magical thing occurs. The
cedure is also a marvel to behold. It is an art form, and, surgeon’s hands create something that is not only good
once learned well, it is a joy to perform. but lovely to behold.
In the hands of a skilled surgeon, phacoemulsifica- More than a dozen of the world’s finest surgeons
tion is a masterful ballet of efficiency and grace. Each have come together in this text to help you learn to
precise and carefully practiced step leads fluidly to the perform phacoemulsification at its highest level. Each
next. For a number of years, I have had the pleasure of has chosen one or more aspects of the procedure and
training a wonderful group of young resident surgeons has carefully analyzed the steps that are critical to the
in phacoemulsification. This experience as a teacher successful performance of that part of the surgery.
and my own 30 years as a phaco surgeon have taught Whenever it is useful, the authors have also provided
me the value of understanding phacoemulsification as narrated video footage that illustrates the key instruc-
a series of thoughtfully considered steps. Each step of tional points made in the text. This complete video
phacoemulsification must be understood thoroughly, reference should prove to be an invaluable resource as
learned perfectly, and practiced repeatedly before the you learn to achieve excellence phacoemulsification.
procedure can be executed with consistent proficiency.
Chapter
1
Local Anesthesia for
Cataract Surgery
Rom Kandavel, MD
In recent years, advances in cataract surgery have Intraocular pressure may be elevated after the
lead to greater levels of refractive precision, faster injection of even modest amounts of anesthetic into
visual rehabilitation, and improved comfort and safety. the orbit. The orbit has an average volume of 30 cc. A
Refinements in phacoemulsification techniques and sudden increase in orbital volume associated with the
intraocular lens (IOL) technology deserve much of the injection of anesthetic results in the transmission of
credit for these advances, but innovations in anesthe- force anteriorly, causing compression of the globe.
sia, especially topical anesthesia, have also played an The floor of the orbit is the shortest of the orbital
important role in improving outcomes and hastening walls and extends only 35 to 40 mm from the orbital
visual recovery. rim. The 38-mm needle used in retrobulbar anesthesia,
While topical anesthesia is favored by many sur- therefore, has the potential to damage the optic nerve
geons for the majority of their cases today, proper in a significant percentage of the population.1
patient screening and careful preoperative planning The abducens, oculomotor, and nasociliary nerves
are necessary in order to choose the best anesthesia pass through the annulus of Zinn. The trochlear nerve
for an individual patient. Mastery of all of the avail- enters outside of the annulus to supply the superior
able techniques—intracameral, topical, parabulbar oblique. Placement of anesthetic within the intramus-
(sub-Tenon’s), peribulbar, and retrobulbar anesthe-
cular cone, whose apex is the annulus of Zinn, typi-
sia—along with an understanding of their advantages
cally results in the paralysis of the oculomotor and the
and disadvantages, is necessary in order to provide the
abducens but not the trochlear. The superior oblique
highest level of care for all patients. The goal of this
is often spared, and cyclotorsion may still occur even
chapter is to define and describe the indications and
with a well-placed retrobulbar injection.
techniques for each of these approaches.
Sensory innervation to the cornea and superonasal
conjunctivae is provided by the nasociliary nerve that
Applied Anatomy is within the muscle cone. The remaining conjunctival
A basic knowledge of orbital anatomy is essential sensation is provided by the remaining branches of
to understand the effects and potential complications ophthalmic nerve (frontal and lacrimal) and two divi-
of orbital anesthesia. sions of the maxillary nerve, which supply the lower
Chapter 1
Relative Relative
Tremor Photophobia
Anxiety Anxiety
Claustrophobia Deafness
Children Long operative time
Poor communication/language barrier/deafness
Long operative time Absolute
Poor communication/language barrier/deafness
Absolute Cannot follow directions
Inability to cooperate (eg, schizophrenia, dementia) Insufficient pain control (as in prior eye surgery)
Uncontrolled coughing/movement disorder
lid and conjunctiva (enters via the inferior orbital fora- have no relevant comorbidities. The surgeon should
men). All of these additional somatosensory nerves lie be experienced and expecting a shorter surgery with-
outside of the muscular cone. For this reason, a retro- out anticipated complications or added procedures.
bulbar block can still leave areas of the conjunctiva Longer procedures that may require iris manipulation
sensitive to pain and touch. or scleral suturing may benefit from retrobulbar or
The dura surrounding the optic nerve is continu- peribulbar anesthesia for improved iris and ciliary body
ous with the dura of the brain. Inadvertent injection anesthesia. While most patients can lie still, some may
of anesthetic into the subdural space within the nerve, not be able to follow directions and are not well suited
therefore, can result in brainstem anesthesia. for topical anesthesia. Patients who have psychiatric
disease or other comorbidities that prevent them from
lying still may be candidates for general anesthesia.
Preoperative Evaluation The information contained in Tables 1-1 and 1-2
Careful patient screening is essential in order to can serve as general guidelines for anesthesia selec-
determine which form of anesthesia is best suited for tion. In some instances, reviewing the procedure and
an individual. A surgeon should develop a checklist different anesthesia approaches with the patient is
to avoid missing data that can influence the choice of useful. This allows the patient to self-assess his or her
anesthesia. A history and physical examination, with preferences. The discussion also allows the patient to
review of medications, is an excellent starting point ask questions and develop greater comfort with the
for evaluation. Particular attention should be given to surgeon and surgery.
the patient’s ability to communicate, lie flat and still,
and follow directions. A history of congestive heart
failure, chronic obstructive pulmonary disease, chronic Retrobulbar Anesthesia
bronchitis, claustrophobia, anticoagulation status, and Multiple protocols have been published with a
use of alpha-blockers (tamsulosin) should be addressed common goal of improving the efficacy and safety of
with each patient. retrobulbar anesthesia. Complications arising from ret-
Retrobulbar and peribulbar anesthesia generally robulbar anesthesia include retrobulbar hemorrhage,
provide excellent intraoperative pain control with globe/nerve perforation, extraocular muscle injury,
the added benefit of complete or partial akinesia and and brainstem anesthesia/death.2 Other disadvantages
visual block. General anesthesia may be utilized when include the need for increased sedation, a postopera-
generalized muscle paralysis is an additional factor to tive eye patch, longer visual recovery, ptosis, chemosis,
ensure surgical success. Topical anesthesia should be subconjunctival hemorrhage, and increased posterior
reserved for communicative and calm patients who pressure during surgery. The most feared complication
Local Anesthesia for Cataract Surgery
of retrobulbar injection, perforation of the globe, is Retrobulbar anesthesia is performed prior to sterile
more common with eyes of higher axial length and/or prep. The patient is positioned flat on the operative
staphyloma.2 bed. At the level of the forehead, 1-inch silk, plastic,
A well-placed retrobulbar block usually results in or paper tape can be used to secure the head to the
excellent akinesia and sensory block with some visual table if an assistant is not available. Intravenous pro-
block also. As previously noted, motor nerves within pofol or Versed (Hospira, Lake Forest, IL) should be
the muscle cone, the abducens, oculomotor, as well as administered in conjunction with an analgesic, such
the sensory nasociliary nerve, are affected, but because as fentanyl, to help prepare the patient for injection.
the trochlear nerve passes outside the cone, superior If propofol is used, time for the medication to take
oblique muscle innervation is usually spared and cyclo- effect should be allowed. Testing the lack of orbicularis
torsion may still occur. Most surgeons supplement contraction by gently brushing the eyelashes can help
retrobulbar blocks with topical anesthesia to complete verify adequate sedation.
anterior segment anesthesia because portions of the Following surgery, the eye should be patched. This
trigeminal, which supply the conjunctiva and lid, also is because the retrobulbar block reduces sensation
pass outside the muscle cone. of the eye (which results in a reduced blink reflex),
The goal of retrobulbar anesthesia is the place- provides akinesia (which causes a transient diplo-
ment of anesthetic into the intramuscular cone located pia), and reduces vision (which is frightening to the
behind the globe and anterior to the orbital apex. patient). The patch may be removed after 4 to 6 hours
Structures traversed by the retrobulbar needle include in patients who have received only lidocaine. When
the skin, orbital septum, periocular tissue/fat, and bupivacaine is used, the patch should remain for not
the intramuscular connective tissue. Structures to be less than 8 hours.
avoided include blood vessels, extraocular muscles,
the globe, and the optic nerve. The technique detailed
below is designed to avoid these structures and give Parabulbar (Sub-Tenon’s)
reliable and reproducible anesthesia. Each surgeon will Anesthesia
develop personal amendments, but the basic tenets Some surgeons have adopted the technique of
apply. using a blunt-tipped cannula intraoperatively to inject
Injectable mixtures should include a total volume the same anesthetic mixture. This is known as a para-
of 10 cc or less composed of 2% lidocaine without bulbar block. Parabulbar blocks can be placed as a
epinephrine mixed 50:50 with 0.75%. Note that this planned anesthesia or can be utilized intraoperatively
50:50 mixture dilutes each component to half the if the patient is uncooperative or has inadequate pain
original concentration. Some surgeons may prefer 4% control with topical/peribulbar anesthesia.
lidocaine, if available, to yield a final effective concen- This technique avoids the hazards of a sharp
tration of 2% lidocaine. The addition of bupivacaine needle placement into this space and is a safer alterna-
increases the duration of action. If hyaluronidase is tive to retrobulbar anesthesia, but it can also result in
available, it can also be added to the mixture to speed increased chemosis, subconjuctival hemorrhage, and
diffusion of the medication and improve akinesia and incomplete anesthesia if the cannula is not advanced in
sensory block. Hyaluronidase can also decrease poste- the sub-Tenon’s space.4 Damage to the vortex veins has
rior pressure by causing the volume to distribute more also been reported.5 Onset is rapid, but the added dis-
quickly. Fifteen to 20 units of hyaluronidase per mL of section can add to operative time. The disadvantages
solution can be used. such as the need for patching with delayed visual reha-
A 38-mm (1.5-inch) 23-gauge needle with a round- bilitation apply, as with retrobulbar anesthesia.
ed point (Atkinson) is preferred. A standard sharp
point needle has the advantage of passing through tis-
sues more easily with less discomfort, but the reduced Peribulbar Anesthesia
sensory feedback during injection and higher potential The injection of anesthesia within the orbit with-
for injury to ocular structures favors the Atkinson or out directing the needle inside the muscle cone
blunt-tipped needle.3 A 10-cc syringe is also preferred reduces the risk of damage to vital structures. The soft
over a 5-cc for better tactile control of injection pres- tissue, intramuscular septae are incomplete and allow
sure and enough volume to change needles and contin- for the diffusion of medication into the cone, resulting
ue with facial nerve blocks after retrobulbar injection in akinesia and visual block, as well as sensory dein-
using the same syringe. nervation to the nasociliary and extraconal divisions
Chapter 1
of first and second divisions of the trigeminal nerve. the anterior segment sensory block. Communicative,
This technique relies on larger volumes (7 to 10 cc) calm, cooperative patients are candidates for topical
and works best if supplemented by 500 units of hyal- anesthesia. Careful patient selection is important.
uronidase. Sedation with propofol, as with retrobulbar Multiple agents are available for topical anesthesia
anesthesia, is preferred. and include tetracaine 0.5% drops, Tetravisc 0.5%
Sensory block and akinesia are dependent on dif- gel (Ocusoft, Richmond, TX), lidocaine 2% jelly,
fusion, therefore this technique requires reassessment Xylocaine 4% (AstraZeneca, Wilmington, DE), and
of akinesia (if desired) after 5 to 7 minutes. If adequate bupivacaine 0.75%. Topical agents are placed at least 5
medial rectus akinesia is not obtained, the peribulbar to 10 minutes prior to surgery. They provide excellent
injection can be repeated using the same technique intraoperative pain control and also allow the patient
targeting the medial fat compartment. Up to 24% to have less discomfort from the Betadine prep prior
of patients will require this supplemental 3- to 5-cc to draping.
block.6 The entrance site for the supplemental block is Drop preparations are generally administered
just nasal to the medial rectus, adjacent to the carun- in two to three repeated doses separated by 5 to 10
cle, and parallel to the medial orbital wall in the same minutes. Gel preparations have the benefit of coating
fashion as described above. Higher volumes overall the eye without requiring repeated doses. If used prior
are used, therefore orbital pressure is increased and to dilating agents, gels can interfere with absorption.
ecchymosis and chemosis are more likely than with the Therefore, many surgeons place a liquid preparation
retrobulbar block.7 Reports of retrobulbar hemorrhage such as proparacaine 0.5% or tetracaine drops first
and globe perforation have also been published but are and then complete the dilation protocols. After the
less common. This technique has reported anesthetic pupil is dilated and 5 to 10 minutes prior to entering
pain control similar to retrobulbar placement, but has the operating room, Tetravisc or lidocaine gel can be
an improved safety profile.8 Overall, the advantages placed into the eye. Lidocaine gel can be more viscous
of peribulbar anesthesia should be weighed against the and at times more difficult to place under the lids to
frequent need for supplemental anesthesia, incomplete anesthetize the superior and inferior conjunctiva and
akinesia, the larger volume of anesthesia, and longer fornices.10 Tetravisc has an intermediate viscosity and
time required for complete diffusion. therefore spreads like a liquid drop but also coats like
a gel. Each surgeon should develop a simple, repro-
ducible protocol for topical anesthesia that can be
Topical and Intracameral performed efficiently by the surgical staff. One other
Anesthesia variant on this form of anesthesia includes soaking
As phacoemulsification techniques have advanced, a sponge with both dilating and/or anesthetic drops
incision size has decreased, the need for iris manipula- (perilimbal anesthesia) and placing it in the inferior
tion has diminished, and operative time has lessened. fornix for 10 to 15 minutes. Anecdotal reports suggest
These changes have resulted in a decrease in the need that soaked pledgets can deliver higher concentrations
for complete akinesia, long duration of ocular anesthe- of both anesthetic and mydriatic medications, but the
sia, and intensity of iris and ciliary body sensory block. actual procedure of sponge placement can be more
Topical and intracameral anesthesia alone can provide intrusive than drops alone.
adequate anterior segment anesthesia for noncomplex Topical anesthesia alone may not provide adequate
phacoemulsification with proper patient selection.9 iris and ciliary body anesthesia. Therefore, many
Use in trabeculectomy, secondary sutured IOLs, and surgeons will supplement with intracameral 1% non-
pterygium excision has also become more common. preserved lidocaine. After the initial paracentesis is
Topical anesthesia avoids the systemic and ocular created, approximately 0.5-cc nonpreserved lidocaine
risks of the previously described modalities. In addi- is instilled into the anterior chamber. Uncomplicated
tion, it allows for quick visual recovery. Monitored cataract surgery can be performed with topical anes-
anesthesia care can be used, but surgery can also thesia alone, but prospective trials suggest an addition-
be performed without intravenous agents (discussed al anesthetic benefit to intracameral lidocaine.11 This
below). It should be noted that many surgeons who additional agent represents a very quick, extra step
use retrobulbar or peribulbar block use topical and/or in cataract surgery. At 1% concentration, endothelial
intracameral anesthetic in addition to help complete cell toxicity has not been demonstrated in humans.
Local Anesthesia for Cataract Surgery
The additional anesthetic effect makes any iris touch The modified Van Lint targets the terminal branch-
or manipulation more comfortable. If a scleral sutured es at the lateral canthus and lid. This technique avoids
posterior chamber lens, pupil expansion device, or iris the paralysis of the other divisions of the seventh nerve
stretching is necessary, intracameral anesthesia can be but can cause lid ecchymosis and edema.
a useful adjunct. Other agents such as epinephrine or Facial nerve blocks are best done with conscious
phenylephrine can also be added to this intracameral sedation usually directly after retrobulbar or peribul-
solution. These and other techniques are discussed in bar block while the propofol is still at maximal effect.
other portions of this text. The same 10-cc syringe can be used if appropriate
by changing the needle to a conventional sharp point
1-inch 30-gauge or 27-gauge needle.
Facial Nerve Blocks
Occasionally, a patient may have difficulty with
relaxing his or her orbicularis oculi muscle. Many Conscious Sedation and
times this is anxiety related, other times it may be
an idiosyncratic reflex specific to that individual.
General Anesthetic Agents
Cataract surgeons should possess a basic under-
Psychiatric disease can be a risk factor. If intravenous standing of common anesthetic agents. Many times
agents fail to reduce squeezing, facial nerve blocks feedback from the patient is only perceived and
in combination with any of the anesthetic modalities communicated to the surgeon intraoperatively. The
above can allow the surgeon to have improved control. surgeon may also better understand the needs of each
Generally, patients who require facial nerve blocks are patient, having treated him or her for many years,
good candidates for retrobulbar/peribulbar anesthesia than the anesthesiologist present for the surgery. An
because of associated Bell’s phenomenon. Facial nerve understanding of the common medications and their
blocks can be performed at any portion of the extra- relative analgesic, anxiolytic, and amnestic properties
cranial course after it exits the stylomastoid foramen. will allow the surgeon to help tailor preoperative plan-
The nerve gives off multiple branches as it courses ning and intraoperative supplementation.
from behind the ear over the angle of the mandible, Monitored anesthesia care involves intravenous
penetrating the parotid gland and dividing into its ter- sedation and analgesia with noninvasive monitoring.
minal branches, including the temporal and zygomatic, This allows for less physical stress on the patient. The
which supply the orbicularis. Types of facial nerve patient is able to respond to commands, facilitating
blocks are differentiated by their location, and each surgery, and recovery is quicker. Conversion to gen-
has inherent advantages and disadvantages. eral anesthesia is still possible. Commonly used single
Careful placement of additional anesthetic in the agents include opiates (fentanyl), benzodiazepines
inferior fornix and anterior lateral orbit as the needle (midazolam), and propofol.
is withdrawn during retrobulbar and peribulbar anes- Propofol (Diprivan [AstraZeneca]) is a short-act-
thesia can also result in seventh nerve block in up to ing induction agent that provides temporary sedation
88% of cases by continued diffusion.12 Although less without analgesia. Propofol can be used prior to ret-
reliable, this can obviate the need for a separate facial robulbar block placement. Although the block can be
nerve block. placed without propofol, this agent provides a short
The Nadbath block is directed at the exit of the duration of deep sedation with amnesia. Testing lack
nerve at the stylomastoid foramen. Respiratory and of orbicularis contraction by gentle eyelash stimulation
vocal chord paralysis have been reported with inadver- can be a helpful measure of adequate sedation prior to
tent injection into the jugular foramen.13-15 Prolonged retrobulbar placement. Hypotension and temporary
facial nerve block has also been reported.15 This tech- apnea are possible, therefore pulsoximetry and blood
nique avoids ecchymosis of the face and is less painful, pressure monitoring are essential.
but also can temporarily paralyze multiple divisions of Fentanyl and midazolam (Versed) can be used
the facial nerve. alone or in conjunction. Fentanyl, a short-acting nar-
The O’Brien is placed more distally just below the cotic, provides analgesia with some mild anxiolysis.
zygomatic arch, anterior to the tragus. This site can Midazolam is an excellent anxiolytic and can also have
be more painful and can also cause paralysis of the lips an amnestic effect. Midazolam is short acting, water
and lower face in addition to the intended superior soluble, and has no analgesic properties. Both have
divisions. a quick onset of action and can be augmented dur-
Chapter 1
Figure 1-1. Proper patient stabilization and position Figure 1-2. The retrobulbar/peribulbar needle
for retrobulbar/peribulbar anesthesia. should enter at the lateral one-third of the lower
eyelid below the globe with the eye in primary
position. Supplemental medial peribulbar blocks
ing surgery for added effect. Midazolam can have a enter between the caruncle and medial rectus.
disinhibiting effect that can result in a lack of patient
cooperation. This disinhibition and confusion is more
common in the elderly and quite rare in younger the globe throughout the procedure (Figure
patients. In some circumstances, patients can attempt 1-1). The surgeon should be on the same side of
to sit up or remove their draping. Therefore, careful the bed as the operative eye. The lower eye lid
attention and communication with the patient and skin should be cleaned with an alcohol swab.
anesthesiologist during surgery should be maintained Step 3. Needle Placement. The needle tip, bevel
in order to continually assess patient comfort and down, is advanced parallel to the orbital floor,
mental status. entering at the lateral third of the inferior lid.
Cooperation with adequate pain and anxiety con- The patient’s eye should be in primary position
trol is the goal of every cataract surgery. The surgeon’s (Figures 1-2 and 1-3).
demeanor and communication can help supplement Step 4. Needle Advancement. The surgeon’s index
pharmacologic anesthesia. Some individuals may expe- finger can be used to palpate and displace the
rience pain but not alert the surgeon for fear of “inter- globe superiorly as the needle is positioned
fering” with the surgery. It can be useful to briefly to create adequate space for the needle to
describe to the patient what to expect in the operating pass inferior to the globe between the lateral
room and encourage him or her to verbally express and inferior rectus muscles. Resistance to the
discomfort so that added analgesia can be provided. rounded needle can be noted when the orbital
septum is reached. Once the needle has passed
the equator of the globe (the halfway point of
Step-by-Step Approach to the needle should be at the level of the iris),
Retrobulbar Anesthesia the needle is then angled superior and slightly
Step 1. Anesthetic Preparation. A 38-mm (1.5-inch) medial toward the muscular cone to a loca-
23-gauge needle with a rounded point (Atkin- tion posterior to the macula. A small amount
son) on a 10-cc syringe is preferred. Ten cc of anesthesia can be injected as the needle is
containing 2% lidocaine without epinephrine advanced.
mixed 50:50 with 0.75% bupivacaine and 10 to Step 5. Entering the Muscle Cone and Injecting. Re-
15 units hyaluronidase per cc (optional) can be sistance and relief can be detected as the needle
used. enters the muscle cone. The syringe plunger
Step 2. Patient Position. The assistant should be pres- should be gently withdrawn to ensure a blood
ent at the head of the bed, facing the feet, hold- vessel has not been entered prior to injection.
ing the head securely with both hands. One Depending on anticipated cone volume, 2.5 to
finger can be used to lift the upper lid of the 4.0 cc should be injected. An additional 1 to 2
operative eye to allow the surgeon to visualize cc can be injected as the needle is withdrawn.
Local Anesthesia for Cataract Surgery
Step-by-Step Approach to
Step 6. Assessment. Gentle “on and off” digital pres-
sure should be used for 2 to 4 minutes to help Parabulbar (Sub-Tenon’s)
facilitate diffusion of the anesthesia. Checking Anesthesia
for the amount of akinesia can help assess the Step 1. Conjunctival Incision. An incision is made
success of the retrobulbar block within a few with a Wescott scissors between the superior
minutes of placement. rectus and lateral rectus 9 to 10 mm posterior
If progressive proptosis, hemorrhagic chemo- to the limbus down to bare sclera. The scissors
sis, or unexplained posterior pressure during are used to bluntly dissect posteriorly to allow
surgery is detected, retrobulbar hemorrhage space to advance the cannula (Figure 1-4).
should be suspected. Immediate lateral can- Step 2. Anesthetic Placement. A 5-cc syringe with a
thotomy and cantholysis is the treatment of blunt-tipped cannula containing a 50:50 lido-
choice. Some surgeons proceed with surgery caine 2% (without epinephrine) and bupiva-
immediately once retrobulbar pressure is re- caine 0.75% mixture is advanced around the
duced. Most surgeons prefer to delay the pro- equator of the globe into the anterior intra-
cedure and wait a sufficient time for recovery conal space. It is important to directly visual-
and reassessment. ize the blunt cannula entering under Tenon’s
capsule (Figure 1-5). The cannula should fol-
low the curve of the globe posteriorly. Two to
Step-by-Step Approach to 3 cc should be injected.
Peribulbar Anesthesia
Step 1. Anesthetic Preparation. A 25-mm 23-gauge
needle with a rounded point (Atkinson) on Step-by-Step Approach to
a 20-cc syringe is preferred. Ten cc contain- Topical and Intracameral
ing 2% lidocaine (without epinephrine) mixed
50:50 with 0.75% bupivacaine and 10 to 15
Anesthesia
units hyaluronidase per cc is used.
Step 2. Patient Position. The assistant should be pres-
Application of Topical Anesthetic
Tetracaine or proparacaine is used in two to
ent at the head of the bed, facing the feet,
three divided doses in each eye prior to sur-
holding the head securely with both hands.
gery. The first dose is given just prior to di-
One finger can be used to lift the upper lid of
lating agents and then repeated every 5 to
the operative eye to allow the surgeon to visu-
10 minutes with each application of dilating
Chapter 1
Figure 1-4. Dissection to bare sclera in the supero- Figure 1-5. The cannula is advanced in sub-Tenon’s
temporal quadrant. (Photo courtesy of Thomas A. space posteriorly hugging the globe. (Photo cour-
Oetting, MS, MD.) tesy of Thomas A. Oetting, MS, MD.)
drops. One additional application just prior to pivacaine is injected. This technique avoids ec-
surgery may be necessary. If Tetravisc is uti- chymosis of the face and is less painful but also
lized, one dose 5 to 10 minutes prior to sur- can temporarily paralyze multiple divisions of
gery is placed in each eye. the facial nerve.
A small astigmatically neutral cataract incision is incision and observed that McFarland’s scleral tunnel
one of the fundamental benefits of phacoemulsifica- involved a dissection into corneal tissue. He theorized
tion and foldable intraocular lenses (IOLs). When that the water-tight nature of the incision was due in
intracapsular and standard extracapsular surgery were large part to an internal corneal flap that behaved like
the mainstay of ophthalmology, the customary surgi- a flutter valve. Ernest subsequently performed cadaver
cal approach was the fornix-based peritomy, followed studies and, utilizing manometric pressure testing,
by a superior limbal or scleral incision, closed with concluded that the strongest and most stable design
interrupted and/or running sutures. Astigmatic insta- for a sutureless incision was one in which the width and
bility, associated with uneven suture tension in the depth of the incision were equal.2 In the early 1990s,
short term and wound separation with flattening of foldable IOL technology had not evolved sufficiently
the corneal curvature in the long term, was an unfor- to allow IOLs to be inserted through incisions smaller
tunate but unavoidable feature of these long incisions. than 3.5 to 4 mm. For this reason, Ernest initially advo-
Phacoemulsification has been embraced by ophthalmic cated scleral- or limbal-based incisions with an internal
surgeons in large part because small incision surgery corneal flap of 1.5 mm or more.3 With improvements
provides patients with an opportunity for more rapid in IOL delivery systems in the mid-1990s, it became
visual recovery and for greater refractive stability. possible to perform the entire phaco procedure with
lens implantation through an incision of 3 mm or less.
Once incisions were of this size, both limbal and “clear
Evolution of the corneal” incisions were found to be of virtually equal
Sutureless Incision in strength as long as the equality of incisional width and
internal length were maintained.4 Topographic studies,
Phacoemulsification moreover, performed by Menapace and his colleagues
McFarland reported the first series of patients under- on a variety of clear corneal incision configurations
going phacoemulsification with a sutureless incision in determined that square incisions in which the internal
1990. His original approach involved a standard scleral length of the incision equaled its width provided the
tunnel technique, performed superiorly with a conjunc- greatest astigmatic stability both in the short and lon-
tival peritomy.1 Ernest analyzed McFarland’s sutureless ger term.5
11
12 Chapter 2
Figure 2-1. Stabilize the globe. Figure 2-2. Using a trapezoidal blade that is pre-
cisely matched in width to your phaco tip, enter
at the end of the terminal vessels in the limbal
and Fine, Hoffman, and Packer have reported a large arcade.
series of sutureless clear corneal incisions over a
10-year period without a single case of endophthalmi-
tis.9 I have had a similar experience. I have performed temporally placed clear corneal incision for the main-
over 8000 clear corneal incisions without a case of tenance of astigmatic neutrality, but they suggest that
postoperative infection. It must be understood, howev- incisions placed superiorly should be scleral corneal.
er, that the threshold for placement of corneal sutures
should be very low. It is impossible for any surgeon
to make a “perfect” clear corneal incision with every Step-by-Step Approach to the
effort. At the end of each case, every incision must Clear Corneal Incision
be critically evaluated and carefully tested. If there is Step 1. Stabilize the Globe. Stabilize the globe using a
evidence that the internal length of the incision is too ring holder placed at the limbus (Figure 2-1).
short or that the incision is poorly constructed in any Step 2. Enter at the Limbal Arcade. Using a trapezoi-
way, the incision should be sutured. dal blade that is precisely matched in width to
your phaco tip, enter at the end of the termi-
nal vessels in the limbal arcade. Placement of
Astigmatic Considerations the entry at this location allows the surgeon
Phacoemulsification surgeons today fall into two to develop an incision that is as long internally
groups: those who always approach the eye from a as it is wide without extending too far into the
temporal location and then perform limbal relaxing anterior cornea and also helps the surgeon to
incisions when necessary in the steep axis, and those avoid cutting through the conjunctiva (Figure
who make their incision on the steep corneal axis and 2-2). If the incision is made too posteriorly, in-
then make limbal relaxing incisions, as needed, oppo- fusion of fluids with the phaco tip can create
site and adjacent to the incision. The temporal clear conjunctival chemosis that can result in pool-
corneal incision is favored by many because of ease of ing of extraocular fluids over the surface of the
access and because of the astigmatic neutrality afford- cornea and reduced visualization. If conjunc-
ed by this approach.21 While studies have shown that tival chemosis occurs, it can be relieved easily
small scleral corneal tunnel incisions made superiorly by snipping through the conjunctiva radially
result in astigmatic changes similar to small temporal and in both lateral directions at the limbus.
clear corneal incisions,22 clear corneal incisions made Step 3. Make the Intrastromal Length of the Incision
superiorly clearly result in greater and less predict- Equal to the Width of the Incision. Direct the
able astigmatic shifts than do temporally placed clear tip of the blade anteriorly under direct visual-
corneal incisions.23,24 It has even been demonstrated ization until the tip of the blade has reached an
that superior oblique clear corneal incisions result in intrastromal length equal to or slightly longer
greater astigmatic shifts than do temporal clear corneal than the width of the blade (Figure 2-3). If the
incisions.25 These studies confirm the usefulness of the incision is made much longer than the width of
14 Chapter 2
Figure 2-3. Direct the tip of the blade anteriorly Figure 2-4. Complete the internal incision by direct-
under direct visualization until the tip of the blade ing the tip of the blade parallel to the iris plane and
has reached an intrastromal length equal to the enter the anterior chamber.
width of the blade.
