General IR400
Specifications NDIR TYPE INFRARED GAS ANALYZER
(5-COMPONENT ANALYZER)
GS 11G02N01-01E

The IR400 infrared gas analyzer is capable of measuring


the concentrations of NO, SO2, CO2, CO, CH4 and O2
components in sample gas.
NO, SO2, CO2, CO and CH4 are measured by non-
dispersive infrared method (NDIR), while O2 is measured
by built-in paramagnetic sensor or external zirconia
sensor. A maximum of 5 components including O2 (up to
4 components except for O2 measurement) are simulta-
neously measurable.
The mass flow type twin detector of high sensitivity and ■ SPECIFICATIONS
reliability adopted in the infrared ray method detection
unit makes the measurement hardly affected by Standard Specifications
interfering components. Measurement principle:
In addition, the IR400 has a large-size liquid crystal NO, SO2, CO2, CO, CH4: Non-dispersive infrared method
display, providing easy operation, high accuracy and Single light source-double beams
multiple functions. O2: Paramagnetic type (built-in), or
zirconia type (external)
Optimum as an analyzer unit of measurement system for
Measurable gas components and measuring ranges:
combustion exhaust gas from refuse incinerator and
boiler, or gas from different industrial furnaces. Range
Minimum range Maximum range
Component
NO 0 – 50 ppm 0 – 5000 ppm
■ FEATURES SO2 0 – 50 ppm 0 – 10 vol%
1. Simultaneous and continuous measurement of up to 5 CO2 0 – 20 ppm 0 – 100 vol%
components including O2 and 4 components selected CO 0 – 50 ppm 0 – 100 vol%
from among NO, SO2, CO, CO2, and CH4. CH4 0 – 200 ppm 0 – 100 vol%
2. Minimal interference from other gas components O2
0 – 5 vol% 0 – 25 vol%
The mass flow type twin detector of high sensitivity and (paramagnetic)
reliability minimizes interference from other gas O2
0 – 5 vol% 0 – 25 vol%
components, ensuring excellent stability. (zirconia)
T01.eps
3. Extensive functions
Incorporating O2 correction, average value computing, • Measurement of up to 5 components including O2.
automatic calibration, one touch calibration, upper/lower • 1 or 2 measuring range per component.
limit alarm, remote measurement range changeover, • Measuring range ratio  1:5 (O2 analyzer)
range identification signal output, etc., the analyzer  1:25 (except O2 analyzer)
accommodates different application requirements. For measurable components and possible combina
4. Easy-to-read, large LCD tions of measuring ranges, refer to Tables 1-(1) to (7).
Large LCD provides clear indications of all measured
Display:
components and computed values and easy interactive
Digital indication in 4 digits (LCD with backlight)
operation.
• Instantaneous value of each component
5. 19-inch rack mounting
• Instantaneous value after O2 correction
Unitized construction of the main body on the 19-inch
(only in NO, SO2, CO with O2 measurement)
rack and of the signal input/output terminal module
• Average value after O2 correction
allows easy configuration of a gas analyzer system.
(only in NO, SO2, CO with O2 measurement)
6. Maximum measuring range ratio
• Average O2 value
A maximum range ratio of 1:25 is achieved.
Analog output signal:
* Input/Output of analog signals is
available in combination with the
input/output terminal module.
4 to 20 mA DC or 0 to 1 V DC,
non-isolated, 12 points max.
Analog output corresponds one-to-one
with measured value indication.

Yokogawa Electric Corporation GS 11G02N01-01E


2-9-32, Nakacho, Musashino-shi, Tokyo, 180-8750 Japan ©Copyright June 2002
Tel.: 81-422-52-5617 Fax.: 81-422-52-6792 1st Edition June 2002
4th Edition June 2006
<<Contents>> <<Index>> 2
Permissible load resistance; Weight: Approx. 22 kg (only analyzer)
550  max. for 4 to 20 mA DC Finish color: Front panel; Off-white (Munsell 10Y7.5/0.5
100 k min. for 0 to 1 V DC or equivalent)
* Refer to Table 8 for the channel numbers Casing; Plating, Steel-blue (gray)
of displayed values and analog output Enclosure: Steel casing, for indoor use
signals. Material of gas-contacting parts:
Analog input signal: Gas inlet/outlet; SUS304
For signal input from external O2 analyzer. Sample cell; SUS304/neoprene rubber
Signal requirement; Infrared-ray transmitting window; CaF2
(1) Signal from Yokogawa’s zirconia O2 O2 sensor sampling cell: SUS316
sensor (Model ZX8D*C or ZX8D*D) Internal piping; Toaron tube, Teflon tube
(2) 0 to 1V DC from an O2 sensor Gas inlet/outlet: Rc1/4 or 1/4NPT internal thread
Input section is not isolated. This Purge gas flow rate: 1 L/min (when required)
feature is effective when built-in O2 Installation altitude: 2000 m or less
sensor is not used. Safety and EMC conforming standards:
(An input signal triggers measured Safety; EN61010-1: 2001
concentration indication and O2 Pollution degree; 2 (Note)
correction.) Installation category; II (Note)
* External O2 sensor should be purchased Note: Installation category, called over-
separately. voltage category, specifies impulse with
Relay contact output: standing voltage. Category II is for
1a contact (250 V AC/2 A, resistive load) electrical equipment. Pollution degree
Instrument error, calibration error, range indicates the degree of existence of
solid, liquid, gas or other inclusions
identification, auto calibration status,
which reduce dielectric strength.
pump ON/OFF, peak alarm. Degree 2 is the nomal indoor environ
1c contact (250 V AC/2 A, resistive load) ment.
Selectable 6 outputs. EMC; EN61326-1: 1997, A1: 1998,
High/Low limit alarm contact output (for A2: 2001 Class A
each channel). EN61000-3-2: 2000
Power disconnection alarm. EN61000-3-3: 1955, A1: 2001
* All relay contacts are isolated mutually
and from the internal circuit. CAUTION
Contact input: Non-voltage contact (ON/0 V, OFF/ 5V DC,
The instrument is a Class A product, and it is designed
5 mA flowing at ON).
Remote range changeover, auto calibra- for use in the industrial environment. Please use this
tion remote start, remote hold, average instrument in the industrial environment only.
value reset, pump ON/OFF.
Isolated from the internal circuit with Standard Functions
photocoupler. Contact inputs are not Output signal hold:
isolated from one another. Output signals are held during manual and
Transmission output: auto calibrations by activation of holding
Solenoid valve drive signal for automatic (turning "ON" its setting).
calibration. The values to be held are the ones just
Transistor output (100 mA or less) before start calibration mode.
* For details, see External Connection Indication values will not be held.
Diagram on page 15. Remote output hold:
Power supply: Voltage rating; 100 to 240 V AC Output signal is held at the latest value or
Allowable range; 85 to 264 V AC setting value by short-circuiting the remote
Frequency; 50/60 Hz output holding input terminals.
Power consumption; 250 VA max. Holding is maintained while the terminals
Inlet; Conform to EN60320 are short-circuited. Indication values will
Protection Class 1 not be held.
Operating conditions: Switch range :
Ambient temperature; –5 to 45˚C The range changeover is available in
Ambient humidity; 90% RH max., non- manual, auto, and remote modes. Only
condensing. preset changeover method is effective.
Storage conditions: Manual; Allows range to switch by key operation.
Ambient temperature; –20 to 60˚C Auto; Allows range to switch from low to high
Ambient humidity; 90% RH max., non- range when 90%FS or more is available in
condensing the low range.
Dimensions (H  W  D): Allows range to switch from high to low
Analyzer main unit; range when 80%FS or less is available in
177  483  578 mm the low range.
Input/Output terminal module;
164  316  55 mm

