General IR400
Specifications NDIR TYPE INFRARED GAS ANALYZER
(5-COMPONENT ANALYZER)
GS 11G02N01-01E
All Rights Reserved. Copyright © 2002, Yokogawa Electric Corporation GS 11G02N01-01E 4th Edition Oct.02.2006-01
<<Contents>> <<Index>> 3
Remote; Non-voltage contact input (for measurable Settable 60 to 900 seconds (in increments
components) of 1 second)
Allows range to switch via an external Upper/lower limit alarm:
signal when remote range changeover Alarm contact output turns on when
input is received. measurement value reach the preset
Range identification signal: upper or lower limit alarm value.
The present measuring range is identified Contacts close when the instantaneous
by a contact signal. value of each component becomes larger
The contact output terminals for each than the upper alarm limit value or smaller
component are short-circuited when the than the lower alarm limit value.
first range is selected, and when the Instrument error contact output:
second range is selected, the terminals are Contacts close at occurrence of analyzer
open. error No. 1, 3 or 10.
Auto calibration: Calibration error contact output:
Auto calibration is carried out periodically Contacts close at occurrence of manual or
at the preset cycle. auto calibration error (any of errors No. 4
When a standard gas cylinder for calibra- to 9).
tion and a solenoid valve for opening/ Auto calibration status contact output:
closing the gas flow line are prepared Contacts close during auto calibration.
externally by the customer, calibration will Pump ON/OFF contact output:
be carried out with the solenoid valve drive During measurement, contacts close.
contacts for zero calibration and each span While calibration gas is flowing, contacts
calibration turned on/off sequentially at the open. Contacts are connected in power
set auto calibration timing. supply of pump, and stop the sample gas
Auto calibration cycle setting; while calibration gas is flowing.
Auto calibration cycle is set.
Setting is variable within 1 to 99 hours (in Optional Functions
increments of 1 hour) or 1 to 40 days (in O2 correction: Conversion of measured NO, SO2 and CO
increments of 1 day). gas concentrations into values at reference
Gas flow time setting; O2 concentration.
The time for flowing each calibration gas in 21 – On
auto calibration is set. Correction formula: C = Cs
21 – Os
Settable within 60 to 900 seconds (in
increments of 1 second) Where:
Auto calibration remote start: C: Sample gas concentration after O2
Auto calibration is carried out only once correction
according to an external input signal. Cs: Measured concentration of sample
Calibration sequence is settable in the gas
same way as the general auto calibration. Os: Measured O2 concentration (limit
Auto calibration is started by opening the setting: 1 to 20% O2, default 17% )
auto calibration remote start input On: Reference O2 concentration (value
terminals after short-circuiting for 1.5 changeable by setting: 0 to 19% O2,
seconds or longer. default 4%)
Auto zero calibration: The result of calculation is indicated and
Auto zero calibration is carried out output in an analog output signal.
periodically at the preset cycle. Average value after O2 correction and O2 average value
This cycle is independent of "Auto calculation:
calibration" cycle. The result of O2 correction or instanta
When zero calibration gas and solenoid neous O2 value can be outputted as an
valve for opening/closing the calibration average value in the determined period of
gas flow line are prepared externally by the time.
customer, zero calibration will be carried Used for averaging is the moving average
out with the solenoid valve drive contact method in which sampling is carried out at
for zero calibration turned on/off at the set intervals of 30 seconds.
auto zero calibration timing.
(Output is updated every 30 seconds. It is
Auto zero calibration cycle setting;
the average value in the determined period
Auto zero calibration cycle is set.
of time just before the latest updating.)
Setting is variable within 1 to 99 hours (in
Averaging time is settable within 1 to 59
increments of 1 hour) or 1 to 40 days (in
minutes (in increments of 1 minute) or 1 to
increments of 1 day)
Gas flow time setting; 4 hours (in increments of 1 hour).
The time for flowing zero gas in auto zero
calibration is set.
All Rights Reserved. Copyright © 2002, Yokogawa Electric Corporation GS 11G02N01-01E 4th Edition Aug.27.2007-01
<<Contents>> <<Index>> 4
Average value reset: Standard Requirements for Sample Gas
The above-mentioned output of average
Flow rate: 0.5 ±0.2 L/min
value is started from the initial state by
Temperature: 0 to 50˚C
opening the average value resetting input
Pressure: 10 kPa or less (Gas outlet side should be
terminals after short-circuiting for 1.5
open to the atmospheric air.)
seconds or longer.
