Calculating Total
Uncertainty of
Temperature
Calibration with
a Dry Block
I
n Calibration World magazine (Spring/Summer 2011) – Display accuracy
there was an article on temperature and the calibration – Axial uniformity
of temperature instruments. In this article we continue – Radial uniformity
with temperature-related articles. We will discuss the various – Loading
uncertainty components related to temperature calibration – Stability over time
using a temperature dry-block. Also, we will discuss how to – Hysteresis
calculate the total uncertainty of a calibration performed with – Sufficient immersion (15 x diameter)
a dry block. – Stem loss for 6 mm or greater probes
– Probe clearance
(<= 0,5 mm at –80 … 660 °C)
Temperature dry block
(<= 1,0 mm at +660 … 1300 °C)
First, let’s discuss what a temperature dry block is:
• consists of a heatable and/or coolable metallic block,
Related uncertainty components
controller, an internal control sensor and optional readout
for external reference sensor. This article will focus on Let’s have a look at the various uncertainty components that
models that use interchangeable metallic multi-hole inserts. are related to temperature calibration done using a dry block.
• There are fast and lightweight dry blocks for industrial field These components are relevant to all manufacturers’ dry
use as well as models that deliver near bath-level stability in blocks. Some manufacturers specify these components and
laboratory use. some do not.
• There are also some work safety issues that favor dry blocks It is possible to use a dry block with the block’s internal
in preference to liquid baths. For example, in temperatures measurement as the reference (true value), or you can also use
above 200 °C liquids can produce undesirable fumes or there an external reference temperature probe inserted in the block
may be fire safety issues. If a drop of water gets into hot as a reference measurement.
silicon oil, it could even cause a small steam explosion which
may splash hot oil on the user.
Internal measurement as reference
• Dry blocks are almost without exception meant to be used
dry. Heat transfer fluids or pastes are sometimes used around First, let’s discuss the uncertainty components related to the
or inside the insert, but they don’t necessarily improve use of a dry block’s internal measurement as reference.
performance. They may actually even impede the dry The following components should be taken into account:
block’s performance and damage its internal components.
• Display accuracy (accuracy of the internal measurement)
– It is important to remember that all of the thermometers
EURAMET
based on thermal contact measure their own temperature.
The EURAMET guideline (EURAMET /cg-13/v.01, July 2007 With dry blocks, the internal control sensor is typically
[previously EA-10/13]): located inside the actual block, whereas the probes to
• This Euramet calibration guide defines a normative way to be calibrated are immersed in the insert. There is always
calibrate dry blocks. As most of the manufacturers nowadays thermal resistance between the internal sensor and the
publish their product specifications including the main probes inside the insert and other sources of uncertainty
topics in the Euramet guide, the products are easier to need to be considered.
compare.
• Main topics in the EURAMET guideline include:
Stem conductance
Sensor to be
calibrated
Reference sensor
Axial uniformity
CALCULATION EXAMPLES