All Rights Reserved. Copyright © 2002, Yokogawa Electric Corporation GS 11G02N01-01E 4th Edition Oct.02.2006-01
<<Contents>> <<Index>> 3
Remote; Non-voltage contact input (for measurable Settable 60 to 900 seconds (in increments
components) of 1 second)
Allows range to switch via an external Upper/lower limit alarm:
signal when remote range changeover Alarm contact output turns on when
input is received. measurement value reach the preset
Range identification signal: upper or lower limit alarm value.
The present measuring range is identified Contacts close when the instantaneous
by a contact signal. value of each component becomes larger
The contact output terminals for each than the upper alarm limit value or smaller
component are short-circuited when the than the lower alarm limit value.
first range is selected, and when the Instrument error contact output:
second range is selected, the terminals are Contacts close at occurrence of analyzer
open. error No. 1, 3 or 10.
Auto calibration: Calibration error contact output:
Auto calibration is carried out periodically Contacts close at occurrence of manual or
at the preset cycle. auto calibration error (any of errors No. 4
When a standard gas cylinder for calibra- to 9).
tion and a solenoid valve for opening/ Auto calibration status contact output:
closing the gas flow line are prepared Contacts close during auto calibration.
externally by the customer, calibration will Pump ON/OFF contact output:
be carried out with the solenoid valve drive During measurement, contacts close.
contacts for zero calibration and each span While calibration gas is flowing, contacts
calibration turned on/off sequentially at the open. Contacts are connected in power
set auto calibration timing. supply of pump, and stop the sample gas
Auto calibration cycle setting; while calibration gas is flowing.
Auto calibration cycle is set.
Setting is variable within 1 to 99 hours (in Optional Functions
increments of 1 hour) or 1 to 40 days (in O2 correction: Conversion of measured NO, SO2 and CO
increments of 1 day). gas concentrations into values at reference
Gas flow time setting; O2 concentration.
The time for flowing each calibration gas in 21 – On
auto calibration is set. Correction formula: C =  Cs
21 – Os
Settable within 60 to 900 seconds (in
increments of 1 second) Where:
Auto calibration remote start: C: Sample gas concentration after O2
Auto calibration is carried out only once correction
according to an external input signal. Cs: Measured concentration of sample
Calibration sequence is settable in the gas
same way as the general auto calibration. Os: Measured O2 concentration (limit
Auto calibration is started by opening the setting: 1 to 20% O2, default 17% )
auto calibration remote start input On: Reference O2 concentration (value
terminals after short-circuiting for 1.5 changeable by setting: 0 to 19% O2,
seconds or longer. default 4%)
Auto zero calibration: The result of calculation is indicated and
Auto zero calibration is carried out output in an analog output signal.
periodically at the preset cycle. Average value after O2 correction and O2 average value
This cycle is independent of "Auto calculation:
calibration" cycle. The result of O2 correction or instanta
When zero calibration gas and solenoid neous O2 value can be outputted as an
valve for opening/closing the calibration average value in the determined period of
gas flow line are prepared externally by the time.
customer, zero calibration will be carried Used for averaging is the moving average
out with the solenoid valve drive contact method in which sampling is carried out at
for zero calibration turned on/off at the set intervals of 30 seconds.
auto zero calibration timing.
(Output is updated every 30 seconds. It is
Auto zero calibration cycle setting;
the average value in the determined period
Auto zero calibration cycle is set.
of time just before the latest updating.)
Setting is variable within 1 to 99 hours (in
Averaging time is settable within 1 to 59
increments of 1 hour) or 1 to 40 days (in
minutes (in increments of 1 minute) or 1 to
increments of 1 day)
Gas flow time setting; 4 hours (in increments of 1 hour).
The time for flowing zero gas in auto zero
calibration is set.

All Rights Reserved. Copyright © 2002, Yokogawa Electric Corporation GS 11G02N01-01E 4th Edition Aug.27.2007-01
<<Contents>> <<Index>> 4
Average value reset: Standard Requirements for Sample Gas
The above-mentioned output of average
Flow rate: 0.5 ±0.2 L/min
value is started from the initial state by
Temperature: 0 to 50˚C
opening the average value resetting input
Pressure: 10 kPa or less (Gas outlet side should be
terminals after short-circuiting for 1.5
open to the atmospheric air.)
seconds or longer.
Dust: 100 µg/Nm3 or less in particle size of 1 µm
Output is reset by short-circuiting and
or less
restarted by opening.
Mist: Unallowable
CO concentration peak count alarm:
Moisture: Below a level where saturation occurs at
(available only for CO + O2 measurement)
2˚C (condensation unallowable).
Alarm output turns on according to the
Corrosive component: HCl 1 ppm or less
preset concentration and count.
Standard gas for calibration:
Whenever the instantaneous value of CO
Zero gas; Dry N2
exceeds the preset concentration value,
Span gas; Each sample gas having
count increments. If the count exceeds the
concentration 90 to 100% of its
preset value in one hour, the alarm
measuring range (recom-
contacts close.
mended).
Communication function:
Gas beyond concentration
RS-232C (9 pins D-sub)
100%FS is unusable.
Half-duplex bit serial
In case a zirconia O2 analyzer is installed
Start-stop synchronization
externally and calibration is carried out on
ModbusTM protocol
the same calibration gas line:
Contents: Read/write parameters
Zero gas; Dry air or atmospheric air
Read measurement concentra-
(provided without CO2 sensor)
tion and instrument status
Span gas; For other than O2 measure-
Remark: When connecting via RS-485
ment, each sample gas having
interface, a RS-232C RS-
concentration 90 to 100% of its
485 converter should be used.
measuring range.
For O2 measurement, O2 gas of
Performance 1 to 2 vol%.