Figure 2-5. Examine the incision carefully to make Figure 2-6. Gently hydrate the margins of the
certain that the architecture of the incision is incision and all side ports with BSS. Make sure
square. that there is no evidence of incarceration of a
Descemet’s flap in the incision and that the incision
is secure even to rigorous external pressure.
the incision, introduction of the phaco tip may
create folds in Descemet’s membrane, which
makes visualization of the anterior cham- Make sure that there is no evidence of incar-
ber difficult. As emphasized above, incisions ceration of a Descemet’s flap in the incision,
shorter than the width of the incision are likely that the architecture of the incision is square,
to leak. and that the incision is secure even to rigorous
Step 4. Complete the Internal Incision. Direct the tip external pressure (Figure 2-5). Gently hydrate
of the blade parallel to the iris plane and enter the margins of the incision and all side ports
the anterior chamber (Figure 2-4). Be sure that with balanced salt solution (BSS) (Figure 2-6).
the internal incision is complete, but be careful; Step 6. If the Incision Is Not “Rock Solid” Perfect,
if you are using a side cutting blade, do not en- Suture It. If the incision is poorly constructed
large the incision inadvertently. This can result or if it can be made to leak with rigorous exter-
in poor fluidics during the procedure and an nal pressure, suture the incision and then reex-
incompetent incision at the end of the case. amine. If a Descemet’s flap is observed, gently
Step 5. Carefully Examine and Hydrate the Incision. irrigate the flap into the anterior chamber and
At the end of the procedure, fill the anterior suture the incision. Use additional sutures, if
chamber and inspect the incision carefully. necessary, to ensure competency.
Incisions 15
Figure 2-7. Create a fornix-based peritomy, remov- Figure 2-8. Make a corneal groove 1 to 2 mm
ing all Tenon’s fibers for better hemostasis. Cauterize posterior to the limbus at a depth of approximately
lightly to avoid scleral shrinkage. 250 microns.
Figure 2-10. Using a keratome that is precisely Figure 2-11. At the end of the procedure, examine
matched to the width of your phaco tip, direct the the incision carefully. The incision should be square
blade parallel to the iris plane and enter the ante- or nearly square with an internal corneal incision
rior chamber at the end of the scleral tunnel. of at least 1.5 mm. If the incision appears well
constructed, gently hydrate the margins of the inci-
sion and all side ports with BSS and fill the anterior
anterior chamber and suture the incision. If the chamber.
incision is poorly constructed or if it leaks with
rigorous external pressure, suture the incision scleral tunnel and clear corneal wounds. Am J Ophthalmol.
and reexamine. Use additional sutures, if nec- 2003;136:300-305.
essary, to ensure competency. 13. Taban M, Behrens A, Newcomb RL, et al. Acute endophthal-
mitis following cataract surgery: a systematic review of the
literature. Arch Ophthalmol. 2005;123:613-620.
References 14. Shingleton BJ, Wadhwani RA, O’Donogue MW, et al. Evalu-
ation of intraocular pressure in the immediate period after
1. McFarland MS. Surgeon undertakes phaco, foldable IOL se-
phacoemulsification. J Cataract Refract Surg. 2001;27:524-527.
ries sans sutures. Ocular Surgery News. 1990;8(5):1,15.
15. Taban M, Sarayba MA, Ignacio TS, Behrens A, McDonnell
2. Ernest PH, Fendzl R, Lavery KT, Sensoli A. Relative stability
PJ. Ingess of India ink into the anterior chamber through
of clear corneal incisions in a cadaver eye model. J Cataract
sutureless clear corneal cataract wounds. Arch Ophthalmol.
Refract Surg. 1995;21(1):39-42.
2005;123(5):643-648.
3. Ernest PH, Lavery KT, Kiessling LA. Relative strength of
16. Miller JJ, Scott IU, Flynn HW, et al. Acute endophthalmitis
scleral corneal and clear corneal incisions constructed in ca-
following cataract surgery (2000-2004): incidence, clinical
daver eyes. J Cataract Refract Surg. 1994;20(6):626-629.
settings, and visual acuity outcomes after treatment. Am J
4. Ernest PH. Wound construction: the state of the art. Review of
Ophthalmol. 2005;139:983-987.
Ophthalmology. 2002;9(4):66-70.
17. Tan DY, Vagefi MR, Naseri A. The clear corneal tongue: a
5. Pfleger T, Skorpik C, Menapace R, et al. Long-term course of
mechanism for wound incompetence after phacoemulsifica-
induced astigmatism after clear corneal incision after cataract
tion. Am J Ophthalmol. 2007;143:526-528.
surgery. J Cataract Refract Surg. 1996;22(1):72-77.
18. Olson RJ. Should I change my incision. J Cataract Refract Surg
6. Fine IH. Clear corneal incisions. Int Ophthalmol Clin.
Today. 2007;7:45-47.
1994;34:59-72.
19. Masket S, Belani S. Proper wound construction to prevent
7. Leaming DV. Practice styles and preferences of ASCRS mem-
short-term hypotony after clear corneal incision cataract sur-
bers—2003 survey. J Cataract Refract Surg. 2004;30:892-900.
gery. J Cataract Refract Surg. 2007;33:383-386.
8. Fine IH. Descriptions can improve communication. Ophthal-
20. Monica ML, Long DA. Nine-year safety with self-sealing cor-
mol Times. 1996;21:30.
neal tunnel incision in clear cornea cataract surgery. Ophthal-
9. Fine IH, Hoffman RS, Packer M. The architecture of clear
mology. 2005;112:985-986.
corneal incisions. J Cataract Refract Surg Today. 2007;7:59-65.
21. Altan-Yaycioglu R, Akouva YA, Akca S, et al. Effect on astig-
10. Wallin T, Parker J, Jin Y, et al. Cohort study of 27 cases of
matism of the location of clear corneal incision in phacoemul-
endophthalmitis at a single institution. J Cataract Refract Surg.
sification of cataract. J Refract Surg. 2007;23(5):515.
2005;31:735-741.
22. Poort-van Nouhuijs HM, Hendricks KHM, van Marle WF,
11. Nagaki Y, Hayasaka S, Kadoi C, et al. Bacterial endophthal-
et al. Corneal astigmatism after clear corneal and corneo-
mitis after small-incision cataract surgery. Effect of incision
scleral incisions for cataract surgery. J Cataract Refract Surg.
placement and intraocular lens type. J Cataract Refract Surg.
1997;23:758-760.
2003;29:20-26.
23. Marek R, Klus A, Pawlik R. Comparison of surgically induced
12. Cooper BA, Holekamp NM, Bohigan G, et al. Case-control
astigmatism of temporal versus superior clear corneal inci-
study of endophthalmitis after cataract surgery comparing
sions. Klin Oczna. 2006;108:392-396.
Incisions 17
24. Simisek S, Yasar T, Demirok A, et al. Effect of superior and 25. Rainer G, Menapace R, Vass C, et al. Corneal shape changes
temporal clear corneal incisions on astigmatism after suture- after temporal and superolateral 3.0 mm clear corneal inci-
less phacoemulsification. J Cataract Refract Surg. 1998;24:515- sions. J Cataract Refract Surg. 1999;25:1121-1126.
518.
Chapter
3
Capsulorrhexis
Cataract extraction by the extracapsular method radial extension of one of the small tear notches. These
requires an opening of the anterior capsule. Before tears then extended to the periphery of the capsule
Charles Kelman’s development of phacoemulsification, and “zipped” around the equator, resulting in a poste-
when standard extracapsular surgery was performed rior capsular tear. As phacoemulsification techniques
routinely, the capsule was opened with a multi-jaw progressed from anterior chamber emulsification to
forceps, pinching it until it tore from both sides.1 in-the-bag disassembly of the nucleus, manipulation
Typically, an Aruga forceps was used (Figure 3-1). As within the capsular bag increased and the need to
large an opening as possible was ideal because the lens develop a more tear-resistant opening in the anterior
nucleus was then expressed from the capsular bag into capsule became increasingly critical. The development
the chamber and out of the incision. Kelman described of a continuous circular capsular tear technique, which
a more controlled opening using a blunt cystotome, came to be called continuous curvilinear capsulor-
creating a triangular tear (the so-called “Christmas rhexis (CCC), greatly increased the safety of cataract
tree” tear) starting from the 6 o’clock position of the surgery.2 Highly resistant to the development of radial
pupil (Figure 3-2). The torn capsular tag was pulled tears, CCC greatly reduced the risk of intraoperative
out of the superior incision and cut off, creating a posterior capsular tears3 and paved the way for the
triangular opening (Figure 3-3). The triangular open- development of a variety of lens disassembly tech-
ing technique was then modified with additional tears niques, which also increased the safety of phacoemul-
(Figure 3-4), and eventually the “can opener” technique sification. CCC also increased the ability of surgeons
evolved with multiple tears placed in a circular fashion. to place both loops of an intraocular lens (IOL) within
This technique helped to reduce the risks of engage- the capsular bag more reliably and, because the circular
ment of the capsular flaps during phacoemulsification capsular opening helped to ensure symmetrical capsu-
and cortex removal because the multiple small tears lar contraction forces, the technique helped to obtain
resulted in smaller capsular flaps (Figure 3-5). The and maintain better centration of the IOL.
mechanical stress on the margin of the can opener After its introduction, CCC gradually became
capsulotomy, caused by sculpting and manipulation adopted by ophthalmic surgeons because of its clear
of the nucleus, however, sometimes resulted in the advantages. The technique, however, is not as easy to
19
20 Chapter 3
Figure 3-8. Microincisional forceps. Figure 3-9. Fibrotic anterior capsule opened with
angled micro-scissors.
These higher molecular weight OVDs are particularly
helpful when opening the capsules of young people and for fibrotic anterior capsules that require cutting
and children where the lens material is soft and the with micro-scissors (Figure 3-9). The fibrosis may be
capsule is very elastic. With a cohesive, high molecular better visualized after soft cortical material has been
weight viscoelastic, one is able to flatten the anterior aspirated, a red reflex obtained, and more viscoelastic
capsule and equalize the pressure on each side of the has been added.
capsule. A flat capsule rather than a dome allows for a
more controlled CCC opening. In addition, these high
molecular weight viscoelastic materials are helpful Step-by-Step Approach to
with intumescent cataracts. Surgeons have long known Continuous Curvilinear
that emptying of the capsular bag of soft cloudy liq-
uid material through a small opening prior to CCC Capsulorrhexis
makes the procedure more controllable. Before stains Step 1. Select the Optimal OVD(s) for the Specific
such as trypan blue (Vision Blue, Dutch Ophthalmic, Case. As noted previously, cohesive high mo-
Exeter, NH) were available, surgeons performed this lecular weight OVDs provide optimal flatten-
aspiration technique when the lens material was mostly ing of the anterior lens capsule and are better at
liquid but completely white and under a lot of pressure. chamber maintenance than are dispersive, low
This approach should be kept is mind for special situ- molecular weight OVDs. In cases of extremely
ations such as a rapidly developing traumatic cataract dense cataracts, or if an endothelial dystrophy
Capsulorrhexis 23
exists, a dispersive OVD may be used to coat rather than the point of the forceps. The vec-
the endothelium for increased protection, and tor force required to direct the tear appropri-
then a more retentive, cohesive OVD may be ately varies as the tear progresses around the
used to fill the anterior chamber for capsulor- circle. The capsule is more elastic and less
rhexis. “brittle” in young eyes, and the direction of
Step 2. Begin the Tear Centrally and Create a Cap- the force of the forceps is usually quite radial
sular Flap. Once the anterior chamber is filled centripetally or toward the center of the pupil.
with viscoelastic, the capsule is punctured cen- It is sometimes necessary to pull in a direction
trally with a sharp bent cystotome needle or 90 degrees from the direction of the tear. In
a sharp capsule forceps and the tear guided these eyes it is important to release frequently
away from the center in a direction that will to verify the diameter of the tear as the zonules
enable the surgeon to easily grasp the develop- and the capsule are so elastic. It is also very
ing flap with the forceps. This may be just a important to regrasp every 1 or 2 clock hours
vertical tear away from or toward the incision to keep control of the direction and diameter
or a curving tear that is directed radially to the of the tear.
desired diameter of the CCC. The tear is then
continued in a circumferential direction to cre- Additional Points of Importance
ate a flap edge. Using the forceps or additional The ideal diameter of the capsular opening is now
OVD under the flap, elevate the flap vertically widely believed to be about 5 mm. This allows the
into the anterior chamber to make it easier to CCC rim to cover the edge of the 6-mm IOL optic.
grasp with forceps Nishi and Nishi have shown that there is less fibrous
Step 3. Grasp the Flap and Begin the Curvilinear opacification of the posterior capsule postoperatively
Tear. Once the flap is elevated, grasp the flap if the margin of the anterior capsule rim does not touch
with the forceps and continue the circular, or the posterior capsule.5 This also makes it important to
at least curvilinear, tear until it is complete. make the capsular opening as central as possible and
When performing capsulorrhexis, the flap not too oval or outside of the diameter of the optic in
must be regrasped or reengaged a few times any meridian.
to control the direction of the tear. The more In 2005, Tassignon et al of Belgium introduced
difficult the direction of the curvilinear tear is a newly designed ring caliper to facilitate the sizing
to control, the more frequently the flap must and centration of the CCC along the alignment of the
be released and regrasped. Each time the flap first and third Purkinje reflexes as observed under the
is released, be sure to elevate the flap edge in microscope.6 Most surgeons do not have this device
order to make the flap easier to grasp again. available and simply make their best effort to center the
Step 4. Optimize Control of the Tear. Control of the CCC and to make it of an optimal size. Some surgeons
direction of the tear is optimized when the flap place a ring mark on the cornea with a zone marker to
is regrasped near the point of tearing. The flap assist in sizing and placement of the CCC.2
is then folded on itself, and a shearing tech- If the CCC ends up too small or is eccentric, a
nique is used to better direct the tear. Tear- technique that I have described as “two-staged CCC”
ing the capsule by simply pulling the capsule may be considered to enlarge it or make it more round.
centrally gives the surgeon less control of the This technique may also be used to start a tear going
direction of the tear. Folding the capsule over in the opposite direction from one that has radialized
itself, moreover, causes less stress on the zo- or to convert a small can opener opening to a CCC
nular ligament and prevents the capsule from (Figure 3-10). To start a new tear in the edge of an
shifting from a tug on the zonules while the existing opening, a scissor is used to make a very short
tear is fashioned. Such a shift can result in an tangential cut. The new beginning flap is then grasped
eccentric opening in the capsule or one that is with forceps and the new tear is continued around the
larger than desired. In routine cases this shear- circle of desired diameter (Figure 3-11). If this new start
ing technique can be combined with direct needs to be in the subincisional area, one may have to
tearing for some portions of the circle. When make another capsule puncture with a sharp needle
direct tearing is used, one has to be aware of or cystotome to start a new tear. Angled vitrectomy
vector forces and watch the point of tearing scissors may be used as well for unusual situations.
24 Chapter 3
Careful placement of a cohesive viscoelastic under and A small snag may be created using the barbed end of
over the capsule is necessary to safely start the new a disposable 27-gauge hypodermic needle. This small
tear. barb may be made by pressing the tip of the needle
A capsule with a fibrotic zone or an entirely fibrot- on the handle of a needle driver or forceps until a
ic anterior capsule presents a challenge. If the fibrosis very small, right angle bend is made in the tip. Once
is just a band in a quadrant easy to reach with a scis- the opening is made, the vitreous is pushed back with
sors, one may cut through the band with the scissors. additional viscoelastic through the opening. The tear-
An elegant technique is to use a Fugo blade instrument ing is then continued and additional viscoelastic is
(Medisurg Research and Management, Norristown, added intermittently to keep viscoelastic under the
PA) which cuts with a plasma field around a fine fila- capsule where the flap of capsule is being regrasped.
ment.7 This device cuts with ease through fibrosis as The size of the PCCC should be at least 3 to 4 mm
well as normal capsule and can be used to enlarge for a good-sized opening that will always be clear of
CCCs or manage uncontrolled capsule tears. The secondary cataract, except in children where cells use
device is also useful in the management of traumatic the intact vitreous face as a scaffold and occlude the
openings in the anterior capsule that cannot be safely visual axis; in this case, when the technique of PCCC
handled with scissors and forceps. optic capture is planned, the opening should be about
Posterior continuous curvilinear capsulorrhexis 4.5 mm (Figure 3-12).9 Openings larger than this may
(PCCC) may also be performed.8 The same principles not capture the haptic optic junctions tightly enough
apply when creating an opening in the posterior cap- to prevent lens epithelial cells from migrating behind
sule as in the anterior capsule. When opening the the IOL to opacify the visual axis by depositing new
posterior capsule, one may decide to not prevent for- lens material on the intact vitreous face (Figure 3-13). If
ward movement of the vitreous because a vitrectomy the opening is smaller than 4.5 mm, and thus too small
is planned and thus take no precautions to protect for optic capture, the two-stage technique may be used
it. If, however, one wishes to avoid a vitrectomy and to enlarge it just as with the anterior capsule.
protect the vitreous, it is important to start the PCCC The PCCC may be used for access to the vitreous
with a hooking snag rather than a cutting puncture. cavity for a number of indications such as the removal
Capsulorrhexis 25
of silicone oil, reduction of asteroid hyalosis, for ante- another fixation technique to utilize for fixation of
rior vitrectomy and antibiotic injection in cases of the IOL to the capsule is capsular membrane suture
presumed postoperative endophthalmitis, and to pre- fixation.10 In this technique the haptics of the IOL are
vent secondary cataract by preemptively opening the sutured to the fibrotic elements of the capsular mem-
posterior capsule. Capsular dyes may be helpful when brane to fix the IOL to the capsular membrane rather
doing PCCC, especially when there is a poor fundus than leave it loose in the sulcus to erode iris pigment,
reflex, as with asteroid hyalosis and endophthalmitis. erode iris vessels, and potentially become eccentric
Indications for PCCC in order to prevent secondary (Figure 3-14).
cataract include severe kyphosis and socioeconomic
barriers to Nd:YAG laser capsulotomy.
Openings in the posterior capsule may be made References
1. Kelman CD. Phaco-emulsification and aspiration. A new
many years after primary surgery. This may be indi- technique of cataract removal. A preliminary report. Am J
cated when IOL removal and replacement or IOL Ophthalmol. 1967;64(1):23-35.
repositioning is needed and membrane optic capture 2. Bond W, Bakewell BK, Agarwal A, Spalton D. Sizing and
is planned to fix the IOL to the capsular membrane. If centering the capsulorrhexis. J Cataract Refract Surg Today.
2007;7(4):27-32.
there is no fibrosis in a 4- to 5-mm zone of the capsular
3. Gimbel HV. Divide and conquer nucleofractis phacoemul-
membrane, the opening may be made with the tearing sification: development and variations. J Cataract Refract Surg.
techniques described above. If there is fibrosis pres- 1991;17(3):281-291.
ent or the single posterior layer of capsule is not large 4. Gimbel HV, Neuhann T. Development, advantages, and
enough for a desired 4- to 5-mm opening, the Fugo methods of the continuous circular capsulorrhexis technique.
J Cataract Refract Surg. 1990;16(1):31-37.
blade or scissors would have to be used with increased 5. Nishi O, Nishi K. Effect of the optic size of a single-piece
risk of disturbing the vitreous and of extension of the acrylic intraocular lens on posterior capsule opacification.
tear by the force required to get optic capture. J Cataract Refract Surg. 2003;29(2):348-353.
In secondary surgery for improved fixation, repo- 6. Tassignon MJ, Rozema JJ, Gobin L. Ring-shaped caliper for
better anterior capsulorrhexis sizing and centration. J Cata-
sitioning, or removal and replacement of an IOL,
26 Chapter 3
ract Refract Surg. 2006;32(8):1253-1255. 9. Gimbel HV, DeBroff BM. Intraocular lens optic capture.
7. Fugo RJ. Fugo blade to enlarge phimotic capsulorrhexis. [Let- J Cataract Refract Surg. 2004;30(1):200-206.
ter] J Cataract Refract Surg. 2006;32(11):1900. 10. Gimbel HV, Shah CR, Venkataraman A, Rattray KM. Capsu-
8. Gimbel HV. Posterior capsule tears using phaco-emulsifica- lar membrane suture fixation of sulcus IOLs. Clinical & Surgical
tion. Causes, prevention and management. Eur J Implant Refract Ophthalmlolgy. 2008;26(2):42-47.
Surg. 1990;2(1):63-69.
Chapter
4
Hydrodissection and
Hydrodelineation
I. Howard Fine, MD; Richard S. Hoffman, MD; and Mark Packer, MD, FACS
27
28 Chapter 4
Figure 4-8. Posterior cortex fully draped on top of Figure 4-9. Posterior cortex now draped back on
capsule and iris (arrows=edges of posterior cortex top of the plate haptic IOL ready for mobilization.
are elevated by viscoelastic).
Part A: but it also held the nucleus tightly within the capsular
bag. The old method of tipping the entire nucleus of
Disassembling the Nucleus— a dense cataract into the pupillary plane could no lon-
An Overview ger be used. The new challenge for surgeons was to
devise a method for removing an 8- to 10-mm nucleus
D. Michael Colvard, MD, FACS through a 4- to 7-mm capsular opening.
Surgeons struggled initially as they attempted to
The introduction of the continuous curvilinear cap- modify their techniques of phacoemulsification to meet
sulorrhexis by Howard Gimbel and Thomas Neuhann1 the challenge posed by capsulorrhexis. “Endocapsular”
in the early 1990s solved two critically important techniques of emulsifying the nucleus entirely within
problems for phacoemulsification surgeons and created the capsular bag were proposed by a number of sur-
a new challenge. The traditional can opener capsuloto- geons,2-4 but the awkwardness and inherent difficulties
my, used for years by phaco surgeons, was prone to the and dangers of performing these techniques prevented
development of radial tears. These tears often “zipped” them from gaining wide acceptance. Gradually, a vari-
around the equator of the lens capsule, extended to ety of techniques for cracking or fracturing the nucleus
the posterior capsule, and were a major source of intra- within the capsular bag evolved, and the approach of
operative complications. Even when the radial tears first disassembling the nucleus and then bringing sec-
remained small and caused no intraoperative problems, tions of the nucleus into a “safe zone” for emulsification
asymmetric capsular contractional forces associated an became the foundation of all phaco techniques used
irregular capsular margin frequently lead to intraocular today.
lens (IOL) decentration. Gimbel and Neuhann’s new The term divide and conquer was first proposed by
technique resulted in greater stability of the capsular Howard Gimbel.5 Dr. Gimbel described a systematic
bag during phacoemulsification and improved centra- fracturing of the nucleus within the capsular bag using
tion of IOLs. The problem with capsulorrhexis was that the phaco tip and a surgical spatula.6 John Shepherd
it restricted access to the nucleus. The anterior capsu- modified Gimbel’s approach and developed an elegant
lar margin resisted the development of radial tears, technique for divide and conquer that is widely used
33
34 Chapter 5
Figure 5-1. Divide and conquer—forces are radial Figure 5-2. Horizontal chop—forces are radial and
and centrifugal. centripetal.
6. Gimbel HV. Divide and conquer nucleofractis phacoemul- dus, especially in denser nuclei. This assessment is a
sification: development and variations. J Cataract Refract Surg. relatively easy one to make, even for the inexperienced
1991;17:281-291.
7. Shepherd JR. In situ fracture. J Cataract Refract Surg. surgeon. Chopping techniques rely more heavily on
1990;16:436-440. kinetic and tactile clues, which must be learned and
8. Nagahara K. Phaco chop film. Presented at: International vary greatly from patient to patient.
Congress on Cataract, IOL, and Refractive Surgery. ASCRS,
Seattle, WA; May 1993.
More Protective of the Corneal Endothelium
Divide and conquer tends to be more gentle on
the corneal endothelium because the technique cre-
Part B: ates space in the posterior chamber and encourages
Divide and Conquer phacoemulsification farther from the endothelium.
Chopping techniques can be performed safely, but
D. Michael Colvard, MD, FACS there is a tendency with these techniques to bring the
nuclear fragments into the anterior chamber where
phaco energy is more damaging to the cornea.
Divide and conquer is the most versatile and reli-
able of all methods for nuclear disassembly. Surgeons Useful for Nuclei of All Densities
new to phacoemulsification often view divide and Divide and conquer is a highly versatile proce-
conquer as a beginning technique, yet because of its dure that can be used effectively for a full range of
many virtues, divide and conquer is widely used by nuclear densities from the very soft to the most dense.
experienced surgeons both as a primary approach and Chopping techniques that depend on the phaco tip to
as a dependable fallback maneuver in difficult cases. impale and stabilize the nucleus work best on medium
to dense nuclei. With divide and conquer, small varia-
Advantages of the tions of technique allow the surgeon to manage cata-
racts of all densities.
Divide and Conquer
Requires Less Bimanual Manipulation and Fundamentals of
Lower Vacuum Setting Divide and Conquer
Divide and conquer is easier to learn than chop- The basic divide and conquer technique involves
ping techniques, and it is a safer procedure for begin- the creation of two deep grooves in the nucleus that
ning surgeons. Chopping techniques require the sur- intersect centrally, followed by the cracking of the
geon to impale the nucleus, using phaco power and nucleus into four separate quadrants. This cracking is
high vacuum, and simultaneously engage and crack accomplished by placing a spatula or chopping device
the nucleus using a chopper in the second hand. The at the base of one side of the groove and the phaco tip
basic fracturing maneuver of divide and conquer is less at the other side. Using either direct or cross-action
demanding. It requires the opposing action of both force with these instruments, the vertical margins of
hands, but this is accomplished under nonturbulent the groove are separated, and the nucleus is fractured
conditions without the use of phaco power or vacuum. along the axis of the groove. Proper placement of the
Divide and conquer, moreover, can be performed cracking instrument and the phaco tip at the base of
effectively with lower levels of vacuum and flow than the groove is critical to the success of this maneuver
chopping techniques. This leads to greater anterior (Figure 5-4). If the instruments are placed along the
chamber stability and prevents structures in the ante- anterior margin of the groove, the forces created by
rior chamber from moving too quickly. the separation of the instruments will tend to compress
rather than crack the nucleus (Figure 5-5). Care should
Relies Primarily on Visual Assessments be taken also to make sure that cleavage of the nucleus
The successful performance of divide and conquer is complete before an attempt is made to remove the
relies primarily on visual, rather than tactile, signs. quadrants. Nuclear fracture should occur centrally
Groove depth, a fundamental assessment in divide first. Repositioning of the instruments may be neces-
and conquer, is judged by the light reflex of the fun- sary to extend the fracture to the peripheral rim of the
36 Chapter 5
Figure 5-4. The phaco tip and the cracking instru- Figure 5-5. Placement of the phaco tip and/or the
ment should be placed at the base of the groove. cracking instrument along the anterior margin of
Separation of the instruments results in bidirectional the groove results in compression, rather than sepa-
forces at the base of the groove and nuclear frac- ration and fracture, of the posterior nucleus.
ture.
nucleus. Once the fracture extends the periphery in all drodissection. This is time well spent. Failure
quadrants, the quadrants should mobilize easily. Each to free the nucleus from its cortical attach-
quadrant is then engaged by the phaco tip and brought ments at this stage of the procedure will force
into the pupillary plane in the center of the pupillary you to struggle with nuclear rotation later and
axis for safe phacoemulsification. increase the risk of zonular injury.
