You are on page 1of 7

Nuclear Instruments and Methods in Physics Research A 486 (2002) 67–73

Comparison of the methods for determination of


scintillation light yield
E. Sysoeva*, V. Tarasov, O. Zelenskaya
STC Institute for Single Crystals of the National Academy of Science of Ukraine, 60 Lenin ave., Kharkov 61001, Ukraine

Abstract

One of the most important characteristics of scintillators is the light yield. It depends not only on the properties of
scintillators, but also on the conditions of measurements. Even for widely used crystals, such as alkali halide
scintillators NaI(Tl) and CsI(Tl), light yield data, obtained by various authors, are different. Therefore, it is very
important to choose the convenient method of the light yield measurements. In the present work, methods for the
determination of the physical light yield, based on measurements of pulse amplitude, single-electron pulses and intrinsic
photomultiplier resolution are discussed. These methods have been used for the measurements of light yield of alkali
halide crystals and oxide scintillators. Repeatability and reproducibility of results were determined. All these methods
are rather complicated in use, not for measurements, but for further data processing. Besides that, they demand a
precise determination of photoreceiver’s parameters, as well as determination of light collection coefficient of
scintillation detector. Thereby, such methods should be used only for special cases of certification of reference detectors.
For routine measurements of scintillation light yields, a simple method based on comparison of signals from tested and
certificated detectors, was developed. r 2002 Elsevier Science B.V. All rights reserved.

Keywords: Scintillation detector; Measurement; Light yield; Light output; Resolution

1. Introduction Technical light yield is a ratio of total energy


(Lp ) of scintillation photons, passing through the
The light yield is the main characteristic of window of the detector, to energy (E), deposited
scintillation detectors. One should differentiate the by ionizing radiation in scintillator:
absolute light yield (scintillation efficiency) from Lp
the technical light yield [1]. Absolute light yield (Z) T¼ : ð2Þ
E
is a ratio of total energy (Ep ) of scintillation
photons to energy (E) deposited by ionizing Between technical and absolute light yield exists
radiation in the scintillator: a relationship:
T ¼tZ ð3Þ
Ep
Z¼ : ð1Þ where t ¼ Lp =Ep is a light collection coefficient in
E
the scintillation detector.
*Corresponding author. Tel.: +380-572-308-380; fax: +380-
Absolute light yield is characterized by the
572-320-207. number of photons (Np ) in scintillation per
E-mail address: sysoeva@isc.kharkov.com (E. Sysoeva). 1 MeV energy radiation absorbed by scintillator,
0168-9002/02/$ - see front matter r 2002 Elsevier Science B.V. All rights reserved.
PII: S 0 1 6 8 - 9 0 0 2 ( 0 2 ) 0 0 6 7 6 - 9
68 E. Sysoeva et al. / Nuclear Instruments and Methods in Physics Research A 486 (2002) 67–73