Repeatability: 0.5% of full scale (1% of full scale for Installation Requirements
range less than 0-50 ppm)
• Indoor use: Avoid exposure to direct sunlight, weather,
Linearity: 1% of full scale
and radiant heat from hot substances. Where exposure
Zero drift: 1% of full scale/week
to such conditions are unavoidable, a protective hood or
(2% of full scale/week for range between
cover should be prepared.
0-50 ppm and 0-200 ppm)
• Minimal vibration
(2% of full scale/day for range less than
• A clean atmosphere
0-50 ppm)
Span drift: 2% of full scale/week
(2% of full scale/day for range less than
0-50 ppm)
Response time (for 90%FS response) :
Within 60 seconds including replacement
time of sample gas (when gas flow rate is
0.5 L/min). Gas replacement time depends
on the number of measuring components,
and measuring range effects of interfering
gases.
Effects of interfering gases
When sample gas contains gas components listed below,
the measurement accuracy may suffer. Consult Yokogawa
for countermeasures or effect on accuracy.

Analyzer Interference gas Effect


SO2 analyzer NO2 50 ppm of NO2 is equivalent to -6 ppm of SO2
CO2 15% of CO2 is equivalent to 7 to 10 ppm of CO
CO analyzer
N 2O 1000 ppm of N2O is equivalent to 80 ppm of CO
CH4 analyzer CO2 15% of CO2 is equivalent to approx. 3 ppm of CH4
T03s.EPS

All Rights Reserved. Copyright © 2002, Yokogawa Electric Corporation GS 11G02N01-01E 4th Edition Jun.20.2006-00
<<Contents>> <<Index>> 5
Diagram of measurement principle of infrared gas analyzer (NO, SO2, CO2, CO, CH4)

Motor Distribution cell


(interference filter) Interference compensating
Trimmer detector
Reference cell

Rotary sector

Sector
sync. signal Infrared (single) Sample cell
light source Sample gas Sample gas
inlet outlet
Gain changeover
AC amplifier

O2 sensor
(optional)
A/D

RAM
CPU
ROM
External input / Contact output D/A Indicator

Indication

Operating keys
Output
0 to 1 V DC or
4 to 20 mA DC F01.EPS

Diagram of measurement principle of paramagnetic oxygen analyzer

Permanent Mirror
magnets

Gas inlet Measuring cell

Magnetic field

Gas outlet

Preamplifier

Permanent
magnets

LED Photodiode

Signal processing
Indication
and calculation block

Output F02.EPS

All Rights Reserved. Copyright © 2002, Yokogawa Electric Corporation GS 11G02N01-01E 4th Edition Jun.20.2006-00
<<Contents>> <<Index>> 6
Example of gas sampling system configuration
The following illustrates a typical system configuration for five component gas measurement for monitoring combustion
exhaust gas from boiler, refuse incinerator, etc. Contact Yokogawa for system configuration matching the particular use or
further information.
1 Probe IR400
7 Membrane filter
2 Mist filter
CO2, CO
Gas inlet tube (O2)
4 Pump
15˚ or larger 8 Flowmeter
Infrared gas
NO2/NO
11 analyzer main
converter unit
for NOx
5 measurement
Electric
gas cooler 7 Membrane
filter NO, SO2

3 Safety drain trap 6 Solenoid


valve
ZERO NO SO2 CO2 CO O2
Pressure
reducing valve O2
Exhaust
N2 NO SO2 CO2 CO O2 10 Zirconia
or /N2 /N2 /N2 /N2 /N2 O2 sensor
Air

Drain 9 Standard gas (3.4L)


F03.EPS

Typical sampling system components


No. Item Description

1 Probe With a heating type stainless steel filter. Standard pore size: 40 m

2 Mist filter Separates drain and removes dust and mist.

3 Safety drain trap Prevents drain from being sucked. Composite operation of
constant-pressure bubbler.

4 Pump Sucks in sample gas. Sample gas flow rate: approx. 2 L/min

5 Electric gas cooler Dehumidifies sample gas.

6 Solenoid valve Used for introducing calibration gas.

7 Membrane filter Glass fiber or PTFE filter removes fine dust. Dust buildup
conditions can be monitored through front panel of analyzer.

8 Flowmeter Adjusts and monitors sample gas flow rate.

9 Standard gas Used for zero/span calibration. Zero, NO, SO2, CO, CO2, and O2
gas cylinders.

10 Zirconia O2 sensor Not required when built-in type is used.
Installed externally. Measures O2 concentration (0 to 25%) of
sample gas.

11 NO2/NO converter Required for NOx measurement. Converts NO2 to NO gas
efficiently using special catalyst.
For each sampling component, consult with Yokogawa. T20.EPS

All Rights Reserved. Copyright © 2002, Yokogawa Electric Corporation GS 11G02N01-01E 4th Edition Jun.20.2006-00
<<Contents>> <<Index>> 7
MODEL AND SUFFIX CODES
[Style:S3]
Model Suffix code Option code Description
IR400 Infrared gas analyzer 19-inch rack mounting type with slide rail
Measurable 1st 2nd 3rd 4th
component -A NO
(note 8) -B SO2
-C CO2
-D CO
-F CH4
-G NO SO2
-H NO CO
-J CO2 CO
-K NO SO2 CO
-L NO SO2 CO2 CO
O2 Analyzer N Without O2 analyzer
1 External zirconia type O2 sensor (purchase separately: ZX8D)
(note 10)
2 External O2 analyzer (note 1)
3 Built-in paramagnetic type O2 sensor
1st Component
V 020 ppm (note 3)
1st Range
A 050 ppm
(note 2)
B 0100 ppm
C 0200 ppm
D 0250 ppm
W 0300 ppm
E 0500 ppm
F 01000 ppm
G 02000 ppm
H 05000 ppm
J 01%
K 02%
L 03%
M 05%
P 010%
Q 020%
R 040%
S 050%
T 070%
U 0100%
1st Component A 050 ppm
2nd Range B 0100 ppm
(note 2) C 0200 ppm
D 0250 ppm
W 0300 ppm
E 0500 ppm
F 01000 ppm
G 02000 ppm
H 05000 ppm
J 01%
K 02%
L 03%
M 05%
P 010%
Q 020%
R 040%
S 050%
T 070%
U 0100%
N Not available
2nd Component A 050 ppm
1st Range B 0100 ppm
(note 2) C 0200 ppm
D 0250 ppm
W 0300 ppm
E 0500 ppm
F 01000 ppm
G 02000 ppm
H 05000 ppm
J 01%
K 02%
L 03%
M 05%
P 010%
Q 020%
R 040%
S 050%
T 070%
U 0100%
N Not available

To be continued.