Step 4. Sculpt the Nucleus. Using moderate flow,
low phaco power, and low vacuum, begin the
Divide and Conquer: sculpting of the grooves. Low vacuum will al-
A Step-by-Step Approach low you to create the grooves without engag-
Step 1. Choose a Keratome That Fits the Phaco Tip. ing and grabbing the nucleus. Start the groove
Make sure that the width of the keratome you at the proximal margin of the capsulorrhexis.
are using is a perfect fit for your phaco tip. Too Carry the groove across to the distal margin.
small an incision can restrict infusion and re- Use phaco power only as you sculpt forward.
sult in an incisional burn. Too large of an in- This will reduce phaco time and help to lim-
cision will result in excessive outflow of fluid it the phaco energy released in the anterior
and chamber instability. chamber (Figure 5-6).
Step 2. Create a 5- to 7-mm Capsulorrhexis. Make Step 5. Deepen the Grooves. Carry the groove pos-
your capsulorrhexis 5 to 7 mm in diameter. If teriorly until you see a good fundus reflex. As
the capsulorrhexis is too small, mobilization a rule of thumb, three times the width of the
of the quadrants of the nucleus is made diffi- phaco tip is usually deep enough for most nu-
cult. This is particularly true of larger, denser clei. Softer nuclei are often not as thick axi-
nuclei. Be careful, however. Attempts to make ally as dense nuclei. Care must be taken not to
a larger capsulorrhexis may increase the risk groove too posteriorly in these eyes.
of “losing the rhexis” and creating a posterior Step 6. Complete Grooves Before Cracking. Before
tear. attempting to crack the nucleus, rotate the nu-
Step 3. Hydrodissect. Hydrodissect thoroughly (see cleus 90 degrees and create the second groove,
Chapter 4). Prior to introducing the phaco intersecting centrally with the first. The nu-
tip, place additional viscoelastic in the ante- clear plate is easier to rotate if both grooves
rior chamber and use a chopper to check the are made prior to cracking. Also make sure
mobility of the nucleus. The nucleus should that you do not leave a mound of unsculpted
spin freely within the capsular bag. If it does nucleus at the intersection of the two grooves.
not, spend the extra time to complete the hy- This can make cracking more difficult and can
Phaco Techniques 37
Figure 5-6. Start the groove at the proximal margin Figure 5-7. Before attempting to crack the nucleus,
of the capsulorrhexis and carry the groove to the rotate the nucleus 90 degrees and create a second
distal margin of the rhexis with the phaco device groove intersecting the first. Make sure that you do
in “sculpt mode” (ie, phaco I—moderate flow, low not leave a mound of unsculpted nucleus at the
vacuum, low phaco power). intersection of the grooves.
Figure 5-8. With the phaco device in irrigation Figure 5-9. Spread apart and gently lift the edges
only, place the phaco tip and a second instrument of the groove until a crack is seen posteriorly in the
at the base of the groove distal to the primary nuclear plate. Carry this crack across the intersec-
incision. tion of the grooves. Repeat this maneuver to disas-
semble completely all four quadrants.
be avoided by simply carrying each sculpting repeat this maneuver until all four quadrants
pass continuously across the intersection (Fig- are freely mobile. Take the time to make sure
ure 5-7). that disassembly is complete before moving to
Step 7. Crack the Nucleus Into Quadrants. With the the next stage of the procedure. As previously
phaco device in irrigation only, place the pha- noted, the most common error made by begin-
co tip and a second instrument at the base of ning surgeons is the failure to place both of
the groove, distal to the intersection of the two the instruments at the base of the groove. The
grooves. Spread apart and gently lift the edges spreading of instruments placed too anteriorly
of the groove until a crack is seen posteriorly in the groove creates vectors that compress
in the nuclear plate. Reposition the instruments rather than separate the posterior aspect of the
as necessary to complete the crack through groove (Figures 5-8 and 5-9).
the peripheral rim of the nucleus. Rotate the Step 8. Phaco the Quadrants. After the nuclear plate
nuclear plate, using a chopper or spatula, and is disassembled, you can begin the process of
38 Chapter 5
Figure 5-10. Engage a quadrant of nucleus distal Figure 5-11. With the removal of the last nuclear
to the incision with the phaco device in “nucleus quadrant and the epinucleus, switch the phaco
removal mode” (ie, phaco II—higher flow, higher device to the “epinucleus mode” (ie, phaco III—
vacuum, higher phaco power). Once the nuclear moderate flow, moderate vacuum, moderate phaco
quadrant is impaled, wait for vacuum to increase, power). This provides maximal control of the cham-
then draw the quadrant to the center of the pupil ber volume and depth and helps to protect poste-
and begin emulsification. rior capsule at the conclusion of the procedure.
removing the individual quadrants of nucleus. lary axis more easily than larger, wider sections of the
Using higher flow and vacuum levels, engage dense nucleus. Fourth, use a chopping device to break
a nuclear quadrant distal to the intersection of larger sections of hard nucleus into more manageable
the grooves. Right-handed surgeons will find it segments. For the surgeon unaccustomed to using a
easier to engage the distal quadrant to the left. chopper, this can be done safely once the nucleus is
Impale the quadrant, wait just a moment for brought into the pupillary plane and away from the
vacuum to increase and then draw the quad- posterior capsule.
rant to the center of the pupil in the pupillary For very soft nuclei, disassembly in the usual man-
plane, and begin emulsification. The second ner can be difficult because the nuclear material tends
instrument should be used to keep the nuclear to crumble rather than crack. Fortunately, soft nuclei
fragment in the pupillary plane well away from usually do not need to be cracked. In these cases, once
the corneal endothelium (Figures 5-10 and 5- the nucleus is freely mobile after hydrodissection and
11). hydrodelineation, and the grooves have been complet-
ed, simply engage each quadrant with the phaco tip in
irrigation and aspiration and fold the nuclear segments
Additional Tips into the pupillary plane. Ultrasonic energy is often not
For very dense nuclei, several modifications of needed for removal of soft nuclei.
technique are useful. First, retract the phaco irrigation
sleeve, exposing slightly more phaco tip. This allows
the phaco tip to impale and to cut dense nuclei more
efficiently. Second, attempt to make the capsulor- Part C:
rhexis wider for very dense nuclei. Dense nuclei also Phaco Chop Techniques
tend to be very large nuclei. A small capsulorrhexis
makes delivery of large dense nuclear sections into the David F. Chang, MD
pupillary plane more difficult. Trypan blue facilitates
visualization of the capsule and makes creation of a
larger capsulorrhexis safer and easier, especially for Phaco chop refers to an advanced set of phaco
beginning surgeons. Third, make the angle of intersec- techniques that should not be attempted until one has
tion of the grooves 60 and 120 degrees, rather than at already mastered the divide and conquer method.1,2
90 degrees. Remove one of the 60-degree “quadrants” Compared to chopping, the latter method is easier to
first. This smaller segment will slide into the pupil- learn because it is much less dependent upon bimanual
Phaco Techniques 39
cases. In contrast to supracapsular techniques such as nuclear equator with the chopper tip within
phaco flip, however, the all-or-none requirement of the epinuclear space of the peripheral capsular
prolapsing the entire nucleus anteriorly through the bag prior to initiating the horizontally directed
capsulorrhexis is avoided. chop (see Figures 5-12A through 5-12C).3 Prior
to placing the chopper, the central anterior epi-
Decreased Reliance on the Red Reflex nucleus should be aspirated with the phaco tip
The increasingly brighter red reflex appearing at (see Figure 5-12A). This allows one to better vi-
the base of the trench during sculpting allows us to sualize and estimate the size of the endonucleus
judge the depth of the phaco tip and its proximity and the amount of separation between the en-
to the posterior capsule. In contrast, the instrument donucleus and the surrounding capsular bag.
maneuvers performed during chopping are more tactile The chopper tip touches and maintains con-
and kinesthetic. Because it is not necessary to directly tact with the anterior endonucleus as it travels
visualize the precise depth of the phaco tip, chopping is peripherally beneath the opposing capsulor-
advantageous in the presence of a poor red reflex, such rhexis edge (see Figure 5-12A). This ensures
as with diminished corneal clarity, smaller pupils, and that the chopper tip stays within the bag as it
mature nuclear or cortical cataracts (see Figure 5-12). descends to hook the endonucleus peripher-
In addition to improved surgical efficiency, safety is ally. Although some surgeons tilt the chopper
enhanced by these aforementioned attributes of reduced tip sideways to reduce its profile as it passes
phaco power, reduced zonular stress, decreased reliance underneath the capsular edge, this is generally
on the red reflex, and the supracapsular and central unnecessary unless the capsulorrhexis diameter
location of emulsification. These universal advantages is small. Instead, the horizontal chopper tip can
that both horizontal and vertical chopping share make remain in a vertical upright orientation because,
them optimal techniques for difficult and compli- like an elastic waistband, the capsulorrhexis will
cated cases. The improved ability to handle brunescent stretch to accommodate it without tearing (see
nuclei, white cataracts, loose zonules, posterior polar Figure 5-12B).
cataracts, crowded anterior chambers, capsulorrhexis Once it reaches the epi/endonuclear junction,
tears, and small pupils should be the primary motiva- the vertically oriented chopper tip descends
tion for a divide and conquer surgeon to transition to into the epinuclear space alongside the edge of
phaco chop.3,4,13,14 the endonucleus (see Figure 5-12C). This step
is easiest to perform with a smaller endonucleus
surrounded by a large epinucleus. Nudging the
Horizontal Phaco Chop nucleus with the chopper confirms that it has
The horizontal chopping technique relies upon hooked the equator and that it is within, rather
compressive force to fracture the nucleus. This exploits than outside, the capsular bag. Trypan blue dye
natural fracture planes in the lens created by the lamel- improves visualization of the anterior capsule
lar orientation of the lens fibers. Hydrodelineation is for this step and is a useful teaching aid (see
particularly important for horizontal chopping because Figures 5-12A through 5-12C).15
it decreases the diameter of the endonucleus that must Step 2. Impale and Immobilize the Nucleus With the
be peripherally hooked and divided by the chopper.3 In Phaco Tip. Next, one must deeply impale and
addition, the soft epinucleus provides a working zone immobilize the nucleus with the phaco tip (see
for the horizontal chopper where it can be manipulated Figure 5-12D). The phaco tip should be di-
peripheral to the endonuclear equator without overly rected vertically downward and positioned as
distending or tearing the capsular bag. Finally, the epi- proximally as possible in order to maximize the
nuclear shell restrains the posterior capsule from tram- amount of nucleus located in the path of the
polining toward the phaco tip as the final endonuclear chopper. If the depth of the phaco tip is too
fragments are emulsified. shallow, sufficient compression of the central
nucleus cannot occur. Once impaled, the phaco
tip holds and stabilizes the nucleus with vacu-
Horizontal Chop: um in foot pedal position 2.
A Step-by-Step Approach Step 3. Execute the First Chop. The chopper tip is
Step 1. Place the Chopper Tip in the Epinuclear drawn directly toward the phaco tip, and upon
Space. The critical first step is to hook the contact, the two tips are moved directly apart
Phaco Techniques 41
Figure 5-12A. Horizontal chop of mature white Figure 5-12B. The horizontal chopper passes
cataract following trypan blue capsular stain- beneath distal capsulorrhexis edge while oriented
ing. Chang microfinger-style horizontal chopper vertically.
(Katena, ASICO) maintaining contact with anterior
endonuclear surface.
Figure 5-12C. The horizontal chopper tip drops Figure 5-12D. Phaco tip impales centrally with 320
into the epinuclear space and hooks the nuclear mmHg vacuum.
equator.
Figure 5-12E. Chopper cuts toward the fixating Figure 5-12F. Instrument tips separate upon contact
phaco tip. to split the nucleus in half.
42 Chapter 5
Figure 5-12G. After rotating the nucleus 45 degrees Figure 5-12H. Chopper tip descends to hook the
clockwise, the chopper makes a second pass nucleus.
beneath the anterior capsule.
from each other (see Figures 5-12E and 5-12F). pression is taking place. This resistance is much
This separating motion occurs along an axis greater while chopping a dense nucleus, where
perpendicular to the chopping path and propa- the compressive force is followed by a sudden
gates the fracture across the remaining nucleus snap as the initial split occurs. To develop suf-
located behind and beneath the phaco tip. ficient compressive force, one must move the
The denser and bulkier the endonucleus, the chopper tip directly toward the phaco tip until
further the hemisections must be separated in they touch before commencing the sideways
order to cleave the connecting nuclear attach- separating motion. Veering the chopper tip to
ments. Thanks to the elasticity of the capsulor- the left as it approaches the phaco tip limits
rhexis, a wide momentary separation of large the compressive force and causes the nucleus
heminuclei will not tear the capsular bag. to swivel.
In order for the initial chop to succeed, enough Step 4. Remove the First Chopped Fragment. Upon
of the central endonucleus must lie within the completion of the initial chop, the nucleus
path of the chopper. Particularly if the anterior should be divided in half. After rotating the bi-
epinucleus has not been removed, it is easy to sected nucleus 30 to 45 degrees in a clockwise
misjudge the depth of the two instrument tips. direction, repeating the same steps of hooking
If the phaco tip is too superficial or too cen- the equator and chopping toward the phaco tip
tral, or the chopper tip is not kept deep enough creates a small, pie-shaped fragment (see Fig-
throughout the course of the chop, the nucleus ures 5-12G through 5-12K). The strong grip
will not fracture.3 Instead, the chopper will afforded by high vacuum facilitates elevation
only score or scratch the anterior surface. The of this first piece out of the bag. Insufficient
larger and denser the nucleus is, the more im- holding force may be the result of inadequate
portant and more difficult proper positioning vacuum settings or failure to completely oc-
of the two instrument tips becomes. A coun- clude the tip. Burst mode enhances the phaco
terproductive but natural tendency to elevate tip’s purchase of a firm nuclear piece by better
the chopper tip during the chop arises from a preserving the initial seal around the opening.
fear of perforating the posterior capsule. Step 5. Chop and Phaco Additional Nuclear Seg-
The tactile “feel” of the horizontal chop will ments. Every subsequent chop is a repetition
vary significantly as one proceeds up along the of these steps, and each wedge-shaped piece is
spectrum of nuclear density. A soft nucleus has emulsified as soon as it is created. Once half of
the consistency of soft ice cream and no resis- the capsular bag is vacated, the phaco tip can
tance is felt as the chopper is moved. With a impale and transport the remaining heminucle-
medium-density nucleus, the chopper encoun- us toward the center of the pupil. This allows
ters slight resistance, indicating that some com- the horizontal chopper tip to be positioned
Phaco Techniques 43
Figure 5-12I. Phaco tip impales and immobilizes Figure 5-12J. The second chop is executed in the
the nucleus with 320 mmHg vacuum. horizontal plane.
Figure 5-13A. Vertical chop. Chang vertical chop- Figure 5-13B. Central nucleus is impaled with
per (Katena, ASICO) in profile following aspiration the phaco tip using burst mode and 399 mmHg
of the anterior epinucleus. vacuum. Note the retraction of the phaco sleeve to
permit deeper penetration.
Figure 5-13C. Sharp vertical chopper tip is posi- Figure 5-13D. As the chopper tip descends into the
tioned just anterior to the phaco tip prior to incising nucleus, the phaco tip lifts slightly.
into the nucleus.
Figure 5-13E. Further penetration of the chopper tip Figure 5-13F. Sideways separation of the tips com-
results in a fracture line. pletes the division of the nucleus.
Phaco Techniques 45
G H
5-13E). This contrasts with the compressive the equator of the nucleus. Therefore, remov-
force produced by horizontal chopping. After ing fragments to vacate space within the cap-
initiating a partial-thickness split, the embed- sular bag early on provides no real advantage
ded instrument tips are used to pry the two (see Figure 5-13K).
hemisections apart (see Figure 5-13F). Just as
with horizontal chopping, this sideways sepa-
ration of the instrument tips extends the frac- Comparing Horizontal and
ture deeper until the remainder of the nucleus Vertical Chop—
is cleaved in half. The vertically chopped edg-
es appear sharp, like pieces of broken glass, be-
Which Technique?
Although I use both techniques with equal fre-
cause there is none of the crushing force that quency, each employs different mechanisms that have
characterizes horizontal chop. complimentary advantages and disadvantages. It is
Step 3. Chop All Fragments Before Removing Them. worth learning and utilizing both variations for this
In horizontal chop, sequentially removing each reason. Vertical chopping requires less dexterity of
newly created fragment provides the chopper the nondominant hand and is therefore easier for most
with increased working space within the cap- transitioning surgeons to learn. Vertical chopping also
sular bag. In contrast, one need not remove the requires a nucleus that is brittle enough to be snapped
vertically chopped pieces until the entire nu- in half, which means that it is ineffective for soft
cleus is fragmented. This is because the pres- nuclei.3 The ability of the horizontal chopper tip to
ence of the adjacent interlocking pieces better easily slice through a soft nucleus instead of fracturing
stabilizes and immobilizes the section that is it makes horizontal chopping the method of choice for
being chopped (see Figures 5-13G through 5- these cases.
13J). In contrast to horizontal chopping, the Horizontal chop is also my preference for loose
vertical chopper is never placed peripheral to zonule cases, such as traumatic cataracts. Because
46 Chapter 5
Figure 5-13J. With high vacuum (400 mmHg) grip Figure 5-13K. After vertically chopping the entire
of one segment, the two instrument tips pry the nucleus into multiple pieces, the fragments are
pieces apart. The sharp edges of the fragments elevated out of the capsular bag.
result from a shearing force.
of the oppositely directed, compressive instrument tor force initiates the fracture. This “diagonal” chop
forces, horizontal chop produces the least amount of therefore combines the mechanical advantages of both
nucleus movement or tilt. Finally, horizontal chop is strategies. With denser nuclei, one should also begin
more effective for subdividing smaller, mobile nuclear by sculpting a small pit or half trench centrally.16-18 By
fragments—particularly brunescent ones. Attempting entering at the base of the pit, the phaco tip can impale
to vertically chop and shear such fragments will often more deeply than would have been possible without
dislodge the small piece instead. Trapping and then this preliminary debulking. Retracting the irrigation
crushing fragments between the horizontal chopper sleeve further maximizes penetration. One should later
and the phaco tip will immobilize and divide them switch to horizontal chopping for subdividing brunes-
most effectively. cent fragments into smaller pieces.18 This will improve
The limitation of horizontal chopping is in its followability and reduce endothelial cell loss due to
relative inability to transect thicker brunescent nuclei. chatter and particle turbulence at the phaco tip.
Indeed, horizontal chopping should never be utilized
in the absence of an epinuclear shell since there will be
insufficient space in the peripheral bag to accommo- Stepwise Game Plan for
date the chopper. Frequently, the horizontally directed Learning Horizontal Chop
path of the chopper is not deep enough to sever the Of the two different techniques, the greater
leathery posterior plate of an ultra-brunescent nucleus. requirement for bimanual dexterity with the chopper
If this occurs, the partially chopped pieces will still be makes horizontal chopping more difficult to learn. The
connected at the apex, like flower petals. In such cases, most difficult steps are the initial ones—the first chop
one should try to inject a dispersive viscosurgical across the entire unsculpted diameter of the nucleus
device (OVD) through one of the incomplete cracks and removal of the first segment. Each subsequent
in the posterior plate to distance it from the posterior step becomes progressively easier as additional space
capsule. Since a dispersive OVD resists aspiration, the is vacated within the capsular bag. Logically, the saf-
surgeon can attempt to carefully phaco through the est strategy would allow surgeons to learn the steps in
remaining connecting bridges. the reverse order, starting with the easiest maneuvers
Because vertical chop is more consistently able to first.3,15 In the proposed game plan, the component
fracture the leathery posterior plate, it is well suited skills can be isolated, developed, and rehearsed while
for denser nuclei.3 With an ultra-brunescent lens, performing divide and conquer or stop and chop cases.
the vertical chopper should approach the embedded These principles and the same stepwise learning pro-
phaco tip from a diagonal angle. This provides more gression are equally applicable to mastering vertical
of a horizontal vector force that pushes the nucleus phaco chop.
against the tip, while the downward vertical vec-
Phaco Techniques 47
Summary 15.
corporated; 2004.
Chang DF. Transitioning to phaco chop. In: Chang DF, ed.
Horizontal and vertical chopping are variations that Phaco Chop: Mastering Techniques, Optimizing Technology, and Avoid-
rely upon different mechanisms to provide complemen- ing Complications. Thorofare, NJ: SLACK Incorporated; 2004.
tary advantages and common benefits. Mastering both 16. Vasavada AR, Singh R. Step-by-step chop in situ and separa-
tion of very dense cataracts. J Cataract Refract Surg. 1998;24:156-
methods affords surgeons greater flexibility in dealing 159.
with the wide range of nuclear densities and other sur- 17. Vanathi M, Vajpayee RB, Tandon R, et al. Crater-and-chop
gical variables.3,17 With dense lenses, one may employ technique for phacoemsulsification of hard cataracts. J Cata-
both techniques during the same case.17 Transitioning ract Refract Surg. 2001;27:659-661.
18. Chang DF. Comparing and integrating horizontal and vertical
surgeons should consider learning vertical chopping chopping. In: Chang DF, ed. Phaco Chop: Mastering Techniques,
first. In addition to increasing surgical efficiency for Optimizing Technology, and Avoiding Complications. Thorofare, NJ:
routine cases, chopping provides an increased margin SLACK Incorporated; 2004.
of safety for complicated cases (see Figure 5-12).3,13,14 19. Dewey S. Transition to chop: a non-impaling technique (vid-
A more detailed discussion of chopping tech- eo). American Academy of Ophthalmology Annual Meeting,
2003.
niques is available in the author’s book Phaco Chop:
Mastering Techniques, Optimizing Technology, and Avoiding
Complications,7 from which much of this content was
excerpted.
Phaco Techniques 49
Figure 5-16.
Step 4. Embed the Nucleus With the Phaco Needle; the nucleus in two pieces. As vacuum builds
Stabilize the Nucleus With the Vertical Ir- to occlusion, the endonucleus is held firmly
rigating Chopper, Then Lift With the Phaco by the phaco needle. At the moment occlu-
Needle to Create a Vertical Chop. The pha- sion is reached, the aspiration flow rate drops
co needle is then embedded proximally with to zero. Then move into foot position two so
high vacuum and 40% power (Figure 5-16). A that high vacuum is maintained and the power
vertical irrigating chopper is then used to split goes to zero (see Figure 5-15). The blade of the
Phaco Techniques 51
Figure 5-17. The phaco needle is pulled up and Figure 5-18. Following a second chop, the first
to the right as the chopper blade slices into the quadrant is lifted and pulled centrally where it
nucleus just in front of the needle and is then pulled is consumed with high vacuum and low levels of
to the left and down, effectively hemisecting the ultrasound power application.
cataract.
Figure 5-20. Following rotation and chopping of Figure 5-21. The final quadrant is consumed. The
the remaining heminucleus, the first quadrant is irrigation flow is maintained posteriorly to keep
consumed. the capsule well away from the phaco needle. The
chop element is turned to avoid contact with the
capsule.
Figure 5-22. The rim of the epinuclear shell is Figure 5-23. Because the cortex was consumed
grasped distally and pulled centrally, flipping the with the epinucleus, the capsule is pristine and
epinucleus and allowing its rapid consumption. ready for IOL insertion. The patient’s asteroid hya-
losis is clearly visible.
Every cataract surgeon should have a game plan This undesirable cascade of events can be
for when and how to perform an anterior vitrectomy averted by filling the anterior chamber with
following posterior capsule rupture. This chapter will a dispersive ophthalmic viscosurgical device
review the goals, the indications, and the techniques of (OVD) such as Viscoat (Alcon, Fort Worth,
an anterior vitrectomy. Understanding and mentally TX) or Healon D (AMO, Santa Ana, CA),
rehearsing these strategies will better prepare cataract prior to removing the phaco tip. Dispersive
surgeons to make correct decisions amidst the stress of OVDs are preferable to cohesive OVDs in the
an unexpected complication. face of a capsule tear. This is because disper-
sive OVDs are better at maintaining space and
resisting aspiration during phacoemulsification
Managing Posterior and irrigation/aspiration ([I/A] see Chapter 11).
Capsular Rupture— As the dispersive OVD is injected through the
side-port opening, the surgeon moves from
A Step-by-Step Approach foot pedal position 1 to 0. Once the chamber
Step 1. Attempt to Avoid Vitreous Loss Following is filled with OVD, the posterior capsule can-
Posterior Capsular Rupture. In many instanc- not bulge forward as the incision is unplugged.
es with a torn posterior capsule, it is possible to If one resumes phacoemulsification or cortical
avoid rupturing the hyaloid face. The surgeon cleanup, the same maneuver must be repeated
must avoid immediately withdrawing the pha- whenever the instruments are removed.
co tip upon recognizing a posterior capsular Step 2. Manage the Nucleus Following Posterior
defect. This abruptly unplugs the incision and Capsule Rupture. Early recognition of pos-
allows the anterior chamber to collapse. The terior capsule rupture is the key to avoiding a
sudden posterior pressure gradient will rupture dropped nucleus. It is much easier to remove
an intact hyaloid face, and vitreous will pro- the nucleus while it remains anterior to the
lapse to the incision, expanding the capsular posterior capsule defect. Because they ap-
rent in the process. proach the nucleus from above, subsequent
53
54 Chapter 6
Figure 6-2. Viscoat is injected via a pars plana scle- Figure 6-3. The Viscoat cannula is used to carefully
rotomy behind the descending nuclear fragments to lift the fragments into the anterior chamber.
provide immediate supplemental support.
(MVR) blade (Alcon, Katena) is used to make a duce the chance of touching the retina with a
pars plana sclerotomy located 3.5 mm behind metal spatula tip.
the limbus. An oblique quadrant should be se- Once a fragment descends into the mid or pos-
lected, and these steps can be performed under terior vitreous cavity, it is dangerous to blindly
topical anesthesia. The Viscoat cannula is then fish for it with any instrument. One should
advanced and aimed behind the nucleus under abandon the dropped nucleus and concentrate
direct visualization. The first step is to inject on removing the residual epinucleus and cortex,
a bolus of dispersive OVD behind the nucleus while preserving as much capsular support as
to provide immediate supplemental support possible. A thorough anterior vitrectomy must
(Figure 6-2). Periodic palpation of the globe be performed prior to inserting the IOL. Since
confirms that overinflation has not occurred. the vitreoretinal surgeon will later use a three-
If the nucleus is subluxated laterally, directing port fragmatome and vitrectomy technique to
OVD toward the region beneath it will often remove any retained nucleus, it is preferable
buoy the nucleus toward a more central posi- to insert an IOL through the cataract incision
tion. This is preferable to blindly probing with during the initial surgery if possible.
a metal spatula. One should not attempt to float Step 4. “Trap” Residual Lens Material in the Anterior
the nucleus into the anterior chamber using a Chamber and Manage Vitreous Loss Using
massive infusion of OVD alone. Unlike using a Dispersive OVD. Any residual nucleus re-
liquid perfluorocarbon in a vitrectomized cav- trieved with the Viscoat PAL technique can be
ity, an excessive injection of viscoelastic may removed using either of two techniques—re-
overinflate the globe and cause vitreous expul- suming phaco over a Sheet’s glide or convert-
sion through the sclerotomy. ing to a large incision manual extracapsular
Instead, the cannula tip itself should be used cataract extraction. At some point during this
to mechanically prop and levitate the nucleus sequence, the phaco or I/A tip may ensnare
into the anterior chamber (Figure 6-3). Small prolapsing vitreous. To avoid vitreous traction,
aliquots of additional dispersive OVD can be the surgeon must stop to perform an anterior
injected to help in the elevation and maneuver- vitrectomy before extraction of the remaining
ing of the nucleus. A small capsulorrhexis or lens material can be resumed.
pupil will stretch to accommodate the levita- The most common practice is to place a sepa-
tion of a greater diameter nucleus. The use of rate self-retaining irrigating cannula though a
the dispersive OVD to first support and then limbal paracentesis and to insert the vitrectomy
to reposition the nucleus prior to definitive probe through the phaco incision. However,
manual levitation is a major advantage of the there are multiple drawbacks to this approach.
