PORTABLE PRESSURE STEAM STERILIZER YX280
I Applications
‘This product user pressure saturation to sterilize object quickly and effectively. It is applied to
hospital health station, clinics of factories and mines, science and technology research units for
sterilizing medical apparatuses, hat, slass-wares, solution and culture mediums. It ean provide the
highland to use as cooker and in factory and mines for making high quality drinking water.
|. Screen plate
. Indictor light
Structure Figure of Product
Safety valve
Pressure guage
Air escape valve
Cover of container
Bolt
Wing screw
Sealing ring
Container
Sterilizing cylinder Vx
Electro-thermal pipe \ + J
Water relief cock
| Wing-shaped nut spanner
[Structure Characters
‘The container and sterilizing cylinder adopt stainless steel to be pressed forming and can be acid,
alkali and tensile strength proof.
‘The safe valve and air escape valve on the lid of the container can release. The excessive pressure
automatically when ten steam pressure exceeds 0.14 MPa. It has douile safety function. Equipped
with pressure gauge of MPa, convenient for customers to read
The product adopts immersion electrothermal tube to heat. Temperature rises up quickly. It can
save energy about 50% compared with electric furnace.III Technical Parameters
1)Sterilizing cylinder valume: 0.024m?
2)Rated working pressure: 0.14-0.165MPa
3)Maximum safety pressure: 0.165MPa
4)Power pressure: ACIIOV60HZ
5)Power of the elector thermal pipe: 2KW.
Parameter Table
Time for Gauge pressure Saturation steam
Sterilizing objects sterilization relative temp,
Pe Kgvem? Ibiin?
(min) ane aa cc)
Rubber: 1s O11 LL 16 121-126
Hat 30-45 o.1os-o4 | 108-14 121-126
Glassware 15 0.105-0.14 1.05~1.4 121~126
Instruments
& 10 0.105-0.14 | 105-14 121-126
equipment
bottle solutions 20-40 o.10s-0.14 | 105-14 | 15-20 121-126
Guide for use
1) Adding 3.5 litres of water in the container each time to protect the thermal tube from being burnt ut
2) Pile up: Wrap up the object and put tem on the screen plate of sterilizing eylinder, left some space
among parcels so s to let the steam penetrate them and ensure the sterilizing quality
3) Put the sterilizing cylinder into the container, insert the flexible pipe which is attached to the lid of
the container imo the semicircle slot of the sterilizing cylinder, tighten up the screws with wing
nutspanner to ensure that the lid and the container are closely sealed.
4) Heating: Connect sterilizer to power AC HOV. After put on the switeh, the indicator light starts to
show the beginning of heating, At this time, open the air escape valve to let cold air escape. When the
rapid steam jets out, close it. During the heating, the pressure gauge will indicate the steam pressure
in sterilizer.
5) Steril
ing: Start to count the time for sterilization when the steam pressure reaches to pressure
constant, IF tis used for sterilization with a pressure below 0.14 MPa, when the pressure reaches the
requested range, control the power switch by hand to keep the pressure constant,6) Cooling: When the sterilization is complete, turns off the power switch.
Do not open the air escape valve on the lid of the container until the sterilizer cools down and the
reading of pressure gauge returns fo Zero so as {0 avoid the boiling liquid to flow out the container or
injure any person,
VSafety and Maintenace
1) Don’t bump the sterilizer when it is in use and forbid to use over pressure. When the pointer of
pressure gauge reaches 0.165 MPa and the safe valve does not release the pressure, switch it off at
once. And allow itto cool and check the safety valve.
2) When not in operation, the pointer of the pressure gauge should be at zero position, otherwise the
pressure gauge is faulty and should be repaired or replaced or replaced. To ensure the sen:
and accuracy. the pressure gauge should be regulated 1-2 times per year.
3) After use, drain the water ftom the container and the condensing water below the screen plate. Wipe
up the descale in the container and electrothermal pipe. In order to improve the sterilizing quality
and prolong the life ofthe sterilizer.
4) Check the seal ring often and when the rubber becomes ageing and worn out, replace it in time.