All Rights Reserved. Copyright © 2002, Yokogawa Electric Corporation GS 11G02N01-01E 4th Edition Mar.09.2012-01
<<Contents>> <<Index>> 8
MODEL AND SUFFIX CODES
Model Suffix code Option code Description
IR400 Infrared gas analyzer 19-inch rack mounting type with slide rail
2nd Component B 0100 ppm
2nd Range C 0200 ppm
(note 2) D 0250 ppm
W 0300 ppm
E 0500 ppm
F 01000 ppm
G 02000 ppm
H 05000 ppm
J 01%
K 02%
L 03%
M 05%
P 010%
Q 020%
R 040%
S 050%
T 070%
U 0100%
N Not Available
3rd Component A 050 ppm
1st Range B 0100 ppm
(note 2) C 0200 ppm
D 0250 ppm
W 0300 ppm
E 0500 ppm
F 01000 ppm
G 02000 ppm
H 05000 ppm
J 01%
K 02%
L 03%
M 05%
P 010%
Q 020%
R 040%
S 050%
T 070%
U 0100%
N Not Available
3rd Component B 0100 ppm
2nd Range C 0200 ppm
(note 2) D 0250 ppm
W 0300 ppm
E 0500 ppm
F 01000 ppm
G 02000 ppm
H 05000 ppm
J 01%
K 02%
L 03%
M 05%
P 010%
Q 020%
R 040%
S 050%
T 070%
U 0100%
N Not Available
4th Component A 050 ppm
1st Range B 0100 ppm
(note 2) C 0200 ppm
D 0250 ppm
W 0300 ppm
E 0500 ppm
F 01000 ppm
G 02000 ppm
H 05000 ppm
J 01%
K 02%
L 03%
M 05%
P 010%
Q 020%
R 040%
S 050%
T 070%
U 0100%
N Not Available

To be continued.

All Rights Reserved. Copyright © 2002, Yokogawa Electric Corporation GS 11G02N01-01E 4th Edition Mar.09.2012-00
<<Contents>> <<Index>> 9
MODEL AND SUFFIX CODES
Model Suffix code Option code Description
IR400 Infrared gas analyzer 19-inch rack mounting type with slide rail
4th Component B 0100 ppm
2nd Range C 0200 ppm
(note 2) D 0250 ppm
W 0300 ppm
E 0500 ppm
F 01000 ppm
G 02000 ppm
H 05000 ppm
J 01%
K 02%
L 03%
M 05%
P 010%
Q 020%
R 040%
S 050%
T 070%
U 0100%
N Not available

O2 Analyzer 1 05%
1st Range 2 010%
(note 2) 3 025%
N Not available
O2 Analyzer 2 010%
2nd Range 3 025%
(note 2) N Not available
Output -4 420 mA DC, non-isolation
-1 01 V DC, non-isolation

Piping R Rc 1/4
T 1/4 NPT
Indication, Power Cable (note 7) J Japanese, Power Cable; rated voltage 125 V AC
E English, Power Cable; rated voltage 125 V AC (UL)
U English, Power Cable; rated voltage 250 V AC (CEE)

Option O2 correction and O2 avarage (note 4) /K With O2 correction and O2 average value
/A With peak count alarm (note 4)
Communication /C RS232C (note 9)
Internal purge (note 5) /P Analyzer internal purging
Relay board (note 6) /R With dedicated cable

Footnotes:
1: A signal from the external O2 analyzer should be 0-1 V DC linear to full scale.
2: Possible combinations of ranges are specified in separate tables.
3: Only available for CO2 measurement. Option code “/P,” Analyzer internal purging, must be specified.
4: O2 correction is available only for NO, CO, and SO2. Both average value output after O2
correction and average O2 value output are provided at the same time.
A peak count alarm can be provided only for CO measurement.
5: When internal purging is specified with 3- or 4-component analyzers, only one set of gas
inlet/outlet can be used and thus NO2/NO converter cannot be connected between two
measurement sections.
6: Should be specified when using a solenoid valve for automatic calibration.
7: Suffix Codes “E” and “U” are power cables with different voltage rating and plug type. Select
appropriate code according to the operating power supply voltage to be used in the field.
Suffix Code “E” is of the North American plug type and “U” of the European type.
8: For NOx measurement, a NO2/NO converter (P/N K9350LE or K9350LF) should be purchased separately.
9: Should be specified when using ModbusTM communication.
10: ZX8D style D is specified for CE marking.

All Rights Reserved. Copyright © 2002, Yokogawa Electric Corporation GS 11G02N01-01E 4th Edition Mar.09.2012-00
<<Contents>> <<Index>> 10
Measurable components and ranges – availability check table –
Table 1. Single-component analyzer (NO, SO2, CO2, CO, CH4)

2nd range A B C D W E F G H J
1st range 050ppm 0100ppm 0200ppm 0250ppm 0300ppm 0500ppm 01000ppm 02000ppm 05000ppm 01%
V 020ppm    
A 050ppm    
B 0100ppm    
C 0200ppm    
D 0250ppm       
W 0300ppm       
E 0500ppm      
F 01000ppm       
G 02000ppm        
H 05000ppm         
J 01%          
K 02%          
L 03%          
M 05%          
P 010%          
Q 020%          
R 040%          
S 050%          
T 070%          
U 0100%          