Dust: 100 µg/Nm3 or less in particle size of 1 µm
Output is reset by short-circuiting and
or less
restarted by opening.
Mist: Unallowable
CO concentration peak count alarm:
Moisture: Below a level where saturation occurs at
(available only for CO + O2 measurement)
2˚C (condensation unallowable).
Alarm output turns on according to the
Corrosive component: HCl 1 ppm or less
preset concentration and count.
Standard gas for calibration:
Whenever the instantaneous value of CO
Zero gas; Dry N2
exceeds the preset concentration value,
Span gas; Each sample gas having
count increments. If the count exceeds the
concentration 90 to 100% of its
preset value in one hour, the alarm
measuring range (recom-
contacts close.
mended).
Communication function:
Gas beyond concentration
RS-232C (9 pins D-sub)
100%FS is unusable.
Half-duplex bit serial
In case a zirconia O2 analyzer is installed
Start-stop synchronization
externally and calibration is carried out on
ModbusTM protocol
the same calibration gas line:
Contents: Read/write parameters
Zero gas; Dry air or atmospheric air
Read measurement concentra-
(provided without CO2 sensor)
tion and instrument status
Span gas; For other than O2 measure-
Remark: When connecting via RS-485
ment, each sample gas having
interface, a RS-232C RS-
concentration 90 to 100% of its
485 converter should be used.
measuring range.
For O2 measurement, O2 gas of
Performance 1 to 2 vol%.
Repeatability: 0.5% of full scale (1% of full scale for Installation Requirements
range less than 0-50 ppm)
• Indoor use: Avoid exposure to direct sunlight, weather,
Linearity: 1% of full scale
and radiant heat from hot substances. Where exposure
Zero drift: 1% of full scale/week
to such conditions are unavoidable, a protective hood or
(2% of full scale/week for range between
cover should be prepared.
0-50 ppm and 0-200 ppm)
• Minimal vibration
(2% of full scale/day for range less than
• A clean atmosphere
0-50 ppm)
Span drift: 2% of full scale/week
(2% of full scale/day for range less than
0-50 ppm)
Response time (for 90%FS response) :
Within 60 seconds including replacement
time of sample gas (when gas flow rate is
0.5 L/min). Gas replacement time depends
on the number of measuring components,
and measuring range effects of interfering
gases.
Effects of interfering gases
When sample gas contains gas components listed below,
the measurement accuracy may suffer. Consult Yokogawa
for countermeasures or effect on accuracy.
All Rights Reserved. Copyright © 2002, Yokogawa Electric Corporation GS 11G02N01-01E 4th Edition Jun.20.2006-00
<<Contents>> <<Index>> 5
Diagram of measurement principle of infrared gas analyzer (NO, SO2, CO2, CO, CH4)
Rotary sector
Sector
sync. signal Infrared (single) Sample cell
light source Sample gas Sample gas
inlet outlet
Gain changeover
AC amplifier
O2 sensor
(optional)
A/D
RAM
CPU
ROM
External input / Contact output D/A Indicator
Indication
Operating keys
Output
0 to 1 V DC or
4 to 20 mA DC F01.EPS
Permanent Mirror
magnets
Magnetic field
Gas outlet
Preamplifier
Permanent
magnets
LED Photodiode
Signal processing
Indication
and calculation block
Output F02.EPS
All Rights Reserved. Copyright © 2002, Yokogawa Electric Corporation GS 11G02N01-01E 4th Edition Jun.20.2006-00
<<Contents>> <<Index>> 6
Example of gas sampling system configuration
The following illustrates a typical system configuration for five component gas measurement for monitoring combustion
exhaust gas from boiler, refuse incinerator, etc. Contact Yokogawa for system configuration matching the particular use or
further information.
