Professional Documents
Culture Documents
Gas Processors
Association
GPA publications necessarily address problems of a general nature and may be used by anyone desiring to do so.
Every effort has been made by GPA to assure accuracy and reliability of the information contained in its
publications. With respect to particular circumstances, local, state, and federal laws and regulations should be
reviewed. It is not the intent of GPA to assume the duties of employers, manufacturers, or suppliers to warn and
properly train employees, or others exposed, concerning health and safety risks or precautions.
GPA makes no representation, warranty, or guarantee in connection with this publication and hereby expressly
disclaims any liability or responsibility for loss or damage resulting from its use or for the violation of any federal,
state, or municipal regulation with which this publication may conflict, or for any infringement of letters of patent
regarding apparatus, equipment, or method so covered.
FOREWARD
GPA 2261 provides the gas processing industry a method for determining the chemical composition of natural gas
and similar gaseous mixtures using a Gas Chromatograph (GC).
The precision statements contained in this standard are based on the statistical analysis of round-robin laboratory
data obtained by Section B.
This standard was developed by the cooperative efforts of many individuals from industry under the sponsorship of
GPA Section B, Analysis and Test Methods.
Throughout this publication, the latest appropriate GPA Standards are referenced
©
“Copyright 2013 by Gas Processors Association. All rights reserved. No part of this Report may
be reproduced without the written consent of the Gas Processors Association.”
Analysis for Natural Gas and Similar Gaseous Mixtures by Gas Chromatography
> 30
SCOPE
40 - 100
The precision statement in this standard utilizes equations
derived from a regression of the data in RR-188 and is
detailed in GPA TP-31. The precision statement criterion N/A
applies only to values listed in Section 10, Table 6. n-Butane
0.01 - 0.25
1.2 Components sometimes associated with natural gases,
i.e., helium, hydrogen sulfide, water, carbon monoxide, 0.25 - 4
hydrogen and other compounds are excluded from the main >4
body of the method. These components may be determined
and made a part of the complete compositional data. Refer
to Appendix A.
0.1 - 10
> 10
Lower
Round
Isopentane
Higher
0.01 - 0.12
Component
0.12 - 1.5
Region
Robin > 1.5
Region
Propane
Nitrogen
0.01 - 0.1
0.01 - 0.1
0.1 - 10
0.1 - 30
> 30
> 10
Carbon Dioxide
3.2 Carrier Gas - The contaminants in the carrier gas Figure 1B Chromatogram of early backflush of
must be limited to levels that are known not to interfere heptanes and heavier (C7+).
with the analysis or cause maintenance problems with
the GC. Refer to manufacturer for recommendations 3.3 Sample Conditioning Systems - GPA 2166 gives
regarding carrier gas quality guidance for proper design and use of sample
conditioning systems. The sample conditioning system
3.2.1 Pressure and Flow Control Devices - These should not cause the GC precision to fall outside the
devices should maintain flow rate consistent enough to requirements in Section 10.
provide repeatable peak retention times and
compositional precision within the limits described in NOTE 4 – Valves and sample introduction system must
Section 10 during the reference standard and be maintained at a temperature above the hydrocarbon
corresponding sample runs. Two Stage regulators with dew point of the calibration blend and unknown
stainless steel diaphragms have been shown to be samples. Supplemental heating may be required to
satisfactory for this purpose.
accomplish this. Refer to GPA 2166 for guidance. Integration systems can not correct for inadequate
component separation. The integration system should
3.4 Integration System - The integration system should not cause the GC precision to fall outside the
be configured to properly integrate all peaks of interest. requirements in Section 10.
2
Control Charts - GPA 2198 describes the use of Control
Charts. Control Charts can be used to monitor each
component in the calibration blend and the GC
performance over time.