2nd range K L M P Q R S T U
1st range 02% 0–3% 0–5% 0–10% 0–20% 0–40% 0–50% 0–70% 0–100%
V 0–20ppm         
A 0–50ppm         
B 0–100ppm         
C 0–200ppm         
D 0–250ppm         
W 0–300ppm         
E 0–500ppm         
F 0–1000ppm        
G 0–2000ppm       
H 0–5000ppm     
J 0–1%    
K 0–2%    
L 0–3%    
M 0–5%   
P 0–10%    
Q 0–20%     
R 0–40%      
S 0–50%       
T 0–70%        
U 0–100%        
:CO2 analyzer measurable range :CO analyzer measurable range :CH4 analyzer measurable range
:SO2 analyzer measurable range :NO analyzer measurable range

All Rights Reserved. Copyright © 2002, Yokogawa Electric Corporation GS 11G02N01-01E 4th Edition Jul 20.2012-07
<<Contents>> <<Index>> 11
Table 2. Two-component analyzer (NO and SO2 )
2nd component (SO2), 1st range SO2
1st component A B C D W E F G H
(NO),1st range 050ppm 0100ppm 0200ppm 0250ppm 0300ppm 0500ppm 01000ppm 02000ppm 05000ppm
A 050ppm 
B 0100ppm 
C 0200ppm 
NO
D 0250ppm 
W 0300ppm 
E 0500ppm
F 01000ppm
G 02000ppm
H 05000ppm     
: Double components measurable range. 1st component ; NO, 2nd component ; SO2. T04.EPS

* 1st range (low range) must meet the combination in above table.
2nd range, both NO and SO2 measurements are available up to 5000 ppm.

Table 3. Two-component analyzer (NO and CO)


Both NO and CO analyzer must meet the range in Table 1; Single component analyzer.
Table 4. Two-component analyzer (CO2 and CO)

2nd component (CO), 1st range CO


1st component A B C D W E F G H J
(CO2),1st range 050ppm 0100ppm 0200ppm 0250ppm 0300ppm 0500ppm 01000ppm 02000ppm 05000ppm 01%
A 050ppm
B
0100ppm
C
0200ppm
0250ppm
D
0300ppm
W
CO2 E 0500ppm
F 01000ppm
G 02000ppm ⴛ 2.5 ⴛ 2.5 ⴛ 2.5 ⴛ 2.5 ⴛ5 ⴛ 10 ⴛ 10
H 05000ppm ⴛ1 ⴛ1 ⴛ1 ⴛ1 ⴛ2 ⴛ4 ⴛ4
J 01% ⴛ1 
ⴛ2 
ⴛ2 ⴛ 10
K 02% ⴛ1 ⴛ1 ⴛ5 ⴛ 10
L 05% ⴛ2 ⴛ5
M 010% ⴛ2 ⴛ2 ⴛ2 ⴛ2 ⴛ2 ⴛ2 ⴛ 2ⴛ 10 ⴛ 2ⴛ 5 ⴛ2 ⴛ5
N 020% ⴛ1 ⴛ1 ⴛ1 ⴛ1 ⴛ1 ⴛ1 ⴛ 1ⴛ 10 ⴛ 1ⴛ 5 ⴛ1 ⴛ 2.5
:Double components measurable. 1st component ; CO2, 2nd component ; CO.
Note: 1st range (low range) must meet the combination in above table. (For 0-200 ppm range, measurement is available up to 25 times.)
2nd range, is specified; both CO 2 and CO measurements are available up to 20 times of the 1st range.
is specified; CO measurement is available up to 20 times of the 1st range.
CO 2 measurement is available up to the ratio written after the mark.
is specified; both CO 2 and CO measurements are available up to ratio written after the mark.
The ratio, first value is for CO 2 , second value is for CO.
example: ⴛ2 ⴛ5 means, 2nd range of CO 2 is available up to double of 1st range, 2nd range of CO is
available up to 5 times of 1st range.
ⴛ 1 means only 1st range.

Table 5. Three-component analyzer (NO + SO2 + CO)


See Table 2 for NO + SO2 measurement of three-component analyzer (NO + SO2 + CO). See Table 1 for CO measurement.
Table 6. Four-component analyzer (NO + SO2 + CO2 + CO)
See Table 2 for NO + SO2 measurement and Table 4 for CO2 + CO measurement.
Table 7. O2 analyzer
2nd range 2 3
1st range 010% 025%
1 05%
2 010% 
3 025%  T07.EPS

: Built-in O2 analyzer measurable range


: External zirconia type O2 analyzer (in this case, Yokogawa’s ZX8D )
measurable range
*O2 analyzer is selectable independently of combination with other components.

All Rights Reserved. Copyright © 2002, Yokogawa Electric Corporation GS 11G02N01-01E 4th Edition May 08.2012-00
<<Contents>> <<Index>> 12
EXTERNAL DIMENSIONS
Analyzer Main Unit

Unit: mm

<Side View>
<Top View>

M4 (For mounting slide rail)

578 Max.

466.7
315.8
288.9
12.7
123.8
(22)

38

26.1
<Front View>

Power switch

Sample gas outlet Rc1/4 or 1/4 NPT


INFRARED GAS ANALYZER

POWER
Sample gas inlet Rc1/4 or 1/4 NPT
(Purge gas inlet Rc1/4 or 1/4 NPT)
MODE ESC ZERO

ENT SPAN

465
483

<Rear View>
429 <Rack mounting hole>

M5
7.5

100.8

177
169

EIA : 101.6mm
JIS : 100mm

Power inlet (100 to 240 V AC, 50/60 Hz)


Connector for imput/output terminal module

All Rights Reserved. Copyright © 2002, Yokogawa Electric Corporation GS 11G02N01-01E 4th Edition Jun.20.2006-00
<<Contents>> <<Index>> 13
Input/Output Terminal Module: K9218SC (Accessory)

Unit: mm
Mounting hole
6- 4.5 hole

316

150 150 8
12

5
(140)

164

154

Terminal Terminal Terminal Terminal Terminal


block block block block block
<TN1> <TN2> <TN3> <TN4> <TN5>
Screw terminals M3.5
Connector <CN2>
29
(54.5)

Input/Output terminal (280) 8


module connector <CN1>
Connector <CN3>

Cable for Connecting Input/Output Terminal: K9218SD (Accessory)

100030

Dimensions for Mounting Input/Output Terminal Module

M4 screw
150 150
154

142

302
Cut M4 screw holes at 6 positions.
Drill a rectangular hole of 302  142 mm or more in the center. F05.EPS

All Rights Reserved. Copyright © 2002, Yokogawa Electric Corporation GS 11G02N01-01E 4th Edition Jun.20.2006-00
<<Contents>> <<Index>> 14
EXTERNAL DIMENSIONS OF ACCESSORY SLIDE RAIL Unit: mm

Model: 305A-24/Accuride International Inc.