1 Probe IR400
7 Membrane filter
2 Mist filter
CO2, CO
Gas inlet tube (O2)
4 Pump
15˚ or larger 8 Flowmeter
Infrared gas
NO2/NO
11 analyzer main
converter unit
for NOx
5 measurement
Electric
gas cooler 7 Membrane
filter NO, SO2
All Rights Reserved. Copyright © 2002, Yokogawa Electric Corporation GS 11G02N01-01E 4th Edition Jun.20.2006-00
<<Contents>> <<Index>> 7
MODEL AND SUFFIX CODES
[Style:S3]
Model Suffix code Option code Description
IR400 Infrared gas analyzer 19-inch rack mounting type with slide rail
Measurable 1st 2nd 3rd 4th
component -A NO
(note 8) -B SO2
-C CO2
-D CO
-F CH4
-G NO SO2
-H NO CO
-J CO2 CO
-K NO SO2 CO
-L NO SO2 CO2 CO
O2 Analyzer N Without O2 analyzer
1 External zirconia type O2 sensor (purchase separately: ZX8D)
(note 10)
2 External O2 analyzer (note 1)
3 Built-in paramagnetic type O2 sensor
1st Component
V 020 ppm (note 3)
1st Range
A 050 ppm
(note 2)
B 0100 ppm
C 0200 ppm
D 0250 ppm
W 0300 ppm
E 0500 ppm
F 01000 ppm
G 02000 ppm
H 05000 ppm
J 01%
K 02%
L 03%
M 05%
P 010%
Q 020%
R 040%
S 050%
T 070%
U 0100%
1st Component A 050 ppm
2nd Range B 0100 ppm
(note 2) C 0200 ppm
D 0250 ppm
W 0300 ppm
E 0500 ppm
F 01000 ppm
G 02000 ppm
H 05000 ppm
J 01%
K 02%
L 03%
M 05%
P 010%
Q 020%
R 040%
S 050%
T 070%
U 0100%
N Not available
2nd Component A 050 ppm
1st Range B 0100 ppm
(note 2) C 0200 ppm
D 0250 ppm
W 0300 ppm
E 0500 ppm
F 01000 ppm
G 02000 ppm
H 05000 ppm
J 01%
K 02%
L 03%
M 05%
P 010%
Q 020%
R 040%
S 050%
T 070%
U 0100%
N Not available
To be continued.
All Rights Reserved. Copyright © 2002, Yokogawa Electric Corporation GS 11G02N01-01E 4th Edition Mar.09.2012-01
<<Contents>> <<Index>> 8
MODEL AND SUFFIX CODES
Model Suffix code Option code Description
IR400 Infrared gas analyzer 19-inch rack mounting type with slide rail
2nd Component B 0100 ppm
2nd Range C 0200 ppm
(note 2) D 0250 ppm
W 0300 ppm
E 0500 ppm
F 01000 ppm
G 02000 ppm
H 05000 ppm
J 01%
K 02%
L 03%
M 05%
P 010%
Q 020%
R 040%
S 050%
T 070%
U 0100%
N Not Available
3rd Component A 050 ppm
1st Range B 0100 ppm
(note 2) C 0200 ppm
D 0250 ppm
W 0300 ppm
E 0500 ppm
F 01000 ppm
G 02000 ppm
H 05000 ppm
J 01%
K 02%
L 03%
M 05%
P 010%
Q 020%
R 040%
S 050%
T 070%
U 0100%
N Not Available
3rd Component B 0100 ppm
2nd Range C 0200 ppm
(note 2) D 0250 ppm
W 0300 ppm
E 0500 ppm
F 01000 ppm
G 02000 ppm
H 05000 ppm
J 01%
K 02%
L 03%
M 05%
P 010%
Q 020%
R 040%
S 050%
T 070%
U 0100%
N Not Available
4th Component A 050 ppm
1st Range B 0100 ppm
(note 2) C 0200 ppm
D 0250 ppm
W 0300 ppm
E 0500 ppm
F 01000 ppm
G 02000 ppm
H 05000 ppm
J 01%
K 02%
L 03%
M 05%
P 010%
Q 020%
R 040%
S 050%
T 070%
U 0100%
N Not Available
To be continued.