technical light yield by the number of photons e the charge of electron, and e the energy of light
(Npt ), passing through the window of the detector photon.
for the same energy: Since e; E and e values are tabular data, the
E0 error of the light yield (DT=T) measurement is
Np ¼ Z ð4Þ determined by the errors of measuring of signal
e
amplitude (DV =V ), of output capacity (DC=C), of
E0 anode sensibility ðDSa =Sa Þ and of amplification
Npt ¼ T ð5Þ factor ðDK=KÞ:
e sffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffi
 2  2  2  2
where e is the average energy of scintillation DT DV DC DSa DK
photons and E0 is an energy of absorbed radiation, ¼ þ þ þ : ð7Þ
T V C Sa K
equal to 1 MeV.
The problem of measurement of the light yield is An estimation of instrumental error of measure-
quite complicated because of the necessity to ment of values V ; C; Sa and K was made by us.
determine values of parameters of detectors and The obtained values were equal to 0.8%, 5.3%,
used photoreceivers precisely, which introduce big 2.6% and 2.2%, respectively. Instrumental error
errors in the result. Ordinary practice includes for the determination of the light yield was equal
usually measurements of light output by relation- to 6.3%. Reproducibility of measurements (statis-
ship of signals amplitude from tested and reference tical error) was equal to 1.5%, therefore, the total
detectors. Detectors on the base of NaI(Tl) [2], error of the light yield measurement by pulse
antracene [3] or stilbene [4] are used as reference method was not more than 6.5%.
detectors. However, because of the fact that To reduce the error of the light yield measure-
detector’s quality significantly varies from sample ment by pulse method, one can use the photo-
to sample, it is necessary to measure the absolute multiplier as a photodiode, by reading the signal
(or technical) light yield of reference sample and to from the first dynode and by shorting the other
indicate its value under certification of detectors. dynode [5]. In this case, the photomultiplier must
In this work, the analysis of direct methods of be calibrated by a silicon diode, under excitation
measuring of detector light yield was made and the by low-energy gamma-radiation. One can reach
method, based on comparison of signals from the same result by using a photodiode as a
tested and reference detectors, was worked out. photoreceiver with the same calibration [5,6]. In
the paper [6], the instrumental error of light yield
measurements was equal to 5%.
2. Analysis of methods for light yield measuring
2.2. Method based on the measuring of intrinsic
resolution of the photomultiplier
2.1. Pulse method
Another direct method of light yield measure-
Direct method for measurement of the light
ment consists in measuring the number of photo-
yield of scintillation detector consists in measuring
electrons (Npe ) collected on the first dynode of the
of the signal amplitude (V ) at the spectrometer’s
photomultiplier:
output (pulse method). Signal amplitude and light
yield are associated by the relationship Npe
Np ¼ ð8Þ
t  Sa
V eC
T¼ ð6Þ
Sa  K  E  e Npe
Npe ¼ : ð9Þ
where C is the output capacity of the photo- Sa
multiplier, Sa the effective anode sensibility of the It is known [7] that because of the statistical
photomultiplier, K the amplification factor of the nature of the processes taking place in the
measuring path, E the energy of gamma-radiation, photomultiplier, the intrinsic resolution (Rp ) of
E. Sysoeva et al. / Nuclear Instruments and Methods in Physics Research A 486 (2002) 67–73 69

the high-quality photomultiplier is determined by tron signal from the noise. This method is very
the number of photoelectrons collected on the first useful for detectors with very low light output,
dynode: when the resolution of spectrometer is big, or
sffiffiffiffiffiffiffiffiffiffiffi absent at all. In the case of high light yield,
8 ln 2 photomultiplier is connected in photodiode mode,
Rp ¼ : ð10Þ
Npe as stated above (the first method).
For the number of photoelectrons per 1 MeV
energy radiation, absorbed by scintillator, from 2.4. Method of comparison
Eq. (10) we get the relation:
The problem of the determination of light yield
5:544
Npe ¼ 2 : ð11Þ may be facilitated significantly, when comparison
Rp  E of signal amplitudes from tested detector and
One can get intrinsic resolution of the photo- reference detector with known light yield is carried
multiplier from the spectrometer’s resolution (R), out. Light yield (Z; T) and light output are
if intrinsic detector’s resolution (Rd ) is known: associated by these relationships:
qffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffi Z0 L
Rp ¼ R2  R2d : ð12Þ Z¼ ð13Þ
t
Intrinsic detector’s resolution may be measured T ¼ Z0 L ð14Þ
by independent methods: method of homogenous
neutral light filters or method of comparison [8]. where Z0 is the light yield of the reference sample.
It should be noted that this method is especially For measuring the light output, the difference of
convenient for scintillators with low light yield or radiation spectra from tested and reference detec-
for scintillators, designed for low-energy radiation tors and their convolution with spectral sensibility
measurements (for example, for X-ray detectors). of photomultiplier photocathode have to be
In this case, a small signal leads us to neglect the considered. Besides that, it is necessary to make
intrinsic resolution of the detector in comparison corrections for decay time of scintillators.
with the resolution of the photomultiplier. How- Verification scheme for measuring of light
ever, this condition should be revised, for example, output in reference units for detectors of various
by measuring of the spectrometer resolution at too types and sizes was worked out by us [9]. Stilbene,
low energies. If values of the number of photons, as the most stable scintillators, was chosen as a
calculated by Eq. (10), are equal in both cases, reference sample for comparison. However, the
then the intrinsic resolution of detector may be same principal approach may be used for other
neglected. reference scintillators.
Estimation of measuring error of light yield by
photomultiplier’s resolution, performed similarly
to the estimation of error of the first method, 3. Experiment and results
indicated that the value of this error was not more
than 5%. In the present work, light yield of standard
samples on the base of single crystals of stilbene
2.3. Single-electron method and NaI(Tl) were determined experimentally.
Samples with a good spectrometric quality of
The third direct method of measuring the light cylindrical form with dimensions +40  40 mm2
yield consists in comparing the signal amplitude were used. Crystals were packed in an aluminum
from the detector with the average value of the alloy container with magnesium oxide reflector.
amplitude of the single-electron distribution of the Measurements were made using gamma-radia-
photomultiplier. It is needed to use a low-noise tion source 137Cs. Signal amplitude of NaI(Tl)
photomultiplier, allowing picking out single-elec- detector was determined by the position of
70 E. Sysoeva et al. / Nuclear Instruments and Methods in Physics Research A 486 (2002) 67–73

