Professional Documents
Culture Documents
Planning radiation therapy treatments for lung tumors with or without inhomogeneity
correction results in differences in dose distribution to the target volume, and also to the
Standard isodose charts and depth dose tables were created in reference to radiation beam
soft tissue. When the beam encounters tissue with density variations, such as air, bone,
metal, or odd shaped combinations of tissues with various densities, it causes a change to
There are two ways in which tissue inhomogeneity affects dose distribution. The first is
that it changes the attenuation of the primary beam the pattern of photon interactions
downstream. The second is the way the electron fluence changes as a result. The greater
the density of the inhomogeneity, the greater the attenuation and loss of primary beam
energy, but also increased secondary electron fluence. Conversely, if the inhomogeneity
is less dense, there is less attenuation to the primary beam, so it maintains higher energy
them. As the beam traverses the lower density of the lung tissue, the beam maintains a
higher energy, or depth dose compared to the standard isodose charts, or what would be
expected in water or soft tissue. This results in higher doses to the lung than would be
expected, unless we corrected for the inhomogeneity with a factor for lung density.
A second issue affects dose distribution to the lung tumor, particularly if the mass is
situated such that the beam traverses lung tissue prior to contact with the mass. In this
case, the secondary electron fluence decreases and requires penetration of some depth,
generally greater than 1.5cm for 6 MV beam energy, before the normal electron fluence
would reach the expected level that it would have in water or soft tissue. In other words,
the electron scatter component of the dose would need to build up with depth upon
A third issue for dose coverage to a lung tumor is that, for a small lung tumor, to be
treated with a small field size, there would be an increased dose fall off near the edges of
the field due to scattered electrons escaping the lateral beam edges before depositing their
energy. This is because they would travel farther in the lower density lung tissue than in
water or soft tissue, allowing more of the scattered electrons to escape the beam laterally.
This decreases the dose due to the scattered electron component around the periphery of
My clinical site, Renown Institute for Cancer, uses Acuros XB to calculate dose for lung
tumors. This algorithm is very strong in its capability to account for dose in
The one lung treatment plan that I was involved in was an SBRT treatment, and
inhomogeneity correction was turned on. With the capabilities of Acuros in predicting
radiation interactions with matter, we rely on the dose calculations produced by Acuros.
We have a high level of trust in the capabilities of Acuros to present accurate predictions
downstream from the inhomogeneity. For this reason, we leave it turned on. The
physicians, in turn, must accept the physical reality of some of the limitations in dose
The only times that we override the inhomogeneity values from the CT scan are either:
dental implant, causing the CT values to misrepresent the real tissue density, and
2. When there is a contrast agent, or other radiopaque material present that will not
Hounsfield unit. In these cases, the volume containing contrast is overridden and
Within the field of radiation oncology, the debate continues, whether or not to use
heterogeneity correction2. Proponents of using the correction argue that more accurate
Those who have reservations about utilizing inhomogeneity corrections express concerns
due to the idea that treatment regimens have been established and refined under
conditions in which inhomogeneity corrections were not used. The concern is that
although inhomogeneity corrections may result in more accurate dose calculations, their
use may still result in dose deliveries that differ from those which were used to establish
the successful treatment regimens, and possibly result in less favorable patient outcomes.
Yet, radiation oncology is a field that is rightfully determined to operate with the greatest
possible accuracy, to the point of sub-millimeter, in some cases. More accurate methods
will prevail and become the norm as time passes, and older, less accurate methods will
One clinical scenario in which severe inhomogeneity exists is when a patient has metal
inside of them as they undergo CT simulation, treatment planning and treatment delivery.
A study by Maerz et al.3 found that severe streaking artifacts caused by metallic objects
within patients impact CT data and inhibit accurate representation of shape and density of
the high density object(s) and the surrounding tissue. This study determined that the
Another study by Ziemann et al.4 found that inaccuracies resulting from streaking related
to high density objects resulted in calculation inaccuracy in the order of 8.4%. They also
significantly improved planning accuracy, and could reduce this number to a much more
reasonable 2.7% inaccuracy due to the same high density artifact (HDA).