All Rights Reserved. Copyright © 2002, Yokogawa Electric Corporation GS 11G02N01-01E 4th Edition Mar.09.2012-00
<<Contents>> <<Index>> 9
MODEL AND SUFFIX CODES
Model Suffix code Option code Description
IR400 Infrared gas analyzer 19-inch rack mounting type with slide rail
4th Component B 0100 ppm
2nd Range C 0200 ppm
(note 2) D 0250 ppm
W 0300 ppm
E 0500 ppm
F 01000 ppm
G 02000 ppm
H 05000 ppm
J 01%
K 02%
L 03%
M 05%
P 010%
Q 020%
R 040%
S 050%
T 070%
U 0100%
N Not available
O2 Analyzer 1 05%
1st Range 2 010%
(note 2) 3 025%
N Not available
O2 Analyzer 2 010%
2nd Range 3 025%
(note 2) N Not available
Output -4 420 mA DC, non-isolation
-1 01 V DC, non-isolation
Piping R Rc 1/4
T 1/4 NPT
Indication, Power Cable (note 7) J Japanese, Power Cable; rated voltage 125 V AC
E English, Power Cable; rated voltage 125 V AC (UL)
U English, Power Cable; rated voltage 250 V AC (CEE)
Option O2 correction and O2 avarage (note 4) /K With O2 correction and O2 average value
/A With peak count alarm (note 4)
Communication /C RS232C (note 9)
Internal purge (note 5) /P Analyzer internal purging
Relay board (note 6) /R With dedicated cable
Footnotes:
1: A signal from the external O2 analyzer should be 0-1 V DC linear to full scale.
2: Possible combinations of ranges are specified in separate tables.
3: Only available for CO2 measurement. Option code “/P,” Analyzer internal purging, must be specified.
4: O2 correction is available only for NO, CO, and SO2. Both average value output after O2
correction and average O2 value output are provided at the same time.
A peak count alarm can be provided only for CO measurement.
5: When internal purging is specified with 3- or 4-component analyzers, only one set of gas
inlet/outlet can be used and thus NO2/NO converter cannot be connected between two
measurement sections.
6: Should be specified when using a solenoid valve for automatic calibration.
7: Suffix Codes “E” and “U” are power cables with different voltage rating and plug type. Select
appropriate code according to the operating power supply voltage to be used in the field.
Suffix Code “E” is of the North American plug type and “U” of the European type.
8: For NOx measurement, a NO2/NO converter (P/N K9350LE or K9350LF) should be purchased separately.
9: Should be specified when using ModbusTM communication.
10: ZX8D style D is specified for CE marking.
All Rights Reserved. Copyright © 2002, Yokogawa Electric Corporation GS 11G02N01-01E 4th Edition Mar.09.2012-00
<<Contents>> <<Index>> 10
Measurable components and ranges – availability check table –
Table 1. Single-component analyzer (NO, SO2, CO2, CO, CH4)
2nd range A B C D W E F G H J
1st range 050ppm 0100ppm 0200ppm 0250ppm 0300ppm 0500ppm 01000ppm 02000ppm 05000ppm 01%
V 020ppm
A 050ppm
B 0100ppm
C 0200ppm
D 0250ppm
W 0300ppm
E 0500ppm
F 01000ppm
G 02000ppm
H 05000ppm
J 01%
K 02%
L 03%
M 05%
P 010%
Q 020%
R 040%
S 050%
T 070%
U 0100%
2nd range K L M P Q R S T U
1st range 02% 0–3% 0–5% 0–10% 0–20% 0–40% 0–50% 0–70% 0–100%
V 0–20ppm
A 0–50ppm
B 0–100ppm
C 0–200ppm
D 0–250ppm
W 0–300ppm
E 0–500ppm
F 0–1000ppm
G 0–2000ppm
H 0–5000ppm
J 0–1%
K 0–2%
L 0–3%
M 0–5%
P 0–10%
Q 0–20%
R 0–40%
S 0–50%
T 0–70%
U 0–100%
:CO2 analyzer measurable range :CO analyzer measurable range :CH4 analyzer measurable range
:SO2 analyzer measurable range :NO analyzer measurable range
All Rights Reserved. Copyright © 2002, Yokogawa Electric Corporation GS 11G02N01-01E 4th Edition Jul 20.2012-07
<<Contents>> <<Index>> 11
Table 2. Two-component analyzer (NO and SO2 )
2nd component (SO2), 1st range SO2
1st component A B C D W E F G H
(NO),1st range 050ppm 0100ppm 0200ppm 0250ppm 0300ppm 0500ppm 01000ppm 02000ppm 05000ppm
A 050ppm
B 0100ppm
C 0200ppm
NO
D 0250ppm
W 0300ppm
E 0500ppm
F 01000ppm
G 02000ppm
H 05000ppm
: Double components measurable range. 1st component ; NO, 2nd component ; SO2. T04.EPS
* 1st range (low range) must meet the combination in above table.