Viscoat PAL variation. Because there is no as- First, the phaco incision is too large for the
piration involved, these PAL maneuvers should sleeveless vitrectomy instrument. This leak-
minimize iatrogenic vitreous traction and re- ing incision affords poor chamber stability and
56 Chapter 6
Figure 6-5. The sleeveless vitrectomy cutter is Figure 6-6. Bimanual irrigation/aspiration instru-
introduced via a pars plana sclerotomy and kept mentation is used to remove residual cortex follow-
behind the plane of the capsulorrhexis and pupil. ing the anterior vitrectomy.
This severs the transpupillary bands, but keeps the
vitrectomy separated from the partitioned anterior
chamber. The self-retaining infusion cannula (not ports again become entangled with vitreous,
shown) is placed through a limbal stab incision. one can repeat the Viscoat Trap maneuver fol-
lowed by additional pars plana anterior vit-
rectomy. Bimanual cortical I/A can then be
posterior vitreous cavity. The main advantage resumed.
is that using a properly sized sclerotomy will Step 7. Implant IOL Following Anterior Vitrectomy.
decrease incisional leak and vitreous prolapse If the capsulorrhexis is still intact, a three-piece
and should provide a better fluidic seal. Un- foldable or nonfoldable posterior chamber
like with a limbal incision, the vitrector need IOL can be placed in the ciliary sulcus. After
not traverse the anterior chamber and disrupt the haptics are first positioned in the sulcus,
the dispersive OVD partition, and it will not the optic should be captured behind the cap-
draw more vitreous forward into the anterior sulorrhexis if possible. This will ensure excel-
chamber. Performing the vitrectomy posterior lent centration because the optic cannot move.
to the plane of the pupil and capsulorrhexis First, one side of the optic is tilted back and
also decreases the chance of inadvertently cut- beneath the capsular rim before repeating the
ting either structure. If the capsulorrhexis is same maneuver for the other side. This maneu-
preserved, a foldable posterior chamber IOL ver can be very challenging following a vitrec-
may still be implanted in the ciliary sulcus. The tomy, however, and may not be possible if the
sclerotomy can be closed with a single inter- capsulorrhexis diameter is too large.
rupted 8-0 Vicryl suture. If the capsulorrhexis is not intact, there may
Following the retropupillary anterior vitrecto- still be enough capsular support to put the pos-
my, one can resume aspiration of the remaining terior chamber IOL in the sulcus. Amidst the
cortex or epinucleus trapped in the dispersive stress of managing an unexpected complication,
OVD-filled anterior chamber (Figure 6-6). some surgeons use the same foldable posterior
Step 6. Bimanual I/A of Residual Cortex. Once the chamber IOL they were planning to implant
capsule or zonules have ruptured, bimanual I/A in the capsular bag. This is not recommended
instrumentation is ideal for epinuclear and cor- for several reasons. First, moving the axial IOL
tical extraction for several reasons. Access to position slightly forward changes the effective
subincisional cortex is improved. The tighter power of the lens; you need to decrease the
paracentesis incisions better restrain vitreous power by about 1 diopter to compensate for
from prolapsing compared to using the phaco this position change. Second, nearly all fold-
incision. Finally, this is a lower flow fluidic sys- able lenses are 13.0 mm or less long, which is
tem compared to coaxial I/A. This permits the too small for the ciliary sulcus in many eyes.
surgeon to work in “slow motion” by lowering Although the lens may center well in the oper-
the irrigation bottle and decreasing the aspira- ating room, if it is too short, it can eventually
tion flow and vacuum settings. If the aspirating rotate and subluxate peripherally over time.
58 Chapter 6
For this reason, it is preferable to have longer rupture. However, there is a potentially fine line divid-
backup IOLs available, such as a 14.0-mm long ing maneuvers that are reasonable and safe from those
three-piece polymethylmethacrylate (PMMA) that are overly aggressive or dangerous. Cataract sur-
IOL, or the STAAR Surgical (Monrovia, CA) geons must be honest in assessing their own level of
AQ-2010 foldable three-piece silicone lens comfort and expertise. Timely surgical management
with an overall length of 13.5 mm. A single- of a dropped nucleus by a vitreoretinal surgeon at a
piece acrylic IOL should never be placed in the later date is always preferable to overstepping this fine
sulcus because the overall length is too short line.10
and the haptics are not rigid. In addition to
poorly centering the lens, the thicker, sharp-
edged haptics will rub against the back surface References
1. Michelson MA. Use of a Sheets’ glide as a pseudoposterior
of the iris, causing iris transillumination defects capsule in phacoemulsification complicated by posterior cap-
and pigmentary glaucoma. sule rupture. Eur J Implant Surg. 1993;570-572.
Although anatomical studies have shown that 2. New PAL method may save difficult cataract cases. Ophthal-
there is no reliable way to gauge the ciliary mology Times. 1994;19:51.
3. Chang DF, Packard RB. Posterior assisted levitation for nu-
sulcus diameter according to external land-
cleus retrieval using Viscoat after posterior capsule rupture.
marks, it is helpful to measure the white-to- J Cataract Refract Surg. 2003;29:1860-1865.
white corneal diameter intraoperatively. If it 4. Chang DF. Managing residual lens material after posterior
measures 11.5 mm or less, a standard 13.0-mm capsule rupture. Techniques in Ophthalmology. 2003;1(4):201-
long foldable IOL will probably center well in 206.
5. Chang DF. Strategies for managing posterior capsular rup-
the sulcus. Absent capsulorrhexis capture, a ture. In: Chang DF, ed. Phaco Chop: Mastering Techniques, Op-
longer IOL should be considered, however, if timizing Technology, and Avoiding Complications. Thorofare, NJ:
the sulcus diameter is 12.0 mm or larger. An SLACK Incorporated; 2004.
advanced option is to anchor the sulcus-fixated 6. Burk S, Sugar J, Farber MD. Comparison of the effects of two
viscoelastic agents, Healon and Viscoat, on postoperative in-
IOL by suturing one haptic to the iris. A single
traocular pressure after penetrating keratoplasty. Ophthalmic
10-0 polypropylene McCannel suture around Surg. 1990;21:821-826.
one of the haptics will be enough to keep the 7. Probst LE, Hakim OJ, Nichols BD. Phacoemulsification
lens from rotating or decentering.9 Finally, a with aspirated or retained Viscoat. J Cataract Refract Surg.
properly sized and well-placed anterior cham- 1994;20:145-149.
8. Burk SE, Da Mata AP, Snyder ME, et al. Visualizing vitreous
ber IOL is always an excellent option if poste- using Kenalog suspension. J Cataract Refract Surg. 2003;29:645-
rior chamber IOL support is not ideal. 651.
9. Chang DF. Siepser slipknot for McCannel iris-suture fixa-
tion of subluxated intraocular lenses. J Cataract Refract Surg.
Final Comments 2004;30:1170-1176.
Cautious adherence to these principles described 10. Scott IU, Flynn HW, Jr, Smiddy WE, et al. Clinical features
and outcomes of pars plana vitrectomy in patients with re-
above may help surgeons to reduce the chance of tained lens fragments. Ophthalmology. 2003;110:1567-1572.
dropping the nucleus following posterior capsular
Chapter
7
Management of the
Small Pupil
Robert H. Osher, MD, and James M. Osher, MD
Managing the Small Pupil work near the iris with diminished risk of inadvertent
iris and capsular damage. This is an approach that has
A small pupil increases the likelihood of intraop-
erative complications and was originally considered stood the test of time.
by Dr. Charles Kelman to be a contraindication to Today’s surgeon has a wide range of choices for
phacoemulsification. In 1985, Dr. Robert Osher pre- handling the small pupil. It is the purpose of this chap-
sented a series of small pupil cases that were success- ter to review these different options and also discuss
fully managed, utilizing a modification of the United the use of reduced phaco parameters in the manage-
Surgical phacoemulsification machine that provided an ment of specific surgical challenges associated with the
improvement in fluidics control. Prior to the introduc- miotic pupil. The second issue of the 2008 Video Journal
tion of this modification, phaco machines had only of Cataract and Refractive Surgery provides video footage
maximum and minimum settings with a bottle height of each of the following categories.
fixed by a cumbersome rod. Osher introduced a new
concept: surgeon-controlled vacuum with continuous Options for Pharmacologic Dilation
irrigation from a bottle of adjustable height, regulated Traditionally, the use of topical mydriatics in
by an automated IV pole. The result of this innovation the form of a sympathomimetic agent combined
was a reduction of anterior chamber turbulence and with a parasympathetic blocker fulfilled the goal of
improved stability. achieving mydriasis. Adding epinephrine to the bal-
Introducing a technique called slow-motion phaco- anced salt solution (BSS) infusion has supplemented
emulsification, Osher explained that, by lowering and maintained dilation. With the introduction of
parameters, he was able to perform phacoemulsifica- Ocufen (Allergan, Irvine, CA), it was realized that a
tion effectively in spite of poorly dilating pupils. He nonsteroidal anti-inflammatory drug (NSAID) also
presented a paper entitled “The GASP Technique” contributed to maintenance of dilation by blocking the
(Golly, Another Small Pupil) at the American Intra- miotic effect of prostaglandins released when the iris
Ocular Implant Society meeting and Dr. Kelman was was manipulated.1 A cotton pledget soaked in neosyn-
a discussant. Dr. Kelman agreed that by modifying ephrine and placed at the limbus was very effective in
fluidic behavior and reducing surge, the surgeon could achieving maximal dilation but fell out of favor because
59
60 Chapter 7
A B
of systemic concerns. However, IOL Tech, a French allowing the pupil to narrow. In order to use Healon 5
company, introduced Mydriasert, a sustained-release most effectively as a viscomydriatic, it is important to
product placed in the inferior fornix. Intracameral understand how to keep the Healon 5 in the anterior
mydriatics were introduced by Dr. Bjórn Lundberg chamber.
and Dr. Andes Behndig2 and by Drs. Cionni, Barros, Healon 5 behaves as a dispersive OVD when the
Kaufman, and Osher.3 The instillation of 1% phen- vacuum is less than about 200 mmHg and the aspira-
ylephrine directly onto the anterior capsule has been tion rate is about 25 cc/min or less. If the vacuum
found to be helpful in maintaining pupil size in patients or aspiration rate exceeds these levels, the material
with intraoperative floppy iris syndrome (IFIS) by behaves in a cohesive manner. This unique variable vis-
Drs. Monvikar and Allen4 and Drs. Gurbaxani and coelastic behavior allows the surgeon to keep Healon
Packard.5 Dr. Joel Shugar advocated “Shugarcaine” 5 in the anterior chamber during the phacoemulsifica-
using 1:1,000 bisulfite-free epinephrine that is mixed tion if vacuum and aspiration rates are maintained at
in a 1:3 dilution with three parts BSS+ and one part low levels, and then remove the OVD material readily
nonpreserved lidocaine 4%. Approximately 1 mL of at the conclusion of the procedure, using higher vacu-
this mixture is slowly injected into the anterior cham- um and aspiration settings. Since ultrasonic energy is
ber before instillation of the ophthalmic viscosurgical capable of shattering the tightly packed, long-chained
device (OVD).6 molecules, it is best to avoid anterior chamber emul-
sification as fractured Healon 5 is less likely to resist
Viscomydriasis With Healon 5 aspiration forces. Moreover, if a high bottle height cre-
Pupillary dilation can be achieved by the injection ates a high pressure in the eye, there is a greater likeli-
of an OVD (viscomydriasis). Healon 5 (AMO, Santa hood that the Healon 5 will be “forced” either into the
Ana, CA), a high molecular weight sodium hyaluronate aspiration port or out through the incision, especially
OVD, is uniquely capable of dilating and maintaining a when the latter is poorly constructed or distorted.
wide pupil.7 By mechanically moving iris tissue toward (See below for a step-by-step explanation of the use of
the angle, Healon 5 is effective in expanding the pupil Healon 5 in IFIS.)
(Figure 7-1). Moreover, Healon 5 behaves as a highly
retentive, semi-solid material that bows the iris poste- Stretching the Pupil
riorly, resulting in a deeper chamber with additional Cutting or stretching the pupillary sphincter is
dilation. Some pupils will enlarge dramatically with another method of obtaining a larger pupil during
Healon 5, while others will dilate less. In virtually all surgery. The use of scissors to create multiple sphinc-
patients, however, Healon 5 produces some useful terotomies was popular during the 1990s before Dr.
viscomydriasis. While it is easy to obtain initial visco- Luther Fry popularized the pupil stretch technique.8
mydriasis, it is often necessary to reinject Healon 5 in Pupil stretching is accomplished with two blunt instru-
order to maintain the effect. This is because either the ments (eg, hooks, collar buttons, retractors, etc). Under
OVD is aspirated or it escapes through the incision, the protection of an OVD, the hooks are introduced
Management of the Small Pupil 61
Peripupillary Membranectomy
Another technique involving the pupillary sphinc-
ter is peripupillary membranectomy, which Osher
described in the early 1980s.9 In cases of uveitis or
chronic pilocarpine usage, the pupil is bound down by
synechiae, which prohibit the pupil from dilating. A
string-like fibrotic membrane at the border of the pupil
can often be stripped, which serves to release the pupil
(Figure 7-3).
C
Iris Hooks
Iris retractors are another option for achieving
mechanical dilation. Prior to the introduction of micro-
scopic hooks, surgeons occasionally used a retraction
suture or deliberately prolapsed the iris into an incision
to achieve a dilatory effect. Later, metal, fine wire, and
prolene retractors were introduced by Drs. Mackool,10
Engels,11 and deJuan,12 respectively. Iris hooks may be
disposable or reusable and may be anchored by adjust-
able corks, sliding tabs, or weights. The technique for
inserting the iris hooks requires a carefully thought-
Figure 7-2. Fry pupil stretch technique. out plan since tenting the iris toward the cornea
should be avoided (Figure 7-4). In cases with a shallow
chamber, an instrument introduced through a remote
and placed in the same meridian 180 degrees away incision may help to usher the hook into position. The
from one another. The iris is then stretched with each surgeon may vary the number of iris hooks as well as
instrument simultaneously toward the angle, momen- the incisions. A recent variant has been introduced by
tarily held, and then released (Figure 7-2). A second Drs. Oetting and Omphrey. These authors described
stretch can be performed 90 degrees away from the the placement of a hook below the primary cataract
first if the surgeon desires. After pupil stretching, when incision, which creates a diamond-shaped pupil for
OVD is reinjected, an enlargement of the pupil usually optimal visualization and manipulation of the ultra-
occurs. While small sphincter ruptures are visible at sound and irrigation and aspiration (I/A) tips.13 While
the slit lamp following surgery, the pupil generally tiny sphincter ruptures can be observed at the slit lamp
retains a physiologic shape and functions normally. following surgery, the pupil usually regains normal
size and function. Although some effort and time is
62 Chapter 7
such as indocyanine green, introduced by Horiguchi et many years, the small pupil was one of Dr. Kelman’s
al,21 and trypan blue, first described by Melles et al.22 contraindications to phacoemulsification. Fortunately,
Staining the anterior capsule with these dyes can be advances in machine technology, viscosurgery, surgi-
accomplished by a number of different techniques, but cal techniques, and devices for mechanical dilation
we prefer a three-step method utilizing Healon 5.23 have made operating within the small pupil far more
safe and compatible with an excellent visual outcome.
Capsule Staining, Utilizing Healon 5:
A Step-by-Step Approach
Step 1. Place Healon 5 over the Anterior Capsule. References
1. Duffin RM, Camras CB, Gardner SK, Pettit TH. Inhibitors of
Healon 5 is injected into the anterior cham- surgically induced miosis. Ophthalmology. 1982;89:966-979.
ber, being careful not to overfill it. Healon 5, 2. Lundberg B, Behndig A. Intracameral mydriatics in phaco-
which is highly retentive under conditions of emulsification cataract surgery. J Cataract Refract Surg.
low flow, provides a very stable chamber for 2003;29:2366-2371.
3. Cionni RJ, Barros MG, Kaufman AH, Osher R. Cataract sur-
intraocular manipulation. gery without preoperative eyedrops. J Cataract Refract Surg.
Step 2. Create a Space Between Healon 5 and the 2003;29:2281-2283.
Anterior Capsule. Inject BSS directly onto the 4. Monvikar S, Allen D. Cataract surgery management in pa-
anterior capsule, elevating the Healon 5 into tients taking tamsulosin staged approach. J Cataract Refract
Surg. 2006;32:1611-1614.
the corneal dome while creating a thin layer of
5. Gurbaxani A, Packard R. Intracameral phenylephrine to pre-
fluid directly over the anterior capsule. vent floppy iris syndrome during cataract surgery in patients
Step 3. Inject Capsular Staining Dye Into the Supra- on tamsulosin. Eye. 2007;21:331-332.
capsular Space. Trypan blue is then placed 6. Shugar J. Use of epinephrine for IFIS prophylaxis. J Cataract
into the thin BSS-filled space. This results in Refract Surg. 2006;32:1074-1075.
7. Osher R. Viscomydriasis. Video J Cataract Refract Surg.
an even stain of the anterior capsule without 2002;XVIII(2).
creating an “ink blot” in the anterior chamber 8. Fry L. Pupil stretching. Video J Cataract Refract Surg.
or forcing dye under pressure through the zon- 1995;XI(1).
ules into the vitreous cavity. The Osher dye 9. Osher R. Peripupillary membranectomy. Video J Cataract Re-
fract Surg. 1991;VII(3).
cannula (Storz [Bausch & Lomb, San Dimas,
10. Mackool R. Small pupil enlargement during cataract extrac-
CA] and Crestpoint Management [St Louis, tion: a new method. J Cataract Refract Surg. 1992;18:523-526.
MO]) has the port on the posterior surface of 11. Engels T. Peripupillary membranectomy. Wire retractor. Vid-
the cannula that allows the dye to be delivered eo J Cataract Refract Surg. 1995;XI(1).
precisely onto the anterior capsular surface. It 12. deJuan E Jr, Hickingbotham D. Flexible iris retractor (letter).
Am J Ophthalmol. 1991;111:776-777.
may be necessary to inject additional BSS or 13. Oetting TA, Omphrey LC. Modified technique using flex-
Healon 5 to gain optimal visualization before ible iris retractors in clear corneal surgery. J Cataract Refract
proceeding with the capsulorrhexis. Surg. 2002;28:596-598.
14. Graether J. Silicone expander. Video J Cataract Refract Surg.
1995;XI(1).
Small Pupil Associated With Iridodialysis 15. Siepser S. Expansile hydrogel ring. Video J Cataract Refract
Rarely, the anterior segment surgeon will encounter Surg. 1995;XI(1).
a traumatic cataract associated with extensive iris disin- 16. Akman A, Yilmaz G, Oto S, et al. Comparison of various
sertion. Depending upon the extent of the damage, the methods of phacoemulsification in eyes with a small pupil sec-
ondary to pseudoexfoliation. Ophthalmology. 2004;111:1693-
pupil may appear miotic and eccentric. The iris must 1698.
be reattached to the sclera by a series of nonabsorbable 17. Auffarth G, Reuland A, Heger T, et al. Cataract surgery in
horizontal mattress sutures. Following the repair of the eyes with iridoschisis using the Perfect Pupil iris extension
iridodialysis, the use of iris hooks or pupil-expanding system. J Cataract Refract Surg. 2005;31:1877-1880.
18. Malyugin B. Russian solution to small pupil phaco and the
devices may be helpful if pharmacologic dilation still
tamsulosin floppy iris syndrome. Video J Cataract Refract Surg.
fails to achieve an appropriate pupil size. 2007;XXIII(1).
19. Chang DF, Campbell JR. Intraoperative floppy iris syn-
drome associated with tamsulosin. J Cataract Refract Surg.
Summary 2005;31:663-673.
Dr. Charles Kelman, the father of phacoemulsi- 20. Osher R. Healon 5 in IFIS. Video J Cataract Refract Surg.
2005;XXI(2).
fication, alerted his disciples to the perils of operat-
21. Horiguchi M, Miyake K, Ohta I, et al. Staining of the lens
ing upon the cataract patient with a small pupil. For
66 Chapter 7
capsule for circular continuous capsulorrhexis in eyes with Cataract Refract Surg. 1999;25:7-9.
white cataract. Arch Ophthalmol. 1998;116:535-537. 23. Marques DM, Marques FF, Osher RH. Three-step technique
22. Melles G, deWaard P, Pameyer J, et al. Trypan blue capsule for staining the anterior lens capsule with indocyanine green
staining to visualize the capsulorrhexis in cataract surgery. J or trypan blue. J Cataract Refract Surg. 2004;30:13-16.
Chapter
8
The Phaco Machine
Understanding the Equipment to Take
Advantage of Contemporary Phaco Techniques
Introduced by Dr. Charles Kelman in 1962, phaco- sandths of an inch). Most machines operate in the 2 to
emulsification machines have undergone constant 4 mil range. One mil is 25 microns. Therefore, most
improvement, ever increasing both their complex- phaco needles travel a distance of 50 to 100 microns.
ity and safety. There is one principle, however, that The longer the stroke length, the greater the physical
remains unchanged. All phaco machines consist of a impact on the nucleus and the greater the generation
computer to generate electrical signals and a trans- of cavitation energy. Longer stroke lengths, like higher
ducer to turn these electronic signals into mechanical frequencies, however, tend to generate extra heat.
energy. The energy thus produced is passed through Stroke length is determined by foot pedal excur-
a hollow needle and is controlled within the eye to sion in position 3 during linear control of phaco. As the
overcome the inertia of the lens and emulsify it. Once foot pedal is depressed, the stroke length and therefore
turned into emulsate, fluidic systems remove the emul- the power increase to the preset maximum. New foot
sate, replacing it with balanced salt solution (BSS). pedals allow the surgeon to control the throw length
in each major division, increasing the capability of
Power Generation the surgeon to manage control of both the fluidic and
ultrasonic components of phaco.
Power is created by an interaction between fre-
quency and stroke length.
Frequency is defined as the speed of the needle Tuning
movement. The manufacturer of the machine deter- The central processing unit (CPU) of modern
mines it. Presently, most machines operate at a fre- phaco machines recognizes when the phaco needle
quency between 27,000 cycles per second (Hz) to passes into different intraocular media. For example,
50,000 cycles per second. This frequency range is the resistance of the aqueous is less than the resistance
efficient for nuclear emulsification. Lower frequencies of the cortex, which in turn, is less than the resistance
become less efficient and higher frequencies create of the nucleus. As the resistance to the phaco tip var-
excess heat. ies to maintain maximum efficiency dependent on
Stroke length is defined as the length of the needle the machine, small alterations in frequency or stroke
movement. This length is generally 2 to 6 mils (thou- length are created by the tuning circuitry in the CPU.
67
68 Chapter 8
This is important to minimize the excessive generation angle from the bevel, and a 15-degree tip 15 degrees
of ultrasonic energy, which is harmful to the intraocu- from the bevel. A 0-degree tip creates the cavitation
lar contents. The surgeon will subjectively determine wave directly in front of the tip and the focal point is
good tuning circuitry by a sense of smoothness and 0.5 mm from the tip. The Kelman tip has a broad band
power. of powerful cavitation that radiates from the area of the
angle in the shaft. A weak area of cavitation is devel-
Phaco Energy oped from the bevel but is inconsequential.
The actual tangible forces that emulsify the nucle- Taking into consideration analysis of enhanced
us are thought to be a blend of the “jackhammer” effect cavitation, it can be concluded that phacoemulsifica-
and cavitational energy.1 The jackhammer effect is the tion is most efficient when both the jackhammer effect
physical striking of the needle against the nucleus. The and cavitation energy are combined. To accomplish
cavitation effect is more convoluted. Recent studies this, the bevel of the needle should be turned toward
indicate that there are two kinds of cavitational energy. the nucleus or nuclear fragment. This simple maneuver
One is transient cavitation and the other is sustained will cause the broad bevel of the needle to strike the
nucleus. This will enhance the physical force of the
cavitation.
needle striking the nucleus. In addition, the cavitation
force is then concentrated into the nucleus rather than
Transient Cavitation
away from it. Finally, in this configuration, the vacuum
The phaco needle, moving through a liquid medi-
force can be maximally exploited as occlusion is encour-
um at ultrasonic speeds, gives rise to intense zones
aged. This causes energy to emulsify the nucleus and
of high and low pressure. Low pressure, created with
be absorbed by it. A 0-degree tip automatically focuses
backward movement of the tip, pulls dissolved gases
both the jackhammer and cavitational energy directly
out of solution, thus producing micro bubbles. Forward
in front of it. When the bevel is turned away from the
tip movement then creates an equally intense zone
nucleus, the cavitational energy is directed up and away
of high pressure. This initiates compression of the
from the nucleus toward the iris and endothelium.
micro bubbles until they implode. At the moment of
implosion, the bubbles create a temperature of 7204˚C Sustained Cavitation
degrees and a shock wave of 5,171,100 mbar. Of the If phaco is energized beyond 4 milliseconds, tran-
micro bubbles created, 75% implode, amassing to cre- sient cavitation with generation of micro bubbles and
ate a powerful shock wave radiating from the phaco shock waves ends. The bubbles then begin to vibrate
tip in the direction of the bevel with annular spread. without implosion. No shock wave is generated.
However, 25% of the bubbles are too large to implode. Therefore, there is no emulsification energy produced.
These micro bubbles are swept up in the shock wave Sustained cavitation is ineffective for emulsification.
and radiate with it. Transient cavitation is a violent Water bath, hydrophonic studies indicate that
event. The energy created by transient cavitation exists transient cavitation is significantly more powerful than
for no more than 4 milliseconds and is present only in sustained cavitation. With this information in mind,
the immediate vicinity of the phaco tip and within its it would appear that continuous phaco is best used
lumen. It is this form of cavitation that is thought to to emulsify the intact nucleus, held in place by the
generate the energy responsible for emulsification of capsular bag, during the sculpting phase of divide and
cataractous material. Additionally, transient cavitation conquer or stop and chop. Jackhammer energy is most
is instrumental in clearing nuclear fragments within the important for emulsification in this setting.
phaco needle, preventing repetitive needle clogging. Transient cavitation is maximized during micro-
The transient cavitational energy can be directed pulse phaco. This is best used during phaco of the
in any desired direction. The angle of the bevel of the nuclear fragments in the later phase of the above two
phaco needle governs the direction of the generation procedures or during phaco chop procedures.
of the shock wave and micro bubbles.
I have developed a method of visualization of these
forces called “enhanced cavitation.” Using this process, Modification of
it can be seen that with a 45-degree tip, the cavitation Phaco Power Intensity
wave is generated at 45 degrees from the tip. Similarly, Application of the minimal amount of phaco power
a 30-degree tip generates cavitation at a 30-degree intensity necessary for adequate emulsification of the
The Phaco Machine 69
nucleus is desirable. Unnecessary power intensity is a allows inflow of irrigating fluid in the micro cavity
cause of heat with subsequent wound damage, endo- between the phaco tip and nuclear fragment. This
thelial cell damage, and iris damage with alteration of renewal of fluid is important to provide new fuel for
the blood-aqueous barrier. Phaco power intensity can transient cavitation as well as for cooling of the phaco
be modified by the following: tip.
Alteration in stroke length The cool phaco tip has been termed cold phaco.
Alteration of duration This is a misnomer as the phaco tip is not cold but
Alteration of emission warm. However, studies indicate that it will not devel-
op a temperature greater than 55°C, the temperature
Alteration in Stroke Length required to create a wound burn. Phaco techniques
Stroke length is determined by foot pedal adjust- such as phaco chop utilize minimal periods of power in
ment. When set for linear phaco, depression of the foot pulse mode, or micro-pulse mode, to reduce superflu-
pedal will increase stroke length and therefore power. ous power delivery to the anterior chamber. In addi-
New foot pedals, such as those found in the AMO tion, the use of pulse mode, or micro-pulse mode, to
(Santa Ana, CA) Sovereign/Signature and the Alcon remove the epinucleus provides for an added margin
(Fort Worth, TX) Infinity, permit surgeon adjustment of safety. When the epinucleus is emulsified, the poste-
of the throw length of the pedal in position 3. This can rior capsule is exposed to the phaco tip and may move
refine power application. toward it due to surge. Activation of pulse phaco, or
The Bausch & Lomb (Rochester, NY) Millennium/ micro-pulse phaco, will create a deeper anterior cham-
Stellaris dual linear foot pedal permits the separation ber to work within. This occurs because, as noted pre-
of the fluidic aspects of the foot pedal from the power viously, each period of phaco energy is followed by an
elements. interval of no energy. The epinucleus is drawn toward
the phaco tip during the interval of absence of energy,
Alteration of Duration producing partial occlusion and interrupting outflow.