SAMPLE INTRODUCTION
ANALYTICAL PROCEDURE
1098 0.57
Ethane Table V Calculation of Concentration
0.0002641 21470 in Unknown Sample
0.0004821
10.35
Hexanes Plus Using Response Factors
0.18
10.34
Component
Propane
Area
749 13623 Resp. Factor
0.0003843 Unnorm.
5.24
0.0002403
Norm. Mole
5.23
Isobutane
100.00 2253
0.0003275
0.74
0.74 Mole %
n-Butane %
4324
0.0003193 Nitrogen
1.38
Table III Calculation of 1050
Molar Concentration in 1.38 0.0006465
Isopentane 0.68
1398 0.68
0.0002849
Unknown Sample Using 0.40 Methane
Response Factors
0.40 95153
n-Pentane 0.0008398
1321 79.91
0.0002641 79.77
0.35
Carbon Dioxide
0.35
Hexanes Plus
1255
2080 0.0004514
0.0002403 0.57
0.50 0.57
0.50 Ethane
21470
0.0004821
10.35
10.33
Unnorm.
Norm. 100.12
Component Hydrogen Sulfide
Area 100.00
Resp. Factor
Mole % where:
Mole % 0.05
Nitrogen
1050 Fnorm = normalization factor
0.05
0.0006465
0.68
Σunnorm = unnormalized total Propane
Isobutane
2253
0.0003275
0.74
0.74
n-Butane
4324
0.0003193
1.38
1.38
Isopentane
1398
0.0002849
0.40
0.40
n-Pentane
1321
0.0002641
0.35
0.35
Hexanes Plus
2080
0.0002403
0.50
0.50
100.12
100.00
9.3 Reporting is
commonly to two
decimal places due to
limitations on equipment.
TCD detectors typically
have a linear dynamic
range of 10,000:1.
Numbers are calculated
to three decimal places
and then rounded up
when the third digit is 5
or higher.
5
PRECISION Iso-pentane
.02 – 4
1/4
The repeatability and reproducibility statements for this 0.009x
1/4
standard are from the statistical data obtained in a GPA 0.025x
RR-188. The testing program included ten samples N-pentane
comprised of ten components analyzed by six laboratories. .02 – 4
1/5
The standard as revised has been statistically evaluated 0.01x
1/3
under ISO and ASTM protocols. The documentation of the 0.026x
statistical evaluation may be found in GPA TP-31. Hexanes Plus
.02 – 2
1/4
To determine the precision for any component at a 0.0135x
1/2
specific concentration, use the formulae shown in Table 0.051x
6 and substitute the mole percent of the component for
x. The following example calculations show the repeatability
and reproducibility for two different blends. The Ranges
Repeatability is the expected precision within a from the previous precision statement are used in the two
laboratory using the same equipment and same analyst. examples. Example 1 lists the lower concentration from the
Repeatability is the difference in analyzed values original precision statement range of each component and
between two sequential runs. Reproducibility is the Example 2 lists the higher concentration from the original
expected precision when the same method is used by precision statement range for each component along with
different laboratories using different equipment and the repeatability and reproducibility calculated for those
different analysts. Reproducibility is the difference values.
between two analyzed values. Neither value represents
the difference between an analyzed value and the The values shown in these calculations are in mol percent.
certified value listed on a blend. (Refer to 10.6 and These values are the mol % of the component plus or
10.7). minus the value determined from the appropriate equation.
That is to say, if the value is 1.00 and the precision value is
Table VI Component Ranges for Precision Limits 0.02, results that are between
and 1.02 are acceptable and values that are above or below
Range that range are not acceptable and fail to meet the precision
Repeatability criteria of this standard. When the result is less than 0.01,
Reproducibility use 0.01 as the lowest precision value.