In order to compare and evaluate the use of inhomogeneity correction in the treatment
planning environment, I will set up a test patient with a lung tumor as the target volume.
The inhomogeneity will involve the presence of lung tissue adjacent to target volume.
With my treatment plan designed, I will calculate the dose in the default setting – with
inhomogeneity correction turned on, and I will calculate again with inhomogeneity
correction turned off. This will reveal the differences between calculations, which
attempt to account for dose distribution with the inhomogeneity scenario involved with a
lung tumor, and calculations, which instead perceive the patient as a singular,
According to Essers et al.5, lung treatments are more accurate when incorporating a
correction, but I will perform a comparison to test and evaluate the differences to enhance
my understanding of the differences between lung plans generated with or without use of
the correction.
turned on, and off, I completed the following steps: I imported the patient’s DICOM data
and structures to Eclipse. In the contouring tab, I placed a carbon support structure under
the patient. In Eclipse, I created a new plan with a dose of 6000 cGy prescribed to the
PTV. I placed an AP beam with 6 MV energy. I created a new MLC and shaped it to the
PTV with a 1cm margin. I created a new opposed field from the PA, with 6 MV energy.
I created a new MLC for the PA field and shaped it to the PTV with a 1cm margin. I
noticed sloping isodose curves toward the patient’s superior and lateral sides, so I placed
10 degree dynamic wedges on both fields with heels toward the superior and lateral sides
to even out the isodose curves. This should level the isodose curves to produce a more
uniform dose distribution, and result in a reduced hot spot value after normalization. I
selected Acuros XB 13.5 for the calculation model and algorithm. I turned off the
heterogeneity correction. I calculated the plan. I observed the dose distribution, then,
adjusted beam weighting to achieve balanced coverage and minimum hot spot. I
normalized 100% dose to 95% of the PTV. I check for optimum beam weighting again
to minimize the hot spot. I copied/pasted the plan and turned on the heterogeneity
correction. I calculated the plan, and then adjusted the beam weighting to achieve
minimum hot spot. Now, I have two plans, with and without heterogeneity correction,
for comparison.
Looking at the dose distribution between the plan with inhomogeneity correction on, and
the plan with inhomogeneity correction off, there are some apparent differences in
appearance. First, the plan with inhomogeneity correction off has much more geometric
and symmetric appearance to the dose distribution. The curves of dose distribution in the
axial views are more smooth and rounded. The dose distributions on the coronal and
sagittal views are more smooth and representative of depth dose curves. The dose
distributions for the plan with inhomogeneity correction on are less smooth, have
attenuation due to the accounting for the range of tissue densities that the beams traverse.
Figure 1: Inhomogeneity ON
By comparing two treatment plans for a lung tumor, with and without inhomogeneity
correction, I was able to enhance my understanding of the differences between the two
surrounding normal tissue. By normalizing 100% of the dose to 95% of the target
volume, both plans achieved identical target coverage values. I was able to identify
differences in hot spot and surrounding normal tissues. I was also able to identify
differences in the appearances of dose distribution between the two plans, with much
more variation and details apparent in the plan with inhomogeneity correction turned on,
and smoother, more symmetrical isodose lines, more closely representing depth dose
curves, in the plan with inhomogeneity correction turned off. These observations have
not changed my opinion that we produce more accurate treatment plans with
1. Khan FM, Gibbons JP. The Physics of Radiation Therapy. Philadelphia: Lippincott
doi:10.1118/1.1287645.
0774-2.
calculation in radiation therapy due to metal artifact correction using the augmented
2018;19(3):227-233. doi:10.1002/acm2.12325.
5. Essers M, Lanson JH, Leunens G, Schnabel T, Mijnheer BJ. The accuracy of CT-
based inhomogeneity corrections and in vivo dosimetry for the treatment of lung cancer.
com.libweb.uwlax.edu/login.aspx?direct=true&AuthType=ip,uid&db=mnh&AN=874658