The duration of application of phaco power has a This allows inflow to deepen the anterior chamber
dramatic effect on overall power delivered. Usage of immediately prior to onset of another pulse of phaco
pulse or burst mode phaco will considerably decrease energy. The surgeon will recognize the outcome as
overall power delivery. New machines allow for a operating in a deeper, more stable anterior chamber.
power pulse of duration alternating with a period of
aspiration only. Burst mode (parameter is machine Pulse Shaping
dependent) is characterized by 80- or 120-millisecond This is a modification of varying power duration.
periods of power combined with fixed short periods By changing the morphology of the power burst in
of aspiration only. Pulse mode utilizes fixed pulses of hyper-pulse phaco, the power can be delivered with
power of 50 or 150 milliseconds with variable short greater effectiveness. Different manufactures have
periods of aspiration only. developed different burst morphology.
AMO (Whitestar/Signature) uses increased con-
Micro-Pulse (Hyper-Pulse) trol and efficiency (ICE). A 1-millisecond punch of
Through the development of highly responsive power with an amplitude of 7% of the preset power
and low mass piezo crystals, combined with software maximum is delivered at the beginning of each burst.
modifications, the manufacturers of phaco machines This “kicker” has two consequences. First, it drives the
have shortened the cycle of on and off time. This pro- nucleus away from the phaco tip sufficiently to aug-
cess, patented by AMO, is called “micro-pulse.” This ment partial occlusion phaco. Second, it allows the
technology is now available in most phaco machines. phaco tip to accelerate to the preset velocity almost
A duty cycle is defined as the length of time of instantly. The result is more effective phaco of the
power on combined with power off. The short bursts fragments.
of phaco energy followed by a short period without Bausch & Lomb (Millennium/Stellaris) has taken a
phaco energy allows two important events to occur. different approach. They bring the power up to maxi-
First, the period without phaco energy permits the mum more slowly. They believe the slow increase in
nuclear material to be drawn toward the phaco tip power enhances partial occlusion by not pushing the
to increase efficiency. Second, the absence of power fragment away from the phaco tip.
70 Chapter 8
or zero vacuum is helpful during sculpting of hard or um from power. In this way, flow or vacuum
large nucleus where the high power intensity of the tip can be lowered before beginning the emulsi-
may be applied near the iris or anterior capsule. Zero fication of an occluding fragment. The emul-
vacuum will prevent inadvertent aspiration of the iris sification therefore occurs in the presence of
or capsule, avoiding significant morbidity. a lower vacuum or flow so that surge is mini-
mized.
ABS: Alcon Infinity/Legacy—The ABS tips
Surge have 0.175-mm holes drilled in the shaft of the
A fundamental limiting factor in the selection needle. During occlusion, the hole provides
of high levels of vacuum or flow is the development for a constant alternate fluid flow. This will
of surge. When the phaco tip is occluded, flow is cause dampening of the surge on occlusion
interrupted and vacuum builds to its preset level. break.
Emulsification of the occluding fragment then clears
the occlusion. Flow instantaneously begins at the pre-
Nonlongitudinal Phaco: Modification of Fluid
set level in the presence of the high vacuum level. In
addition, if the aspiration line tubing is not reinforced Control by Power Modulations
to prevent collapse (a function of tubing compliance), Three significant, trend-setting technologies have
the tubing will have constricted during the occlusion. revolutionized the way power is modulated. When
It then expands on occlusion break. This expansion employing these power modulations, the duration
is an additional source of vacuum production. These of power operation and the motion of needle move-
factors trigger a rush of fluid from the anterior seg- ment are significant on their effect on fluid flow and
ment into the phaco tip. The fluid in the anterior occlusion. These modulations have an effect on the
chamber may not be replaced rapidly enough by infu- fluidic balance during phaco, which is as important to
sion to prevent shallowing of the anterior chamber. chamber maintenance and ease of removal of nuclear
Therefore, with sudden volume reduction in the ante- fragments as the preset vacuum and flow.
rior chamber there is succeeding rapid anterior move- Micro-Pulse Phaco—Discussed previously,
ment of the posterior capsule. This abrupt forceful the rapid 4-millisecond power on cycle maxi-
stretching of the bag around nuclear fragments (espe- mizes the development of transient cavitation-
cially if the fragment is hard with jagged edges) may al energy. All cavitational energy in the 4-mil-
be a cause of capsular tears. In addition, the posterior lisecond burst is capable of emulsifying tissue.
capsule can be literally sucked into the phaco tip, tear- The ensuing 4-millisecond period of aspiration
ing it. The magnitude of the surge is contingent on replenishes fluid at the phaco tip and cools it.
the duration of occlusion and the pre-surge settings The use of micro-pulse phaco is necessary to
of flow and vacuum. create the shift in phaco technique from post-
Classically selecting lower levels of flow and vac- occlusion phaco to partial-occlusion phaco.
uum control surge. The phaco machine manufacturers Torsional Phaco (Alcon Infinity Ozil Hand-
help to decrease surge by providing noncompliant piece)—Classic phaco has utilized a phaco tip
aspiration tubing that will not constrict in the pres- that moves forward and backward, or longitu-
ence of high levels of vacuum. More important are the dinally. Torsional phaco is defined as a 32-kHz
following noteworthy new technologies: oscillatory movement of an angled (Kelman)
CASE: AMO Sovereign/Signature—Micro- phaco tip. This can be combined with longitu-
processors sample vacuum and flow param- dinal movement of the needle at 44 kHz. The
eters 50 times a second, creating a “virtual” torsional component is linear and the longitu-
anterior chamber model. At the moment of dinal component can be micro-pulse. The po-
occlusion, the computer senses the decrease tential flexibility of this system is enormous.
in flow and instantaneously slows the pump Ellips Phaco (AMO Signature)—In this sys-
to stop surge production. The Alcon Infinity tem the longitudinal movement of the phaco
works in a similar manner. tip at 38 kHz is combined with a transversal
Dual Linear: Bausch & Lomb Millennium/Stel- motion at 26 kHz. The resultant movement of
laris—The dual linear foot pedal can be pro- the needle can be described as prolate-spher-
grammed to separate both the flow and vacu- oid (shaped much like an egg cut in half).
72 Chapter 8
is applied at a level high enough to emulsify the frag- in the divide and conquer section. Each fragment
ment without driving it away from the phaco tip. and the remaining heminucleus are removed in turn.
“Chatter” is defined as a fragment bouncing away Epinucleus and cortex removal are also performed as
from the phaco tip due to excessively aggressive appli- noted above.
cation of phaco energy.
Phaco Chop
Epinucleus and Cortex Removal Phaco chop requires no sculpting. Therefore,
If cortical cleaving hydrodissection has been per- the procedure is initiated with high vacuum and
formed, the endonucleus is removed first as noted flow and linear pulsed or micro-pulse phaco power.
above. The result is a shell of epinucleus and cortex. Nonlongitudinal phaco does not work well for the
For removal of epinucleus and cortex, the vacuum is actual chopping as the shaving movement of the phaco
decreased while flow is maintained. This will allow tip prevents an adequate vacuum seal to assist chop-
for grasping of the epinucleus just deep to the ante- ping and fragment mobilization. For a 0-degree tip,
rior capsule. The low vacuum will help the tip hold especially when emulsifying a hard nucleus, a small
the epinucleus on the phaco tip without breaking off trough may be required to create adequate room for
chunks. High vacuum results in breaking off pieces the phaco tip to push deep into the nucleus. For a
of epinucleus and cortex, making it more difficult to 15- or a 30-degree tip, the tip should be rotated bevel
remove. With the fluid parameters balanced, the epi- down to engage the nucleus. The phaco tip should
nucleus/cortex scrolls around the equator and can be be encased within the endonucleus with the minimal
pulled to the level of the iris. There, low power pulsed amount of power necessary. All chopping proce-
or hyper pulse phaco is employed for emulsification. dures require 1 mm of exposed phaco tip to create
adequate holding power for chopping. If the phaco
Stop and Chop Phaco tip is inserted into the nucleus with excess power, the
Groove creation is performed as noted above adjacent nucleus will be emulsified, creating a poor
under divide and conquer sculpting techniques. Once seal between nucleus and tip. This will make it impos-
a single deep groove is adequate vacuum and flow sible to remove fragments, as the tip will just “let go” of
are increased to improve holding capability of the the nuclear material. Additionally, the bevel should be
phaco tip. The nucleus is rotated 90 degrees and the turned toward the fragment to create a seal between
phaco tip is driven into the mass of one heminucleus tip and fragment, allowing vacuum to build and create
using pulsed linear phaco. The sleeve should be 1 mm holding power.
from the base of the bevel of the phaco tip to create
adequate exposed needle length for sufficient holding Horizontal Chop
power. Excessive phaco energy application is to be A few bursts or pulses of phaco energy will allow
avoided, as this will cause nucleus immediately adja- the tip to be encased within the nucleus. It then can
cent to the tip to be emulsified. The gap thus created be drawn toward the incision to allow the chopper
in the vicinity of the tip is responsible for interfering access to the epi-endo nuclear junction. The chopping
with the seal around the tip and therefore the capabil- instrument is passed over the nucleus and under the
ity of vacuum to hold the nucleus. The nucleus will anterior capsule into this junction. It may be helpful
then pop off the phaco tip, making chopping more to rotate the chopper to horizontal as it passes below
difficult. With a good seal, the heminucleus can be the anterior capsule. If the nucleus comes off the phaco
drawn toward the incision and the chopper can be tip, excessive power has produced a space around the
inserted at the endonucleus-epinucleus junction. The tip, impeding vacuum holding power as noted above.
chopper is then drawn down and left, while the phaco Pulling the chopper down and left and pushing the
tip is pushed up and right. This will result in chopping phaco tip up and right will generate the first chop.
of the heminucleus. Minimal rotation of the nucleus will allow for creation
After the first chop, a second similar chop is of the second chop. The first pie-shaped piece of
performed so the heminucleus is divided into three nucleus is mobilized with high vacuum and elevated
pieces. One pie-shaped piece of nucleus thus created to the iris plane. There it is emulsified with low linear
is elevated to the iris plane (occlusion is utilized to hyper-pulse or nonlongitudinal power, high vacuum,
move fragments) and removed with low power hyper- and moderate flow.
pulsed phaco or nonlongitudinal phaco as discussed
74 Chapter 8
Alternatively, the vitrector can be inserted through the capability of the surgeon for appropriate response
a pars plana incision 3 mm posterior to the limbus. to this requirement. It is this crucial attitude that
Recently, 25-gauge vitrectomy instruments have been through relentless evaluation of the interaction of the
introduced. Their ultimate utility, however, is not yet machine, and the phaco procedure, the skillful surgeon
clear. In an effort to better visualize the vitreous for will find innovative methods to enhance technique.
thorough vitrectomy, unpreserved sterile prednisone
acetate (Kenalog), previously purchased from a formu-
lating pharmacy, can be injected into the vitreous. The Bibliography
Buratto L, Osher RH, Masket S, eds. Cataract Surgery in Complicated
prednisone particles adhere to the vitreous strands, Cases. Thorofare, NJ: SLACK Incorporated; 2001.
making the invisible visible. Fishkind WJ, ed. Complications in Phacoemulsification: Recognition,
Avoidance, and Management. New York, NY: Thieme Publish-
ers; 2001.
Summary Fishkind WJ. Pop Goes the Microbubbles. ESCRS Film Festival
Grand Prize Winner, 1998.
The phaco process is a balance of technology and Fishkind WJ, Neuhann TF, Steinert RF. The Phaco Machine in Cata-
technique. Awareness of the principles that influence ract Surgery Technique: Complications & Management. 2nd ed. Phila-
phaco machine settings is a prerequisite for the perfor- delphia, PA: WB Saunders Co; 2004.
mance of a proficient and safe operation. Additionally, Miyoshi T. From phaco-cutting to true phacoemulsification. Vid-
eo competition grand prize winner ASCRS, ESCRS 2007.
often during the procedure, there is a demand for Seibel BS. Phacodynamics: Mastering the Tools and Techniques of Phaco-
modification of the initial parameters. A thorough emulsification Surgery. 3rd ed. Thorofare, NJ: SLACK Incorpo-
understanding of fundamental principles will enhance rated; 2000.
Chapter
9
Setting Phaco Parameters
Mark Packer, MD, FACS; I. Howard Fine, MD; and Richard S. Hoffman, MD
is mobilized and emulsified. Alleviating the repulsive and flow restriction device in the aspiration line. The
force of longitudinal tip motion has been the impetus capacious filter element traps emulsate so that it will
behind the development of nonlongitudinal sonic and not clog the small diameter flow restrictor, which is
ultrasonic energy delivery systems, such as oscillatory, placed just up the aspiration line. The inner diameter
torsional, or transverse tip motions. The balancing of the flow restrictor is about the same as the aperture
of these competing forces at the phaco tip underlies on an aspiration tip used for removing cortex and vis-
much of the logic of setting parameters for efficient coelastic (0.2 mm). This device has the effect of greatly
surgery. reducing or eliminating surge because it limits the rate
at which fluid can move up the line. Fortunately, the
flow restriction does not impact flow at the usual rates
Vacuum and Venturi Pumps applied during phaco. In this light, it is interesting to
In a Venturi pump (named for the Italian physicist note that one of the situations where Venturi is most
Giovanni Battista Venturi), the foot pedal directly safe and efficient is during irrigation/aspiration.
controls the application of vacuum; aspiration flow
occurs in response to vacuum pressure. According
to the classic Venturi principle, it is the flow of pres- Power and
surized gas through a narrowed tube that creates the
vacuum. Unlike a peristaltic pump, with which vacuum
Power Modulations
The ability to variably control the application of
does not exist until there is resistance to flow, with a ultrasound power to within a period of several mil-
Venturi pump vacuum is always present. The surgeon liseconds has revolutionized phaco technology. The
sets the maximum vacuum level as one of the param- first generation of phaco machines only allowed appli-
eters. There is no setting for aspiration flow. The cation of continuous power at a fixed level. Following
vacuum increases in a linear fashion as the foot pedal the development of linear power control, the first
is depressed in foot position 2. Machines that feature power modulations were developed, pulse and burst
a bidirectional foot pedal also allow control of vacuum modes. In 2001, we showed how application of these
with yaw (ie, movement of the foot pedal in a direction modulations reduces the use of ultrasound energy and
parallel [rather than perpendicular] to the floor). This permits rapid visual rehabilitation after surgery.7 We
feature permits greater flexibility in separately control- also showed in that reduction of effective phaco time
ling the application of vacuum and ultrasound power. correlates with improved uncorrected visual acuity at
Conventional wisdom regards Venturi pumps as the first visit after surgery. Subsequently, the intro-
more aggressive than peristaltic pumps. This percep- duction of millisecond level control and variable duty
tion comes about primarily because of surge. Vacuum cycle applications has permitted further reduction of
increases as the surgeon depresses (or yaws) the foot ultrasound energy and eliminated the risk of thermal
pedal in order to evacuate material, and the vacuum injury to the cornea, paving the way for the adoption
remains high even after the material is evacuated of biaxial microincision cataract surgery.8 Surgeons
unless the surgeon actively reduces the vacuum by should try a variety of power settings, including pulse
moving the foot pedal. This process is in contradistinc- and burst modes, variable duty cycles, and percentage
tion to a peristaltic pump in which the vacuum will power ceilings, in order to develop parameters best
drop to zero once the occlusion has passed regardless suited to their individual techniques. Machines also
of foot pedal action. Of course, surge can still occur feature standard longitudinal, torsional, and transverse
with a peristaltic pump because of stored energy in tip motions that can be customized in amplitude to suit
the tubing and cassette (low compliance systems are a variety of techniques.
designed to reduce this problem). Nevertheless, with Intraoperative awareness and moment-to-moment
an appropriate initial Venturi vacuum setting and a assessment of surgical success offer the best opportu-
good foot pedal control, one can maintain a stable nity to alter settings and improve results. The surgeon
chamber. Therefore, not only do Venturi pumps have a should recognize that insufficient holding implies a
reputation for being more aggressive, they also have a need for greater vacuum, whereas an uncomfortable
reputation for allowing exceptionally rapid clearing of amount of surge calls for a reduction in vacuum. Poor
material, excellent followability, and fast surgery. followability may require increased aspiration flow or
One of the technological advances that has made vacuum if the problem is bringing material to the tip,
Venturi pumps safer involves the insertion of a filter or higher power if material comes to the tip but then
80 Chapter 9
bounces off when ultrasound is applied. A shallow NA. Role of corneal elasticity in damping of intraocular pres-
chamber indicates a need to check the irrigation bottle sure. Invest Ophthalmol Vis Sci. 2007;48(6):2540-2544.
4. Pallikaris IG, Kymionis GD, Ginis HS, Kounis GA, Tsilim-
height and the continuity and patency of the irrigation baris MK. Ocular rigidity in living human eyes. Invest Oph-
tubing from the bottle to the eye. Understanding the thalmol Vis Sci. 2005;46(2):409-414.
roles of flow, vacuum, and power will allow the sur- 5. Khng C, Packer M, Fine IH, Hoffman RS, Moreira FB. Intra-
geon to make machine adjustments that vastly improve ocular pressure during phacoemulsification. J Cataract Refract
Surg. 2006;32(2):301-308.
the surgical experience.
6. Zacharias J, Zacharias S. Volume-based characterization of
postocclusion surge. J Cataract Refract Surg. 2005;31(10):1976-
1982.
References 7. Fine IH, Packer M, Hoffman RS. The use of power modula-
1. Osher RH, Barros MG, Marques DM, Marques FF, Osher tions in phacoemulsification of cataracts: the choo choo chop
JM. Early uncorrected visual acuity as a measurement of the and flip phacoemulsification technique. J Cataract Refract Surg.
visual outcomes of contemporary cataract surgery. J Cataract 2001;27:188-197.
Refract Surg. 2004;30(9):1917-1920. 8. Packer M, Fine IH, Hoffman RS. Bimanual ultrasound phaco-
2. Fine IH, Packer M, Hoffman RS. Power modulations in new emulsification. In: Fine IH, Packer M, Hoffman RS, eds. Re-
phacoemulsification technology: improved outcomes. J Cata- fractive Lens Surgery. Heidelberg, Germany: Springer-Verlag;
ract Refract Surg. 2004;30(5):1014-1019. 2005:193-198.
3. Johnson CS, Mian SI, Moroi S, Epstein D, Izatt J, Afshari
Chapter
10
Foldable Intraocular Lens
Implantation
Richard S. Hoffman, MD; I. Howard Fine, MD; and Mark Packer, MD, FACS
81
82 Chapter 10
Step-by-Step Approach to
Folding and Implanting
Three-Piece Intraocular
Lenses
Step 1. Place IOL in Folder, Held in Nondominant
Hand. The lens is purchased with the inser-
tion forceps held in the dominant hand, then
placed into the folder or on the surface from
which it can be purchased by the folder in the
nondominant hand.
Step 2. Fold the IOL and Then Grasp the IOL With
Figure 10-3. Configuration of IOL folded across 10
the Insertion Device in the Dominant Hand.
and 4 o’clock axis demonstrating crossed swords
The lens is folded, and then the insertion de-
configuration.
vice, in the dominant hand, holds the folded
lens, which is now ready for insertion. When
lenses are folded across the 12 and 6 o’clock Step 3. Insert the IOL. In general, folded lenses should
axis, they are oriented in the holding or inser- be inserted through the incision with the fold
tion instrument with a leading and trailing hap- to the right unless they are folded across the 12
tic (Figure 10-2). This orientation is extremely and 6 o’clock axis, in which case they should
useful for implantation of an IOL into the cili- be inserted with the fold to the left. This spe-
ary sulcus in the presence of a compromised cial consideration ensures that lenses folded
capsular bag. In contrast, folding across the 10 with a leading and trailing haptic (6 and 12
and 4 o’clock axis (oblique axis) (Figure 10-3) o’clock fold) do not flip upside down because
or across the 9 and 3 o’clock axis (Figure 10-4) of the orientation of the leading haptic under
yields a folded configuration with both haptics the capsulorrhexis or under the iris for sulcus
pointed inferiorly with the fold superiorly (see implantation. The hand is then brought into
DVD). a proper position so that the fold is superior.
Foldable Intraocular Lens Implantation 83
Figure 10-4. Configuration of IOL folded across 9 acrylic IOLs have a cartridge with a 45-degree bevel-
and 3 o’clock axis with both haptics pointing infe- down configuration, which can implant foldable IOLs
riorly with the fold superiorly. into the capsular bag with ease. The tip of the Silver
Series insertion rod has a Teflon cap so that tearing of
the lens is avoided.
After the leading haptic has been delivered un-
der the distal capsulorrhexis, the forceps are
slowly opened (direct-acting forceps) or closed Step-by-Step Approach to
(reverse-acting forceps), allowing the lens to Implantation of Three-Piece
unfold. The trailing haptic is then usually di-
aled into the capsular bag to the left. Using the
Intraocular Lenses Using
folded orientation with both haptics directed the AMO Silver and Emerald
inferiorly negates the need for dialing in the Unfolder
trailing haptic because both haptics unfold into Step 1. Load IOL in Cartridge With Viscoelastic
the capsular bag, pulling the optic through the Material. Line the cartridge with viscoelastic,
capsulorrhexis. fold the IOL in the cartridge, and load the IOL
into the injector.
Step 2. Insert Injector Tip Into Incision. The injec-
Cartridge Injector Systems tor tip is inserted through the incision into the
In general, cartridge injector systems are now the
anterior chamber with the bevel down. The
standard for foldable IOL implantation. There are
bevel is then rotated slightly to the surgeon’s
many advantages of implanting foldable IOLs with
left so that the leading haptic is pointing to
injector systems as compared to folding forceps. These
the surgeon’s left as the optic is advanced with
advantages include the possibility of greater sterility,
the handpiece rod. The leading loop of the
ease of folding and insertion, and implantation through
IOL should always point to the surgeon’s left
smaller incisions. Every ophthalmic company that pro-
throughout the entire procedure.
duces IOLs has its own injector system. Each injector
Step 3. Insert IOL, Gradually Rotating the Tip Bevel
system has its own nuances for loading and implant-
Counterclockwise as the Leading Haptic and
ing the lens within the capsular bag, but for the most
IOL Enter the Capsular Bag. As the optic is
part the systems are more similar then dissimilar. The
advanced, the bevel needs to be rotated down
DVD and Appendix (see page 88) contain detailed
and then to the surgeon’s right to keep the lens
instructions for loading each of the current popular
in the proper orientation. The leading haptic is
IOL models.
placed in the bag as the IOL is released. Once
The AMO (Santa Ana, CA) Silver Series Unfolder
the optic is completely out of the cartridge, the
(Figure 10-5) for three-piece silicone IOLs and the
handpiece rod is retracted proximal to the end
Emerald Series Unfolder (Figure 10-6) for three-piece
of the trailing haptic, then advanced with the
84 Chapter 10
Figure 10-7. Following injection of the optic into the Figure 10-8. Dialing in of the trailing haptic is
capsular bag, the injector rod is withdrawn proxi- accomplished by placing a hook at the junction of
mal to the trailing haptic and then inserted into the the lens optic and the trailing haptic and pushing
eye carrying the trailing haptic through the anterior downward slightly with a 1 to 3 clock hour clock-
rhexis and into the capsular bag. wise rotation.
bevel down to place the trailing haptic within An intact capsular bag with compromised zonules
the bag (Figure 10-7). Placing the bevel com- can be easily addressed with a CTR and in-the-bag
pletely within the capsulorrhexis at this stage IOL implantation (Figures 10-9A through 10-9C).
of insertion keeps the optic in place and en- Utilizing a three-piece IOL with rigid haptics offers
sures placement of the trailing loop. the advantage of positioning the IOL haptics in the
Step 4. Place the Trailing Haptic in Capsular Bag. meridian of zonular weakness or dehiscence in order
Placement of the trailing loop of a foldable to support this region of the capsular bag and prevent
acrylic IOL usually requires implantation uti- further lens decentration during postoperative capsule
lizing a Lester hook to dial in the trailing hap- fibrosis. In instances of moderate zonular weakness,
tic. This is easily accomplished by removing implantation of a single-piece acrylic IOL offers the
the injector cartridge from the incision and advantage of inducing less zonular stress during
placing a hook at the junction of the trailing implantation of the IOL but little capsular support
haptic and the lens optic (Figure 10-8). With from the acrylic haptics. When severe or complete
slight downward pressure, the hook is pushed zonular dehiscence is present, it is best to implant a
distally and under the anterior capsulorrhex- three-piece IOL in the sulcus with concurrent iris or
is until the end of the trailing haptic is seen scleral fixation to ensure centration.
to pass under the rhexis. A 1 to 3 clock hour When the posterior capsule is intact, but a rent or
clockwise rotation of the hook will usually fa- tear in the anterior capsule may compromise IOL inser-
cilitate placement of the trailing haptic. tion, implantation of a single-piece acrylic IOL will
ensure that the anterior rhexis tear does not extend out
to the posterior capsule. These lenses typically unfold
Intraocular Lens after implantation in a very slow, controlled manner
Implantation in the that places little if any stress on the capsule. A three-
Presence of a Compromised piece IOL can also be implanted in these compromised
capsular bags; however, care should be taken to not
Capsular Bag stress the capsule in the location of the anterior tear.
When faced with a compromised capsular bag Dialing in of these IOLs should be accomplished with
following nucleus and cortex removal, the surgeon all forces directed in a location distant from the tear.
must decide the best lens model and location of final When the posterior capsule is torn, IOLs can be
implantation to optimize the postoperative result. placed within the capsular bag if the tear is small or
Foldable Intraocular Lens Implantation 85
Figure 10-9A. Large zonular dehiscence (arrow). Figure 10-9B. Implantation of CTR with vector
forces directed toward area of zonular dehiscence
to prevent dehiscence extension.
Figure 10-9C. Dialing in of three-piece IOL with Figure 10-10. Optic capture through the anterior
PMMA haptics with vector forces directed toward capsulorrhexis accomplished with downward pres-
the region of zonular dehiscence. sure on the edge of the IOL optic with a blunt Lester
hook.
converted to a continuous posterior capsulorrhexis. and smaller than 6 mm, the optic can be captured
However, under most circumstances it is safer to behind the rhexis after sulcus implantation by pressing
implant the IOL in the ciliary sulcus to prevent lens down on one edge of the optic until it prolapses behind
decentration or subluxation into the posterior cham- the rhexis and then pressing the other edge, 180
ber. Single-piece acrylic IOLs have been reported to degrees away from the first location, until the entire
cause pigment dispersion and glaucoma secondary to optic is posterior to the anterior rhexis (Figure 10-10).
iris chaffing from the sharp anterior edges of the hap- Implantation of an IOL into the sulcus in the presence
tics.1,2 For this reason, it is preferred to implant three- of a torn posterior capsule is a simple procedure and
piece IOLs with a rounded anterior optic edge in the can be accomplished with standard cartridge injector
ciliary sulcus. In addition, if the anterior rhexis is intact systems.