Nitrogen
.02 – 15
0.039x
1/4 Example 1
1/2
0.158x
Methane
50 – 100 Mol %
1/3 Repeatability
0.0079x
-3 Reproducibility
91000x
Nitrogen
CO2
1
.02 – 15
1/3 0.04
0.0042x 0.16
1/3
0.12x Methane
Ethane 71.6
.02 – 15 0.03
1/3 0.25
0.0124x
0.0315x
1/3 CO2
Propane
0.14
.02 – 15 0.01
1/8 0.06
0.0084x
1/2
Ethane
0.026x 4.9
Iso-butane 0.02
.02 – 8 0.05
1/5
0.01x Propane
0.018x
1/2 2.3
N-butane 0.01
.02 – 8 0.04
2/5 Iso-butane
0.0117x
1/2 0.26
0.033x 0.01
0.01 0.01
N-butane 0.05
0.6 Iso-butane
0.01 1
0.03 0.01
Iso-pentane 0.02
0.12 N-butane
0.01 1.9
0.01 0.02
N-pentane 0.05
0.14 Iso-pentane
0.01 0.45
0.01 0.01
Hexanes Plus 0.02
0.1 N-pentane
0.01 0.42
0.02 0.01
0.02
Hexanes Plus
0.35
Example 2 0.01
0.03
Mol %
Repeatability Performance evaluations commonly use the repeatability
Reproducibility
Nitrogen
and reproducibility of laboratory results compared to a
7.7 certified blend. This precision statement is based on the
0.06 data contained in GPA RR-188 and the statistical
0.44 evaluation described in GPA TP-31. This treatment of data
Methane compared laboratory results independent of the certified
86.4 blend values. Therefore, performance evaluations must
0.03 either compare the laboratory results in the same manner
0.14 by using the reproducibility values described in Table 6 and
CO2
subsequent example calculations, or use the Performance
7.9
0.01 Evaluation Acceptance Criteria listed below.
0.24
Ethane The ability of an instrument to match the certified values
9.7 from a gravimetric blend referred to as is the Performance
0.03 Evaluation Acceptance Criteria. The blend uncertainty
0.07 must be known to use this approach. The reproducibility
Propane
and the uncertainty of the calibration blend are used to
4.3
determine the Performance Evaluation Acceptance Criteria.
6
Where:
7.70
0.077
CVB is the certified value of component in blend PE is the 0.44
Acceptance Criteria for component 0.45
Iso-pentane
0.45
0.0045
0.020
0.02
N-pentane
0.42
0.0042
0.020
0.02
Hexanes
Plus
0.35
0.0035
0.030
0.03
DEFINITIONS
8
APENDIX A - Calibrations
(calibratio desired
n curve) or component
secondary should be
analysis. made. If the
A-1 values on
Linearity duplicate
A-1.1
Calibratio runs agree
Section n Curves within the
4, (Multi- tolerances
Appendix level in Section
C and Calibratio 10,
GPA n) “Repeatabil
2198 ity”, the
discuss response
Linearity A-1.1.1 factor
and list Calibratio should be
procedur n Curves calculated
es to Using
as follows
determin Multiple for each
e the Calibratio
n Blends concentrati
linear on level:
range and
calibratio Once
n linearity K= Cn An
requirem has been where
ents of determined
GC for a GC, K=
systems. as Response
When it described factor
is in Section Cn =
anticipate 4, and the Concentrati
d that the linear on of
range of range is component
concentra found to be n
tions of inadequate An = Peak
compone for the area in
nts in the range of arbitrary
unknown unknown units of
samples sample component
will not concentrati n
fall in the ons
linear anticipated
Calibration
calibrated ,
curves may
range of calibration
curves for now be
the
developed
instrume any
by plotting
nt, it is component
response
necessary may be
factors
to make determined
versus
correctio by using
multiple concentrati
ns for
on. Any
this. Two calibration
program
means of blends.
capable of
accompli
generating
shing this Duplicate
a
are injections
polynomial
through of at least
curve fit
multi- three
may be
level concentrati
used.
calibratio on levels
n for the
A-1.1.2 least three area of pure
Calibrati partial component
on pressures in arbitrary
Curves in units
Using duplicate
Partial and Calibration
Pressures capture curves can
of Pure data now be
Compone including developed
nts Barometric by plotting
Pressure at response
Once the time of factors
linearity the versus
has been injection. concentrati
establishe When on. Most
d for the concentrati integration
instrume ons on software
nt as duplicate packages
described runs meet have this
in the criteria feature
Section listed in built-in, but
4, Section 10, if this
calibratio “Repeatabi feature is
n curves lity”, not
for any calculate available,
compone the other
nt to be response programs
measured factor as capable of
in the follows: generating
unknown a
sample K= (Pi) polynomial
may be (100) curve fit
determin may be
ed by (Po) (A) used.