2nd range, both NO and SO2 measurements are available up to 5000 ppm.
All Rights Reserved. Copyright © 2002, Yokogawa Electric Corporation GS 11G02N01-01E 4th Edition May 08.2012-00
<<Contents>> <<Index>> 12
EXTERNAL DIMENSIONS
Analyzer Main Unit
Unit: mm
<Side View>
<Top View>
578 Max.
466.7
315.8
288.9
12.7
123.8
(22)
38
26.1
<Front View>
Power switch
POWER
Sample gas inlet Rc1/4 or 1/4 NPT
(Purge gas inlet Rc1/4 or 1/4 NPT)
MODE ESC ZERO
ENT SPAN
465
483
<Rear View>
429 <Rack mounting hole>
M5
7.5
100.8
177
169
EIA : 101.6mm
JIS : 100mm
All Rights Reserved. Copyright © 2002, Yokogawa Electric Corporation GS 11G02N01-01E 4th Edition Jun.20.2006-00
<<Contents>> <<Index>> 13
Input/Output Terminal Module: K9218SC (Accessory)
Unit: mm
Mounting hole
6- 4.5 hole
316
150 150 8
12
5
(140)
164
154
100030
M4 screw
150 150
154
142
302
Cut M4 screw holes at 6 positions.
Drill a rectangular hole of 302 142 mm or more in the center. F05.EPS
All Rights Reserved. Copyright © 2002, Yokogawa Electric Corporation GS 11G02N01-01E 4th Edition Jun.20.2006-00
<<Contents>> <<Index>> 14
EXTERNAL DIMENSIONS OF ACCESSORY SLIDE RAIL Unit: mm
Closed
(24.4)
466.7–0.3
15.9–0.5
F06.EPS
19.1–0.3 606.6–0.8
Reinforcement plate
101.6
100
100
or
or
465 465
Guide rails
For the guide rail mounted type, a maintenance
space (200 mm or more) should be provided on top of
the main unit.
All Rights Reserved. Copyright © 2002, Yokogawa Electric Corporation GS 11G02N01-01E 4th Edition Jun.20.2006-00
<<Contents>> <<Index>> 15
All Rights Reserved. Copyright © 2002, Yokogawa Electric Corporation GS 11G02N01-01E 4th Edition Jun.20.2006-00
<<Contents>> <<Index>> 16
STANDARD ACCESSORIES
Response time: Approx. 20 seconds (for 90% response)
Measured gas flow rate: 0.5 ± 0.25 L / min
Name Part Number Description Qty
Remark: The Zirconia system, due to its principle,
may produce a measuring error due to
Power cable K9218SA standard inlet type (2.5 m ) 1
relative concentration versus the com-
replacement fuse (250 V AC, 2 bustible O2 gas concentration. Also, a
Fuse K9218SB
3.15 A, delay type) 1 corrosive gas (SO2 of 250 ppm or more,
Input/output
K9218SC External terminal module 1 etc.) may affect the life of the sensor.
terminal module Gas inlet/outlet size: Rc1/4
Connection cable between main unit Power supply: 90 to 126 V AC or 200 to 240 V AC,
Cable K9218SD 1
and input/output terminal module (1 m) 50/60 Hz
Slide rail K9218SE Slide rail 2 Enclosure: Steel casing, for indoor application
T09.EPS
Indication: Temperature indication (LED)
Note: Quantity in this is number of accessories supplied as Temperature alarm output:
standard. For instance, two K9218SE parts, i.e., two slide Contact output 1a contact,
rails, are supplied as standard. When ordering separately, Contact capacity 220 V AC, 1 A (resistive
the required number of should be considered. load)
Safety and EMC conforming standards:
Dedicated Zirconia O2 Sensor Safety; EN61010-1
(to be purchased separately) Pollution degree; 2
For O2 correction, the IR400 can accept linearized 0 to 1 V Installation category; II
DC signal coming from an analyzer calibrated to 0 to 25% EMC; EN61326
O2 of full scale. Dedicated zirconia O2 sensor , Model EN61000-3-2
ZX8D, is available from Yokogawa. EN61000-3-3
Dimensions (H W D):
Measuring method: Zirconia system
140 170 190 mm
Measurable component and measuring range:
Weight: Approx. 3 kg
Measurable component Minimum range Maximum range
Finish color: Munsell 5Y 7/1
Oxygen (O2) 05 vol% 025 vol% Model Suffix code Option code Description
T11.EPS
ZX8D Dedicated zirconia O2 sensor
Repeatability: Within ± 0.5% of full scale Power 5 90126 V AC, 50/60 Hz
Linearity: Within ± 1% of full scale supply 3 200240 V AC, 50/60 Hz
Zero drift: Within ± 1% of full scale/week Style code C Non-CE conformity
D
Span drift: Within ± 2% of full scale/week CE conformity
T13.EPS
All Rights Reserved. Copyright © 2002, Yokogawa Electric Corporation GS 11G02N01-01E 4th Edition Feb.08.2011-04