86 Chapter 10
Figure 10-11. Injection of ophthalmic viscoelastic Figure 10-12. Placement of the optic under the dis-
device in the subincisional region between the tal iris as it is unfolding will prevent the IOL from
anterior capsulorrhexis and the iris to facilitate sul- flipping upside-down.
cus placement of the IOL.
available. In addition, when capsular bags are compro- intraoperatively, in addition to a thorough knowledge
mised from posterior or anterior tears, or from zonular of the options for IOL implantation, will ensure the
dehiscences, the lens implant technique needs to be best possible surgical result in both routine and chal-
modified to ensure a good postoperative result with an lenging cases.
adequately centered IOL. Adjunctive capsular devices
such as the CTR have improved our ability to address
zonular weakness prior to IOL insertion within the References
1. Iwase T, Tanaka N. Elevated intraocular pressure in second-
capsular bag but even in the best of circumstances and ary piggyback intraocular lens implantation. J Cataract Refract
with the best surgical technique, IOLs may need to Surg. 2005;31:1821-1823.
be implanted within the ciliary sulcus. Understanding 2. Micheli T, Cheung LM, Sharma S, et al. Acute haptic-in-
the limitations and risks inherent in various clinical duced pigmentary glaucoma with an AcrySof intraocular
lens. J Cataract Refract Surg. 2002;28:1869-1872.
settings preoperatively and as they are developing
88 Chapter 10
Appendix
Lens Loading Steps
AMO Three-Piece Silicones AMO Three-Piece Acrylics
Clariflex, SA40, Z9002 ZA9003, AR40e, NXG1
Silver Series Unfolder “pscst” Cartridge and Emerald Series Unfolder Cartridge and Injector
Injector 1. Fill cartridge barrel and two channels with vis-
1. Fill cartridge barrel and two channels with vis- coelastic, adding a dollop on the central ridge.
coelastic, adding a dollop on the central ridge. 2. Using a smooth lens loading forceps, lift the
2. Using a smooth lens loading forceps, lift the lens from its packaging and transfer it to the
lens from its packaging and transfer it to the center of the two channels.
center of the two channels. 3. Initially holding the wings of the two channels
3. Initially holding the wings of the two channels spread wide, depress the sides of the optic un-
spread wide, depress the sides of the optic un- der the ledges of the two channels, then bring
der the ledges of the two channels, then bring the wings partially together to hold them
the wings partially together to hold them there there. There may be a need to depress the op-
(Figure 10-13). tic centrally with the lens loading forceps to
4. Place the leading haptic inside the barrel, trap- help it fold center downward.
ping it in an extended position. 4. Place the leading haptic inside the barrel, trap-
5. Check that the trailing haptic exits the rear of ping it in an extended position.
the cartridge. 5. Check that the trailing haptic exits the rear of
6. Insert the cartridge firmly into the injector, en- the cartridge.
suring that the trailing haptic rests to the side 6. Insert the cartridge firmly into the injector,
of the injector, out of the way of the plunger. ensuring that the trailing haptic rests to the
7. Advance the plunger, moving the lens forward outside of the injector, out of the way of the
toward the tip of the cartridge, stopping be- plunger (Figure 10-14).
fore the leading haptic exits the tip. The action 7. Depress the plunger, moving the lens forward
should be very smooth which is characteristic toward the tip of the cartridge. The action
of the silicone lenses. will be very stiff due to the lens material be-
ing acrylic. As always, stop before the leading
haptic exits the tip.
Figure 10-13. AMO Silver Series Cartridge. Depress Figure 10-14. AMO Emerald Series Cartridge.
the sides of the optic under the ledges of the two After inserting the cartridge firmly into the injector,
channels, then bring the wings partially together to ensure that the trailing haptic rests to the outside of
hold the optic edges in place. the injector, out of the way of the plunger.
Foldable Intraocular Lens Implantation 89
A B
Alcon Single-Piece Acrylics 5. Using the top edge of the entry port of the
cartridge, fold the leading haptic over its optic
SA60AT et al (Figure 10-15A) as the optic is slid forward into
Green Monarch C Cartridge and Injector the cartridge (Figure 10-15B).
1. Orient the C cartridge with lens outline up- 6. Releasing the lens from the forceps can be
ward, tail on bottom side. tricky at this point.
2. Fill the cartridge approximately two-thirds full 7. Grasp the trailing haptic and fold it over its op-
with viscoelastic. tic while sliding the entire lens fully into the
3. Place dots of viscoelastic on each haptic and cartridge (Figure 10-15C).
centrally on the optic. 8. Snap the cartridge firmly into the injector.
4. With a smooth forceps, lift the lens from its 9. Advance the lens (twist action) within the car-
packaging, avoiding the center of the optic. tridge until it is visible near the tip.
90 Chapter 10
Figure 11-1A. Cohesive OVDs behave like jelly or Figure 11-1B. Dispersive OVDs behave like honey
jam under conditions of zero shear (when there is at zero shear. The short-chained molecules of dis-
no fluid movement in the eye). The long-chained persive OVDs tend to slide over one another and
molecules of cohesive OVDs tend to intertwine and puddle. In general, dispersive OVDs are good at
“lock in position.” This creates a scaffolding effect coating intraocular structures but are not as effec-
intraocularly that helps these OVDs to maintain tive as cohesive materials at maintaining surgical
space very effectively. space.
“dispersive” because their short molecular chains do intraocularly that helps these OVDs to maintain space
not interlink or become entangled easily. This causes very effectively. This quality makes cohesive OVDs
the molecules of these OVDs to separate from one more retentive when there is no fluid movement in
another and behave in a dispersive manner. the eye, and ideal for surgical challenges that require
difficult intraocular maneuvers such as intraocular lens
(IOL) exchanges. Higher molecular weight cohesive
Defining Zero Shear vs OVDs such as Healon (AMO) and Healon GV are
High Shear Conditions more retentive than lower molecular weight cohesive
Cohesive and dispersive OVDs behave predict- OVDs such as Provisc (Figure 11-1A).
ably under different conditions of fluid movement
within the eye. Zero shear is a term used to describe a Dispersive Ophthalmic Viscoelastic Devices
condition when there is no fluid movement within the The short-chained molecules of dispersive OVDs
eye. Capsulorrhexis performed in an anterior chamber tend to slide over one another and create a puddle
filled with an OVD is an example of a zero shear con- under conditions of zero shear. The concentrations
dition. High shear describes a condition when there of dispersive low molecular weight OVDs are gener-
is a high rate of fluid movement within the eye. High ally higher (range ~3% to 4%) than that of cohesive
shear conditions exist during phacoemulsification and longer chained OVDs (range ~1% to 2.3%). The
during aspiration/irrigation. concentration is increased in an effort to increase the
zero shear viscosity of these shorter chained materials.
An increase in concentration increases to some degree
Ophthalmic Viscoelastic the retentiveness of short-chained OVDs and prevents
Device Behaviors at these materials from being excessively “runny.” In gen-
Zero Shear eral, however, dispersive OVDs are not as effective as
cohesive materials at maintaining surgical space under
conditions of zero shear (Figure 11-1B).
Cohesive Ophthalmic Viscoelastic Devices
During conditions of zero shear, the long-chained
molecules of cohesive OVDs tend to intertwine and
“lock in position.” This creates a scaffolding effect
Understanding the Clinical Behavior of Ophthalmic Viscoelastic Devices 93
Figure 11-2A. Long-chained molecules of cohesive Figure 11-2B. Short-chained molecules of dispersive
OVDs behave like spaghetti during conditions of OVDs under conditions of high shear behave like
high shear. They tend to entangle and leave the penne. The molecules do not tend to interlink and
eye as a bolus. This makes cohesive materials much stay in the anterior chamber much more effectively
easier to remove at the conclusion of a procedure. than do cohesive OVDs. For this reason, dispersive
OVDs of sodium hyaluronate provide superior
endothelial protection in conditions of high shear,
Ophthalmic Viscoelastic especially during phacoemulsification.
Device Behaviors at
High Shear The Advantages of Using
Cohesive Ophthalmic Viscoelastic Devices
Dispersive and Cohesive
The long-chained molecules of cohesive OVDs, Ophthalmic Viscoelastic
under conditions of high shear, tend to entangle and Devices Together
leave the eye as a bolus. This makes cohesive materials Many surgeons, in order to obtain the best quali-
much easier to remove at the conclusion of a proce- ties of both dispersive and cohesive OVDs, use the
dure, but less protective than dispersive OVDs during two types of OVDs in combination. A dispersive
phacoemulsification (Figure 11-2A). OVD, such as Healon D, may be used at the beginning
of a case to provide maximum endothelial protection
Dispersive Ophthalmic Viscoelastic Devices during phacoemulsification. A cohesive OVD, such as
Dispersive OVDs of sodium hyaluronate provide Healon, may then be used at the time of IOL implan-
superior endothelial protection during phacoemulsi- tation both to provide excellent chamber maintenance
fication. The short-chained molecules tend to slide and easy removal of the OVD, thereby reducing the
over themselves and coat intraocular structures in a chances of elevated intraocular pressure (IOP) post-
honey-like fashion. In addition, because the short- operatively.
chained molecules do not tend to interlink, dispersive
OVDs tend to stay in the anterior chamber much more
effectively during phacoemulsification than cohesive The Special Qualities of
OVDs. This dispersive quality, however, ceases to Healon 5
be an asset when it is time to remove the viscoelas- Healon 5 is an unusual OVD that cannot be
tic material at the end of a case. Complete removal described adequately as either a cohesive or a disper-
of sodium hyaluronate dispersive OVDs is difficult sive agent. Depending on conditions of flow, it has
because the molecules do not tend to join together and qualities of both. Healon 5 has both a high molecu-
do not aspirate as a unit (Figure 11-2B). lar weight (4,000,000 Daltons) and a relatively high
molecular concentration (2.3%).
At zero shear, the long-chained molecules of
Healon 5 lock together readily and act as a unit once
94 Chapter 11
they are injected into the anterior chamber. The link- During irrigation/aspiration, high shear conditions
age of the long-chained molecules is enhanced by may be created by increasing both flow and vacuum,
increased concentration of the material. As a result, and Healon 5 becomes a super cohesive material
Healon 5 is the most highly retentive of all OVDs at that leaves the anterior chamber readily in a bolus.
zero shear, allowing this material not only to maintain Special care must also be taken at the end of the case
anterior chamber volume extraordinarily well but to to remove all of the high molecular weight Healon 5
move tissue and enlarge poorly dilating pupils as well. from beneath the IOL in order to prevent an elevation
Under conditions of low shear, when phaco- of IOP. Because of its unique clinical features, Healon
emulsification parameters are set to low flow and low 5 is very useful in the management of some of the most
vacuum, the Healon 5 material can be fractured and difficult challenges in cataract surgery, including small
compartmentalized. An arching dome of Healon 5 pupils unresponsive to mydriatics, iris prolapse, floppy
can be left above the pupillary plane which provides iris syndrome, and the mature intumescent cataract.
both endothelial protection and pupillary dilatation, To understand the clinical applications of Healon 5
while nucleus removal is accomplished in the posterior more completely, see Chapter 7 for a detailed descrip-
chamber. tion of its uses and indications.
Chapter
12
Intraocular Lens
Materials and Design
Oliver Findl, MD, MBA
Figure 12-2.
Figure 12-3. Fusion of capsule at haptic-optic junction for different haptic designs. (Left) Acrysof multi-
piece with nearly complete fusion. (Middle) Acrysof single-piece with incomplete fusion which may serve
as one entry site (arrows) for regenerating LECs and no sharp edge at junction. (Right) The Tecnis 1-Piece
IOL incorporates a new feature of the ProTec 360-degree barrier edge.
Figure 12-5. Problems with accommodating IOLs. Infolding of haptics due to capsule constriction with
1CU (left); early PCO due to missing barrier along broad haptic-optic junctions for 1CU (middle) and
Crystalens AT-45 (right).
of IOL power. Should buttonholing not be possible, eye of normal dimensions. Accordingly, in a short eye,
about 0.5 diopters should be deducted from the cal- such a shift would cause more refractive change. These
culated power since the IOL will be more anteriorly IOLs have in common a hinge-like junction of haptics
placed in the eye. to optic that should allow the shifting of the optic when
In cases where no capsule support is given, apart the haptics are compressed. Measurements of IOL
from the classical angle-supported anterior chamber shift with current models have shown only very small
IOL, iris-supported IOLs and scleral-sutured IOLs are amounts of IOL movement and to be very variable
the most popular options (see Figure 12-2). In the case among eyes, both when stimulated with a near target
of the iris-supported IOL with lobster-claw haptics that or pilocarpine-induced ciliary muscle contraction.19-21
are “clipped” onto iris stroma, they can be clipped onto Apart from lacking evidence of their function, these
the iris from the anterior side or from the posterior IOL designs have had significant amounts of PCO with
side—so-called retropupillary fixation. This IOL style most patients needing Nd:YAG capsulotomies within
has a long track record in aphakic eyes and appears to the first 2 years after surgery (Figure 12-5).22
have a low rate of endothelial cell loss, but do require a
6-mm incision since the aphakic style is currently only
available in PMMA. Intraocular Lens
In the case of scleral suturing of a posterior cham- Optic Design
ber IOL, both foldable and rigid IOLs can be used.
However, there have been several reports of long-term Edge Design
knot erosion resulting in decentration or even sublux- During the past decade it has become clear that
ation of these IOLs as well as late endophthalmitis.18 optic edge design plays an important role in the pre-
The trend is away from sutured IOLs back to modern vention of PCO. When the Acrysof lens (Alcon) was
anterior chamber IOLs and iris-fixated IOLs. introduced in the early 1990s, several studies showed
that PCO development was significantly less than with
other IOLs.23-25 This first was attributed to the acrylic
Special Haptics— material and to the surface properties of the IOL.26
Accommodating Intraocular Later it could be shown that the sharp-edge design of
Lenses the lens seemed to be the key factor for this effect.27
Currently available accommodating IOLs are sup- The sharp IOL edge was a result of the manufactur-
posed to work according to the optic shift principle. ing process, and its blocking effect on LEC migra-
Ciliary muscle contraction should result in an anterior tion, therefore, rather coincidental. Further studies
shift of the optic, resulting in an overall increase in confirmed that the rectangular shape of the IOL rim
refractive power of the eye. A 0.7-mm shift would be with its sharp edges, in combination with the acrylic
predicted to achieve 1 diopter of accommodation in an material, was in fact the main reason for the reduced
Intraocular Lens Materials and Design 101
Figure 12-6. Blocking of LEC migration at posterior sharp optic edge due to bending of the capsule (left)
compared to round edge IOL (right).
formation of PCO.28 Studies by Nishi revealed that the of the front and back surface are identical. Some
discontinuous capsular bend seems to be a key factor manufacturers have an asymmetric biconvex optic,
for the preventative effect of a sharp-edge optic.27,29 where the back surface curvature is relatively flat and
The capsular bend at the posterior optic edge causes constant throughout most of the power range and
mechanical pressure and/or contact inhibition of LEC the anterior curvature is varied for IOL power. This
growth on the posterior capsule (Figure 12-6). causes a slight shift of the principal optical plane of
As a result of these findings, several new IOLs with the IOL and also implies that the lens should not be
a sharp optic edge design were introduced in the past implanted front to back in the eye, apart from the
years and compared in clinical trials. In a meta-analysis angulation of the haptics being backward as well. In
of the randomized controlled trials comparing round a symmetrically biconvex lens with no angulation,
and sharp-edge IOLs,30 there was a clear beneficial the IOL could be implanted front to back without a
effect of sharp-edge IOLs concerning inhibition of change in optical power.
PCO. This also confirmed that the sole modification
of the posterior optic edge from a round edge to a Optical Zone
sharp edge leads to a significant reduction of PCO by Most IOLs have a full-size effective optical zone
inducing a discontinuous bend at the posterior capsule of 6 mm in the main range of IOL powers. Therefore,
(Figure 12-7).31,32 the higher powered IOLs will have a thicker optic
Unfortunately, sharp optic edges of IOLs may also than the lower powers. This has the advantage of a
have disadvantages. As described previously, in some full optic zone, but can make folding of the IOL or
cases with implantation of lenses with a rectangular injecting with a shooter variable depending on IOL
edge shape combined with a high refractive index, power. Some IOLs keep a constant center thick-
such as found with the Acrysof lens, an increased ness of the optic and vary the effective optical zone,
incidence of persistent edge-glare phenomena was thereby varying the curvature of the optic and, there-
reported.33,34 Sharp-edge IOL designs cause the light fore, optic power. To my knowledge, there was only
rays that are refracted through the peripheral IOL to one manufacturer (Dr. Schmidt) that actually varied
be more intense on the peripheral retina. Round-edge refractive index of the silicone material used for dif-
IOL designs disperse the rays of light over a larger sur- ferent powers, thereby keeping a constant effective
face area of the retina, leading to less glare. However, optical zone and center thickness.
the half-rounded edge profile of some newly developed
IOLs with a round anterior and sharp posterior optic Special Optics
edge seems to avoid this disturbing side effect.35
Aspherical Intraocular Lenses
Optic Geometry This topic is covered extensively in Chapter 13.
In short, these IOLs are either neutral concerning
Biconvexity SA, therefore not adding SA to the eye, or like most
Most IOLs on the market have a symmetrically models currently on the market have a prolate sur-
biconvex optic, meaning that the radius of curvature face inducing negative SA, which should neutralize
102 Chapter 12
Figure 12-7. PCO 1 year (upper) and 3 years (lower) after surgery for round (left) and sharp (right) edge
optic design for a hydrophobic acrylic IOL.
the positive SA of the average cornea. The aim is to steep axis, adding an opposite clear cornea incision
increase contrast sensitivity under mesopic conditions (OCCI) on the same axis, or making limbal relaxing
where the pupil is dilated. The IOLs have little to no incisions (LRIs) on the steep axis. Most surgeons will
effect when the pupil is small. use a 600-micron knife to perform LRIs. LRIs are able
to reduce corneal astigmatism by as much as 3 diop-
Toric Intraocular Lenses ters. This topic is covered at length in Chapter 16. The
With cataract surgery we can attempt to reduce variability of the outcome is mainly due to interpatient
preexisiting corneal astigmatism using incisional tech- differences in scarring of the corneal tissue, corneal
niques, such as placing the corneal incision on the rigidity, and corneal thickness.
Intraocular Lens Materials and Design 103
life.38 However, good refractive outcome and low
residual astigmatism after surgery are key to success.
Therefore, meticulous biometry and power calculation
are needed. Additionally, since the light is divided and
also some light (about 20%) is lost to higher orders of
diffraction, patients have reduced contrast sensitivity.
Small amounts of PCO may cause substantial loss in
visual functions and Nd:YAG capsulotomy may need
to be performed earlier than usual. Additionally, the
blurred nonfocused image will overlay the focused
image and can cause the photic phenomenon of halos
seen around light sources especially at night with a
larger pupil. These can be disturbing to patients and
are the main reason for explantation of mIOLs.
There are two types of mIOLs: diffractive and
Figure 12-8. Toric IOL with marks for alignment. refractive. Diffractive mIOLs (Figure 12-9) use the
entire optical zone for the creation of two foci and are,
therefore, bifocal mIOLs. The focal points are created
An effective and quite predictable method of using constructive and destructive interference of light
neutralizing corneal astigmatism is the use of toric rays. These phase differences are induced by small
IOLs. The steep axis of the eye needs to be marked steps that are about one-half of the incident light wave-
in the sitting position before surgery since the eye length. In refractive mIOLs, several foci are created by
will undergo some cyclotorsion in the supine position. zones of different surface curvatures of the lens. These
The mean cyclotorsion was reported to be 2 degrees, IOL models will differ according to the distribution of
however, can vary between patients and be up to 10 the zones on the optic surface, and the light distribu-
degrees in individual cases.36 Accurate axis placement tion onto the different foci is pupil size dependent.
of the toric IOL is critical to the outcome since 3% of In general, diffractive mIOLs usually have very
the toric correction is lost for every degree off axis. good near vision outcomes, however, intermediate
Toric IOLs have marks on the IOL optic for alignment vision is poor. In contrast, refractive mIOLs usually
(Figure 12-8). Being 10 degrees off the desired axis have good intermediate vision but relatively poor near
results in about one-third of the toric correction lost. vision. In an attempt to get the best of both worlds, a
Being 30 degrees off results in no toric correction and strategy called “mix-and-match” with implantation of a
a shift of the axis, and errors beyond that result in an refractive mIOL into one eye and a diffractive mIOL
increase in astigmatism of the eye, being more than
into the contralateral eye has been developed. To date
preoperatively and at a completely different axis (axis-
there are little published data available, but this strat-
flip). Since it is crucial that the IOL does not rotate
egy appears promising in some patients.
inside the capsule bag during capsule shrinkage, there
Another strategy to avoid mIOLs and their poten-
are several different special haptic designs that should
ensure stability. Clinical outcomes with modern toric tial drawbacks as mentioned above is monovision
IOLs have been very promising and rotational stability where both eyes receive standard monofocal IOLs.
appears to be within 2 degrees.37 Good planning and The dominant eye receives an IOL power to achieve
precision during surgery seem to be key to the success good distance vision and the contralateral eye is made
with these IOLs. about 1.25 diopters more myopic to allow intermedi-
ate vision. With both eyes open, the patients usually
Multifocal Intraocular Lenses have satisfactory near vision, at least under good light-
Multifocal IOLs (mIOL) are designed to overcome ing conditions.
the postoperative lack of accommodation by dividing Whether using mIOLs or monovision, patient
the incoming light onto two or more focal points. selection and extensive preoperative counseling are
One of these is used for distance vision, the other for key factors for a good outcome. It appears that patient
near or intermediate vision. These IOLs have shown motivation to achieve spectacle independence may be
to reduce the need for spectacle correction in daily the critical deciding factor for success.38
104 Chapter 12
Clinical Performance of an
Intraocular Lens
Biocompatibility
Phacoemulsification and foldable IOL technology
have permitted the use of small incisions, which results
in less trauma caused by cataract surgery. Immediate
postoperative inflammation is mainly attributed to
surgical irritation of the anterior uvea, which causes
changes in the blood-aqueous barrier.39 Long-term
postoperative inflammation is caused by other factors
such as immunological reactions.
The performance of an IOL is determined by sev-
eral factors such as the surgical technique,40 the peri-
operative treatment,41 the IOL biomaterial and design,
and the host reaction to the lens.
The cellular reaction seen on an IOL is an impor-
tant indicator of the IOL’s biocompatibility. On the Figure 12-9. Diffractive mIOL with PCO.
one hand, it consists of macrophages in the form of
small, round cells and foreign body giant cell on the
IOL surface. On the other hand, the cells are LECs entiation into myofibroblasts and the synthesis of col-
after the capsule comes into contact with the foreign lagen fibers.43 These cytokines may act in an autocrine
body IOL. Accordingly, the biocompatibility of an and paracrine fashion, influencing the postoperative
IOL can be divided into two parts—the uveal and the proliferation of LECs in the capsular bag. Thus, Nishi
capsular reaction, as described by Amon.42 and coauthors postulated that fibrous proliferation of
Uveal biocompatibility is defined as the reaction LECs with anterior capsule fibrosis is often associated
of the uvea to the IOL. As a result to the surgical trau- with blood-aqueous barrier disruption, clinically vis-
ma and the IOL, monocytes and macrophages migrate ible as flare in the anterior chamber.43
through the uvea’s vessel walls into the aqueous and
then onto the IOL surface. Monocytes transform
into small, round cells and macrophages transform Posterior Capsule
into epithelioid and foreign body giant cells that are Opacification
responsible for the phagocytosis of debris. These cells PCO (or after cataract) remains a common prob-
constitute the natural immunological process in a for- lem after cataract surgery with implantation of an IOL.
eign body reaction. It resulted from the transition from intracapsular cata-
Capsular biocompatibility is defined as the reac- ract extraction (ICCE) to ECCE, where the posterior
tion of LECs and the capsule to the IOL material and lens capsule is left intact during surgery. Patients with
design. This encompasses LEC ongrowth, anterior PCO suffer from decreased visual acuity, impaired
capsule opacification, and PCO. The LECs residing contrast sensitivity, and glare disability.
on the posterior side of the anterior capsule (Figure Clinically, two different components of PCO can
12-10), the so-called A-cells, can proliferate onto the be differentiated, namely a regeneratory and a fibrotic
IOL optic from the anterior capsular rim (ongrowth) component (Figure 12-12). Regeneratory PCO is much
and lay down collagen which results in whitening of more common; it is caused by residual LECs from the
the capsule as well as contraction of the capsule, which lens equator region, the so-called E-cells, migrating
in turn may cause rhexis contraction or even phimosis, and proliferating into the space between the posterior
decentration of the IOL, or buttonholing of the IOL capsule and the IOL, forming layers of lens material
(Figure 12-11). and Elschnig pearls. Fibrotic PCO is caused by LECs
LECs also express cytokines, such as interleukin-1, from the anterior capsule that undergo transforma-
interleukin-6, and transforming growth factor , that tion to myofibroblasts and gain access to the poste-
are responsible for LEC proliferation and transdiffer- rior capsule, causing whitening and wrinkling of the
Intraocular Lens Materials and Design 105
Figure 12-11. Complications of extensive fibrotic reaction of capsule: rhexis contraction (left), IOL decen-
tration (middle), partial buttonholing with IOL tilt (right). Arrows indicate location where the rhexis has
“slipped” behind the optic.
How to Achieve a Low Posterior Capsule As a result, round-edge IOLs have practically disap-
peared from the market. However, although drastically
Opacification Rate
reduced, the problem of PCO has not been eliminated.
Meticulous surgical technique is a prerequisite for
The role of IOL optic material remains unclear; while
low PCO rates. A well-centered capsulorrhexis where
hydrogel lenses have been shown to have a high PCO
the rhexis edge overlaps the IOL optic edge around the
incidence, there is still an ongoing debate about which
entire circumference is necessary to ensure a bending
of the hydrophobic material—hydrophobic acrylic or
effect on the posterior capsule to act as a barrier to
silicone—should be preferred with respect to PCO
invading LECs. Concerning IOL design, the concept
inhibition. Single-piece IOLs with an incomplete sharp
of a sharp posterior optic edge has been proven to be
optic rim have not shown significantly higher PCO
the most effective method to reduce PCO up to now.
Intraocular Lens Materials and Design 107
Figure 12-13. Examples of the dynamic changes of Elschnig pearls within a month in eyes with PCO; birth
and death (upper), questionable fusion of two pearls and then disappearance (lower).
rates than multipiece IOL designs. However, new Refract Surg. 2008;34(4):677-686.
ultrathin IOLs that are currently being developed for 8. Frohn A, Dick HB, Augustin AJ, Grus FH. Late opacification
of the foldable hydrophilic acrylic lens SC60B-OUV. Oph-
microincision surgery might perform worse concerning thalmology. 2001;108(11):1999-2004.
PCO inhibition, due to their thin optic rim and there- 9. Habib NE, Freegard TJ, Gock G, et al. Late surface opacifica-
fore possibly weaker barrier effect at the optic edge. tion of Hydroview intraocular lenses. Eye. 2002;16(1):69-74.
10. Schmidbauer JM, Werner L, Apple DJ, et al. Postoperative
opacification of posterior chamber intraocular lenses—a re-
References view. Klin Monatsbl Augenheilkd. 2001;218(9):586-594.
11. Findl O, Menapace R, Sacu S, et al. Effect of optic material
1. Findl O, Buehl W, Menapace R, et al. Long-term effect of
sharp optic edges of a polymethyl methacrylate intraocular on posterior capsule opacification in intraocular lenses with
lens on posterior capsule opacification: a randomized trial. sharp-edge optics: randomized clinical trial. Ophthalmology.
Ophthalmology. 2005;112(11):2004-2008. 2005;112(1):67-72.
2. Shah A, Spalton DJ, Gilbert C, et al. Effect of intraocular lens 12. Mainster MA. Violet and blue light blocking intraocular lens-
edge profile on posterior capsule opacification after extra- es: photoprotection versus photoreception. Br J Ophthalmol.
capsular cataract surgery in a developing country. J Cataract 2006;90(6):784-792.
Refract Surg. 2007;33(7):1259-1266. 13. Izak AM, Werner L, Apple DJ, et al. Loop memory of haptic
3. Alio JL, Chipont E, BenEzra D, Fakhry MA. Comparative materials in posterior chamber intraocular lenses. J Cataract
performance of intraocular lenses in eyes with cataract and Refract Surg. 2002;28(7):1229-1235.
uveitis. J Cataract Refract Surg. 2002;28(12):2096-2108. 14. Leydolt C, Davidovic S, Sacu S, et al. Long-term effect of
4. Leaming DV. Practice styles and preferences of ASCRS mem- 1-piece and 3-piece hydrophobic acrylic intraocular lens on
bers—2003 survey. J Cataract Refract Surg. 2004;30(4):892- posterior capsule opacification: a randomized trial. Ophthal-
900. mology. 2007;114(9):1663-1669.
5. Farbowitz MA, Zabriskie NA, Crandall AS, et al. Visual com- 15. Bender LE, Nimsgern C, Jose R, et al. Effect of 1-piece and 3-
plaints associated with the AcrySof acrylic intraocular lens. J piece AcrySof intraocular lenses on the development of pos-
Cataract Refract Surg. 2000;26(9):1339-1345. terior capsule opacification after cataract surgery. J Cataract
6. Findl O, Leydolt C. Meta-analysis of accommodating intra- Refract Surg. 2004;30(4):786-789.
ocular lenses. J Cataract Refract Surg. 2007;33(3):522-527. 16. Findl O, Drexler W, Menapace R, et al. Accurate determina-
7. Nanavaty MA, Spalton DJ, Boyce J, et al. Edge profile of com- tion of effective lens position and lens-capsule distance with
mercially available square-edged intraocular lenses. J Cataract 4 intraocular lenses. J Cataract Refract Surg. 1998;24(8):1094-
1098.
108 Chapter 12
17. Vass C, Menapace R, Schmetterer K, et al. Prediction of optic edge design in an acrylic intraocular lens on posterior
pseudophakic capsular bag diameter based on biometric vari- capsule opacification. J Cataract Refract Surg. 2005;31(5):954-
ables. J Cataract Refract Surg. 1999;25(10):1376-1381. 961.