Closed

Cabinet member 632–3


Closed position flush or under
1.52t
(22.7) (9.5) Open
Intermediate
member
609.6–0.8
Drawer member
577.8–0.3
1.27t 15.9–0.5
565.1–0.3
454–0.3
111.1–0.3
25.4–0.3
12.7–0.3
35.3–0.3

(24.4)

12.7–0.3 4.5  5.3 4.5  5.3


123.8–0.3
288.9–0.3

466.7–0.3
15.9–0.5
F06.EPS
19.1–0.3 606.6–0.8

Reinforcement plate

19-inch rack mounting method:


The instrument weight should be supported at the bottom of the unit (or the side of the unit when mounted with the slide rails).
For easy maintenance, it is recommended to select the method to allow withdrawing along the slide rail.

Slide rail mounted type Guide rail mounted type

Rack size Rack size


101.6

101.6
100

100
or

or

450 or more 450 or more

465 465

Mounting diagram Mounting diagram

Slide rails Guide rails

Guide rails
For the guide rail mounted type, a maintenance
space (200 mm or more) should be provided on top of
the main unit.

All Rights Reserved. Copyright © 2002, Yokogawa Electric Corporation GS 11G02N01-01E 4th Edition Jun.20.2006-00
<<Contents>> <<Index>> 15

EXTERNAL CONNECTION DIAGRAM

Terminal block 1 Terminal block 2


<TN1> <TN2>
Ch5 output  1 11  Ch10 output O2 sensor input*  1 11 Ch5 remote range
switch input
(Ch5_OUT)
2 12

(Ch10_OUT) (O2_IN)
2 12 (R_RNG_Ch5)

Ch4 output  3 13  Ch9 output Ch12 output 3 13 Ch4 remote range


switch input
(Ch4_OUT)
4 14

(Ch9_OUT) (Ch12_OUT) 4 14 (R_RNG_Ch4)

Ch3 output  5 15 5 15 Ch3 remote range


 Ch8 output Ch11 output
switch input
(Ch3_OUT)
6 16

(Ch8_OUT) (Ch11_OUT) 6 16 (R_RNG_Ch3)

Ch2 output  7 17 7 17 Ch2 remote range


 Ch7 output Unassigned switch input
(Ch2_OUT)
8 18

(Ch7_OUT)
8 18 (R_RNG_Ch2)

Ch1 output  9 19 9 19 Ch1 remote range


 Ch6 output Unassigned switch input
(Ch1_OUT)
10 20

(Ch6_OUT)
10 20 (R_RNG_Ch1)
(M3.5 screw) (M3.5 screw)
* For external O2 sensor input.

Terminal block 3 Terminal block 4


<TN3> <TN4>
1 11 Ch5 range identification Peak count alarm 1 11
Unassigned output (PEAK_ALM) Unassigned
signal output (RNG_ID_Ch5)
2 12 2 12
3 13 Ch4 range identification Auto calibration status 3 13
Unassigned Unassigned
4 14 signal output (RNG_ID_Ch4) contact output (ACAL/MNT) 4 14

Remote hold input 5 15 Ch3 range identification Pump ON/OFF 5 15


contact output (PUMP) Unassigned
(R_HOLD) 6 16 signal output (RNG_ID_Ch3) 6 16

Average value reset 7 17 Ch2 range identification Calibration error 7 17


contact output Unassigned
input (RESET) 8 18 signal output (RNG_ID_Ch2) (CAL_ALM) 8 18
Auto calibration 9 19 Ch1 range identification Instrument error 9 19
remote start contact output Unassigned
10 20 signal output (RNG_ID_Ch1) (FAULT) 10 20
input (R_CAL)
(M3.5 screw) (M3.5 screw)

Terminal block 5 Connector


<TN5> <CN3>
Unassigned 1 11 Unassigned Solenoid valve drive signal output for calibration

2 12 Contact output for


Alarm3 output Alarm6 output sample gas selecting 1
3 13 Contact output for Ch3
(ALM3) (ALM6) 6 span calibration
2
4 14 Contact output for Ch4
7 span calibration
Contact output for
5 15 zero calibration 3 Contact output for Ch5
Alarm2 output Alarm5 output 8 span calibration
6 16 Contact output for Ch1 4
(ALM2) (ALM5) span calibration
9 Power for relay drive
7 17 Contact output for Ch2 5 5V DC
span calibration
8 18
Alarm1 output Alarm4 output <D-sub 9P>
(ALM1) 9 19
(ALM4)
10 20
(M3.5 screw)

Note 1: Unassigned terminals are used for internal connection.


So they should not be used as repeating terminals either.
Note 2: The allocation of each channel (Ch1 to Ch12) depends on
measured gas components. Refer to the table on the next page.
F08.EPS

All Rights Reserved. Copyright © 2002, Yokogawa Electric Corporation GS 11G02N01-01E 4th Edition Jun.20.2006-00
<<Contents>> <<Index>> 16

Table 8. Measurable Components and Their Corresponding Channel Numbers


Suffix/Option Code Output and Corresponding Channel
Measurable component O2 analyzer O2 correction CH1 CH2 CH3 CH4 CH5 CH6 CH7 CH8 CH9 CH10 CH11 CH12
-A N Not specified NO
-B N Not specified SO2
-C N Not specified CO2
-D N Not specified CO
-F N Not specified CH4

-G N Not specified NO SO2


-H N Not specified NO CO
-J N Not specified CO2 CO
-K N Not specified NO SO2 CO
-L N Not specified NO SO2 CO2 CO
-A 1, 2, 3 /K NOx O2 Correct NOx Correct NOx av. O2 av.
-B 1, 2, 3 /K SO2 O2 Correct SO2 Correct SO2 av. O2 av.
-D 1, 2, 3 /K CO O2 Correct CO Correct CO av. O2 av.
-F 1, 2, 3 /K CH4 O2 O2 av.
-G 1, 2, 3 /K NOx SO2 O2 Correct NOx av. Correct SO2 Correct NOx av. Correct SO2 av. O2 av.
-H 1, 2, 3 /K NOx CO O2 Correct NOx Correct CO Correct NOx av. Correct CO av. O2 av.
-J 1, 2, 3 /K CO2 CO O2 Correct CO2 av. Correct CO av. O2 av.
-K 1, 2, 3 /K NOx SO2 CO O2 Correct NOx Correct SO2 Correct CO Correct NOx av. Correct SO2 av. Correct CO av. O2 av.
-L 1, 2, 3 /K NOx SO2 CO2 CO O2 Correct NOx Correct SO2 Correct CO Correct NOx av. Correct SO2 av. Correct CO av. O2 av.
-D 1, 2, 3 except /K CO O2
-H 1, 2, 3 except /K NO CO O2
-J 1, 2, 3 except /K CO2 CO O2
-K 1, 2, 3 except /K NO SO2 CO O2
-L 1, 2, 3 except /K NO SO2 CO2 CO O2
T08.EPS
: No measurement in this area is displayed as NOx.
Notes: Peak count alarm is a contact out put.
"Correct XX" means an instantaneous XX value after O2 correction, "Correct XX av." an average XX
value after O2 correction, and O2 av." an average O2 value.