using
pure Table A-1
compone
nts. Where
Component
Partial
Attach K = Barometric
Pi/Po * 100
the pure Response
compone factor Pressure mm
nt to the Pressure mm
sample Pi = Partial of
entry pressure of
system pure
and component of Hg (Pi)
evacuate in mm of Hg (Po)
the entry mercury to
system to nearest 0.5 Oxygen
less than mm 100
750
1 mm of 13.33
mercury. Po = Nitrogen
Using the Barometric 100
partial pressure in 750
13.33
pressure mm of Methane
range mercury to 500
suggested nearest 0.5 750
in Table mm 66.67
Carbon
A-l, 100
inject at A = Peak 750
13.33 response analysis or
Monoxide factors
run. When
caused by
compressibili more than
ty of the pure one
component. component
Carbon is
Dioxide A-1.2 determined,
100 (650)*
Secondary add all
750
13.33 Analyses component
Ethane concentrati
200 (450)* ons and
750
Secondary
26.67 analyses normalize
Propane may be to 100%.
100 (200)* used When a
750 single
13.33
instead of
Isobutane calibration component
100 (100)* curves (as is
750 in the case determined,
13.33 it is
n-Butane
of Carbon
100 (100)* Dioxide on acceptable
750 a Porous to keep that
13.33 Polymer component
Isopentane concentrati
50 (50)*
column.)
750 or for on whole as
6.67 determinati described
n-Pentane on of below:
50 (50)*
750
compound
6.67 s not Fnorm= 100-
determined
Cn
*Partial by the
Pressures partition
in column 100
parentheses
are the run. The
maximum secondary Where
pressures to analysis or Fnorm =
be used to run may
determine
Normalizati
response occur on Factor
factors. separately Cn =
Exceeding or
these Concentrati
simultaneo on of
pressures
could result usly to the component
in low primary n
9
All components determined in the primary analyses or run
are then multiplied by N, and the single component held
whole.
10
APENDIX B – Linearity Discussion
Mole % linear:
An ideal GC
detector will
Methane In this
provide a linear ÷ Peak example,
response across
all sample Calibrating to
component
Area =
80 Mole %
concentrations. 80 ÷ would yield
In this case, a 80,000 = the following
calibration response
standard with 0.001 factor:
any
concentration When the
of the detector was KFMethane
component of perfectly = Mole
interest could linear, and an
be run and a unknown %
calibration sample was Methane
response factor run and
could be generated a
÷ Peak
determined: peak area for
Methane of Area
40,000, the
Example: A
Methane
Methane
calibration
standard has 80 concentration = 80 ÷
Mole % in Mole % 96,000 =
Methane. Whenwould be: 0.000833
the sample is
run on the GC, However,
it generated a many Using the
peak area of chromatograp above
calibration
80,000. The h detectors
factor to
response factor are not linear analyze a
for Methane at in their sample with
80% response. A 40% Methane
concentration graph of would yield
is: Mole % the following
concentration result:
to peak area
Kmethane = would not be
Met straig
ht
Peak han line):
Area e 100000
x Line
60000
A ar
Kmeth graph Meth
of ane
ane = Mole Resp
40,00 % onse
0 x conce
ntratio
0.001 n to 120000 Area
= peak 80000 Peak
area
would
40 be
Mole linear
(a
%
53.33 The
linearit
3 y check
Mole is used
to
% determi
40000 ne the
Meth number
of
100000 0 ane calibrat
ion
Since standar
the ds that
differen are
ce needed
between to
the analyze
20000
actual all the
Area
80000
value expecte
0 (40%)
20
d
40
and the sample
60 reported compos
80 value itions.