<<Contents>> <<Index>> 17
External Dimensions of ZX8D Unit: mm
170
152
OXYGEN ANALYZER
800
TEMP
140
131
110
7-M4
POWER FUSE3A 1 2 3 4 5 6 7
OFF + -
AC E OUT ALM
1 2 3 4 5 6 7
+ -
External Dimensions
Unit: mm
15 Max.
Mounting hole ø4.5
CN13
65
75
CN9
CN1 CN2 CN3 CN4 CN5 CN6 CN7 CN8
140
150
F10.EPS
All Rights Reserved. Copyright © 2002, Yokogawa Electric Corporation GS 11G02N01-01E 4th Edition Jan.20.2007-02
<<Contents>> <<Index>> 18
Connections NO2/NO Converter
CN9 CN1 CN2 CN3 CN4 CN5 CN6 CN7 CN8
Part number: K9350LE (Non-CE conformity)
K9350LF (CE conformity)
Reserved
Mounting: Indoor surface mounting
Target Gases: General boiler exhaust gas,
atmosphere
Catalyst: Amount; 2 cm3
Solenoid valve
Solenoid valve
driving power For switching For zero gas CH1 CH2 CH3 CH4 CH5
Replacement cycle; Approx. 8 months
sample
For span gas
(at flow rate of 0.5 L/min with 5% O2, 10
ppm NO2)
Temperature set-point; 220 10C
CN13
(Sensing tip: K thermocouple)
Wetted materials: Ceramic, Viton, glass filter, SUS316
I/O terminal module
Conversion efficiency: 90% or higher, conforms to JIS
Gas Flow Rate: 0.5 L/min
Relay board CN3
Ambient Temperature: -5 to +45C
Dedicated cable (D-sub 9-pin straight cable 1.5 m)
F11.EPS
Power Supply: 100 VAC, 50/60 Hz (K9350LE)
100 to 240 VAC, 50/60 Hz (K9350LF)
Contact action Safety conforming standards:
• During measurement: CN1; ON Safety; EN61010-1
Others; OFF Pollution degree; 2
• During calibration: CN1; OFF Installation category; II
Others; Contact corresponding to Weight: Approx. 1.1 kg (K9350LE)
calibration timing is ON Approx. 1.2 kg (K9350LF)
Recommended Connector Sample gas requirements:
• CN1 to CN9: Housing; VHR-2N Dust/drain removed, gas temperature at
(Japan Solderless Terminals (JST)) 150C or less
Contact; SVH-21T-1.1
One-year-Use Spare Parts
(Japan Solderless Terminals (JST))
Item Part No. Qty
Catalyst for NO2/NO converter K9350LP 2
Glass wool for NO2/NO converter K9350LQ 2
Fitting for NO2/NO converter K9350LV 2
T15.eps
External Dimensions
Unit: mm
(150)
135
Wiring Diagram
(210)
(120)
105
7 8
4- 5
100V AC
Approx.130
K9350LE K9350LF
All Rights Reserved. Copyright © 2002, Yokogawa Electric Corporation GS 11G02N01-01E 4th Edition Jan.20.2007-02
<<Contents>> <<Index>> 19
Inquiry Sheet for IR400 NDIR Type Infrared Gas Analyzer
Place a checkmark ✔ in the appropriate box and fill in the specific information in the blanks for your reference.