STANDARD ACCESSORIES
Response time: Approx. 20 seconds (for 90% response)
Measured gas flow rate: 0.5 ± 0.25 L / min
Name Part Number Description Qty
Remark: The Zirconia system, due to its principle,
may produce a measuring error due to
Power cable K9218SA standard inlet type (2.5 m ) 1
relative concentration versus the com-
replacement fuse (250 V AC, 2 bustible O2 gas concentration. Also, a
Fuse K9218SB
3.15 A, delay type) 1 corrosive gas (SO2 of 250 ppm or more,
Input/output
K9218SC External terminal module 1 etc.) may affect the life of the sensor.
terminal module Gas inlet/outlet size: Rc1/4
Connection cable between main unit Power supply: 90 to 126 V AC or 200 to 240 V AC,
Cable K9218SD 1
and input/output terminal module (1 m) 50/60 Hz
Slide rail K9218SE Slide rail 2 Enclosure: Steel casing, for indoor application
T09.EPS
Indication: Temperature indication (LED)
Note: Quantity in this is number of accessories supplied as Temperature alarm output:
standard. For instance, two K9218SE parts, i.e., two slide Contact output 1a contact,
rails, are supplied as standard. When ordering separately, Contact capacity 220 V AC, 1 A (resistive
the required number of should be considered. load)
Safety and EMC conforming standards:
Dedicated Zirconia O2 Sensor Safety; EN61010-1
(to be purchased separately) Pollution degree; 2
For O2 correction, the IR400 can accept linearized 0 to 1 V Installation category; II
DC signal coming from an analyzer calibrated to 0 to 25% EMC; EN61326
O2 of full scale. Dedicated zirconia O2 sensor , Model EN61000-3-2
ZX8D, is available from Yokogawa. EN61000-3-3
Dimensions (H  W  D):
Measuring method: Zirconia system
140  170  190 mm
Measurable component and measuring range:
Weight: Approx. 3 kg
Measurable component Minimum range Maximum range
Finish color: Munsell 5Y 7/1

Oxygen (O2) 05 vol% 025 vol% Model Suffix code Option code Description
T11.EPS
ZX8D Dedicated zirconia O2 sensor
Repeatability: Within ± 0.5% of full scale Power 5 90126 V AC, 50/60 Hz
Linearity: Within ± 1% of full scale supply 3 200240 V AC, 50/60 Hz
Zero drift: Within ± 1% of full scale/week Style code C Non-CE conformity
D
Span drift: Within ± 2% of full scale/week CE conformity
T13.EPS

All Rights Reserved. Copyright © 2002, Yokogawa Electric Corporation GS 11G02N01-01E 4th Edition Feb.08.2011-04
<<Contents>> <<Index>> 17
External Dimensions of ZX8D Unit: mm

170

152

OXYGEN ANALYZER

800
TEMP
140

131

110
7-M4

POWER FUSE3A 1 2 3 4 5 6 7

OFF + -
AC E OUT ALM

17 Gas inlet 172.3 Gas outlet


5
F09.EPS

External Connection Diagram

1 2 3 4 5 6 7

+ -

AC power supply E Output Temperature


50/60Hz to IR400 alarm output F09-2.EPS

Dedicated relay board (Option code: /R) Part Numbers


This relay board receives signals from connector CN3 of
the IR400 I/O terminal module and activates the calibration Item Part No. Description Qty
solenoid valve directly.
Relay board K9218SF For external contact point 1
• Relay contact : 1 normally closed contact,
contact capacity; 250 V AC/2 A Cable K9218SG For relay board 1
T10.EPS
(resistive load)

External Dimensions
Unit: mm

15 Max.
Mounting hole ø4.5

CN13
65
75

CN9
CN1 CN2 CN3 CN4 CN5 CN6 CN7 CN8

140
150
F10.EPS

All Rights Reserved. Copyright © 2002, Yokogawa Electric Corporation GS 11G02N01-01E 4th Edition Jan.20.2007-02
<<Contents>> <<Index>> 18
Connections NO2/NO Converter
CN9 CN1 CN2 CN3 CN4 CN5 CN6 CN7 CN8
Part number: K9350LE (Non-CE conformity)
K9350LF (CE conformity)
Reserved
Mounting: Indoor surface mounting
Target Gases: General boiler exhaust gas,
atmosphere
Catalyst: Amount; 2 cm3
Solenoid valve
Solenoid valve
driving power For switching For zero gas CH1 CH2 CH3 CH4 CH5
Replacement cycle; Approx. 8 months
sample
For span gas
(at flow rate of 0.5 L/min with 5% O2, 10
ppm NO2)
Temperature set-point; 220 10 C
CN13
(Sensing tip: K thermocouple)
Wetted materials: Ceramic, Viton, glass filter, SUS316
I/O terminal module
Conversion efficiency: 90% or higher, conforms to JIS
Gas Flow Rate: 0.5 L/min
Relay board CN3
Ambient Temperature: -5 to +45 C
Dedicated cable (D-sub 9-pin straight cable 1.5 m)
F11.EPS
Power Supply: 100 VAC, 50/60 Hz (K9350LE)
100 to 240 VAC, 50/60 Hz (K9350LF)
Contact action Safety conforming standards:
• During measurement: CN1; ON Safety; EN61010-1
Others; OFF Pollution degree; 2
• During calibration: CN1; OFF Installation category; II
Others; Contact corresponding to Weight: Approx. 1.1 kg (K9350LE)
calibration timing is ON Approx. 1.2 kg (K9350LF)
Recommended Connector Sample gas requirements:
• CN1 to CN9: Housing; VHR-2N Dust/drain removed, gas temperature at
(Japan Solderless Terminals (JST)) 150 C or less
Contact; SVH-21T-1.1
One-year-Use Spare Parts
(Japan Solderless Terminals (JST))
Item Part No. Qty
Catalyst for NO2/NO converter K9350LP 2
Glass wool for NO2/NO converter K9350LQ 2
Fitting for NO2/NO converter K9350LV 2
T15.eps