100 from
120
calibrati
When
on to
more
80%
Mole% Methan than
0
Metha one
e
60000
ne calibrat
(53.333
ion is
%)
Mol exceeds require
d, this
e% the can be
reprodu
Met cibility achieve
d by
han limits having
e = establis
hed in
a
0 separat
20 Peak Pea Section
e
40 k 10 it calibrat
60
80 Are would
be
ion
100 method
120
a necessa for
Met ry to each
have
han separate expecte
d
e * calibrati sample
Mole%
Methane KF on standar
compos
ition, a
40000 Meth ds for multi-
samples
level
Non - ane containi calibrat
Linear ng 80%
ion for
Methan Methan
e = e and all
Respon compo
se
64,0 samples nents
00 * containi
ng 40%
or a
multi-
0.00 Methan level
120000
20000
083 e. calibrat
3= ion for
the it can establiMole% various
compone be shed is partial
nts that analyz by plotted pressur
are not ed runni against es. In
linear. within ng the peak this
the multiparea. case,
Whether repro le the
a GC ducibi calibr Linearit Mole%
detector lity ation y curves value is
is linear limits standa can also determi
for a outlin rds of be ned by
compone ed in variou establis the
nt or not Sectio s hed by followi
linear for n 10. comp running ng
a ositio the formul
compone Linear ns. In same a (see
nt is ity this calibrati Appen
determin curve case, on dix A,
ed by s can the standar A-
whether be actual d under 1.1.2):
11
Partial Pressure Mole % Equivalent =
Where:
12
APENDIX C – Supplementary Procedures
sieve column. precision of
C-1 Run Analysis for However, measurement
Nitrogen, Methane, should a gas will be poor
Carbon Dioxide, blend not be under these
and Ethane available, a conditions due
The porous polymer calibration to similar
column must curve should be thermal
completely separate developed using conductivity
methane, carbon pure carbon values for
dioxide and ethane monoxide to hydrogen and
to baseline as shown determine the helium. If a
in the example extent of the calibration gas
chromatogram. The nonlinearity, if blend is
linearity of this present. available
system must be containing
determined to be C-3 hydrogen and
linear to be an Determination
acceptable of Hydrogen helium, it
alternative to the and Helium should be used
calibration curve to obtain
technique described When hydrogen response factors
in Appendix B. This is to be however, if this
system can be used separated from is not the case,
as part of a multi- helium, a 20’ the pure
column GC, as in the molecular sieve components,
case of some 5A column hydrogen and
portable GC’s. using nitrogen helium, may be
or argon as a used to develop
C-2. Determination carrier gas may response factors
of Carbon Monoxide be used. Low in the manner
temperature, set forth in
This component is 40°C (104°F) or Appendix A, A-
encountered in less is necessary 1.
association with to effect this
oxygen, nitrogen, separation. C-4.
carbon dioxide and Determination
the conventional When hydrogen of Hydrogen
hydrocarbons in the is present, it will Sulfide
effluent streams elute on the
from combustion standard As indicated
processes such as molecular sieve earlier in this
insitu combustion, run, using text, to be
manufactured gas helium as a absolutely sure
and many varied carrier gas, just of the hydrogen
types of stack gases. before oxygen. sulfide content of
No extra equipment The hydrogen a gas,
is necessary to response is determinations
should be made
determine carbon downscale
at the sample
monoxide since it (negative) rather source. However,
elutes shortly after than upscale in the case where
methane on the (positive). a field
molecular sieve run. Signal polarity measurement has
If a calibration gas is must be not been made
available containing reversed for the and although
carbon monoxide, hydrogen peak corrosion of the
obtain a response to be recorded sample bottle
factor as for methane upscale. The may have
on the molecular sensitivity and resulted in some
loss of hydrogen determining the protection
sulfide, a hydrocarbons in equipment is
measurement of the a natural gas 10 ppm for an
in-place component analysis. eight hour
may be made by gas Hydrogen sulfide working
chromatography. It is elutes between period. When
necessary to charge a ethane and the exposure
sample of pure propane with lasts through
hydrogen sulfide to good resolution.
the working
the column prior to
day,
charging the unknown CAUTION -
gas. As soon as the Extreme care concentrations
pure hydrogen sulfide must be taken as low as 15
has cleared the ppm may cause
when working severe
column, the unknown
with hydrogen irritation to the
gas should be
charged. (All sulfide due to eyes and
calibrations should be the very toxic respiratory
done the same way, nature of the tract. Exposure
that is, each partial gas. The best of 800 to 1,000
pressure charge of ventilation ppm may be
pure hydrogen sulfide possible must fatal in a few
must be preceded by a be maintained minutes. The
full sample loop of in the nose must not
pure hydrogen laboratory. The be depended
sulfide.) A column Maximum upon to detect
that has proved Allowable the presence of
satisfactory for this Concentration
hydrogen
type of analysis is the that a person
sulfide, as 2-15
Silicone 200/500 may be
column. It is most minutes of
exposed to exposure will
convenient since this without
is the recommended cause the loss
approved of smell.
column for
respiratory
13