18. Guell JL, Barrera A, Manero F. A review of suturing tech- 33. Holladay JT, Lang A, Portney V. Analysis of edge glare phe-
niques for posterior chamber lenses. Curr Opin Ophthalmol. nomena in intraocular lens edge designs. J Cataract Refract
2004;15(1):44-50. Surg. 1999;25(6):748-752.
19. Findl O, Kriechbaum K, Koeppl C, et al. Laserinterferomet- 34. Erie JC, Bandhauer MH, McLaren JW. Analysis of postopera-
ric measurement of the movement of an ‘accommodative’ in- tive glare and intraocular lens design. J Cataract Refract Surg.
traocular lens. In: Guthoff R, Ludwig K, eds. Current Aspects of 2001;27(4):614-621.
Human Accommodation II. Heidelberg, Germany: Dr. Reinhard 35. Buehl W, Findl O, Menapace R, et al. Effect of an acrylic in-
Kaden Verlag; 2003. traocular lens with a sharp posterior optic edge on posterior
20. Koeppl C, Findl O, Menapace R, et al. Pilocarpine-induced capsule opacification. J Cataract Refract Surg. 2002;28(7):1105-
shift of an accommodating intraocular lens: AT-45 Crystal- 1111.
ens. J Cataract Refract Surg. 2005;31(7):1290-1297. 36. Chernyak DA. Cyclotorsional eye motion occurring between
21. Hancox J, Spalton D, Heatley C, et al. Objective measure- wavefront measurement and refractive surgery. J Cataract Re-
ment of intraocular lens movement and dioptric change with fract Surg. 2004;30(3):633-638.
a focus shift accommodating intraocular lens. J Cataract Re- 37. Weinand F, Jung A, Stein A, et al. Rotational stability of a
fract Surg. 2006;32(7):1098-1103. single-piece hydrophobic acrylic intraocular lens: new meth-
22. Findl O. Intraocular lenses for restoring accommodation: od for high-precision rotation control. J Cataract Refract Surg.
hope and reality. J Refract Surg. 2005;21(4):321-323. 2007;33(5):800-803.
23. Hollick EJ, Spalton DJ, Ursell PG, et al. The effect of poly- 38. Leyland M, Zinicola E. Multifocal versus monofocal intra-
methylmethacrylate, silicone, and polyacrylic intraocular ocular lenses after cataract extraction. Cochrane Database Syst
lenses on posterior capsular opacification 3 years after cata- Rev. 2003(3):CD003169.
ract surgery. Ophthalmology. 1999;106(1):49-54. 39. Miyake K, Ota I, Miyake S, Maekubo K. Correlation between
24. Ursell PG, Spalton DJ, Pande MV, et al. Relationship between intraocular lens hydrophilicity and anterior capsule opacifi-
intraocular lens biomaterials and posterior capsule opacifica- cation and aqueous flare. J Cataract Refract Surg. 1996;22(Suppl
tion. J Cataract Refract Surg. 1998;24(3):352-360. 1):764-769.
25. Sundelin K, Friberg-Riad Y, Ostberg A, Sjostrand J. Posterior 40. Pande MV, Spalton DJ, Kerr-Muir MG, Marshall J. Postoper-
capsule opacification with AcrySof and poly(methyl methac- ative inflammatory response to phacoemulsification and ex-
rylate) intraocular lenses. Comparative study with a 3-year tracapsular cataract surgery: aqueous flare and cells. J Cataract
follow-up. J Cataract Refract Surg. 2001;27(10):1586-1590. Refract Surg. 1996;22(Suppl 1):770-774.
26. Hollick EJ, Spalton DJ, Ursell PG, Pande MV. Lens epithelial 41. Nishi O, Nishi K, Morita T, et al. Effect of intraocular sus-
cell regression on the posterior capsule with different intra- tained release of indomethacin on postoperative inflamma-
ocular lens materials. Br J Ophthalmol. 1998;82(10):1182-1188. tion and posterior capsule opacification. J Cataract Refract
27. Nishi O. Posterior capsule opacification. Part 1: experimental Surg. 1996;22(Suppl 1):806-810.
investigations. J Cataract Refract Surg. 1999;25(1):106-117. 42. Amon M. Biocompatibility of intraocular lenses. J Cataract
28. Nishi O, Nishi K, Sakanishi K. Inhibition of migrating lens Refract Surg. 2001;27(2):178-179.
epithelial cells at the capsular bend created by the rectangular 43. Nishi O, Nishi K, Imanishi M. Synthesis of interleukin-1 and
optic edge of a posterior chamber intraocular lens. Ophthalmic prostaglandin E2 by lens epithelial cells of human cataracts.
Surg Lasers. 1998;29(7):587-594. Br J Ophthalmol. 1992;76(6):338-341.
29. Nishi O, Nishi K, Mano C, et al. The inhibition of lens epi- 44. Heatley CJ, Spalton DJ, Kumar A, et al. Comparison of pos-
thelial cell migration by a discontinuous capsular bend cre- terior capsule opacification rates between hydrophilic and
ated by a band-shaped circular loop or a capsule- bending hydrophobic single-piece acrylic intraocular lenses. J Cataract
ring. Ophthalmic Surg Lasers. 1998;29(2):119-125. Refract Surg. 2005;31(4):718-724.
30. Findl O, Buehl W, Bauer P, Sycha T. Interventions for pre- 45. Hayashi K, Hayashi H. Posterior capsule opacification after
venting posterior capsule opacification. Cochrane Database Sys- implantation of a hydrogel intraocular lens. Br J Ophthalmol.
tem Rev. 2007(3). 2004;88(2):182-185.
31. Buehl W, Menapace R, Findl O, et al. Long-term effect of 46. Abela-Formanek C, Amon M, Schild G, et al. Uveal and cap-
optic edge design in a silicone intraocular lens on posterior sular biocompatibility of hydrophilic acrylic, hydrophobic
capsule opacification. Am J Ophthalmol. 2007;143(6):913-919. acrylic, and silicone intraocular lenses. J Cataract Refract Surg.
32. Buehl W, Findl O, Menapace R, et al. Long-term effect of 2002;28(1):50-61.
Chapter
13
Aspheric Intraocular Lenses
Y. Ralph Chu, MD
With continued improvements in surgical tech- sion of the vision of frogs and eagles. Glasser and
niques, biometry, and intraocular lens (IOL) technol- Campbell had shown that spherical aberration (SA)
ogy, cataract surgeons have for some time been capable of the crystalline lens changes considerably with age,
of consistently achieving highly accurate quantitative moving from a negative SA value to a positive one.1
refractive results following cataract/lens replacement Jack Holladay further demonstrated that side effects
surgery. We know we can improve an individual’s of myopic LASIK were likely due to the fact that the
vision from 20/400 to 20/20, for example. procedure turned a prolate human eye into an oblate
The modern cataract surgeon, however, is now one, with a sphericity or Q-value more akin to that of
embarking on the quest for “perfect vision” beyond a a frog than of a predator eagle.2 The role of SA in the
simple 20/20 standard. This does not necessarily mean aging eye suddenly became much more interesting.
getting the patient to 20/10. Rather, it means that we The average sphericity of the normal human
have started paying attention to other aspects of vision cornea is positive and remains stable throughout life,
beyond Snellen acuity, such as contrast sensitivity and but the lens SA changes with age. In the young eye,
wavefront error, in order to achieve the highest pos- the negative SA of the crystalline lens balances the
sible quality of vision. Cataract surgeons are becoming positive SA of the cornea, resulting in zero or very
refractive surgeons, and IOL manufacturers are begin- low total ocular SA.3 Light is sharply focused on the
ning to incorporate advanced refractive technology retina, producing a quality image and good functional
toward the same objective. vision (Figure 13-1). But in older eyes, the crystalline
Aspheric IOLs are the first new technology IOLs lens loses the ability to compensate for corneal SA,
to reflect the refractive shift in cataract surgery. total ocular SA becomes increasingly positive, and the
resulting aberrations cause blurred vision and reduced
contrast sensitivity, affecting functional vision (Figure
The Importance 13-2).
of Asphericity We also know that with age, contrast sensitivity
A decade ago, Jack Holladay introduced us to decreases, first at the higher spatial frequencies, then
the importance of asphericity in his famous discus- at all the spatial frequencies4 (Figure 13-3).
109
110 Chapter 13
Figure 13-1. The young eye has essentially zero Figure 13-2. The aging eye has positive spherical
spherical aberration at age 19. aberration reducing functional visual acuity.
older implanted with the Tecnis aspheric IOL, total improvement in functional vision may improve patient
ocular SA was not significantly different from zero, so safety for other situations in which visibility is low.
the lens is effective in reaching the intended target. Since then, it has been shown that this lens provides
A prospective randomized study showed a nearly uncorrected and distance-corrected near visual acuity
78% gain in peak contrast sensitivity with the new similar to that obtained with standard spherical mono-
lens, with mesopic contrast sensitivity approximately focal lenses, so there does not appear to be any loss of
equivalent to photopic contrast sensitivity with a depth of focus from correction of the positive SA.15
spherical lens11 (Figure 13-4). Early European studies
also showed that it could improve visual quality.12,13
In controlled, multicenter, US clinical trials (n=78), An Evolving Market
SA was significantly less 3 months post-implantation Since 2004, other lens manufacturers have intro-
of the Tecnis lens than after implantation of a spheri- duced other concepts of asphericity, with new aspheric
cal acrylic IOL. The benefit was independent of age14 lenses of their own.
(Figure 13-5). The Acrysof IQ IOL (SN60WF, Alcon, Fort
Driving simulations were also conducted as part of Worth, TX) was designed to partially compensate for
the US Food and Drug Administration (FDA) clinical the SA of a model eye. The lens has an aspheric poste-
trials to determine the impact of the lens on func- rior optic design with a thinner center. It induces -0.15
tional vision. Patients viewing a simulated nighttime microns of SA, compared to the -0.27 microns induced
rural road through a Tecnis aspheric lens identified a by the Tecnis lens, leaving approximately 0.1 microns
pedestrian in the road significantly faster than patients of positive SA in the average cornea.
viewing through a spherical lens.14 On average, Some studies have shown that Navy aviators with
patients with Tecnis lenses saw the pedestrian 0.50 excellent visual abilities have small amounts of SA, so
seconds sooner than the spherical IOL patients, which in theory, leaving a small amount of residual SA might
gave them a 45-foot advantage to react to the hazard be a good thing. However, Steve Schallhorn, who
in the road. Many recent vehicular safety improve- conducted the pilot studies, continues to believe that
ments that are now standard on automobiles improve striving for zero SA remains the most effective target.
braking time by just 0.11 to 0.35 seconds. In his aviator studies, those subjects with SA closer to
The FDA approved the Tecnis lens in 2004, with zero had better mesopic contrast acuity than their fel-
the unprecedented claim that it was likely to offer a low pilots with higher SA.16
meaningful safety benefit for elderly drivers and oth- Other human studies have also shown that superior
ers with whom they share the road. Moreover, the youthful vision is associated with zero SA. Pablo Artal
112 Chapter 13
Conclusion
Aspheric IOLs are here to stay and are rap-
idly becoming the standard of care because they can
potentially provide superior optical quality, especially
in low light and low contrast situations.
I believe that aspheric IOLs represent the first
truly refractive IOLs. They offer an easy way for
the general cataract surgeon to begin making the
transition to refractive cataract surgeon. Once one
has implemented the steps necessary for implanting
aspheric lenses (eg, precision biometry, correction of
preoperative astigmatism at the time of surgery), one
can more easily consider other premium IOLs, includ-
Figure 13-7. Photopic contrast sensitivity results from ing those with multifocal or accommodating surfaces.
our clinical comparison of three different aspheric As we develop better ways of measuring preop-
lenses. erative corneal SA, we may find ourselves custom-
izing the IOL to not only the axial length, but also to
the patient’s individual corneal SA, in an attempt to
highest quality vision—and may even improve Snellen optimize vision. And, farther in the future, we may be
visual acuity, as our anecdotal experience seems to sug- customizing IOLs to a whole range of quality of vision
gest. The one exception to this rule is the patient who factors as the quest for “perfect vision” evolves.
has had previous hyperopic laser refractive surgery.
If the correction was for significant hyperopia (+2.0
diopters or greater), the cornea will already have low or References
negative SA, and an aspheric lens implant can actually 1. Glasser A, Campbell MC. Presbyopia and the optical
changes in the human crystalline lens with age. Vision Res.
increase the total negative SA of the ocular system.
1998;38(2):209-229.
Some surgeons may prefer to measure corneal SA 2. Holladay JT, Dudeja DR, Chang J. Functional vision and cor-
preoperatively and base their lens decision on which neal changes after laser in situ keratomileusis determined by
of the three aspheric lenses is the most likely to bring contrast sensitivity, glare testing, and corneal topography. J
the patient’s total SA back into balance at zero. For the Cataract Refract Surg. 1999;25(5):663-669.
3. Guirao A, Redondo M, Artal P. Optical aberrations of the
majority of patients, the Tecnis IOL is the most likely human cornea as a function of age. J Opt Soc Am A Opt Image
to achieve the zero SA target. Sci Vis. 2000;17(10):1697-1702.
Of course, in addition to choosing an appropri- 4. Owsley C, Sekuler R, Siemsen D. Contrast sensitivity
ate aspheric IOL, surgeons should also take care to throughout adulthood. Vision Res. 1983;23:689-699.
5. Holladay JT. Spherical aberration: the next frontier. Cataract
maximize visual function with these lenses by fully
Refract Surg Today. 2006;Nov-Dec:95-106.
correcting lower order aberrations. This requires the 6. Nishi T, Nawa Y, Ueda T, et al. Effect of total higher-order
use of optimized IOL constants for biometry and cor- aberrations on accommodation in pseudophakic eyes. J Cata-
rection of astigmatism with limbal relaxing incisions or ract Refract Surg. 2006;32:1643-1649.
by other means. 7. Wang L, Koch DD. Custom optimization of intraocular lens
asphericity. J Cataract Refract Surg. 2007;33:1713-1720.
A good surgical technique with appropriate capsu- 8. Holladay JT, Piers PA, Koranyi G, et al. A new intraocular
lorrhexis and in-the-bag positioning is important. One- lens design to reduce spherical aberration of pseudophakic
piece and three-piece aspheric lenses are available. To eyes. J Refract Surg. 2002;18(6):683-691.
maintain adequate centration during a complicated 9. Wang L, Dai E, Koch DD, Nathoo A. Optical aberra-
tions of the human anterior cornea. J Cataract Refract Surg.
case, such as one in which there is a break in the cap-
2003;29(8):1514-1521.
sular bag that requires sulcus placement, a three-piece 10. Guirao A, Tejedor J, Artal P. Corneal aberrations before and
lens is necessary. after small-incision cataract surgery. Invest Ophthalmol Vis Sci.
Because these lenses have aspheric surfaces on 2004;45(12):4312-4319.
only one side, I have often been asked what would hap- 11. Packer M, Fine IH, Hoffman RS, Piers PA. Prospective ran-
domized trial of an anterior surface modified prolate intra-
pen if the lens is implanted upside down. The answer is ocular lens. J Refract Surg. 2002;18:692-696.
that the patient will still benefit from the asphericity of 12. Mester U, Dillinger P, Anterist N. Impact of a modified optic
the lens, although the refraction may be off. design on visual function: clinical comparative study. J Cata-
114 Chapter 13
ract Refract Surg. 2003;29(4):652-660. aspherical intraocular lens: prospective randomized masked
13. Bellucci R, Morselli S, Piers P. Comparison of wavefront ab- clinical trial. J Cataract Refract Surg. 2007;33(4):603-610.
errations and optical quality of eyes implanted with five dif- 19. Altmann GE, Nichamin LD, Lane SS, Pepose JS. Optical
ferent intraocular lenses. J Refract Surg. 2004;20(4):297-306. performance of 3 intraocular lens designs in the presence of
14. Package insert. TECNIS™ foldable posterior chamber intra- decentration. J Cataract Refract Surg. 2005;31(3):574-585.
ocular lens. Advanced Medical Optics, Inc. 20. Piers PA, Weeber HA, Artal P, Norrby S. Theoretical com-
15. Packer M, Fine IH, Hoffman RS. Visual acuity at distance parison of aberration-correcting customized and aspheric
and near with monocular and binocular monofocal aspheric intraocular lenses. J Refract Surg. 2007;23(4):374-384.
Tecnis IOL. Paper presented at: American Society of Cata- 21. Akkin C, Ozler SA, Mentes J. Tilt and decentration of bag-
ract and Refractive Surgery; San Diego, CA; 2007. fixated intraocular lenses: a comparative study between
16. Schallhorn SC, Tanzer DJ. Ideal spherical aberration to opti- capsulorrhexis and envelope techniques. Doc Ophthalmol.
mize visual outcome. Paper presented at: Annual Symposium 1994;87(3):199-209.
of the American Society of Cataract and Refractive Surgery; 22. Mutlu FM, Bilge AH, Altinsoy HI, Yumusak E. The role of
San Diego, CA; April 2007. capsulotomy and intraocular lens type on tilt and decentra-
17. Awwad ST, Lehmann JD, McCulley JP, Bowman RW. A com- tion of polymethylmethacrylate and foldable acrylic lenses.
parison of higher order aberrations in eyes implanted with Ophthalmologica. 1998;212(6):359-363.
AcrySof IQ SN60WF and AcrySof SN60AT intraocular 23. Hayashi K, Harada M, Hayashi H, et al. Decentration and
lenses. Eur J Ophthalmol. 2007;17(3):320-326. tilt of polymethyl methacrylate, silicone, and acrylic soft in-
18. Pandita D, Raj SM, Vasavada VA, et al. Contrast sensitivity traocular lenses. Ophthalmology. 1997;104(5):793-798.
and glare disability after implantation of AcrySof IQ Natural
Chapter
14
Capsular Tension Rings
115
116 Chapter 14
Indications
Although zonular laxity can be encountered in
any patient, there are common conditions that should
be recognized preoperatively as having a higher risk
Figure 14-1. Standard CTR. A single-piece PMMA for bag instability. Pseudoexfoliation is by far the most
semicircular device with blunt tip eyelets. (Photo common of these conditions. Other conditions include
courtesy of Morcher GmbH, Stuttgart, Germany.) uveitis, Marfan syndrome, homocystinuria, hyperma-
ture cataracts, microspherophakia, iatrogenic or trau-
Presently in the United States there are two Food matic zonular laxity, retinitis pigmentosa, myotonic
and Drug Administration (FDA) approved CTRs. One dystrophy, and eyes that have previously undergone
is made by Morcher GmbH (Stuttgart, Germany) and vitrectomy or filtering surgery.
the other by Ophtec (Groningen, The Netherlands). Careful attention during the preoperative exami-
The Morcher ring (marketed in the United States as nation can often identify mild iridodonesis or phaco-
the Reform Ring and distributed exclusively by FCI donesis. With the judicious use of a CTR, such cases
Ophthalmics of Marshfield Hills, Mass) is available in often proceed without complication and result in
three different sizes. The Type 14 CTR has an uncom- a well-centered and stable capsular bag and IOL.
pressed diameter of 12.3 mm and can be compressed Moderate or severe irido or phacodonesis (or frank
to a diameter of 10 mm. The Type 14C has an uncom- lens subluxation) are signs of significant zonule com-
pressed diameter of 13 mm and can be compressed to promise and alternative methods of cataract removal
11 mm. The Type 14A has an uncompressed diameter and lens implantation should be considered. In cases
of 14.5 mm with a compressible diameter of 12 mm. of severe bag/lens instability, the capsular tension seg-
The Ophtec CTR is distributed by AMO (Santa Ana, ment (CTS) and the modified capsular tension ring
CA) and marketed under the name StabilEyes. It is (M-CTR) may be utilized. Descriptions and use of
available in uncompressed diameters of 12 mm and these devices will be discussed later in this chapter.
13 mm (compressed diameters of 10 mm and 11 mm In cataract surgeries requiring a CTR, the device
respectively). Both the Reform Ring and the StabilEyes can be implanted at any point after the capsulorrhexis
can be implanted using a universal injector made by is made.14 To facilitate the remaining steps in the sur-
Ophtec (Figure 14-2) or bimanually with forceps. gery, the CTR should be implanted as late as possible
In order to get maximum zonular support for 360 but as soon as necessary during a case with compro-
degrees, the ends of a CTR should overlap slightly mised zonules. A CTR should never be used if there is
after being inserted. Since the CTR cannot be visual- a tear in the anterior or posterior capsule, as the tear
ized once it is inserted, some surgeons advocate white- will almost certainly extend given the force placed on
to-white measurements or axial eye length to predict the capsular bag during insertion. Although there is no
the diameter of the capsular bag.13 There appears to be consensus, most surgeons believe that a CTR should
no disadvantage to having too large a CTR in an eye, not be utilized if there are more than 4 clock hours of
so many surgeons opt for placing the largest available zonulysis or more than mild diffuse zonular laxity.15
CTR in all cases (author’s preference). Generally, a standard CTR will have no beneficial
effect in an eye with severe laxity or significant loss
of zonules.
Capsular Tension Rings 117
Insertion Techniques
There are several techniques described for CTR
insertion, but the authors have found the universal
CTR injector (Ophtec) to be a simple and predictable
means of CTR placement.
6. Kim JH, Kim H, Joo CK. The effect of capsular tension ring timing of capsular tension ring implantation: Miyake-Apple
on posterior capsular opacity in cataract surgery. Korean J video analysis. J Cataract Refract Surg. 2005;31(9):1809.
Ophthalmol. 2005;19(1):23. 15. Hasanee K, Ahmed II. Capsular tension rings: update on endo-
7. Lee DH, Shin SC, Joo CK. Effect of the capsular tension ring capsular support devices. Ophthalmol Clin N Am. 2006;19:508.
on intraocular lens decentration and tilting after cataract sur- 16. Anqunawela RI, Little B. Fishtail technique for capsular ten-
gery. J Cataract Refract Surg. 2002;28(5):843. sion ring insertion. J Cataract Refract Surg. 2007:33(5):767.
8. Boomer JA, Jackson DW. Effect of the Morcher capsular 17. Lee V, Bloom P. Microhook capsular stabilization for phaco-
tension ring on refractive outcomes. J Cataract Refract Surg. emulsification in eyes with pseudoexfoliation-induced lens
2006;32(7):1180. instability. J Cataract Refract Surg. 1999;25:1567.
9. Scherer M, Bertelmann E, Rieck P. Late spontaneous in- 18. Cionni,RJ, Osher RH. Management of profound zonular di-
the-bag intraocular lens and capsular tension ring disloca- alysis or weakness with a new endocapsular ring designed for
tion in pseudoexfoliation syndrome. J Cataract Refract Surg. scleral fixation. J Cataract Refract Surg. 1998;24:1299.
2006;32(4):672. 19. Little BC, Richardson T, Morris S. Removal of the capsular
10. Oner FH, Kocak N, Saatci AO. Dislocation of the capsular tension ring from the anterior chamber angle. J Cataract Re-
bag with intraocular lens and capsular tension ring. J Cataract fract Surg. 2004;30(9):1832.
Refract Surg. 2006;32(10):1756. 20. Levy J, Klemperer I, Lifshitz T. Posteriorly dislocated capsu-
11. Goldman JM, Karp CL. Adjunct devices for managing chal- lar tension ring. Ophthal Surg Lasers Imaging. 2005;36(5):416.
lenging cases in cataract surgery: pupil expansion and stabili- 21. Jehan FS, Mamalis N, Crandall AS. Spontaneous late disloca-
zation of the capsular bag. Curr Opin Ophthalmol. 2007;18:44. tion of the intraocular lens within the capsular bag in pseudo-
12. Boomer JA, Jackson DW. Anatomic evaluation of the Morcher exfoliation patients. Ophthalmology. 2001;108(10):1727.
capsular tension ring by ultrasound biomicroscopy. J Cataract 22. Ahmed II, Chen SH, Kranemann C, Wong DT. Surgical re-
Refract Surg. 2006;32(5):846. positioning of dislocated capsular tension rings. Ophthalmol-
13. Vass C, Menapace R, Schetterer K, et al. Prediction of pseu- ogy. 2005;112(10):1725.
dophakic capsular bag diameter based on biometric variables. 23. Ma PE, Kaur H, Petrovic V, Hay D. Technique for removal
J Cataract Refract Surg. 1999;25:1376. of a capsular tension ring from the vitreous. Ophthalmology.
14. Ahmed II, Cionni RJ, Kranemann C, Crandall AS. Optimal 2003;110(6):1142.
Chapter
15
Preventing Postoperative
Infection and Inflammation
Nick Mamalis, MD
a high level of antibiotics in the cornea and anterior then be used to “tuck” the plastic drape around the lid
chamber.6 The use of preoperative antibiotics in a margin and lashes. This way, there will be no direct
loading fashion prior to surgery may also provide high contact of potentially contaminated eyelid margin or
levels of antibiotics. Patients may receive four sets of lashes with the surgical field.
antibiotic drops during their preoperative preparation
for surgery at the time that dilating drops are given to Intraoperative
the patient. It is critical to maintain antiseptic techniques
Topical fourth-generation fluoroquinolones have throughout the entire procedure. Instruments that are
low ocular toxicity, superior penetration through the used within the eye should be carefully sterilized and
cornea, and higher minimum inhibitory concentration care should be taken not to break the sterility of the
(MIC) levels in aqueous compared with third-gen- surgical field or the instrument tray at any time during
eration fluoroquinolones.7,8 In addition, the broad- the surgery.
spectrum coverage provided by these agents against The construction of the clear corneal wound used
both gram-positive and gram-negative organisms make for the majority of cataract surgeries is critical. Several
them theoretically ideal for prevention of postopera- relatively recent studies have raised a concern that
tive endophthalmitis.9,10 Fourth-generation fluoroqui- postoperative endophthalmitis following cataract sur-
nolones such as moxifloxacin (Vigamox, Alcon, Fort
gery is more likely with clear corneal incisions.15-17 This
Worth, TX) and gatifloxacin (Zymar, Allergan, Irvine,
concern is also backed by evidence of an increased rate
CA) have been shown to provide excellent prophylaxis
of post-cataract endophthalmitis since 1994, which
for most bacteria that are responsible for postopera-
coincides with the timeline for widespread use of
tive endophthalmitis and can rapidly attain high levels
unsutured clear corneal cataract incisions.3 Proper
within the cornea and the anterior chamber prior to
construction of the clear corneal wound is important
surgery.
to ensure a water-tight closure at the conclusion of the
Preoperative preparation of the patients for sur-
case. Studies by Ernest and co-authors18 have shown
gery is also very important as reports have shown that
that the clear corneal cataract wounds that are square
normal ocular flora from the eyelids or conjunctiva
are the most common bacteria causing endophthal- or nearly square in architecture are significantly more
mitis.11,12 Skin preparation should be undertaken with resistant to external deformation than those that are
10% Betadine on the lid skin and lashes surrounding more rectangular. In addition, Masket has shown that
the eye. Vigorous scrubbing of the lashes should not the design and length of the clear corneal incision is
be undertaken immediately before cataract surgery as critical to ensure that the incision seals at the conclu-
this may actually liberate bacteria from the eyelashes. sion of the case.19 Meticulous construction of a clear
One of the most important factors in the preopera- corneal incision to ensure adequate sealing of the
tive sterilization of the surface of the eye is the use incision at the conclusion of the case should have an
of 5% Betadine on the cornea and conjunctiva during acceptably low risk of postoperative endophthalmi-
preparation of the eye for surgery. This preoperative tis.20 Newer methods of evaluating the clear corneal
povidone-iodine antisepsis (when combined with pre- incision architecture have been developed using opti-
operative topical antibiotics therapy) has been shown cal coherence tomography. This imaging technol-
to markedly decrease the bacteria that are present on ogy allows evaluation of the architectural features of
the surface of the eye prior to cataract surgery.13,14 the clear corneal wound in patients postoperatively.