External Dimensions
Unit: mm
(150)
135

Gas Outlet 9.5/ 5.5

Wiring Diagram
(210)
(120)
105

7 8

4- 5

Gas Inlet 9.5/ 5.5

100V AC
Approx.130

(Shared with internal wiring)


110

K9350LE K9350LF

All Rights Reserved. Copyright © 2002, Yokogawa Electric Corporation GS 11G02N01-01E 4th Edition Jan.20.2007-02
<<Contents>> <<Index>> 19
Inquiry Sheet for IR400 NDIR Type Infrared Gas Analyzer

Place a checkmark ✔ in the appropriate box and fill in the specific information in the blanks for your reference.
1. General Information
Company: Delivery destination:
Responsible person: Section: (Phone No. )
Plant name: Measurement location:
Purpose:  Indication reading,  Recording,  Alarm,  Control

2. Requirements
Measurable component:
O2 Analyzer:
 Without O2 analyzer
 External zirconia type sensor (purchase separately:
ZX8D Style C)
 External O2 analyzer
 Built-in paramagnetic type O2 sensor

NO2/NO Converter
 With NO2/NO converter
 Without NO2/NO converter

Range:
1st component, 1st range 1st component, 2nd range 2nd component, 1st range 2nd component, 2nd range
 0 – 20 ppm  0 – 50 ppm  0 – 50 ppm  0 – 100 ppm
 0 – 50 ppm  0 – 100 ppm  0 – 100ppm  0 – 200 ppm
 0 – 100 ppm  0 – 200 ppm  0 – 200 ppm  0 – 250 ppm
 0 – 200 ppm  0 – 250 ppm  0 – 250 ppm  0 – 500 ppm
 0 – 250 ppm  0 – 500 ppm  0 – 500 ppm  0 – 1000 ppm
 0 – 500 ppm  0 – 1000 ppm  0 – 1000 ppm  0 – 2000 ppm
 0 – 1000 ppm  0 – 2000 ppm  0 – 2000 ppm  0 – 5000 ppm
 0 – 2000 ppm  0 – 5000 ppm  0 – 5000 ppm  0 – 1%
 0 – 5000 ppm  0 – 1%  0 – 1%  0 – 2%
 0 – 1%  0 – 2%  0 – 2%  0 – 3%
 0 – 2%  0 – 3%  0 – 3%  0 – 5%
 0 – 3%  0 – 5%  0 – 5%  0 – 10%
 0 – 5%  0 – 10%  0 – 10%  0 – 20%
 0 – 10%  0 – 20%  0 – 20%  0 – 40%
 0 – 20%  0 – 40%  0 – 40%  0 – 50%
 0 – 40%  0 – 50%  0 – 50%  0 – 70%
 0 – 50%  0 – 70%  0 – 70%  0 – 100%
 0 – 70%  0 – 100%  0 – 100%  Not available
 0 – 100%  Not available  Not available

All Rights Reserved. Copyright © 2002, Yokogawa Electric Corporation GS 11G02N01-01E 4th Edition Jun.20.2006-00
<<Contents>> <<Index>> 20
3rd component, 1st range 3rd component, 2nd range 4th component, 1st range 4th component, 2nd range
 0 – 50 ppm  0 – 100 ppm  0 – 50 ppm  0 – 100 ppm
 0 – 100 ppm  0 – 200 ppm  0 – 100 ppm  0 – 200 ppm
 0 – 200 ppm  0 – 250 ppm  0 – 200 ppm  0 – 250 ppm
 0 – 250 ppm  0 – 500 ppm  0 – 250 ppm  0 – 500 ppm
 0 – 500 ppm  0 – 1000 ppm  0 – 500 ppm  0 – 1000 ppm
 0 – 1000 ppm  0 – 2000 ppm  0 – 1000 ppm  0 – 2000 ppm
 0 – 2000 ppm  0 – 5000 ppm  0 – 2000 ppm  0 – 5000 ppm
 0 – 5000 ppm  0 – 1%  0 – 5000 ppm  0 – 1%
 0 – 1%  0 – 2%  0 – 1%  0 – 2%
 0 – 2%  0 – 3%  0 – 2%  0 – 3%
 0 – 3%  0 – 5%  0 – 3%  0 – 5%
 0 – 5%  0 – 10%  0 – 5%  0 – 10%
 0 – 10%  0 – 20%  0 – 10%  0 – 20%
 0 – 20%  0 – 40%  0 – 20%  0 – 40%
 0 – 40%  0 – 50%  0 – 40%  0 – 50%
 0 – 50%  0 – 70%  0 – 50%  0 – 70%
 0 – 70%  0 – 100%  0 – 70%  0 – 100%
 0 – 100%  Not available  0 – 100%  Not available
 Not available  Not available

O2 Analyzer, 1st range O2 Analyzer, 2nd range


 0 – 5%  0 – 10%
 0 – 10%  0 – 25%
 0 – 25%  Not available
 Not available

Output:  4 – 20 mA DC  0 – 1 V DC  RS-232C
O2 correction and O2 average:  Yes  No
Peak count alarm:  Yes  No
Relay board:  Yes  No

3. Sample gas conditions


Fuel:  Gas,  Oil,  Coal,  Refuse,  Other fuel
(1) Temperature: to , Normal temperature [°C]
(2) Pressure: to , Normal pressure [MPa]
(3) Humidity: [vol%]
(4) Dust: [mg/Nm3]
(5) Corrosive gas :  Yes  No

Composition (Detailed composition of sample gas should be provided. This is important for the purpose of
knowing the effect of interference gases)
Contents Concentration range
CO to %  ppm
CO2 to %  ppm
CH4 to %  ppm
H2 to %  ppm
O2 to %  ppm
N2 to %  ppm
SO2 to %  ppm
NOx to %  ppm
H2O to %  ppm
to %  ppm

All Rights Reserved. Copyright © 2002, Yokogawa Electric Corporation GS 11G02N01-01E 4th Edition Jun.20.2006-00
Subject to change without notice.