1. General Information
Company: Delivery destination:
Responsible person: Section: (Phone No. )
Plant name: Measurement location:
Purpose: Indication reading, Recording, Alarm, Control
2. Requirements
Measurable component:
O2 Analyzer:
Without O2 analyzer
External zirconia type sensor (purchase separately:
ZX8D Style C)
External O2 analyzer
Built-in paramagnetic type O2 sensor
NO2/NO Converter
With NO2/NO converter
Without NO2/NO converter
Range:
1st component, 1st range 1st component, 2nd range 2nd component, 1st range 2nd component, 2nd range
0 – 20 ppm 0 – 50 ppm 0 – 50 ppm 0 – 100 ppm
0 – 50 ppm 0 – 100 ppm 0 – 100ppm 0 – 200 ppm
0 – 100 ppm 0 – 200 ppm 0 – 200 ppm 0 – 250 ppm
0 – 200 ppm 0 – 250 ppm 0 – 250 ppm 0 – 500 ppm
0 – 250 ppm 0 – 500 ppm 0 – 500 ppm 0 – 1000 ppm
0 – 500 ppm 0 – 1000 ppm 0 – 1000 ppm 0 – 2000 ppm
0 – 1000 ppm 0 – 2000 ppm 0 – 2000 ppm 0 – 5000 ppm
0 – 2000 ppm 0 – 5000 ppm 0 – 5000 ppm 0 – 1%
0 – 5000 ppm 0 – 1% 0 – 1% 0 – 2%
0 – 1% 0 – 2% 0 – 2% 0 – 3%
0 – 2% 0 – 3% 0 – 3% 0 – 5%
0 – 3% 0 – 5% 0 – 5% 0 – 10%
0 – 5% 0 – 10% 0 – 10% 0 – 20%
0 – 10% 0 – 20% 0 – 20% 0 – 40%
0 – 20% 0 – 40% 0 – 40% 0 – 50%
0 – 40% 0 – 50% 0 – 50% 0 – 70%
0 – 50% 0 – 70% 0 – 70% 0 – 100%
0 – 70% 0 – 100% 0 – 100% Not available
0 – 100% Not available Not available
All Rights Reserved. Copyright © 2002, Yokogawa Electric Corporation GS 11G02N01-01E 4th Edition Jun.20.2006-00
<<Contents>> <<Index>> 20
3rd component, 1st range 3rd component, 2nd range 4th component, 1st range 4th component, 2nd range
0 – 50 ppm 0 – 100 ppm 0 – 50 ppm 0 – 100 ppm
0 – 100 ppm 0 – 200 ppm 0 – 100 ppm 0 – 200 ppm
0 – 200 ppm 0 – 250 ppm 0 – 200 ppm 0 – 250 ppm
0 – 250 ppm 0 – 500 ppm 0 – 250 ppm 0 – 500 ppm
0 – 500 ppm 0 – 1000 ppm 0 – 500 ppm 0 – 1000 ppm
0 – 1000 ppm 0 – 2000 ppm 0 – 1000 ppm 0 – 2000 ppm
0 – 2000 ppm 0 – 5000 ppm 0 – 2000 ppm 0 – 5000 ppm
0 – 5000 ppm 0 – 1% 0 – 5000 ppm 0 – 1%
0 – 1% 0 – 2% 0 – 1% 0 – 2%
0 – 2% 0 – 3% 0 – 2% 0 – 3%
0 – 3% 0 – 5% 0 – 3% 0 – 5%
0 – 5% 0 – 10% 0 – 5% 0 – 10%
0 – 10% 0 – 20% 0 – 10% 0 – 20%
0 – 20% 0 – 40% 0 – 20% 0 – 40%
0 – 40% 0 – 50% 0 – 40% 0 – 50%
0 – 50% 0 – 70% 0 – 50% 0 – 70%
0 – 70% 0 – 100% 0 – 70% 0 – 100%
0 – 100% Not available 0 – 100% Not available
Not available Not available
Output: 4 – 20 mA DC 0 – 1 V DC RS-232C
O2 correction and O2 average: Yes No
Peak count alarm: Yes No
Relay board: Yes No
Composition (Detailed composition of sample gas should be provided. This is important for the purpose of
knowing the effect of interference gases)
Contents Concentration range
CO to % ppm
CO2 to % ppm
CH4 to % ppm
H2 to % ppm
O2 to % ppm
N2 to % ppm
SO2 to % ppm
NOx to % ppm
H2O to % ppm
to % ppm
All Rights Reserved. Copyright © 2002, Yokogawa Electric Corporation GS 11G02N01-01E 4th Edition Jun.20.2006-00
Subject to change without notice.