full-absorption peak (0.662 MeV); signal ampli- Thus, the method of comparison of signal
tude of stilbene was determined by the position of amplitude from tested and reference detectors
the edge of Compton distribution (0.477 MeV). can be used for ordinary measurements of the
Effective anode sensibility of photomultiplier was light yield.
determined by a calibrated light-measuring lamp Light yield of the CsI(Tl) crystal with dimen-
SI-10-300 with a strip tungsten fiber. In addition, sions +25  25 mm2 was measured for checking
the reconciliation of spectrum of scintillator the adequacy of this method. The crystal was
radiation with spectral sensibility of photocathode arranged in optical contact on the photocathode of
of the photomultplier was taken into considera- the photomultiplier (type Hamamatsu R1307).
tion. The output capacity of the photomultiplier The crystal was surrounded with Teflon reflector
was determined by droop of pulse of scintillation on the side. A Mylar reflector was placed on the
for NaI(Tl) with oscillograph. Amplification factor crystal top. The resolution of the spectrometer was
of measuring path was measured by using stan- measured on the crystal absorption peak for
dard pulse generator BGA-05F. Intrinsic resolu- gamma-radiations of 137Cs, 60Am, 109Cd and
55
tion of NaI(Tl) detector was determined with the Fe. Intrinsic resolution of the crystal was
help of homogenous neutral light filters for 137Cs determined for gamma-radiation of 137Cs by the
gamma-radiation, then the intrinsic resolution of method of homogenous neutral light filters. The
photomultiplier was calculated for the same energy obtained value of intrinsic resolution for the
by Eq. (11). Intrinsic resolution of photomultiplier crystal (4.5%) was accepted to be the same for
for an energy corresponding to Compton edge was all energies, since energy dependence of intrinsic
calculated by using the ratio of peak and Compton resolution may be neglected for these energies. In
edge energies. Light collection coefficients were Table 2, resolution of the spectrometer (R),
estimated by data [9] for both detectors and resolution of the photomultiplier (Rp ) and number
accepted to be equal to 0.7. of photoelectrons (Npe ), collected at the first
In Table 1, results of determination of absolute dynode of the photomultiplier for an absorbed
light yield in percentage (Z) and in photons per energy of 1 MeV are shown.
MeV (Np ) and of technical light yield in percentage As one can see from Table 2, results of
(T) and in photons per MeV (Npt ) of reference measurements of light yield at different energies
detectors by three methods are shown. of absorbed radiation are in a good coincidence. In
As one can see from Table 1, reference detectors this case, intrinsic resolution of the crystal may be
have a high light yield. Light output of detectors is neglected for measurements with the radiation
determined relative to the standard sample of source 55Fe only. Absolute light yield of CsI(Tl)
stilbene (in reference units) [1]. As one can see crystal was calculated from the data from Table 2
from Table 1, one reference unit of light yield is by Eq. (8). The coefficient of light collection is
equal to technical light yield of 2.3%. If the light equal to 0.7 in this case, and the effective anode
output of the detector in reference units (L) is sensibility is equal to 0.15. The number of photons
known, then it is easy to calculate the light yield by in the crystal per 1 MeV energy absorbed radiation
using Eqs. (4), (5) and (13), (14). is equal to 42 830.