Great care should be taken during the draping Endothelial gapping and loss of coaptation postopera-
of the patient prior to surgery to ensure that there tively has been shown in some patients. This can be
is a barrier between the lid margin and lashes and potentially important at times immediately following
the surgical field. There are many plastic drapes that cataract surgery when the intraocular pressure (IOP)
are available that can perform this important func- is low, which would significantly increase the risk for
tion. With the use of topical anesthesia, the patient endophthalmitis.21
is instructed to widely open his or her eyes and the The role of antibiotics in the irrigating solution to
sticky plastic drape is then placed over the eyelids and try and prevent endophthalmitis during the procedure
lashes. A sharp scissors can then be used to make an is quite controversial. Surgeons have advocated the use
opening through the center of the drape overlying the of antibiotics in the irrigating solution for prevention
cornea. Either an open- or closed-loop speculum may of endophthalmitis in the past.22 Gentamicin sulfate as
Preventing Postoperative Infection and Inflammation 125
well as vancomycin in the irrigating solution has been roquinolones as an agent for intracameral prophylaxis
advocated. However, antibiotics within the irrigating of endophthalmitis.27
solution provide a relatively low dose of antibiotics An additional concern at this time is that no com-
for a short period of time, which would not render a mercially available, Food and Drug Administration
bacteriostatic antibiotic useful in the killing of bacteria approved antibiotics are available to the ophthalmic
during cataract surgery. There is also a concern about surgeon in a unit dose delivery device for the use of
the possibility of toxicity from intraocular gentamicin these antibiotics intracamerally. These antibiotics have
use.23 In addition, the misdosing of the antibiotic with- to be custom mixed for injection into the anterior
in the irrigating solution has the potential for causing chamber at the conclusion of the case. This raises the
postoperative inflammation or toxic anterior segment potential for problems regarding the administering of
syndrome (TASS). Therefore, antibiotics within the “homemade” intracameral antibiotics. Possible dilution
irrigating solution during cataract surgery are not errors, bacterial contamination, or even the creation of
recommended. TASS is a concern. A recent survey of members of the
Another way of attaining a high dose of antibiotics American Society of Cataract and Refractive Surgery
at the immediate conclusion of cataract surgery is the (ASCRS) found that this was a significant concern to
use of intracameral antibiotics. Intracameral cefurox- 45% of surgeons currently not using intracameral anti-
ime has been evaluated for endophthalmitis prophylax- biotics. At present, more than 80% of ASCRS mem-
is and has gained widespread acceptance in countries bers expressed a need for a commercially approved
such as Sweden. Montan and coauthors have shown preparation at a reasonable cost that would lead to
that a 1.0 mL dose of intracameral cefuroxime appar- routine injection of intracameral antibiotics.28
ently has no signs of toxicity on the corneal endothe-
lium or on the anterior segment.24 The decreased
rates of postoperative endophthalmitis in Sweden
Postoperative
Use of postoperative antibiotics for the prevention
since the adaptation of cefuroxime helped stimulate
of endophthalmitis following cataract surgery has now
the European Society of Cataract and Refractive
become routine and some may argue that this is the
Surgery (ESCRS) to perform a prospective, investiga-
tor-masked, placebo-controlled multicenter clinical standard of care. However, there is very limited and
trial to evaluate the use of cefuroxime intracamerally in often indirect evidence regarding the efficacy of the
the prevention of endophthalmitis. The ESCRS study use of postoperative antibiotics in the prevention of
was a large multicenter study that eventually included endophthalmitis. The huge numbers necessary to per-
greater than 16,000 patients. They found that risk for form a study as well as the ethical issues involved with
presumed infectious endophthalmitis postoperatively the use of a placebo make randomized, prospective,
was increased nearly five-fold in patients who did controlled studies very difficult to perform to confirm
not receive intracameral cefuroxime (0.30%) com- the efficacy of postoperative antibiotics in the preven-
pared to those receiving the intracameral antibiotic tion of endophthalmitis. When postoperative antibiot-
(0.06%).25,26 However, there have been several ques- ics are used, it is very important that they be used in
tions raised about the limitations of the ESCRS study a proper manner. There has been a marked increase
following publication. First of all, levofloxacin was used in resistance to second generation fluoroquinolones
for the topical antibiotic prophylaxis. Since that study noted over the past decade.29-31 The rapid increase in
began, the fourth-generation fluoroquinolones moxi- resistance to so-called second-generation fluoroqui-
floxacin and gatifloxacin have gained widespread use nolones has rendered these drugs much less useful in
in the United States and there is evidence to support the prophylaxis of postoperative endophthalmitis. The
the fact that these fourth-generation fluoroquinolones most common bacteria implicated in endophthalmitis
are a better choice for topical antibiotic prophylaxis to are coagulase-negative staph, Staph aureus, and strep
prevent endophthalmitis. Other methods of providing species. The availability of fourth-generation fluoro-
intracameral antibiotic prophylaxis are being evaluated quinolones gatifloxacin and moxifloxacin has lead to
at the moment. Fourth-generation fluoroquinolones their widespread use for postoperative prophylaxis of
such as moxifloxacin, which have potent and rapid endophthalmitis. The incidence of resistance to these
bacteriocidal activity against common gram-positive new fluoroquinolones is much decreased compared to
pathogens, have been evaluated. There are theoretic older generations. However, resistance even to fourth-
advantages to the use of these fourth-generation fluo- generation fluoroquinolones is now being reported.32
126 Chapter 15
At the conclusion of the surgical procedure, two be quiet preoperatively for a minimum of 6 weeks
drops of fourth-generation fluoroquinolone should be prior to contemplating cataract surgery. In patients
placed onto the cornea while the patient is still in the with a history of uveitis, it is recommended that anti-
operating room. The patient should then be instructed inflammatory drops be started at least 1 week prior to
to use this antibiotic every 2 hours for the first day surgery. Prednisolone acetate (Pred Forte [Allergan,
following surgery. The fourth-generation fluoroquino- Irvine, CA]) as well as a nonsteroidal anti-inflamma-
lone antibiotic should then be used four times per day tory drug (NSAID) should be used four times per day
for 7 days following surgery and should be abruptly for the week prior to surgery. In patients with a history
discontinued in routine cases. There is no place for the of severe uveitis, oral prednisone in a moderate dose
tapering of antibiotics in the postoperative period as of 40 to 50 mg per day may also be started during this
this may increase the risk of formation of resistance to period of time.
these antibiotics. In a routine cataract patient without a history
The prevention of postoperative endophthalmitis of preexisting uveitis, it is unclear how soon prior to
following cataract surgery is a multi-faceted procedure. surgery that NSAID use should be started. There
This begins with a thorough preoperative evaluation of are advantages in beginning NSAID therapy prior to
the patient including treatment of any preexisting dac- surgery so that there is adequate blockage of pros-
ryocystitis and blepharitis. Preoperative preparation of taglandins release at the time of surgery. In addition,
the patient including the use of antibiotics and povi- use of NSAIDs preoperatively will help to prevent
done-iodine is essential. Careful attention to draping progressive pupil miosis during the surgical procedure.
and preparing of the patient’s eye with adherence to Preoperative NSAID regimens for the treatment of
aseptic techniques is important. The design and con- anterior segment inflammation vary from beginning
treatment 1 to 3 days prior to surgery to starting with
struction of a clear corneal wound is critical to allow
a dose immediately before surgery. This is quite similar
sealing of the wound at the conclusion of the case to
to the use of preoperative antibiotics for the prevention
decrease the potential risk of ingress of bacteria. Lastly,
of endophthalmitis. It is reasonable to begin NSAID
the use of antibiotics intracamerally at the conclusion
treatment when the patient is in the preoperative hold-
of the surgery as well as postoperatively should help to
ing area with three drops of NSAID given at the same
decrease the risk for postoperative endophthalmitis.
time as the antibiotic and dilating drops. Some would
argue that preoperative treatment with NSAIDs fol-
Preventing Postoperative lowed by combination therapy with NSAIDs and cor-
ticosteroids postoperatively has become the standard
Inflammation of care in cataract surgery.33,34
Control of postoperative inflammation following
cataract surgery is important to prevent sequelae of
chronic inflammation such as corneal decompensation,
Postoperative
The most common postoperative regimen for the
glaucoma, synechiae formation, and cystoid macular
treatment of inflammation in a routine cataract patient
edema (CME). Control of postoperative inflamma-
is the use of 1% prednisolone acetate four times per
tion becomes even more important in patients with
day for 2 weeks with tapering depending on the condi-
conditions that predispose them to breakdown of the tion of the patient and any preexisting conditions that
blood aqueous barrier such as diabetes, and a history would cause a breakdown of the blood-aqueous barrier
of preexisting iritis or uveitis. As with the prevention postoperatively. This can be supplemented by NSAID
of infection, the prevention of postoperative inflam- treatments, which are once again used four times per
mation begins in the preoperative period and extends day with a similar tapering dose. Postoperative use of
through the surgery to the postoperative period. anti-inflammatory medications such as corticosteroids
or NSAIDs may help reduce inflammation and pre-
Preoperative vent possible postoperative complications.35 The use
Patients with a history of uveitis, iritis, or any of NSAIDs in addition to corticosteroids or used by
inflammatory condition should be carefully evaluated themselves prophylactically may help prevent postop-
in the clinic prior to consideration of cataract surgery. erative inflammation and sequelae such as CME.36
It is essential that there is no active uveitis present at There are many different NSAIDs available for
the time of cataract surgery. The patient’s eye should the prevention of postoperative inflammation as well
Preventing Postoperative Infection and Inflammation 127
as to help minimize pain in the postoperative period. ondary to widespread endothelial damage, as well as
These include such NSAIDs as ketorolac trometh- marked anterior segment inflammation with hypopyon
amine (Acular, Acular LS, Allergan) and diclofenac and fibrin formation. Finally, TASS can cause diffuse
sodium 0.1% (Voltaren ophthalmic, Novartis, Duluth, iris damage as well as damage to the trabecular mesh-
GA). In addition, there are some newly available work leading to glaucoma.39,40
NSAIDs that may require less frequent dosages and Potential etiologic factors involved in TASS are
have some potential advantages regarding penetra- extremely broad and include problems with intraocular
tion and onset of anti-inflammatory effect. Nepafenac irrigating solutions such as balanced saline solution
ophthalmic suspension 0.1% (Nevanac, Alcon, Fort (BSS). This includes abnormalities of pH, osmolarity,
Worth, Texas) is a very effective NSAID with inhibi- ionic composition, problems with contaminants, medi-
tion of cyclooxygenase 1 and 2. It also has a relatively cations added to the solution, or potential endotoxin
long duration of action. Nevanac crosses the cornea contamination. Any medications that are used intra-
rapidly and then undergoes bioactivation within ocular ocularly including analgesics and antibiotics have the
tissue to amfenac. The dosing regimen of nepafenac potential to cause inflammation. It is important that
0.1% three times a day starting 1 day prior to surgery any medications used have the proper concentration
and continuing for 14 days after surgery used as a and be preservative free if they are injected into the
sole postoperative treatment was found to prevent as eye. Problems with ophthalmic viscosurgical devices
well as treat ocular inflammation and pain associated (OVDs) can cause postoperative inflammation and
with cataract surgery in a large multicenter study.37 TASS. An emerging issue that is of critical importance
Another newer NSAID is bromfenac ophthalmic solu- in the causation of TASS is the cleaning and steriliza-
tion 0.09% (Xibrom, ISTA Pharmaceuticals, Irvine, tion of ophthalmic instruments.
CA), which similarly acts to prevent inflammation in The most important factor in the prevention of
the arachidonic acid cascade through the inhibition TASS is the recognition of possible factors that may
of cyclooxygenase. Bromfenac sodium is available in a be involved in causing postoperative inflammation and
0.09% solution and may be dosed two or three times elimination of as many factors as possible. The corneal
per day postoperatively for the treatment and preven- endothelium as well as the trabecular meshwork and
tion of anterior segment inflammation and reduction cells within the iris are very sensitive to any toxic
of ocular pain following cataract surgery. Two large insult. This may lead to corneal edema due to acute
phase-three studies confirmed that bromfenac effec- breakdown of endothelial junction and loss of bar-
tively and rapidly cleared ocular inflammation as well rier function. In addition, there may be a broad based
as reduced ocular pain following cataract surgery breakdown of the blood-aqueous barrier leading to
with no serious ocular adverse events.38 The use of increased inflammation in the anterior segment. It is
NSAIDs and prednisolone acetate are essential in the important that any solution used during cataract sur-
prevention of postoperative inflammation and pain fol- gery, especially the BSS, be of the proper composition
lowing cataract surgery. These medications may help chemically. Incorrect pH as well as incorrect osmolar-
to decrease the potential for postoperative inflamma- ity or problems with additives may cause postoperative
tory complications following cataract surgery such as inflammation and TASS.
CME. NSAIDs may also be helpful for the prevention Preservatives in ophthalmic medicines that are
of intraoperative miosis. used either intraocularly or at the conclusion of the
case postoperatively are potentially toxic, especially to
the corneal endothelium. There have been reports of
Toxic Anterior medications with preservatives inadvertently injected
Segment Syndrome into the eye during the anterior segment surgery, which
TASS is an acute, sterile anterior segment inflam- may cause TASS.41 Many ophthalmic medications are
mation following any anterior segment surgery. The preserved with benzalkonium chloride (BAK). The
most common hallmark of TASS is markedly blurred corneal endothelium is quite sensitive to any medica-
vision, which patients often note almost immediately tions that have BAK preservatives within them. In addi-
after cataract surgery with many signs and symptoms tion, it is important to recognize that some medications
appearing within 12 to 48 hours of surgery. The do not necessarily have a preservative but have a sta-
most common clinical findings include diffuse corneal bilizing agent added to them that may be toxic. The
edema, which has been called “limbus-to-limbus” sec- epinephrine that is used in BSS during the procedure to
128 Chapter 15
help prevent pupil miosis needs to be preservative free. The cleaning and sterilization of ophthalmic
This includes bisulphites and metabisulphites, which instruments has become a very important factor when
are technically stabilizers rather than preservatives, but analyzing outbreaks of TASS. Many centers are using
may still be toxic to the corneal endothelium. enzymes and detergents in the cleaning of reusable
Intraocular anesthetics that are used during cata- ocular instruments between cases. Any residue of
ract surgery once again need to be preservative free. detergent or enzyme on the instruments is potentially
In addition, any intraocular anesthetic should be of inflamagenic. Enzymes or active ingredients in these
the proper concentration. Preservative free lidocaine detergents are often not deactivated in standard auto-
at a 1% dose appears to be safe for cataract surgery. claves and may cause significant inflammation when
However, dosages higher than 2% have been found to flushed into the eye when the instruments are used
cause significant corneal thickening and opacification again.47 Detergent residues left on ophthalmic instru-
postoperatively.42-44 Therefore, intraocular anesthetics ments can cause toxicity to the corneal endothelium.
should not only be preservative free but of the proper Breebaart and coauthors described severe toxic endo-
concentration. thelial cell destruction following surgery with deter-
The use of intraocular antibiotics has been dis- gent residues found on reusable cannulas.48
cussed previously. The use of gentamicin and van- In addition to possible residues of detergent or
comycin in irrigating solutions has been discouraged enzymes, outbreaks of TASS have been found to be
due to potential problems with toxicity, especially related to endotoxin contamination of the instruments
involving gentamicin. While intracameral antibiotics that occurs during sterilization. Ultrasounds or water
such as cefuroxime have been shown to be safe when baths that are used for the treatment of instruments
properly mixed, concerns have been raised with poten- following surgery may grow gram-negative bacteria.
tial problems involving “kitchen pharmacies.” Incorrect Although the bacteria are destroyed during heat ster-
dosage, problems with sterility, and other issues with ilization in autoclaving, heat stable lipopolysaccharide
the customer mixing of intracameral antibiotics may endotoxins from the gram-negative bacteria cell wall
potentially lead to issues with TASS. remain active and may be attached to the instruments
OVDs are a potential source of TASS. It is essential following stabilization. Injection of the endotoxin into
that the OVDs be completely removed at the conclu- the eye during the surgery may cause significant ante-
sion of the surgical procedure and that large amounts rior segment inflammation.49
of OVDs are not left within the capsular bag or the The potential etiologic factors involved in an out-
posterior chamber. This could lead to increased post- break of TASS are extremely broad. Analysis of TASS
operative inflammation and difficult to control IOP. outbreaks often reveals multiple potential sources
In addition, OVD residues on reusable cannulas and rather than a single point source associated with the
irrigation/aspiration tips that are not properly flushed outbreak.50 The increased incidence of TASS over the
following cataract surgery may be associated with past 2 years has lead to the formation of an ASCRS-
TASS. This retained OVD may become broken down sponsored TASS task force to evaluate outbreaks of
or altered during sterilization, which can cause toxic TASS. Educational materials from the task force are
inflammation when this is subsequently flushed into available including a video symposium involving mem-
the eye.45 bers of the task force with input from nursing organiza-
Another potential source of TASS that may occur tions involved in ophthalmology (www.tassfacts.com).
either acutely or on a delayed onset basis is the ingress In addition, reports from the task force are available
of topical ophthalmic ointment, used postoperatively, on the ASCRS Web site (www.ascrs.org) as well as on
into the anterior segment of the eye. Many ophthal- the American Academy of Ophthalmology Web site
mic ointments are petroleum based and deposition (www.aao.org). A complete guideline for the cleaning
of hydrocarbon material within the vehicle of these and sterilization of ophthalmic instruments is also avail-
postoperative ointments may cause toxicity within the able on the ASCRS Web site and has been published
eye.46 This ingress of ointment is only possible through recently.51 The prevention of TASS is a team effort
a clear corneal wound that is not water tight or incom- involving not only the surgeon but the entire operating
petent at the conclusion of the surgery and once again room staff including nurses and those involved in the
brings forth the importance of a well-constructed clear cleaning and sterilization of instruments as well as the
corneal wound. ordering of ophthalmic medications.
Preventing Postoperative Infection and Inflammation 129
35. Rowen S. Preoperative and postoperative medications used 44. Kim T, Holley GP, Lee JH, et al. The effects of intraocu-
for cataract surgery. Curr Opin Ophthalmol. 1999;10:29-35. lar lidocaine on the corneal endothelium. Ophthalmology.
36. Flach AJ. Topical non-steroidal anti-inflammatory drugs in 1998;105:120-125.
ophthalmology. Int Ophthalmol Clin. 2002;42:1-11. 45. Kim JH. Intraocular inflammation of denatured viscoelastic
37. Lane SS, Modi SS, Lehmann RP, Holland EJ. Nepfenac oph- substance in cases of cataract extraction and lens implanta-
thalmic suspension 0.1% for the prevention and treatment of tion. J Cataract Refract Surg. 1987;13:537-542.
ocular inflammation associated with cataract surgery. J Cata- 46. Werner L, Shear JH, Taylor JR, et al. Toxic anterior segment
ract Refract Surg. 2007;33:53-58. syndrome and possible association with ointment in the an-
38. Donnenfeld ED, Holland EJ, Stewart RH, et al. Bromfenac terior chamber following cataract surgery. J Cataract Refract
ophthalmic solution 0.09% (Xibrom) for postoperative ocular Surg. 2006;32:227-235.
pain and inflammation. Ophthalmol. 2007;114:1654-1662.
47. Parikh C, Sippy BD, Martin DF, Edelhauser HF. Effects of
39. Mamalis N, Edelhauser HE, Dawson DG, et al. Toxic anterior
enzymatic sterilization detergents on corneal endothelium.
segment syndrome–Review/Update. J Cataract Refract Surg.
Arch Ophthalmol. 2002;120:165-172.
2006;32:324-333.
48. Breebaart AC, Nuyts RM, Pels E, et al. Toxic endothelial cell
40. Mamalis N. Toxic anterior segment syndrome (Editorial). J
destruction of the cornea after routine extracapsular cataract
Cataract Refract Surg. 2006;32:181-182.
41. Liu H, Routley I, Teichmann KD. Toxic endothelial cell de- surgery. Arch Ophthalmol. 1990;108:1121-1125.
struction from intraocular benzalkonium chloride. J Cataract 49. Kreissler KR, Martin SS, Young CW, et al. Postoperative in-
Refract Surg. 2001;27:1746-1750. flammation following cataract extraction caused by bacterial
42. Kadonosono K, Ito N, Yazama F, et al. Effect of intracameral contamination of the cleaning bath detergent. J Cataract Re-
anesthesia on the corneal endothelium. J Cataract Refract Surg. fract Surg. 1992;18:106-110.
1998;24:1377-1381. 50. Mamalis N. Anatomy of a TASS outbreak (Editorial). J Cata-
43. Guzy M, Satici A, Dogan Z, Karadede S. The effect of bupi- ract Refract Surg. 2007;33:357-358.
vacaine and lidocaine on the corneal endothelium when ap- 51. ASCRS and ASORN. Recommended practices for cleaning
plied into the anterior chamber at the concentration supplied and sterilizing intraocular surgical instruments (Special Re-
commercially. Ophthalmologica. 2002;216:113-117. port). J Cataract Refract Surg. 2007;33:1095-1100.
Chapter
16
Optimizing Refractive
Outcomes
H. John Shammas, MD; Eric Donnenfeld, MD; and Renée Solomon, MD
131
132 Chapter 16
Regression Formulas Please note that P, the IOL power for emmetropia,
During the same period, Sanders, Retzlaff, and varies on a one-to-one ratio with A. This can be very
Kraff reviewed their results and thought that they helpful when switching between implants. For exam-
could get better results with a regression equation. ple, an IOL with an A constant of 118.9 will require a
Their equation became known as the SRK formula: 0.50 D stronger power than an IOL with an A constant
of only 118.4.
P = A – 2.5L – 0.9K Please also note that the A constant, among other
things, relates to the position of the IOL within the
where P is the power for emmetropia, L is the axial eye. This can be very helpful if the surgeon encoun-
length in mm, K the corneal power in diopters, with A ters complications during surgery. Let us suppose that
being a constant. the surgeon is planning to insert a one-piece IOL in
In 1988, the authors of the SRK formula realized the capsular bag and that implant has an A constant
that their regression equation did not perform well in of 118.9. If the posterior capsule is compromised, a
very long and very short eyes. They modified it by three-piece IOL has to be inserted in the sulcus; the
fudging the calculations in these very long and very A constant drops to 117.5 requiring a 1 to 1.5 diopters
short eyes, and calling it the SRK II formula. The weaker IOL. If vitreous is lost, and an anterior chamber
results improved but they were still not satisfactory. IOL is to be used, the A constant drops to 115.3 requir-
ing a 3.5 diopters weaker IOL.
Modern Theoretical Formulas
After 1988, three modern theoretical formulas
were introduced. These are the Holladay formula
Measuring the Axial Length
(1988), the SRK/T formula (1990), and the Hoffer Q
formula (1993). In all three formulas, ELP varies not
A-Scan Biometry
The axial length is conventionally measured with
only with axial length but also with the corneal curva-
ultrasonography, using a biometry unit. An immer-
ture. However, each formula uses a different constant:
sion technique is recommended where the ultrasound
SF (surgeon factor) for the Holladay formula, A for the
probe remains 5 to 8 mm away from the cornea.
SRK/T formula, and ACD (anterior chamber depth)
It is important to recognize the A-scan pattern of a
for the Hoffer Q formula.
normal phakic eye examined with an immersion tech-
The last decade saw the introduction of some
nique. The following echospikes are displayed from
advanced theoretical formulas, mainly the Holladay II
left to right (Figure 16-1):
and the Haigis formulas. In the Holladay II formula,
The initial spike is produced at the tip of the
ELP varies with the axial length, corneal curvature,
probe. It has no clinical significance.
white-to-white measurement, anterior chamber depth,
The corneal spike is double peaked represent-
lens thickness, and age. This formula requires the
ing the anterior and posterior surfaces of the
purchase of a special software program to run it. The
cornea.
Haigis formula is widely available on the IOLMaster
The anterior lens spike is generated from the
(Carl Zeiss Meditec, Dublin, CA).
anterior surface of the lens.
The posterior lens spike is generated from the
Clinical Pearl
posterior surface of the lens.
For your IOL power calculations, only use a mod-
The retinal spike is generated from the anterior
ern or advanced theoretical formula. Avoid the SRK
surface of the retina. It is straight, highly reflec-
and SRK II formulas.
tive, and tall whenever the ultrasound beam is
perpendicular to the retina, as it should be dur-
Clinical Pearl
ing axial length measurement.
The A constant has become a value character-
The scleral spike is another highly reflective
izing each IOL, and every manufacturer prints an A
spike generated from the scleral surface, right
constant on the box holding the IOL. Although this A
behind the retinal spike, and should not be
constant has to be personalized for each surgeon and
confused with it.
for each lens model, the value given by the manufac-
The orbital spikes are low reflective behind the
turer is often very close.
scleral spike.
Optimizing Refractive Outcomes 133
K = (1.3375 – 1) ÷ r
Part B:
Limbal Relaxing Incisions
Eric Donnenfeld, MD, and Renée Solomon, MD
A B
C
D
Figure 16-4. In with-the-rule astigmatism, the steep axis is vertical (A). In against-the-rule astigmatism, the
steep axis is horizontal (B). Oblique astigmatism occurs when the steep axis is neither vertical or horizon-
tal (C). Irregular astigmatism occurs when the steep and flat axis are not at a 90-degree angle (D).
Figure 16-5. LRIs relax the steep axis of the astigmatism and allow the eye to heal into a more spherical
shape.
138 Chapter 16
A B
Figure 16-9. An astigmatism marker can be used and the cornea can be marked.
Improving refractive outcomes is an important ract surgery. Cataract Refract Surg Today. 2006;7:41-44.
goal for cataract surgeons today and learning to per- 10. Tejedor J, Murube J. Choosing the location of corneal inci-
sion based onpreexisting astigmatism in phacoemulsification.
form LRIs is a useful step in achieving this end. The Am J Ophthalmol. 2005;139(5):767-776.
good news is that LRIs are not difficult to learn and, 11. Kaufmann C, Peter J, Ooi K, et al. Limbal relaxing inci-
when performed properly, they are both predictable sions versus on-axis incisions to reduce corneal astigma-
and uniformly successful. tism at the time of cataract surgery. J Cataract Refract Surg.
2005;31(12):2261-2265.
12. Muller-Jensen K, Fischer P, Siepe U. Limbal relaxing inci-
References sions to correct astigmatism in clear corneal cataract surgery.
J Refract Surg. 1999;15(5):586-589.
1. Nichamin LD. Nomogram for limbal relaxing incisions. J
Cataract Refract Surg. 2006;32:1048. 13. Akura J, Matsuura K, Hatta S, Otsuka K, Kaneda S. A
2. Nichamin LD. Astigmatism control. Ophthalmol Clin North Am. new concept for the correction of astigmatism: full-arc,
2006;19:485-493. depth-dependent astigmatic keratotomy. Ophthalmology.
3. Wang L, Misra M, Koch DD. Peripheral corneal relaxing in- 2000;107(1):95-104.
cisions combined with cataract surgery. J Cataract Refract Surg. 14. Faktorovich EG, Maloney RK, Price FW Jr. Effect of as-
2003;29:712-722. tigmatic keratotomy on spherical equivalent: results of
4. Budak K, Friedman NJ, Koch DD. Limbal relaxing incisions the Astigmatism Reduction Clinical Trial. Am J Ophthalmol.
with cataract surgery. J Cataract Refract Surg. 1998;24:503. 1999;127(3):260-269.
5. Gills JP. Treating astigmatism at the time of cataract surgery. 15. Price FW, Grene RB, Marks RG, Gonzales JS. Astigmatism
Curr Opin Ophthalmol. 2002;13:2-6. reduction clinical trial: a multicenter prospective evaluation
6. Oshika T, Shimazaki J, Yoshitomi F, et al. Arcuate keratom- of the predictability of arcuate keratotomy. Evaluation of sur-
etry to treat corneal astigmatism after cataract surgery: a gical nomogram predictability. ARC-T Study Group. Arch
prospective evaluation of predictability and effectiveness. Ophthalmol. 1995;113(3):277-282. Erratum in: Arch Ophthalmol.
Ophthalmology. 1998;105:2012. 1995;113(5):577.
7. Maloney WF, Grindle L, Sanders D, et al. Astigmatism con- 16. Oshika T, Shimazaki J, Yoshitomi F, et al. Arcuate keratotomy
trol for the cataract surgeon: a comprehensive review of to treat corneal astigmatism after cataract surgery: a prospec-
surgically tailored astigmatism reduction (STAR). J Cataract tive evaluation of predictability and effectiveness. Ophthal-
Refract Surg. 1989;15:45. mology. 1998;105(11):2012-2016.
8. Price FW, Greene RB, Marks RG, et al. Astigmatism reduc- 17. Bradley MJ, Coombs J, Olson RJ. Analysis of an approach to
tion clinical trial: a multicenter prospective evaluation of astigmatism correction during cataract surgery. Ophthalmo-
the predictability of arcuate keratotomy. Arch Ophthalmol. logica. 2006;220(5):311-316.
1995;113:277. 18. Package insert. Acrysof Toric TM SA60T4IOL, Alcon Labora-
9. Devgan U. Corneal correction of astigmatism during cata- tories, Inc.