You are on page 1of 2

SS-CRP DAIICHI

PENTRA 400

General Parameter

Test Name Code Channel Release Modified On


CRP CRP * √ Enable *

Characteristics Pre-Dilution Result


Sample Type Serum / Plasma Diluent Name Unit mg/dL
Number of reagent 2 Factor Decimal Position 2

Reagent Incub.time ( in Cycle ) Manual Patient Validation


Reagent Short Name Reag.Number
CRP *
On Board Stability Correlation
√ Linearity ( mg/dL ) Slope Intercept
Cassette Low Limit High Limit 1 0
0.03 30

Automatic Rerun
Post Dilution
Dilution Factor Delta Check
Post Concentration Delta Check Validity Absolute Validation Relative Validation
Concentration Factor

Reference Range
Man/Default Woman Child
Low Check
√ High Check 0.3 0.3

* Data to be chosen by the user


Calibration Parameters
Test Name Code Channel
CRP CRP *

Check
Predilution
Type √ Reagent limit absorbance check
Reagent range low -3.00000
Factor 1 Factor 2 Factor 3 Factor 4 Reagent range high 3.00000

Factor 5 Factor 6 Factor 7 Factor 8 Reagent blank limit absorbance check


Blank range low limit
Blank range high limit
Calibration Validity
Calibration Mode Liniar inter. On Request √
Level 5 Time Validity
Calibration Factor √ Control Required
Run (s) 2 Interval Time Limit
√ Deviation 30 Control Used
Calibration Used CRP Calib Seronorm level 1 / 2

* Data to be chosen by the user

Analysis Parameters
Test Name Code Channel
CRP CRP *

Cleaner Wavelength ( nm ) Blank


Cleaner Solution Primay Wavelenght 570 Diluent
√ H2O √ Reagent Blank
Before Secondary Wavelenght 800
After
Analysis Sequence
Cycle Reagent Needle Vol ( uL ) Sample Needle Vol ( uL ) H2O Vol ( uL ) Mixing Speed
1 R1 150 Sample 3.0 10.0 40
25 R2 50

* Data to be chosen by user


Calculation Parameters
Test Name Code Channel
CRP CRP *

Correlation Factor
Slope 1 Reaction Direction Sample Limit Check
Intercept 0 √ Reaction Direction Check
Sample Limit ( ∆OD )
Reaction Direction Increase Sample Limit Cycle

Definition

Calculation Type End Point

Reaction Limit Check First Reading Last Reading


Reaction Limit Absorbance Cycle 28 Cycle 50
Cycle

* Data to be chosen by user

Unit Parameters

Test Name CRP


Code CRP
Channel *

Unit Conversion Factor

Modified on *

* Data to be chosen by user