Table 1
Light yield of the reference detectors

Method Stilbene NaI(Tl)

Z (%) Np (ph/MeV) T (%) Npt (ph/MeV) Z (%) Np (ph/MeV) T (%) Npt (ph/MeV)

Pulse 3.3 10,300 2.3 7200 14.4 48,000 10.1 33,700


By Rp 3.4 10,600 2.4 7500 16.1 53,700 11.3 37,700
Single electron 3.1 9700 2.2 6900 15.0 50,000 10.5 34,000
Average values 3.370.2 10,2007500 2.370.1 72007300 15.270.8 50,60073100 10.670.7 35,10072600
E. Sysoeva et al. / Nuclear Instruments and Methods in Physics Research A 486 (2002) 67–73 71

The light output of this CsI(Tl) crystal was energy. Difference of results, obtained by method
determined by the method of comparison of of comparison and by method using photomulti-
signals from tested and reference detectors also. plier resolution is equal to 3.4%, which is quite
The light output was determined by certified satisfactory.
reference NaI(Tl) detector taking into considera- As an example of applicability of these methods
tion the agreement of radiation spectra of crystals for the determination of the light yield for crystals
and spectral sensibility of photomultiplier, and with very low light yield, results of light output
time of signal collection on the photomultiplier measuring for CsI crystals with dimensions
anode. Relative light yield of the CsI(Tl) crystal 25  25  130 mm3 and PWO with dimensions
was equal to 3.1 reference units. The value of light 20  20  40 mm3 were adduced. Crystals were
output was calculated from Eqs. (4) and (13). It is surrounded with white reflector on the sides and
equal to 44 300 photons per MeV of absorbed at the top and were placed on the photomultiplier
photocathode PMT R1307 with optical contact.
The resolution of the spectrometer was measured
at absorption by CsI crystal of gamma-radiation
Table 2
Resolution and light yield of CsI(Tl) crystal
of 137Cs, 60Am and 109Cd. Because of the fact that
the light yield of PWO is very low, resolution was
Source Energy (MeV) R (%) Rp (%) Npe (pe/MeV) determined only at absorption of 137Cs gamma-
137
Cs 0.662 6.2 4.3 4529 radiation. Amplitude distributions of pulses from
60
Am 0.060 15.1 14.4 4456 PWO and single-electron pulse of photomultiplier
109
Cd 0.022 24.5 23.7 4486 R1307 are drawn in Fig. 1. Pulse amplitude ratio
55
Fe 0.006 45.8 45.6 4516
from PWO and single-electron pulse is equal to 13,
Average value 4500750
which corresponds to 19.6 photoelectrons per

137
Fig. 1. Amplitude spectrum of single-electron pulse of photomultiplier R1307 and gamma-radiation Cs absorbed in the PWO
crystal.
72 E. Sysoeva et al. / Nuclear Instruments and Methods in Physics Research A 486 (2002) 67–73

MeV of absorbed energy. It should be noted that As one can see from Table 4, results obtained by
for such low energies the ‘‘peak-to-valley’’ ratio all the three methods are in a quite fair agreement.
for full absorption peak is less than two, so the
resolution was determined from the right half of
the peak. The same spectrum was obtained for CsI 4. Conclusion
crystal at 109Cd radiation. In this case, signals ratio
is equal to 4.5, which corresponds to 205 photo- Thus, analysis of the three methods for the
electrons per MeV of absorbed energy. In Table 3, determination of light yield of scintillation detec-
resolution of the spectrometer (R) and number the tors: pulse method, single-electron method and
of photoelectrons (Npe ), collected at the first method using photomultiplier resolution was
dynode of the photomultiplier at absorption of carried out. It was shown that the error of
the energy equal to 1 MeV, are shown. determination of the light yield by these methods
As one can see from Table 3, in this case (when was up to 6.5%.
signal from the crystal is weak) the resolution of Method of determination of the light yield by
the spectrometer is almost equal to the photo- comparison of pulse amplitude of tested detector
multiplier resolution. Thus, the light yield of CsI and reference sample, certified with a given
and PWO crystals was determined by using three accuracy, was worked out. This method provides
methods: by spectrometer resolution, single-elec- results, comparable with results obtained by direct
tron peak and by method of comparison. The methods, and may be applicable for certification of
estimated values of effective anode sensibility were scintillation production.
equal to 0.15 and 0.25 for CsI and PWO,
respectively. The light output in reference units
was equal to 0.16 and 0.01 for CsI and PWO, Acknowledgements
respectively. Results of light output determination
for CsI and PWO in photoelectrons (Npe ) and in Authors express their thanks to Dr. A. Kudin
photon (Np ) per MeV of absorbed energy are and Dr. M. Kosmyna for CsI and PWO crystals,
shown in Table 4. used in the experiments.

Table 3 References
Resolution and light yield of CsI crystal
[1] Yu.A. Tsirlin, M.E. Globys, E.P. Sysoyeva, Optimizatsia
Source Energy (MeV) R (%) Npe (pe/MeV) detectirovania gamma-izlucheniya scntillyatzion-numi
137 kristallami, Energoatomizdat, Moskow, 1991, p.153 (in
Cs 0.662 21.2 186
60
Am 0.060 71.5 181 Russian).
109
Cd 0.022 112.5 199 [2] Scintillation Detectors. Crismatec, Catalog, Saint-Gobain
ceramiques industrielles, March 1992.
Average value 189710
[3] Catalog Bicron, 1993.

Table 4
Light yield of CsI and PWO crystals

Method CsI PWO

Npe (pe/MeV) Np (ph/MeV) Npe (pe/MeV) Np (ph/MeV)

By Rp 189 2362 18.5 297


Single electron 205 2562 19.6 314
By comparison — 2300 — 303
Average value — 24107150 — 305710
E. Sysoeva et al. / Nuclear Instruments and Methods in Physics Research A 486 (2002) 67–73 73

[4] GOST 23077-78 Detectory ioniziruyushchikh izlucytniy [8] A.Ya. Berlovskiy, E.P. Mokhir, Yu.A. Tsirlin, V.A. Shvets,
scintillyatsionnye. Terminy, opredeleniya i bukvennye Patent 240118 USSR, Byulleten’ izobreteniy, 30, 1969.
oboznacheniyap, Moskow, 1978. [9] M.E. Globus, B.V. Grinev, Neorganicheskie scintillyatory.
[5] E. Sakai, IEEE Trans. Nucl. Sci. NS-34 (1987) 418. Novye i traditsionnye materialy, Khakov, Acta, 2000, p.
[6] I. Holl, E. Lorentz, G. Mageras, IEEE Trans. Nucl. Sci. 400.
NS-35 (1988) 105.
[7] J.B. Birks, The Theory and Practice of Scintillation
Counting, Pergamon Press, Oxford, 1964, p. 513.

You might also like